SSI BINARY SOLVENT DELIVERY MODULE 90-2581 REV B, BINARY SOLVENT DELIVERY MODULE Operator's Manual

Binary Solvent Delivery Module
Dual Series-III Pumps with Serial PC and Voltage Control
Operator's Manual
90-2581 rev B
Scientific Systems, Inc. 349 N. Science Park Road State College,
6803
yp
V
V
y
1. INTRODUCTION
This operator's manual contains information needed to install, operate, and perform minor maintenance on the Binary Solvent Delivery Module. The figures below are for reference of items described in the manual.
Power
Switch
“T” With
Outlet
Left
Drawer
Control
ads
Ke
Prime Purge
alves
Cooling
Fan
Outlet Filters
Right
Drawer
Power
Entr
Auxiliary I/O
oltage Control, etc.
1-1
RS-232
Serial Ports
1.1 Description of the Binary Solvent Delivery Module
The system consists of two Series III high performance metering pumps, in a compact package with pulse dampener and an off-line pressure transducer. The pumps use a drawer system for easy access to internal components for troubleshooting & service. These features allow for:
f Fast & easy setup:
Two inlet connection (from solvent reservoirs) One outlet connection (to injector)
All connections accessible from front panel
f Modular:
Can be added to any system
f All Stainless steel or PEEK Fluid path, including pumps, valves
and fittings
f Modular pump bays for easy replacement and maintenance
f Self flushing pump heads for extended seal life and reduced
maintenance
f Pulse Dampener for reduced pulsation.
f Very high performance/price ratio
f Easily user adjustable process set points (flow rate, pressure. etc.)
via front key pads f Digital readout of process parameters f User settable upper/lower pressure and temperature limits f RS-232 serial PC interface f Compact size—Requires only 11 inches of bench space
The low pulsation flow produced by the reciprocating, single­piston pump is achieved by using an advanced rapid-refill cam design, programmed stepper motor acceleration, and an internal pulse damper.
1.1.1 Pump Features
They include:
Rapid refill mechanism to reduce pulsation
Stainless Steel or PEEK™ pumps head
1-2
LED front panel readout of flow rate, pressure and upper/lower
pressure limits
Flow adjustment in 0.001 ml increments, from 0.001 to 5.000
ml/min with a precision of 0.2% RSD (5mL/min heads)
Flow adjustment in 0.01 ml increments, from 0.01 to 10.00
ml/min with a precision of 0.2% RSD (10mL/min heads)
Microprocessor advanced control
Digital stepper motor design to prevent flow rate drift
over time and temperature
Back panel RS232 serial communications port for
complete control and status monitoring
Remote analog inputs (e.g. voltage) to control flow rate
1.1.2 Wetted Pump Materials
Pump heads, check valve bodies, and tubing are made out Stainless Steel or PEEK™. Other materials are synthetic ruby and sapphire (check valve internals and piston).
1.1.3 Self-Flushing Pump Head
Self-flushing pump heads provide continuous washing of the piston surface without the inconvenience of a manual flush or gravity feed arrangement. The self-flushing pump head uses a diaphragm and secondary set of check valves to create a continuous and positive flow in the area behind the high pressure pump seal. The flushing solution washes away any buffer salts that have precipitated onto the piston. If not removed, these precipitates can abrade the high pressure seal and cause premature seal failure, leakage, and can possibly damage the pump.
MOBILE
PHASE
SAPPHIRE
PISTON
1-3
FLOW OF
FLUSH SOLUTION
HIGH-PRESSURE SEAL
PRIMARY
SECONDARY
SELF-FLUSH SEAL
Figure 1-1. Self-Flushing Pump Head
PISTON MOVEMENT
1.1.4 Self-Flush and Seal Life
It is recommended that the Self Flush feature be used to improve seal life in a number of applications. In particular, (as stated above) if pumping Buffers, Acids/Bases or any inorganic solution near saturation, the pump should utilize the Self Flush feature. With every piston stroke, an extremely thin film of solution is pulled back past the seal. If this zone is dry (without use of Self Flush), then crystals will form with continuous operation, which will ultimately damage the seal.
