SPE ware CEREX System-48-II Operating Instructions Manual

OPERATING INSTRUCTIONS
CEREX PRESSURE PROCESSOR
System
48-II
INSTALLATION
CAUTION
The SPE Processor does not include a filter on the source gas input. A clean, oil-free gas source is essential to prevent sample contamination.
Compressed nitrogen is the recommended pressure source in order to assure inert atmosphere operation. If this is not a consideration, the unit can operate using high purity filtered instrument air.
This instrument is supplied with an installation kit which includes 6 feet of 1/8” i.d. (1/4” o.d.) polyethylene tubing and connectors to 1/8” NPT or 1/4” NPT.
Using the tubing and fittings provided connect the unit to a regulated oil free gas source.
Optimum Source Pressure is 80 psig.
Min Pressure is 50 psig
Max Pressure is 100 psig
SPE RACK CONFIGURATIONS and USE
The System 48-II Pressure Processor provides the means to batch process solid phase extraction columns in the 48-position format. The CEREX column racks interface to several collection tube racks which are also in the 48-position format. Depending upon the volume of the effluent, the appropriate collection rack can be selected from the ancillary products available with the Pressure Processor.
The processor collection tube racks are optimally designed for 75mm or 125mm high collection tubes. Ancillary collection racks are available to accommodate common collection tube sizes (Auto sampler vial racks, 10x75, 12x75, 13x100, 13 x 125, 16 x 100 and 16x125). Contact customer service for further details and availability of other sizes.
Standard SPE C olum n & Collection Tube Combinations...
10
x
75
1cc
12
x
75
13
x
100
16
x
100
3cc 6cc
12
10
x
75
x
75
13
x
100
16
x
100
12
10
x
x
75
75
13
x
100
16
x
100
The waste bin and tube racks are physically keyed to the platform by three elevated pins on the platform that recess into the bottom of the rack. When installed appropriately, the tub and rack should sit level on the platform.
System 48 column racks are keyed to the collection racks so that sample position is preserved throughout the SPE sequence. The column rack is then positioned directly on top of the waste bin or collection tray. Note that the When appropriately placed, the stack should sit level on the platform. For each extraction step, slide the platform toward the back until it stops on the guides.
All collection racks are compatible with the system 48 sample concentrator, allowing for samples to be placed in the concentrator without the timely transfer of samples from rack to rack.
Loading...
+ 7 hidden pages