5. Operating elements and connections of the scanner 8
5.1 Description of operating elements 8
6. Setup and installation 10
6.1 Unpacking of the scanner 10
6.2 Setup of the computer and monitor 10
6.3 Setup of the scanner 11
7. Getting started 12
7.1 my.Straumann
®
CARES® De sk to p 12
8. Setup of the internet connection 13
8.1 Configuration of the internet connection 13
8.2 Configuration of an internet connection with a dynamic
IP address (DHCP) 13
8.3 Configuration of an internet connection with a static IP address 15
9. Troubleshooting 17
10. Changing fine fuses 17
11. Cleaning notes 18
12. Preparation for transportation (packing of the scanner) 20
13. Technical data 25
14. Explanation of symbols 26
1. FOREWORD
Dear User,
Thank you for your interest and welcome to Straumann!
These installation instructions show you how to install the Straumann
®
CARES®
Scan CS2 scanner.
Please read these installation instructions and the provided Software User
Manual carefully to become familiar with the instructions prior to use. Please
keep these installation instructions safe in case you require them again in the
future.
This device was manufactured and packed with the greatest care; however, should you experience problems with it or if a part listed in chapter 3
is missing, please contact your local Straumann representative immediately.
The information in this manual corresponds to the latest state of the art; nevertheless, it must be noted that the specifications of the scanner may be changed
without previous announcement. The relationship between Straumann and the
user is exclusively governed by the General Terms and Conditions of Straumann.
Straumann products must be used in accordance with their accompanying
instructions for use.
If you would like additional information on the proper use of Straumann products,
please contact your local Straumann distributor.
2
2. INTENDED USE
The Straumann® CARES® Scan CS2 is a 3D scanner of high precision for objects
of the dental area. It is used for 3D measurements of single preparations, abutments, models of the jaw with up to 14 elements, and different impressions.
Together with the corresponding software, the scanner gives dental technicians
the ability to scan models, design a restoration (crown, bridge, abutment, etc.
in a wide range of materials) and send the resulting data file directly to the
manufacturing facility.
Notes
p The surface of the materials must be scannable and therefore it must be
matte (not shiny/reflecting).
p In addition, the surfaces should have a homogeneous color gradient.
Please note
Practitioners must have appropriate knowledge and instruction in the handling
of the Straumann product described herein (“Straumann Product”) for using the
Straumann Product safely and properly in accordance with these instructions.
The Straumann Product must be used in accordance with the instructions for
use provided by the manufacturer. It is the practitioner’s responsibility to use
the device in accordance with these instructions for use and to determine if the
device fits to the individual patient situation.
The Straumann Product is part of an overall concept and must be used only
in conjunction with the corresponding original components and instruments
distributed by Straumann USA, LLC, its ultimate parent company and all affiliates
or subsidiaries of such parent company (“Straumann”), except if stated otherwise in these instructions. If use of products made by third parties is not recommended by Straumann in these instructions, any such use will void any warranty
or other obligation, express or implied, of Straumann.
Validity
Upon publication of these instructions, all previous versions are superseded.
3
3. SCANNER, COMPUTER, MONITOR AND ACCESSORY KIT
The Straumann® CARES® Scan CS2 scanner is delivered
in transport packaging, which is designed to prevent any
damage of the scanner during transport. Please check the
packaging for external damage. Should you discover such
damage, please inform your local Straumann representative
immediately, as potential complaints will otherwise expire.
Please store the transport packaging in case you need to
safely transport your scanner.
Contents of the delivery include:
1. Straumann
2. Computer box
3. Monitor box
4. Accessory kit
®
CARES® Scan CS2 scanner
3.1 Straumann
The illustration may differ from the current
configuration
®
CARES® Scan CS2 scanner
4
3.2 Computer box
The contents of the computer box may vary in different
countries. For further information, please contact your local
Straumann representative.
Content of the computer box:
p Computer
3.3 Monitor box
The illustration may differ from the current
configuration
The contents of the monitor box may vary in different countries. For further information, please contact your local
Straumann representative.
Content of the monitor box:
p Monitorp Monitor signal cable(s) (DVI/VGA)p Monitor manual
The illustration may differ from the current
configuration
5
3.4 Accessory kit
The contents of the accessory kit may vary in different
countries. For further information, please contact your local
Straumann representative.
