Sper scientific AquaShock 850046K User Manual

Page 1
AquaShock® DO Kit
850046K
Page 2
2 3
AquaShock® DO Kit 850046K
Copyright ©2015 by Sper Scientic
ALL RIGHTS RESERVED
The contents of this manual may not be reproduced or transmitted in any form or by any means electronic, mechanical, or other means that do not yet exist or may be developed, including photocopying, recording, or any information storage and retrieval
system without the express permission from Sper Scientic.
Sper Scientic 8281 E. Evans Rd., Suite #103, Scottsdale, AZ 85260 Tel: (480) 948-4448 Fax: (480) 967-8736 Web: www.sperscientic.com
Page 3
3
INTRODUCTION .....................................4
FEATUR ES..........................................5
MATERIALS SUPPLIED................................5
OPTIONAL ACCESSORIES ............................5
LCD DISPLAY .......................................6
POWER SUPPLY .....................................7
KEYPAD ............................................8
SETUP MODE .......................................9
PROBE SETUP .....................................17
CALIBRATION ......................................18
MEASUREMENT PROCEDURES .......................23
SPECIFICATIONS ...................................27
WARRANTY .......................................28
TABLE OF CONTENTS
Page 4
4 5
INTRODUCTION
The Sper Scientic AquaShock® DO Meter is the most waterproof, rugged and reliable DO meter on the market. It is fully waterproof, including probe connections. As with all AquaShock® meters, the DO Meter is completely shockproof, oats, and features a double
injection molded case with an integral protective soft grip outer
layer, high end antishock, LED backlight display, and is \powered
by the latest environmentally friendly, long lasting, rechargeable
Lithium Ion battery. The meter is highly accurate and stable, very intuitive and includes all of the functions required by most users. Sper Scientic guarantees you will not nd any meters more rugged and reliable than AquaShock®.
Page 5
5
FEATURES
• Floats
• Shockproof
• IP67 Waterproof
• Protective soft grip outer layer
• Rechargeable Lithium Ion battery
• Multi-line display LCD screen
• Reads DO (Dissolved Oxygen)
• Auto Ranging
• Selectable Automatic or Manual Temperature Compensation
• Celsius or Fahrenheit selectable
• 99 data points per scale for analysis
• Calibration point review
• Internal clock and calendar
• Maximum, minimum and average
• Hold function
• Low battery indicator
• Backlight
MATERIALS SUPPLIED
• Meter
• Waterproof Probe Input Covers
• Dissolved Oxygen/ATC Temperature Probe
• Electrolyte
• 1 Lithium 7.4 V Battery (1450 mAh)
• DC Power Adapter
• Instruction Manual
• Hard Carrying Case
• 840093 Field Tripod
850046P Replacement AquaShock® DO Probe 10’ Cable
850046P25 Replacement AquaShock® DO Probe 25’ Cable
850046P50 Replacement AquaShock® DO Probe 50’ Cable
OPTIONAL ACCESSORIES
Page 6
6 7
LCD DISPLAY
Name Description
1 Ready Displayed when measuring
results are stable 2 Scale Unit Current measurement scale unit 3 Measuring Results Current measurement value 4 ATC/MTC Automatic/Manual Temperature
Compensation indicator 5 Temperature Scale Current temperature measurement unit 6 Temperature Value Current temperature
measurement value 7 Battery Icon Current battery power level
1 2
3
4
5
6
7
Page 7
7
Battery voltage is more than 7.7V shows full.
Battery Power Icon:
Battery voltage between 7.2V and 7.7V shows 2 cells.
Battery voltage between 6.6V and 7.2V shows 1 cell.
Battery voltage less than 6.6V shows 0 cells and ashes.
Display showing battery charging from
less than 6.6V to fully charged.
More than 90% life remaining
More than 85– 90% life remaining
More than 80- 85% life remaining
Less than 80% life remaining
Probe Life Icon:
LCD DISPLAY
POWER SUPPLY
This meter is powered by one rechargeable Lithium 7.4V (1450 mAh) battery. The DC power adapter can be used to charge the battery even when the meter is turned o󰀨. This meter includes a
charge protection function and will automatically stop charging
when the battery is full. A complete re-charge takes approximately 8 hours.
