Sperian ToxiLtd Reference Manual

Page 1
ToxiLtd®
Single Gas
Detector
Reference Manual
Sperian Instrumentation
651 South Main Street Middletown, CT 06457 (800) 711-6776 (860) 344-1079 Fax (860) 344-1068 Part number 13-266 Version 2.00 23AUG2007
0
Page 2
THE TOXILTD® PERSONAL PORTABLE GAS DETECTORS HAVE BEEN DESIGNED FOR THE DETECTION AND MEASUREMENT OF POTENTIALLY HAZARDOUS ATMOSPHERIC CONDITIONS.
IN ORDER TO ASSURE THAT THE USER IS PROPERLY WARNED OF POTENTIALLY DANGEROUS ATMOSPHERIC CONDITIONS, IT IS ESSENTIAL THAT THE INSTRUCTIONS IN THIS REFERENCE MANUAL BE READ, FULLY UNDERSTOOD, AND FOLLOWED.
Sperian Instrumentation Part Number 13-266
Sperian Protection Instrumentation, LLC
No page or part of this operation manual may be reproduced in
any form without written permission of the copyright owner
Sperian Instrumentation reserves the right to correct
ToxiLtd® Reference Manual
Version 2.00
Copyright 2007
by
Middletown, Connecticut 06457
All rights reserved.
shown above.
typographical errors.
1
Page 3
Table of Contents
Certifications........................................................................................ 3
ATEX Label Requirement.................................................................... 3
Warnings and Cautions ...................................................................... 3
A. Signal Words .........................................................................................................3
B. Warnings................................................................................................................4
1. Overview......................................................................................... 5
1.1 Methods of sampling .......................................................................................... 5
1.2 Sensors..............................................................................................................5
1.2.1 Sensor ranges and resolutions................................................................... 5
1.2.2 ToxiLtd O
1.2.3 F1: Missing sensor during startup............................................................... 5
1.3 Alarm logic......................................................................................................... 6
1.3.1 Gas alarms................................................................................................. 6
1.3.2 Silence warning alarms .............................................................................. 6
1.3.3 Calibration due notices............................................................................... 6
1.3.4 Low battery alarms ..................................................................................... 6
1.4 Design components ........................................................................................... 6
1.5 Standard accessories.........................................................................................7
1.6 Value pack kits................................................................................................... 7
2. Field Operation .............................................................................. 7
2.1 Turning the ToxiLtd on ....................................................................................... 7
2.1.1 O
2.1.2 Toxic version start up sequence................................................................. 7
2.1.3 Start up sequence ......................................................................................7
2.2 Functions ...........................................................................................................8
2.2.1 View max gas readings ..............................................................................8
2.2.2 View alarm levels ....................................................................................... 8
2.3 Calibration due notices.......................................................................................9
2.4 Low battery alarms............................................................................................. 9
2.5 Turning the ToxiLtd off ....................................................................................... 9
2.6 Sampling.......................................................................................................... 10
2.6.1 Sample draw kit usage .............................................................................10
version start up sequence......................................................................7
2
3. Calibration.................................................................................... 10
3.1 Verification of accuracy .................................................................................... 10
3.1.1 Verifying accuracy and response: ToxiLtd O
3.1.2 Verifying toxic sensor performance...........................................................11
3.2 Effect of contaminants on ToxiLtd sensors.......................................................11
3.2.1 Effects of contaminants on oxygen sensors..............................................11
3.2.2 Effects of contaminants on toxic gas sensors........................................... 11
3.3 Fresh air/zero calibration..................................................................................12
3.4 Functional (bump) testing for ToxiLtd with toxic gas sensor..............................12
3.5 Span Calibration for ToxiLtd with toxic gas sensor ...........................................13
3.6 Failure to calibrate............................................................................................14
3.6.1 Fresh air/zero calibration failure................................................................14
3.6.1.1 Reason for fresh air/zero calibration failures....................................................14
3.6.2 Span calibration failure............................................................................. 14
3.6.2.1 Sensor out of range (no CAL)..........................................................................14
3.6.2.2 No gas............................................................................................................14
warm-up phase .........................................................................5
2
...........................................10
2
2
Page 4
3.6.2.3 Causes for span cal failure.............................................................................. 14
3.6.3 Forced fresh air/zero calibration ...............................................................14
3.7 Fresh air/zero calibration in a contaminated atmosphere .................................15
4. Event logging............................................................................... 15
4.1 Initiating communications................................................................................. 15
Appendices ........................................................................................ 16
Appendix A: Sensor Cross-Sensitivity Chart................................................................ 16
Appendix B: Default toxic sensor calibration gas settings ............................................ 16
Appendix C: Calibration Frequency ............................................................................. 17
Sperian Instrumentation Warranty Gas Detection Products ........ 18
Certifications
The ToxiLtd® carries the following certifications:
UL and c-UL, Class I, Division 1, Groups A,B,C,D, Temp Code T4. UL Class II, Division 1, Groups E,F,G ATEX Certification: II 2 G EEx ia IIC T4 UL International DEMKO A/S 04 ATEX 0327520 IECEx Ex ia IIC T4 CQST (China)
ATEX Label Requirement
Warnings and Cautions
A. Signal Words
The following signal words, as defined by ANSI Z535.4-1998, are used in the ToxiLtd Operator’s Guide.
indicates an imminently hazardous situation which, if not
avoided, will result in death or serious injury.
indicates a potentially hazardous situation which, if not
avoided, could result in death or serious injury.
indicates a potentially hazardous situation, which if not
avoided, may result in moderate or minor injury.
CAUTION used without the safety alert symbol indicates a potentially
hazardous situation which, if not avoided, may result in property damage.
3
Page 5
B. Warnings
1.
