Stephanix BRS Preinstallation

Technical Publication
PI-1005R3
Pre-Installation
HF Series Generators
REVISION HISTORY
0 JUL 15, 2002 First edition
1 SEP 20, 2003 kW corrections for SHF-5xx model
2 FEB 17, 2004 New equipments
HF Series Generators
Pre-Installation
3
This Document is the English original version, edited and supplied by the manufacturer.
The Revision state of this Document is indicated in the code number shown at the bottom of this page.
SEP 15, 2005 Revision of environmental and electrical requirements
ADVISORY SYMBOLS
The following advisory symbols will be used throughout this manual. Their application and meaning are described below.
DANGERS ADVISE OF CONDITIONS OR SITUATIONS THAT IF NOT HEEDED OR AVOIDED WILL CAUSE SERIOUS PERSONAL INJURY OR DEATH.
ADVISE OF CONDITIONS OR SITUATIONS THAT IF NOT HEEDED OR AVOIDED COULD CAUSE SERIOUS PERSONAL INJURY, OR CATASTROPHIC DAMAGE OF EQUIPMENT OR DATA.
PI-1005R3
Advise of conditions or situations that if not heeded or avoided could cause personal injury or d amage to equipment or data.
Note . Alert readers to pertinent facts and conditions. Notes represent
information that is important to know but which do not necessarily relate to possible injury or damage to equipment.
HF Series Generators
Pre-Installation
TABLE OF CONTENTS
Section Page
1 INTRODUCTION 1.........................................................
1.1 Responsibility of Purchaser 1..........................................
2 PRE-INSTALLATION DATA 3...............................................
3 ROOM REQUIREMENTS 11.................................................
3.1 Environmental Requirements 11.........................................
3.2 Electrical Requirements 11.............................................
3.3 Line Powered Generators -- Power Line Requirements 12...................
3.4 Line Powered Generators -- Recommended Wire Size 15...................
3.5 Capacitor Assisted Generator s -- Power Line Requirements 17..............
3.6 Capacitor Assisted Generators -- Recommended Wire Size 17...............
3.7 Battery Powered generators -- Power Line Requirements 18.................
3.8 Battery Powered Ggenerators -- Recommended Wire Size 18...............
3.9 Interconnection and Grounding Requirements 19..........................
3.10 Safety Devices 22.....................................................
i
PI-1005R3
HF Series Generators
Pre-Installation
PI-1005R3
ii
SECTION 1 INTRODUCTION
This Pre-Installation document provides the information and data needed to plan and qualify the customer site prior to equipment delivery and installation.
This document considers only the Generator and its associated components. Product information, environmental and electrical requirements are specified.
For system-related requirements, such as room layout, and system interconnections, refer to documentation provided with other subsystems.
1.1 RESPONSIBILITY OF PURCHASER
HF Series Generators
Pre-Installation
Note .
Site planning and preparation are the responsibilities of the purchaser. The following points should be considered fundamental to the customers Pre-Installation activities; addition work may be needed depending on specific site circumstances:
Install required material prior the delivery of the system components.
Complete room floor, ceiling and wall finish.
Install c onduit, duct, and raceway.
Install proper size junction boxes with covers at locations specified in the
installation plan.
Install mains power of proper voltage output and adequate kVA rating.
Install all safety devices according to this document and Local Codes.
Provide current room dimensions, including hall way and entry door
sizes.
Complete and proper Pre-Installation will avoid delays and confusion.
PI-1005R3
1
HF Series Generators
Pre-Installation
This page intentionally left blank.
2
PI-1005R3
SECTION 2 PRE-INSTALLATION DATA
This section provides product information and illustrations showing physical dimensions, weight, mounting holes and normal access areas for cabling and service. (Refer to Illustration 2-1.)
PHYSICAL CHARACTERISTICS
HF Series Generators
Pre-Installation
COMPONENT
LINE POWERED GENERATORS
Vertical Generator Cabinet
with LF-RAC (LSS) or DRAC (HSS)
Compact Generator Cabinet
(for only 1 Tube (LSS))
Compact Generator Cabinet
(for 1 or 2 Tubes (LSS or HSS))
CAPACITOR ASSISTED GENERATORS
Compact Generator Cabinet
with Capacitors Module
DIMENSIONS
WEIGHT
Length Width Height
506 mm 468 mm 1101 mm 148 kg
445 mm 360 mm 568 mm 72 kg
592 mm 360 mm 690 mm 95 kg
500 mm 360 mm 790 mm 108 kg
BATTERY POWERED GENERATORS
Compact--ESM Generator Cabinet
with Batteries Module
Compact--ESM Generator Cabinet
with Batteries Module (17 A/h batteries)
PI-1005R3
813 mm 436 mm 948 mm 235 kg
813 mm 436 mm 1223 mm 372 kg
3
HF Series Generators
TouchScreenConsol
e
Pre-Installation
PHYSICAL CHARACTERISTICS
COMPONENT
STANDARD CONTROL CONSOLES
RAD Control Consoles
R&F Control Consoles
RAD Console Graphic Display
Touch Screen Console
(TPC 10” or 12”)
Optional Pedestal for
RAD Console Graphic Display or Touch Screen Console
Note.-- Dimensions for no-standard Consoles are not indicated in this document.
w/o Pedestal 433 mm 298 mm 123 mm 8kg
with Pedestal 433 mm 298 mm 1023 mm 22 kg
w/o Pedestal 554 mm 360 mm 124 mm 12 kg
with Pedestal 554 mm 360 mm 1010 mm 35 kg
with Handswitch support 545 mm 290 mm 50 mm 6kg
w/o Handswitch support 430 mm 290 mm 50 mm 6kg
with Handswitch support 468 mm 290 mm 11 4 m m 8kg
w/o Handswitch support 360 mm 290 mm 11 4 m m 8kg
DIMENSIONS
WEIGHT
Length Width Height
298 mm 236 mm 930 mm 10 kg
TOUCH SCREEN CONSOLE AND PC UNIT
Touch Screen Console 400 mm 200 mm 400 mm 5kg
PC Unit 480 mm 200 mm 400 mm 15 kg
PC Interface Box 130 mm 140 mm 46 mm 0.6 kg
Note.-- Specifications of Touch Screen Consoles and PC Units subject to c hange without notice.
4
PI-1005R3
METHOD OF MOUNTING
COMPONENT NORMAL METHOD OF MOUNTING
HF Series Generators
Pre-Installation
Generator Cabinet
Control Consoles Desk freestanding, wall mounted or anchor to an optional pedestal.
Note: Anchoring hardware should be field supplied. For seismic areas all components must be anchored,
Local Standards should be applied.
Floor freestanding, wall mounted or anchor to floor with four M10 (3/8”) bolts.
MINIMUM RECOMMENDED FREE AREA FOR SERVICE ACCESS
COMPONENT
Generator Cabinet
Control Consoles
Note: Ventilation conditions requires to keep a minimum free distance of 15 cm (6”) from both lateral sides of the Generator Cabinet and also the same distance from the rear side when the Generator is provided with High Speed Starter (fans for the starter module).
Left Side Right Side Front Rear Top Bottom
50 cm
(20”)
10 cm
(4”)
50 cm
(20”)
10 cm
(4”)
SURFACE
100 cm
(40”)
Completely
free
--
(see note)
10 cm
(4”)
Completely
free
Completely
free
--
--
PI-1005R3
5