Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
1. Introduction
SeaStar is supporting software for the Starmon mini temperature recorder. SeaStar
can run under versions newer than Win 2000. This manual takes you step-by-step
through the hardware connections and use of the SeaStar Program.
Please note that the downloaded file has to be installed into the existing SeaStar
folder. The update alone is not a complete program and therefore will not work like
that. Follow the instructions on the update website for correct installation.
This manual guides the users of Starmon mini through the operation of the recorder.
Setting the operating parameters as well as data retrieval requires access to a PC
computer with a standard RS-232C serial interface or USB.
Starmon mini is a compact microprocessor-controlled temperature recorder with
electronics and probe housed in a strong waterproof cylinder made of either plastic or
titanium. The plastic housing allows for an external pressure of 40 bars, which equals
a depth of approximately 400m. The titanium housing has a pressure tolerance of
1100 bar (11000m).
The serial name of the recorder is T-nnnn. nnnn is a four-digit number.
The recorder is powered by a small 3.6V lithium battery. The battery can be replaced
with the use of a soldering iron. The recorder can be sent to Star-Oddi for battery
replacement and recalibration.
Starmon mini is calibrated at Star-Oddi, and each recorder has its calibration
constants stored in its internal memory. It is recommended that the recorder be
recalibrated every 1-2 years.
To download SeaStar visit www.star-oddi.com/downloads.To install the program,
follow the automatic InstallShield Wizard. Follow the instructions on the screen. You
will be prompted for a directory name for your SeaStar program. Type in your product
key. Please keep the product key in a safe location as you may need it for future use.
The product key works for multiple computers on the same network.
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
2. Preparation
This chapter describes preparations that should be carried out before connecting to
the recorder.
Before connecting to the recorder, it is important that the clock on the PC computer is
correct. This is important since the recorder automatically downloads the PC clock
settings.
Start SeaStar and the following window will appear.
Figure 2.1 SeaStar main Window
If you are starting the software for the first time, you will be asked if you want the
Connection Wizard to open automatically each time you start SeaStar. If you change
your mind you can always change the settings under Settings-Options-Startup.
2.1 Select Recorder Type
SeaStar is compatible with other Star-Oddi instruments. A recorder type must be
selected before connecting to the recorder. The default setting in the software is set
to Starmon mini as a recorder type. Change the recorder type to Starmon mini by
doing the following:
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
Connect the USB cable and the selected port will appear in the wizard pop-up
window (see figure 3.3).
The serial port can also be defined manually in the following way:
1. Choose the Communication command in the Settings menu.
The Communication definition dialog box appears.
2. Select a serial port for communications.
3. Choose the OK button.
Figure 2.4 Communication Port Configuration
After you plug in the USB cable you will see a new communication port in your device
manager, called Prolific USB-to-Serial Comm Port. Make sure that the port is
enabled and select the appropriate port in SeaStar.
Figure 2.5 USB Communication Port in Device Manager
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
Using the Serial Port
To check which communication ports are available on your computer, open your
Device Manager. On Windows XP you can do it following way: Control panel >
system > hardware > device manager.
Under Ports you can view all available ports. Make sure that the port you are intending
to use is enabled: right-click on the appropriate port and select properties. Under
Device usage it should state use this device (enable).
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
3. Hardware Connections
This chapter provides details concerning the communication cable and how to
connect the cable to the recorder.
3.1 Communication Cable
The Starmon communication cable is an RS-232C compatible cable, specially
designed to communicate with Starmon recorders. The cable will feed power to the
recorder while communicating with the PC. A USB serial converter cable is also
available.
If the battery inside the recorder is dead, the PC and cable are still able to connect to
the recorder and retrieve data.
3.2 Connecting
Unscrew the end cap on Starmon mini. The socket becomes accessible as shown in
the following figure:
Figure 3.1 Connector Socket under the Cap
Connect the communication cable between the serial port of the computer and the
socket inside the recorder.
When data upload is completed and/or new measurement settings have been
downloaded to the recorder, the user should disconnect the recorder and close the
housing. It should be enough to tighten the cap firmly by hand. Occasional lubrication
of the o-ring seal is recommended.
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
4. Using Starmon mini - Tutorial
This chapter guides users through the process of utilizing SeaStar to set up Starmon
mini prior to recording and data retrieval.
4.1 Starmon mini Set Up
Step 1 - Start Program
Start the SeaStar software. Remember to carry out the software preparation outlined
in chapter 2.
Step 2 - Connect
Unscrew the end cap on Starmon mini. Connect the communication cable between
the serial port of the computer and the socket inside the recorder.
Step 3 - Connect to the Recorder
Connect the recorder, the selected com port will turn green in the Connection Wizard
window.
Figure 4.1 Connection Wizard
You can either click once on the selected port and select Next or double-click on the
port. If you are not using the wizard choose the Recorder menu and the Connect
command.
Battery life. Place
the mouse here to
see the remaining
battery life in %
Amount of
memory used
Recorder Mode
Status of Recorder
Serial number
Program and
start recorder
Retrieve
Recorder Data
Disconnect Recorder
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
If you have selected to use the Connection Wizard, the following window appears:
Figure 4.4 Recorder Wizard
SeaStar checks which version of firmware is in the recorder. If the software does not
recognize the firmware the user will be asked to update SeaStar as there might be a
new functionality in the firmware which SeaStar does not support.
Connection time registration
Each time SeaStar connects to a Recorder, the PC time and recorder time is
registered to a binary file, with a CLK extension. This information is also saved to a text
file with the extension CCT.
Another text file is created with the extension CCD, containing various information, for
example on previous connection and time comparisons, depending on the status of
recorder.
If a previous CCT file exists and the recorder was in sleep or slumber mode, a PC vs.
recorder clock comparison is calculated, indicating difference in seconds and ppm.
If the recorder was in measurement mode, then a drift calculation is performed, based
upon PC-
The RIT file contains partly the same information as the CCD file.
Upon retrieving date from a recorder and converting the data, the last measurement
retrieved time and the next scheduled measurement time is compared to the PC time.
The results are added to the CCD file and the MIT file.
Note: If the retrieved number of data does not match the last measured number of
data, erroneous results are to be expected.
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
Step 4 - Set New Measurement Parameters
(If the recorder has already been set up to record data and you want to retrieve data,
go directly to step 6).
Select Program and start recorder in the wizard and select which interval, single or
multiple, you want to use (see figure 4.5).
Figure 4.5 Intervals
If you are not using the wizard choose the Edit menu and the New Measurement
Sequence Definition command. The following window appears:
Figure 4.6 Measurement Sequence Definition
Select the start time of recordings (date-month-year, hours-minutes). The start time
of the new measurement sequence must be set at a minimum of three minutes in
advance. Set the sampling interval (hours-minutes-seconds).
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
If the user wants to program many recorders with the same sampling interval and start
time, it is recommended to select Multi mode. By choosing Multi mode, the settings
are fixed; giving each recorder connected to thereafter, the same settings. The
recorders that are connected to, following the first recorder, automatically get the
same settings as the first recorder. This saves time in the programming process. The
multi mode can be deactivated by clicking Cancel when connecting to the recorder
(see figure 4.7).