Another application where Self Flush is highly recommended is when pumping Tetrahydrofuran (a.k.a. THR, Diethylene Oxide) or other volatile solvents such as acetone (Note: THF and most solvents are compatible only with all-Stainless Steel systems. THF will attack PEEK). Volatile solvents will dry rapidly behind the seal (without the use of Self Flush), which will dry and degrade the seal.
IPA, Methanol, 20% IPA/water mix or 20% Methanol/water mix are good choices for the flush solution. Consult the factory for specific recommendations.
1-4
1.2 Specifications for the Binary Solvent Delivery Modules
Flow Rates ............0.000 to 5.000 mL/min for 5mL/min heads
0.00 to 10.00 mL/min for 10mL/min heads
Pressure ................ 0 to 6,000 p.s.i. for SS pump heads,
0 to 5,000 p.s.i. for PEEK™ heads,
Pressure Accuracy… .± 1% of full-scale pressure Pressure Zero Offset. .± 2 p.s.i.
Flow Accuracy........± 2% for a flow rate of 0.20 mL/min and
above.
Flow Precision .......0.2% RSD
Dimensions ............9" high x 10.5" wide x 18.5" deep
Weight ...................34 lb
Power .................... 90-260 VAC, 50-60 Hz, 45W (The main
voltage supply shall not exceed ±10%) Environmental……..Indoor use only Altitude……………..2000 M Temperature………10 to 30° C Humidity……………20 to 90% Relative humidity
Remote Inputs . . . . RS-232, Voltage, Frequency
1-5
2. INSTALLATION
2.1 Unpacking and Inspection
Prior to opening the shipping container, inspect it for damage or evidence of mishandling. If it has been damaged or mishandled, notify the carrier before opening the container. Once the container is opened, inspect the contents for damage. Any damage should be reported to the carrier immediately. Save the shipping container. Check the contents against the packing list.
2.2 Location/Environment
The preferred environment for the Binary Solvent Delivery Module is normal laboratory conditions. The area should be clean and have a stable temperature and humidity. The specific temperature and humidity conditions are 10 to 30 °C and 20% to 90% relative humidity. The instrument should be located on a stable flat surface with surrounding space for ventilation and the necessary electrical and fluid connections.
2.3 Fluid Connections & Priming There are only four connections to be made.
1. Connect left drawer pump solvent line supplied
2. Connect right drawer pump solvent line supplied
3. Connect line to injection valve from outlet “T”
4. Next, install the self flush. Connect the self flush inlet and outlet (opaque) tubing as shown on the the next page.
Details on the proper installation of tubing and priming of the pump and self flush is shown on the next page.
3-1
SELF-FLUSH
PUMP
Connect wash solution inlet and outlet tubing (opaque) to the flush housing as
shown.
Make sure ferrule is in the correct position.
Screw syringe on luer fitting. Insert wash outlet tubing end into clear adapter.
Connect pump inlet tubing as shown.
Make sure ferrule is in the correct position.
Ensure inlet line filter is submersed into solvent.
Attach syringe to Prime-Purge valve.
Open Prime-Purge valve by turning knob counterclockwise two turns.
Draw syringe back to prime. Draw on syringe until no
3-2
2.4 Electrical Connection
The system utilizes Universal Switching Power Supply, and will accept voltages from 90 – 260 VAC, 50-60 Hz.
WARNING: Do not bypass the safety ground connection as a serious shock hazard could result.
2.5 Solvent Preparation
Proper solvent preparation will prevent a great number of pumping problems. The most common problem is bubble formation, which may affect the flow rate consistency. Aside from leaky fittings, the problem of bubble formation arises from two sources: solvent out-gassing and cavitation. Filtration of HPLC solvents is also required.
2.5.1 Solvent Out-gassing and Sparging
Solvent out-gassing occurs because the mobile phase contains dissolved atmospheric gases, primarily N2 and O2. These dissolved gases may lead to bubble formation and should be removed by degassing the mobile phase before or during use. The best practical technique for degassing is to sparge the solvent with standard laboratory grade (99.9+%) helium. Helium is only sparingly soluble in HPLC solvents, so other gases dissolved in the solvent diffuse into the helium bubbles and are swept from the system. Solvent filtration is not an effective alternative to helium degassing.