The ToxiLtd personal, portable gas detectors have been designed for the detection of either oxygen deficiencies or specific toxic gas accumulations. An alarm condition indicates the presence of a potentially life­threatening hazard and should be taken very seriously.
2.
In the event of an alarm condition it is important to follow established procedures. The safest course of action is to immediately leave the affected area, and to return only after further testing determines that the area is once again safe for entry. Failure to immediately leave the area may result in serious injury or death.
3.
The accuracy of ToxiLtd instruments equipped with a toxic
gas sensor should be checked periodically with known concentration calibration gas. Failure to check accuracy can lead to inaccurate and potentially dangerous readings. ToxiLtd instruments equipped with an oxygen sensor should be periodically calibrated in fresh air.
4.
A ToxiLtd equipped with a toxic gas sensor that fails
calibration may not be used until testing with known concentration test gas determines that accuracy has been restored, and the instrument is once again fit for use. Instruments equipped with an oxygen sensor that fail calibration may not be used until testing with fresh air determines that accuracy has been restored and the instrument is once again fit for use. An instrument that cannot be calibrated or is found to be out of tolerance must be replaced immediately.
5.
Do not reset the calibration gas concentration setpoints in
the ToxiLtd unless the concentrations of your calibration gas differ from the concentrations of the calibration gas that is normally supplied by Sperian for use in calibrating the ToxiLtd.
6.
Use of non-standard calibration gas and/or calibration kit
components when calibrating the ToxiLtd can lead to dangerously inaccurate readings and may void the standard Sperian warranty.
Sperian offers calibration kits and long-lasting cylinders of test gas specifically
developed for easy calibration. Customers are strongly urged to use only Sperian calibration materials when calibrating the ToxiLtd.
7.
8.
Substitution of components may impair intrinsic safety.
For safety reasons the ToxiLtd must be operated by
qualified personnel only. Read, understand and follow the directions set forth in this reference manual before operating the ToxiLtd.
The ToxiLtd has been tested for intrinsic safety in
9. Explosive Gas/AIR (max. 21.0% O
).
2
10. parts inside of the instrument.
Do not open the ToxiLtd. There are no user serviceable
4
Page 6
1. Overview
The ToxiLtd is a single sensor gas detector that is designed to detect either oxygen (O gases. The ToxiLtd’s sensor type is shown on the front of the instrument and is also shown on the display during the startup sequence.
ToxiLtd gas detectors are disposable instruments that are designed to operate for 2 full years from the date of activation. ToxiLtd models equipped with a toxic gas sensor (CO, H activated within 6 months of the manufacturing date. Oxygen versions must be activated within 1 month of the manufacturing date. If the instrument is not turned on by the “must be activated by date”, the instrument will activate itself and the 2-year countdown will begin automatically.
Upon delivery, the screen of the ToxiLtd will be blank except for the negated horn icon.
If the “must be activated by date” has passed and the instrument has activated itself, the instrument will show three lines on the screen with the negated horn icon.
Contact Sperian or your distributor if three lines are displayed when you first receive the detector.
The ToxiLtd is delivered with a pre­installed lithium thionyl chloride (LiSOCl battery capable of powering the instrument for the full 2 years of the instrument’s operating life. After the 2­year instrument life cycle is up, the instrument will turn itself off.
Note: Early production runs of the oxygen version of the ToxiLtd were referred to as the “OxyLtd”. The instructions contained in this manual that refer to the ToxiLtd Oxygen (O
) or one of a variety of toxic
2
S or SO2) must be
2
)
2
)
2
are valid for the OxyLtd with similar instrument firmware.
1.1 Methods of sampling
The ToxiLtd may be used in diffusion mode, or with the manual sample draw kit that is available separately. In either mode, the atmosphere must reach the sensor for the instrument to register a reading. In diffusion mode, the atmosphere reaches the sensor by diffusing through the sensor port on the front of the instrument. Normal air movements are enough to carry the sample to the sensor. During remote sampling, the gas sample is drawn into the sensor compartment through the probe assembly and a length of tubing.
See section 2.6 for more details on sampling the atmosphere.
1.2 Sensors
Toxic sensor versions of the ToxiLtd use an electrochemical toxic gas sensor. The ToxiLtd O
uses a galvanic oxygen
2
sensor. Both types of sensor have been designed to minimize the effects of common interfering gases. These sensors provide accurate, dependable readings for gases commonly encountered in industrial applications.
1.2.1 Sensor ranges and resolutions.
Sensor Range Resolution
CO 0-1000 PPM 5 PPM H2S 0 -200 PPM 2 PPM SO2 0 – 20 PPM 0.1 PPM O2 0-30%/Vol. 0.2%/Vol.
1.2.2 ToxiLtd O2 warm-up phase
The ToxiLtd O
requires a one-hour
2
warm-up phase from the moment of activation until actual use. See section
2.1.1 for further details.
1.2.3 F1: Missing sensor during startup
If the ToxiLtd fails to detect the sensor during startup, it will show “F 1” with the caution symbol following the software version screen at startup. This is covered in greater detail in section 2.1.3.
5
Page 7
1.3 Alarm logic
1.3.1 Gas alarms
When an alarm set point is exceeded a loud audible alarm sounds, the bright red LED alarm light flashes and the display will indicate whether it is a warning alarm or a more serious danger alarm. If the ToxiLtd is equipped with the optional vibrating alarm, it will also be activated during an alarm condition. For oxygen sensor-equipped instruments, the warning alarm is sounded for oxygen enrichment and the danger alarm is sounded for oxygen deficiency. Instruments equipped with a toxic gas sensor have two ascending alarm set points. The lower alarm set point is the warning alarm, while the higher alarm set point is the danger alarm.
Sensor Warning Alarm Danger Alarm
CO 35 PPM 100 PPM H2S 10 PPM 20 PPM SO2 2 PPM 5 PPM O2 23.5%/Vol. 19.5%/Vol.
