Sperian ToxiPro Operating Manual

ToxiPro®
Single Gas
Reference Manual
Honeywell Analytics 651 South Main Street Middletown, CT 06457 (800) 711-6776 (860) 344-1079 Fax (860) 344-1068 Part number 13-264-N3 Version 1.30
03-June-2011
0
THE TOXIPRO®-NGSG PERSONAL PORTABLE GAS DETECTORS HAVE BEEN DESIGNED FOR THE DETECTION AND MEASUREMENT OF POTENTIALLY HAZARDOUS ATMOSPHERIC CONDITIONS.
IN ORDER TO ASSURE THAT THE USER IS PROPERLY WARNED OF POTENTIALLY DANGEROUS ATMOSPHERIC CONDITIONS, IT IS ESSENTIAL THAT THE INSTRUCTIONS IN THIS REFERENCE MANUAL BE READ, FULLY UNDERSTOOD, AND FOLLOWED.
®
ToxiPro
-NGSG Reference Manual
Honeywell Analytics Part Number 13-264-N3
Version 1.30
Copyright 2011
by
Honeywell Analytics, Inc
Middletown, Connecticut 06457
All rights reserved.
No page or part of this operation manual may be reproduced in
any form without written permission of the copyright owner
shown above.
Honeywell Analytics reserves the right to correct typographical errors.
1
Table of Contents
Operating Temperature Range ............................................................................3
Certifications ......................................................................................................... 3
Warnings and Cautions........................................................................................ 4
A.Signal Words ................................................................................................................. 4
B.Warnings ....................................................................................................................... 4
1.Overview......................................................................................................... 5
1.1 Methods of sampling .................................................................................................. 5
1.2 Sensors ...................................................................................................................... 5
1.2.1 ToxiPro sensor ranges........................................................................................ 5
1.2.2 ToxiPro O2 sensor ranges................................................................................... 5
1.2.3 ToxiPro O2 warm-up phase................................................................................. 5
1.3 Alarm and warning logic ............................................................................................. 5
1.3.1 Gas alarms.......................................................................................................... 5
1.3.2 Calibration and bump test due notices................................................................ 5
1.3.3 Battery replacement Warning.............................................................................. 5
1.3.4 Low battery alarms.............................................................................................. 6
1.3.5 Missing sensor during startup ............................................................................. 6
1.3.6 Corrupt memory .................................................................................................. 6
1.3.7 Vibrating Alarm Error .......................................................................................... 6
1.3.8 Heartbeat Feature............................................................................................... 6
1.4 Functions.................................................................................................................... 6
1.4.1 Security beep ...................................................................................................... 6
1.4.2 Latching alarms................................................................................................... 6
1.4.3 Enhanced Latching Alarm ................................................................................... 6
1.4.4 Silence warning alarms....................................................................................... 6
1.5 IQ Dock Compatibility................................................................................................. 6
1.6 Design components.................................................................................................... 7
1.7 Standard accessories ................................................................................................. 7
1.8 Value pack kits ........................................................................................................... 7
2.Field Operation .............................................................................................. 7
2.1 Turning the ToxiPro on ............................................................................................... 7
2.1.1 Calibration due lockout ....................................................................................... 8
2.2 Backlight..................................................................................................................... 8
2.3 Functions.................................................................................................................... 8
2.4 Calibration and bump test due notices ....................................................................... 9
2.5 Low battery alarms ..................................................................................................... 9
2.6 Sensor Over Range.................................................................................................... 9
2.7 Turning the ToxiPro off ............................................................................................. 10
2.8 Always On Mode ...................................................................................................... 10
2.8.1 Turning Off in Always On Mode ........................................................................ 10
2.9 Sampling .................................................................................................................. 10
2.9.1 Sample draw kit usage...................................................................................... 11