It is recommended that you sparge the solvent vigorously for 10 to 15 minutes before using it. Then maintain a trickle sparge during use to keep atmospheric gases from dissolving back into the mobile phase. The sparged solvent must be continually blanketed with helium at 2 to 3 psi. Non- blanketed, sparged solvents will allow atmospheric gases to dissolve back into the mobile phase within four hours.
Solvent mixtures using water and organic solvents (like methanol or acetonitrile) hold less dissolved gas than pure solvents. Sparging to reduce the amount of dissolved gas is therefore particularly important when utilizing solvent mixture.
Even with sparging some out-gassing may be occur. A back pressure regulator installed after the detector flow cell will help prevent bubbles from forming and thus limit baseline noise.
WARNING: Always release pressure from the pump slowly. A rapid pressure release could cause the pulse damper diaphragm to rupture.
3-3
2.5.2 Cavitation
Cavitation occurs when inlet conditions restrict the flow of solvent and vapor bubbles are formed during the inlet stroke. The key to preventing cavitation is to reduce inlet restrictions. The most common causes of inlet restrictions are crimped inlet lines and plugged inlet filters. Inlet lines with tubing longer than 48" (120 cm) or with tubing of less than 0.085" (2 mm) ID may also cause cavitation.
Placing the solvent reservoirs below the pump level also promotes cavitation. The optimal location of the reservoirs is slightly above the pump level, but it is adequate to have them on the same level as the pump.
2.5.3 Filtration
Solvent filtration is good practice for the reliability of the Binary Solvent Delivery Module and other components in a HPLC system. Solvents should always be filtered with a 0.5 micron filter prior to use. This ensures that no particles will interfere with the reliable operation of the piston seals and check valves. Solvents in which buffers or other salts readily precipitate out will need to be filtered more often. After filtration, the solvents should be stored in a closed, particulate-free bottle.
2.5.4 Initial system pressurization (Daily)
IMPORTANT: To maximize accuracy at all pressures and flows this pump contains a pulse damper for each solvent used. It is important that the pump be brought to pressure with both pumps running to insure that each pulse damper has the correct solvent in it. It is recommended that the flow be set to the operational flow with an equal mixture of the two solvents. Once the system has reached a constant pressure the composition should be set to the initial conditions.
2.6 Instrument Installation
2.6.1 Mobile Phase Reservoirs
The mobile phase reservoir should be placed at the same level or slightly higher than the pump, never below the pump, and the inlet tubing should be as short as practical. These steps minimize pressure losses on the inlet side of the pump during refill and help to avoid bubble formation. These steps are particularly important when using high vapor pressure solvents (hexane, methylene chloride, etc.). Mobile phases should be degassed, filtered and covered. (See Section
2.4.)
3-4
2.6.2 Self-Flush Solution
Self-flush heads require 250-500 mL of flushing solution. See section 1.1.4 for self-flush solution recommendations. A pH indicator that will indicate the concentration of salts in the solution is recommended as a reminder to change the solution. This flush solution should be replaced with a fresh solution weekly to avoid frequent pump maintenance.
2.6.3 Inlet Tubing and Filters
The table below shows the inlet tubing and filter used in the Binary Solvent Delivery Module. All inlet lines are supplied in a 30" (76 cm) length and are made of a fluoropolymer material.
2.6.4 Priming the Pump and the Flushing Lines
Connect a syringe to the outlet tubing. Run the pump at a flowrate of 3 to 5 ml/min. Prime the pump by pulling mobile phase and any air bubbles through the system and into the syringe (a minimum of 20 ml).
To prime the flush lines for a self-flush head, simply place the inlet line in the flush solution and connect a syringe to the outlet line and apply suction until the line is filled with flush solution. Place the outlet line in the flush solution. Secure both flush lines in the flush solution container so they stay immersed during pump operation.