Default Warning and Danger Alarm Settings
In the event of an alarm condition it is important to follow established procedures. The safest course of action is to immediately leave the affected area, and to return only after further testing determines that the area is once again safe for entry. Failure to immediately leave the area may result in serious injury or death.
1.3.2 Silence warning alarms
The ToxiLtd’s audible and vibrating (if so equipped) alarms can be turned off during an alarm condition by pressing the MODE button if this function has been enabled with BioTrak software. The visual warning alarm light and readings will continue to indicate the alarm.
1.3.3 Calibration due notices
The ToxiLtd includes calibration due notices that are displayed when the instrument is due for calibration.
For more information on the calibration due notices, see section 2.3 below.
1.3.4 Low battery alarms
The battery in the ToxiLtd is designed to power the instrument for a full two years
from the date of activation. In the unlikely event that the battery runs out of power prior to the completion of the 2-year operational lifespan, the ToxiLtd includes two battery warning alarms that are activated when the battery voltage is reduced to specific levels.
For more details on the battery alarm, see section 2.4 below.
1.4 Design components Case: The instrument is enclosed in a
solid PC (polycarbonate) case with TPE (rubber) overmold. Front face: The front face of the instrument houses the LCD display, MODE button, sensor port, LED and audible alarm port. LCD display: The liquid crystal display (LCD) displays gas readings, messages and other information. Visual alarm (LED) / IrDA port: A bright red LED (Light-Emitting Diode) alarm light provides a visual indication of the alarm state. The LED also functions as the IrDA port. MODE button: The large push-button on the front of the instrument is called the MODE button. The MODE button is used to turn the ToxiLtd on and off, to view the MAX screen and to initiate the automatic calibration sequence.
Figure 1. Exterior front view.
Sensor port: The sensor port is located
at the upper left corner of the instrument. A filter prevents unwanted contaminants from entering the sensor. Audible alarm orifice: A cylindrical resonating chamber contains the loud audible alarm.
6
Page 8
Bottom surface: The belt clip attaches to the bottom surface of the instrument.
1.5 Standard accessories
Standard accessories with every ToxiLtd include installed sensor and battery, reference manual and calibration/sample draw adapter.
1.6 Value pack kits
ToxiLtd value packs include all standard accessories, calibration fittings, 34-liter cylinder of calibration gas, and fixed flow rate regulator in a foam-lined, hard-shell carrying case.
2. Field Operation
Field operation of the ToxiLtd is controlled entirely through the MODE button, which is located on the front of the instrument. The MODE button is used to turn the ToxiLtd on and off (if desired), to access MAX gas readings for the current session and to initiate the fresh air/zero and span calibration sequences.
2.1 Turning the ToxiLtd on
The ToxiLtd is effectively disabled when it leaves the Sperian Factory. Upon arrival, the screen will be blank other than the negated horn icon.
Note: If the ToxiLtd arrives with three lines showing on the screen, it is an indication that the “must be activated by date” has passed and the instrument has already turned itself on. Contact Sperian or your distributor for further instructions.
The ToxiLtd is a disposable instrument. Once activated, it will operate continuously for 24 months. In the unlikely event that the battery runs out of power prior to the completion of the 2-year operating cycle, the ToxiLtd must be returned to Sperian or an authorized service center for the installation of a new battery.
There are no user serviceable parts inside of the ToxiLtd.
To initialize the ToxiLtd press and hold the MODE button for 5 seconds until the instrument chirps, then release the MODE button.
2.1.1 O
ToxiLtd O
version start up sequence
2
models require a one-hour
2
warm-up period prior to use. Once the MODE button is pressed and held for 5 seconds, the instrument will begin a 60­minute countdown. The instrument will be ready to use in 60 minutes.
When the warm-up period concludes, the instrument will show three dashes with the negated horn symbol.
Once three lines are shown, press and hold the MODE button for 1 second to start the instrument. The start-up sequence will begin. Proceed to section
2.1.3.
2.1.2 Toxic version start up sequence
ToxiLtd models equipped with a toxic gas sensor do not require a warm up period. Upon arrival, press and hold the MODE button for 5 seconds. The ToxiLtd will immediately proceed to the start-up sequence.
2.1.3 Start up sequence
During the start-up sequence, the instrument performs an electronic self test that will take approximately 30 seconds to complete. During the self-test sequence, the audible alarm will “chirp” and all sections of the display will be lit.
The ToxiLtd will then display the “on” screen followed by software version.
7
Page 9
If the sensor has become disconnected or removed from the instrument during shipping, the screen will show “F 1” with the caution symbol following the software version screen. If “F 1” is shown, the instrument will automatically shut down and show three lines on the screen. If “F 1” is shown and the ToxiLtd shuts itself down, contact Sperian or your local distributor for further information. Once the sensor is recognized, the serial number screens will be shown. Note that the 6 digit serial number is shown on two screens. In this case, the instrument serial number is 123456.
The sensor type will be shown next.
During the display of the warning alarm level, the LED alarm light will flash twice and the audible warning alarm will sound twice. The audible warning alarm is slow and warbling. During the display of the danger alarm level, the LED visual alarm will flash twice and the audible danger alarm will sound twice. The audible danger alarm is louder and a higher frequency than the warning alarm. Following the danger alarm level, the ToxiLtd will begin standard operation and will display the number of months
remaining in the 24-month lifecycle of the instrument. During an alarm condition, the instrument will display the current gas reading. For the first full month of use following initialization the display will show 23 full months remaining. Once the instrument determines that there are 90 days of operation remaining, the instrument will indicate 90 days and proceed to count the days down until the instrument reaches 0 days and turns itself off.
