Stephanix BRS Service manual

Technical Publication SM-0521R3
Service Manual
BRS
Basic Radiographic System
0459
rue Jean Moulin
Zone Industrielle du Bayon
42150 La Ricamarie
FRANCE
Tel. : 00 33 4 77 47 81 60 ; Fax : 00 33 4 77 37 55 19
Le marquage CE qui se trouve sur ce produit indique que celui-ci est en conformité avec la directive européenne des dispositifs médicaux 93/42/CEE
This product bears a CE marking in accordance with the provisions of the 93/42/EEC MDD dated June 14, 1993.
Basic Radiographic System
REVISION HISTORY
REVISION DATE REASON FOR CHANGE
0 April 22, 2003 First edition
1 30 March, 2004 Installation and dimensions modified.
2 07 May, 2005 Installation improvements.
Service Manual
3
This Document is the English original version, edited and supplied by the manufacturer.
The Revision state of this Document is indicated in the code number shown at the bottom of this page.
12 Jul, 2005 New schematics and updated photos.
ADVISORY SYMBOLS
The following advisory symbols will be used throughout this manual. Their application and meaning are described below.
DANGERS ADVISE OF CONDITIONS OR SITUATIONS THAT IF NOT HEEDED OR AVOIDED WILL CAUSE SERIOUS PERSONAL INJURY OR DEATH.
ADVISE OF CONDITIONS OR SITUATIONS THAT IF NOT HEEDED OR AVOIDED COULD CAUSE SERIOUS PERSONAL INJURY, OR CATASTROPHIC DAMAGE OF EQUIPMENT OR DATA.
SM-0521R3
Advise of conditions or situations that if not heeded or avoided could causepersonal injury or damageto equipment or data.
Note . Alert readers to pertinent facts and conditions. Notes represent
information that is important to know but which do not necessarily relate to possible injury or damage to equipment.
Basic Radiographic System
Service Manual
SAFETY SYMBOLS
The following safety symbols will be used in the equipment. Their meaning are described below.
Attention, consult accompanying documents.
Ionizing radiation.
Type B equipment.
Dangerous voltage.
Ground.
This symbol indicates that the waste of electrical and electronic equipment must not be disposed as unsorted municipal waste and must be collected separately. Please contact an authorized representative of the manufacturer or an authorized waste management company for information concerning the decommissioning of your equipment.
SM-0521R3
Basic Radiographic System
Service Manual
TABLE OF CONTENTS
Section Page
1 INTRODUCTION 1.........................................................
1.1 Tools 1..............................................................
1.2 Pre-Installation Checks 1..............................................
2 UNPACKING 3............................................................
3 INSTALLATION 5..........................................................
1.2 Positioner Installation overview 5.......................................
4 ADJUSTMENTS 15........................................................
4.1 Adjustment tools 15....................................................
4.2 Alignment of X-ray Beam 18............................................
4.2.1 Alignment of Light Field with X-ray Field 22.........................
4.2.2 Perpendicularity Adjustment of X-ray Beam with Image Receptor 25...
4.2.3 Centering of X-ray Field and Image Receptor Bucky Assembly 27.....
4.3 Field Size Indicator Test 28.............................................
4.4 X-ray Tube or Collimator Replacement 30.................................
4.5 Adjustment of rotation Plate 30..........................................
5 RENEWAL P ARTS 31......................................................
6 MANTEINANCE 33.........................................................
7 INTERCONNECTION MAPS 35..............................................
SM-0521R3 i
Basic Radiographic System
Service Manual
ii
SM-0521R3
SECTION 1 INTRODUCTION
1.1 TOOLS
The following hand tools are required for the Installation:
Standard service engineers tool kit.
Electric drill motor and assorted bits.
