Sper scientific AquaShock Bluetooth pH Kit 850057K User Manual

Page 1
AquaShock®
Bluetooth®
pH Kit
850057K
Page 2
2 3
AquaShock® Bluetooth® pH Kit 850057K
ALL RIGHTS RESERVED
Printed in the USA
The contents of this manual may not be reproduced or transmitted in any form or by any means electronic, mechanical, or other means that do not yet exist or may be developed, including photocopying, recording, or any information storage and retrieval
system without the express permission from Sper Scientic.
EPA Equivalency Method (with Glass Probe): 150.1
FCC ID: 2AK5U-1300
This device and enclosed comply with part 15 of the FCC Rules.
Operation is subject to the two following conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may
cause undesired operation.
Sper Scientic
8281 E. Evans Rd., Suite #103, Scottsdale, AZ 85260
Tel: (480) 948-4448
Fax: (480) 967-8736
Web: www.sperscientic.com
Page 3
3
INTRODUCTION .....................................4
FEATURES..........................................5
MATERIALS SUPPLIED................................5
LCD DISPLAY .......................................6
POWER SUPPLY .....................................7
KEYPAD ............................................8
SETUP MODE .......................................9
CALIBRATION ......................................11
MEASUREMENT PROCEDURES .......................12
BLUETOOTH® DATALOGGING .........................14
OPTIONAL ACCESSORIES ...........................14
SPECIFICATIONS ...................................15
WARRANTY .......................................16
TABLE OF CONTENTS
Page 4
4 5
INTRODUCTION
This Sper Scientic AquaShock® Bluetooth® pH Meter
(850057K) is the industry’s leading cross-over meter that
brings lab quality capability to the portable meter industry. The AquaShock® line can withstand rigorous use, oats, and can be completely submerged in up to a meter of water. The fast calibration calibrates up to ve-points with auto-recognition technology. Data can be stored in the meter and later streamed
to the included software for graphing, manipulation, and sharing
via Bluetooth® technology. It can also stream in real-time allowing users to wirelessly collect data from up to 100 feet away. Data gathered with the AquaShock® is precise, fast, and secure making it possible for the same meter to be used in multiple
applications including laboratory settings, manufacturing, and
environmental testing.
Page 5
5
FEATURES
• Floats
• Shockproof
• IP67 Waterproof
• Bluetooth® streaming up to 100 feet
• Protective soft grip outer layer
• Rechargeable Lithium Ion battery
• Also accepts standard BNC probes (Waterproof seal is compromised with non-AquaShock® probes)
• Multi-line display LCD screen
• Reads pH and mV
• Automatic or manual temperature compensation
• Automatic bu󰀨er recognition
• Celsius or Fahrenheit selectable
• 99 data points for analysis
• Calibration point review
• Internal clock and calendar
• Maximum, minimum and average
• Hold function
• Low battery indicator
• Backlight
MATERIALS SUPPLIED
• Meter
• Bluetooth® Dongle
• Dual pH/ATC Probe
• pH Bu󰀨ers, Distilled Water
• 1 Lithium 1450 mAh 7.4 V Battery
• DC Power Adapter
• Instruction Manual
• Hard Carrying Case
Page 6
6 7
Name Description
1 Ready Displayed when measuring
results are stable
2 Electrode
Durability Icon
Current electrode durability
3 Scale Unit Current measurement scale unit 4 Measuring Results Current measurement value 5 ATC/MTC Automatic/Manual Temperature
Compensation indicator
6 Temperature Scale Current temperature
measurement unit
7 Temperature Value Current temperature
measurement value
8 Battery Icon Current battery power level
1
2
3
4
5
6
7
8
LCD DISPLAY
Page 7
7
Battery voltage is more than 7.7V shows full.
Battery Power Icon:
Battery voltage between 7.2V and 7.7V shows 2 cells.
Battery voltage between 6.6V and 7.2V shows 1 cell.