2.2 Functions
As stated above, the ToxiLtd will show the amount of time remaining in the product’s lifecycle during normal operation unless there is an alarm present. The ToxiLtd will display gas readings while it is in alarm.
2.2.1 View max gas readings
From the months remaining or the current gas readings screen, press the MODE button once to view the MAX gas values screen. The MAX figure represents the highest gas value reading that has been recorded by the instrument during the current operating session or during the last 24 hours of operation.
Note: To reset the max readings level, fresh air/zero calibrate the instrument as discussed in section 3.3.
The MAX screen will be shown for 5 seconds. The instrument will then automatically return to the months or days remaining screen (unless an alarm condition is present).
2.2.2 View alarm levels
To view the current gas alarm levels, from the current gas reading screen, press and hold the MODE button for about 5 seconds until the display test screen is shown.
8
Page 10
Release the MODE button. The software will proceed through the software version and serial number screens before showing the alarm levels.
The instrument will then return to normal operation and display the months remaining screen.
2.3 Calibration due notices
When the calibration due notice is enabled and the ToxiLtd is due for fresh air calibration “0-CAL will be displayed at the bottom of the current gas readings screen along with the triangular warning symbol.
When the calibration due notice is enabled and the ToxiLtd is due for span calibration the calibration bottle icon will be displayed at the bottom of the current gas readings screen along with the warning symbol. The bump test due notice is designed for instruments that are processed in an IQ Express Dock. When the bump test due notice is enabled and the ToxiLtd is due for a bump test, the calibration bottle icon will be displayed at the bottom of the current gas readings screen without the triangular warning symbol.
Note: If an IQ Express Dock is unavailable, a manual span calibration will also reset the bump test due notice.
The default fresh air/zero calibration interval for the O thirty days. Different versions of software may have different calibration intervals.
version of the ToxiLtd is
2
2.4 Low battery alarms
When the ToxiLtd determines that there are less than 7 days of battery life remaining, the battery icon will be lit.
Note that “MONTHS” or “DAYS” and the number of months or days of operating life remaining are still shown when the battery icon is lit.
When the ToxiLtd determines that there are less than 8 hours of battery life remaining, the battery icon and the triangular warning symbol on the LCD will be shown. At this point the audible alarm will begin to chirp once per minute.
When the battery reaches a level where it can no longer power the instrument, the ToxiLtd will sound the high/danger audible alarm while displaying OFF with the danger, caution and battery icons. Press MODE to turn the instrument off. If the MODE button is not pressed, the instrument will remain in alarm for as long as it can before shutting itself down. The following screen will be shown after the instrument is shut down. In the event that the instrument actually shuts itself down due to low battery levels, it will automatically launch the IrDA connection subroutine so that data in the instrument may be downloaded to BioTrak.
2.5 Turning the ToxiLtd off
To turn the ToxiLtd off, press and hold the MODE button. After 5 seconds, the instrument will light up all segments on the display. After 3 more seconds, the display will show “IrDA”. Continue to hold the MODE button for an additional 15 seconds after “IrDA” is shown. The following screen will be shown after the instrument is shut down.
Note: Turning the ToxiLtd off will have no effect on the operational life of the instrument. The ToxiLtd will operate
9
Page 11
for 24 months from the activation date regardless of whether the instrument is turned on or off.
2.6 Sampling
The ToxiLtd may be used in either diffusion or sample-draw mode. In either mode, the gas sample must enter the sensor compartment for the instrument to register a gas reading. In diffusion mode, the atmosphere reaches the sensor by diffusing through the sensor port on the front of the instrument. Normal air movements are enough to carry the sample to the sensor. The sensor reacts quickly to changes in the concentration of the gas being measured. It is also possible to use the ToxiLtd to sample remote locations with the hand­aspirated sample-draw kit that is available separately. During remote sampling, the gas sample is drawn into the sensor compartment through the probe assembly and a length of tubing.
2.6.1 Sample draw kit usage
1. Connect the shorter piece of hose
from the squeeze bulb to the sample draw adapter. Connect the longer end of the hose to the sample probe.
2. Attach the sample draw adapter to the
ToxiLtd.
3. Cover the end of the sample draw
probe assembly with a finger, and squeeze the aspirator bulb. If there are no leaks in the sample draw kit components, the bulb should stay deflated for a few seconds.
4. Insert the end of the sample probe into
the location to be sampled.
5. Squeeze the aspirator bulb several
times to draw the sample from the remote location to the sensor compartment. Allow one squeeze of the bulb for every one foot of sampling hose for the sample to reach the sensors. Then continue to squeeze the bulb for an additional 45 seconds or until readings stabilize.
6. Note the gas measurement readings.
CAUTION: Hand aspirated remote sampling only provides continuous gas readings for the area in which the probe is located when the bulb is being continuously squeezed.
Note: Each time a reading is desired, it is necessary to squeeze the bulb a sufficient number of times to bring a fresh sample to the sensor compartment and to then continue squeezing the bulb for an additional 45 seconds or until readings stabilize.
3. Calibration
The ToxiLtd features fully automated calibration functions. Instruments equipped with an oxygen sensor only require the fresh air/zero calibration, while toxic sensor-equipped instruments require both fresh air/zero and span calibrations. The MODE button is used to initiate the automatic calibration sequence. Calibration adjustments are made automatically by the instrument.
3.1 Verification of accuracy
Sperian recommends regular verification of accuracy for all of our gas detection products in order to maximize worker safety. For a discussion of Sperian’s Calibration Recommendations, see Appendix C.
3.1.1 Verifying accuracy and response: ToxiLtd O
The ToxiLtd O
uses a galvanic oxygen
2
sensor that puts out a continuous electrical signal (as long as oxygen is present), which the detector converts into a reading. As the oxygen sensor ages, the output signal tends to diminish gradually, which can eventually lead to erroneous readings.