Page 8
8 9
DO Meter
1
3
5
7
4
6
8
2
KEYPAD
Name 1st Function 2nd Function
1 Power/Backlight/
Low Battery
Power on/o󰀨 Backlight on/o󰀨
2 Setup Setting parameters Conrm setting
3 MI↑ Store data
4 MR↓ Recall data
5 Max/Min/Ave Max/Min/Ave
6 Range/Temp Switch measuring
scale
Set temperature value
7 Hold/Exit Hold Exit
8 CAL Calibration
Page 9
9
SETUP MODE
The Setup Mode allows you to customize the following meter preferences and defaults:
• Temperature Scale
• Clock Setting
• Restore Factory Defaults
• MTC (Manual Temperature Compensation)
Temperature Coe󰀩cient
• View Calibration Points
• View Probe Parameters
• Barometric Pressure Parameters
• RH Parameters
• Salinity Compensation
• Oxygen Content (mg/L (ppm)) Unit
• % O󰀨set Setting
Press POWER to turn the meter on before performing
any setup function.
Temperature Units
1. Press SETUP to enter the Setup Mode.
2. Press SETUP to enter the Set Temperature screen.
3. The unit °C or °F will ash on the LCD.
4. Press or to select the °C or °F temperature unit.
5. Press SETUP to save and return to Normal Mode.
6. Press EXIT to return to Normal Mode without saving the selection.
Note…
When the temperature reading is outside the range, the display will
show “HHH” (high) or “LLL” (low).
Real Time Clock
1. Press SETUP to enter the Setup Mode.
2. Press or to select the Set Clock screen (see Fig. A).
3. Press SETUP to set the year (see Fig. B). The value will ash
on the LCD.
Page 10
4. Press to increase the value by 1. Hold down to continually increase the value.
Press to decrease the value by 1. Hold down to continually decrease the value.
Press SETUP to save and the meter automatically
moves to the month setting (see Fig. C.).
5. Repeat step 4 to adjust the month, day, hour, and minute (see Fig. D, E & F).
6. Press EXIT to return to Normal Mode without saving.
Note...
This procedure adjusts the meter’s internal clock which is a 24 hour format only. The internal clock will function when the meter is turned o󰀨 if the battery is installed.
Fig. A
Fig. D
Fig. B
Fig. E
Fig. C
Fig. F
Page 11
11
Restore Factory Settings
1. Press SETUP to enter the Setup Mode.
2. Press ↑ or ↓ to select the Restore Factory Settings screen (see Fig. A).
3. Press SETUP to conrm selection. “NO” will ash on the LCD. (see Fig. B).
4. Press or to select “YES” or “NO”. (see Fig. C).
5. When “YES” is displayed, press SETUP to conrm and all parameters will restore to factory settings (except the clock) and the meter returns to Normal Mode.
6. Press EXIT to return to Normal Mode without saving.
Fig. A Fig. B
Fig. C
Factory Defaults
Parameter Name Parameter Value
Date 2011-1-1 0:0
CAL Standard
CAL Point
O󰀨set
Temperature 77.0°F (25.0°C)
Temperature Unit °F
Bar. Unit mmHg
Bar. Value 760
% O󰀨set Value 0.0ppt
mg/L (ppm) 0
% saturation 0
RH 100%
Memory
Page 12
Set MTC (Manual Temperature Compensation)
Temperature Coe󰀩cient
Note...
Only perform this setting when ATC Temperature Probe
is not available.
1. Press RANGE for 2 seconds to enter the Set
Temperature Mode (see Fig. A). The temperature value will ash (see Fig. B).
2. Press ↑ to increase the Temperature Value. Hold down ↑ to continually increase the value.
Press ↓ to reduce the Temperature Value. Hold down ↓ to continually reduce the value.
If the temperature unit is °C, the range for adjustment is 0.0 to 50.0°C.
If the temperature unit is °F, the range for adjustment is 32.0 to 122.0°F.
3. Press SETUP to conrm change and return to Normal Mode (see Fig. D).
Press EXIT to return to Normal Mode without saving change
(see Fig. C).
Fig. A Fig. B
Fig. D
Fig. C
Page 13
13
View the Probe Data
1. Press SETUP once to enter the Set up mode.
2. Press ↑ or ↓ to select the View Probe screen (see Fig. A).
3. Press SETUP once to enter the View Slope screen which shows the electrode slope (see Fig B).
4. Continue to press ↑ or ↓ to view % O󰀨set Value, 100%
O󰀨set nA Value, 0% O󰀨set nA Value screens (see Fig. C, Fig. D & Fig. E).
