SRS Labs SR400 User Manual

MODEL SR400
GATED PHOTON COUNTER
Phone: (408) 744-9040 • Fax: (408) 744-9049
email: info@thinkSRS.com • www.thinkSRS.com
1290-D Reamwood Avenue
Sunnyvale, California 94089
All Rights Reserved.
Revision 2.7 (11/2018)
Table of Contents
TABLE OF CONTENTS
CONDENSED INFORMATION
Safety and Preparation for Use 3 Specifications 5 Menu Display List 7 Abridged Command List 9
INSTRUMENT OVERVIEW
Signal Inputs / Discriminators 11 Counting 12 Gating 13 Count Modes 14 Outputs 14 Measurement Examples 17
OPERATION
Front Panel Summary 21 Power 21 Liquid Crystal Display 21 Menu Keys 21 Spin Knob 21 Start, Stop, and Reset 21 Instrument Status 22 Counter LED's 22 Inputs 22 Outputs 22 Rear Panel Summary 24 Power Entry Module 24 IEEE-488 Port 24 RS-232 Connector 24 Input 24 Output 24 Front Panel Menus 25 Mode Menu 25 Level Menu 27 Gate Menus 30 Setup Menu 31 Com Menu 31
INSTRUMENT SETUPS
Default Setup / Power On Clear 33 Quick Check 33 2 Photon Pile Up Correction 34 Gated Photon Counting 34 Synchronous or Chopped Photon Counting 34
REMOTE PROGRAMMING
Communications 37
Command Syntax 37 Status LED's 38 RS-232 Echo 38 Try-out with ASCII Terminal 38 Detailed Command List 39 Mode 39 Levels 40 Gates 41 Front Panel 42 Interface 44 Store/Recall 45 Polled Data 45 Continuous Data 45 Examine Data 46 Status Byte 46 Secondary Status Byte 47 Errors / Data Window 47 Trouble Shooting Interface Problems 47 Common Hardware Problems 47 Common Software Problems 48
RS-232 INTERFACE
Data Communications Equipment 49 Echo 49 Wait Command 49 Termination Sequences 49 Example Programs 49 Microsoft BASIC 50 Microsoft FORTRAN 51
GPIB (IEEE-488) INTERFACE
GPIB Capabilities 53 Serial Polls and Service Requests 53 Example Programs 53 Microsoft BASIC 54 Microsoft C 56
TROUBLESHOOTING
LCD Contrast 59 Fan 59 Front Panel Test 59 Power On Reset 59 Quick Test 59 Counting 60
CALIBRATION AND REPAIR
Signal Input Offset 61 Signal Input Repair 61
1
Table of Contents
USING PHOTOMULTIPLIER TUBES
PMT Selection for Photon Counting 63 Geometry 63 Spectral Response 63 Gain and Risetime 63 Dark Counts 64 PMT Base Design 64 Dynode Biasing 64 Snubbing 66 PMT Base Conclusions 67 Cathode Shielding 67 Preamplifiers 67 Gain Requirement 67 Plateauing a PMT 67 The 'Correct' Way 68 The 'Fast and Pretty Good' Way 68
TYPICAL EXPERIMENT
Energy Levels of Ruby 69 Apparatus 69 Operation 69 Instrument Configuration 70 Computer Data Acquisition 70
CIRCUIT DESCRIPTION
Microprocessor Control 73 Microprocessor System 73 RS-232 Interface 73 GPIB Interfac e 73 Input Ports 73 Slow Counters 73 Output Ports 73 Front Panel 73 D/A Converter 74 Fast Counters 74 Signal Inputs 74
Discriminators 74 A and B Counters 74 T Counter 75 Start / Stop and Dwell 75 Gate Generators 76 Digital Delay 76 Analog Delay 76 Delay Reset 77 Gates 77 Power Supply 77
PARTS LISTS
Main Assembly 79 Power Supply 93 Miscellaneous 95
SCHEMATIC DIAGRAMS
Signal Amplifiers SR400-1 Discriminators SR400-2 Fast Counters A and B Prescallers SR400-3 Fast Counters T Prescaller SR400-4 Slow Counters (A, B, T) SR400-5 Dwell Timer: Start/Stop SR400-6 Fast Gate Delay Ramps SR400-7 Slow Gate Delay Counters SR400-8 Fast Gate Ramps SR400-9 Width Oscillators SR400-10 Slow Gate Width Counters SR400-11 Microprocessor System SR400-12 Counters and Timers: Digital Interface 1 SR400-13 Output Ports: Digital Interface 2 SR400-14 Digital to Analog Converter SR400-15 Power Supply
2
Safety and Preparation for Use
SAFETY AND PREPARATION FOR USE
**** CAUTION ****
This instrument may be damaged if operated with the LINE VOLTAGE SELECTOR set for the wrong ac line voltage or if the wrong fuse is installed.
LINE VOLTAGE SELECTION
The SR400 operates from a 100V, 120V, 220V, or 240V nominal ac power source having a line frequency of 50 or 60 Hz. Before connecting the power cord to a power source, verify that the LINE VOLTAGE SELECTOR card, located in the rear panel fuse holder, is set so that the correct ac input voltage value is visible.
Conversion to other ac input voltages requires a change in the fuse holder voltage card position and fuse value. Disconnect the power cord, open the fuse holder cover door and rotate the fuse-pull lever to remove the fuse. Remove the small printed circuit board and select the operating voltage by orienting the printed circuit board to position the desired voltage to be visible when pushed firmly into its slot. Rotate the fuse-pull lever back into its normal position and insert the correct fuse into the fuse holder.
LINE FUSE
Use 1 Amp slow blow fuse for 100V/120V as well as 220V/240V.
LINE CORD
The SR400 has a detachable, three-wire power cord for connection to the power source and to a protective ground. The exposed metal parts of the instrument are connected to the outlet ground to protect against electrical shock. Always use an outlet which has a properly connected protective ground.
WARNING REGARDING USE WITH PHOTOMULTIPLIERS
The signal inputs may be damaged if a photomultiplier is used improperly with the photon counter. When left completely unterminated, a PMT will charge a cable to a few hundred Volts in a very short time. If this cable is connected to a signal input, the stored charge may damage the front-end transistors. To avoid this problem,
provide a leakage path of about 100 KΩ to ground
inside the base of the PMT to prevent charge accumulation.
CONNECTION TO OTHER INSTRUMENTS
All front panel BNC shields are connected to the chassis ground and to the power outlet ground via the power cord. Do not apply any voltage to either the shields or to the outputs. The outputs are not protected against connection to any potential other than ground.
3
SRS Symbols
Symbols that may be found on SRS products
4
SR400 Specifications
SR400 SPECIFICATIONS
OVERVIEW
There are three complete counters identified as A, B, and T. All counters count up to 200 MHz.
