Sperian SURVIVAIR Opti-Fit 7641, SURVIVAIR Opti-Fit 7540, SURVIVAIR Opti-Fit 7640, SURVIVAIR Opti-Fit 7741, SURVIVAIR Opti-Fit 7740 Operation And Maintenance Manual

...
Page 1
SURVIVAIR Opti-Fit™ Convertible Facepiece
Models 7540/7640/7740/7541/7641/7741
OPERATION and MAINTENANCE MANUAL
for Air Purifying Respirator (APR), Continuous Flow Supplied Air Respirator
(CF–SAR), and Powered Air Purifying Respirator (PAPR) Congurations
Air Purifying Respirator
Front Adapter Cap
Air Purifying Respirator
Low Prole Nozzle Plug
CF-Supplied Air Respirator
Side-mount Breathing Tube
SURVIVAIR Opti-Fit
Convertible Facepiece
CF-Supplied Air Respirator
Front-mount Breathing Tube
Powered Air Purifying
Respirator Mask-mount Blower
Powered Air Purifying
Respirator Belt-mount Blower
CF-Supplied Air Respirator
P100 Filter Escape
DO NOT USE your respirator until you read and completely understand this instruction manual. You are required to inspect the adapter prior to putting it into service. Please refer to the inspection procedures in this manual. Failure to comply with this warning may result in personal injury, illness, or death.
©
Copyright SPERIAN Respiratory Protection USA, LLC April 2008 All Rights Reserved P/N 764099 REV D 03/09
WARNING
AN ISO 9001
REGISTERED
COMPANY
SPERIAN Respiratory Protection USA, LLC 3001 South Susan St., Santa Ana, CA 92704 Toll-Free 888.APR.SCBA or 714.545.0410 Fax 714.850.0299 www.sperianprotection.com
Page 2
NIOSH APPROVAL CAUTIONS AND LIMITATIONS
The Cautions and Limitations applicable to the air purifying respirator (APR), continuous ow supplied air respirator (CF-SAR), and powered air purifying respirator (PAPR) congurations are listed below:
Operating Conguration Cautions and Limitations
Air Purifying Respirator (APR) ABCHJLMNOPS
Continuous Flow Supplied Air Respirator (SAR) ABCDEGHJLMNOPS
Powered Air Purifying Respirator (PAPR) ABCFHIJLMNOPS
Cautions and Limitations Key:
A - Not for use in atmospheres containing less than 19.5 percent oxygen.
B - Not for use in atmospheres immediately dangerous to life or health.
C - Do not exceed maximum use concentrations established by regulatory standards.
D - Air line respirators can be used only when the respirators are supplied with respirable air meeting the requirements
of CGA G-7.1 Grade D or higher quality.
E - Use only the pressure ranges and hose lengths specied in the User’s Instructions.
F - Do not use respirator if airow is less than four cfm (115 lpm) for tight tting facepieces or six cfm (170 lpm) for hoods
and/or helmet.
G - If airow is cut off, switch to lter and/or cartridge and immediately exit to clean air.
H - Follow established cartridge and canister change schedules or observe ESLI to ensure that cartridges and canisters
are replaced before breakthrough occurs.
I - Contains electrical parts which have not been evaluated as an ignition source in ammable or explosive atmospheres
by MSHA/NIOSH.
J - Failure to properly use and maintain this product could result in injury or death.
L - Follow the User’s Instructions for changing cartridges, canisters and/or lters.
M - All approved respirators shall be selected, tted, used, and maintained in accordance with MSHA, OSHA, and other
applicable regulations.
N - Never substitute, modify, add, or omit parts. Use only exact replacement parts in the conguration as specied by
the manufacturer.
O - Refer to Users Instructions, and/or maintenance manuals for information on use and maintenance of these respirators.
P - NIOSH does not evaluate respirators for use as surgical masks.
S - Special or critical User’s Instructions and/or specic use limitations apply. Refer to User’s Instructions before donning.
i
Page 3
SURVIVAIR Opti-Fit™ Convertible Facepiece
TABLE OF CONTENTS
NIOSH APPROVAL CAUTIONS AND LIMITATIONS ....................................................i
TABLE OF CONTENTS .........................................................................ii
I. INTRODUCTION.............................................................................1
II. SAFETY PRECAUTIONS......................................................................1
III. DESCRIPTIONS OF RESPIRATOR OPERATING CONFIGURATIONS .................................1
A. Air Purifying Respirator (APR) Conguration ...................................................1
B. Supplied Air Respirator (CF-SAR) Conguration ................................................1
C. Powered Air Purifying Respirator (PAPR) Conguration ...........................................2
IV. General Warnings for All Respirator Congurations .................................................2
V. RESPIRATOR OPERATING CONFIGURATIONS SCHEMATIC ........................................5
VI. PARTS LIST ...............................................................................6
Table 1. Air Purifying Respirator (APR) Filters and Cartridges ........................................8
Table 2. Continuous Flow Supplied Air Respirator (CF-SAR) Air Hose Components .......................8
Table 3. Powered Air Purifying Respirator (PAPR) Belt-Mounted Cartridges .............................8
VII. FACEPIECE PREPARATION USING INCLUDED ITEMS............................................9
VIII. OPERATING INSTRUCTIONS FOR SPECIFIC RESPIRATOR CONFIGURATIONS .....................10
A. Air Purifying Respirator (APR) Conguration ..................................................10
1. Installing Chemical Cartridges and Particulate Filters ......................................... 11
2. Donning Instructions for APR Conguration.................................................12
3. Fit Check Instructions for APR Conguration ................................................13
4. How to Use the Respirator ..............................................................13
5. Dofng Instructions for APR Conguration..................................................13
6. Chemical and Particulate Filter Service Life and Replacement ..................................13
B. Continuous Flow Supplied Air Respirator (CF-SAR) Congurations.................................15
1. Air Supply ...........................................................................16
2. Quick Coupler Assembly ...............................................................16
3. Breathing Tube Setup..................................................................17
4. Donning Instructions for CF-SAR Congurations .............................................18
5. Fit Check Instructions for CF-SAR Congurations ............................................19
6. How to Use the Respirator ..............................................................20
7. Dofng Instructions for CF-SAR Congurations ..............................................21
8. Pressure Tables for Continuous Flow Supplied Air Respirator (CF-SAR) ..........................21
9. P100 Filter Service Life and Replacement ..................................................22
C. Powered Air Purifying Respirator (PAPR) Congurations .........................................22
1. Battery Pack Charging .................................................................22
2. Cartridge and Filter Installation...........................................................23
3. Pre-operational Flow Check .............................................................24
4. Donning Instructions for PAPR Congurations ...............................................26
5. Fit Check Instructions for PAPR Congurations ..............................................27
6. How to Use the Respirator ..............................................................28
7. Dofng Instructions for PAPR Congurations ................................................28
8. Chemical and Particulate Filter Service Life and Replacement ..................................28
9. Decontamination......................................................................30
IX. MAINTENANCE ...........................................................................30
A. Cleaning ..............................................................................30
B. Inspection .............................................................................31
C. Repair ................................................................................32
D. Storage ...............................................................................32
X. SHIPMENT................................................................................32
XI. OVERHAUL FREQUENCY...................................................................32
XII. ADDITIONAL INFORMATION ................................................................32
XIII. WARRANTY AND LIMITATION OF LIABILITY ...................................................33
XIV. RESPIRATOR INSPECTION TABLES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
ii
Page 4
SURVIVAIR Opti-Fit™ Convertible Facepiece
I. INTRODUCTION
This manual provides instructions for the use and main­tenance of the various respirator operating congura­tions achievable with the SURVIVAIR Opti-Fit™ Convert­ible full facepiece respirator models 7540, 7640, 7740,
7541, 7641, and 7741. The base conguration of the
SURVIVAIR Opti-Fit facepiece allows use as an air pu­rifying respirator (APR) in conjunction with S Series par-
ticulate lters, gas and vapor cartridges, and cartridge/ lter combinations as required for specic contaminant
protection. Additional components can be attached to the SURVIVAIR Opti-Fit Convertible to easily convert it to a front or side hose connection continuous ow sup­plied air respirator (CF-SAR) and mask or belt mounted
powered air purifying respirator (PAPR) congurations.
Before using a respirator in a contaminated work place, you must, at a minimum, understand the ele­ments of a respirator program, your work place, and the contaminants you are trying to protect yourself against. Determining the applicability of a respirator
is difcult. You must be able to calculate the service
life of the cartridges, determine the type and con­centration of contaminants, and determine whether the contaminants have poor warning properties, etc. If you do not have a complete understanding of the use of respirators, you must seek the aid of
a certied industrial hygienist or contact SPERIAN.
You must read and understand this manual and be trained in the proper use of the respirator before wearing it in a contaminated atmosphere.
II. SAFETY PRECAUTIONS
The Warnings, Cautions, and Notes contained in this
instruction sheet have the following signicance
WARNING
Maintenance or operating procedures and techniques that may result in seri­ous personal injury, serious personal ill­ness, or death if not carefully followed.
CAUTION
Maintenance or operating procedures and techniques that may result in damage to equipment and/or minor to moderate per­sonal injury if not carefully followed.
NOTE
Maintenance or operating procedures and
techniques or information considered impor-
tant enough to emphasize.
III. DESCRIPTIONS OF RESPIRATOR OPERATING CONFIGURATIONS
The SURVIVAIR Opti-Fit full facepiece contains two connectors on each side of the lens for attaching par-
ticle lters, gas/vapor cartridges or air hoses, as well
as a front adapter for a supplied air breathing hose or powered air purifying blower. This range of connections
allows the SURVIVAIR Opti-Fit Convertible facepiece to be congured, through the use of additional compo­nents, as a supplied air respirator (CF-SAR) or powered air purifying respirator (PAPR).
The full facepiece is designed with a wide lip silicone
skirt and ve point head strap harness (models 7540,
7640, and 7740). The facepiece is also available with a Headnet™ retention system (models 7541, 7641, and
7741). All facepieces contain a lens treated with an abrasion resistant coating. The facepiece is available in three sizes, small, medium and large. While a nose cup is standard, the facepiece may be equipped with an optional spectacles kit, clear or tinted lens covers and neck strap kit.
The connection versatility of the SURVIVAIR Opti-Fit™ Convertible full facepiece and available components and accessories (see Section VI, Parts List) allow it to be as-
sembled into the following operating congurations:
A. Air Purifying Respirator (APR) Conguration
The side port connectors are used to attach S series
particulate lters, gas and vapor cartridges, and car­tridge/lter combinations as required for a specic con-
taminant protection. The front adapter can be removed
and replaced with a plug to reduce the front prole of
the mask.
B.Continuous Flow Supplied Air Respirator
(CF-SAR) Conguration
With one side port connector port capped with a sup­plied screw cap, a high volume, low pressure breath­ing tube can be attached to the second side connector
to provide a constant ow of air to the facepiece when
connected to an external air source. An alternate con­nection option uses a breathing tube connected to the front adapter with both side port connectors capped. An optional Vortex air cooler may also be connected in-line to provide cool air to the facepiece.
NOTE
The Vortex air cooler cannot be used in
conjunction with ambient air pump supply
systems.
The SURVIVAIR Opti-Fit™ Convertible can also be con-
gured as a CF-SAR with P100 lter cartridges and the
breathing tube connected to the front adapter. When the air line hose is disconnected from the air line quick cou­pler, a check valve in the air line couple assembly pre­vents leakage of contaminated air through the facepiece
hose, and the wearer’s air is drawn through the P100 lters. This feature allows for safe entry, exit or escape
from a dangerous area in case of air supply failure.
For all CF-SAR congurations, when used at pressures specied in the tables in Section VIII, Part B.8, air is de-
livered to the facepiece at a rate between 115 and 425 liters per minute (4-15 cubic feet per minute [cfm]).
1
Page 5
SURVIVAIR Opti-Fit™ Convertible Facepiece
WARNING
The pressure within the full facepiece
continuous ow supplied air respirator
(CF-SAR) remains positive under most working conditions, but as with all respi­rators, negative pressure excursions are possible. Conditions when a respirator can experience negative facepiece pres­sures include, but are not limited to:
1) the respirator is improperly worn,
2) the respirator is not used in accor­dance with the instructions, or
3) the respirator is over-breathed during heavy work rates. The respirator will provide reduced protection when oper­ated in a negative pressure mode. Fail­ure to comply with this warning may re­sult in personal injury, illness, or death.
The ANSI Z88.2-1992 assigned protection factor (APF) for full facepiece continuous ow atmosphere supply­ing respirators is 1000. An APF of 1000 means that this device, when properly used, will reduce exposure by a factor of 1000.
C. Powered Air Purifying Respirator (PAPR)
Conguration
The SURVIVAIR Opti-Fit™ Convertible full facepiece can be congured into a mask-mounted or belt-mount­ed powered air purifying respirator. In the mask-mount-
ed conguration, a high efciency (HE) lter/blower
mounts directly to the nozzle of the facepiece and deliv-
ers air to the facepiece at a ow rate greater than 115
liters per minute (4 cubic feet per minute). The blower is powered by a belt mounted rechargeable nickel cad­mium battery pack equipped with an ON/OFF switch.