5. Press SETUP or EXIT to return to Normal Mode.
Fig. A
Fig. D
Fig. B
Fig. E
Fig. C
Page 14
14 15
Set Barometric Pressure Parameters
1. Press SETUP to enter the Setup Mode.
2. Press ↑ or ↓ until the Bar. Pressure screen is seen (see Fig. A).
3.
SETUP to enter the Select Bar. Unit screen (see Fig. B or Fig. C).
4. Press ↑ or ↓ to scroll through the available barometric
measurement scales.
5. Press SETUP to save the displayed measurement scale and enter the Set Bar. Value screen (see Fig. D or Fig. E). The current Bar. Value number will ash.
6. Press ↑ to increase the Bar. Value.
Hold down ↑ to continually increase the value.
7. Press ↓ to reduce the Bar. Value.
Hold down ↓ to continually reduce the value.
If the Bar. Value Measurement scale is mmHg the available range is 500 to 1499.
If the Bar. Value measurement scale is kPa the available range is 66.6 to 199.9
8. Press SETUP to save the changes and return to Normal Mode.
9. Press EXIT to return to Normal Mode without saving changes.
Fig. A Fig. B
Fig. C
Fig. D
Page 15
15
Set RH Parameters
1. Press SETUP to enter the Setup mode.
2. Press ↑ or ↓ to display the Set RH screen (see Fig. A)
3. Press SETUP to enter the Set RH screen (see Fig. B). The current set value will ash.
4. Press ↑ to increase the RH Value.
Hold down ↑ to continually increase the value. Press ↓ to reduce the RH Value. Hold down ↓ to continually reduce the value. RH Value range may be set from 0 to 100%.
5. Press SETUP to save the change and return to Normal Mode.
6. Press EXIT to return to Normal Mode without saving change.
Fig. E
Fig. A Fig. B
Page 16
Set Salinity Compensation
1. Press RANGE to select mg/L (ppm) measurement scale.
2. Press SETUP to enter the Setup Mode.
3. Press ↑ or ↓ to display the Set Salinity screen (see Fig. A)
4. Press SETUP to enter the Set SAL. Value screen (see Fig. B).
5. Press ↑ to increase the SAL. Value. Hold down ↑ to continually increase the value. Press ↓ to reduce the SAL. Value. Hold down ↓ to continually reduce the value. Salinity compensation value range is 0.0 to 50.0 ppt.
6. Press SETUP to save the value and return to Normal Mode.
7.
Press EXIT to return to Normal Mode without saving the change.
Fig. A Fig. B
Set Oxygen Content (mg/L (ppm)) Unit of Measure
1. Press RANGE to select mg/L (ppm).
2. Press SETUP to enter the Setup Mode.
3. Press ↑ or ↓ to display the Select Unit screen (see Fig. A).
4. Press SETUP to enter the Select Unit screen (see Fig. B).
5. Press ↑ or ↓ to select either mg/L or ppm scale.
6. Press SETUP to save the change and return to Normal Mode.
7.
Press EXIT to return to Normal Mode without saving the change.
Fig. A Fig. B
Page 17
17
Set % O󰀨set
1. Press RANGE to select % measuring scale.
2. Press SETUP to enter the Setup Mode.
3. Press ↑ or ↓ to display Set % O󰀨set screen (see Fig. A).
4. Press SETUP to enter the Set % O󰀨set mode.
The current value will ash (see Fig. B).
5. Press ↑ to increase the % O󰀨set Value.
Hold down ↑ to continually increase the value. Press ↓ to reduce the % O󰀨set Value. Hold down ↓ to continually reduce the value. The % O󰀨set range is ±10.0. It cannot exceed 0.0 to199.9%.
6. Press SETUP to save the changes and return to Normal Mode.
Press EXIT to return Normal Mode without saving changes.
Fig. A Fig. B
PROBE SETUP
IMPORTANT: THIS DO PROBE WILL NOT FUNCTION UNTIL ELECTROLYTE IS ADDED
Before using the probe please follow the instructions, below.
1. Remove the protective cap from the DO probe by unscrewing it.
2. Fill the cap with the included electrolyte solution. The cap
should be at least 75% full.
3. Securely screw the cap back onto the probe. Wipe any excess
electrolyte from the outside of the probe.
NOTE...