Figure 4.7 Multi Start Mode
The three buttons Use Template, Use Sequence, and Previous are not commonly
used. They can be used when one wants to use a previously programmed sequence.
The Use Template can be used if older definitions have been saved by enabling the
been set up in the past. With the Use Sequence button it is possible to get same
settings as from other recorders, by browsing for their serial no. By hitting the
Previous button the settings go back to previous, before changes were made. To read
more about templates see chapter 7.
After the settings have been selected, press the OK button.
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
Step 5 - Start a New Measurement Sequence
If you are using the wizard, the software will automatically start the new measurement
sequence (start recording) and disconnect the logger. If you are not using the wizard
choose the Recorder menu and Start New Measurement Sequence command. The
window in figure 4.8 appears.
Figure 4.8 Start New Measurement Sequence
Once the new measurement sequence has been downloaded into the recorder, a
window appears on the screen in order to verify the settings (see figure 4.9):
Figure 4.9 Start New Measurement Sequence
Press OK and the recorder will be set in Measurement Mode.
NMS Calculation
Upon setting up a NMS the estimated measurement sequence is calculated and
memory and battery energy usage predicted. The results are written to a text file,
bearing the recorder name with a NMS extension. If a recorder has not been selected
then the file bears the recorder type name, still the NMS extension.
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
An example of the *.NMS file is shown in fig. 4.10. The file is placed in the
SeaStar\NMS directory.
Figure 4.10 Results from NMS Calculation
When starting the recorder in a NMS, the *.RDT file now shows the sequence
settings, plus information on the sequence time duration and battery usage (same as
in the *NMS file).
Step 6 - Retrieve Data
When connecting to a logger after a measurement period, the following window
appears:
Figure 4.11 Retrieve recorder data
Click Yes if you wish to retrieve the data, click No if you would like to retrieve it later.
If you want the software to automatically retrieve data upon connection, check the
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
Choose the Recorder menu and the Retrieve Data command. The following window
appears:
Figure 4.12 Retrieve Data from Recorder
The window in figure 4.12 shows the data retrieval progress in bytes, time elapsed
and number of packs received.
Data will be uploaded from the recorder and presented in graphical and tabular form.
With default settings, SeaStar automatically creates a sub-directory in SeaStar
Starmon miniStarmon mini
nnnn) is created. All files related to the recorder
are automatically saved in the serial number directory. The default data directory can
be viewed in the Settings menu, and Directories. The user has the option to define a
new data directory path if desired.
The data is retrieved in a non-destructive manner. This means that you can repeatedly
read the data without erasing it from the memory. Each time data is retrieved, a new
file will automatically be created. The name of the data file consists of the
measurement sequence number and the serial number of the recorder. The recorder
cannot be loaded with new settings if it contains data that has not been retrieved. This
is a safety feature which prevents data from being accidentally lost.
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
Once the data has been retrieved the following window appears:
Figure 4.13 Recorder wizard
If you wish to start a new measurement sequence you can either program the logger
with a new measurement sequence by selecting Program and start recorder or
Restart recorder with the same sampling interval.
Step 7 - Disconnect
Disconnect the cable and close the housing. The cap should be tightened firmly.
Occasional lubrication of the O-ring seal is recommended.
when being disconnected from the computer but will continue consuming battery. To
put your Starmon into sleep mode select Disconnect Recorder in the wizard or
Recorder > Disconnect to go into off-line mode.
Step 8 - View Data
When data is retrieved from the recorder, a new file will automatically be created and
opened in graphical and tabular form. To open existing data files on the computer,
choose the File menu, Select Recorder command, and then the Select
Measurement command. When a data file has been opened, a new window appears
with the data shown as a chart and a table. The next chapter will describe in detail
how to view and process data.
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
5. View Data
This chapter describes how to open a data file and the options available in SeaStar for
viewing the data.
5.1 Open Data
To open an existing data file that has already been retrieved, do the following:
1. Choose the File menu and the Select Recorder command. When a recorder
has been selected, its serial number will be displayed in the blue top bar of the
software.
2. Select the data file by choosing the File menu and the Select Measurement
command.
The name of the data file consists of the number of times the recorder has been set up
to measure plus the serial number of the recorder. When the data file (.mid) has been
selected, the chart and data tables are opened. The following figure shows a chart
and data table which have been opened:
Figure 5.1 Chart and Data Table of an open File
At the bottom of the SeaStar window you can view the measurement sequence
number, which indicates the number of times the recorder has been set up to
measure.
5.2 Import
Data files from other recorders than Starmon can be imported into SeaStar and
viewed as charts and tables. The imported file must be a text file with the extension
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
DAT, and with SeaStarata files
and headers in the Appendix.
5.3 Buttons
When the chart is opened, buttons related to the chart are displayed.
Active Series List
This button can be useful when working with a multi-sensor recorder, in order to
select which parameters to display in the chart. Click the button and the following
window appears:
Figure 5.2 Series list
By selecting Only show active table columns only the selected series will be
displayed in the chart table.
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
Under Bin Settings there are two options, Default and User. Choose User to select
the number of bins and the minimum and maximum values you wish to display.
The Histogram can be printed out by clicking the Print button.
Save Window Data
Retrieved data is automatically saved, but if changes are made with the graph, for
example with the time scale, the save button enables the user to save the graph (as a
DAT file).
Print Chart
The chart can be printed. For more details see chapter 6.Printing.
Zoom Out
Click this button to zoom out the chart.
Zoom In
Click this button to zoom in on a previous view.
Adjust Axis
This button adjusts the axis in order to display all data within the zoomed timeline.
Open file in PatternFinder
Click this button to transfer data to PatternFinder
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
Display table completely
Partial table display
Hide table
5.4 Using the Mouse
Min and Max Values
Click once on the temperature line in the chart, and the maximum and minimum values
are shown. A straight vertical line should be visible in the graph (to the left). By
clicking on this line and holding the left button on your mouse down, the line can be
dragged through the chart and the data-table will scroll along with the position of the
line.
Zoom
The chart can be zoomed in on by holding down the left button of the mouse, and
using the cursor to create a box to the right, around the desired area to zoom in on.
Scroll the Chart
The chart can be scrolled upward, downwards and to the sides, by holding down the
right button of the mouse and moving the cursor in the desired direction.
Go Back to Initial View
To go back to the initial view of the chart, simply hold down the left button of the
mouse and make a box to the left with the cursor. The position of the box on the chart
is irrelevant.
Change the Axis
By double clicking on the X and Y-axis, the measurement time and temperature range
can be changed.
Change the Title
To change the title of the chart, double click on the title of the chart.
5.5 The View Menu
When a recorder has been selected, the files in the View menu can be opened.
Data Trend and Table
This view is automatically selected when data is retrieved or a data file is opened. If
the chart is closed, but a recorder has been selected, this command can be selected
to view the chart. The software will open the latest chart that was viewed for that
recorder.
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
Recorder Information (RIT)
This is a text file containing basic information about the recorder, serial number,
production date, estimated battery life etc.