This conguration is NIOSH approved and SPERIAN
recommended for respiratory protection against dusts, fumes and mists, including asbestos-containing dusts and mists, particulate radionuclides, and radon daugh­ters attached to dusts, fumes and mists.
The belt-mounted PAPR respirator uses the SURVIVAIR Opti-Fit™ Convertible facepiece, a rubber breathing tube, a belt mounted blower with three (3) cartridges or lters and a rechargeable Nickel Cadmium (NiCad) bat­tery pack. Filtered breathing air is provided to the face-
piece at a ow rate greater than 170 liters per minute
(6 cubic feet per minute). To maintain NIOSH approval, only the equipment and accessories listed on the ap­propriate NIOSH approval label may be used together.
IV. GENERAL WARNINGS FOR ALL RESPIRATOR CONFIGURATIONS
The following general warnings must be observed for proper operation of all air purifying (APR), continuous
ow supplied air (CF-SAR) and powered air purifying (PAPR) respirator congurations. Additional warnings for specic operating congurations are listed in Section VIII,
Parts A, B, and C, and must also be observed for proper
operation of the respective respirator conguration.
General Warnings for All Respirator Congurations:
WARNING
● Your employer (or you, if you do not
have an employer) is responsible for establishing that this respirator is suitable for the user’s application.
● SPERIAN cannot predict what will
happen to this respirator in every po­tential environment. Materials can be chemically attacked if exposed to the wrong environment and may exhibit excessive corrosion or other forms of damage. Permeation or penetration of gases, liquids or particles through the materials could be excessive. Ex­tremes of temperature might cause thermal degradation. Each of these things, or a combination of them, could create conditions in which this respi­rator would be dangerous to use.
● Before entering a hazardous environ­ment while wearing this respirator,
you must conduct safe, scientic
tests to determine if the environment could render the equipment unsafe. Results of this testing should be well documented. Seek the help of a certi-
ed safety professional or industrial
hygienist. DO NOT USE this equip­ment if the user would be endangered in any way through environmentally induced degradation of the materials in the apparatus.
● All persons using this SPERIAN
breathing apparatus must be made aware of its limitations. We cannot be responsible for any damage to property, personal injury, or death in which environmental exposure is a contributing factor.
● Do not use this respirator if anything
comes between the facepiece and your face (e.g., hair, bandanna, head covering, etc.). This condition will prevent a good facepiece seal and may allow the leakage of contami­nants into the facepiece.
● Do not use this respirator in environ­ments where the concentrations of con­taminants are unknown or are immedi­ately dangerous to life or health (IDLH).
IDLH atmospheres are dened as:
a. Those that the wearer could not
breathe for short periods.
b. Those from which the wearer
could not escape without the aid of a respirator.
c. Those which have an immediate or
delayed adverse effect on health.
● Do not use this respirator at ambient
temperatures above 130°F (55°C).
2
Page 6
SURVIVAIR Opti-Fit™ Convertible Facepiece
WARNING─Continued
● Do not use this respirator where en-
vironmental conditions could cause the respirator facepiece to be dis­lodged or slip.
● Immediately return to a non-contam­inated area if:
You taste or smell contaminants, a. or if your eyes, nose, or throat become irritated. Breathing becomes difcult.b. The air you are breathing be-c. comes uncomfortably warm. You feel nauseous or dizzy.d. You notice a loss or decrease in e.
airow.
The facepiece moves, slips, or f. leaks.
● This respirator does not protect ex­posed areas of the body. Some con­taminants can be absorbed directly through the skin while others may irritate exposed areas.
● This respirator does not provide pro­tection from hazardous rays or harm-
ful noise. Always wear proper head and ear protection.
● This respirator must not be used
underwater, for fumigation, inte­rior structural fire fighting, abrasive blasting operations, or in areas where high heat, sparks, or flames could contact the respirator. If this respi­rator will be used during welding, it is the user’s responsibility to obtain the proper equipment for protection against sparks, optical radiation, and impact. This respirator will not pro-
tect you against these hazards.
● This respirator must be used in con-
junction with a written respirator program meeting the requirements of the OSHA Standard for Respira­tory Protection, 29 CFR 1910.134, available from the U.S. Department of Labor, Occupational Safety and Health Administration. The program must include procedures for evalu­ating air contaminants and selecting appropriate respirators; procedures for proper use of respirators; proce­dures for testing the facepiece fit of respirators; procedures for cleaning, disinfecting, inspecting, maintaining, and storing respirators; procedures for determining if workers are physi­cally and medically capable of wear­ing respirators; and procedures for training employees in the use and
care of respirators and in recogniz­ing the hazards associated with con-
taminants in the work place.
WARNING─Continued
● Do not wear this respirator if a satis-
factory fit, as determined by a quali­tative or quantitative fit test, cannot be obtained. See ANSI Z88.2, latest issue, and OSHA 29 CFR 1910.134, latest edition. Beards, stubble, or sideburns will prevent a good face­piece seal, and facial hair may inter­fere with valve function. Do not use this respirator unless you are clean shaven. Absence of one or both den­tures can seriously affect the fit of the respirator.
● This respirator will reduce, but will
not eliminate the inhalation of con­taminants. Some sensitive individu­als may experience health prob­lems when exposed to even minute amounts of contaminants. This res­pirator will not prevent health prob­lems to those individuals.
● Use of components other than those
listed on the NIOSH approval label
and in this manual, or as authorized
by SPERIAN, or modification of this respirator in any manner will void the NIOSH certification and invali­dates all SPERIAN’s warranties for the respirator.
● Always read and follow the instruc­tions listed in the Material Safety Data Sheet for the chemicals that are present in the work area.
● Do not use if you have a preexisting
skin condition (for example, folliculi­tus or vitiligo) until you obtain clear­ance from a medical doctor.
● This respirator must be worn and
used as specified in SPERIAN’s in­structions. No respirator can pro­vide complete protection from all conditions. Use extreme care for emergency conditions.
● Some individuals are sensitive to
chemicals (e.g., isocyanates, latex, oil mists, etc.) or may have some type of respiratory disorder (e.g. asthma, chronic obstructive airway disease, etc.). If you are sensitive to any chemical or have a respiratory disorder, you may have a severe reac­tion at contaminant levels well below accepted health levels, such as the OSHA Permissible Exposure Limit (PEL), ACGIH Threshold Limit Value (TLV), or the NIOSH Recommended Exposure Limits (REL). Many chemi­cals (e.g., isocyanates, Mercury, etc.) have no physical warning properties and you cannot taste or smell the contaminants even though they may
3
Page 7
SURVIVAIR Opti-Fit™ Convertible Facepiece
WARNING─Continued
be present in the respirator face­piece. This respirator will reduce, but will not eliminate the possibility of contaminants entering the facepiece and causing a severe reaction. Do not use this respirator until you obtain clearance from a medical doctor.
● Users must clean and maintain this
respirator only in accordance with SPERIAN’s instructions. Accessories not offered by SPERIAN may degrade performance, and will void NIOSH certification.
● The respirator facepiece assembly
contains natural rubber latex, which may cause allergic reactions in some individuals. Discontinue use if you experience an allergic reaction.
● Discontinue use if you experience
skin irritation or discoloration.
● This product may provide less than
adequate protection if improperly used, which may result in personal injury, illness, or death.
● For use by trained, qualified personnel
only.
● Do not use this respirator if it has
been left exposed in a contaminated environment as contaminants could collect on the inside of the facepiece and be inhaled when the gas mask is donned.
● Routinely used respirators must be
inspected before and after each use. Stored emergency respirators must be inspected after each use and at least every 30 days. A written record must be kept of emergency respira­tor inspections.
FAILURE TO OBSERVE ALL WARN­INGS MAY RESULT IN PERSONAL IN­JURY, ILLNESS, OR DEATH.
4
Page 8
SURVIVAIR Opti-Fit™ Convertible Facepiece
V. RESPIRATOR OPERATING CONFIGURATIONS SCHEMATIC
Cartridges and
Filters See
cartridge table
in Section VI
for available
cartridge part
numbers.
Side-mount
Tube and Belt
Kits*:
Light Duty P/N 981415 Heavy Duty P/N 981416
SURVIVAIR Opti-Fit
Convertible Facepiece
5-Pt. Strap
Small/Medium/Large
754000/764000/774000
Mesh Headnet
Small/Medium/Large
754100/764100/774100
APR
Air Purifying
Respirator (APR)
SURVIVAIR Opti-Fit
Facepiece
Side-mount Tube
CF-Supplied Air
Respirator (CF-SAR)
PAPR
CF-SAR
Front- mount
Tube and
Screw Caps
Kit*:
P/N 981205
Front- mount
Tube and P100
Filters
Kit*:
981209
Mask- mount
PAPR
Kit
520075
SURVIVAIR Opti-Fit
Facepiece Front-
mount Tube CF-SAR
5-Pt. Strap S/M/L
981500/981501/981503
Mesh Headnet™ S/M/L
981550/981551/981553
SURVIVAIR
Opti-Fit Facepiece
P100 Escape
CF-Supplied Air
Respirator (CF-SAR)
SURVIVAIR Opti-Fit
Facepiece
Mask-mount PAPR
5-Pt. Strap S/M/L
520500/520600/520700
Mesh Headnet™ S/M/L
520520/520620/520720
* All continuous ow supplied air respirator (CF-SAR) and powered air purifying respirator (PAPR) kit components are
available individually. See Parts List in Section VI.
Belt- mount
PAPR
Components
Blower Kit*:
500085
Breathing Tube:
B420030
5
SURVIVAIR Opti-Fit
Facepiece Kit Belt-mount
PAPR 5-Pt. Strap S/M/L 540510/540610/540710
Mesh Headnet™ S/M/L
541510/541610/541710
Blower/Battery Kit 500085
Page 9
SURVIVAIR Opti-Fit™ Convertible Facepiece
VI. PARTS LIST
A. Please refer the Parts List Matrix in Part B for component descriptions and identication of their use with the various operating congurations of the SURVIVAIR Opti-Fit Convertible Facepiece.
2b
APR
29
Headnet
Air Purifying Respirator
Cartridges
CF-SAR
SURVIVAIR Opti-Fit Convertible
Facepiece Models
5 Pt. Headstrap 7540/7640/7740 Mesh Headnet
7541/7641/7741
CF-Supplied Air
Respirator
Side-mount Breathing Tube
PAPR
CF-Supplied Air
Respirator
Front-mount Breathing Tube
CF-Supplied Air
Respirator
P100 Escape
Powered Air Purifying Respirator
Mask-mounted Blower
Powered Air Purifying Respirator
Belt-mounted Blower
6
Page 10
SURVIVAIR Opti-Fit™ Convertible Facepiece
B. Parts List Matrix
SURVIVAIR
7
Page 11
SURVIVAIR Opti-Fit™ Convertible Facepiece
Table 1. Air Purifying Respirator (APR) Filters and Cartridges
Part No. Protection/Description
Filters
1050 P100
1060 Filter Pad, N95
1070 Filter Pad, N99/R95
1190 P100 Slimline
1195 P100 Disk
Cartridges
100100 Organic Vapors
100200 Acid Gases
100300 Organic Vapors and Acid Gases
100400 Ammonia and Methylamine
100600 Mercury Vapor and Chlorine
100800 Multi-contaminant
1051 Combination Organic Vapors and P100
1052 Combination Acid Gases and P100
1053 Combination Organic Vapors, Acid Gases, and P100
1054 Combination Ammonia, Methylamine, and P100
1056 Combination Multi-contaminant and P100
1058 Combination Multi-contaminant and P100
Table 2. Continuous Flow Supplied Air Respirator (CF-SAR) Air Hose Components
Part No. Item Description
930861 High performance hose—25 feet (requires quick coupler kit)
930862 High performance hose—50 feet (requires quick coupler kit)
930864 High performance hose—100 feet (requires quick coupler kit)
930801 Light weight hose—25 feet (requires quick coupler kit
930802 Light weight hose—50 feet (requires quick coupler kit
930804 Light weight hose—100 feet (requires quick coupler kit
985241 Light weight hose with OBAC couplings—25 feet
985242 Light weight hose with OBAC couplings—50 feet
985244 Light weight hose with OBAC couplings—100 feet
930810 Quick coupler kit—style: Foster
930820 Quick coupler kit—style: Schrader
930830 Quick coupler kit—style: Hansen
930840 Quick coupler kit—style: OBAC
Table 3. Powered Air Purifying Respirator (PAPR) Belt-Mounted Cartridges
Part No. Item Description
150100 OV Cartridge
150200 AG Cartridge
150300 OV/AG Cartridge
150400 AM/MA Cartridge
158100 OV/HEPA Cartridge
158200 AG/HEPA Cartridge
158300 OV/AG/HEPA Cartridge
158400 AM/MA/HEPA Cartridge
108000 HEPA Filter
APR Cartridge Considerations:
WARNING
The following is a partial list of materials that have either poor warning proper­ties, will develop high heats of reaction with the sorbent in the canister, have extremely short service times, or will not be absorbed at all. Respirators can­not be used to protect you against many
of these chemicals. Contact a certied
industrial hygienist or SPERIAN if you have any of these materials present in the work place. The use of a respirator for these materials may lead to personal injury, illness, or death.