Content of the accessory kit:
p Power cable for scannerp Power cable for computerp Power cable for monitorp USB cable for scanner – computer p Complete set-up aid kit
(set-up aid kit, putty holder and putty)
p One copy of Software User Manualp Installation manual for Straumann
®
CARES® Scan
CS2 scanner
p Keyboardp Mousep Network cable (blue), length 10 m (optional) p Socket strip (optional)p One small can of CopyCAD wax (optional)
6
4. WARNINGS: PRECAUTIONARY MEASURES
If the unit develops smoke, smells burnt or produces unusual noises, immediately pull out the main power cable and ask your local Straumann
representative or technical support for advice.
Attempting to work with a faulty device can be dangerous.
Keep small objects and liquids away from the unit.
Small objects may inadvertently drop through the ventilation slots of the
unit`s housing and thereby cause fire, electric shock or damage of the unit.
Should an object or liquid get into the housing, immediately pull out the
main power cable of the unit. Have the unit checked by a qualified Straumann
technician before you work with it again.
Place the unit on a suitable and stable surface.
Otherwise, the unit could be damaged and there is a risk of fire and electric
shock.
p Do not use the unit outdoors. p Do not use the unit in a moving vehicle (ship, aircraft, train, car, etc.).p Do not install it in dusty or damp environment.p Do not place it in the vicinity of a heat-generating device or an air humidifier
p After unpacking and before making any electrical connections, allow the
unit to acclimate to ambient temperature and humidity for 2 hours.
Use the enclosed main power cable for connection to the national standard
line voltage. Make sure that the nominal voltage of the main power cable
is not exceeded.
Otherwise, there is a risk of fire and electric shock.
The unit must be connected to a correctly grounded wall outlet.
Otherwise, there is a risk of fire and electric shock.
Connect the main power cable to an easily accessible wall outlet.
Make sure that you can quickly pull out the main power cable in case of an
emergency.
Caution
p Do not block the ventilation slots of the housing (on the back of the scanner).p Do not place any books, papers or other objects on any ventilation slots.p Do not install the unit in a constricted space without ventilation. Blocking
the ventilation slots so that air can no longer circulate may cause fire or
other damage.
7
5. OPERATING ELEMENTS AND CONNECTIONS OF THE SCANNER
5.1 Description of operating elements
Position of indication lights
Scanner status Laser status
Scanner as illustrated from the front
Status of scanner and laser based on indication lights
Indication lights and meanings:
StatusScanner status LED
(left)
Scanner offOffOff
Scanner and PC turned on, setting up the connectionYellow, flashing slowlyOff
Connection established and HighSpeed
USB active
Scanner in standbyYellowGreen
Ready to load/unload scanner
(loading position, unloading position)
ScanningGreenYellow, flashing
ReferencingGreenYellow, flashing
PositioningGreenYellow, flashing
StopGreenOff
Problem, please contact service supportRed, flashingRed, flashing
GreenOff
GreenGreen
Laser status LED
(right)
8
Description of ports
Ports on the back of the scanner
(1) USB connection to computer
USB connection for scanner control which must be
connected to the computer with the supplied serial
cable.
1
2
3
4
(2) Port of power supply
This port of the scanner must be connected to the power
supply with the supplied main power cable.
(3) Fine fuse holder
For changing the fine fuses (insert), please refer to
chapter 10.
(4) On/off switch for scanner power supply
Rocker switch for switching the scanner on and off
Warning
p The fine fuse insert may be opened only with the main
power cable pulled out.
p If the line voltage is selected incorrectly, the unit may
be destroyed and there is a risk of fire.
9
6. SETUP AND INSTALLATION
The scanner works within a voltage range of 100–240 V AC.
Arrangement/setup of the system
p All plug connections should be easily accessible.
6.1 Unpacking of the scanner
Please remove the individual components (foam parts) while
unpacking the scanner. Store the foam components in a safe
place in case you need to ship the scanner in the future.
Please refer to chapter 12 to see the individual packaging
components.
6.2 Setup of the computer and monitor
1. Plug the keyboard (preferably via the PS/2 connector)
into the designated port on the back of the computer.
2. Plug the mouse (preferably via the PS/2 connector) into
the designated port on the back of the
computer.
3. Plug the scanner into the designated port on the back of
the computer (USB).
4. Plug the monitor signal cable (preferably the
supplied DVI cable) into the designated port on the back
of the computer.
5. Connect the computer to a power outlet with the
enclosed main power cable.
5
12
3
4
10
6.3 Setup of the scanner
Follow these steps to set up the scanner (refer to the
illustration on the right):
1. Release the transportation lock on the right side
of the scanner and the two transportation locks on
the back before the scanner is turned on.