Counter A can count INPUT 1 or the 10 MHz internal clock. Counter A is gated by AGATE, and has a count capacity of 10
9
.
Counter B can count INPUT 1 or INPUT 2. Counter B is gated by BGATE, has a count capacity of 10
9
and may be used as the preset
counter to determine the count interval.
Counter T can count the internal 10 MHz clock, INPUT 2, or the gate trigger. This counter may be preset to determine the count period for up to 25 hours or 9 x 10
11
counts, gates, or triggers. The
timer has an accuracy or 25 ppm from 0 to 50∞C.
SIGNAL INPUTS
Bandwidth: dc to 300 MHz Input impedance: 50 Ohms Linear range: ± 300 mV (at input) Input protection: ± 5 Vdc, 50 V for 1 µs Overload recovery: 5 ns for <10 µs duration
overload
DISCRIMINATORS
Counters A, B, and T have independent discriminators when counting the signal inputs. All discriminator levels may be set to a fixed level or scanned. A rear panel INHIBIT input (TTL active high) can inhibit the discriminators to stop the count.
Referenced to the signal inputs: Discriminator range: -300 mV to +300 mV. Discriminator slope: Rising or Falling Resolution: 0.2 mV Input offset voltage: 1 mV Minimum pulse input: 10 mV Pulse pair resolution: 5 ns. DISC outputs: NIM levels into 50 Ohms
TRIGGER INPUT
Impedance: 10 KOhms Threshold: ±2.000 Vdc in 1 mV steps Slope: Rising or Falling
Protection: 15 Vdc, 100 V for 1 µs
GATE GENERATORS
There are two independent gates, AGATE and BGATE, which enable the A and B counters. Both gates are triggered by the TRIGGER input. The gates may be fixed in time or scanned. The GATE outputs show the positions of the gates with respect to the discriminator outpu ts.
Insertion delay: 25 ns Maximum delay: 999.2 ms Minimum gate width: 5 ns Maximum gate width: 999.2 ms or CW Resolution: 0.1%, 1 ns minimum Accuracy: 2 ns +1% Jitter: 200 ps rms +100 ppm Maximum trigger rate:1 MHz GATE view outputs: NIM levels into 50 Ohms GATE view error: < 2 ns
COUNTING MODES
A and B for a period determined by T preSET and T input. A ± B for a period determined by T preSET and T input. A for a period determined by B preSET and B input.
SCAN AND DWELL
The number of count PERIODS or data points in a scan may be set from 1 to 2000. The duration of one count period is determined by the preset condition.
The time between consecutive count periods is the DWELL TIME and can be set from 2 ms to 60 s. The DWELL output will be TTL high during the DWELL time. This output can be used to trigger external devices.
At the end of a scan (of 1 to 2000 count periods) counting may be programmed to STOP or START the scan over again.
The START key begins the first count period of the programmed scan.
5
SR400 Specifications
The STOP key terminates the current count period and pauses the scan. If scanning, gates and disc levels are held at their last value. The STOP key pressed while in a paused condition will reset the scan and all scanned parameters will return to their start values. The START key pressed while paused resumes the scan by starting the next count period.
The DWELL TIME may also be set to EXTERNAL. In this mode, count periods begin with the START key or EXTERNAL START INPUT (TTL rising edge). Count periods terminate with the preset condition, the STOP key, or the EXTERNAL STOP INPUT (TTL rising edge). A STOP key while not counting resets the scan. This allows completely variable dwell times and/or counting periods.
All count data is internally buffered for one scan. Data may be read over the computer interfaces during or after a scan.
DISPLAY MODE
CONTINUOUS: Displays current counter value HOLD: Displays final count value
D/A OUTPUT
The front panel D/A Output is proportional to A, B, A-B, or A+B depending upon the counting mode selected. The D/A output is updated at the end of each count period.
There are two rear panel D/A outputs, PORT 1 and PORT 2. These outputs may be set or scanned from the front panel or via the computer interface.
Full scale: ± 10 Vdc Resolution: 12 bits (5 mV) Current rating: 10 mA Output Impedance: < 1 Ohm Accuracy: 0.1% + 5 mV
COMPUTER INTERFACE
IEEE-488 (GPIB) and RS-232 (up to 19.2 kbaud). Full instrument control and data transmission.
GENERAL
Dimensions: 16" x 13" x 3.5" Weight: 10 lbs Power: 35 Watts from 100, 120,
220, or 240 Vac
Warranty: One year parts and labor
on materials and workmanship
6
Menu Display List
MENU DISPLAY LIST
MODE
COUNT=A,B FOR T PRESET A=INPUT 1 B=INPUT 2 B SET=1E3 T=10MHZ T SET=1E0 s N PERIODS=2000 [at 1234] AT N=STOP DWELL=1E0s D/A=A D/A RANGE=LOG DISPLAY=CONTINUOUS
LEVEL
TRIG SLOPE=RISE TRIG LVL=+2.000 V A DISC SLOPE=FALL A DISC=SCAN ∆=-1.0 mV A DISC LVL=-10.0 mV B DISC SLOPE=FALL B DISC=SCAN ∆=-1.0 mV B DISC LVL=-10.0 mV T DISC SLOPE=FALL T DISC=SCAN ∆=-1.0 mV T DISC LVL=-10.0 mV PORT1=SCAN ∆=+0.100 V PORT1 LVL=+0.500 V PORT2=SCAN ∆=+0.100 V PORT2 LVL=+0.500 V
A GATE
AGATE=SCAN ∆=1.000 µs A DELAY=1.000 µs A WIDTH=1.000 µs
SETUP
LCD CONTRAST=12 STORE=3 EXECUTE=> RECALL=3 EXECUTE=>
B GATE
BGATE=SCAN ∆=1.000 µs B DELAY=1.000 µs B WIDTH=1.000 µs
COM
GPIB ADDR=23 RS-232 BAUD=9600 RS-232 BITS=8 RS-232 PARITY=NONE RS-232 WAIT=6 RS-232 ECHO=OFF DATA=
7
Menu Display List
8
Abridged Command List
ABRIDGED COMMAND LIST
The parameter i is 0,1,or 2 to select counter A,B, or T
MODE
CM j Set COUNT to mode j; A,B(0), A-B(1), A+B(2) for preset T,or A FOR B preset(3). CI i,j Set counter i to input j; 10 MHz(0), INP 1(1), INP 2(2), TRIG(3). CP i, n Set counter i preSET to 1 <= n <= 9E11. NP m Set Number of PERIODS in a scan to 1 <= m <= 2000. NN Read current count period number or scan position. NE j Set end of scan mode to mode j; START(1) or STOP (0). DT x Set DWELL time to 2E-3 <= x <= 6E1 s or EXTERNAL(0). AS j Set D/A to source j; A(0), B(1), A-B(2), A+B(3). AM j Set front panel D/A to RANGE j; LOG(0) or LINEAR (1-7). SD j Set DISPLAY to mode j; CONTINUOUS(0) or HOLD(1).