Acroline Carbonyls Hydrogen
Aniline Cyanogen Isocyanates:
Arsine Dimethylaniline Methanol Phosphine
Boron Hydrides
Bromine Ethyl Cyanide Nitro
Carbon Dioxide
Carbon Monoxide
Dimethyl Sulfate Methyl
Fluorine Sulfur Chloride
Hydrogen Cyanide
Selenide
TDI, HDI, MDI
Bromide Methyl Chloride, Methylene Chloride
Compounds:
Nitrogen Oxides Nitroglycerin Nitrobenzene Nitromethane
Ozone
Phosgene
Phosphorous Trichloride
Stibine
Vinyl Chloride
VII. FACEPIECE PREPARATION USING INCLUDED ITEMS
The SURVIVAIR Opti-Fit Convertible is shipped with side connector port screw caps and a nozzle plug kit
that must be used to congure the facepiece for the various operating congurations as detailed below.
8
Page 12
SURVIVAIR Opti-Fit™ Convertible Facepiece
A. Air Purifying Respirator (APR)
The SURVIVAIR Opti-Fit Convertible is shipped from the factory with the nozzle adapter (item 20, P/N 702045) installed in the facepiece nozzle. This adapter must be capped using the included adapter cap (item 24, P/N
420020) for the facepiece to be used as an APR respi­rator. When attaching the adapter cap, an audible click should be heard (Figure 1).
Figure 1. Adapter Cap
Figure 4. Insert Nozzle Gasket into Facepiece
If a low prole nozzle setup is desired, the nozzle
adapter can be removed from the facepiece and the nozzle plug kit installed in its place. Unscrew the nozzle adapter from the facepiece by pressing down on the adapter locking tab from inside the lens and turning the adapter counterclockwise (Figures 2 and 3). Insert the nozzle gasket (item 25, P/N 763016) into the nozzle port (Figure 4). Thread in the nozzle plug (item 26, P/N
962066), using a quarter or similar item in the molded slot to hand-tighten (Figure 5).
Figure 2. Push Down on Adapter Locking Tab
Figure 5. Tighten Nozzle Plug with Coin
in Molded Slot
WARNING
The included screw caps (item 27, P/N
985020) are not to be used on the face­piece side ports when using the SUR­VIVAIR Opti-Fit Convertible in the air puri-
fying respirator (APR) conguration. See
Section VIII, Part A.1 for installing chemi-
cal cartridges and particulate lters to
the facepiece side connector ports. Fail­ure to comply with this warning may lead to personal injury, illness, or death.
Figure 3. Remove Adapter
9
Page 13
SURVIVAIR Opti-Fit™ Convertible Facepiece
B. Continuous Flow Supplied Air Respirator(CF-SAR)
1. Side-mounted Breathing Tube
When using the SURVIVAIR Opti-Fit Convertible as a supplied air respirator (CF-SAR) with side­mounted breathing tube, the nozzle adapter and the side port not in use by the breathing tube must be capped for proper operation. Attach the nozzle adapter cap (item 24, P/N 420020) to the nozzle adapter, making sure to hear an audible click has the cap is pushed into place (Figure 1).
If a low prole nozzle setup is desired, the noz­zle adapter can be removed from the facepiece and the nozzle plug kit installed in its place. Un­screw the nozzle adapter from the facepiece by pressing down on the adapter locking tab from inside the lens and turning the adapter counter­clockwise (Figures 2 and 3). Insert the nozzle gasket (item 25, P/N 763016) into the nozzle port (Figure 4). Thread in the nozzle plug (item 26, P/N 962066), using a quarter or similar item in the molded slot to hand-tighten (Figure 5).
Thread one of the provided screw caps (item 27, P/N 985020) onto the side connector port not in use by the side-mounted breathing tube and hand-tighten (Figure 6).
C. Powered Air Purifying Respirator (PAPR)
Both the mask-mounted and belt-mounted powered air
purifying respirator (PAPR) congurations require both
side connector ports to be capped with the included screw caps (item 27, P/N 985020). Thread the screw caps onto the side connector ports and hand-tighten (Figure 6). Use of the nozzle adapter is necessary for connecting the mask-mounted blower or breathing tube and there-
fore the low-prole nozzle plug kit cannot be used.
WARNING
● Do not attempt to use the SURVIVAIR
Opti-Fit Convertible with both the nozzle adapter and the side connec­tor ports capped.
● Failure to securely cap unused
threaded side port connectors and/ or the front adapter port for the APR, CF-SAR or PAPR configuration you are using can allow contaminants to leak into the respirator and cause personal injury, illness, or death.
VIII. OPERATING INSTRUCTIONS FOR SPECIFIC RESPIRATOR CONFIGURATIONS
Figure 6. Thread Screw Cap onto Side Connection
Port (CF-SAR and PAPR Only)
2. Front-mounted Breathing Tube and P100
Escape
When using the SURVIVAIR Opti-Fit Convert­ible as a supplied air respirator (CF-SAR) with front-mounted breathing tube, both of the side connector ports must be capped with the includ­ed screw caps (item 27, P/N 985020). Thread the screw caps onto the side connector ports and hand-tighten (Figure 6). If the facepiece
is to be congured as an CF-SAR with P100
escape, the screw caps are not required, and
a P100 lter cartridge (item 38, P/N 105001)
should be screwed onto each side connector port. Use of the nozzle adapter is necessary for connecting the breathing tube and therefore
the low-prole nozzle plug kit cannot be used.
A. Air Purifying Respirator (APR) Conguration
In addition to the General Warnings listed in Section IV, the following warnings apply to air purifying respirator
(APR) congurations:
WARNING
● When using this respirator for protec­tion against any gas or vapor, includ­ing gases or vapors with poor warn­ing properties (e.g., HDI, MDI, or TDI isocyanates, paint hardeners, etc.), OSHA 29 CFR 1910.134 requires the canister to have an end-of-service­life indicator (ESLI). If the canister does not have an ESLI, OSHA re­quires the canister be changed at intervals derived from a change-out schedule that is based on objective information or data that ensures the canister is changed before the end of its service life.
● Do not use for protection against
gases or vapors which generate high heats of reaction with the sorbent material or against gases or vapors which are not readily absorbed by the sorbent (e.g., methanol).
● There are gases and vapors (e.g., ac­etone, acetaldehyde, etc.) that have a boiling point below 149ºF (65ºC). Gas masks should not be used for protec­tion against these chemicals
10
Page 14
SURVIVAIR Opti-Fit™ Convertible Facepiece
WARNING─Continued
for more than 8 hours or one shift, even though the calculated service life estimate is greater that 8 hours or one shift. These contaminants can mi­grate through the canister even when the canister is not in use (e.g., over­night) and collect in the facepiece.
● Do not use this respirator without the
appropriate SPERIAN cartridges and/
or lters securely attached onto the
facepiece. Always read cartridge la­bels prior to use and be certain that
you have cartridges and/or lters
that will provide the required protec­tion. This respirator must be used for protection against only those air contaminants listed on the air purify-
ing cartridge, lter and/or the NIOSH
approval label. Respirators labeled for protection against particulates only shall not be used for gases/va­pors. Respirators labeled for protec­tion against gases/vapors only shall not be used for particulates.
● This respirator does not supply ox­ygen. Do not use the respirator in atmospheres containing less than
19.5% oxygen by volume.
● Failure to comply with this warning
may lead to personal injury, illness, or death.
1. Installing Chemical Cartridges and Particulate Filters
a. Mercury Vapor Cartridges
Special or Critical User’s Instructions: The mer­cury vapor cartridges incorporate passive end­of-service-life indicators (ESLI). Each ESLI, yellow in color when new, turns gray when exposed to mercury vapor. Do no use these cartridges unless you can distinguish between the safe and discard colors of the ESLI. The indicators must be visible when wearing the respirator without having to manipulate either the facepiece or the indicator. If the indicators cannot be seen, do not wear the respirator. The cartridges must be replaced when the ESLI change color from yellow to gray, when the ESLI become dirty or damaged, after 30 days of use, or if removed from their original packag­ing and not used within 30 days. Never enter or remain in a hazardous atmosphere if one or more of the ESLI is gray.
b. Particulate Filters
Before using a particulate lter, you must con­rm the atmosphere is non-IDLH and not oxy­gen decient, the contaminant is a particulate
hazard, and you must determine the hazard ra­tio (hazard concentration/exposure limit). If the hazard ratio is less than 10, you may use a half
mask respirator with a lter efciency of 95% or
higher. If the hazard ratio is less than 20, you
may use a full facepiece respirator with a lter efciency of 95% or higher. If the hazard ratio is
less than 100, you may use a full facepiece res-
pirator with a lter efciency of 99% or higher. If
the hazard ratio is greater than 100, you must use another type of respirator.
N-series lters must only be used for non-oil
containing aerosols or particulates. R or P-
series lters can be used if the atmosphere contains oil (e.g., lubricants, cutting uids, glycerin, etc.). An R-series lter used in an oil-
containing atmosphere must be replaced after
each 8-hour shift. A P-series lter can be used
for more than one 8-hour shift. Please refer to Section VIII, Part A.6, Chemical and Particulate Filter Service Life and Replacement, for impor­tant additional information. Refer to the NIOSH publication Guide to the Selection and Use of
Particulate Respirators Certied Under 42CFR
84 for additional guidelines on use limitations
for lters.
WARNING
Filters improperly used or improperly assembled to the facepiece could cause leakage, which could lead to personal injury, illness, or death.
c. Cartridge/Filter Installation
Ensure that you have the correct cartridge and/
or lter type for the contaminants in your work
area. If you are not sure, consult your supervi­sor or safety professional.
Install cartridges and lters by threading i. them onto the facepiece connectors and hand tightening. Do not overtighten.
If using lter pads, insert the appropriate ii. pad into the lter retainer so that the side
stamped with the model number will face towards the facepiece when installed.
Completely snap the lter retainers onto the cartridges or lter holders.
Inspect the lter through the retainer to iii. ensure that the entire edge of the lter is
making contact with the retainer and the
lter is not bunched or folded. The outer
edge of the pad must be compressed be-
tween the top edge of the cartridge or lter
holder and the inner sealing rings of the
retainer. Install lter holders in the same
manner as cartridges.
11
Page 15
SURVIVAIR Opti-Fit™ Convertible Facepiece
2. Donning Instructions for APR Conguration
WARNING
● Your employer (or you, if you do not
have an employer) is responsible for establishing that this respirator is suitable for the user’s application.
● Always read cartridge labels prior to
use to be certain that you have car-
tridges and/or lters that will provide
the required protection. Filters la­beled for protection against particu­lates only shall not be used for gas­es/vapors, and cartridges labeled for protection against gases/vapors only shall not be used for particulates.
Both cartridges/lters must be of the
same type.
● Ensure that the o-ring (Item 23, P/N
824014), is in the cap (Item 24, P/N
420020) and the cap plug is securely
seated on the adapter. If the nozzle
plug (Item 26, P/N 962066) is to be used instead of the cap, be sure
that the Nozzle Gasket (Item 25, P/N
763016) is installed with an inhala­tion valve.
● The failure to comply with this warn­ing may lead to personal injury, ill­ness, or death.
7. Hold the facepiece against your face and pull the straps over your head.
8. Tighten the two lower straps, temple straps, then the top straps until a good seal is obtained and all straps lie at on your head. Do not overtight­en the straps. See Figure 8a. For respirators with a Headnet, tighten the two lower straps and the temple straps until a good seal is obtained
and all straps lie at on your head. Flatten the
Headnet with a wiping motion toward the back of your head. See Figure 8b. Retighten straps as necessary. Do not overtighten.
Figure 8a. Adjust Lower Straps First
1. Inspect the respirator as described in Section IX, Part B.
2. Verify that unused ports are capped as detailed in
Section VII for the APR operational conguration.
3. Verify that two lters and/or cartridges are at-
tached to the threaded side connectors of the facepiece.
4. Adjust the top straps until one inch of strap ex­tends through each buckle. Fully loosen the lower straps. For respirators featuring a Head­net, fully loosen all straps.
5. Place the neck strap (if installed) over your head.
6. Place your chin against the chin cup and center the facepiece on your face. See Figure 7.
Figure 8b. Flatten Headnet with Wiping Motion
NOTE
When properly adjusted, the head strap hub should be centered on the back of your head and the lower straps should be below your ears. See Figure 9a. For the Headnet ver­sion, the backstrap should be at the nape of
the neck. The Headnet should lie at on the
head and not bunch up. See Figure 9b.