Battery voltage less than 6.6V shows 0 cells and ashes.
Display showing battery charging from
less than 6.6V to fully charged.
More than 90% life remaining
More than 85– 90% life remaining
More than 80- 85% life remaining
Less than 80% life remaining
Probe Life Icon:
LCD DISPLAY
POWER SUPPLY
This meter is powered by one rechargeable Lithium 7.4V (1450 mAh) battery. The DC power adapter can be used to charge the battery even when the meter is turned o󰀨. This meter includes a
charge protection function and will automatically stop charging
when the battery is full. A complete re-charge takes approximately 8 hours.
Page 8
8 9
Name 1st Function 2nd Function
1 Power/Backlight/
Low Battery
Power on/o󰀨 Backlight on/o󰀨
2 Setup Setting parameters Conrm setting
3 MI↑ Store data
4 MR↓ Recall data
5 Max/Min/Ave Max/Min/Ave
6 Range/Temp Switch measuring
scale
Set temperature value
7 Hold/Exit Hold Exit
8 CAL Calibration
1
3
5
7
4
6
8
2
KEYPAD
Page 9
9
SETUP MODE
The Setup Mode allows you to customize the following meter preferences and defaults:
• Temperature Units
• Clock Setting
• Restore Factory Defaults
• View Calibration Data
• View Probe Parameters
1. Press POWER for 1 second to turn the meter on before
performing any setup function.
Temperature Units
1. Press SETUP to enter the Setup Mode.
2. Press SETUP to enter the Set Temperature screen.
3. The unit °C or °F will ash on the LCD.
4. Press or to select the °C or °F temperature unit.
5. Press SETUP to save and return to Normal Mode.
6. Press EXIT to return to Normal Mode without saving the selection.
Real Time Clock
1. Press SETUP to enter the Setup Mode.
2. Press or to select the Set Clock screen.
3. Press SETUP to set the year. The value will ash on the LCD.
4. Press to increase the value by 1. Hold down to continually increase the value.
Press ↓ to decrease the value by 1. Hold down to continually decrease the value.
Press SETUP to save and the meter automatically
moves to the month setting.
5. Repeat step 4 to adjust the month, day, hour, and minute.
6. Press EXIT to return to Normal Mode without saving.
Note...
This procedure adjusts the meter’s internal clock. The internal clock will function when the meter is turned o󰀨 but only when the battery is installed.
Page 10
10 11
Restore Factory Settings
1. Press SETUP to enter the Setup Mode.
2. Press or to select the Restore Factory Settings screen.
3. Press SETUP to conrm selection. “NO” will ash on the LCD.
4. Press or to select “YES” or “NO.”
5. When “YES” is displayed, press SETUP to conrm and all parameters will restore to factory settings (except the clock) and the meter returns to Normal Mode.
6. Press EXIT to return to Normal Mode without saving.
View Calibration Data
1. Press SETUP to enter the Setup Mode.
2. Press or to select the View CAL Data screen.
3. Press SETUP to enter the mode.
4. Press to view the previous calibration point. Press ↑ to view the next calibration point. The date and time is on cyclical display.
5. Press EXIT to return to Normal Mode.
View Probe Parameters
1. Press SETUP to enter the Setup Mode.
2. Press or to select the View Probe Parameters screen.
Note...
If the meter is not calibrated, the screen will show:
Press SETUP or EXIT to
return to Normal Mode.
Page 11
11
CALIBRATION
Calibration should be performed whenever the probe has been
out of use for 48 hours or longer. To prepare for calibration:
Calibrate With Temperature Sensor/ATC pH Probe
1. Press CAL for 2 seconds to enter Calibration Mode.
2. The meter will automatically recognize the calibration bu󰀨er. The value and “Finished” will display on the LCD when the procedure is complete.
3. Press EXIT to return to Normal Mode.
4. Repeat Steps 1-2 to calibrate other points.
Note...
All other keys will be inoperable while using the calibration function.