1.2 PRE-INSTALLATION CHECKS
Basic Radiographic System
Service Manual
Prior to beginning installation it is recommended to inspect the site and verify that the X-ray room complies with requirements such as:
The maximum dimensions and system movements to plan the minimum space required in the room:
Maximum Height 2500 mm
Maximum Width 1550 mm
Maximum Length 1340 mm
Conduits and walls are ready to install the System.
Electricity installation:
Power Line (for Collimator Lamp): 24 VAC, 50/60 Hz, 6.5 A
Power Line (for Movement Locks): 24 VDC, 0.2 kVA
ACCORDING TO MDD93/42/CEE, THIS UNIT IS EQUIPPED WITH EMC FILTERS. THE LACK OF THE PROPER GROUNDING MAY PRODUCE ELECTRICAL SHOCK TO THE USER.
SM-0521R3
1
Basic Radiographic System
Service Manual
Illustration 1-1 Dimensions
566
2148
1795
707 mm.
SID = 1400 mm.
Max. Width = 1595 mm. Anchura Max. = 1595 mm.
885
1400
Max. Travel = 1400 mm.
440
Recorrido Max. = 1400 mm.
Column Max. height = 2240 mm.
Altura Max. Columna = 2240 mm.
497
Max. Height = 2500 mm.
610
Altura Max. = 2500 mm.
700
(*) Dimensions Subject to change depending on the type of tube
(*) Dimensiones sujetas a cambi o dependiendo del tipo de tubo
2
SM-0521R3
SECTION 2 UNPACKING
The Unit is shipped in one box to facilitate transport and installation. Upon receipt of the X-ray unit and associated equipment, inspect all shipping containers for signs of damage. If damage is found, notify the carrier or his agent immediately.
1. Place the shipping pallet near its final site in the room and remove all its laterals. Do not discard any packing material such as envelopes, boxes, bags until all parts are accounted for as listed on the packing list.
A T LEAST THREE PEOPLE ARE REQUIRED TO REMOVE ALL HEAVY COMPONENTS FROM THE SHIPPING PALLET.
Basic Radiographic System
Service Manual
2. When the equipment is unpacked, check part numbers and serial numbers of each component with its identification labels. Inspect all pieces for visible damages. If any damaged part is found, repair it or order its replacement to prevent unnecessary delay in installation.
3. Verify that all items on the customer order are present.
4. Leave a working area around equipment until its final installation is complete.
SM-0521R3
3
Basic Radiographic System
Service Manual
This page intentionally left blank.
4
SM-0521R3
SECTION 3 INSTALLATION
3.1 POSITIONER INSTALLATION OVERVIEW
TICK BOXES WHEN DONE
Basic Radiographic System
Service Manual
1.-- Install the Support of the Tube--Collimator Assembly
and the Control panel (in some cases it is factory installed)
2.-- Install the Wall Support at the back of the Column
(the Column laid on the pallet)
3.-- Drill WALL HOLES and anchor the Wall Support .
4.-- Install the Collimator and the X-ray tube.
5.-- Remove the Safety Locking Rod/Central carriage Screw.
6
6.-- Install all the counterweight plates supplied.
1
2
4
5
7.-- Install the Oscillating Grid Assembly or Fixed Grid
8.-- Drill FLOOR HOLES and anchor the unit to the floor.
9.-- Connect cables (HV, Tube, Stator, Bucky or Grid)
10.-- Check levels on Column, Tube-Collimator Assembly
SM-0521R3
Assembly. (level it)
and guide them .
and Receptor Assembly.
9
10
7
5
Basic Radiographic System
Service Manual
Note .
Note .
Perform Installation following the order described below.
DO NOT REMOVE SAFETY LOCKING ROD FROM THE COUNTERWEIGHT CARRIAGE AND/OR THE CENTRAL CARRIAGE SCREWS (2) BEFORE SPECIFICALLY INSTRUCTED IN THIS DOCUMENT.
Due to packaging requirements, some Units may not include the Tube Collimator Assembly and the Control panel factory pre-installed. If that is the case, follow steps 1 and 2, if not, go to step 3.