Recorder Calibration Information (RCI)
This text file contains information about the calibration constants of the recorder,
calibration range of the parameters. The range is both showed as decimal values (DV)
and unit values (UV). The survival depth of the recorder is also shown.
Recorder Download Information (RDT)
This text file contains information about the measurement sequence that was last
downloaded into the recorder, including the start time, sampling interval and the
measurement sequence number.
Viewing Text Files
Under <View\Text files> the user can select a text file, via a file browser, for viewing.
This option is meant to give the (advanced) user a chance to view files that are not
directly supported for viewing in SeaStar, but are a part of the SeaStar
documentation system.
The files that the file browser accepts must have the following extensions:
TXT
Frequently, users copy SeaStar data to text processors and spreadsheets, like
Word and Excel, and later save the data as TXT files for viewing in other
applications.
INI
There are a number of INI files that SeaStar uses for store settings. These files
The latest members in
the INI files family are the Template files.
DAD
This is the original data file. It contains the raw data (in Ascii format) retrieved
from recorder. The DAD together with the MID are converted to the DAT file.
A word of warning: The DAD file should never be tempered with, as that might
give cause to erroneous data conversion.
CCT
Every time SeaStar
clock data are stored in a <recorder name>.CCT file.
CCD
Every time SeaStar connects to a recorder, the
time are compared to the last CCT, if found, and stored in <recorder
name>.CCD file, <measurement sequence name>.CCD.
NMS
Upon completion of a NMS editing session, estimation of memory and battery
usage is calculated and the results placed in a NMS file, which is then
displayed to the user.
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
SNMS LOG File
success or failure, a log file for each recorder type is placed in the SeaStar directory.
The log file name, for Starmon mini, is DSTT.LOG, i.e. the same naming convention
as for the type INI files.
Each time a SNMS is performed then the appropriate LOG file is updated.
\
In addition to the LOG file, if an error occurs during the SNMS, the MDD file is
created as a text file, with a description of what went wrong. This MDD file is
displayed as a clear message for the user (figure 5.8).Usually the MDD file is a binary
file, and the MDT is created from the MDD. If the SNMS is repeated successfully the
MDD file is overwritten and the MDT file created.
Figure 5.8 The MDD SMNS Error Message
Measurement Data
Measurement Download Definition (MDT)
This text file contains information on data downloaded into the recorder for the
measurement file being viewed, including the start time and sampling interval.
Measurement Information (MIT)
This text file contains information concerning the measurement upload time, start date,
start time and number of measurements in the measurement data file.
Conversion Information (CIT)
This text file contains conversion information on the units and the data file format.
Statistics in CIT:
The CIT, has been enhanced with more detailed information on the converted data.
As can been seen in figure 5.9, the statistical data, that have been tabulated, are split
into two sections, the Unit values statistics and the Decimal value statistics. Values are
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
A one line description/comment is available for the user to fill out in the Reconvert
data menu. This description is placed in line two of the CIT file.
Figure 5.9 Statistics in CIT File
Out of Range (OOR)
This text file shows the number of measurements recorded that were outside the
calibration range.
Binary Data (DAB)
This file contains raw data of the measurements in the data file. Calibration files are
not calculated into the data.
View DAB Files as Trend
When converting the DAD -> DAT the DAB file is created simultaneously.
-4095.
These values originate from the recorders 12 bit ADC (Analog to Digital Converter).
It can be of interest to view the data in its native form, for example to spot
saturation/out of range intervals.
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
Figure 5.10 DAB Trended
SeaStar cannot view older DAB files, and will inform the user to that effect, if an
attempt to view an older DAB file is performed. By reconverting the data, a viewable
file is produced.
Binary Pressure Temperature Corrected Data (BPD)
Pressure measurement circuitries are calibrated against measurement errors due to
temperature changes.
Raw decimal values from the recorder can be seen in the DAB file. The first step in
converting decimal values to unit values is the circuitry temperature correction. The
result of this first correction can now be seen in a new file in SeaStar, *.BPD. The file
contains the corrected decimal value and the difference between the original value
and the corrected one.
The file can be viewed graphically or as tabulated lists.
Converted Data (DAT)
This is the text file for the converted data. The DAT file as well as the other text files
can be opened in other programs (Excel, Notepad, Word etc.).
Time Synchrone Data (TSD) and Rate of Change (ROC)
If the TSD and/or ROC file creation has been enabled under the Settings menu, then
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
5.6 Opening Data in Excel
When retrieving data, the SeaStar software automatically creates a text file with the
data, called *.DAT. This text file can be opened in Excel and in other programs.
Joined Date and Time
option is not disabled, the following is
performed, for opening data in Excel:
1. Excel is started.
2. The \
3.
a. No changes should be made in first step, so hit the next button.
b.
c.
4. You can now create the graph with columns B and C.
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
6. Printing
All charts and text files can be printed from the SeaStar software.
6.1 Charts
Open the chart you want to print out and select File-Print on the menu bar. The
following window appears:
Figure 6.1 Print Preview
Orientation
The chart can be printed in either portrait or landscape format.
Margins
-scaled. The margins
can be varied by either changing the numbers given for the left, right, top and bottom
margins, or by moving the margin lines with the mouse by pressing and holding the left
button as the lines are dragged.
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
Move
When the cursor is located on the chart, a hand becomes visible. By clicking on the
chart and holding down the left button of the mouse, the chart can be moved around
the page to a desired position.
Details
The size of horizontal and vertical gridlines as well as the texts on the X and Y-axis can
be re-scaled by using the scroll bar. The number of gridlines is increased when the
bar is moved to the left.
Another way of printing out a chart is opening the chart you want to print out and
clicking the Print Chart button on the chart bar.
6.2 Histogram
To print out a histogram, open the chart which histogram you want to print out and
click on the Histogram button on the chart bar. The following window appears:
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
Click Print and the window in figure 6.3 appears:
Figure 6.3 Output Options
To change the selected printer, click Setup and select another printer.
To see a preview of the print out, select Preview under Report Destination and click
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
6.3 Files
Activate the file you want to print out and select File-Print on the menu bar. The
following window appears:
Figure 6.5 Output Options
The dialog window in figure 6.5 has the same functions as the one in figure 6.3.
Note that as some of the files are very large, it is recommended to highlight the data
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
7. Settings
This chapter describes the options in the Settings menu. These settings are
automatically saved when exiting the program.
7.1 Directories
The path definition can be set as Default SeaStar path structure or User defined
paths. Choose the Settings menu and the Directories command. A dialog box
appears:
Figure 7.1 Directory Path Definition
Default SeaStar Path Structure
With default settings, SeaStar Starmon mini
SeaStareated when the software was installed on the
computer. After connecting to a recorder, another directory is automatically created in
data files and other files relate
directory. Each Starmon mini recorder has its own data directory.
User Defined Paths
If this option is selected, the User definition data directory becomes active as shown
in figure 7.1. The data path/directory can be defined by the user. For defining a path
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
on the network, the particular drive must be mapped (under Windows Explorer) in
Create recorder directories
This option is default on, and means that a recorder type and serial number directory
is created for every recorder connected to.