It is better to have excess electrolyte in the probe’s cap than not enough. Once the cap is secure, excess electrolyte can be wiped away with DI water to prevent crystallization.
Page 18
18 19
4. Recalibrate the probe using a DO meter.
This procedure can also be used when replacing the membrane
caps. Extra caps are included with this probe because they may need to be replaced occasionally.
Fig. A
CALIBRATION
Note...
While performing calibration, if the measurement value reads from
10% to 50% the display will show “Error” as in Fig. A. Check probe for damage. Repair or replace probe.
Press EXIT to return to Normal Mode.
All other keys except POWER ON/OFF and HOLD are inoperable while performing calibration.
Page 19
19
100% Calibration
Note...
Only when the measurement value is above 50% can the meter be
calibrated at 100%. All other keys except POWER ON/OFF and HOLD are inoperable while performing calibration.
1. Clean the probe thoroughly with deionized water.
2. Press RANGE to select the % saturation scale (see Fig. A).
3. Wrap the probe with a wet cloth being sure not to touch the probe’s oxygen permeation membrane.
4. Press CAL for 2 seconds to enter the calibration mode
(see Fig. B).
5. When calibration is complete the display will read “Finished” (see Fig. C).
6. Press EXIT to return to Normal Mode (see Fig. D).
Fig. A
Fig. D
Fig. B
Fig. C
Page 20
20 21
Fig. A Fig. B
Fig. C Fig. D
0% Calibration
Note...
The measurement value must read below 10% to calibrate the
meter at 0%. All other keys except POWER ON/OFF and HOLD are inoperable while performing calibration.
1. Clean the probe thoroughly with deionized water.
2. Press RANGE to select the % saturation scale (see Fig. A).
3. Place the probe into 0 Oxygen Solution. The solution should have a ow rate of at least 0.3m/s
4. Press CAL for 2 seconds to enter the calibration mode
(see Fig. B)
5. Once the reading is steady the display will read “Finished” (see Fig. C).
6. Press EXIT to return to Normal Mode (see Fig. D).
Page 21
21
Calibrate dissolved oxygen in mg/L (ppm) mode
Note...
mg/L (ppm) calibration can only be performed after 100% calibration
has been completed.
1. Clean the probe thoroughly with deionized water.
2. Put the probe into a sample where the oxygen content is known. The solution should have a ow rate of at least 0.3m/s.
3. Press RANGE to select the mg/L(ppm) scale (see Fig. A).
4. Press CAL to enter calibration mode and begin calibration.
The measurement value will ash (see Fig. B).
5. Press ↑ to increase the Measurement Value. Hold down ↑ to continually increase the value. Press ↓ to reduce the Measurement Value. Hold down ↓ to continually reduce the value. The Measurement Value range is 0.00 to 19.99mg/L.
6. Press SETUP to save the calibration value and exit to
Normal Mode (see Fig. C).
7. Press EXIT to return to Normal Mode without saving.
Fig. A Fig. B
Fig. C
Page 22
22 23
View Calibration Points
1. Press RANGE chose the scale you want view.
2. Press SETUP to enter the Setup mode.
3. Press ↑ or ↓ to select the View Cal Data screen (see Fig. A).
4. Press SETUP to enter the mode. (The 100% scale will display
as in Fig. B and Fig. C. The 0% scale will display as in Fig. E and Fig. F). The date and time is a cyclic display.
If there are no calibration points set the display will
show “- - - -” (see Fig. D).
5. Press ↑ once to view the previous Cal point. Repeat to view all previous Cal points. Press ↓ once to view the next Cal point. Repeat to view all additional Cal points.
6. Press SETUP or EXIT to return to Normal Mode.
Fig. A Fig. B
Fig. C Fig. D
Page 23
23
Fig. E Fig. F
Auto Ranging
During measurement and calibration this meter will automatically
switch to the correct range.
MEASUREMENT PROCEDURES
Note...
Before beginning measurements, connect the probe to the meter,
turn the meter on and polarize the electrode for 15 minutes.
Turning the Meter On/O󰀨
1. Press POWER to turn the meter on. The meter will default to
the last used measurement and temperature scales.
2. Press and hold POWER for 2 seconds to turn the meter o󰀨.
Indicator Light Descriptions
Meter Status Indicator Status
Power o󰀨 Light o󰀨
Power on Green light on for 5 seconds, then light o󰀨
Low battery Red light ashing every 5 seconds
Charging battery Blue light on
Page 24
24 25
Scale Selection
1. Press RANGE to switch between the % of Saturation and
Milligrams per Liter (mg/L) scales (see Fig. A and Fig. B).
