TABLE OF CONTENTS
I. INTRODUCTION ............................................................................. 2
II. PANEL DESCRIPTION ................................................................... 2
III. MEASUREMENT PROCEDURES .................................................. 3
IV. SPECIFICATIONS .......................................................................... 3-4
VII. WARRANTY .................................................................................... 4
I. INTRODUCTION
This IR Probe is heavy duty and compact and the microprocessor circuit
assures a high degree of accuracy. Measures °C and °F, with a preset
emissivity value of 0.95, that accommodates most applications. With the
plug-and-play feature of this probe, calibration procedures are not required. The meter features Hold, Max./Min, RS232 and Auto Shut Off.
II. PANEL DESCRIPTION
11
9
8
10
12
7
1. LCD Display 5. °C/°F (A) Button 9. Probe Lock
2. Power Button 6. K/J (B) Button 10. Probe Input Socket
3. REC. Button 7. Battery 11. IR Probe Sensing Head
4. Hold Button 8. Temperature Input 12. Temperature Adjustment
- 2 -

III. MEASUREMENT PROCEDURES
•
Plug probe into the PROBE INPUT SOCKET.
•
Make sure the PROBE LOCK is in the lock position
•
Turn on the instrument by pressing the POWER button.
•
Point the probe toward the source to be measured. The distance-tospot (D/S) ratio is approximately 7 to 1. Press the
°C/°F (A)
button to
select Centigrade or Fahrenheit. The temperature is displayed on the
LCD.
Refer to the Environmental Quality Meter 850071 Instruction Manual for
complete instructions about Hold, Max./Min, RS232, Auto Shut Off, and
other functions.
Temperature Adjustment
Environmental conditions such as low temperatures and rapid temperature
changes may cause the reading to drift by 1 to 3 degrees. If a deviation
occurs, slightly turn the temperature adjustment until a correct reading is
obtained.
Field Distance/Spot (D/S)
The distance-to-spot factor is approximately 7-to-1. The object must be
larger than the spot size as calculated by the Distance/Spot ratio. For an
accurate measurement, a 1.5 times distance/spot size is recommended.
Emissivity
The probe has an emissivity value of 0.95, which will cover 90% of the
typical applications. If the emissivity value of the object is not 0.95 or if it's
surface is highly reflective, apply masking tape or a paint with a 0.95 emis-
sivity value. Clean the surface if it is covered with frost or other matter
.
IV. SPECIFICATIONS
Unit of Probe Resolution Accuracy
ºC -10 ~ 300
ºF 14 ~ 572
1°
The greater of the two:
±3°C, or ±3% of the reading.
Operating Temperature (meter) 0°C to 50°C (32°F to 122°F).
Operating Humidity Max. 80% RH.
IR Sensor Thermocouple pie.
- 3 -
.

Emissivity Setting 0.95
Measurement Wave Length 6 ~ 12 micro meters
Field Distance/Spot Approximately 7:1
Probe Size Diameter: 1.6" (40mm).
Length: 6" (150mm).
Refer to the Environmental Quality Meter 850071 Instruction Manual for
meter specifications.
ONE YEAR WARRANTY
Sper Scientific warrants this product against defects in materials and
workmanship for a period of one year from the date of purchase, and
agrees to repair or replace any defective unit without charge. If your
model has since been discontinued, an equivalent Sper Scientific product
will be substituted if available. This warranty does not cover damage
resulting from accident, misuse, or abuse of the product. In order to obtain
warranty service, simply ship the unit postage prepaid to:
SPER SCIENTIFIC LTD.
7720 East Redfield, Suite 7
Scottsdale, Arizona 85260
(480) 948-4448
www.sperscientific.com
- 4 -