The display on the ToxiLtd O a reading when it is in alarm, so it is difficult to know exactly when to fresh air/zero calibrate the detector. Sperian recommends that the fresh air/zero calibration be performed regularly on the ToxiLtd O
. The fresh air/zero calibration
2
takes less than ten seconds to perform and does not require an external gas source when it is performed in a fresh air environment.
Simply take the ToxiLtd O where the atmosphere is known to be fresh and perform a fresh air/zero calibration adjustment as discussed below in section 3.3. If fresh air is not available, see section 3.7 below for instructions on fresh air calibration in contaminated air.
10
2
only shows
2
to an area
2
Page 12
Sperian also recommends that the response of the oxygen sensor be regularly verified by any of the three following methods:
x Expose the O
sensor to a known
2
concentration of gas containing less than 19% oxygen. If the descending oxygen alarm is set to 19.5% the instrument should go into alarm a few seconds after the gas reaches the sensor face.
x Process the ToxiLtd O
in an IQ
2
Express Dock that has a cylinder of calibration gas containing 18.0% oxygen (or less) connected to the gas port and proceed with the oxygen bump test.
x Breath test: Hold your breath for 10
seconds, then slowly exhale directly onto the face of the sensor (in the same way you would to fog up a piece of glass). If the descending oxygen alarm is set to 19.5%, the instrument should go into alarm after a few seconds.
3.1.2 Verifying toxic sensor performance
Verification is a two-step procedure for ToxiLtd instruments equipped with a toxic gas sensor. Step one is to take the ToxiLtd to an area where the atmosphere is known to be fresh and perform a fresh air/zero calibration as discussed below in section
3.3. The fresh air/zero calibration takes
less than ten seconds to perform and does not require an external gas source when it is performed in a fresh air environment. Step two is to test sensor response by exposing the sensor to a test gas of known concentration. This is known as a functional (bump) test. Readings are considered to be accurate when the display is between 90%* and 120% of the expected values as given on the calibration gas cylinder. If readings are accurate, there is no need to adjust your gas detector. See section 3.4 for further details concerning the functional/bump test. If the readings are inaccurate, the instrument must be span calibrated before further use as discussed in section 3.5.
The Canadian Standards Association
* (CSA) requires the instrument to undergo calibration when the displayed value during a bump test fails to fall between 100% and 120% of the expected value for the gas.
Accuracy of
ToxiLtd instruments equipped with a toxic gas sensor should be checked periodically with known concentration calibration gas. Failure to check accuracy can lead to inaccurate and potentially dangerous readings.
See Appendix C for a discussion of calibration frequency.
3.2 Effect of contaminants on ToxiLtd sensors
The atmosphere in which the ToxiLtd is used can have lasting effects on the sensors. Sensors may suffer losses in sensitivity leading to degraded performance if exposed to certain substances. There are two basic types of sensors that may be installed in the ToxiLtd: galvanic oxygen or electrochemical toxic. Each type of sensor uses a slightly different detection principle, so the kinds of conditions that affect the accuracy of the sensors vary from one type of sensor to the next.
3.2.1 Effects of contaminants on oxygen sensors
Oxygen sensors may be affected by prolonged exposure to "acid" gases such as carbon dioxide. The oxygen sensors used in Sperian instruments are not recommended for continuous use in atmospheres containing more than 25%
.
CO
2
3.2.2 Effects of contaminants on toxic gas sensors
Sperian’s “substance-specific” electrochemical sensors have been carefully designed to minimize the effects of common interfering gases. “Substance-specific” sensors are designed to respond only to the gases that they are supposed to measure. The higher the specificity of the sensor, the less likely the sensor will react to other gases, which may be incidentally present in the environment. For instance, a
11
Page 13
“substance-specific” carbon monoxide sensor is deliberately designed not to respond to other gases that may be present at the same time, such as hydrogen sulfide (H
).
(CH
4
Although great care has been taken to reduce cross-sensitivity, some interfering gases may still have an effect on toxic sensor readings. In some cases the interference may be positive and result in readings that are higher than actual. In other cases the interference may be negative and produce readings that are lower than actual and may even cause the instrument to display negative readings for the target gas.
See Appendix A for cross-sensitivity data.
3.3 Fresh air/zero calibration
To initiate the fresh air/zero calibration procedure:
1. From the current gas reading screen, press the MODE button three times within two seconds. The ToxiLtd will briefly display “CAL” and then begin a 5-second countdown with the 0-CAL icon lit.
S) and methane
2
o
2. Press the MODE button before the end of the 5-second countdown to begin the fresh air/zero calibration. The fresh air/zero calibration has been successfully initiated when the ToxiLtd alternates between the following two screens:
when the instrument returns to the current gas reading screen.
Note: The maximum reading value (MAX) will be automatically reset following a successful fresh air/zero calibration.
Fresh air/zero calibrations may only be performed in an atmosphere that is known to contain 20.9% oxygen and 0 PPM toxic gas. Performing fresh air/zero calibrations in an atmosphere that is not comprised of 20.9% oxygen and 0 PPM toxic gas may lead to inaccurate and potentially dangerous readings.
3.4 Functional (bump) testing for ToxiLtd with toxic gas sensor
The accuracy of the ToxiLtd with a toxic gas sensor may be verified at any time by a simple functional (bump) test. To perform a functional (bump) test, do the following:
1. Turn the ToxiLtd on and wait at least
three minutes to allow the readings to fully stabilize.
2. Make sure the instrument is located in
fresh air.
3. Perform a fresh air/zero calibration as
discussed in section 3.3 and allow the instrument to return to the normal operation.
l
3. For instruments equipped with a toxic gas sensor, the fresh air/zero calibration is complete when the instrument begins a second 5-second countdown for the span calibration. If span calibration is not required, allow the countdown to reach 0 without pressing the MODE button. If the instrument is equipped with an oxygen sensor, calibration is complete
Figure 3.4: Bump-test/span calibration set­up for toxic sensor-equipped instruments.