LEVELS
TS j Set TRIG to SLOPE j; RISE(0) or FALL(1). TL v Set TRIG LVL to -2.000 <= v <= 2.000 V. DS i,j Set DISC i to SLOPE j; RISE(0) or FALL(1). DM i,j Set DISC i to mode j; FIXED(0) or SCAN(1). DY i,v Set DISC i scan step to -0.0200 <= v <= 0.0200 V. DL i,v Set DISC i LVL to -0.3000 <= v <= 0.3000 V. DZ i Read current DISC i LVL (during scan). PM k, j Set PORT k (1 or 2) to mode j; FIXED(0) or SCAN(1). PY k,v Set PORT k (1 or 2) scan step to -0.500 <= v <= 0.500 V. PL k,v Set PORT k (1 or 2) LVL to -10.000 <= v <= 10.000 V. PZ k Read current PORT k (1 or 2) LVL (during scan).
GATES
GM i,j Set GATE i to mode j; CW(0), FIXED(1), or SCAN(2). GY i,t Set GATE i DELAY scan step to 0 <= t <= 99.92E-3 s. GD i,t Set GATE i DELAY to 0 <= t <= 999.2E-3 s. GZ i Read current GATE i DELAY position (during a scan). GW i,t Set GATE i WIDTH to 0.005E-6 <= t <= 999.2E-3 s.
FRONT PANEL
CS Count start, same as START key. CH Count pause, same as STOP key while counting. CR Count reset, same as STOP key pressed twice. CK j Simulate key press j. RR Rotate knob right (CW or UP) 1 step. RL Rotate knob left (CCW or DOWN) 1 step. SC Read cursor position. Left(0), right(1), or inactive(2). MI j Set front panel to mode j; local(0), remote(1), locked-out(2). RS-232 only. MS string Display string on menu line. MS Return menu line to normal display. MD j,k Display line k of menu j. MM Read menu number of display (j). ML Read menu line of display (k).
9
Abridged Command List
STORE/RECALL
ST m Store instrument settings to location m (1 to 9). RC m Recall instrument settings from location m (1 to 9). RC 0 Recall default settings.
INTERFACE
CL Clear instrument. SV m Set GPIB SRQ mask to 0 <= m <= 255. SS Read status byte. SS j Read bit j(0-7) of status byte. SI Read secondary status byte. SI j Read bit j (0-2) of secondary status byte. SW m Set RS-232 character wait interval to m*3.33 ms 0 <= m <= 25. RS-232 only. SE j,k,l,m Set RS-232 terminator sequence to j,k,l,m (ASCII codes). RS-232 only. SE Clear RS-232 terminator sequence to defaults. RS-232 only.
DATA
QA Read last count in counter A. QB Read last count in counter B. QA m Read from scan buffer point m(1-2000) for counter A. QB m Read from scan buffer point m(1-2000) for counter B.
EA Send entire counter A buffer. EB Send entire counter B buffer. ET Send entire counter A and B buffer.
FA Start scan and send N PERIODS data points from counter A. FB Start scan and send N PERIODS data points from counter B. FT Start scan and send N PERIODS data points from both counters.
XA Read current contents of counter A. XB Read current contents of counter B.
STATUS BYTE
Bit Description
0 Parameter changed 1 Data ready 2 Scan finished 3 Counter overflow 4 Rate error 5 Recall error 6 Service request 7 Command error
SECONDARY STATUS BYTE
Bit Description
0 Triggered 1 Inhibited 2 Counting 3-7 Unused
10
Instrument Overview
INSTRUMENT OVERVIEW
SR400 OVERVIEW
The SR400 architecture is diagrammed above. There are three fast counters identified as counters A, B, and T. All three counters operate at rates up to 200 MHz. The input to each counter is selected from a number of sources, including the two analog signal inputs, the crystal timebase, and the external trigger input. Counter T is presettable and determines the measurement period. For pulsed experiments, counters A and B may also be synchronized to external events via the two independent gate generators. The gate generator provides gates from 5 ns to 1 s in duration with a delay from external trigger ranging from 25 ns to 1 s.
SIGNAL INPUTS AND DISCRIMINATORS
There are two independent analog signal inputs labelled INPUT 1 and INPUT 2. They are
internally terminated into 50 Ω. The inputs can
accept signals of either polarity up to ±300 mV and are protected to ±5V dc. Each input is followed by a dc to 300 MHz amplifier. This allows detection of pulses as s mall as 10 mV. If greater sensitivity is required, the SR445A remote amplifier is available. The SR445A can provide gains from 5 to 625 at bandwidths up to 350 MHz.
There are three discriminators, one for each counter. Each discriminator has selectable slope and a threshold programmable from -300 mV to +300 mV in 0.2 mV steps (referenced to the inputs). Pulse-pair resolution is 5 ns and pulses of either polarity may be detected. Each threshold
11
Instrument Overview
may be programmed to scan in either direction with selectable step size.
COUNTING
Each counter's input may be selected from a number of sources. All multiplexing of counter inputs is done internally to provide accurate timing and ease of operation. Input cables should rarely need to be swapped or disconnected to change measurement modes.
The actual inputs to the counters can be viewed as NIM level pulses from the DISC outputs on the front panel. The DISC pulses are negative going from 0 to -0.7 V. The falling edge is the active or counted edge regardless of the discriminator slope setting. The DISC outputs are very useful when adjusting discriminator thresholds or gate timing.
When the START key is pressed, a START command is received from the computer interface, or the EXTERNAL START input is pulsed, counter T is enabled to count. The COUNT PERIOD begins with the first pulse from T counter's input after START. The COUNT PERIOD enables counters A and B to count their respective inputs. If the A GATE and B GATE are CW or continuous, all pulses that occur during the COUNT PERIOD are counted. If the gates are enabled, only the pulses that occur during those gates which fall within the COUNT PERIOD are counted.
Counter T is presettable from 1 to 9E11. When counter T reaches the preset count, the COUNT PERIOD is terminated and counting is halted on all counters. When the input to counter T is the 10 MHz crystal timebase, the COUNT PERIOD is a fixed time interval from 100 ns to 25 hours. When the input is INPUT 2, the COUNT PERIOD is externally determined and can be as short as 5 ns and indefinitely long. This mode is used for source compensation or reciprocal measurements. If the input is TRIGGER, the COUNT PERIOD is a fixed number of gates (each trigger provides one gate). This results in a constant gate aperture time during each COUNT PERIOD. N+1 Triggers are required for N gates per period. If the input is B PRESET, then counter T counts the B DISC
output gated by B GATE. This is for gated ratio or reciprocal counting.
