Starlight Xpress SXVF-H35 User Manual

Handbook for the SXVF-H35 Issue 1 August 2007
Starlight Xpress Ltd
SXVF-H35
CCD camera user manual
36.1mm x 24.05mm active area. The use of high performance microlenses on the CCD surface gives the greatest possible throughput of light to the pixels and the resulting QE is very good over the entire visible spectrum. Our new ‘F’ type USB2 interface hardware gives an exceptionally fast download speed of about 2 megapixels per second, and so the SXVF-H35 can download a full resolution 16 bit image in only 8 seconds.
1
Handbook for the SXVF-H35 Issue 1 August 2007
Please take a few minutes to study the contents of this manual, which will help you to get the camera into operation quickly and without problems. I am sure that you want to see some results as soon as possible, so please move on to the ‘Quick Start’ section, which follows. A more detailed description of imaging techniques will be found in a later part of this manual.
‘Quick Starting’ your SXVF-H35 system
In the shipping container you will find the following items:
1) The SXVF-H35 camera head.
2) A power supply module and cable.
3) A 3 metre USB2 camera cable.
4) An adaptor/spacer for 72mm x 1mm thread ‘Takahashi’ focuser assemblies.
5) An adaptor for 2” drawtubes and M42 ‘Pentax’ thread lenses.
6) A guider cable for ‘ST4’ style mount guiding inputs.
7) A CD with the ‘SXVF_H35’ software and manual.
You will also need a PC computer with Windows XP or Windows Vista. This machine must have at least one USB2.0 port and at least 500 Megs of memory. If you intend to view the finished images on its screen, then you will also need a graphics card capable of displaying an image in a minimum of 1600 x 1200 pixels and 32 bit colour. A medium specification Pentium with between 1GHz and 3GHz processor speed is ideal. Please note that the SXVF-H35 is not designed for USB1.1 operation and will give inferior results if used on USB1.1.
Connecting up:
Plug the 5 pin DIN connector into the socket on the power supply box, and plug the power supply into the wall socket. The yellow LED on the power supply should light.
2
Handbook for the SXVF-H35 Issue 1 August 2007
A rear view, showing the input and output connectors
3
Handbook for the SXVF-H35 Issue 1 August 2007
Connect the miniature 4 way power plug to the socket on the rear of the camera and screw the retaining ring into place. The LED on the rear of the camera will light a dim yellow. The other connections should not be attached until after the software has been installed.
Installing the software:
Switch on the computer and allow it to ‘boot up’. Once you have the system ready to run, insert the program disk into your CD drive and select ‘Setup.exe’ if the disk does not autostart. The initial installation is to set up the USB drivers required by the SXVF electronics. The files SXVIO.sys and Generic.sys are copied to your Windows\System32\Drivers folder and SXVIO_H35.inf is copied to Windows\Inf. After this, the program ‘SXV_H35_usb.exe’ will be installed into your ‘CCD’ directory and a new directory called ‘Autosave’ will now exist on the same drive. ‘Autosave’ is where SXV_H35 will normally store its configuration file, ‘SXVH35.ini’, and any image files, which are recorded using the ‘Autosave’ mode in SXV_H35 and saved in FITs format.
Please note that the version of SXVIO.sys supplied with your H35, is an improved issue that should replace any copy that is already resident on your machine. Failure to update will usually result in a tendency for white spots and streaks to appear in your images.
4
Handbook for the SXVF-H35 Issue 1 August 2007
You now need to set up the camera control defaults (shown above), as follows: Start SXV-Hmf by clicking on the icon and select the ‘File’ menu. Now select ‘Set program defaults’ and a window, which contains the various software settings, will appear. Suggested starting defaults are as follows:
1) Background Image area Red (or as preferred)
2) FITS Unsigned Integer format Off
3) Star mask size (area used for photometry and guiding) 8 pixels
4) Telescope guiding output to ‘autoguider socket’
The other default settings are not important for current purposes and may be left as the software start-up values for now.
Recording your first image:
We now have the camera and computer set up to take pictures, but an optical system is needed to project an image onto the CCD surface. You could use your telescope, but this introduces additional complications, which are best avoided at this early stage. There are two simple options, one of which is available to everyone:
1) Attach a standard ‘M42’ SLR camera lens to the SXVF-H35, using the 27mm
spacer/adaptor to achieve approximately the correct focal distance.
2) Create a ‘Pin hole’ lens by sticking a sheet of aluminium baking foil over the end
of the lens adaptor and pricking its centre with a small pin.
5
Handbook for the SXVF-H35 Issue 1 August 2007
If you use a normal lens, then stop it down to the smallest aperture number possible (usually F22) as this will minimise focus problems and keep the light level reasonable for daytime testing. The pin hole needs no such adjustments and will work immediately, although somewhat fuzzily!