Figure 7. Place Chin in Chin Cup
12
Page 16
SURVIVAIR Opti-Fit™ Convertible Facepiece
whether the cartridges are correctly seated against the facepiece. It is your responsibility to ensure that the face seal is not disturbed when reinstalling the cartridges and that each cartridge is sealed against the facepiece.
Inhale gently and hold your breath to slightly collapse the facepiece. If air leakage occurs, readjust the
facepiece and head strap, then retest. If lter retain­ers are used, do not remove them after the t check is completed. Removing the lter retainers may shift
the facepiece and break the face seal.
Figure 9a. Properly Center Head Strap Hub
Figure 9b. Properly Fitted Headnet
3. Fit Check Instructions for APR Conguration
You must perform the following t check each time
the respirator is worn or before entering the contami­nated area.
WARNING
If a leak-tight seal is not obtained during
the following t check, do not wear the
respirator. Ask your supervisor or safety professional for assistance. The failure to comply with this warning may lead to personal injury, illness, or death.
4. How to Use the Respirator
To receive the maximum protection available from your respirator, you must read, understand, and follow all the warnings, limitations, and instructions
contained in this manual and follow your employer’s
instructions on the use and maintenance of the res­pirator. Read and understand the following warnings prior to using the respirator.
WARNING
● Never remove the respirator for any
reason while in the work area.
● You must leave the work area imme­diately if the facepiece-to-face seal is disturbed for any reason such as:
a. Slippage due to sweating or ex-
cessive head movement.
b. The facepiece becomes dislodged
as a result of being knocked.
c. Sneezing or coughing while wear-
ing the facepiece.
d. You need to blow your nose,
scratch your face or adjust your spectacles.
● For any other reason that would cause
the facepiece seal to be disturbed.
● You must restore the facepiece-to-
face seal and perform a fit check in
a non-hazardous environment before
re-entering the work area.
● Failure to comply with this warning
may lead to personal injury, illness, or death.
Negative Pressure Fit Check
Cover the inlet of the cartridges and/or lters with the
palms of your hands or another suitable item, such as a thin sheet of plastic or rubber. When the respi-
rator is equipped with cartridges, lter retainers may
be used as an aid. Alternatively, the cartridges may
be removed to perform the t check. If removed,
SPERIAN recommends someone other than you
reinstall them after a leak-tight t is achieved. It is difcult for the wearer to reinstall the cartridges
without disturbing the face seal and to determine
5. Dofng Instructions for APR Conguration
1. Loosen the two lower head straps completely
(for ve strap head harness or Headnet).
2. Grasp the nozzle rmly, and pull the facepiece
upward over the head.
6. Chemical and Particulate Filter Service Life and Replacement
a. Chemical Cartridge Service Life and Replacement
13
Page 17
SURVIVAIR Opti-Fit™ Convertible Facepiece
WARNING
● OSHA allows the use of air purify­ing respirators for protection against contaminants with poor warning properties. You must replace car­tridges when the end-of-service-life indicator (ESLI) has changed color or in accordance with an OSHA com­pliant cartridge change-out schedule that is based on objective information or data that ensures the cartridges are changed before the end of their service life.
● You must immediately leave the con­taminated area if you taste or smell contaminants, or if your eyes or throat become irritated. Replace car­tridges and/or filters before re-enter­ing the contaminated area.
● Establishing the cartridge service
life for mixtures of contaminants is a complex task, and one that requires considerable professional judgment to create a reasonable change-out schedule. OSHA provides a “rule-of­thumb” method for determining the cartridge service life for mixtures. The method addresses two situa­tions. The first is where the individual compounds of the mixture have simi­lar breakthrough times (i.e., within one order of magnitude). In this case, OSHA recommends adding the con­centrations of all the components of the mixture to determine a “total con­centration,” then determine the mix­ture service life by applying the total concentration to the component with the shortest service life. The second is where the individual compounds have breakthrough times that vary by two orders of magnitude or great­er. In this case, OSHA recommends the mixture service life be based on the contaminant with the short­est breakthrough time. The OSHA rule-of-thumb methods may not be applicable for all mixtures. In some cases, the actual mixture service life may be much lower than the one cal­culated. For this reason, SPERIAN recommends that the cartridge ser­vice life for mixtures be determined using experimental methods. If you are unsure of how to determine the cartridge service life for mixtures, please contact SPERIAN.
● Failure to comply with this warning
may lead to personal injury, illness, or death.
The cartridge service life depends on the user’s
breathing rate, the characteristics of the contami­nant, and the environmental conditions such as temperature and humidity. Saturated cartridges will leak trace amounts of contaminant to the wearer, which may be detected by odor, taste, and/or ir­ritation. Replace cartridges when the end-of-ser­vice-life indicator (ESLI) has changed color, at the
rst trace of the characteristic contaminant odor
or taste, at regularly scheduled intervals recom­mended by your safety professional, if they show any signs of damage, or in accordance with an OSHA compliant cartridge change-out schedule. The service life of the cartridge can be estimated
using SPERIAN’s cartridge service life estimation
program, available from SPERIAN. Always replace cartridges in pairs.
Replace cartridges as follows:
1. Return to fresh air.
2. Remove cartridges by turning counter­clockwise. Dispose of used cartridges in accordance with Federal, state, and local guidelines.
3. Install new cartridges, taken only from sealed packages, as described in Sec­tion VIII, Part A.1.C.
4. Perform a facepiece t check as de­scribed in Section VIII, Part A.3.
b. Particulate Filter Service Life and Replacement
WARNING
You must replace the lters when breath­ing becomes uncomfortable or difcult. As particulates collect on the lter sur-
face, the breathing resistance of the facepiece increases. If you wait too long
to replace the lters, the particulates
may leak past the face seal instead of
being collected on the lters. Therefore,
SPERIAN strongly recommends that the
lters be changed at least daily.
Failure to comply with this warning may lead to personal injury, illness, or death.
The service life of the lter depends on the lter
series, the way in which the lter captures the con-
taminant (either mechanically or by electrostatic means), the characteristics of the contaminant,
the user’s breathing rate, and the environmental
conditions such as humidity, temperature, etc. SPERIAN’s N95, N99, R95 and Slimline P100 l­ters use electrostatic charges to capture particles. High humidity and temperatures, as well as con­tact with oil or non-oil based liquid aerosols, may
affect the capture of contaminants by these lters. All new SPERIAN lters must be maintained in fac-
tory sealed bags and must only be removed when
14
Page 18
SURVIVAIR Opti-Fit™ Convertible Facepiece
they are to be used. Used lters must be stored in
a protective bag, and in a clean, dry environment. Detailed guidelines for each lter series are pro­vided below.
1. N-Series Filters
These lters are only NIOSH approved for use
against solid and water-based particulates and must not be used in any area that con­tains oil aerosols. N-series lters capture par­ticles using electrostatic charges, and these
types of lters may be subject to some loss of efciency if they are exposed to certain liquid aerosols. The N-series (N95 and N99) lters must be replaced immediately whenever:
● they are damaged, soiled, soaked with
liquids such as water or alcohol, or ap­pear to be suspect or damaged in any manner, or
● the user notices an increase in breathing
resistance such that the respirator be­comes uncomfortable to wear.
If none of the above occurs and there is a respiratory protection program per OSHA
29 CFR 1910.134 in place, then these lters
may be used continually, but only for a maxi­mum of 30 days or 40 total hours, whichever
comes rst. However, SPERIAN recommends that all lters be replaced daily because the
user may be unable to detect small defects
resulting in a loss of lter efciency, and the
determination of when breathing resistance becomes too uncomfortable is subjective. Further, SPERIAN recommends that these
lters be replaced daily when there are any
liquid aerosols present and there is no respi­rator program in place.
2. R-Series Filters
These lters are NIOSH approved for use
against any type of particulate, except that
for atmospheres that contain oil. The lters
can only be used for 8 hours or a single shift,
whichever comes rst. When used in either
oil-containing or non-oil containing atmo-
spheres, these lters must be replaced im­mediately whenever:
● they are damaged, soiled, soaked with liq-
uids such as water or alcohol, or appear to be suspect or damaged in any manner, or
● the user notices an increase in breathing re­sistance such that the respirator becomes uncomfortable to wear.
If none of the above occurs, the lter is used
in an area that does not contain oil, and there is a respiratory protection program per OSHA
29 CFR 1910.134 in place, then these lters
may be used continually, but only for a maxi­mum of 30 days or 40 total hours, whichever comes rst. Even when used in non-oil con­taining atmospheres, SPERIAN recommends
that all lters be replaced daily because the
user may be unable to detect small defects
resulting in a loss of lter efciency, and the
determination of when breathing resistance becomes too uncomfortable is subjective.
3. P-Series Filters
These lters are NIOSH approved for long
duration use against any type of particulate. When used in either oil-containing or non-oil
containing atmospheres, these lters must be replaced immediately whenever:
they are damaged, soiled, soaked with liquids such as water or alcohol, or ap­pear to be suspect or damaged in any manner, or
● the user notices an increase in breathing
resistance such that the respirator be­comes uncomfortable to wear.
If none of the above occurs and there is a respiratory protection program per OSHA
29 CFR 1910.134 in place, then these lters
may be used continually, but only for a maxi­mum of 30 days or 40 total hours, whichever
comes rst. However, service time for the Model 1050 P100 lter can be extended if l-
ter performance to 42 CFR Part 84 can be demonstrated. Whether used in oil-containing or non-oil containing atmospheres, SPERIAN
recommends that all lters be replaced daily
because the user may be unable to detect small defects resulting in a loss of lter ef­ciency, and the determination of when breath­ing resistance becomes too uncomfortable is subjective.
WARNING
Failure to properly assemble lter pads to cartridges or lter holders can cause
leakage, which could lead to personal injury, illness or death.
1. Return to fresh air.
2. Remove cartridges or P100 lters by
turning them counterclockwise. Dispose of used cartridges and lters in accor­dance with Federal, state, and local guidelines.
3. If using lter pads, remove the lter re­tainers from the cartridges or lter hold­ers. Remove the used lter pads and
dispose of them in accordance with Fed­eral, state, and local guidelines. Clean
the lter retainer and cartridge or holder
if necessary.
4. Install new P100 lters or lter pads,
taken only from sealed packages, as described in Section VIII, Part A.1.C.
5. Perform a facepiece t check as de­scribed in Section VIII, Part A.3.
15
Page 19
SURVIVAIR Opti-Fit™ Convertible Facepiece
B. Supplied Air Respirator (CF-SAR) Congurations
In addition to the General Warnings listed in Section IV, the following warnings apply to supplied air respirator
(CF-SAR) congurations:
WARNING
● SPERIAN respirators, accessories,
and associated equipment should not be used in atmospheres that may contain contaminant concentra­tions above the lower explosive level (LEL). Intrinsic safety certification of electronic components does not eliminate potential danger from igni­tion in these atmospheres.
● Do not exceed maximum use con­centrations established by regula­tory standards.
● Air line respirators can be used only
when the respirators are supplied with respirable air meeting the re­quirements of CGA-7.1, Grade D or higher quality.
● Use only the pressure ranges and
hose lengths specified in these user instructions.
● Do not use this respirator in confined
spaces.
● For the P100 escape configuration:
Never use this respirator without two SPERIAN Model 1050 P100 filters se­curely threaded onto the facepiece.
● For the P100 escape configuration:
The P100 escape filter functionality of the P100 escape CF-CF-SAR con­figuration is not to be used in atmo­spheres containing less than 19.5% oxygen.
● FAILURE TO COMPLY WITH THIS
WARNING MAY LEAD TO PERSONAL INJURY, ILLNESS, OR DEATH.
1. Air Supply (applicable to side-mounted tube, front-
mounted tube and P100 escape congurations)
The following warnings regarding air supply are ap-
plicable to all continuous ow supplied air respirator (CF-SAR) congurations (side-mounted tube, front-
mounted tube, P100 escape), and should be fol­lowed for safe operation of a supplied air respirator.
WARNING
● Compressors, storage cylinders,
valves, regulators, fittings, and oth­er hardware must be large enough to deliver the air volume required by all users at peak demand.
● You are responsible for air quality and
compliance with safety and health codes applicable to your area.
● See the tables in Section VIII, Part B.8
for hose length, air supply pressure, and quick coupler requirements.
● Do not mix hose types or coupler as­semblies except as specified in the tables in Section VIII, Part B.8.
● Be sure that all respirable air system
piping, tubing, fittings, and couplings are incompatible with non-respirable gas systems.