1. Assemble the Support of the Tube-Collimator Assembly to the Upper area of the Swivel Arm using the four Safety Screws installed in the Arm.
SAFETY SCREWS
SUPPORT OF TUBE-COLLIMATOR ASSEMBLY
2. Assemble the Control Panel in the Support with the two screws, washers and nuts attached to the Control Panel.
CONTROL PANEL
SCREWS, WASHERS AND NUTS
6
SUPPORT OF TUBE-COLLIMATOR ASSEMBLY
SM-0521R3
Basic Radiographic System
Service Manual
3. Assemble the Upper Wall Support to the Column. Lift the Column as much as necessary as to mount the Upper Wall Support in its upper part.
SAFETY SCREWS
Note .
At least two people are required to perform the next operation.
4. Position the column against the wall on its final site in the room. While one person is holding up the Column, the other one should level it vertically on both lateral sides and on the front. Also use the leveling screws at the base of the column.
5. Mark its anchoring positions on the wall (not to the floor yet).
SM-0521R3
Note .
Distances shown on Illustration 3-1 may also be taken as reference to mark the anchoring holes position.
7
Basic Radiographic System
Service Manual
Illustration 3-1 Drill Template
FOUR WALL MOUNTING HOLES
FOR SCREWS 8 mm.
160
1839
FOUR FLOOR MOUNTING HOLES FOR SCREWS10 mm.
8
SM-0521R3
Basic Radiographic System
Service Manual
6. Move the Column and prepare the anchorages.
BEFORE ANCHORING THE UNIT, CHECK THAT THE WALL AND THE ANCHORING SYSTEM ARE STRONG ENOUGH (2000 NEWTON TRACTION FORCE) TO ENSURE A SAFE INSTALLA TION. SOME NON BRICK WALLS MA Y REQUIRE ADDITIONAL ANCHORAGE INSTALLATION.
7. Drill holes and anchor the Wall Support. Make sure the Column is firmly anchored and leveled at front, sides and base.
8. Install the X-ray Tube in the Upper Support of the Column using the Collimator Adaptation Ring and its four Safety Screws (Allen).
9. Before installing the Collimator, unscrew the same number of turns the four Centering Adjustment and Safety Screws (Allen) to allow the Collimator installation in the Collimator Adaptation Ring. Adjust the Collimator Blades to their widest setting and carefully install the Collimator centering it with the X-ray Tube window.
Tighten carefully the four Centering Adjustment and Safety Screws (Allen) equally (same number of turns) until Collimator is centered and held firmly on the Coupling Ring (support). (Also, refer to Collimator Manual).
Illustration 3-2 Installation of X-ray Tube and Collimator
X-RAY TUBE COVER
CENTERING/ADJUSTMENT SCREWS
10. Remove the X-ray Tube and Collimator covers and connect the wires from the upper harness to the X-ray Tube and Collimator according to their identification (refer to 54301065).
X-RAY TUBE CABLE
COLLIMATOR COVER
COLLIMATOR CABLE
SM-0521R3
9
Basic Radiographic System
Service Manual
1 1. Remove now the Safety Locking Rod from the Counterweight Carriage
12. Add all the counterweigh pieces to both entrances at the back side of the
and the Central Carriage Screws (2). The carriage may go down.
Safety locking Rod
Central Carriage Screw
Column and secure them with the provided nuts. The unit is factory counterweighted, so the counterweights provided are to be installed.
Counterweight entrances
at Column Back
13. Add all the counterweigh pieces to both entrances at the back side of the Arm and secure them with the provided nuts. The unit is factory counterweighted, so all the counterweights provided are to be installed.
Counterweight entrances
at Arm Back
14. Now install the Oscillating Grid Assembly or the Fixed Grid Assembly. Follow steps 14 to 17 for Oscillating Grid Assembly or steps 18 to 21 for Fixed Grid Assembly.
10
SM-0521R3