Point the camera + lens or pinhole, towards a well-lit and clearly defined object some distance away. Now click on the camera icon in the toolbar of the SXV-H35 software and the camera control panel will appear (see above). Select an exposure time of 0.1 seconds and press ‘Take photo’. After the exposure and download have completed (between 8 and 10 seconds) an image of some kind will appear on the computer monitor. It will probably be poorly focused and incorrectly exposed, but any sort of image is better than none! In the case of the pinhole, all that you can experiment with is the exposure time, but a camera lens can be adjusted for good focus and so you might want to try this to judge the image quality that it is possible to achieve.
One potential problem with taking daylight images is the strong infrared response of the SXVF-H35 as this will cause ‘soft focus’ with camera lenses. Soft focus is much reduced by keeping the aperture setting below F8. Also, IR blocking filters are available from various suppliers (True Technology, Edmunds etc.) and are recommended for the best results when using a lens.
If you cannot record any kind of image, please check the following points:
1) Is the power LED on?
2) Does the software indicate that the camera is successfully connected? An attempt to take a picture will fail with an error message if the USB is not properly installed. In this case, try unplugging the USB cable and then reconnecting it after about 5
6
Handbook for the SXVF-H35 Issue 1 August 2007
seconds. Restart the camera software and see if it can link now. If not, check in Windows device manager (via ‘System’ in ‘Control Panel’) and see if the BlockIOClass device is installed properly. If all looks OK, try checking the ‘Disable VID/PID detection’ in the ‘Set program defaults’ menu and try again.
3) If you cannot find any way of making the camera work, please try using it with another computer. This will confirm that the camera is OK, or otherwise, and you can then decide how to proceed. Also check on our web site to see if there are any updates or information about your camera software that might help. The message board might prove useful to ask for help with getting your camera operating properly.
Our guarantee ensures that any electrical faults are corrected quickly and at no cost to the customer.
Enhancing your image:
Your first image may now be reasonably good, but it is unlikely to be as clear and sharp as it could be. Improved focusing and exposure selection may correct these shortcomings, and you may like to try them before applying any image enhancement with the software. However, there will come a point when you say, ‘That’s the best that I can get’ and you will want to experiment with various filters and contrast operations. In the case of daylight images, the processing options are many, but there are few that will improve the picture in a useful way.
The most useful of these are the ‘Normal Contrast Stretch’ and the ‘High Pass Low Power’ filter. The high pass filter gives a moderate improvement in the image sharpness, and this can be very effective on daylight images. Too much high pass filtering results in dark borders around well-defined features and will increase the ‘noise’ in an image to unacceptable levels, but the ‘Low Power’ filter is close to optimum and gives a nicely sharpened picture.
The ‘Contrast’ routines are used to brighten (or dull) the image highlights and shadows. A ‘Normal’ stretch is a simple linear operation, where two pointers (the ‘black’ and ‘white’ limits) can be set at either side of the image histogram and used to define new start and end points. The image data is then mathematically modified so that any pixels that are to the left of the ‘black’ pointer are set to black and any pixels to the right of the ‘white’ pointer are set to white. The pixels with values between the pointers are modified to fit the new brightness distribution. Try experimenting with the pointer positions until the image has a pleasing brightness and ‘crispness’.
At this point, you will have a working knowledge of how to take and process an SXVF-H35 image. It is time to move on to astronomical imaging, which has its own, unique, set of problems!
*********************************************************************
7
Handbook for the SXVF-H35 Issue 1 August 2007
Astronomical Imaging with the SXVF-H35
1) Getting the image onto the CCD:
It is essential to set up a good optical match between your H35 and your telescope. The H35 has a very large CCD area and so many of the popular ‘SCT’ ‘scopes are unable to provide good quality star images over the large chip. Because of this limitation, the H35 was designed for use with a wide field highly corrected refractor, such as the Takahashi FSQ106 or similar, and is supplied with an M72 spacer ring to screw into the FSQ106 focus assembly. Here is a view of an SXVF-H35 attached to the author’s FSQ, via the adaptor ring.
As a general guide, most CCD astronomers try to maintain an image scale of about 2 arc seconds per pixel for deep sky images. This matches the telescope resolution to the CCD resolution and avoids ‘undersampling’ the image, which can result in square stars and other unwanted effects. To calculate the optimum focal length required for this condition to exist, you can use the following simple equation:
F = Pixel size * 205920 / Resolution (in arc seconds)
In the case of the SXVF-H35 and a 2 arc seconds per pixel resolution, we get
F = 0.009 * 205920 / 2 = 927mm
8