Browse first in user directory
With default off, the user is first directed to the path where SeaStar was installed, to
look for the INI file for a particular recorder. If this option is enabled, SeaStar will start
looking for the INI file in the user-defined directory. More information on files and their
structure can be found in the Appendix.
7.2 Preferences
Units
By choosing the Settings menu, the Preferences command, and Units, a dialog box
appears:
Figure 7.2 Temperature Unit Dialog Box
Select Units
The user defines whether degrees Celsius or Fahrenheit are used as a unit for
temperature measurements.
Select Number of Decimals
The user defines the number of decimals for temperature values displayed in SeaStar
and for text files that are created.
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
7.4 Connection
Choose the Settings menu and the Connection command. A dialog box appears:
Figure 7.4 Connect and Retrieve Data Options
Perform CRC Test When Retrieving RID
The CRC test is a safety check performed by the software when retrieving Recorder
Information Data (RID). If the CRC test is not successfully carried out, it means that
some errors have occurred in the transfer of files or constants, between the recorder
and the computer. If a CRC test failure occurs, the user is notified. Normally, this
option should always be selected.
If data is in the recorder then:
If this option is enabled, data retrieval is done
manually via the recorder wizard or by choosing the File menu, and Retrieve Data
command.
Automatic data retrieval upon connection: By enabling this option, the software will
automatically check for data in the recorder upon connection and retrieve it.
Ask user whether to retrieve the data: By enabling this option, you will be asked
whether you want to retrieve the data or not each time you connect to the logger.
Perform Comparison Test on RID in SNMS
The comparison test takes place when the PC has sent all the settings data to the
recorder and it reads the whole file back and compares to what was sent. Bad
communication can cause errors in the information file when read back. If the data
echo check is enabled it is assured that the information file was transferred correctly
to the recorder, therefore it is OK to disable the comparison test.
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
7.5 Retrieve Data
Choose the Settings menu and the Retrieve Data command. A dialog box appears:
Figure 7.5 Retrieve Data Options
Retrieve Number of Data
With default settings, the software retrieves the last measurement sequence recorded.
By choosing the User defined number of data, the number of measurements
retrieved from the recorder can be set.
As the complexity of measurement intervals give incoherence in the calculation from
measurement number to the number of data byte to retrieve, the user has the option to
define number of bytes to retrieve.
If for some reason or another it becomes necessary to retrieve data more than once or
even several times, the previous DAD and MIT files are not overwritten before a
or
the last DAD file.
A backup number system is in use that marks the first retrieved file as number one, the
second number two etc. The number sign #, is used to identify the backup up files.
Thus if a data file has for example the name 25M1106.DAD then the backup file, that
was the first retrieval, gets the name 1#25M1106.DAD, and the 25M1106.MIT file
that is associated with the DAD file, is backed up as 1#25M1106.MIT.
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
7.6 View Data
Choose the Settings menu and the View Data command. A dialog box appears:
Figure 7.6 View Data
Select View Data Trend after data retrieval/conversion to view the data right after
data retrieval.
Select View Data Trend after Measurement Sequence Selection to work with
measurement data while offline.
With all the options in SeaStar, the user may want to do a lot of reconverting
(<File\Reconvert Data>) on data files after defining the preferences; changing units,
data formats, OOR settings, TSD settings, ROC settings, offset settings and
converting different sections of the data file, all of which result in numerous files. To
ease the task of remembering settings and results and/or renaming and copying files,
a backup system of reconverted files has been implemented.
A reconvert number is now associated with each sequence. This number is saved in
made of the DAT, CIT, DAB, OOR, TSD and the ROC files with the reconvert number
attached to the filename.
Thus if a data file is for example named 5M1024.DAT, then after the first reconvert, a
copy of the DAT is named 1-5M1024.DAT, and the same naming convention is
applied to the other files.
Each time data retrieval is performed, a subdirectory, which name is the sequence
numbe
files are placed in the sequence directory. The reconvert number system is optional,
and the user can disable it in the reconvert menu. When the reconvert number system
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
7.7 Out Of Range
If measurements are outside the calibrated range, they may seem inaccurate. The out
of range measurements are shown in the OOR file (View Measurement Data Out
Of Range Data). The min. and max. calibration points are given in the RID file, and in
Reconvert Definition Data
See chapter 9 for more.
The user can also define the ranges. Choose the Settings menu, Out Of Range - UV
and DV definition command. A dialog box appears:
Figure 7.7 Set Temperature Out Of Range (OOR) Parameters
Decimal Values
Decimal Values (DV) are raw measurements, sometimes referred to as binary data.
The decimal value is between 0 and 4095, which is the range for the 12 bit AD
define the decimal value range, unless he has obtained these values for that particular
recorder from Star-Oddi.
Unit Values
The decimal values are converted to Unit Values (UV), for example a decimal value of
2000 could give 20°C. The user can select a specified range for temperature and
pressure. If unit values in the data are outside the measurement calibration range of
the recorder, the user has the option to set a max and min range for these unit values.
After selecting the unit value ranges and pressing the OK button, the following
window appears:
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
Figure 7.8 Out Of Range (OOR) Correction
If no values have been selected for the Decimal Value (DV), no correction should be
selected. If Unit Values (UV) ranges were specified, the user has the option to change
the out of range measurements to the last value that was recorded before the out of
range measurement was recorded, or have it according to the max/min unit value as
specified earlier. In this case the software will know if the value is under the minimum
or maximum limit. After the correction has been made, the data needs to be
reconverted under File Reconvert Data, and data will be changed accordingly.
7.8 Time Synchronize Data (TSD)
Choose the Settings menu, TSD definition command. A dialog box appears:
Figure 7.9 Time Synchronous Data Settings
If the user wants to time synchronize two or more data files, it is possible to select
three options: 1 hour, 10 minutes, or 5 minutes. The restriction is that the sampling
interval selected must be greater than the initial interval, and it must also be dividable
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
by the initial interval. For example, if the initial interval was 15 minutes, only the 1 hour
option can be selected for time synchronous.
measurements in the original file that are not to be shown.
Re
7.9 Rate Of Change (ROC)
This new feature gives the user a chance to see how fast the signal is changing, i.e.
the estimated velocity of the signal. This applies to all measured parameters. The ROC
is calculated via a two point numerical differentiation:
ROC= (P2-P2)/(t2-t1)
P: Parameter expressed in a user defined unit
t: Time (hour, minute or second)
The way the ROC is set up and viewed is very similar to the TSD.
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
The ROC is only computed when a DAD->DAT conversion is performed, and when
the user wishes. The time units are selectable as hour, minute or seconds. For small
changes it can be convenient to express the ROC as a milli (10-3) unit. For example if
the depth is in meters (m) then millimeters are expressed as (mm).
\Data Trend and Table\Rate Of
\Measurement Data\
7.10 Chart
Title
To change the title of the chart, click on the title name. A dialog box appears:
Figure 7.11 Chart Title Dialog Box
The chart is automatically assigned the data file number as a chart title. To change the
chart title, enter a new text string and press the OK button.
Time Axis
To change the time scale of the chart click on the X axis. A dialog box appears:
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
By disabling the Automatic option, the minimum and maximum time values become
active. By clicking on the Change buttons for the maximum and minimum values, the
time range can be specified. By enabling the Automatic option, the measurement time
will go back to the initial range.
