Work surface (Table Mat)
When the path of the ground to the work surface (Table Mat) is broken or if the resistance of the path (Work surface
(Table Mat) to ground) is higher or lower than the preset value, the green LED of its testing side will change from a
green LED to a flashing red LED and buzzer alarm.
SPECIFICATION
Power source: 9VDC adaptor.
Preset impedance (Operator): The wrist strap person adjustment can be changed The monitor can be changed by
turning the adjustment screw to another value (FIGURE 2.) depending on the weight and/or height of the user.
Preset resistance (Work surface (Table Mat)): This is 10 Meg ohms. This can be adjusted to a different value.
(FIGURE 2.) Merely turn the adjustment from 10Meg to 5 Meg or 11.5 Meg Ohm.
ADJUSTMENT
Operator
A 5 Meg Ohm (original setting) was used to set up and calibrate the CM-1702. In case the user has lost or wants to
recalibrate the setting connect a 5meg ohm resistor between the ground and banana jack for wrist strap. Then turn the
calibration knob in Figure 2 until the LED “JUST” changes from flashing red LED to solid green LED to a flashing
red LED.
Work surface (Table Mat)
A 10 Meg Ohm (original setting) was used to calibrate the CM-17as example in case the user has lost on setting.
Firstly, connect a 10 Meg Ohm resistor between the ground and banana jack for work surface, then turn the knob in
Figure 2 until the LED has “JUST” change from flashing red to green or green to flashing red.
TROUBLESHOOTING
Operator
The red LED is always flashing red with a buzzer alarms even the user wears in specification wrist straps.
1. The wrist strap is broken in the coil cord.
2. The wristband on the operator is not connected correctly to the skin of the wrist. Check to see if the strap is in
contact with the skin of the operator. Use an ESD hand lotion if necessary
3. The equipment is not connected correctly to earthy/electrical ground.
4. The impedance setting is not calibrated correctly to the operator due to the difference in weight or height Turn and
adjust the impedance setting pot inside the rectangular hole in the front panel (FIGURE 2.) to a lower or higher
sensitivity.
5. The building ground wire of the building is not in good contact with the to the earth ground..
The LED is always on the green “PASS” state even the user is not wearing the wrist strap but has laid the wrist strap
on the ESD table mat or ESD work surface.
1. The sensitivity of the impedance is too low. You need to turn and adjust the impedance setting pot inside the
rectangular hole in the front panel (FIGURE 2.) to higher sensitivity.
Work surface (Table Mat)
The LED is on flashing a RED LED and alarms the buzzer even if the connection is well attached as in Figure 1.
1. The work surface (Table Mat) is not well connected correctly to the earth or electrical ground.
2. The connection wire between the CM-1701 and work surface (Table Mat) is broken or contains some internal
resistance.
3. The resistance setting does not fit to the resistance of your work surface; just turn the resistance setting pot to lower
sensitivity.
4. The earth wire of the building is not connected correctly to the earth.