The SR400 may be programmed to cycle through 1 to 2000 COUNT PERIODS in a single scan. At the end of the programmed scan, the counters may be stopped or the scan may be restarted. If the end of scan mode is STOP, then after the last COUNT PERIOD in the scan, the counters are halted indefinitely. A counter RESET is required to rearm the counters and a START command starts another scan. If the Number of PERIODS in the scan is 1, then each START command results in one COUNT PERIOD. If N PERIODS is greater than 1, then consecutive COUNT PERIODS are separated by the DWELL time. The DWELL time may be set from 2 ms to 60 s. During the DWELL time, counting is disabled and data may be transferred or external parameters scanned. The DWELL OUTPUT is a TTL output which is high during the DWELL time. At the end of the DWELL time, the next COUNT PERIOD begins and the scan continues until N PERIODS have been measured. If the end of scan mode is START, one DWELL time separates the last COUNT PERIOD of the completed scan and the first COUNT PERIOD of the next scan.
Another DWELL mode is EXTERNAL. In this mode, the EXTERNAL START input enables counter T. The COUNT PERIOD begins with the first pulse to counter T after START. Counting then proceeds normally until the preset is reached. At the end of the preset COUNT PERIOD, the DWELL time is indefinite and the next COUNT PERIOD starts with t he n ext EXTE RNAL ST ART pulse. In this way, preset COUNT PERIODS may be synchronized to external events. If externally determined COUNT PERIODS are desired, then the preset COUNT P ERIOD can be set to a value much larger than actually required and the EXTERNAL STOP input used to terminate the COUNT PERIOD. The COUNT PERIOD then starts with the first pulse to counter T after EXTERNAL START and ends with EXTERNAL STOP. The next EXTERNAL START be gins the next COUNT PERIOD in the scan. When the scan is finished and the scan end mode is STOP, a RESET from the front panel or computer interface is required to reset the scan. If the scan end mode
12
Instrument Overview
is START, the next EXTERNAL START begins the first COUNT PERIOD of the next scan.
SCANNING
In all scan modes, a number of parameters may be scanned. These parameters are the three discriminator thresholds (Pulse Height Analysis), the two gate delays (Boxcar mode), and the two D/A output ports (X-axis of recorder, scope, or analog control of other apparatus).
After each COUNT PERIOD, each scanning parameter is adjusted by one step. All changes are made during the DWELL time so that all values are stable during the COUNT PERIODS. The scan limits are determined by the start position, the step size, and the number of PERIODS in a scan. The range of a scan is N PERIODS times the step size. When the counters are reset, all scanned parameters return to their start positions.
GATING
The gate generators may be disabled by setting the gate modes to CW. In this mode, the gates are continuously open and counting is enabled solely by the COUNT PERIOD. This is appropriate for CW sources or long counting intervals. If the signal is pulsed or so weak that synchronous (lock­in) measurement is required, then the gate generators are required. The gates prevent pulses that occur outside the time interval of interest from being counted and improve signal to noise ratio. When the gates are scanned, the time behavior of the signal may be recovered.
The gates are triggered by the TRIGGER input. The TRIGGER threshold is adjustable from -2V to +2V and slope may be either RISE or FALL. The discriminated trigger pulse may be viewed through the T DISC output when the input to counter T is set to TRIG. This can be helpful when setting the trigger threshold.
The minimum insertion delay from trigger to gate is 25 ns. The additional delay may be programmed from 0 ns to 999.2 ms with a resolution of 1 part in 1000 or 1 ns, whichever is greater. The gate width is adjustable from 5 ns to 999.2 ms with the same resolution. The two gates are output as NIM pulses
from the GATE outputs. The gates are open when the outputs are low. The gate outputs have a timing accuracy with respect to the discriminator outputs of 2 ns. This allows accurate timing of fast gates relative to the signal.
The two gates have independent modes, delays, and widths. This allows for a variety of measurements. For example:
1) "Boxcar" mode. The gates are used to enable the counters only when the signal is present. If the gates are scanned, then the time profile of the signal is recovered. Background subtraction can be included by using one gate for the signal and an equal width gate on the background, either before or after the signal. The signal gate may be scanned while the background gate held fixed. Gated source compensation is also possible since counters A and B may count different sources at the same time or the same source at different times. In the compensation mode, the COUNT PERIOD is determined by the gated output of the B discriminator and the COUNT MODE is A FOR B PRESET.
2) "Lock-in" or "Chopped" mode. In this case, a cw signal is chopped on and off. This can be accomplished with a light chopper such as the SR540. The light chopper reference output is used to trigger the gates. The A GATE is positioned during the "open" cycle and counts signal plus noise. The B GATE is positioned during the "closed" cycle and only counts the noise. The counter outputs are subtracted to give only the signal. The widths of both gates should be equal and less than half of the chopper period. This prevents frequency jitter in the chopper from affecting the overlap of the gates with the open and close cycles of the chopper.
When counter T is preset and its input is set to TRIG, then each COUNT PERIOD consists of a constant number of gates. N+I triggers are needed for N gates per preiod. This is the usual mode of gated operation. However, a number of other modes are available. For example, if the input is set to 10 MHz, each COUNT PERIOD consists of the number of gates which occur during the COUNT PERIOD time interval. If the input is a signal input, then counting is both gated and
13
Instrument Overview
source compensated. If the DWELL mode is EXTERNAL, an EXTERNAL START p ulse s tart s counting and an EXTERNAL STOP pulse stops counting. The counters are enabled only during those gates which occurred between the two EXTERNAL pulses.
COUNT MODES
The length of a COUNT PERIOD is the time it takes counter T to reach its preset count. If N the preset count in counter T, and R
and Rt are the
a
is
t
mean pulse rates of the inputs to counters A and T, then the COUNT PERIOD is
T = N
/ Rt
t
The number of counts in counter A will be
A = R
T = Ra Nt / Rt
a
Rate:
If R
= 10 MHz, then the COUNT PERIOD is a
t
constant amount of time and
A = R
( Nt / Rt ) ~ Ra
a
and counter A measures the mean rate of its input.
Gated:
If R
is the gate trigger rate and counter A is
t
gated, then the COUNT PER IOD is N "effective" time the counter is enabled is N where W
is the width of A gate and
a
but the
t/Rt
tWa
A = R
( Nt Wa ) ~ Ra
a
In this case, counter A measures the mean rate during the A gate.