Important: All transportation locks have to be
released before operation.
This symbol indicates that the scanner is in a locked position.
Transportation locks:
One lock on the side of the scanner
2. Use the supplied USB cable to connect the scanner
to the computer (port 3 on the back of the computer).
3. Connect the scanner to the power outlet with the
enclosed
main power cable.
For more information regarding the connections and operating
elements of the scanner, please refer to chapter 5.
Caution
Place the scanner on a stable and flat surface. Ensure
that no external vibration is influencing the scanner
while in operation. Do not continue operation in the
event of scan objects or any other objects being lost inside
the scanner.
Transportation locks:
Two locks on the back of the scanner
11
7. GETTING STARTED
After you have carefully connected all components according to the instructions, the Straumann
monitor can be switched on. Directly after switching on the system, the left
(green) LED on the scanner will light up. The right LED on the scanner will not
light up since the scanner is not in operation.
After registration/login on the start screen, the scanner is ready to use. Please
refer to the Software User Manual for more information.
®
CARES® Scan CS2 scanner as well as the computer and
7.1 my.Straumann
®
CARES® Desktop
As a Straumann customer, the default option to set up the my.Straumann
CARES
tomatically started and guides you immediately to the my.Straumann
Desktop which contains individual tiles with Straumann
®
Desktop is activated. With the pre-setting, the server and client are au-
®
CARES®
®
applications.
®
12
8. SETUP OF THE INTERNET CONNECTION
Provided that you did not order a special system configuration before delivery, the default setting is routing via DHCP.
8.1 Configuration of the internet connection
To establish an internet connection, it is necessary to
connect the network card on the back of the computer
with a router via the network cable. For information on configuring and setting up further connections for your router,
please read the relevant operating instruc tions or contact
your service technician.
8.2 Configuration of an internet connection with
a dynamic IP address (DHCP)
1. Log into the Windows operating system.
2. Click on the Start button.
3. Enter Network connections in the search field and
then click on Show network connections.
4. Right-click on Local Area Connection and then
select Properties.
13
5. Select Internet Protocol (TCP/IPv4) using the left
mouse key and then click on Properties.
6. Select both Obtain an IP address automatically
and Obtain DNS server address automatically.
7. Then click OK.
8. Start Internet Explorer to test the connection; if the
home page is displayed, your computer is connected
to the internet.
14
8.3 Configuration of an internet connection with a
static IP address
1. Log into the Windows operating system.
2. Click on the Start button.
3. Enter Network connections in the search field and
then click on Show network connections.
4. Right-click on Local Area Connection and then select
Properties.
5. Select Internet Protocol (TCP/IPv4) using the left
mouse key and then click on Properties.
15
6. Select both Use the following IP address and
Use the following DNS server addresses.
7. Enter valid IP addresses in both fields.
8. Then click OK.
9. Start Internet Explorer to test the connection; if the
home page is displayed, your computer is connected
to the internet.
16
9. TROUBLESHOOTING
Please refer to the Software User Manual to find potential
causes for errors and their solutions.
10. CHANGING FINE FUSES
If necessary, remove the insert with an appropriate screwdriver and change the corresponding fine fuses (5x 20 mm,
2.0 AT 250 V AC compliant with IEC 60127-2 SS 3).
Note
Be sure to unplug the power before changing fuses.
17
11. CLEANING NOTES
The following recommendations refer to the scanner, the standard PC housing,
the plastic housing, and the LCD monitor with surface made of coated and/or
painted plastic.
Cleaning of scanner
p Use a damp towel to clean the outside of the scanner. Please make sure
that the scanner is unplugged while doing so.
p Use a dry towel to clean the inside of the scanner. Do NOT use any
wet or damp towels to clean the inside. Do NOT use compressed air to
clean the inside.
Damp cleaning of monitor
p If you damp clean an electrical device, always switch the device off first
and disconnect it from the power supply. After cleaning let the device dry.
p Never use aggressive cleaning agents such as thinners, scouring liquid,
wax, petroleum, spirit, spray cleaners, acids or alkalis. Do not use any hard
or rough cleaning media such as scouring sponges. These can destroy
the coating of the screens.
p A damp cloth is sufficient for cleaning (microfiber cloth recommended).
Be sure to wring out the cleaning cloths well so that no excess water can
get to sensitive electronic components. The screen surface should be
cleaned with as little pressure as possible to avoid abrasion of the coating.
p If you use a special screen cleaner, never spray it directly on the device,
but spray on the cloth instead. Fine spray mist could possibly reach sensitive components. Under no circumstances should you wipe off the screen
surface with paper towels since these are too hard and scratch
the surface.