Value Axis
To change the scale settings for the value axis, click on the Y axis scale. A dialog box
appears:
Figure 7.13 Y-Axis Settings Dialog Box
By disabling the Automatic option, the minimum and maximum measurement values
become active. By clicking on the Change buttons for the maximum and minimum
values, the temperature range to be shown on the chart can be specified. By enabling
the Automatic option, the temperature will go back to the initial range.
7.11 Chart Editor
The Chart Edit button (on every graph) gives access to the Chart Editor, a multioption menu, now with a Tree View interface, as can be seen in figure 7.14.
There are five main branches on the tree:
Series
Chart
Data
Tools
Export
The tree menu gives access to the same features as the tab menu described in the
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
Chart
The chart branch has six sub branches
Axis
Titles
Legend
Panel
Paging
Walls
Axis
There are 8 axes in a normal SeaStar chart, the number in use depends on the
recorder type. The Left Axis and Custom 1 axis are always in use, the latter is reserved
for the timeline.
Figure 7.16 Axis Formatting
Figure 7.17 shows a six tab Axis menu. There are numerous options, some of which
can drastically change the behavior of the graph. The Minimum and Maximum tabs
give options similar to the menu offered when clicking an axis on the graph, the ticks
tab gives access to
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
Titles
The Titles menu shown in figure 7.17, gives access to 4 tiles, each title a multi line text
window, which can be positioned anywhere on the chart. There are multiple formatting
options, six tabs, for each title.
Figure 7.17 Title Menu
Legend
The legend menu (see figure 7.18) gives the user an option of showing a series
legend.
The 8 tab menu gives options in styling positioning and formatting the legend.
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
Panel
The Panel menu (see figure 7.19) offers panel format options. Images, pictures and
icons can be imported onto the panel as a back image.
Figure 7.19 Panel Menu
Paging
With the Paging menu, shown in figure 7.20, the data series can be chopped into
pages, with a user defined number of points per page. Only one page is shown at a
time, and the user can scroll through the pages via the page buttons, that appear (are
enabled) on the chart to the right of the Edit Chart button, when paging is selected.
Paging can be useful for viewing distinct time periods, one at a time, days or hours for
example. No offset is offered and sampling periods and start time can limit the
usefulness of this option.
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
Walls
The walls menu, as seen in figure 7.21, gives access to configuration of four walls, but
as SeaStar
Figure 7.21 Walls Menu
The Walls menu, has similar options as the Panel menu, when defining a pattern, via
on the panel, or the image on the panel can be seen through the wall with the
Transparency option.
Data
In the data menu, seen in Figure 7.22, gives the user access to the database, upon
which the chart rests. Here text labels can be set to each measurement point.
Changes can be made to the time and measurement values. Changes made to the
data, are included when the data is exported, but if the data are viewed again or
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
Tools
Under 'Tools' you will find several options to customize the setting of the cursor and
the appearance of the graph, such as page numbering, font, font size etc. To
active/deactivate the chosen setting check/uncheck the 'Active' box on top of the first
page of this menu.
Figure 7.23 First Page of Tools Menu
Export
Under the Export menu, three tabs offer as many menus for saving data as
Pictures
Native (data)
Data (file)
Pictures
Figure 7.24 shows options when saving the chart as a picture to a file, or copy the
picture to the clipboard. There are seven format possibilities.
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
Figure 7.26 Export Menu: Data
There are four Include options:
1. Point Index
The point index is the same as the measurement number, but offset by -1, as it
starts in 0 instead of 1.
2. Point Labels
If the user has added text labels to points in the DATA menu, these can be
included.
3. Header
The same headers as are in the DATA menu.
4. Point Colors
This option is not relevant in SeaStar.
Another method for exporting is to use the Alt+PrintScreen keyboard buttons to copy
an image of the SeaStar window to the clipboard, thereafter to paste it to another
Windows application. When a dialog box is open, then only that frame is copied to the
clipboard.
To copy a part of a data file, for example from a DAT file, then highlight the part you
want to copy and use the Ctrl+C keyboard buttons to copy the data snippet to the
clipboard. Then use the paste-option in another Windows application to transfer the
data from the clipboard.
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
Figure 7.27 Highlighted Text to be Copied
7.12 Timeline
The timeline is a graphical viewing aide. Every time a file is loaded for trending, an
extra time value is calculated for every measurement point. These time values form the
Timeline. The Timeline is an extra series that follows the rules of the other
measurement series. When the graph first appears, the Timeline is inactive.
By selecting the Timeline in the Series List (see fig. 7.28), the Timeline becomes
visible as shown in figure 7.29.
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
Fig. 7.29 Timeline on a Graph SY
The Timeline is defined under Settings\Preferences\Timeline.
There are several options for the Timeline, as can be seen in fig. 7.30.
Figure 7.30 Timeline Options
There is no need to reconvert the data, as every time a data file is loaded the Timeline
is calculated in accordance with Timeline settings. The Timeline is not a part of the *:
DAT or the *.TSD files. If the data is exported (see the Export section), then the
Timeline (if selected) will be included. Thus, the Timeline can also be used for
statistical purposes.
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
7.13 Templates
There are two scenarios where templates come in handy:
1. When many recorders are programmed with the same settings (multi mode),
and circumstances are in a way that the user cannot complete the setup of all
the recorders in one session, and he has to exit SeaStar (to continue later).
2. When the user wants to reuse, partly or completely, presumably a complex
and/or lengthy measurement setting.
Saving a Template
The user can now, when editing New Measurement Settings (NMS), save his work by
plate (fil1), before
acknowledging OK from the Edit NMS menu. Having selected this option and the
NMS are OK, the user has an option to write a description for the template (see fig.
7.33).
INI file and saved in the SeaStar\NMS directory. The name of the file is
MST<date><type letter>time<Interval mode>.INI.
Fig. 7.31 Template Options in Edit\New Measurement Sequence
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
Figure 7.32 Saving NMS as a Template
Using a Template
Templates can be retrieved at a later time to be used directly or modified. The list of
templates is kept in the SeaStar\templets.ini file.
There are three buttons in the Edit NMS window, see fig. 7.31. The top button is used
for retrieving a template.
As can be seen in fig. 7.33, templates are selected via template name or description.
Parallel to selecting a template, descriptions can be edited and eventually templates
deleted.
The recorder type and interval type, which the template was based on, are visible for
each template.
Figure 7.33 Template Selection
Templates can be used across recorder type boundaries and interval types.
Things to keep in mind when selecting a template are:
Single interval templates converted to a multiple interval NMS. All seven
intervals are the same.
Multiple interval templates converted to a single interval NMS, set the interval
as interval one.
Start time in the past is not allowed as will be apparent when exiting the Edit
NMS window.
Some recorder types are less compatible than others, an example of that is a
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
After selecting a template, the user is returned to the edit window. If the selection is
not
The user has one more option to quickly set up the NMS. The button in fig. 1, marked
and return four different settings depending on the status of recorder and
measurement selection.