Source Compensated:
If Rt is INPUT 2, then the COUNT PERIOD is determined by the pulse rate of INPUT 2 and the T discriminator. Count A will be
A = N
Ra / Rt ~ Ra / Rt
t
and is proportional to the ratio of the rates of counters A and T. If INPUT 2 monitors source
intensity, then the COUNT PERIOD will adjust for fluctuations in the source. Counter A measures the ratio of its mean input rate and the mean rate of INPUT 2.
Reciprocal:
If R
is INPUT 2, then the COUNT PERIOD is
t
N
. If Ra is 10 MHz, then counter A measures
t/Rt
A = R
Nt / Rt = ( Ra Nt ) / Rt ~ 1 / Rt
a
and is proportional to the reciprocal of the count rate of INPUT 2. Not e that ea ch COUNT PE RIOD is determined by N
counts of INPUT 2, thus the
t
measurement time is determined by the signal strength. Measurements of both strong and weak signals are made to the same accuracy for constant signal to noise ra tio .
Counter B:
Counter B may be used in all of the above modes except Reciprocal. Since counter B is completely independent of counter A, 2 sources of counts may be measured simultaneously. These sources may be the same signal or 2 different signals; the sources may be discriminated at separate thresholds; they may be non-gated or gated; if gated, the 2 gates may be independently set. Almost any measurement can be made.
In addition, counter B has a preset mode called A FOR B PRESET. In this mode, the input to counter B is sent to counter T. The COUNT PERIOD is now determined by the B discriminator and the B gate. If R
is INPUT 1,
a
then counter A is source compensated by the gated input of counter B. This allows source compensation of gated signals and sources. If R
a
is 10 MHz, then counter A measures the reciprocal of the rate of counter B's gated input.
OUTPUTS
The front panel can display counts up to 10
9
- 1. Counters A and B can be displayed as separate counters or combined as A-B or A+B. When A FOR B PRESET is the count mode, then only A data is displayed.
14
Instrument Overview
The front panel D/A output provides an analog output proportional to A, B, A-B, or A+B depending on the counting mode. The scale may be logarithmic (1V/decade) or linear to any 3 digits.
If a chart recorder is used, the D/A output should drive the Y axis. The X axis can be recorder driven (strip chart mode) or the PORT1 or PORT2 outputs can be used to drive the X axis (scans). This latter method allows accurate determination of the X value of each point. If the DWELL output is used as the pen lift, the points will be unconnected.
If count rates are high and count periods short, then scans may be displayed on an X-Y scope. The D/A output is the Y drive and PORT1 or PORT2 is the X drive. The DWELL output should be used as the blanking pulse.
15
Instrument Overview
16
Measurement Examples
MEASUREMENT EXAMP L ES
Example 1: Internally Triggered Gate Mode
This example, described under the heading “QUICK TEST” on page 59 of this manual, is easy to perform and will quickly acquaint the user with the basic operation of the SR400.
Example 2: Externally Triggered Gate Mode
In this example, the SR400 is programmed to make 10 counting measurements of a 10 kHz pulse train. For each measurement, a gate will be delayed 500 ms from an external trigger pulse and will open of 500 ms duration. Following the gate, there will be a 6 second dwell period* during which the counter will be disarmed. After the dwell period, the counter will reset to 0 counts, re-arm and repeat the measurement. After 10 measurement cycles, the SR400 will stop counting and disarm. You will need a 10 kHz signal source and a
0.5 Hz square wave generator for this example.
*It is important to note that the dwell period begins at the first external trigger pulse arriving after the gate is closed. Thus, even though there is only one gate opened for each measurement cycle, a minimum of two external triggers are needed: one to trigger the gate and another to begin the dwell period.
The following timing diagram illustrates this example.
17
Measurement Examples
1. Turn the SR400 off. Hold down the STOP (RESET) button while turning the SR400 on. Wait three seconds and release the button.
2. Press the MODE key. Using the four arrow keys and the front panel knob, select the following parameters. Unless otherwise indicated, leave all other settings at their default values:
T = TRIG TSET = 1E0 N PERIODS = 10 [AT 0] AT N = STOP DWELL = 6E0 S
3. Press the A GATE key and select the following parameters:
A GATE = FIXED A DELAY = 500 mS A width = 500 mS
4. Press the LEVEL key and select the following parameters:
A DISC SLOPE = RISE A DISC LVL = +150 mV
5. Using a function generator or a pulse generator, apply a 0.5 Hz, 5 V (TTL) signal to the TRIGGER INPUT on the SR400.
6. Using a second signal source, apply a 10 kHz square wave (or pulse train) with an amplitude of approximately 500 mV to the SIGNAL INPUT 1 on the SR400.
7. To view the COUNT PERIODS, press the MODE key and use the up/down arrow keys to display “N PERIODS = 10 [at 0]”
Arm the counter by pressing the START button. (Alternatively, a TTL pulse applied to the START BNC on the front panel will arm the counter.) Each measurement cycle should result in approximately 5,000 counts.
As an extension of the above example, the SR400 can be programmed to accumulate counts over numerous gates. This is often done to improve the signal to noise ratio of low light level experiments. To accumulate counts, set TSET (MODE menu) to the desired number of gates. For example, if you wish to accumulate over five gates, set TSET – 5E0. In this case, six external triggers are required for each measurement cycle: five to trigger the five gates and one to start the DWELL PERIOD. Any additional triggers received during the DWELL PERIOD are ignored.
18
Measurement Examples
Example 3: Externally Triggered Gate Mode
This example is a variation of Example 2, but requires no additional trigger pulses to begin the DWELL PERIOD. In this example, the SR400 is programmed to make 10 counting measurements of a 10 kHz pulse train. The counting gates, each 500 ms wide, are delayed 300 ms from external triggers arriving at a 1 Hz rate. In this case, the SR400 requires only one trigger per COUNT PERIOD, however the trigger configuration is a bit trickier than the previous example.
The following timing diagram illustrates this example.
To avoid the triggered dwell period, the T counter is logically ANDed with the A gate. This is accomplished by letting the T counter count the SR400’s internal 10 MHz oscillator signal for a period extending our beyond the A gate, but not exceeding the trigger repetition period (1 second in this case). Mathematically, this is expressed:
A DELAY + A WIDTH < TSET < (1/Trigger Rate)
Furthermore, DWELL must be set to EXTERNAL in the MODE menu. This enables the SR400 to reset and begin each COUNT PERIOD upon receiving a TTL pulse at the START BNC input.
19
Measurement Examples
1. Turn the SR400 off. Hold down the STOP (RESET) button while turning the SR400 on. Wait three seconds and release the button.
2. Press the MODE key. Using the four arrow keys and the front panel knob, select the following parameters. Unless otherwise indicated, leave all other settings at their default values:
T = 10 MHz TSET = 9E -1 s N PERIODS = 10 [AT 0] AT N = STOP DWELL = EXTERNAL (Turn the knob counterclockwise until “DWELL = EXTERNAL” is displayed.)