3.Calibration .................................................................................................... 11
3.1 Verification of accuracy ............................................................................................ 11
3.1.1 Verifying accuracy and response: ToxiPro O2................................................... 11
3.1.2 Verifying accuracy: ToxiPro with toxic gas sensor ............................................ 11
3.2 Effect of contaminants on ToxiPro sensors .............................................................. 12
3.2.1 Effects of contaminants on O2 sensors ............................................................. 12
3.2.2 Effects of contaminants on toxic gas sensors ................................................... 12
3.3 Fresh air/zero calibration .......................................................................................... 12
3.4 Functional (bump) testing (toxic sensor versions) .................................................... 13
3.5 Span Calibration (toxic sensor versions) .................................................................. 13
3.6 Failure to calibrate.................................................................................................... 14
3.6.1 Fresh air/zero calibration failure........................................................................ 14
3.6.1.1 Causes of fresh air/zero calibration failures............................................... 14
3.6.1.2 Forced fresh air/zero calibration ................................................................ 14
2
3.6.2
Span calibration failure ..................................................................................... 14
3.6.2.1 Sensor out of range (no CAL).................................................................... 14
3.6.2.2 Causes for span cal failure ........................................................................ 14
3.7 Fresh air/zero calibration in a contaminated atmosphere ......................................... 15
4.Maintenance ................................................................................................. 15
4.1 Replacing batteries................................................................................................... 15
4.2 Replacing sensors .................................................................................................... 16
4.3 Proper Cleaning ....................................................................................................... 16
4.4 Storage..................................................................................................................... 16
5.PC-Instrument Communications ............................................................... 16
5.1 Event logging............................................................................................................ 16
5.2 Instrument firmware upgrades.................................................................................. 17
5.3 Initiating communications ......................................................................................... 17
6.Exploded view and basic parts list............................................................. 17
Appendices.......................................................................................................... 18
Appendix A: Sensor Cross-Sensitivity Chart ....................................................................... 18
Appendix B: Replacement Sensor List ................................................................................ 18
Appendix C: Calibration Frequency Recommendation....................................................... 19
Honeywell Analytics Warranty Gas Detection Products .................................20
Operating Temperature Range
The safe operating temperature range of the gas detector is printed on the back label of the instrument. Use of Honeywell Analytics Gas Detectors outside of the instrument’s specified operating temperature range may result in inaccurate and potentially dangerous readings.
Certifications
The ToxiPro®-NGSG carries the following certifications:
UL and c-UL, Class I, Division 1, Groups A,B,C,D, Temp Code T4.
UL Class II, Division 1, Groups E,F,G
ATEX Certification:
II 1 G Ex ia IIC T4
UL International DEMKO A/S 09 ATEX 0823124
IECEx Ex ia IIC T4 Ga
3
Warnings and Cautions
A. Signal Words The following signal words, as defined by ANSI Z535.4-1998, are used in the ToxiPro-
NGSG Operator’s Guide.
indicates an imminently hazardous situation which, if not avoided,
will result in death or serious injury.
indicates a potentially hazardous situation which, if not avoided,
could result in death or serious injury.
indicates a potentially hazardous situation, which if not avoided, may
result in moderate or minor injury.
CAUTION used without the safety alert symbol indicates a potentially hazardous
situation which, if not avoided, may result in property damage.
B. Warnings
1.
the detection of either oxygen deficiencies or specific toxic gas accumulations. An alarm condition indicates the presence of a potentially life-threatening hazard and should be taken very seriously.
2.
established procedures. The safest course of action is to immediately leave the affected area, and to return only after further testing determines that the area is once again safe for entry. Failure to immediately leave the area may result in serious injury or death.
3.
4. The accuracy of ToxiPro instruments equipped with toxic gas
sensors should be checked periodically with known concentration calibration gas. Failure to check accuracy can lead to inaccurate and potentially dangerous readings. The ToxiPro O
5. A sensor that cannot be calibrated or is found to be out of
tolerance must be replaced immediately. An instrument equipped with a toxic gas sensor that fails calibration may not be used until testing with known concentration test gas determines that accuracy has been restored, and the instrument is once again fit for use. Instruments equipped with an oxygen sensor that fail calibration may not be used until testing with fresh air determines that accuracy has been restored and the instrument is once again fit for use.