● If an air pump or compressor is used
to supply breathing air, the following precautions must be taken:
The air pump or compressor in-1. take must not be located in an area that could possibly contain con­taminants, such as but not limited to engine exhaust, spray booth exhaust, or other toxic gases. The air pump or compressor must 2. have intake and exhaust filters capable of removing dirt, carbon dust, oil, oil vapor, and excess water mist that may be present in the area of the intake or may be generated within the air pump or compressor itself. Carbon monox­ide alarms should be installed on oil lubricated compressors and in any other situation where carbon monoxide may be present. Frequently test the purity of the air 3. delivered by the air pump or com­pressor. Air pumps and compres­sors deteriorate with age, as does the quality of the air they deliver. Set the pressure regulator and/or 4. relief valve to maintain the pres­sure within the range specified in the tables in Section VIII, Part B.8.
● FAILURE TO COMPLY WITH THIS
WARNING MAY LEAD TO PERSONAL INJURY, ILLNESS, OR DEATH.
Ensure that the air supply is type 1, Grade D or bet­ter, as described in the Compressed Gas Association
Commodity Specication for Air, G-7.1.
Set relief valves at a maximum pressure of 150 psig (1.03 Mpa).
16
Page 20
SURVIVAIR Opti-Fit™ Convertible Facepiece
2. Quick Coupler Assembly (applicable to side­mounted tube, front-mounted tube and P100 es-
cape congurations)
A male plug from one of the air line coupler kits listed in the Parts List, Section VI, must be assem­bled to the open end of the breathing tube. This operation requires a torque wrench.
WARNING
Do not use pipe thread sealing tape on the quick coupler or air line hose threads. The tape could restrict air-
ow or tape fragments could enter the
air system. Failure to comply with this warning may lead to personal injury, ill­ness, or death.
The steps and considerations required for com-
pleting the Quick Coupler assembly are as follow:
1. Breathing Tube
a. Apply Loctite No. 567 PST pipe sealant
or equivalent to the ¼ NPT threads of the male quick coupler plug. Ensure that no sealant enters the quick coupler or air line hose.
b. Thread the male plug into the female t-
ting on the breathing tube. Torque to 150 ±5 in-lb.
NOTE
The Vortex air cooler is not approved for use with the low pressure ambient air pump.
2. Air Supply Hose
Each length of hose is equipped with a ¼ inch female NPT swivel nut at each end and a male to male ¼ inch NPT adapter. The male to male adapter may be used to connect two lengths of hose or connect the hose to the air supply. Four approved methods of connecting hose lengths and air supply are illustrated in Figure 10.
WARNING
● The full facepiece is NIOSH approved
only when supplied with respirable breathing air through no more than three (3) lengths of 9308 Series hose, or two (2) lengths of 9852 Series air line hose at the pressures listed in Tables I and II, located in Section VIII, Part B.8. Do not mix hose types from different columns.
● The third pressure column in Table
I should be used when using a low pressure ambient air pump. No more than two (2) lengths of 9852 Series air line hose may be used. The air supply must be coupled to the air source and the breathing tube with the OBAC couplers and plugs.
● FAILURE TO COMPLY WITH THIS
WARNING MAY LEAD TO PERSONAL INJURY, ILLNESS, OR DEATH.
CAUTION
Do not exceed the maximum continuous duty or maximum intermittent duty operat­ing pressure of the low pressure ambient air pump. Refer to your pump user manual.
Figure 10. Methods of Hose Connection
3. When assembling the quick disconnect ttings
of the coupler or the male to male adapters to the hose, verify that the hose gaskets are in place and torque to 90 ±5 in-lb. Thread sealant is not required.
4. The air supply connection using the ¼ inch male to male adapter illustrated in Method 2 is also approved for use with Methods 3 and 4.
Method 1 is preferred when only one length
of hose is required since it allows the hose to be disconnected from the air supply.
Method 3 is preferred when multiple lengths
of hose (maximum of three) are required, eliminating the cost of quick disconnect cou­plers at the hose junctions.
5. Pressurize and check for leaks with a bubble
type leak detector. If leaks are detected:
Disassemble and repeat steps 3 through 5.
● If the leaks persist, remove the hose from ser-
vice and have repairs made by a SPERIAN
certied technician.
Disassemble and repeat steps 3 through 5.
If the leaks persist, remove the hose from ser-
vice and have repairs made by a SPERIAN
certied technician.
17
Page 21
SURVIVAIR Opti-Fit™ Convertible Facepiece
WARNING
Always verify that the swivel nut gaskets are in place and undamaged before as­sembly. Missing or damaged swivel nut gaskets may allow contaminants to leak into the system, causing illness or death.
3. Breathing Tube Setup (applicable to side-mount­ed tube, front-mounted tube and P100 escape
congurations)
1. Inspect the equipment as described in Section IX, Part B.
2. Verify that the O-ring is installed at the base of the nozzle inlet. See Figure 11.
5. Attach the breathing tube to the facepiece.
● For side-mounted tube CF-SAR conguration:
Thread one end of breathing tube onto one side connector of the facepiece. Verify that the other side connector is securely capped with a screw cap.
● For front-mounted tube and P100 escape CF­SAR congurations: An audible click should
be heard when attaching the breathing tube to the front adapter port on the facepiece nozzle. See Figure 13.
Figure 13. Front Adapter Port and Screw Caps
Figure 11. Inlet O-ring
3. Verify that unused ports are capped as detailed in Section VII for the applicable CF-SAR opera-
tional conguration.
4. Verify that the o-ring is installed into the end of the breathing tube as shown in Figure 12.
6. Connect the air line hose to the air supply.
7. Adjust the regulator at the air supply to the prop­er pressure. Refer to the tables in Section VIII, Part B.8 for determining the required pressure.
8. Don the waist belt, positioning the snap hook over the left hip. Tighten snugly.
9. Attach the waist belt snap hook to the breathing tube clip.
10. Plug the male quick coupler of the breathing tube into the air line quick coupler, and verify
that breathing air ows into the facepiece. From
the inside of the facepiece, you can verify air-
ow by placing your hand near the back side of
the nozzle inlet.
11. With the air owing, verify that the air supply
pressure is within the appropriate pressure range listed in the tables in Section VIII, Part B.8. Adjust if necessary.
12. Disconnect the breathing tube quick coupler from the air line and proceed with donning the facepiece.
4. Donning Instructions for CF-SAR Congurations
Figure 12. Breathing Tube O-ring
WARNING
Always don, remove, and t check the
respirator in a safe, uncontaminated area. Failure to comply with this warn­ing may lead to personal injury, illness, or death.
18
Page 22
SURVIVAIR Opti-Fit™ Convertible Facepiece
1. Inspect the respirator as described in Section IX, Part B.
NOTE
If the waist belt was not previously donned during the Breathing Tube Setup procedure of Section VIII, Part B.3, don the waist belt now. Position the snap hook over the left hip and attach it to the breathing tube.
2. Verify that unused ports are capped as detailed in Section VII for the applicable CF-SAR opera­tional conguration. (See Figure 14 for a rep­resentation of the side-mounted tube CF-SAR
conguration).
Figure 15a. Adjust Lower Straps
Figure 14. Place Chin in Chin Cup
3. Adjust the top straps until one inch of strap ex­tends through each buckle. Fully loosen the lower straps. For respirators featuring a Head­net, fully loosen all straps.
4. Place the neck strap (if installed) over your head.
5. Place your chin against the chin cup and center the facepiece on your face. See Figure 14.
6. Hold the facepiece against your face and pull the straps over your head.
7. Tighten the two lower straps, temple straps, then the top straps until a good seal is obtained and all straps lie at on your head. Do not over­tighten the straps. See Figure 15a. For respi­rators with a Headnet, tighten the two lower straps and the temple straps until a good seal
is obtained and all straps lie at on your head.
Flatten the Headnet with a wiping motion to­ward the back of your head. Retighten straps as necessary. Do not overtighten the straps. See Figure 15b.
Figure 15b. Flatten Headnet with a Wiping Motion
NOTE
When properly adjusted, the head strap hub should be centered on the back of your head and the lower straps should be below your ears. See Figure 16a. For the Headnet ver­sion, the backstrap should be at the nape of
the neck. The Headnet should lie at on the
head and not bunch up. See Figure 16b.
Figure 16a. Properly Center Head Strap Hub
19
Page 23
SURVIVAIR Opti-Fit™ Convertible Facepiece
Figure 16b. Properly Fitted Headnet
5. Fit Check Instructions for CF-SAR Congurations
You must perform the following t checks appro­priate for the CF-SAR conguration you are using
each time the respirator is worn or before entering the contaminated area.
WARNING
If a leak-tight seal is not obtained during
the following t check, do not wear the
respirator. Ask your supervisor or safety professional for assistance. The failure to comply with this warning may lead to personal injury, illness, or death.
A. Positive Pressure Fit Check for Side-mounted
Tube CF-SAR Conguration
Hold the palm of your hand gently over the exhala­tion valve outlet of the nozzle cover as shown in Figure 17 and cover the open end of the breathing tube with your other hand. Do not distort the natu­ral shape of the facepiece. Exhale gently and hold your breath to create and maintain a slight positive pressure inside the facepiece. If air leakage oc­curs, readjust the facepiece and head straps, then retest.
NOTE
The negative pressure t check detailed be­low can be performed as an alternate (or ad-
ditional) t check method for the side-mount­ed tube CF-SAR conguration.
B. Negative Pressure Fit Check for Side-mounted Tube, Front-mounted Tube, and P100 escape CF-
SAR Congurations
For all CF-SAR congurations: Plug the end of the breathing tube with a suitable plug.
For P100 escape CF-SAR congurations:
Cover the inlets of the lters with the palms of
your hands or another suitable item such as a thin sheet of plastic or rubber.
NOTE
Alternatively, the P100 lters may be re­moved to perform the t check. If removed,
SPERIAN recommends someone other
than you reinstall them after a leak-tight t is achieved. It is difcult for the user to reinstall the lters without disturbing the face seal and to determine whether the lters are correctly
installed. It is your responsibility to ensure that the face seal is not disturbed when re­installing the lters and that each lter is in­stalled correctly.
Inhale gently and hold your breath to slightly col­lapse the facepiece. The facepiece should remain collapsed on your face without leaking. If the face­piece leaks, reposition, check the straps, and repeat
the t check.
If a good t was achieved, remove the plug from the
breathing tube. If using the P100 escape CF-CF-
SAR conguration, reinstall the P100 lters.
Attach the breathing tube to the air line hose. The
continuous ow supplied air respirator (CF-SAR) is
now ready for use.
WARNING
Never use this respirator unless the quick connect is fully engaged in the locked position. Failure to comply with this warning may lead to personal inju­ry, illness, or death.
Figure 17. Positive Pressure Fit Check
If a good t was achieved, attach the breathing tube to the air line hose. The continuous ow supplied
air respirator (CF-SAR) is now ready for use.
6. How to Use the Respirator
Before entering the work area for the rst time, or
for re-entry, you must verify proper operation of the respirator in accordance with Section VIII, Parts B.1, B.2, and B.3, and follow the donning and t check­ing procedures described in Section VIII, Parts B.4 and B.5.
To receive the maximum protection available from your respirator, you must read, understand, and follow all the warnings, limitations, and instructions contained in this manual and follow your employ-
20
Page 24
SURVIVAIR Opti-Fit™ Convertible Facepiece
er’s instructions on the use and maintenance of
the respirator. Read and understand the following warnings prior to using the respirator.
If you are using an ambient air pump, turn off the pump when the respirator is not in use. Operating the pump without a connected supply hose and respira­tor will cause the pump to overheat. Turn the pump on before reentering the work area and verify that the pump outlet pressure is correct (See Tables I and II), and that the respirator is providing air as described in Section VIII, Part B.8.
WARNING
● The respirator consists of a full face­piece constructed of silicone with a polycarbonate lens, EPDM or EVA breathing tube, and PVC or rubber air line hoses. It is the user’s respon­sibility to verify the respirator materi­als are acceptable for their intended use. If you are unsure, consult your local safety professional to verify that no possible contaminants and/ or liquids will permeate through any of the respirator materials.
● Do not don, doff, or store the respi­rator in an area where contaminants can contact or accumulate inside any component of the respirator. Contaminants inside the facepiece, breathing tube, or air line hoses may be inhaled or absorbed upon reuse of the respirator.
● Never remove the respirator for any
reason while in the work area.
● You must leave the work area imme-
diately if the facepiece-to-face seal is disturbed for any reason such as:
Slippage due to sweating or ex-1. cessive head movement. The facepiece becomes dislodged 2. as a result of being knocked. Sneezing or coughing while wear-3. ing the facepiece. You need to blow your nose, 4. scratch your face, or adjust your spectacles. For any other reason that would 5. cause the facepiece seal to be dis­turbed.
● You must restore the facepiece-to-
face seal and perform a fit check in
a non-hazardous environment before
re-entering the work area.
● FAILURE TO COMPLY WITH THIS
WARNING MAY LEAD TO PERSONAL INJURY, ILLNESS, OR DEATH.
7. Dofng Instructions for CF-SAR Congurations
a. Loosen the two lower head straps completely
(for ve strap head harness or Headnet).
b. Grasp the nozzle rmly, and pull the facepiece
upward over the head.
