The electrode (ELEC60/ELEC50) is
designed and manufactured by Steeper. It
is the same size and shape as electrodes
from other manufacturers and can be
directly fitted into the cavities prepared
for these electrodes. For all related
products please refer to Steeper Upper
Limb Prosthetic Components Catalogue.
i
NOTE
Ensure the patient receives the information
card included in the product packaging.
WARNING
Please read the information contained in this
manual carefully before using the product.
The electrode is to be used as the control input(s) for a myoelectric
prosthesis. It is suitable for direct skin contact:
–This electrode will control prosthetic devices from most major manufacturers.
–Proportional output signals.
–Suitable for child or adult systems.
–High sensitivity (2000-100,000 fold) and range (90-450Hz).
–Able to function with threshold control using signals as low as 10V.
–Interference protection from common power sources, depending on electrode
type, and high frequency emitting devices.
–An Electrode Gain Control (EGC) is included to allow control of the myoelectric
signal providing optimal control of the terminal device.
–Two fabrication options included in the kit.
–The contacts are coated with pure titanium, a biocompatible material and an
excellent conductor of signal.
–No known contraindications.
Page 3
Safety information
Standard compliance
–Before use, check for any visible electrode damage, particularly the titanium
contacts.
–The electrode should be placed against undamaged skin, avoiding scarred or
unhealed areas.
–Do not disassemble the electrode.
–The electrode must only be used within the power range recommended in this
technical literature.
–Do not immerse the electrode in water.
–Do not attempt to modify the product, seek specialist advice if required.
–Please ensure the patient receives the information card included in the product
packaging.
–Sensitive to EMC radiation.
–In case of an emergency remove forearm immediately. For internal batteries turn
charge point switch to OFF position. For external batteries disconnect battery
from housing.
8747
IEC 60417-5333
The CE-marking has been carried out in accordance with the requirements
of the 93/42/EEC guidelines.
Steeper operate a UKAS approved quality management system
and fully complies with the requirements of BS EN ISO 9001:2008,
ISO27001:2005, ISO50001 & TickIT Issue 5. This certifies that Steeper
meets the appropriate international quality standards for design,
manufacture and supply of prosthetic products and user software.
Steeper is registered with both the Medicines and Healthcare
Regulatory Authority in the UK and the Food and Drugs Administration
of the United States Government for the manufacture and supply of
prosthetics and orthotics products.
MHRA Registration N°: CA001031 FDA Registration N° : 9612243
BF Approved part.
Operating instruction manual.
Page 4
Product details
Mains power system map
ELEC50
Light grey
–Provides interference protection
in countries operating 50Hz mains
power systems (see map).
–Appropriate control device for
prosthetic devices from all major
manufacturers.
–Titanium contacts for optimum
conductivity of signal and
biocompatibility.
ELEC60
Dark grey
–Provides interference protection
in countries operating 60Hz mains
power systems (see map).
–Appropriate control device for
prosthetic devices from all major
manufacturers.
–Titanium contacts for optimum
conductivity of signal and
biocompatibility.
50Hz electrode (ELEC50) should
be used in these areas.
60Hz electrode (ELEC60) should
be used in these areas.
Page 5
In the box
Principal dimensions
1.
E25591
Vacuum kit
2.
E25590
Lamination kit
3.
1. Vacuum kit
2. Lamination kit
3. Electrode (ELEC50 or
ELEC60)
5.
4.
4. Adjustment tool
5. Spare suspension legs
6. User guide card
7. Technical information
A
C
B
ImperialMetric
A - Electrode width
B - Electrode length
C - Electrode height
Weight
Temperature range
Max humidity
Maximum atmospheric pressure
Power supply
Bandwidth
11
⁄16”18mm
11⁄16”27mm
3
⁄8”10mm
1
⁄6 oz4.4g
5oF to 140oF-15oC to 60oC
95%
104Pa
5V to 16V*
90-450Hz
*Steeper batteries recommended
Page 6
Further information
Gain level adjustment
A guide for using the lamination kit and
vacuum forming kit is available at:
www.steepergroup.com or
www.steeperusa.com
A printed copy is also available upon
request.
Spares of the following parts can be
ordered using the appropriate part
number.
Cable Retaining PlugB30819
Suspension LegsB31032
Lamination Kit
Vacuum Kit
Cable Retaining Plug
Gain Level
Adjustment Tool
ELECSP
To adjust and balance the electrode, turn the dial on the back of the product using the
tool provided. Number 1 is the lowest setting, number 7 is the highest setting.
1. 2.3.
WARNING
Do not force the dial
beyond the end stops.
Page 7
Connecting electrode
1.
2.
3. 4. 5.6.
WARNING
Ensure blue side
of electrode cable
(ECA=300, 600 or
1000) is facing towards
the electrode.
Use NLU0010 silicone
grease to seal the
cable retaining block.
Remove any excess
after connection.
Page 8
Electrode suspension leg removal
Electrode suspension leg reattachment
If the electrode suspension legs become worn over time, they may need replacing. There
are two replacements included. Firstly remove them as shown and then replace as shown.
1.
2.3.
Apply a small amount
of NAD0107 glue to
the underside of the
electrode suspension
leg in the area shown.
2.
Insert the suspension
leg into the cavity.
3. 1.
Remove any excess
adhesive after insertion.
Page 9
Returns
Warranty
Returns - If you need to return your product for any reason please contact
your local distributor or [email protected] or inquiries@
steeperusa.com providing full details. A returns number will be issued allowing
your items to be dealt with promptly.
Disposal - If you no longer require your product please return it to your local
supplier or Steeper/SteeperUSA with a request to dispose of it safely.
Legal - The design and manufacture of Steeper equipment and components
are subject to a policy of continuous reappraisal. The company therefore
reserves the right to introduce changes and withdraw products without prior
notice.
Warranty - This product is covered by a 12 month warranty period against
defective design and manufacture.
The electrode is a solid state maintenance free device created using ABS
plastic and pure titanium. Store in a cool dry place.
Customer Services / Technical Support - For technical assistance/support