4. Apply the calibration gas as shown in figure 3.4.
5. Wait for the readings to stabilize. Forty-five seconds to one minute is usually sufficient.
12
Page 14
6. Note the readings. Readings are considered accurate if they are between 90%* and 120% of the expected value. If the readings are considered accurate, then the instrument may be used without further adjustment. If readings are considered inaccurate, the instrument must be adjusted using the “span” calibration procedures discussed in section 3.5 before further use.
*The Canadian Standards Association (CSA) considers bump test readings accurate when the displayed values fall between 100% and 120% of the expected value for the gas. To meet the CSA requirement, an instrument must undergo calibration when the displayed value during a bump test fails to fall between 100% and 120% of the expected value for the gas.
3.5 Span Calibration for ToxiLtd
with toxic gas sensor
Span calibration should be performed when a functional (bump) test has shown that the instrument’s gas readings are not between 90%* and 120% of the expected values as given on the calibration gas cylinder (as discussed in section 3.4). *To meet the CSA requirement, an instrument must undergo calibration when the displayed value during a bump test fails to fall between 100% and 120% of the expected value for the gas. Prior to performing a span calibration, perform a fresh air/zero calibration as discussed in section 3.3. After successful completion of the fresh air/zero calibration, the instrument will begin a second five-second countdown with the calibration gas bottle icon highlighted.
1. Press the MODE button before the countdown is complete to initiate the span calibration. The display will alternate between “GAS” and the expected concentration of calibration gas.
l
2. Apply calibration gas as shown above in figure 3.4. Once calibration gas is detected, the readout will change to show the gas reading. Note that the negated horn symbol is shown to indicate that the alarms are not operating because the instrument is in calibration mode.
3. The calibration is fully automatic from this point on. Once the instrument successfully completes the span calibration, it will emit three short beeps and display the maximum span calibration adjustment value for two seconds.
Note: The maximum span calibration adjustment value shown is an indication of the relative health of the sensor. As a sensor loses sensitivity, the maximum adjustment level will approach the calibration gas concentration, letting you know when the sensor is losing sensitivity.
4. The instrument will then return to normal operation.
Note: Once the calibration cycle is completed, the ToxiLtd automatically returns to normal operation and the gas alarms may be activated. Disconnect the calibration assembly immediately after calibration.
Use of non­standard calibration gas and/or calibration kit components when calibrating the ToxiLtd can lead to inaccurate and potentially dangerous readings, and may void the standard Sperian Warranty.
13
Page 15
3.6 Failure to calibrate
3.6.1 Fresh air/zero calibration failure
In the event of fresh air/zero calibration failure, the “no” and “CAL” screens will be alternately displayed as shown below with the “0-CAL” segment lit. The instrument will then return to the gas reading screen.
l
Following a fresh air/zero calibration failure, the triangular warning symbol and “0-CAL” indicator will remain lit until a successful fresh air/zero calibration is performed.
3.6.1.1 Reason for fresh air/zero calibration failures
Fresh air/zero calibration failures often result from the attempt to calibrate the instrument in a contaminated atmosphere. Fresh air/zero calibration failures can also result from sensor failures.
3.6.2 Span calibration failure
The ToxiLtd is designed to recognize two distinct types of span calibration failure: failures that occur due to sensor response outside the sensor’s normal range for calibration and failures that occur when the instrument fails to recognize any calibration gas whatsoever.
3.6.2.1 Sensor out of range (no CAL)
If the instrument recognizes calibration gas, but the sensor response is not within the range to calibrate the instrument, span calibration will fail and the “no” and “CAL” screens will be alternately displayed.
l
After displaying “no” and “CAL” three times, the instrument will return to the current gas readings screen and the warning symbol and the calibration bottle icon will be shown, which signifies that the
instrument failed the last attempt to span calibrate.
3.6.2.2 No gas
In the case of a span calibration failure in which calibration gas is not detected, the “no” and “GAS” screens will be alternately displayed as shown below with the calibration bottle icon highlighted.
l
When the instrument returns to the current gas readings screen, the warning symbol and the calibration bottle icon will be shown, which signifies that the instrument failed the last attempt to span calibrate.
3.6.2.3 Causes for span cal failure
Span calibration failures can be caused by the following:
1. Expired calibration gas.
2. Calibration gas whose concentration
fails to match the concentration expected by the instrument.
3. Inappropriate regulator. The ToxiLtd
must be calibrated using a 1.0 liter/minute fixed flow regulator.
4. Sensor failure.
3.6.3 Forced fresh air/zero calibration
If a fresh air/zero calibration fails in an atmosphere known to be fresh, the ToxiLtd can be forced to fresh air calibrate as follows.
1. Follow instructions 1 and 2 in section
3.3 to begin the fresh air/zero calibration sequence.
2. As soon as the alternating right and
left 0’s are shown on the screen, press and hold the MODE button.
14
Page 16
l
3. The forced fresh air/zero calibration is complete when the instrument emits three short beeps. Units equipped with an oxygen sensor will then return to gas detection mode. Units equipped with a toxic gas sensor will then move on the span calibration procedure.
Performing the forced fresh air/zero calibration in a contaminated atmosphere may lead to inaccurate and potentially dangerous readings.
3.7 Fresh air/zero calibration in a contaminated atmosphere
To perform the fresh air/zero calibration in a contaminated atmosphere, it is necessary to use special calibration gas, whose composition is identical to that of fresh air. Sperian offers the “Zero Air” calibration gas cylinder as part number 54-9039, which contains 0 PPM toxic gas and 20.9% oxygen.