S – Special or Critical User’s Instructions
8. Pressure Tables for Supplied Air Respirators (SAR)
Table I in Section VIII, Part B.8 provides the pres­sure range settings for the air supply regulator for the SURVIVAIR Opti-Fit Continuous Flow Sup­plied Air Respirator. Refer to Table II when using the vortex air cooler. Refer to Section VI, Parts List, for a listing of the approved air line hose types and coupler kits. No more that three (3) lengths of 9308 Series hose or two (2) lengths of 9852 Series hose at the listed pressures may be used. Do not mix hose types from different columns.
WARNING
● Compressors, storage cylinders,
valves, regulators, fittings, and oth­er hardware must be large enough to deliver the air volume required by all users at peak demand.
● You are responsible for providing
CGA G-7.1, Type 1, Grade D or bet­ter air and ensuring compliance with safety and health codes applicable to your area.
● FAILURE TO COMPLY WITH THIS
WARNING MAY LEAD TO PERSON­AL INJURY, ILLNESS, OR DEATH.
CAUTION
Do not exceed the maximum continuous duty or maximum intermittent duty operat­ing pressure of the low pressure ambient air pump. Refer to your pump user manual.
NOTE
The third column in Table I should be used when using a low pressure ambient air pump. No more than two (2) lengths of 9852 Series air line hose may be used. The air supply line must be coupled to the air source and the breathing tube with OBAC couplers and plugs.
The vortex air cooler is not approved for use with the low pressure ambient air pump.
21
Page 25
SURVIVAIR Opti-Fit™ Convertible Facepiece
TABLE I
SURVIVAIR Opti-Fit Full Facepiece
Hose
Length
(feet)
0* 1 to 4* 2 to 14*
25 5 to 12 9 to 12
50 7 to 17 13 to 23
75 10 to 22 17 to 30
100 15 to 25 20 to 25
150 19 to 35 25 to 35
200 19 to 35 25 to 25
250 20 to 45 30 to 45
300 20 to 50 30 to 50
9308 Series
Hose w/OBAC
Couplings
Pressure (psig)
9308 Series Hose
w/Foster/Hansen/
Schrader Couplings
9852 Series
Hose
(985241)2.5
to 8
(985242)3.6
to 12
(985244)6.3
to 21
*Indicates the pressure range required when no sup-
plied air line hose is being used, and the respirator breathing tube is connected directly to the air supply source. The breathing tube must be connected to the
air supply with a quick-disconnect tting.
TABLE II
SURVIVAIR Opti-Fit Full Facepiece
with Vortex Air Cooler (P/N 981847)
Hose
Length
(feet)
25 45 to 75 (985241) 40 to 75
50 45 to 80 (985242) 40 to 80
75 50 to 80
100 50 to 85 (985244) 40 to 85
150 55 to 85
200 55 to 90
250 60 to 100
300 60 to 100
9308 Series Hose
w/Foster/ Hansen/Schrader/
OBAC Couplings
9852 Series Hose
9. P100 Filter Service Life and Replacement
WARNING
You must replace the lters when breath­ing becomes uncomfortable or difcult. As particulates collect on the lter sur-
face, the breathing resistance of the facepiece increases. If you wait too long
to replace the lters, the particulates
may leak past the face seal instead of
being collected on the lters. Therefore,
SPERIAN strongly recommends that the
lters be changed at least daily. Failure
to comply with this warning may lead to personal injury, illness, or death.
Before using a particulate lter, you must conrm
that the atmosphere is non-IDLH and not oxygen de-
cient, the contaminant is a particulate hazard, and
you must determine the hazard ratio (hazard concen­tration/exposure limit). If the hazard ratio is less than 100, you may use a full facepiece respirator with a
lter efciency of 99% or higher. If the hazard ratio
is greater than 100, you must use another type of respirator.
The service life of the lter depends on the charac­teristics of the contaminant, the user’s breathing rate,
and the environmental conditions such as humidity,
temperature, etc. All new SPERIAN lters must be
maintained in factory sealed bags and must only be
removed when they are to be used. Used lters must
be stored in a protective bag, and in a clean, dry en­vironment.
SPERIAN Model 1050 P100 lters are NIOSH ap­proved for long duration use against any type of particulate. When used in either oil-containing (e.g.,
lubricants, cutting uids, glycerin, etc.) or non-oil­containing atmospheres, these lters must be re­placed immediately whenever:
they are damaged, soiled, soaked with liquids such as water or alcohol, or appear to be suspect or damaged in any manner, or
the user notices an increase in breathing resis­tance such that the respirator becomes uncom­fortable to wear.
If none of the above occurs, and there is a respira­tory protection program per OSHA 29 CFR 1910.134
in place, then these lters may be used continually,
but only for a maximum of 30 days or 40 total hours,
whichever comes rst. However, service time for the Model 1050 P100 lter can be extended if lter per-
formance to 42 CFR Part 84 can be demonstrated. Whether used in oil-containing or non-oil containing
atmospheres, SPERIAN recommends that all lters
be replaced daily because the user may be unable
to detect small defects resulting in a loss of lter ef­ciency, and the determination of when breathing re-
sistance becomes too uncomfortable is subjective.
a. Return to fresh air.
b. Unscrew the used P100 lters from the face-
piece connectors and discard them in accor­dance with applicable local, state and Federal guidelines.
c. Securely install new P100 lters onto the face-
piece threaded connectors as described in Section VIII, Part B.3.
d. Perform a facepiece t check as described in
Section VIII, Part B.5.
C. Powered Air Purifying Respirator (PAPR)
Congurations
In addition to the General Warnings listed in Section IV, the following warnings apply to powered air puri-
fying respirator (PAPR) congurations:
22
Page 26
SURVIVAIR Opti-Fit™ Convertible Facepiece
WARNING
● SPERIAN respirators, accessories,
and associated equipment should not be used in atmospheres that may contain contaminant concen­trations above the lower explosive level (LEL). Intrinsic safety certifica­tion of electronic components does not eliminate potential danger from ignition in these atmospheres.
● This respirator does not supply ox­ygen. Do not use the respirator in atmospheres containing less than
19.5% oxygen by volume.
● For the mask mounted blower con­figuration: NEVER use this respira­tor without a SPERIAN HE filter, P/N 108000, securely threaded into the blower.
● For the mask mounted blower con­figuration: The respirator does not protect against gases or vapors. It is only for protection against particu­late contaminants.
● FAILURE TO COMPLY WITH THIS
WARNING MAY LEAD TO PERSON­AL INJURY, ILLNESS, OR DEATH.
1. Battery Pack Charging (Mask-mounted and
Belt-mounted PAPR Congurations)
The battery pack must be charged before the rst
use and should be recharged after each use.
WARNING
● The SPERIAN PAPR chargers, both
individual and gang-type, are for in­door use only and must not be ex­posed to outdoor elements.
● Battery pack charging must be done in a safe, non-hazardous location.
Failure to do so may result in injury or death.
● Only SPERIAN P/N 520060 or 522000
Nickel-Cadmium battery packs can be charged with a SPERIAN P/N 520073 charger. Charging other types of batteries could cause explo­sions resulting in injury or death.
● FAILURE TO COMPLY WITH THIS
WARNING MAY LEAD TO PERSON­AL INJURY, ILLNESS, OR DEATH.
NOTE
The battery switch may be left in the ON or OFF position during charging.
The battery is equipped with an automatic fault protection circuit; no fuse is required.
1. Plug the battery charger into a 110 to 120 VAC power outlet.
2. Connect the charger outlet cable to the battery pack. The red light on the battery charger will illuminate when an electrical connection to the battery is made. If the light does not come on, the charger or battery pack may be damaged.
3. Allow the battery pack to charge for 14 to 16 hours. The charger light will remain on when­ever the battery is being charged.
NOTE
SPERIAN does not recommend charging the battery for more than 24 hours. Charging the battery pack for longer periods will not nor­mally damage the battery pack but may re­duce the life of the battery
4. If the charger (and light) cycles on and off, there is a problem with the battery pack and it should be replaced.
2. Cartridge and Filter Installation
a. Mask-mounted PAPR Conguration – Filter In-
stallation
WARNING
● Use only SPERIAN HE filters, P/N
108000. Use of any other filter will void the equipment’s NIOSH approv­al and may allow leakage that could result in serious illness or death.
● Never use the PAPR without the
blower inlet gasket (Figure 18) in place. Use of the PAPR without the blower inlet gasket could allow con­taminants to leak into the facepiece, resulting in illness or death.
● Use only components identified in
this manual and on the NIOSH ap­proval label. Substitute components will void approval and may result in illness or death.
CAUTION
SPERIAN recommends that the battery pack be charged at temperatures between 40°F and 80°F (4.5°C and 26.7°C). Charg­ing at higher temperatures may cause a re­duction in capacity and may reduce battery pack life.
1. Verify that the rubber gasket is in place at the bottom of the threaded blower inlet as shown in Figure 18.
23
Page 27
SURVIVAIR Opti-Fit™ Convertible Facepiece
2. If installing chemical cartridges, remove the screw cap and plug from each cartridge. Retain for reuse.
3. Install the three (3) cartridges or lters and
hand tighten.
4. Connect the power cord between the blower and battery pack.
3. Pre-operational Flow Check
a. Mask-mounted PAPR Conguration–Pre-opera-
tional Flow Check
Figure 18. Blower Inlet Gasket
2. Install a SPERIAN HE Filter, P/N 108000, and hand tighten. A SPERIAN shower
cap, P/N 981177, may be tted over the lter inlet.
b. Belt-mounted PAPR Conguration–
Cartridge and Filter Installation
WARNING
● Always read cartridge labels prior
to use to be certain that you have cartridges and/or filters that will provide the required protection. Fil­ters labeled for protection against particulates only shall not be used for gases/vapors, and cartridges la­beled for protection against gases and vapors only shall not be used for particulates.
● Only use three (3) cartridges or filters
of the same type.
● Never use the PAPR without all four
(4) blower inlet gaskets in place (Fig­ure 19). Use of the PAPR without the blower inlet gasket could allow con­taminants to leak into the facepiece.
● FAILURE TO COMPLY WITH THIS
WARNING MAY LEAD TO PERSON­AL INJURY, ILLNESS, OR DEATH.
WARNING
Flow check, facepiece t check, don­ning, and dofng must be done in a safe,
uncontaminated area. Failure to comply with this warning may lead to personal injury, illness, or death.
The airow must be veried before each use.
Use the following procedure to verify that the
blower is supplying adequate airow.
1. Detach the blower from the facepiece by pressing down the catch on top of the blower and pulling the blower away from the facepiece. See Figure 20.
Figure 20. Blower Catch
1. Verify that the rubber gaskets are in place at the bottom of the threaded blower inlets as shown in Figure 19.
Figure 19. Blower Inlet Gaskets
2. Connect the power cord to the battery pack and blower assembly and start the blower by pressing the battery ON/OFF switch.
3. With the blower outlet facing upward, in­sert the Flow Meter, P/N 580009, with the arrow pointing upward, being careful not
to cover the slots in the side of the ow
meter tube. See Figure 21.
24
Page 28
SURVIVAIR Opti-Fit™ Convertible Facepiece
b. Belt-mounted PAPR Conguration–Pre-opera-
tional Flow Check
Flow check, facepiece t check, don­ning, and dofng must be done in a safe,
uncontaminated area. Failure to comply with this warning may lead to personal injury, illness, or death.
WARNING
Figure 21. Flow Meter Attached
to Mask-mounted Blower
4. If the ball rises above the line on the ow me­ter indicating that the ow exceeds 4 cubic
feet per minute, proceed to the step 6.
WARNING
If the ow meter ball does not rise above
the line, DO NOT USE the PAPR until the cause can be determined and corrected. Use of a PAPR with a low ow may re­sult in illness or death.
5. If the ball is below the line indicating low
ow, the most probable causes are:
a. The lter is blocked by dust or water.
Replace with a new SPERIAN HE lter,
P/N 108000.
b. Low battery voltage. Recharge the bat-
tery for 14 to 16 hours.
Identify and correct any low ow problems
before proceeding to step 6.
6. Remove the ow meter and detach the
power cord from the battery pack. Reat­tach the blower to the facepiece, ensuring that the catch snaps into the slot on the facepiece nozzle.
7. Slip the belt through the loops on the bat­tery pack and adjust the battery pack so that the belt rides comfortably on the hips. Tighten the belt.
8. Attach the power cord to the battery pack.
WARNING
If changing the lter and/or recharging
the battery pack does not restore normal
airow, DO NOT USE the PAPR. Have the
PAPR repaired by the SPERIAN factory repair center. Using a PAPR that is operating improp­erly may result in illness or death.
The airow must be veried before each use.
Use the following procedure to verify that the
blower is supplying adequate airow.
1. Verify the tightness of all three (3) cartridges
or lters.
