Sperry instruments 3121, 3123, 3122 Operating Instructions Manual

MODEL 312131223123
Model-31213122 Model-3123
HIGH VOLTAGE INSULATION TESTER
OPERATING INSTRUCTIONS
7
/
05 FORM#341
CONTENTS
Page
1Features …………………………………………………………1
2Specifications ……………………………………………2 3
3Instrument Layout ………………………………………………4
4Operating Instructions …………………………………………5
4-1Mechanical Zero Adjustment ……………………………5
4-2Battery Check ……………………………………………5
4-3Insulation Resistance Measurement ……………………6
4-4Continuous Measurement ………………………………6
4-5Use of Guard Terminal  …………………………………7
5Battery Replacement  ………………………………………8
6Cleaning Meter Cover  ………………………………………8
1
1. Features
●  Battery powered, the instruments test insulation up to
100000MΩ at 2500V for Model 3121, 200000MΩ at 5000V for Model 3122 and 200GΩ at 5000V and 400GΩ at 10000V for Model 3123.
  Suited for heavy duty electrical maintenance and servicing of
industrial installations, cables, transformers, generators and switchgear where high voltage insulation tests are required.
  Dual s c a l e s for low and h i g h ranges which c h a n g e
automatically. Colour coded scales for easy reading plus LED's that illuminate in matching colour.
  Drip proof construction. The case is sealed with rubber
gaskets to protect internal circuit against rain.
  Hard carrying case furnished as standard accessory. Houses
both instrument and test leads in compact form. Made of plastic, it is highly water resistant.
  Designed for low power consumption. Since the maximum current consumption is 90mA eight pieces of 1.5V SUM-3or equivalent ) permit about 6 hours of continuous operation even when the instrument is used on maximum load or twice longer on minimum load.
  Rated output voltage is maintained down to 100MΩ for Model 3121,200MΩ for Model 3122 and 0.2GΩ/0.4GΩ for Model
3123. This permits accurate measurements of low insulation
resistance.
2
Model 3121 Model 3122 Model 3123
DC Test Voltage 2500V 5000 V 5000V 10000V
Measuring Ranges
02000MΩ/
1000100000MΩ
automatic change
05000 MΩ/
2000 200000MΩ
(automatic change)
Accuracy
Insulation Resistance
±5% of reading
10050000
±10% of reading
or 0.5% of scale length
ranges other than
listed above
at 23 ±5
±10% of reading
10050000
±20% of reading
or 1.0% of scale length
ranges other than
listed above
at 1040
±5% of reading 200100000MΩ ±10% of reading
or 0.5% of scale length
( ranges other than
listed above at 23℃ ±5 ±10% of reading 200100000MΩ ±20% of reading
or 1.0% of scale length
( ranges other than
listed above
at −1040
Output Voltage
2500V ±5%
10050000MΩ
5000V ±5 200100000MΩ
Operating Temperature & Humidity
−10℃ 40 at 85% max. relative humidity
Storage Temperature & Humidity
−20℃ 60 at 90% max. relative humidity
Insulation Resistance 1000MΩ max./ 1000V between electrical circuit & housing case
Withstand Voltage 5000V AC for one minute between electrical circuit & housing case Dimensions 200L×140W×80Dmm Weight Approx. 1kgincluding batteries & line probe
Power Source 8 pcs of 1.5V SUM-3 battery or equivalent Accessories          Hard Carrying Case, Batteries, Test Leads earth & guard leads   
2. Specifications
3
05GΩ/2200GΩ
(automatic change)
010GΩ/4400GΩ
(automatic change)
±5% of reading
0.2100
±10% of reading
or 0.5% of scale length
ranges other than
listed above
at 23 ±5
±10% of reading
0.2100
±20% of reading
or 1.0% of scale length
ranges other than
listed above
at 1040
±5% of reading
0.4200
±10% of reading
or 0.5% of scale length
ranges other than
listed above
at 23 ±5
±10% of reading
0.4200
±20% of reading
or 1.0% of scale length
ranges other than
listed above
at 1040
5000V ±5%
0.2~100GΩ
10000V ±5%
0.4~200GΩ
Optional adaptor model 8020 is available for connection to recorder
4
3. Instrument Layout
① Line Probe ③ Earth Terminal ⑤ Press to Test Button ⑥ Function Switch
Fig.1
② Guard Terminal ④ LED's for High&
Low Range Indication
⑦ Battery Compartment   Cover
5
CAUTION: BE CAREFUL ABOUT HIGH VOLTAGE PRESENT ACROSS LINE AND EARTH TERMINALS OF INSTRUMENT WHEN PRESS TO TEST BUTTON IS OPERATED. MAKE SURE TO EARTH CIRCUIT UNDER TEST. ALWAYS CONNECT EARTH TERMINAL OF INSTRUMENT TO EARTH. THE BUZZER WILL KEEP SOUNDING DURING INSULATION RESISTANCE MEASUREMENT.