6.
ToxiPro unless the concentrations of your calibration gas differ from the concentrations of the calibration gas that is normally supplied by Honeywell Analytics for use in calibrating the ToxiPro.
7.
components when calibrating the ToxiPro can lead to dangerously inaccurate readings and may void the standard Honeywell Analytics warranty.
Honeywell Analytics offers calibration kits and long-lasting cylinders of test gas
specifically developed for easy calibration. Customers are strongly urged to use only Honeywell Analytics calibration materials when calibrating the ToxiPro.
8.
9.
personnel only. Read, understand and follow the directions set forth in this reference manual before operating the ToxiPro.
10.
Gas/AIR (max. 21.0% O
ToxiPro personal, portable gas detectors have been designed for
In the event of an alarm condition it is important to follow
Use only Panasonic #CR2 in the ToxiPro.
should be periodically calibrated in fresh air.
2
Do not reset the calibration gas concentration setpoints in the
Use of non-standard calibration gas and/or calibration kit
Substitution of components may impair intrinsic safety.
For safety reasons the ToxiPro must be operated by qualified
The ToxiPro has been tested for intrinsic safety in Explosive
).
2
4
1. Overview
The ToxiPro is a single sensor gas detector that can be configured to detect either oxygen
) or one of a variety of toxic gases. The
(O
2
ToxiPro’s sensor type is shown on the front of the instrument and is also shown on the display during the start up sequence. The ToxiPro includes numerous features designed to meet specific user requirements.
1.1 Methods of sampling
The ToxiPro may be used in diffusion mode, or with the manual sample draw kit that is available separately. In either mode, the atmosphere must reach the sensor for the instrument to register a reading. In diffusion mode, the atmosphere reaches the sensor by diffusing through the sensor port on the front of the instrument. Normal air movements are enough to carry the sample to the sensor. During remote sampling, the gas sample is drawn into the sensor compartment through the probe assembly and a length of tubing.
See section 2.9 for more details on sampling the atmosphere.
1.2 Sensors
All versions of the ToxiPro except the O model use an electrochemical toxic gas sensor. The ToxiPro O
uses a galvanic
2
oxygen sensor. Both types of sensor have been designed to minimize the effects of common interfering gases. These sensors provide accurate, dependable readings for gases commonly encountered in industrial applications. A sensor cross sensitivity chart is provided in Appendix A at the back of this manual.
1.2.1 ToxiPro sensor ranges
Specific toxic sensor ranges and resolutions are provided in the sensor replacement chart in Appendix B.
1.2.2 ToxiPro O
sensor ranges
2
The oxygen sensor used in the ToxiPro O has a range of 0-30% by volume.
1.2.3 ToxiPro O
The ToxiPro O
warm-up phase
2
requires a one-time, 15-
2
minute warm-up phase prior to initial activation. See section 2.1 for further details.
1.3 Alarm and warning logic
2
2
warning state the LED’s will flash, yellow. In the danger state the LED’s will flash, red. Two oxygen alarm set points have been provided for ToxiPro O
instruments. The
2
danger alarm is sounded for oxygen deficiency and the warning alarm is sounded for oxygen enrichment. ToxiPro instruments equipped with a toxic gas sensor have up to four alarm set points: Warning, Danger, Custom, TWA (Time Weighted Average) (if enabled) and STEL (Short Term Exposure Limit) (if enabled). The custom alarm level can be set to a level greater than/equal to the danger alarm level for toxic sensors and less than/equal to danger level for oxygen sensors using the IQ System software. The custom alarm is a latching alarm. The custom alarm looks and sounds similar to the DANGER alarm but, with the warning symbol turned on and the DANGER icon not being set. If the custom alarm level has been reached then the alarm can only be silenced by putting the instrument in the dock (successful IrDA connection) or successful connection to BioTrak. If the custom alarm value is zero then the custom alarm is disabled and “OFF” will appear on the instrument’s LCD during the startup sequence. If the danger alarm has been disabled then the custom alarm is also disabled.