2. Insert the Flow Meter, P/N 580009, into the blower outlet with the arrow pointing upward as shown in Figure 22.
Figure 22. Flow Meter Attached
to Belt-Mounted Blower
3. Connect the power cord to the battery pack and blower assembly, and start the blower by pressing the battery ON/OFF switch.
NOTE
Do not cover the inlet area of any cartridges,
lters, or slots in the side of the ow meter tube when performing the airow check.
4. Verify the airow is greater than 170 liters
per minute (6 cfm) at the blower outlet
as shown in Figure 22. The ow meter is
designed so that a reading of 170 lpm or greater at the blower outlet will ensure the
airow is greater than 115 lpm at the full
facepiece. If the ball rises above the line
on the ow meter indicating that the ow
exceeds 170 lpm, proceed to step 6.
25
Page 29
SURVIVAIR Opti-Fit™ Convertible Facepiece
WARNING
If the ow meter ball does not rise above
the line, DO NOT USE the PAPR until the cause can be determined and corrected. Use of a PAPR with a low ow may re­sult in illness or death.
5. If the ball is below the line indicating low
ow, the most probable causes are:
a. The lter is blocked by dust or water.
Replace with a new SPERIAN HE lter,
P/N 108000.
b. Low battery voltage. Recharge the battery
for 14 to 16 hours.
Identify and correct any low ow problems
before proceeding to step 6.
6. Remove the ow meter from the blower
housing.
7. Slip the belt through the loops on the blower, then through the loops on the bat­tery pack. If you prefer to have the battery pack located on your left side, feed the
belt through the battery pack rst.
8. Attach power cord to the battery pack.
NOTE
Do not attach the breathing tube to the blow­er at this point. Instructions for attaching the breathing tube to the blower are listed in the Fit Check instructions found after the instruc­tions for donning the respirator.
WARNING
If changing the lter and/or recharging
the battery pack does not restore normal
airow, DO NOT USE the PAPR. Have the
PAPR repaired by the SPERIAN factory repair center. Using a PAPR that is op­erating improperly may result in illness or death.
4. Donning Instructions for PAPR Congurations
WARNING
● Always don, remove, and fit check
the respirator in a safe, uncontami­nated area.
● Prior to donning the SPERIAN mask-
mounted PAPR, ensure that the blower inlet gasket is in place (Fig­ure 18), and that the blower o-ring is in place in the blower outlet. Use of the PAPR without the blower in­let gasket and/or blower o-ring in place could allow contaminants to leak into the facepiece, resulting in illness or death.
● Prior to donning the SPERIAN belt-
mounted PAPR, ensure that the blower outlet gasket is in place, and that the connector O-ring is in place in the breathing tube (Figure 24). Use of the PAPR without the blower out­let gasket and /or connector o-ring in place could allow contaminants to leak into the facepiece, resulting in illness or death.
● When the PAPR is used in a contami­nated environment, both screw caps (Item 27, P/N 985020) must be se­curely installed on the gray threaded connectors as shown in figure 1. For the mask-mounted PAPR, the HE fil­ter, P/N 108000, must be installed on the blower assembly. Improperly installed screw caps or filter could allow contaminants to leak into the facepiece, causing illness or death.
1. Inspect the respirator as described in Section IX, Part B.
2. Verify that unused side connector ports are capped as detailed in Section VII for the PAPR
operational conguration. See Figure 23.
Figure 23. O-ring and Screw Caps
26
Page 30
SURVIVAIR Opti-Fit™ Convertible Facepiece
For belt-mounted setup: Also detach the breath- ing tube from the facepiece, if attached, and verify that an o-ring is installed at the base of the adapter inlet. See Figure 24.
Figure 24. Breathing Tube O-ring
3. Verify that the power cord is connected between
the battery pack and blower.
4. If the waist belt was not donned while perform-
ing the pre-operational ow check, don the
waist belt now and tighten snugly. Secure any excess belt length.
5. Slide the battery pack and blower into position
around your waist.
For mask-mounted setup: Position the battery
pack to your right or left side.
For belt-mounted setup: Position the blower
around to your back and the battery pack to your right or left side.
6. Adjust the top straps until one inch of strap ex-
tends through each buckle. Fully loosen the lower straps. For respirators with a Headnet, loosen all straps.
7. Place the neck strap (if installed) over your head.
8. Place your chin against the chin cup and center
the facepiece on your face. See Figure 25.
9. Tighten the two lower straps, temple straps, then the top straps until a good seal is obtained
and all straps lie at on your head. See Figure
26a. For respirators with a Headnet, tighten the two lower straps, then the temple straps, until
a good seal is obtained and all straps lie at on
your head. Flatten the Headnet with a wiping motion toward the back of your head. Retight­en straps as necessary. Do not overtighten the straps. See Figure 26b.
Figure 26a. Adjust Lower Straps
Figure 26b. Flatten Headnet with a Wiping Motion
NOTE
When properly adjusted, the head strap hub should be centered on the back of your head and the lower straps should be below your ears. See Figure 27a. For the Headnet ver­sion, the backstrap should be at the nape of
the neck. The Headnet should lie at on the
head and not bunch up. See Figure 27b.
Figure 25. Place Chin in Chin Cup
Hold the facepiece against your face and pull the straps over your head.
Figure 27a. Properly Center the Head Strap Hub
27
Page 31
SURVIVAIR Opti-Fit™ Convertible Facepiece
Figure 27b. Properly Fitted Headnet
5. Fit Check Instructions for PAPR Congurations
You must perform the following t check each time
the respirator is worn or before entering the con­taminated area.
WARNING
If a leak-tight seal is not obtained during
the following t check, do not wear the
respirator. Ask your supervisor or safety professional for assistance. The failure to comply with this warning may lead to personal injury, illness, or death.
1. Turn off the blower
For mask-mounted setup: Remove the lter from
the blower.
For belt-mounted setup: Remove the breathing tube from the facepiece nozzle.
2. Cover the blower inlet with your hand and inhale gently, slightly collapsing the facepiece.
3. Hold your breath for approximately 5 seconds. The facepiece should remain slightly collapsed. If it does not, readjust the facepiece and perform
the t check again. If a seal is still not achieved,
do not wear the respirator. Ask your supervisor or safety professional for assistance.
4. If a good t was achieved:
For mask-mounted setup: Reattach the lter
onto the blower assembly and press the ON/OFF switch on the battery pack to start the blower.
For belt-mounted setup:
Attach the breathing tube to the facepiece noz-
zle: Push the facepiece connector end of the
breathing tube onto the nozzle of the facepiece. Align the catch on the connector with the slot in the nozzle. Push the connector until the latch completely engages with the nozzle.
Screw the breathing tube into the blower outlet
and tighten securely.
Press the ON/OFF switch on the battery pack to
start the blower.
NOTE
In the event of a decrease or loss of airow
from the blower for both the mask-mounted
and belt-mounted PAPR congurations, the
wearer can safely breathe through the blow­er/lter assembly until reaching an uncon­taminated area.
WARNING
Verify that the blower is providing air to the facepiece before entering a hazard­ous environment. You should feel air blowing onto your face. Improper air­ow may result in inhalation of contam­inants. The failure to comply with this warning may lead to personal injury, ill­ness, or death.
6. How to Use the Respirator
Before entering the work area for the rst time or re-
entering for any reason, you must verify proper op­eration of the respirator in accordance with the Pre­operational Flow Check procedure of Section VIII, Part C.3, and also follow the donning procedures described in Section VIII, Part C.4.
To receive the maximum protection available from your respirator, you must read, understand, and follow all the warnings, limitations, and instructions
contained in this manual and follow your employer’s
instructions on the use and maintenance of the res­pirator. Read and understand the following warnings prior to using the respirator.
WARNING
● The respirator consists of a full face­piece constructed of silicone with a polycarbonate lens, blower, filter, battery, power cord, and a belt. It is the user’s responsibility to verify that the respirator materials are ac­ceptable for their intended use. If you are unsure, consult you local safety professional to verify that no possible contaminants and/or liq­uids will permeate through any of the respirator materials.
● Do not don, doff, or store the respi­rator in an area where contaminants can contact or accumulate inside any component of the respirator. Contaminants inside the facepiece may be inhaled or absorbed upon reuse of the respirator.
● Never remove the respirator for any
reason while in the work area.
28
Page 32
SURVIVAIR Opti-Fit™ Convertible Facepiece
WARNING─Continued
● You must leave the work area imme-
diately if the facepiece-to-face seal is disturbed for any reason, such as:
a. Slippage due to sweating or ex-
cessive head movement.
b. The facepiece becomes dislodged
as a result of being knocked.
c. Sneezing or coughing while wear-
ing the facepiece.
d. You need to blow your nose,
scratch your face, or adjust your spectacles.
e. For any other reason that would
cause the facepiece seal to be dis­turbed.
● You must restore the facepiece-to­face seal and perform a t check in a non-hazardous environment before
re-entering the work area.
● FAILURE TO COMPLY WITH THIS
WARNING MAY LEAD TO PERSONAL INJURY, ILLNESS, OR DEATH.
7. Dofng Instructions for PAPR Congurations
1. Loosen the two lower head straps completely
(for ve strap head harness or Headnet).
2. Grasp the nozzle rmly, and pull the facepiece
upward over the head.
8. Chemical and Particulate Filter Service Life and Replacement
1. Chemical Cartridge Service Life– Belt-mounted PAPR Conguration
WARNING
● OSHA allows the use of air purifying
respirators for protection against contaminants with poor warning properties. You must replace car­tridges when the end-of-service-life indicator (ESLI) has changed color or in accordance with an OSHA compliant cartridge change-out schedule that is based on objective information or data that ensures the cartridges are changed before the end of their service life.
● You must immediately leave the
contaminated area if you taste or smell contaminants, or if your eyes or throat become irritated. Replace cartridges and/or filters before re­entering the contaminated area.
WARNING─Continued
● Establishing the cartridge service
life for mixtures of contaminants is a complex task, and one that requires considerable professional judgment to create a reasonable change-out schedule. OSHA provides a “rule­of-thumb” method for determining the cartridge service life for mix­tures. The method addresses two situations. The first is where the in­dividual compounds of the mixture have similar breakthrough times (i.e., within one order of magnitude). In this case, OSHA recommends adding the concentrations of all the components of the mixture to deter­mine a “total concentration,” then determine the mixture service life by applying the total concentration to the component with the shortest service life. The second is where the individual compounds have breakthrough times that vary by two orders of magnitude or greater. In this case, OSHA recommends the mixture service life be base on the contaminant with the shortest breakthrough time. The OSHA rule­of-thumb methods may not be ap­plicable for all mixtures. In some cases, the actual mixture service life may be much lower than the one cal­culated. For this reason, SPERIAN recommends that the cartridge ser­vice life for mixtures be determined using experimental methods. If you are unsure of how to determine the cartridge service life for mixtures, please contact SPERIAN.
● FAILURE TO COMPLY WITH THIS
WARNING MAY LEAD TO PERSONAL INJURY, ILLNESS, OR DEATH.
The cartridge service life depends on the user’s
breathing rate, the characteristics of the contami­nant, and the environmental conditions such as temperature and humidity. Saturated cartridges will leak trace amounts of contaminant to the wearer, which may be detected by odor, taste, and/or ir­ritation. Replace cartridges when the end-of-ser­vice-life indicator (ESLI) has changed color, at the
rst trace of the characteristic contaminant odor
or taste, at regularly scheduled intervals recom­mended by your safety professional, if they show any signs of damage, or in accordance with an OSHA compliant cartridge change-out schedule. The service life of the cartridge can be estimated
using SPERIAN’s cartridge service life estimation
program, available from SPERIAN. Always replace all cartridges at the same time.
29
Page 33
SURVIVAIR Opti-Fit™ Convertible Facepiece
Replace cartridges as follows:
1. Return to fresh air.
2. Remove cartridges from the blower by turning counterclockwise. Dispose of used cartridges in accordance with Federal, state, and local guidelines.
3. Install new cartridges, taken only from sealed packages, as described in Section VIII, Part C.2.B.
4. Perform a pre-operational ow check as
described in Section VIII, Part C.3.B
2. Particulate Filter Service Life– Mask-mounted and Belt-mounted PAPR Congurations
WARNING
You must replace the combination car­tridges or HE lters in a non-contam­inated area when they become con­taminated such that they do not permit
adequate ow during the ow check
or while in use. As particulates collect
on the lter surfaces, the resistance of the lter increases, causing the ow
of clean air into the facepiece to de­crease. If you wait too long to change
the combination cartridges or lters,
particulates may leak past the face seal
instead of being collected on the lters.
Therefore, SPERIAN strongly recom­mends that combination cartridges or
lters be changed at least daily. Failure
to comply with this warning may lead to personal injury, illness or death.