4-1. Mechanical Zero Adjustment
With the function switch set at OFF position, adjust the meter pointer to ∞ mark on the scale. Use a screwdriver to turn the zero adjust screw located at the center of the front panel.
4-2. Battery Check
With the function switch set at BATT. CHECK position, operate the press to test button. The batteries are good when the pointer stays in BATT. GOOD area or to the right of this area. If not, replace them.
Note:  Refrain from holding down or locking the press to
test button during this as it will result in current consumption larger than insulation resistance measurement while the batteries are still new.
4. Operating Instructions
6
4-3. Insulation Resistance Measurement
With the function switch set at OFF position, always connect the circuit under test to earth. Attach the test lead to the earth terminal of the instrument and connect to the earthed side of the circuit under test. With the function switch set at MΩ position for Model 3121 and 3122 or GΩ position for Model 3123, place the line probe in contact with the circuit under test and operate the press to test button. When the green LED illuminates, read insulation resistance on the outer scale(for high range). Use the inner scale where the red LED illuminates. For insulation testing at 5000V and
10000V, read the black and red scales respectively (for Model
3123). After a test, release the press to test button and wait for
several seconds without disconnecting the line probe from the circuit tested. This is intended to discharge the charge stored in the circuit tested.
4-4. Continuous Measurement
Make sure that the circuit under test is earthed and that the test lead attached to the earth terminal of the instrument is connected to the earthed side of the circuit under test. Push the press to test button and turn clockwise to lock for continuous measurement. When making this measurement, good care must be taken against the high voltage continuously present across the line and earth terminals of the instrument.
Note:  Make certain that the circuit under test does not
include components which will be damaged by the high voltage applied.
7
4-5. Use of Guard Terminal
Illustrated in Fig. 2 is an example of the insulation resistance measurement of an electric wire. If the line probe is simply connected to the wire conductor and the earth lead to the immersion liquid container as shown, a measurement error will be introduced as this results in the measurement of the combined resistance of insulation resistance and the surface leakage resistance at the cut end of the electric wire. In order to remove this surface leakage current, wide a guard wire around the cut end of the conductor and connect it to the guard terminal of the instrument using the guard lead. Then, the surface leakage current will bypass the indicating meter of the insulation resistance tester.
GUARDWIRE
ELECTRIC WIRE
IMMERSIONLIQUID
EARTH GUARD LINE
Fig.2
8
Remove the battery compartment cover by loosening the screw located on the back of the housing case. Replace the whole battery pack. Alkaline batteries are recommended where the instrument is used at a temperature below the freezing point. The ordinary manganese batteries will deteriorate below the freezing point.
For your information, optional adaptor Model 8020 is available to allow
Model 3121,3122 and 3123 to be connected to a recorder for recording insulation resistance. It provides an output signal of 1 μA/10mV DC.
5. Batttery Replacement
6. Cleaning Meter Cover
Do not try to remove dirt on the meter cover by rubbing hard with a dry cloth. This can remove anti-electrostatic agent applied to the surface of the surface of the meter cover.
When the meter reading is affected by electrostatic build up on the meter cover, wipe the meter cover surface using a cloth dampened with off-the shelf anti-static agent or detergent. To avid possible deforming or discoloring, do not use solvents.
To clean the body of the instrument, use cloth dampened with detergent.
CAUTION Never use paint thinner, benzene or other solutions containing solvents for cleaning the instrument. Otherwise, deforming or discoloring of the instrument body or the meter cover may result.
Note: Handle the instrument with care and follow the instructions in order
to maintain it in good condition for a long period of time.
9
Lifetime Limited Warranty
The attention to detail of this fine snap-around instrument is further enhanced by the application of A.W. Sperrys unmatched service and concern for detail and reliability. These A.W. Sperry snap-arounds are internationally accepted by craftsmen and servicemen for their unmatched performance. All A.W. Sperrys snap-around instruments are unconditionally warranted against defects in material and workmanship under normal conditions of use and service; our obligation under this warranty being limited to repairing or replacing free of charge, at A.W. Sperry snap-around instrument that malfunctions under normal operating conditions at rated use.
1
Replacement procedure
Securely wrap the instrument and its accessories in a box or mailing bag and ship prepaid to the address below. Be sure to include your name and address, as well the name of the distributor, with a copy of your invoice from whom the unit was purchased, clearly identifying the model number and date of purchase.
A.W.SPERRY INSTRUMENTS INC.
ATT: Customer service dept.
2150 Joshuas Path, Suite 302,
Hauppauge, NY 11788
1
The warranty is not applicable if the instrument has been: misused, abused, subjected to loads in excess of specifications, has had unauthorized repair or has been improperly assembled or used.
*Note: Recommended calibration interval should not exceed one year. Calibration service charges are not covered terms and conditions of warranty.