Calibrate Without Temperature Sensor/MTC pH Probe
1. Press CAL for 2 seconds to enter Temperature Setup Mode. The temperature value will ash on the LCD.
2. Press to increase the value. Hold down to continually increase the value.
Press ↓ to decrease the value. Hold down ↓ to continually decrease the value.
For °C, the range is 0 to 60.0°C; for °F, the range is 32 to 140.0°F.
Press SETUP and the meter will automatically recognize the
calibration bu󰀨er. The value and “Finished” will display on the LCD when the procedure is complete.
Repeat Step 2 to calibrate other points.
3. Press EXIT to return to Normal Mode.
Page 12
12 13
MEASUREMENT PROCEDURES
Turning the Unit On/O󰀨
1. Press POWER to turn the meter on.
2. Press and hold POWER for 2 seconds to turn the meter o󰀨.
Backlight
1. With the meter on press POWER to turn the backlight on/o󰀨.
Scale Selection
1. Press RANGE to switch between the pH and mV scales.
Setting the MTC Temperature
Do NOT connect the temperature sensor or non-ATC probe
while setting the temperature.
1. From Normal Mode, press RANGE for 2 seconds to enter Set Temperature Mode. The temperature value will ash on the LCD.
2. Press to increase the value. Press and hold to increase the value continuously.
Press ↓ to decrease the value. Press and hold ↓ to decrease the value continuously.
For °C, the range is 0 to 60.0°C; for °F, the range is 32 to 140.0°F.
3. Press SETUP to save and the meter returns to Normal Mode.
4. Press EXIT to return to Normal Mode without saving.
Saving to Memory
1. Press MI to save the current value and view the data point number. There are 99 data points available.
Page 13
13
Recall Memory
1. Press RANGE to select the scale (pH or mV) for recall.
2. Press MR to enter Recall Mode and view the last measured value. The date and time are on cyclical display.
3. Press or to scroll saved data points.
4. Press EXIT to return to Normal Mode.
Clear Memory
1. Press MI and MR simultaneously for 2 seconds to clear all memory. The meter will return to Normal Mode.
Maximum, Minimum, and Average
1. Press RANGE to select the scale.
2. Press MAX/MIN/AVE, the maximum recorded value will appear on the LCD. The date and time is on cyclical display.
3. Press MAX/MIN/AVE, the minimum recorded value will appear on the LCD. The date and time is on cyclical display.
4. Press MAX/MIN/AVE, the average recorded value will appear on the LCD. The date and time is on cyclical display.
5. Press MAX/MIN/AVE to return to Normal Mode.
Hold
1. Press HOLD to hold the value.
2. Press HOLD to release the hold value.
Note...
When in Hold Mode, all other functions will be disabled
except turning the meter o󰀨, backlight function, saving, and exiting Hold Mode.
Page 14
14 15
BLUETOOTH® DATALOGGING WITH SOFTWARE
Software Installation
Note…
Di󰀨erent computer systems may require slightly di󰀨erent installation steps than those below.
First-Time Setup of meter and Software:
1. Download the software to your computer from our website Go to www.sperdirect.com/software.htm and nd your meter or search for you meter (850057K) and download from product page. The Software is compatible with Windows XP through
Windows 10, including all versions in between.
2. Insert the included Bluetooth® Dongle (See image 1) into any USB port. The
red light on the dongle should begin to
ash within a few seconds.
3. Turn the meter on and Press the yellow SETUP button to enter SETUP mode.
4. Using the UP and DOWN arrows, scroll
through the functions until BLUETOOTH
Setting Appears on the LCD screen.
5. Press the yellow SETUP button to enter this feature. Use the UP and DOWN arrows until YES appears on the screen. Press
the yellow SETUP button once to conrm. The meter will return to normal measuring mode and you will see the Bluetooth®
icon in the lower left hand corner of the LCD screen.
6. Enter the control panel on your computer and click “Add a Bluetooth Device.” Double-Click the AquaShock_pH device from the Bluetooth choices.