1. Use Default: No recorder selected.
2. Use Recorder: Recorder selected but not a measurement sequence.
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
8. Multiple Intervals
SeaStar has the option for setting up more than one sampling interval within a
measurement sequence. Reasons for wanting multiple intervals can be to save
memory, but still be able to have frequent measurements in between.
Select Multiple recording intervals in the wizard or choose the Edit menu and the
New Measurement Sequence Definition command. Select Multiple Intervals and
the following dialog box appears:
Figure 8.1 Multiple Interval Selection
8.1 Start Time
The user defines the date and time when recordings should start.
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
8.2 Measurement Interval
The first and the second interval is defined. Number of measurements must be defined
within each interval. The recorder will first take the defined number of recordings
within the first interval, and then shift to the second interval to record the defined
number of recordings for that interval. The recordings will shift between two intervals
and continue until memory is filled or connection with recorder is establised.
8.3 Set Mode
to enable the user to fix the settings, in case many
option has been selected, the same settings will automatically be downloaded into the
following recorders that connection is made with.
The Use Template can be used if older definitions have been saved by enabling the
been set up in the past. With the Use Sequence button it is possible to get same
settings as from other recorders, by browsing for their serial no. By hitting the
Previous button the settings go back to previous, before changes were made.
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
9. Reconvert Data
In SeaStar you have the option of reconverting data that has already been retrieved.
After re-
main reasons for reconverting data are:
Incorrect PC-time at measurement start-time.
Need for temperature unit alteration.
Need for data file format alteration.
9.1 Data Definition
By choosing the Edit menu and Reconvert Data Definition, a dialog box appears:
Figure 9.1 Reconvert Data Definition
Set Reconversion Measurement Sequence Start-Time
If the user notifies from retrieved data that the date and clock are incorrect at the start
of measurements, then it is possible here to specify a new start date and time.
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
In Reconversion Definition Use Measurement Intervals from
The measurement interval is either the interval settings as shown in the measurement
(MID) file, or if that interval is incorrect, a new and correct sampling interval is
In Reconversion Use Constants from
The user can choose to use calibration constants from:
MID file. This option should normally be selected, as it contains information on
the constants from the measurement sequence.
RID file. The recorder information file contains the same calibration constants
as the MID file, except if it has been re-calibrated. If it has been re-calibrated,
and the user wants to use the new constants on older measurements, the RID
option should be selected.
RBD file. This option rarely needs to be selected. If the constants from the MID
and RID file are corrupt, the user should contact Star-Oddi for an RBD file.
See chapter A.3 for instructions on how to transmit the RBD files to the
recorder.
Convert Number of Data According To
You can reconvert all the data in the last measurement sequence, or you can define
the number of measurements to be reconverted. The user has the option to define
what measurement range should be reconverted (for example to exclude data points
when the recorder was above water).
Time Linear Adjustment
The real time clock inside Starmon has an accuracy of approximately +/- 1 minute per
month. This slight variation might accumulate over time and influence the
the measuring period.
Out of Range Decimal Value Settings
Measurements that are outside the calibration range are inaccurate, and may give
erroneous results. The out of range (OOR) measurements can be seen in the OOR
file. The user has the following decimal value out of range detection options when
reconverting:
Full scale (no boundaries).
made.
User definition. - Out Of
Range (DV) are not available in the RID file. This would normally require the user to contact
Star-Oddi for the decimal values.
RID values. The calibration decimal range is given in the RID file. This is the default
option and should normally be selected.
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
If an RBD file exists in the \Seastar\RBD folder, then SeaStar recognizes this file as a
primary source for DV calibration range. In this case, the fourth option appears and is
set as default.
9.2 Temperature Definition
Under the Edit menu and Reconvert Definition, choose the Temperature Definition
command, and the following dialog box appears:
Figure 9.2 Temperature Reconvertion Definition
9.3 Offset Adjustment
When Starmon mini recorders are used over a longer period of time the temperature
measurements can start to deviate from correct values. This is usually due to aging
and Star-Oddi recommends the recorders to be calibrated after 2-3 years.
When recalibrating, the battery can be replaced as well, so the recorders are good as
new. When checking older recorders, it has come apparent that the error in the
measurements is mainly due to an offset. Some users feel they can benefit from an
offset correction option in SeaStar, for example when a recorder has been in the
water for a long time and is later tested to show an offset.
To correct for a temperature offset, use the temperature reconvert definition menu.
Using the Settings menu and Preferences, you can also change the format of the
temperature unit and the data file. For reconverting the data, the data file must be
open. To start reconverting data according to the new settings, choose the File menu
and the Reconvert Data command. The data file is reconverted and conversion data
files are created (CIT, DAB, DAT). A new chart file is now open.
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
Single spike correction
Check Use Single Spike Correction. There are two variables, Time and Limit. The
Limit defines how much difference can be between two points over a defined Time.
If a point is outside the range, +/- the limit of adjacent points. That point is then
calculated as the average of the adjacent points.
Temperature response shift
It is possible to set the desired start/end point to be shifted by X seconds. This shift
can only be performed on data with an interval of 1 second.
Linear estimation outside calibration range
Check this option to linearly estimate temperature measurements outside of
calibration range.
View Previous Conversions
Each time a reconversion is performed, a numbered file is created in the sequence
folder. To view previous conversions select a file under View and choose Previous as
can be seen in figure 9.3.
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
The window in figure 9.4 appears. Select a numbered file from the drop-down list.
Figure 9.4 Select Previous Conversions
Reconvert Previously Retrieved Data
If data from the same sequence is retrieved more than once, a bookmarked data
sequence file (DAD) is created. These files can be accessed and reconverted by
choosing File-Reconvert data-Previous as can be seen in figure 9.5.
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
10. Recorder Modes
10.1 PC Mode
If the recorder is plugged into a PC computer with the communication cable, by
choosing the Recorder menu and the Connect command, the recorder is put into PC
Mode. At the bottom of the SeaStar window, the text On-line is displayed, indicating
that the recorder is on-line and in PC Mode.
While in PC Mode, the recorder will remain in an active listening stage. The
communication protocol is a simple master/slave arrangement, in which the recorder
is the slave. The slave will perform every command the PC issues, and after execution
of a command, it will return to listening.
10.2 Sleep Mode
The user receives the recorder in Sleep Mode, but it can easily be awakened by the
Connect command in the Recorder menu. The recorder is put into Sleep Mode when
the recorder is disconnected without putting it into Measurement Mode or by
choosing the Recorder menu and then Put Recorder in Sleep Mode. When the
recorder is in this mode, less power is depleted from the battery than in other modes.
In this case select Recorder > Disconnect.
10.3 Measurement Mode
The recorder is put into Measurement Mode by choosing the Recorder menu, and
Start New Measurements Sequence. The recorder starts recording measurements
according to the start time (date/month/year) and sampling interval (hours/min.)
defined by the user.
When the memory is full, the recorder shifts automatically into Sleep Mode. Upon
connection and command from the PC, the recorder is put into PC Mode.
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
11. Memory and Battery
11.1 Memory Organization
Data is kept in an EEPROM memory and will not be lost if the battery fails or expires.