2.6.6 Long Term Pressure Calibration Accuracy
This note applies if your pump is equipped with an electronic pressure transducer. The transducer has been zeroed and calibrated at the factory. Over the life of the pump, some drift may occur. For example, it is typical for the zero to drift < 10 p.s.i. after about 1 year of operation (i.e., with no back pressure on the pump a reading of 1-9 p.s.i. may be displayed). A similar drift may also occur at higher pressures, and are typically less than 1% (e.g. <50 p.s.i. at 6,000 p.s.i. back pressure).
If pressure calibration and/or drift is a concern, consult the factory. The pump can be shipped back to SSI for recalibration. Alternatively, written calibration and zero-reset procedures are available. Consult the factory to receive these instructions.
3-5
2.7 Preparation for Storage or Shipping
2.7.1 Isopropanol Flush
Disconnect the outlet tubing from the pump. Place the inlet filter in isopropanol. Use a syringe to draw a minimum of 50 ml through the pump. Pump a minimum of 5 ml of isopropanol to exit. Leave the inlet tubing connected to the pump. Place the inlet filter in a small plastic bag and attach it to the tubing with a rubber band. Plug the outlet port with the shipping plug or leave a length of outlet tubing on the pump or cover the outlet port with plastic film.
2.7.2 Packaging for Shipping
CAUTION: Reship in the original carton, if possible. If the original carton is not available, wrap the pump in several layers of bubble wrap and cushion the bottom, top, and all four sides with 2" of packaging foam. Although heavy, an HPLC pump is a delicate instrument and must be carefully packaged to withstand the shocks and vibration of shipment.
3-6
3. OPERATION
3.1 Pump Front Panel Controls and Indicators
3.1.1 Control Panel
3.1.1.1 Digital Display
The 3-digit display shows the pump flow rate (mL/min), system pressure (psi), or the set upper or lower pressure limit (psi) when operating. Choice of display is selected with the MODE key.
3.1.1.2 Digital Display Pump Keypad
MODE
RUN
STOP
Fast And Slow Button Repeat On The Up And Down Arrow Buttons: If the UP-ARROW or DOWN-ARROW button is held down for more than approximately one half of a second, the button press will repeat at a slow rate of approximately 10 times a second. Once slow button repeat has begun, fast button repeat can be initiated by using a second finger to press down the second arrow button. During fast button repeat, the button press will repeat at a rate of approximately 100 times a second. Switching back and forth between repeat speeds can be accomplished by pressing and releasing the second arrow button while keeping the first arrow button held down.
When pressed, this button increases the flow rate.
When pressed, this button decreases the flow rate.
Use this button to cycle through the four display modes: flow rate, pressure, upper pressure limit, or lower pressure limit. A status LED to the right of the digital display indicates which mode is active.
When pressed, this button alternately starts and stops the pump.
3-7
3.1.1.3 Status LEDs
Q ml/min
Q PSI
Q HI PR
Q LO PR
Q RUN
Q FAULT
When lit, the digital display shows flow rate in mL/min.
When lit, the digital display shows system pressure in psi.
When lit, the display shows the user-set upper pressure limit in psi.
When lit, the display shows the user-set lower pressure limit in psi.
Lights to indicate that the pump is running.
Lights when a fault occurs and stops the pump.
3.1.1.2 Power-up Configuration
Non-volatile Memory Reset: If the pump is operating erratically, there is the possibility that the memory has been corrupted. To reset the memory and restore the pump to it's default parameters, press and hold the UP-ARROW button when the power is switched on. Release the button when the display reads "rES". The parameters stored in non-volatile memory, i.e., the flowrate, the pressure compensation, the voltage/frequency select, the lower pressure limit, and the upper pressure limit will be set to the factory default values. The head type setting is the only parameter not changed by the non-volatile memory reset function. If the firmware is upgraded to a newer version, a non­volatile memory reset will automatically occur the first time the power is switched on.
3.1.1.5 Power-Up Tests
Display Software Version Mode: The software version can be displayed during power-up by pressing and holding the RUN/STOP and the UP-ARROW buttons when the power is switched on. Release the buttons when the display reads "UEr". The decimal point number displayed on the display is the software version. To exit this mode, press the RUN/STOP button.
3-8
Loading...
+ 33 hidden pages