Standard Electric Works 6500 LC Instruction Manual

LOOP CALIBRATOR
INSTRUCTION MANUAL
Distributed by: Mitchell Instrument Co. Inc.
Call Toll Free (888) 270 2690 - www.mitchellinstrument.com
Table of Contents
mA Output Selection............................A.
Percentage(%) Input............................B.
Adjusting the Percentage Up andC.
Down....................................................
Auto Step Function..............................D.
Auto Ramp Function............................E.
0~24V Voltage Output..........................F.
Voltage Output(0-24V)........................G.
Replacing the Battery.............................10.
1 1 1 2 3 4
5-7
8-9
10-11
12-13 14-15 16-17
18 19 20 20
-1-
Introduction11
This tester has been designed and tested in accordance with the CE safety requirements for electronic measuring apparatus EN61326-1 and other safety standards. Follow all warnings to ensure safe operation.
Safety Notes21
Read the following safety information carefully before attempting to operate or service the meter. Use the meter only as specied in this manual. Otherwise, the protection provided by the meter may be impaired. Rated enviromental conditions: (1) Indoor use. (2) Installation Category I (3) Pollution Degree 2. (4) Altitude up to 2000 meters. (5) Relative humidity 80% max. (6) Ambient temperature 0~40℃.
Features:31
(1) 4-20mA (1KΩ load, 24V Loop Supply)
(2) 0.025% Basic Accuracy (3) Simple Operation Interface (4) Auto Ramp and Step Functions (5) 0-20mA, 0-24mA selectable (6) Incremental percentage setting: 0-100% (7) Warning beeper when output is open (8) 0-24 V output
-2-
Specication (2541
±5
≤85%R.H.)
DC CURRENT (1kΩ Max Load, 24v Loop Supply)
RANGE RESOLUTION ACCURACY
0-4mA 1uA ±0.025%±10uA
4-20mA 1uA ±0.025%±5uA
20-24mA 1uA ±0.025%±5uA
DC VOLTAGE
RANGE
RESOLUTION ACCURACY
0-4V
1mV ±0.05%±10mV
4-20V
1mV ±0.05%±5mV
20-24V
1mV ±0.05%±5mV
GENERAL
Power Consumption 150mA at 24V 1kΩ load Operating Temperature 0º C - 50º C Operating Humidity ≤80% R.H. Storage Temperature -10º C - 60º C Storage Humidity ≤85% R.H. Power Source 9V x 1 alkaline battery Dimensions 160 (L) x 100 (W) x 38 (D) mm Weight Approx. 350g (battery includ-
ed)
Accessories DC adapter, pouch, battery,
instruction manual
-3-
Panel Description51
1.LCD Display 6.ENTER Button
2.ON/OFF Button 7.Numerical & Function Keypad
3.Shift Button 8.DC Adaptor Input Socket
4.Voltage Button 9.Stand
5.Current Button 10.Output Terminal
4-20mA LOOP CALIBRATOR
1 2 3
4-20mA
4
7 8
0
9
V
I
START
5
END
%
%
-
+
6
SEC.
ENTER
Shift
0-20mA 0-24mA
ON
OFF
%
1.
7.
6.
2.
8.
9.
3.
4.
5.
10.
-4-
Display61
1.mVA: Units
2.0-20mA: Output Range of mA
3.%: Percentage
4. : Ramp
5. : STEP
6.OL: Overload, output open
7.SHIFT:SHIFT functions
8. : low Battery
1.
2. 3.
4.
5.
6.
7.8.
-5-
Functions71
Power ON/OFFa.
Press the SHIFT button and select the desired mA b.
range by pressing one of these three buttons.
Press the SHIFT button then press START (#4) button c.
to begin Auto Ramp Function.
Press the SHIFT button then press the END (#5) d.
button to stop Auto Ramp Function.
1 2 3
4-20mA
4
7 8
0
9
V
I
START
5
END
%
%
6
SEC.
ENTER
Shift
0-20mA 0-24mA
ON
OFF
%
-6-
Press the SHIFT button then press the SEC. (#6) e. button to enter the duration in seconds for Auto Ramp Function.
Press the SHIFT button then press this button (#7) f. to begin Auto Ramp Function. Press again to end Auto Ramp Function.
Press the SHIFT button then press the % button to g. select the percentage range.
After the percentage range has been entered,use
these / buttons to increase or decrease the
percentage incrementally.
Press the SHIFT button then press this button (#9) to h. start the STEP function. Press again to stop the function.