1. Apply “Zero Air” calibration gas to the
instrument as shown above in figure
3.4 for at least 15 seconds or until the readings fully stabilize.
2. Perform the fresh air/zero calibration
procedure as described in section 3.3.
3. Once the fresh air/zero calibration is
complete, disconnect the calibration assembly and move on to the span calibration if necessary as described in section 3.5.
4.1 Initiating communications
To initiate communications, simply hold the MODE button as if you are turning the instrument off. The ToxiLtd will briefly light all segments in the display after about 5 seconds. Continue to hold the MODE button for an additional 2-3 seconds until the “IrdA” screen is shown and the LED/IrDA port blinks. Then release the MODE button.
Align the infrared port at the top front of the ToxiLtd with the PC’s infrared port and proceed with the download. The ToxiLtd should be located 1-2” (2.5-5cm) from the IrDA Port.
Position of IrDA Module and ToxiLtd
During download.
If 30 seconds pass and the ToxiLtd has not detected an infrared interface, it will return to normal operation.
4. Event logging
Each ToxiLtd includes a built-in event logger that stores instrument readings and other data whenever a reading exceeds a pre-set alarm point.
Incident information can be extracted through a PC with BioTrak software and a functioning IrDA port. The ToxiLtd may also be returned to Sperian to have the information extracted.
If you need to return the instrument to Sperian, call Sperian’s Technical Service Department for a Return Authorization (RA#) prior to shipping the instrument.
15
Page 17
2
H
2
C
4
H
2
C
3
HCN HCl NH
2
H
2
ClO
2
Calibration Gas Part Number
Default
58 or 103 Liter 34 Liter
Setting
54-9037 54-9058
2
S 54-9034 54-9057
2
16
Cl
2
NO NO
2
S SO
2
CO H
Appendices
Appendix A: Sensor Cross-Sensitivity Chart
The table below provides the cross-sensitivity response of the ToxiLtd toxic gas sensors to common interference gases. The values are ToxiLtd as
a percentage of the primary sensitivity or the reading of the sensor when exposed to 100ppm of the interfering gas at 20ºC. These values are
approximate. The actual values depend on the age and condition of the sensor. Sensors should always be calibrated to the primary gas type.
Cross-sensitive gases should not be used as sensor calibration surrogates without the written consent of Sperian.
SENSOR
100 10 5 10 -15 -5 -15 50 15 3 0 75 250
Carbon
Monoxide
(CO)
0.5 100 20 2 -20 -20 -60 0.2 0 0 0 n/d n/d
S)
ulfide
Hydrogen
(H
2
1 1 100 1 -100 -50 -150 0.2 n/d n/d 0 (+) (+)
ulfur
Dioxide
)
2
(SO
Calibration Gas
Description
ToxiLtd
Appendix B: Default toxic sensor calibration gas settings
n/d = no data, (+) undetermined positive, (-) undetermined negative n/d = no data, (+) undetermined positive, (-) undetermined negative
Part Number
S Hydrogen sulfide 25 PPM H
2
54-45-01 CO Carbon monoxide 50 PPM CO 54-9033 54-9036
54-45-02 H
Sulfur dioxide 10 PPM SO
2
54-45-03 SO
Page 18
Appendix C: Calibration Frequency
One of the most common questions that we are asked at Sperian Instrumentation is: “How
often should I calibrate my gas detector?”
Sensor Reliability and Accuracy
Today’s sensors are designed to provide years of reliable service. In fact, many sensors are designed so that with normal use they will only lose 5% of their sensitivity per year or 10% over a two-year period. Given this, it should be possible to use a sensor for up to two full years without significant loss of sensitivity.
Verification of Accuracy
With so many reasons why a sensor can lose sensitivity and given the fact that dependable sensors can be key to survival in a hazardous environment, frequent verification of sensor performance is paramount.
There is only one sure way to verify that a sensor can respond to the gas for which it is designed. That is to expose it to a known concentration of target gas and compare the reading with the concentration of the gas. This is referred to as a “bump” test. This test is very simple and takes only a few seconds to accomplish. The
safest course of action is to do a “bump” test prior to each day’s use. It is not necessary to
make a calibration adjustment if the readings fall between 90%* and 120% of the expected value. As an example, if a CO sensor is checked using a gas concentration of 50 PPM it is not necessary to perform a calibration unless the readings are either below 45 PPM or above 60 PPM.
*The Canadian Standards Association (CSA) requires the instrument to undergo calibration when the displayed value during a bump test fails to fall between 100% and 120% of the expected value for the gas.
Lengthening the Intervals
between Verification of
Accuracy
We are often asked whether there are any circumstances in which the period between accuracy checks may be lengthened.
Sperian Instrumentation is not the only manufacturer to be asked this question! One of the professional organizations to which Sperian Instrum entation belongs is the Industrial Safety Equipment Association (ISEA). The “Instrument Products” group of this organization has been very active in developing a protocol to clarify the minimum conditions under which the interval between accuracy checks may be lengthened.
A number of leading gas detection equipment manufacturers have participated in the development of the ISEA guidelines concerning calibration frequency. Sperian Instrum entation’s procedures closely follow these guidelines.
If your operating procedures do not permit daily checking of the sensors, Sperian Instrumentation recommends the following procedure to establish a safe and prudent accuracy check schedule for your Sperian instrum ents:
1. During a period of initial use of at least 10 days in the intended atmosphere, check the sensor response daily to be sure there is nothing in the atmosphere that is poisoning the sensor(s). The period of initial use must be of sufficient duration to ensure that the sensors are exposed to all conditions that might have an adverse effect on the sensors.
2. If these tests demonstrate that it is not necessary to make adjustments, the time between checks may be lengthened. The interval between accuracy checking should not exceed 30 days.