The memory of the recorder is organized as shown in the figure below:
Figure 11.1 Data Memory Organization
When data is retrieved from the recorder, the following occurs:
1. The computer retrieves the data from the recorder and writes the data to the
The recorder has stopped recording, but still holds the recorded data.
When new settings are sent to the recorder, the following will occur:
New settings are sent to the recorder.
The data memory pointer is reset to the beginning of the data memory and starts to
write over older recordings.
The organization of the memory is now as follows:
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
11.2 Battery Calculations
The voltage characteristics of the batteries used in our recorders make it almost
impossible to measure the energy available in the batteries, therefore we estimate the
energy consumption by calculation. All the energy consuming functions of the
recorder, like sleep, decay, running (microprocessor awake), measurements and
saving data are measured beforehand and used in our calculations. The calculations
recorders themselves do not perform any energy calculations, but in order to start the
recorder in one computer and retrieve data in another while still having the battery
energy information, they carry the percentage of remaining energy with them in a
register file (in the RID). At the time the recorder is in production (when the battery is
new) the energy number is set to 100%.
Each time the recorder is started up in a new measurement sequence, the energy
consumption, since last sequence-start, is calculated and subtracted from the energy
number retrieved from the recorder, and this new energy-left number is downloaded
into the recorder.
The energy-left register in the recorder can be corrupted for reasons such as:
The CRC check is turned off when connecting to the recorder and a
communication error occurs
When a write error occurs when starting up a new sequence or updating a
status to the recorder and a reconnection is performed instead of performing
the operation again.
Energy estimation calculations are not in accordance with actual energy
consumption.
To change/correct the battery energy numbers in the recorder refer to chapter 13.5.
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
12. Wizards
There are seven wizards in SeaStar:
12.1 Connection Wizard
Figure 12.1 Connection Wizard
Choose the Wizards menu and the Connection Wizard command to enter the
connection wizard (see figure 12.1).
12.2 Cable Connection
Figure 12.2 Cable Connection Wizard
Choose the Wizards menu and the Cable Connection command to enter the Com
Port wizard (see figure 12.2). The wizard goes through available ports and activates
them one at a time and asks the user to confirm that the communication cable is
connected to this specific port (see figure 12.3).
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
12.5 Quick Retrieve and Restart Recorder
Choose the Wizards menu and the Quick Retrieve command to enter the wizard
(see figure 12.6).
Figure 12.6 Quick Retrieve Wizard
This wizard retrieves data from the recorder without graphically displaying it and offers
the user to start it again with a new measurement sequence.
12.6 Change Battery Life
A new wizard has been added to the Wizards menu, the Change battery life wizard.
Choose the Wizards menu and the Change battery life command to enter the
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
Select the recorder type and insert the recorder in the Communication Box. Once a
connection has been established a window appears (see figure 12.8). If the recorder
is already on-line then the wizard will start in step 4 (Set New Battery Life).
Figure 12.8 Set Battery Life
The top line shows who long the recorder has been in use since (last) calibration, and
how much energy sleep and decay combined have used up over that time.
The second line shows Battery left based on the register in the recorder.
There are two ways to set the battery life, Direct definition and Calculate from NMS.
Direct definition:
Check the Direct definition option and set the battery life (in %) manually based on
used since then. The user can also refer to other same type recorders working under
similar conditions.
Calculate from NMS:
Check the Calculate from NMS option and click on Calculate NMS. The user is now
transferred to the Set New Measurement Definition window where a sequence can
be defined. Once the measurement sequence has been defined an Estimated
measurement battery usage pr week (%) is calculated and shown in the Set New
battery Life window, note that sleep and decay energy use is included in this number.
Also the Calculated total battery used (%), which is based on the NMS definition
times number of weeks, plus the energy used by sleep and decay.
Now insert for how many weeks the recorder (since produced) has been measuring.
Each time the number of weeks is changed, the Calculated total battery used(%) is
recalculated, note that the Calculated total battery used(%) value will not show a
directly weeks * energy pr week, as the sleep + delay values must be accounted for.
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
12.7 Insert New Calibration Constants
Placing new calibration constants in a Starmon mini recorder is a 2 step operation.
1) The user makes a NCT file and places it in the SeaStar\RBD\ folder.
2) Run the Insert New Calibration Constants wizard
New Calibration Results -> NCT file
The Starmon mini recalibration should produce a 5° polynomial
y =C0+C1*x C2*x^2+C3*x^3+C4*x^4+C5*x^5
y = temperature in °C
x = decimal values (recorder raw measurement values)
Having recalibrated a Starmon mini recorder the results must be placed in a text file
named after the recorder and with the extension NCT. For example, for recorder
T4605 the NCT file should be named T4605.NCT.
The NCT should contain 8 lines, six lines containing the calibration constants C0, C1,
C2, C3, C4 and C5 in that order one constant per line. The two last lines are the
recorder s calibration decimal range values. First the maximum value and the minimum
value last.
Figure 12.9 shows a NCT file.
Figure 12.9 NCT file
When the NCT file is ready for use, place it in the following directory SeaStar\RBD.
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
Insert New Calibration Constants
Choose the Wizards menu and the Insert New Calibration Constants command to
enter the wizard (see figure 12.10).
Figure 12.10 Insert New Calibration Constants
The wizard is a five step operation:
1) Select Starmon mini as recorder type
2) Connect the cable to the Starmon mini recorder
3) Connect to the recorder and retrieve the RID file.
4) Having connected to a Starmon mini (like T6405) the wizard will look for a NCT file
with the same name, like T6406.NCT, in the SeaStar\RBD\ folder. Having found the
matching NCT file, the wizard will read the NCT file and merge it with the existing RID
and save as a new RBN file in the SeaStar\RBD\ folder. The RBN can be used later
as a backup file.
If this merger goes well the wizard shows the following message:
Figure 12.11
5) Finally the wizard transfers the RBN, with the new calibration data, to the Starmon
mini recorder. The wizard uses the Backup function in SeaStar and the user is
therefore prompted for a new sequence number and battery life value(%).
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
Figure 12.12 Set sequence number and battery life
is set too low, previous data can be overwritten. If the battery was replaced then set
battery life to 100%, otherwise to the current % that can be read in the RIT file.
Check the results
Having run the wizard, the user can check if the new calibration constants were
inserted correctly. First connect to the recorder again. The RCI file should now show
the new calibration constants and range values. Also the calibration date and number
should be updated.
Figure 12.13 RCI file
The values in the RCI can now be compared to the NCT file.
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
13. Group
Creating a group is a way to gather multiple measurement sequences which are part
of the same research project into one group. In a group the user can track when the
logger was started, the data retrieved and the data converted/reconverted. Groups
can be transferred to PatternFinder for further data analysis
Figure 13.1 Group menu
New Group
Select New Group to create a new group. Please note that it is also possible to
create a new group by clicking the New Group button in the group view pane.
The following window appears:
Figure 13.2 New Group
Name the group and enter a description. The group file will be saved under the
directory: C:\SeaStar\Groups. Each group has its own subfolder where data files from
each sequence in the group are stored, this way the data can be accessed from
PatternFinder and Gná.