3. Press the A GATE key and select the following parameters:
A GATE = FIXED A DELAY = 300 mS A width = 500 mS
4. Press the LEVEL key and select the following parameters:
A DISC SLOPE = RISE A DISC LVL = +150 mV
5. Using a function generator or a pulse generator, apply a 1 Hz, 5 V (TTL) signal to the EXTERNAL START input on the SR400. Use a BNC ‘T; connector at the EXTERNAL START input and extend a cable from the EXTERNAL START input to the TRIG input.
6. Using a second signal source, apply a 10 kHz square wave (or pulse train) with an amplitude of approximately 500 mV to the SIGNAL INPUT 1 on the SR400.
7. To view the COUNT PERIODS, press the MODE key and use the up/down arrow keys to display “N PERIODS = 10 [at 0]”
8. Arm the SR400 by pressing the STOP (RESET) button.
Each counting period should result in approximately 5,000 counts. After 10 counting measurements the SR400 will disarm. To re-arm, press the STOP (RESET) button.
20
Guide to Operation
GUIDE TO OPERATION
FRONT PANEL SUMMARY
POWER BUTTON
The SR400 is turned on by depressing the POWER button. All instrument settings are stored in nonvolatile memory and are retained when the power is turned off. The model, firmware version, and serial numbers of the unit will be displayed briefly when the power is turned on. The recall status is also dipslayed. If recall is OK, then the instrument returned to the settings in effect when the power was last turned off. If recall is ERR, then the stored settings were lost and the default settings are used. If recall is DEFAULT, then the instrument is resetting to default conditions. This is done if the STOP (RESET) key is held down when the power is turned on. The default setup is listed in the next chapter.
LIQUID CRYSTAL DISPLAY
The 2 line by 24 character LCD is the user interface for data display and front panel programming operations. The top line always shows count data. The lower line is the menu line and displays one line of setup information. The LCD is backlit by an electroluminescent panel for convenient operation in low light level conditions. When in the SETUP menu, the contrast of the LCD may be adjusted for optimum viewing.
MENU KEYS
All operating parameters of the SR400 are grouped into six function menus. The six menu
keys select the group of functions to be programmed. The four arrow keys select individual parameters within each menu. The selected parameter is displayed on the menu line of the display.
The MODE menu selects the counting mode, counter inputs, counter presets, scan mode and length, dwell time, analog output, and display mode. The LEVEL menu adjusts the discriminator, trigger, and D/A levels. The A GATE and B GATE menus program the gate generators. The SETUP menu provides STORE and RECALL of 9 different instrument setups and adjustment of the LCD contrast. The COM menu sets up the computer interface parameters and displays received data. Detailed descriptions of the menus are provided at the end of this section.
SPIN KNOB
All parameters are altered using the spin knob. Whenever the menu line has an active or flashing cursor, the displayed parameter can be adjusted with the knob.
START, STOP, AND RESET
The two count keys, START and STOP, are used to start, pause, and reset the counters.
A scan can be started only if the counters are in the reset condition; all scanned parameters are at their st art va lu es a nd the c ount er s ar e res et t o zero .
21
Guide to Operation
In general, pressing the STOP key twice in succession puts the counters into the reset condition.
START begins the first COUNT PERIOD of the programmed scan. If the DWELL time is programmed, then the scan will continue automatically. A STOP will pause the scan; the current COUNT PERIOD is terminated and ignored. A START while paused will resume the scan. A STOP while paused resets the scan. If the DWELL time is EXTERNAL, then a STOP during the COUNT PERIOD terminates the COUNT PERIOD and advances the scan. START begins the next COUNT PERIOD i n the scan. A START during a COUNT PERIOD or programmed DWELL period is ignored.
A STOP while in a dwell period or pause resets the scan; all scanned parameters return to their start values and the counters are reset. In general, 2 consecutive presses of the STOP key will reset the counters.
The SCAN led is on when a scan is in progress. The SCAN led is off when the scan is paused or the counters are reset.
INSTRUMENT STATUS
Six LED's indicate the instrument status. REM is on when the front panel is locked out, ACT flashes when there is activity on the computer interfaces, and ERR flashes when there is a computer interface error. COUNT is on during each COUNT PERIOD, OVRN indicates that a counter has exceeded 10
9
- 1 counts during a COUNT PERIOD, and RATE flashes whenever a gate delay or width exceeds the trigger period minus 1 µs. The RATE led also flashes if the trigger rate exceeds 1 MHz, even if the gates are in the CW mode.
COUNTER LED'S
The source LED's in the counter sections indicate what each counter's input is. The counter inputs are output through the DISC outputs. The SET LED's indicate which counter, B or T, is the preset counter. The B SET LED is on when the mode is
A FOR B PRESET, the T SET LED is on otherwise.
INPUTS
The rising or falling edge of the TRIGGER input triggers the gate generators. The input impedance is 10 KΩ and the minimum pulse width is 10 ns. The trigger level is adjustable from -2V to +2V with either slope. The minimum pulse amplitude is
0.1 V.
The SIGNAL INPUTS, 1 and 2, are 50 Ω analog
inputs. Photomultiplier or preamplifier outputs are connected to these inputs. These inputs are linear between -300 mV to +300 mV and protected to ± 5 Vdc.
EXTERNAL START is a TTL input with the rising edge active. The minimum pulse width is 20 ns. When the DWELL time is EXTERNAL, the COUNT PERIOD begins with the rising edge of START and ends with the preset condition or the rising edge of STOP. When the DWELL time is programmed, EXT START begins the scan.
EXTERNAL STOP is a TTL input with the rising edge active. The minimum pulse width is 20 ns. When the DWELL time is EXTERNAL, EXT STOP ends each count period. Reset must be accomplished via the STOP key or computer interface. When the DWELL time is programmed, EXT STOP resets the scan.
OUTPUTS
The DISC outputs are 50 Ω, NIM level, outputs
showing the actual inputs to the three counters. These outputs are active low (-.7 V) and counting occurs on the falling edge. The outputs are low when the signal inputs exceed their threshold levels with the correct slope.
Cables attached to the DISC outputs should be kept far from the cables which carry the SIGNAL inputs. The 0.7 V DISC pulses can be picked up by the SIGNAL cables and discriminated. This causes the discriminators to oscillate. This is not a problem if the threshold is greater than 10 mV. If smaller thresholds are required, careful arrangement of the cables is necessary, or the
22
Guide to Operation
DISC output should just be terminated at the front panel.
The GATE outputs are 50 Ω, NIM level, outputs
showing the actual gates to counters A and B. The GATE outputs are low (-.7 V) during the gate. The timing accuracy between the GATE outputs and DISC outputs is 2 ns.
