SPX APV DELTA Operating Manual

Page 1
Read and understand this manual prior to operating or servicing this product.
Operating Manual
DELTA
Inline Measuring Techniques
Page 2
Page 3
Table of Contents Page
2. Safety Instructions 2
3. Field of Application 2
4. Measuring Devices 3 - 4
5. Cleaning 4
6. Installation 4
7. Maintenance 5
UK
8. Materials 5
9. Dimesions / Weights 6
10. Technical Data 6
11. Assembly Instructions 7
12. Trouble Shooting 7
13. Spare Parts Lists RN 01.256
Inline Measuring Techniques Operating Manual Rev.0
1
Page 4
Page 5
1. General Terms
This operating manual has to be read carefully and observed by the competent operating and maintenance personnel.
We have to point out that we will not accept any liability for damage or malfunctions resulting from the non-compliance with this operating manual.
Descriptions and data given herein are subject to technical changes.
2. Safety Instructions
DANGER!
- The line system must be depressurized before any maintenance.
- Observe assembly instructions to ensure safe maintenance of the measuring device.
UK
- Moreover, the special safety instructions of the individual measuring devices (e.g.for electric connections) must be observed.
3. Field of Application
- The APV Inline housings with different measuring devices are used in applications in which hygienic process conditions must be provided by crevice-free sealing and reliable cleaning.
- Two different measuring devices can usually be used in one housing.
- With larger nominal diameters special housings can be equipped with 4 connections for measuring devices.
- For applications with increased line pressures special housing are available.
Inline Measuring Techniques Operating Manual Rev.0
2
Page 6
4. Measuring Device
All measuring devices have the same dimensions of flange connection and can, thus,be used independently of the nominal diameter.
Selection can be made from different measuring devices:
Temperature measurement
- Thermometer:
bimetal thermometer according to type sheet TM 52.01 type series R 5216 accuracy class: 1
- Temperature sensor: resistance thermometer PT 100 according to DIN/IEC 751 cl. B in twin core wiring electric terminal box out of polycarbonate protective type: IP 67
Pressure measurement
- Pressure gauge:
pressure transmitter for front flush stainless steel membrane radial output with different connecting directions filling: glycerine quality class: 1.0
UK
- Pressure measuring transducer:
pressure measuring transducer 141 GB with front flush stainless steel membrane with or without LCD temperature: + 120
o
C filling: Silicone oil protective type: IP 66
- Pressure sensor pressure transmitter type CERABAR with ceramic measuring cells and potential-free analog output, pressure transmitter with front flush stainless steel membrane filling: Silicone oil quality class: 0.2 protective type: IP 65
Flow measurement
- Flow controller immersion sensor ST 74614 according to calorimetric measuring principle measuring range: 1 ...150 cm/s different analysis devices protective type: IP 67
Conductivity measurement
- Conductivitiy transmitter LMIT 08:
compact device with electrodeless measuring cell out of Teflon, different measuring range change-overs with and without LCD protective type: IP 67
Inline Measuring Techniques
3
Operating Manual
Rev.0
Page 7
4. Measuring Device
Optical measuring technique:
turbidity measurement colour measurement concentration measurement
Ultrasonic measuring technique:
concentration measurement density measurement
Sight glass:
with and without illumination.
5. Cleaning
As a result of the design of the Inline housing being free of dead spaces (no sump and no dome) all product-wetted parts can be cleaned properly during the cleaning process.
UK
6. Installation
The housing must be installed in such a way that a faultless function of the measuring device is given.Moreover, the installation must allow for fluids to drain off the housing.
With horizontal installation, measuring devices must be installed vertically.
Attention!
- Observe welding instructions!
6.1 Welding Instructions
Before welding, the measuring devices and the housing cover must be removed. See to a careful handling to avoid damage.
- Welding may only be carried out by certified welders (EN 287-1). (Seam quality EN 25817 „B“).
- The welding of the valve housings must be effected in such a way that deformation strain cannot be transfered to the valve body.
Inline Measuring Techniques Operating Manual Rev.0
4
Page 8
6.1 Welding Instructions
- The preparation of the weld seam up to 3 mm thickness must be carried out in butt manner as an square butt joint without air. (Consider shrinkage!)
- TIG orbital welding should be aimed at!
- After welding of the valve housings, the pipelines must be cleaned from welding residues and soiling in order to prevent other components from being damaged.
- Any damage resulting from the nonobservance of these welding instructions is not subject to our guarantee.
7. Maintenance
- The maintenance intervals depend on the individual application and are to be determined by the operator himself by temporary checks.
UK
- Installation of measuring devices see assembly instructions.
- All seals must be slightly greased before their installation.
Recommendation: APV food grade grease for EPDM,HNBR and FPM
(0,75 kg /tin - ref.-No.000 70-01-019/93) (60 g /tube - ref.-No.000 70-01-018/93)
! For all applications,use only those greases which are suited for the
corresponding seal material !
8. Materials
Product-wetted parts : 1.4571 / 1.4404 Seals : standard EPDM
option HNBR, FPM
Measuring devices : see technical documents
of manufacturer or item 4.
Inline Measuring Techniques
5
Operating Manual
Rev.0
Page 9
9. Dimensions / Weights
DN
A
˘˘˘
C
D
weight in
kg
2568262925,5
4067384131,5
5072505337,5
6585667045,5
8098818553,0
100
111
100
104
62,5
125
130
125
129
75,0
150
150
150
154
87,5
UK
Inline Measuring Techniques Operating Manual Rev.0
10. Technical Data
max. line pressure : 10 bar
(higher pressure on request)
max. operating temperature :140oC EPDM, HNBR
135o C FPM
sterilization temperature :150o C EPDM, HNBR (short-term) : 140o C FPM
For information about the measuring devices refer to the technical documents of the manufacturers.
6
Page 10
11. Assembly Instructions
The item numbers refer to the spare parts list RN 01.256.
11.1 Dismantling from the line system
a. Shut off line pressure in the product line and discharge line if
possible.
b. Disconnect electric and pneumatic connecting lines. c. Remove hex. screws (3).Screw two screws into the threaded holes
of the housing cover (2), thus,pressing the measuring device and the housing cover off.
11.2 Dismantling of wear parts
- If the measuring device has additional wear parts, the corresponding operating instruction must be observed.
a. Pull off housing seal (4).
UK
11.3 Installation of seals and measuring devices
a. Pull the slightly greased seal on the flange of the measuring
device or of the housing cover (2).
See to a correct fit of the seal.
b. Place the measuring device or housing cover straightly on the
housing flange and tighten it by the hex. screws (3).
c. Connect electric and pneumatic lines.
12. Trouble Shooting
- leakage between housing flange and flange of measuring device or housing cover : replace housing seal.
13. Spare Parts Lists
(see annex)
Inline Measuring Techniques
7
Operating Manual
Rev.0
Page 11
Page 12
Your local contact:
APV Zechenstraße 49 D-59425 Unna
Phone: +49(0) 23 03/ 108-0 Fax: +49(0) 23 03 / 108-210
For more information about our worldwide locations, approvals, certifications, and local representatives, please visit www.apv.com.
Copyright © 2008 SPX Corporation
The information contained in this document, including any specifications and other product details, are subject to change without notice. While we have taken care to ensure the information is accurate at the time of going to press, we assume no responsibility for errors or omissions nor for any damages resulting from the use of the information contained herein.
UK
Translation of original manual
rev. 0
BA INLM 000002 ID-No.: H175760
Page 13
Page 14
Page 15
Page 16
Page 17
Page 18
Page 19
Page 20
Page 21
Page 22
Page 23
Loading...