Note: In the standard ToxiPro configuration with toxic gas sensors, the STEL and TWA alarms are not enabled. The STEL and TWA alarms may be enabled at the factory or through BioTrak software. Call Honeywell Analytics for more details.
1.3.2 Calibration and bump test due notices
The ToxiPro includes bump test and calibration due notices.
For more information on the calibration due notices, see section 2.4 below.
1.3.3 Battery replacement Warning
If a new battery has been installed or the existing battery was removed and replaced the ToxiPro will show “F 0” on the display and will beep 6 times upon startup. It will then go into the set time/date mode.
1.3.1 Gas alarms
ToxiPro gas alarms are user-adjustable and may be set anywhere within the range of the specific sensor. When an alarm set point is exceeded a loud audible alarm sounds. In the
5
1.3.4 Low battery alarms
The ToxiPro is designed with two battery warning alarms that are activated when the battery voltage is reduced to specific levels.
For more details on the battery alarm, see section 2.5 below.
1.3.5 Missing sensor during startup
If the ToxiPro fails to detect a sensor during startup, it will show “F 1” with the caution symbol and then shut itself off. For more details, see section 2.1.
1.3.6 Corrupt memory
The ToxiPro continuously monitors its onboard memory. If the instrument determines that the memory is corrupt, it will display “F 4” and proceed to shut itself off. If “F 4” is displayed, contact Honeywell Analytics for further details.
1.3.7 Vibrating Alarm Error
If the vibrating alarm fails “F 6” will be shown on the display. Contact Honeywell Analytics for further details.
1.3.8 Heartbeat Feature
A blinking heart is an active indicator, notifying the user that the instrument is operating normally.
1.4 Functions
1.4.1 Security beep
The ToxiPro includes a security beep that can be enabled or disabled with BioTrak software through the PC’s IrDA port. If the security beep is enabled, the ToxiPro will emit a short beep coupled with an LED flash at a specific interval to remind the user that the instrument is active.
1.4.2 Latching alarms
The ToxiPro includes latching alarms that can be enabled or disabled with BioTrak software through the PC’s IrDA port. With the alarm latch enabled, the audible and visible alarms will continue to sound even after the atmospheric hazard has cleared. To turn the
alarm off once the hazard is no longer present, simply press the MODE button.
1.4.3 Enhanced Latching Alarm Note: V9.20 or greater ToxiPro firmware and
V8.00 or greater is required of all IQ System software to support the Enhanced Latching Alarm. The enhanced latching capability can be enabled for the danger alarm using an IQ System template and apply the template in ToxiPro IQ Express. The user can specify the duration for latching of the Danger alarm. When the instrument goes into the danger alarm for greater than or equal to the user defined duration for latching then the alarm is latched. When the alarm is latched and if the gas level goes below the alarm level, the audible and visual alarms will continued. If alarm latching is enabled and the instrument is in alarm for less than the latching duration, the alarm will not latch when the gas level drops below the alarm level.
Latch Acknowledgement
A latched alarm latch can be acknowledged by two methods. The first is either by mode press or inserting the instrument into the dock. The second is by dock insertion only. The method is selectable in the IQ System template.
Note: For more information on creating and assigning templates, please refer to the Database Manager Reference manual.
1.4.4 Silence warning alarms
The ToxiPro’s audible and vibrating (if so equipped) alarms can be turned off during an alarm condition by pressing the MODE button if this function has been enabled with BioTrak software. The visual warning alarm light and readings will continue to indicate the alarm.
1.5 IQ Dock Compatibility
The ToxiPro must be equipped with instrument firmware version 4.30 or higher to be compatible with the IQ Express Docking Station. The IQ Express Dock is an automatic calibration station coupled with a data management system. The IQ Express Dock must be equipped with firmware version
6.53 or higher. Instrument firmware in the ToxiPro may be upgraded at any time. See
section 5.2 below for details on software upgrades.
6
Loading...
+ 14 hidden pages