The D/A analog output provides a -10V to +10V output proportional to the count data with linear or
logarithmic scaling. Output impedance is < 1 Ω
and the maximum current output is 10 mA.
The DWELL output is a TTL level output which is high whenever the counters are paused. The rising edge may be used to trigger changes in external parameters or data acquisition.
23
Guide to Operation
REAR PANEL SUMMARY
POWER ENTRY MODULE
The power entry module is used to fuse the line, select the line voltage, and block high frequency noise from entering or exiting the instrument. Refer to the first page of this manual for instructions on selecting the correct line voltage and fuse.
IEEE-488 STD PORT
The 24 pin IEEE-488 connector allows a computer to control the SR400 via the IEEE-488 or GPIB instrument bus. The address of the instrument is set from the front panel using the COM menu.
RS-232 CONNECTOR
The RS-232 interface connector is configured as a DCE. The baud rate, parity, character length, and echo mode are programmed from the front panel using the COM menu.
INPUT
INHIBIT is a TTL input which disables the A,B, and Tdiscriminators when it is high. The minimum pulse width is 10 ns.
OUTPUT
The PORT1 and PORT2 D/A analog outputs provide -10 V to +10V, user programmable outputs. The outputs may be scanned to drive the X axis of a recorder or scope during scans. Output
impedance is < 1 Ω and the maximum current
output is 10 mA.
24
Guide to Operation - Menus
FRONT PANEL MENUS
The lower line of the LCD display is the menu line. Menu lines are grouped into six menus. To select a new group or menu, press a new menu key. The menu line that will be displayed is the most recently active menu line of the new menu.
Each menu contains several lines. Each line may consist of 1 or 2 displayed parameters. The DOWN and UP keys change which menu line within a menu will be displayed. The DOWN key displays the next menu line and the UP key, the previous line. Pressing either key many times in succession will scroll through all of the menu lines within a menu.
If the menu line contains two parameters, the LEFT and RIGHT keys select between the two. On some lines, the right hand parameter is active only for certain values of other parameters. On these lines, the right hand parameter may be inactive.
The spin knob is used to modify the active parameter. A flashing cursor precedes the value which the knob will alter.
The following section details each menu line. The lines are listed in the order in which the DOWN key scrolls through each menu. For parameters with limited choices, all selections are shown. For numerical parameters, example values are shown.
MODE MENU
COUNT= A,B FOR T PRESET A-B FOR T PRESET A+B FOR T PRESET A FOR B PRESET
This line selects the data display and the preset counter. The upper line of the LCD is the count display. There are four displays available; A and B separately, A-B, A+B, or only A, depending upon the COUNT selection. All count data is right justified with leading zeroes off. The preset counter is either T or B. The appropriate SET LED will be on indicating which counter is preset. Note that the choice of inputs to the counters determines
whether the count data measures rate, ratio, or reciprocal.
Changing this line causes the counters to reset.
A= INPUT 1 10 MHz
The input to counter A can be either the internal 10 MHz timebase or the signal at INPUT 1 discriminated by the A discriminator. The selected counter input can be viewed at the A DISC output. The 10 MHz source is used for reciprocal measurements or for quick testing of the counter operation. The appropriate A counter source LED is also lit.
Changing this line during a COUNT PERIOD may result in a bad data point.
B= INPUT 2 BSET= **** INPUT 1 1E0 9E11
The input to counter B can be either the signal at INPUT 1 or INPUT 2 discriminated by the B discriminator. The selected counter input can be viewed at the B DISC output. The appropriate B counter source LED is lit. Changing the counter input during a COUNT PERIOD may result in a bad data point.
If counter T is preset, the B SET parameter is inactive (indicated by the **** on the display). If counter B is preset (ratio or reciprocal measurement), B SET is the B preset count from 1E0 to 9E11. Changing B SET during a scan causes the counters to pause; any COUNT PERIOD in progress is terminated and ignored. A START begins the next COUNT PERIOD with the new preset value.
T= 10 MHz TSET= 1E0 s INPUT 2 1E7 TRIG 1E3 ****
25
Guide to Operation - Menus
The input to counter T can be the internal 10 MHz timebase (rate measurement), the signal at INPUT 2 discriminated by T discriminator (ratio or reciprocal measurement), or the TRIGGER input discriminated by the trigger discriminator (constant gate aperture). The selected counter input can be viewed at the T DISC output. The appropriate T counter source LED is lit.
If counter B is preset, this line is inactive (indicated by the **** on the display). If counter T is preset, both parameters on this line are active. T SET is the preset value of counter T from 1E0 to 9E11. If the T input is 10 MHz, then T SET is displayed as real time = T SET cycles of 10 MHz. ChangingT SET during a scan causes the counters to pause; any COUNT PERIOD in progress is terminated and ignored. A START begins the next COUNT PERIOD with the new preset value.
N PERIODS=2000 [at 1234]
This line programs the number of COUNT PERIODS in one scan from 1 to 2000. N PERIODS determines the length of the internal data buffer and the range of any scanned parameters. The quantity in brackets is the number of completed COUNT PERIODS in the current scan.
Changing N PERIODS during a scan changes the length of the current scan. Reducing N PERIODS past the current scan position will cause an end of scan condition.
AT N= STOP DWELL= 1E0 s START 2E-3 s EXTERNAL
This line sets the end of scan mode and the DWELL time. When a scan reaches N PERIODS, the counters may be paused or restarted. AT N=STOP causes an end of scan pause. All scanned parameters are held at their final values. A counter reset is required before another scan can begin. If N PERIODS is 1, then each START results in one COUNT PERIOD. If N PERIODS is gre ater than 1, then consecutive COUNT PERIODS are separated by the DWELL time. During the DWELL time, counting is disabled and scanned
parameters updated. The DWELL OUTPUT is TTL high during the DWELL time.
If AT N=START, the counters will wait one DWELL time at the end of each scan, during which all scanned parameters return to their start positions, and then another scan begins. This is used for continuous data taking with the measurement rate determined by the COUNT PERIOD plus the DWELL TIME.
The DWELL time may be programmed from 2E-3 s to 6E1 s (1 minute) or set to EXTERNAL. The internal DWELL time is based on the 10 MHz timebase and may have 200 ns of indeterminacy. Changing the DWELL time during a scan causes the counters to pause; any COUNT PERIOD in progress is terminated and ignored.
Setting the DWELL time below 2E-3 s results in an EXTERNAL dwell. In this mode, the START key, EXT START, or remote start command starts the COUNT PERIOD. The COUNT PERIOD las t s until the preset condition is reached. The following DWELL time is indefinite; the next COUNT PERIOD begins with another START. A STOP key, EXT STOP, or remote stop command during the COUNT PERIOD terminates the COUNT PERIOD. Thus by programming the preset COUNT PERIOD much longer than required, the COUNT PERIOD may be determin ed by START and STOP. Scanning proceeds as normal.