3. When the interval has been
extended the toxic and combustible gas sensors should be replaced immediately upon warranty expiration. This will minimize the risk of failure during the interval between sensor checks.
4. The history of the instrument response between verifications should be kept. Any conditions, incidents, experiences, or exposure to contaminants that might have an adverse effect on the calibration state of the sensors should trigger
imm ediate re-verification of accuracy before further use.
5. Any changes in the environment in which the instrum ent is being used, or changes in the work that is being performed, should trigger a resum ption of daily checking.
6. If there is any doubt at any time as to the accuracy of the sensors, verify the accuracy of the sensors by exposing them to known concentration test gas before further use.
Gas detectors used for the detection of oxygen deficiencies, flammable gases and vapors, or toxic contaminants must be maintained and operated properly to do the job they were designed to do. Always follow the guidelines provided by the manufacturer for any gas detection equipment you use!
If there is any doubt regarding your gas detector's accuracy, do an accuracy check! All it takes is a few moments to verify whether or not your instrum ents are safe to use.
One Button Auto Calibration
While it is only necessary to do a “bump” test to ensure that the sensors are working properly, all current Sperian gas detectors offer a one-button auto calibration feature. This feature allows you to calibrate a Sperian gas detector in about the same time as it takes to complete a “bump” test. The use of automatic bump test and calibration stations can further simplify the tasks, while automatically maintaining records.
Don't take a chance
with your life.
Verify accuracy frequently!
Please read also Sperian Instrumentation’s application note: AN20010808 “Use of
‘equivalent’ calibration gas mixtures”. This application note
provides procedures to ensure safe calibration of LEL sensors that are subject to silicone poisoning.
Biosystems website is located at
http://www.biosystems.com
17
Page 19
Sperian Instrumentation Warranty Gas Detection Products
General
Sperian Protection Instrumentation, LLC (hereafter Sperian) warrants gas detectors, sensors and accessories manufactured and sold by Sperian, to be free from defects in materials and workmanship for the periods listed in the tables below.
Damages to any Sperian products that result from abuse, alteration, power fluctuations including surges and lightning strikes, incorrect voltage settings, incorrect batteries, or repair procedures not made in accordance with the Instrument’s Reference Manual are not covered by the Sperian warranty.
The obligation of Sperian under this warranty is limited to the repair or replacement of components deemed by the Sperian Instrument Service Department to have been defective under the scope of this standard warranty. To receive consideration for warranty repair or replacement procedures, products must be returned with transportation and shipping charges prepaid to Sperian at its manufacturing location in Middletown, Connecticut, or to a Sperian Authorized Warranty Service Center. It is necessary to obtain a return authorization number from Sperian prior to shipment.
THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND REPRESENTATIONS, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO, THE WARRANTY OF FITNESS FOR A PARTICULAR PURPOSE. SPERIAN WILL NOT BE LIABLE FOR LOSS OR DAMAGE OF ANY KIND CONNECTED TO THE USE OF ITS PRODUCTS OR FAILURE OF ITS PRODUCTS TO FUNCTION OR OPERATE PROPERLY.
Instrument & Accessory Warranty Periods
Product(s) Warranty Period
PhD5, PhD Lite, PhD Plus, PhD Ultra, Cannonball3, MultiVision, Toxi, Toxi/Oxy Plus, Toxi/Oxy Ultra, ToxiVision, Ex Chek
ToxiPro®, MultiPro
ToxiLtd®
Toxi3Ltd®
Mighty-Tox
Mighty-Tox 2 Prorated credit is given towards repair or purchase of a new unit of the same type.
IQ Systems, Series 3000, Airpanel, Travelpanel, ZoneGuard, Gas9Chek1 and Gas9Chek4
Battery packs and chargers, sampling pumps and other components, which by their design are consumed or depleted during normal operation, or which may require periodic replacement
As long as the instrument is in service
2 years from date of purchase
2 years after activation or 2 years after the “Must Be Activated By” date, whichever comes first
3 years after activation or 3 years after the “Must Be Activated By” date, whichever comes first
90 days after activation or 90 days after the “Must Be Activated By” date, whichever comes first
0 – 6 months of use 100% credit 6 – 12 months of use 75% credit 12 – 18 months of use 50% credit 18 – 24 months of use 25% credit
One year from the date of purchase
One year from the date of purchase
Sensor Warranty Periods
Instrument(s) Sensor Type(s) Warranty Period
PhD Plus, PhD Ultra, PhD5, PhD Lite, Cannonball3, MultiVision, MultiPro, ToxiVision,
ToxiPro®, Ex Chek
Toxi, Toxi/Oxy Plus, Toxi/Oxy Ultra
All Others All Sensors 1 Year
** Damage to combustible gas sensors by acute or chronic exposure to known sensor poisons such as volatile lead (aviation gasoline additive), hydride gases such as phosphine, and volatile silicone gases emitted from silicone caulks/sealants, silicone rubber molded products, laboratory glassware greases, spray lubricants, heat transfer fluids, waxes & polishing compounds ( neat or spray aerosol s), mold release agen ts for plastics inject ion molding operations, waterproofing formulations, vinyl & leather preservatives, and hand lotions which may contain ingredients listed as cyclomethicone, dimethicone and polymethicone (at the discretion of Sperian’s Instrument Service department) void Sperian Instrumentation’s Standard Warranty as it applies to the replacement of combustible gas sensors.
O2, LEL**, CO, CO+, H2S & Duo-Tox
All Other Sensors 1 Year CO, CO+, H2S 2 Years All Other Sensors 1 Year
2 Years
18
Page 20
®
ToxiLtd
c-UL Version
ATEX / CQST Version
IECEx Version
Gas O2 CO H
S SO2
2
Firmware Ver. _______
Vibrating Motor Belt Clip
Serial # ______________
Activation Date
19
Loading...