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
Open Group
Select Open Group to open the Group directory. It is also possible to open the
directory by clicking the Open Group button in the group view pane.
Once a group has been selected the group view pane appears:
Figure 13.3 Group view
Adding sequences to group
To add a sequence to the group connect to a logger and define and start a new
measurement sequence. When starting the new sequence the following window
appears:
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
Set an alias for the sequence and enter a description. Click OK and the sequence will
be added to the group (see figure 13.5).
Figure 13.5 Group view pane
If a group is open when starting up a logger the new measurement sequence will be
automatically registered in the group. If no group is open the sequence will not be
registered in any group.
Retrieving data
When retrieving data from a logger and the sequence is already in a group the data
retrieval time is registered in the group. Also when the data is converted/reconverted a
copy of the *DAT file is stored in the group folder.
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
Appendix
A.1 Directories and Files
Directory Structure
The user can choose between SeaStar default data directory structure (path) or
define his own data path. As the SeaStar program is a multi recorder type program,
the directory structure branches from the directory where SeaStar is positioned down
Starmon mini
recorder directories, named after the serial numbers of the recorders, designed to
hold data for each individual recorder (data directory). All in all it is a three level
directory structure.
An example of this directory structure for Starmon mini is:
C:\Seastar\ The installation directory, SeaStar.exe location.
C:\Seastar\Starmon mini\ The type directory for Starmon mini.
C:\Seastar\Starmon mini\T0801\ The directory for the T0801 recorder.
When connecting to a recorder for the first time, SeaStar creates the recorder
directory, named after the serial number of the recorder.
Information Files
SeaStar uses information files to store and retrieve settings. There are two types of
information files:
The type-INI files, (for example StarmonT.ini for Starmon mini).
The information type files are all in the installation directory, they contain
information on specific settings for each type.
The type-INI files for Star-Oddi products are:
The recorder-INI files, for example T0801.ini contains settings information and data file
positioning for the T0801 recorder. The recorder-INI files are located in the
Files Locations
As stated previously, the type-SeaStar
recorder-INI files are placed in the recorder type directory. All other recorder and
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
measurement related files created by SeaStar, are placed in the appropriate
C:\SeaStar\Starmon mini\T0801\ is a directory for
the T0801 recorder and all files belonging to T0801 are placed there.
If the user whishes to use his own defined data path, for example C:\Data\, then all the
files for all the recorders, except for the INI files, are placed there. If the user has
on, individual
recorder directories, irrelative of type, are created under the user defined path, for
example: C:\DATA\T0801\ for the T0801 recorder. The T0801.ini file (located in the
C:\SeaStar\Starmon mini\) contains the path to the C:\DATA\T0801\ directory, and
all files related to the T0801 recorder will be placed in that particular directory.
Browsing for a Recorder
the default method is to find a recorder-INI file in the type directory. The INI file will
user defined directory, where a *.RID file must be selected.
Recorder Related Files
Every file for Starmon mini that is recorder related has the following name format: Txxxx
- The name is the recorder ID (serial no.)
*SNMS: Start New Measurement Sequence
The recorder ID (RIT) can at a later time be opened, and the related measurement
data can be viewed, printed and reconverted. The RDD, RDT and DIR are download
information files, created when a new measurement sequence is started.
Measurement Related Files
Every file that is specifically measurement related has the following name format:
mmmTxxxx - The name is the measurement number + serial no.
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
The MID, MIT and DAD files are created upon measurement data retrieval from the
recorder. The DAB, DAT and CIT files are created when converting the data. The
MDD and MDT are download information files and created when a new measurement
sequence is started.
Data Files
Upon data retrieval, three main data files are created.
The *.DAD File
This file is the raw data source file. The data is in mixed binary form. The user cannot
access this file in the SeaStar program. When converting these raw data into
measurements, the *.MID file is needed. The result from the conversion are two data
files, the *.DAB and the *.DAT.
The *.DAB File
This file contains sequential binary representation of the data in columns. The first
column is the measurement number. The other columns contain the measured
parameters in a binary form.
The *.DAT File
This is the actual result file, where the measurements are converted to their natural
units and timed. This is a text file with columns. The first column is the measurement
number, the second column the date and the time, depending on the set-up. The third
column is the time or the first measured parameter, depending on set-up. The
following column(s) contain the converted measured parameters with units and
number of decimals according to set-up. Number of parameters can range from 1-3,
and number of columns 3-6 accordingly.
Data Header
The *.DAT file contains a data header, which is a detailed description of how the
*.DAT file is configured. This feature is used by SeaStar when graphically
representing the data, and is practical if the user wants to import the data into other
applications or databases. Most of these descriptive items are derived from the
Each header item is contained in one line, and all header lines start with a #
(bookmark) and a number. Then follows a description of the header item, and then 1-4
directives, all separated by tabs. Eventually a comment trails the directives, preceded
by a semicolon (;).
The following is a description of the directives contained in the data header.
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
A.2 Help Menu
Choose Help and the following window appears:
Figure A.1 Help Menu
User Manual
Select User Manual to open the Starmon mini user manual.
Troubleshooting
Select Troubleshooting to open the Trouble-Shooter.
Release Notes
Select Release Notes to view a description of the latest features added to SeaStar,
the date when they were added and a reference to the chapter number of each
manual where the additions can be found.
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
Check for Updates
Select Check for Updates to manually update the latest version of SeaStar, User
Manual, Trouble-Shooter and Release Notes. The following window appears:
Figure A.2 Select files to download
Select the files you want to download and click OK. The program downloads and
installs the latest updates to the software. Please note that this can only be done
when the program is offline i.e. when no recorder is connected to the program.
Change Log is displayed if a new application version is available while updating (see
figure A.3)
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
By selecting Settings-Options-Update you can now choose whether you want to
update SeaStar manually, every X number of days or every time SeaStar is run.
Figure A.4 Update
Submit ticket
Tickets can now be sent directly from the application to our support department by
selecting Help-Submit Ticket. A shortcut key, F1, can also be used at any time while
in the application and is especially useful when the menu bar is disabled.
In the ticket form (see figure A.5) you can add your name and email. Select a category
that best describes your problem and write a subject and description. If you like you
can select Images (screenshots) and browse for images in the screenshots folder
and send them with the ticket. By selecting Other files you can also add files to the
ticket. To automatically add recorder data use the Include Recorder Data option. By
default the following files are included: screenshot, system information, logs and
settings. These can be excluded with options in Other files.
To receive a response we require that you include a valid email address.
Tickets can also be created by e-mailing support@star-oddi.com with the problem
description and files attached.
If you want to add more information to a previously created ticket you can reply to the
support response.
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524
Figure A.5 Submit ticket
About
Select About for general information about SeaStar.
A.3 Recorder Backup
When replacing the battery, new recorder sessions have to be transmitted to the
recorder. The user should contact Star-Oddi for a RBD file.The name of the backup
file is Txxxx.RBD (serial no. of the recorder plus the .RBD ending).
Downloading the backup file:
1. Copy the backup files (RBD) into the SeaStar RBD directory.
2. Open the SeaStar software.
3. Plug a recorder to the PC computer.
4. Select the 'File' menu and 'Recorder backup' command and the following window