Backlight
1. With the meter on press POWER again to turn the
backlight on/o󰀨.
Fig. A Fig. B
Saving to Memory
1. Press MI to save the current measured value and view the
data point number (see Fig. A).
2. Up to 99 Memories can be saved for both Conductivity and TDS. If the saved memory exceeds 99, the new data will be written over saved memories beginning with #1.
Fig. A
Page 25
25
Recall Memory
1. Press RANGE to select the scale for recall.
2. Press MR to enter the Recall mode and view the last saved reading (see Fig. A and Fig. B). The date and time are on cyclical display. If there is no saved data “- - - -” is displayed (see Fig. C).
3. Press ↑ or ↓ to view the saved data points. Press ↑ to view the data points increasing by 1. Press ↓ to view the data points decreasing by 1.
4. Press EXIT to exit the Recall Mode and return to Normal Mode.
Fig. A Fig. B
Fig. C
Clear Memory
1. Press MI and MR simultaneously for 2 seconds to clear all
memory. The meter will return to Normal mode.
Page 26
26 27
Hold
1. Press HOLD to hold the value (see Fig. A).
2. Press HOLD to release the hold value.
Note…
When in Hold Mode, all other functions will be disabled except
turning the meter o󰀨, backlight function, saving and exiting Hold Mode.
Fig. A
Maximum, Minimum and Average
1. Press RANGE to select the scale.
2. Press MAX/MIN/AVE, the maximum recorded value based
on the saved measurements will appear on the LCD
(see Fig. A and Fig. B.) The date and time are on cyclical display, If no measurements have been taken, “- - - -” is displayed (see Fig. C).
3. Press MAX/MIN/AVE, the minimum recorded value based
on the saved measurements will appear on the LCD
(see Fig. D and Fig. E). The date and time are on cyclical display. If no measurements have been taken, “- - - -” is displayed (see Fig. F).
4. Press MAX/MIN/AVE, the average recorded value based
on the saved measurements will appear on the LCD
(see Fig. G and Fig. H). The date and time are on cyclical display. If no measurements have been taken, “- - - -” is displayed.
5. Press MAX/MIN/AVE to return to the regular mode.
Page 27
27
Fig. A Fig. B
Fig. C
Fig. D
Fig. E Fig. F
Fig. G Fig. H
Mode mg/L(ppm) % Saturation Temperature
Range 0.00 to 19.99
mg/L (ppm)
0.0 to 199.9% 32 to 122°F (0 to 50°C)
Resolution 0.01mg/L (ppm) 0.1% 0.1°F/°C
Accuracy ±1.0% Full Scale ±1.0% Full Scale ±0.9°F/±0.5°C
Temperature Compensation
Automatic (ATC) or manual (MTC) from
32 to 122°F or 0 to 50°C
SAL Correction
Range: 0.0 to 50.0 ppt
Resolution: 0.1 ppt
Method: Manual
Barometric Pressure Correction
Range: 500 to 1499 mmHg / 66.6 to 199.9 kPa
Resolution: 1 mmHg / 0.1 kPa
Method: Manual
Electrode
Polarographic Electrode
Memory
99 data points per scale with time and date stamp
Operating Temperature
32 to 122°F (0 to 50°C)
Battery
Specications
Rechargeable Lithium battery 7.4V (1450mAh)
Approximate Battery Life when Fully Charged
>200 hours
SPECIFICATIONS
Page 28
28 PB
WARRANTY
Sper Scientic warrants this product against defects in materials and workmanship for a period of ve (5) years from the date
of purchase, and agrees to repair or replace any defective unit
without charge. If your model has since been discontinued, an equivalent Sper Scientic product will be substituted if available. This warranty does not cover probes, batteries, battery leakage,
or damage resulting from accident, tampering, misuse, or abuse of
the product. Opening the meter to expose its electronics will break the waterproof seal and void the warranty. To obtain warranty
service, ship the unit postage prepaid to:
SPER SCIENTIFIC LTD.
8281 E. Evans Rd., Suite #103
Scottsdale, AZ 85260
(480) 948-4448
The defective unit must be accompanied by a description of the
problem and your return address. Register your product online at www.sperwarranty.com within 10 days of purchase.
Revised 7/12/2017
Loading...