7. A pop-up window will appear, prompting for a passcode. Enter
the number 1234.
8. Once the unit is paired, it will automatically recognize the device on future log-ins from the same computer.
Bluetooth® Dongle
Page 15
15
General Software Use:
1. Open the Software and click CONNECT in the upper left hand corner (See Image 2, below).
2. A pop-up window will open. Choose your desired meter from the AquaShock choices and click “OK.”
Note...
All connected Bluetooth devices will show in the popup window,
including non-AquaShock® devices.
3. The software will now display the Workstation (See image 3).
Page 16
16 17
For Live Streaming
1. Click the Start Button in the upper right hand corner. A live
graph will appear showing both the pH value and the mV
value from the meter (See image 4). Press STOP to stop the live stream. The data can now be Exported to Excel.
Downloading Internal Memory Points
1. Click the LOAD data button on the upper left hand side of
the Workstation.
2. All Internal memory points gathered through the “MI” function will load in the lower left quadrant of the Workstation. See Image 5.
Page 17
17
Exporting Data
Note...
The software will only export the most recent data that has been
streamed or loaded from internal memory. If you are downloading
both a stream and a set of internal memory points, you must
export them individually.
1. Click the LOAD data button on the upper left hand side
of the Workstation.
2. Choose your save location from the Save As popup window.
3. Your data is now available as an Excel le.
Page 18
18 19
OPTIONAL ACCESSORIES
850057P Replacement AquaShock® ATC pH Probe
850057PG AquaShock® Rellable ATC Glass pH Probe
840016 Standard pH Probe (non-ATC)
840049 Spear Tip pH Probe (non-ATC)
840093 Field Tripod
860008 pH 4 Bu󰀨er Solution - 3 bottles of 40ml each
860009 pH 7 Bu󰀨er Solution - 3 bottles of 40ml each
860010 pH 10 Bu󰀨er Solution - 3 bottles of 40ml each
860011 Deionized Water - 3 bottles of 40ml each
860012 pH Bu󰀨er Set - 1 each pH 4, 7, 10
Page 19
19
Power O󰀨 Light o󰀨
Power On Green light for 5 seconds, then light o󰀨 Low Battery Red light ashes every 5 seconds Charging Battery
Blue light on
Mode pH Temperature mV Range 0 to 14 pH 32 to 140°F
(0 to 60°C)
-1999 to
1999 mV Resolution 0.01 pH 0.1°F (0.1°C) 0.1 mV Accuracy ±0.01 pH ±0.9°F (±0.5°C) ±0.3 mV Input BNC
Connector
Temperature Probe Connector
BNC
Connector Calibration
Up to 5 points
Memory
Up to 99 data points
pH Slope Range
75% to 110%
Temp Compensation
Automatic (ATC) or manual (MTC) from
32 to 140°F or 0 to 60°C
Operating Temp
32 to 122°F (0 to 50°C)
Battery Lithium battery 7.4 V (1450 mA/h) Battery Life
> 200 hours
Bluetooth® Range
100 feet
Power Button Indicator Light
SPECIFICATIONS
Page 20
WARRANTY
Sper Scientic warrants this product against defects in materials and workmanship for a period of ve (5) years from the date
of purchase, and agrees to repair or replace any defective unit
without charge. If your model has since been discontinued, an equivalent Sper Scientic product will be substituted if available. This warranty does not cover probes, batteries, battery leakage,
or damage resulting from accident, tampering, misuse, or abuse of
the product. Opening the meter to expose its electronics will break the waterproof seal and void the warranty.
To obtain warranty service, ship the unit postage prepaid to:
SPER SCIENTIFIC LTD.
8281 East Evans Road, Suite #103
Scottsdale, AZ 85260
The defective unit must be accompanied by a description of the
problem and your return address. Register your product online at www.sperwarranty.com within 10 days of purchase.
Rev 2/22/2018