NOTE: If AT N=STOP, the counters will pause at the end of a scan. A reset (STOP key, EXT STOP, or remote reset command) is required to rearm the counters before another scan can begin. In general, to reset the counters at any time, the STOP ke y is pressed twice in succession.
D/A= A B A+B A-B
The D/A output is proportional to the count data on the top line of the LCD. When the COUNT is A-B, A+B, or A only, the D/A output can only follow the data display. In these modes, the cursor is inactive and the knob does nothing. If the
26
Guide to Operation - Menus
COUNT is A,B FOR T PRESET, then the D/A output can follow either A or B.
D/A RANGE= LOG ##765####
The D/A output scale can be either logarithmic or linear. If the output is logarithmic, the scale is 1V per decade with 1 count being 0V, 0 counts being ­1V, and negative counts being -2V. If the scale is linear, the output is proportional to 3 digits of the count. These digits are shown while the ignored digits are represented by #'s.
The D/A output is updated at the end of each COUNT PERIOD. If the D/A output is driving a chart recorder, the end of scan mode should be STOP or the pen reset will mar the graph. If the DWELL output is connected to the recorder pen lift, data will be plotted as unconnected points, otherwise, the discrete points will be connected.
DISPLAY= CONTINUOUS HOLD
The count data on the upper line of the LCD can be updated in two ways. CONTINUOUS mode updates the display 20 times a second during the COUNT PERIOD. Data is held during the DWELL time at the final value reached during the previous COUNT PERIOD. HOLD updates the display only at the end of each COUNT PERIOD.
CONTINUOUS is useful when COUNT PERIODS are long, HOLD is better when COUNT PERIODS are short.
LEVEL MENU
TRIG SLOPE= RISE FALL
The gate generator may be triggered on either the rising or falling edge of the TRIGGER input.
TRIG LVL=+2.000 V
The gate TRIGGER threshold level may be adjusted from -2.000 V to +2.000V. The
discriminated trigger pulses may be viewed through the T DISC output when counter T is preset and its input is TRIG. This can be helpful in selecting a trigger level. The minimum pulse amplitude required to trigger is 100 mV.
A DISC SLOPE= RISE FALL
When A counter's input is INPUT 1, A discriminator may be triggered by rising or falling edges of the INPUT 1 signal. If the signal pulses are negative going, then the slope should be FALL. If the signal pulses are positive, then the slope should be RISE. In either case, the A DISC output is always active low; i.e., a discriminated rising edge of the signal will be output as a negative going NIM pulse at the A DISC output. The falling edges of the A DISC output are counted by counter A.
A DISC= FIXED SCAN ∆= -1.0 mV
The threshold level of A discriminator may be FIXED or SCANned. If the level is fixed, the right hand parameter is not displayed. If the level is scanned, then the right hand parameter is the step size. At the end of each count period in a scan, the level is incremented by one step size. The step size can be varied from -20.0 mV to + 20.0 mV. The range of a scan is N PERIODS times the step size. The starting level is set in the next menu line.
The scan mode may be changed during a scan. If it is changed to SCAN, the level begins scanning from the start value on the next count period. If it is changed to FIXED, the level returns to the start level immediately.
The step size may also be changed during a scan. The new step size will be in effect at the end of the next count period. The scan is not reset or paused.
A DISC LVL = -10.0 mV
: -11.0 mV
The threshold level of A discriminator may be adjusted from -300 mV to + 300 mV. This level is
27
Guide to Operation - Menus
the fixed threshold when A DISC is FIXED and the starting level when A DISC is SCANned.
When the A DISC mode is SCAN and a scan is in progress, this line displays the current position of the A DISC level. The cursor is inactive and replaced by a ':'. The spin knob cannot adjust the threshold while it is scanning. When the mode is SCAN, the starting level may only be adjusted when the counters are in reset.
B DISC SLOPE= RISE FALL
B discriminator may be triggered by rising or falling edges of the INPUT 1 or INPUT 2 signal. If the signal pulses are negative going, then the slope should be FALL. If the signal pulses are positive, then the slope should be RISE. In either case, the B DISC output is always active low; i.e., a discriminated rising edge of the signal will be output as a negative going NIM pulse at the B DISC output. The falling edges of the B DISC output are counted by counter B.
B DISC= FIXED
SCAN ∆= -1.0 mV
The threshold level of B discriminator may be FIXED or SCANned. If the level is FIXED, the right hand parameter is not displayed. If the level is scanned, then the right hand parameter is the step size. At the end of each count period in a scan, the level is incremented by one step size. The step size can be varied from -20.0 mV to +
20.0 mV. The range of a scan is N PERIODS times the step size. The starting level is set in the next menu line.
The scan mode may be changed during a scan. If it is changed to SCAN, the level begins scanning from the start value on the next count period. If it is changed to FIXED, the level returns to the start level immediately.
The step size may also be changed during a scan. The new step size will be in effect at the end of the next count period. The scan is not reset or paused.
B DISC LVL = -10.0 mV : -11.0 mV
The threshold level of B discriminator may be adjusted from -300 mV to + 300 mV. This level is the fixed threshold when B DISC is FIXED and the starting level when B DISC is SCANned.
When the B DISC mode is SCAN and a scan is in progress, this line displays the current position of the B DISC level. The cursor is inactive and replaced by a ':'. The spin knob cannot adjust the threshold while it is scanning. When the mode is SCAN, the starting level may only be adjusted when the counters are in reset.
T DISC SLOPE= RISE FALL
When T counter's input is INPUT 2, T discriminator may be triggered by rising or falling edges of the INPUT 2 signal. If the signal pulses are negative going, then the slope should be FALL. If the signal pulses are positive, then the slope should be RISE. In either case, the T DISC output is always active low; i.e., a discriminated rising edge of the signal will be output as a negative going NIM pulse at the T DISC output. The falling edges of the T DISC output are counted by counter T.
T DISC= FIXED SCAN ∆= -1.0 mV
The threshold level of T discriminator may be FIXED or SCANned. If the level is FIXED the right ha nd parameter is not displayed. If the level is scanned, then the right hand parameter is the step size. At the end of each count period in a scan, the level is incremented by one step size. The step size can be varied from -20.0 mV to +
20.0 mV. The range of a scan is N PERIODS times the step size. The starting level is set in the next menu line.
The scan mode may be changed during a scan. If it is changed to SCAN, the level begins scanning from the start value on the next count period. If it is changed to FIXED, the level returns to the start level immediately.
28
Loading...
+ 68 hidden pages