18
Keyboard
p Keyboards should be cleaned with the computer switched off or
disconnected from the computer.
p Dust and other dirt particles between the key gaps can be removed
with a soft and dry long-fiber brush. The use of compressed air is preferred.
A hair dryer which can blow cold air may also be used.
p The keys should then be freed from dirt and grease traces with a slightly
moistened cloth. No moisture should penetrate the keyboard, as this could
damage the built-in electronics or lead to errors in data transmission to
the PC at a later stage. In case of doubt let the keyboard dry or dry it with
warm (not hot) air.
Dust in the interior of the computer
If you suspect dust in the scanner, PC housing, plastic housing, or the LCD monitor
do not open. Contact Straumann technical support. Opening the housing of
any hardware will void the warranty.
Caution
Do not open the housing while the computer is connected to the power supply.
Otherwise, there is a risk of electric shock.
19
12. PREPARATION FOR TRANSPORTATION (PACKING OF THE SCANNER)
Important: The scanner needs to be in parking position and locked to ensure safe transportation (please
note that the actual transportation packaging may vary
from the following illustrations).
1. In order to park the scanner, please open the software
and go to Scanner in the task line. Click on
Scanner in parking position. This puts the scanner
in the parking position for transportation.
2. Lock the scanner with the one transportation lock on
the right side and the two locks on the back of the
scanner (see description in chapter 6.3, including the
symbol indicating the correct locking position).
Foam parts for transportation
It is necessary to use the original packaging components
for the safe transportation of the scanner.
Straumann
®
CARES® Scan CS2 – Packaging overview
Transportation box for the Straumann
scanner
®
CARES® Scan CS2
20
Foam components number 1 to 3 (2 for each side; left and
right) and component number 4 (one piece for the top).
Please note that the actual pieces are not numbered and
that this is an illustration only.
Foam components for the bottom of the box (numbers
5–7). Please note that the actual pieces are not numbered
and that this is an illustration only.
Step-by-step
Please follow the steps below to pack the scanner.
1. Insert components 2 and 3 into 1 as shown in the
picture on the right.
21
2. Insert component 6 into component 5.
3. Place component 7 on the bottom of the box as
illustrated.
4. Insert components 5 – 6 (as illustrated in step 2) as
protection on the back of the box.
22
5. Carefully place the scanner into the box.
6. Cover the scanner with a protective plastic wrap.
7. Position components 1–3 (prefabricated as described
in step 1).
23
8. Place component number 4 across as shown in
the illustration on the right.
9. Close the lid of the box and tape it shut.
24
13. TECHNICAL DATA
Technical data
Model designation: CARES
Connections: USB 2.0 High-speed/power supply
Power: 100–240 V AC 50–60 Hz
Power Input max.: 900 mA
Laser protection class: Class 1, 1M
IP protection: IP20
– 10 moving axes for scanning dental preparations by
means of laser light sectional method
Scan time
Single die: 25– 45 sec. (Smartscan)
Jaw model: 20– 90 sec. (Smartscan)
Scan method: Automated scanning of dies and
jaw models by multiple
scanning angles. 35°– 75°
®
Scan CS2
Tested according to IEC 60825-1 (TÜV), Standard: 2007
Additional Information
We ig ht: 15.2 kg
Degree of pollution: Degree of pollution 2:
Laboratory environment
Altitude: Recommended altitude below
2’000 m, 6’60 0 ft
Environment: Recommended temperature
15°–35 °C, 59– 95 ° F,
humidity (5°– 60 °)
Theo. Measuring field Width: 16 mm
Height: 42 mm
Max. die height 22.3 mm
Max. jaw height 30 mm (preparation share)
Laser classification 1M
Measuring tolerance ± 10 µm (pot object)
Structure of the serial number
SN= ABBBBCCCC:
A = Scanner generation
B = Year of manufacturing
C = Number
The serial number 220101001 means: 2
nd
generation
scanner, manufactured in 2010, as scanner number 1001
25
14. EXPLANATION OF SYMBOLS
Protective Earth
Laser radiation
LASER RADIATION
DO NOT VIEW DIRECTLY WITH
OPTICAL INSTRUMENTS
CLASS 1M LASER PRODUCT
Laser warning
Caution – Risk of electrical shock
SN
Serial number
Order code
Fuse
Consult instructions for use
Do not dispose of with household waste
Caution: Federal law restricts this device to sale
by or on the order of a dental professional.