The combination cartridge and HE lter are NIOSH
approved for long duration use against any type of particulate. When used in either oil-containing
(e.g., lubricants, cutting uids, glycerin, etc.) or
non-oil-containing atmospheres, the combination
cartridges or lters must be replaced immediately whenever:
they are damaged, soiled, soaked with liquids
such as water or alcohol, or appear to be sus­pect or damaged in any manner, or
the user notices an increase in breathing resis-
tance such that the respirator becomes uncom­fortable to wear.
they become contaminated such that they do not
permit adequate ow during the ow check or
while in use.
Replace the combination cartridges or HE lter
as follows:
1. Return to fresh air.
2. Remove the combination cartridges or HE lter from the blower by turning counter­clockwise. Dispose of used cartridges in accordance with Federal, state, and local guidelines.
3. Install new cartridges, taken only from sealed packages, as described in Section VIII, Part C.2.A or Part C.2.B.
4. Perform a pre-operational ow check as
described in Section VIII, Part C.3.A or Part C.3.B.
9. Decontamination – Mask-mounted and Belt-
mounted PAPR Congurations
CAUTION
Do not allow water to enter the blower. If water is allowed to remain in the blower for more than a few minutes, perma­nent damage will occur. Should water accidentally enter the blower, start the blower and shake the water out. Before storing the PAPR, allow the blower to run for a least ten (10) minutes. Verify that the PAPR is dry.
To prevent water from entering the blower during decontamination, leave the blow­er attached to the facepiece, and the l­ter securely attached to the blower.
Cartridges and/or lters that have been
soaked cannot be used.
Ensure that water does not enter the blower from inside the facepiece during removal or decontamination.
The battery cable must remain attached to the blower and battery pack during de­contamination to prevent water from en­tering the battery pack case or blower.
NOTE
For mask-mounted PAPR setup: Leave the blower attached to the mask and the lter
securely attached to the blower. Inhaled air will be drawn through the lter during decon­tamination.
1. Adjust the shower caps (Item 44, P/N 981177)
on the lters so that the opening is pointing
away from the water spray.
2. Proceed through the shower, following approved decontamination methods.
IX. MAINTENANCE
NOTE
Inspect this respirator for defects before and after each use, and at least once monthly if not used. Repair as necessary, clean and disinfect after each use, and store properly to assure that the respirator is maintained in satisfactory working condition. Keep a record of inspection and repair dates and results. Refer to the Inspection Tables in Section XIV of this manual.
30
Page 34
SURVIVAIR Opti-Fit™ Convertible Facepiece
A. Cleaning (applicable for all APR, SAR,and PAPR
congurations)
Clean respirators after each day’s use by washing with
mild soap and warm water. Then disinfect the respira­tors with a suitable sanitizing solution. OSHA 1910.134, latest edition, and ANSI Z88.2, latest edition, also pro­vide information and guidelines on the cleaning and sanitizing of respirators.
WARNING
● It is the user’s responsibility to en­sure that the cleaning process cho­sen provides adequate decontami­nation and disinfection.
● Specialized processes are required
to disinfect and decontaminate a respirator. You must follow the in­structions of the manufacturer of the equipment and chemicals.
● In the absence of a commercial sani­tizing product, the hypochlorite so-
lution described below will eliminate
many, but not all biohazards.
● FAILURE TO COMPLY WITH THIS
WARNING MAY LEAD TO PERSONAL INJURY, ILLNESS, OR DEATH.
1. To Clean the Facepiece:
a. Make a cleaning solution of warm water and
mild detergent.
b. Disconnect the blower assembly from the face-
piece by depressing the hook latch and pulling the blower free of the facepiece. Take care not to lose the o-ring inside the blower connector.
c. Immerse the facepiece top rst in the solution
until the exhalation valve is covered.
d. Agitate the facepiece and gently clean with a
soft brush.
e. Thoroughly rinse the facepiece in fresh water,
paying particular attention to removal of all soap residue from the exhalation valve. If possible, direct running water onto the exhalation valve.
f. Disinfect the facepiece in a warm, (120°F or
48°C maximum), suitable sanitizing solution, such as “hypochlorite solution” for 2 to 3 min­utes. A 50 ppm hypochlorite solution can be prepared by mixing 2 tablespoons (0.13 ounce) of chlorine bleach per each gallon of water. Rinse thoroughly with fresh warm (120°F or 48°C maximum) water.
g. Allow the facepiece to drip dry, or dry with a lint-free
cloth. Warm air may be used to speed up drying.
CAUTION
For PAPR Configurations: DO NOT
place the battery pack, blower assem­bly, or battery cord in water or solvents. Damage to the equipment will result.
Do not use ultrasonic cleaners to clean respirators equipped with speaking dia­phragms.
Do not use gasoline, organic based sol-
vents, or chlorinated degreasing fluids to clean any part of this respirator.
Do not clean the facepiece with the
blower attached.
You must ensure that this respirator is not
damaged when using disinfecting or de­contamination equipment or chemicals.
The facepiece lens can be scratched
through careless or abusive handling. Do NOT use abrasive cleaners or pads. Do not towel dry.
NOTE
Silicone and rubber parts of the facepiece may be cleaned between washings with SPERIAN mask wipes, P/Ns 140096 and
140082.
NOTE
Washing the lens will remove any user ap­plied anti-fog coating. Recoat with SPERIAN Anti-fog solution, P/N 951015 (1 oz.), or 951016 (16 oz.), or use SPERIAN single use anti-fog wipes, P/N 981803.
h. After cleaning, apply three drops of anti-fog so-
lution as required to the inner surface of the lens
and spread with your nger. Buff with a lint-free
cloth. If using an ant-fog wipe, follow the instruc­tions on the wipe packet.
I. Hold the facepiece rmly against your face and
exhale several times to ensure that the exhala­tion valve functions smoothly.
j. Carefully inspect the respirator as described in
Section IX, Part B.
k. Reconnect breathing tube or cartridges to the
facepiece before use.
For APR Congurations: Reconnect cartridges
and/or lters to threaded side connectors.
For CF-SAR and PAPR Congurations: Re­connect the breathing tube to either threaded side connector or front adapter on the face­piece. For breathing tubes connected to the front adapter on the facepiece nozzle, make sure the
latch locks rmly when reconnecting the tube.
2. To Clean Breathing Tubes and Blowers:
a. Wash the breathing tube, avoiding wetting the
inside of the tube as much as possible.
31
Page 35
SURVIVAIR Opti-Fit™ Convertible Facepiece
b. Clean the blower assembly, battery, and battery
cord with a brush or damp cloth, then dry with a clean, lint-free cloth. Do not allow water to enter the blower.
CAUTION
To prevent mildew, ensure that the in­side of the breathing tube is completely dried after washing.
If the possibility exists that water may have entered the blower, run the blower for a least 10 minutes afterward to dry the inside. The filter should remain on the unit whenever the blower is operat­ing in an atmosphere with a high water content.
B. Inspection (applicable for all APR, SAR, and
PAPR congurations)
Inspect the respirator before and after each use as de­scribed in the Inspection Tables of Section XIV and re­place any damaged component. See Section IX, Part C for repair instructions.
WARNING
● Do not use the respirator with dam-
aged or improperly operating valves.
● Ensure that the twin seal is in the
o-ring groove between the lens and connector. Ensure that the part num­ber tab on the o-ring is visible and in the notch provided on the lens.
● For APR and CF-SAR side-mounted
hose configurations: Ensure that the o-ring (See Figure X in Section VIII, Part A.2) is in the adapter cap (item 24, P/N 420020) and the cap is se­curely seated on the adapter. Using the facepiece in applicable configu­rations without the adapter cap se­curely seated could allow contami­nants to leak into the facepiece.
● For PAPR configurations: Do not at-
tempt to disassemble the blower or battery pack. Disassembly by other than SPERIAN certified technicians might render the equipment unsafe or void NIOSH approvals and your warranty.
● Failure to comply with this warning
may lead to personal injury, illness or death.
C. Repair (applicable for all APR, SAR, and PAPR
congurations)
Repair by the user is limited to replacement of com­ponents listed in the Repair Tables of Section XV and as listed on the NIOSH approval lavel. Disassembly should be performed only to the extent necessary to re-
place the components. To protect your warranty and the
NIOSH certication of the equipment, all other repairs
must be done only by authorized SPERIAN technicians and repair centers. Consult your SPERIAN distributor
for the one nearest you. You must perform a t check
as described in this user manual after any repair and before respirator use.
WARNING
For CF-SAR and PAPR Configurations: Do not disasseble this respirator until the air supply has been disconnected or shut off. Failure to comply with this warning may lead to personal injury, illness, or death.
D. Storage (applicable for all APR, SAR, and PAPR
congurations)
After inspection and cleaning, store your SURVIVAIR Opti-Fit facepiece to protect it against dust, sunlight, ex­treme heat and cold, excessive moisture, or damaging chemicals.
WARNING
The respirator does not have a de­fined storage life. Carefully inspect the respirator before each use. Failure to comply with this warning may lead to personal injury, illness, or death.
CAUTION
The maximum storage temperature for
the respirator is 140°F (60°C). Long­term storage at elevated temperatures may cause premature deterioration.
For PAPR Congurations:
Before storing the PAPR, allow the blower to a. run at least 10 minutes. Check to ensure that the PAPR is dry. After inspection and cleaning, the SPERIAN Mask Mounted PAPR should be stored to protect it against dust, sunlight, extreme heat and cold, excessive moisture, or damaging chemicals.
Follow the charging instructions after removing b. the PAPR batteries from long term storage.
X. SHIPMENT
All products returned to SPERIAN’s factory, distribu­tors, or repair centers must be decontaminated prior to shipment.
PRODUCTS CONTAMINATED WITH DANGEROUS SUBSTANCES WILL BE REFUSED AND RETURNED FREIGHT COLLECT.
32
Page 36
SURVIVAIR Opti-Fit™ Convertible Facepiece
XI. OVERHAUL FREQUENCY
The SPERIAN air purifying, supplied air and powered air purifying respirator (APR, SAR, PAPR) congura­tions do not have an overhaul requirement other than that required in Section IX, Maintenance.
XII. ADDITIONAL INFORMATION
If you need assistance or additional information on any SPERIAN product, consult your local distributor
or contact:
SPERIAN
3001 S Susan Street
Santa Ana, CA 92704
Tel: (714) 545-0410 or (888) APR-SCBA
FAX: (714) 850-0299
XIII. WARRANTY AND LIMITATION OF LIABILITY
LIMITED WARRANTY: SPERIAN warrants this product
to be free from defects in materials and workmanship for one (1) year from the date of purchase. During this period, SPERIAN will repair or replace defective parts,
at SPERIAN’s option. Freight charges to and from the
SPERIAN factory shall be paid by the purchaser.
EXCLUSIONS: NOT WITH STANDING ANY CON­TRARY TERM IN THE PURCHASER’S PURCHASE ORDER OR OTHERWISE, THE ONLY WARRANTY EXTENDED BY SPERIAN IS THE EXPRESSED LIM­ITED WARRANTY DEFINED ABOVE. THIS WARRAN­TY IS EXCLUSIVE AND IN LIEU OF ANY IMPLIED WARRANTY OF MERCHANTABILITY, OR FITNESS FOR A PARTICULAR PURPOSE.
CONDITIONS: To maintain this warranty this product
must be used, maintained, and inspected as prescribed in the owner’s instruction manual, including prompt re­placement or repair of defective parts and such other necessary maintenance and repair as may be required. Normal wear and tear; parts damaged by abuse, mis­use, negligence, or accidents; and batteries or damage
from battery leakage are specically excluded from this
warranty.
LIMITATION OF LIABILITY: No other oral warranties, representations, or guarantees of any kind have been made by SPERIAN, its distributors, or the agents of either of them, that in any way alter the terms of this warranty.
EXCEPT AS HERE IN PROVIDED, SPERIAN SHALL HAVE NO LIABILITY FOR ANY LOSS OR DAMAGE, WHETHER DIRECT, INDIRECT, INCIDENTAL, OR CONSEQUENTIAL, TO ANY PURCHASER OR USER OF THIS PRODUCT ARISING FROM THE SALE, USE, OR OPERATION OF THIS PRODUCT.
WARNING
The failure to use and maintain this equipment in strict conformance with the applicable instruction manual may result in personal injury, illness, or death. The equipment’s use in any man-
ner that is not expressly authorized
pursuant to the applicable instruction manual may result in severe adverse impacts to human health.
33
Page 37
SURVIVAIR Opti-Fit™ Convertible Facepiece
XIV. RESPIRATOR INSPECTION TABLES
General Inspection Table
The following table lists general inspection steps to be followed for all respirator congurations (APR, SAR, PAPR).
COMPONENT LOOK FOR
FACEPIECE LENS 1. Nicks, scratches, or abrasions that could impair visibility.
2. Deep gouges or cracks which could reduce impact resistance.
3. Anti-fog coating in need of replacement.
FACEPIECE RIMS 1. Deformed, cracked, or broken.
2. Loose screws. Do not overtighten.
FACEPIECE SKIRT 1. Cuts, gouges, or punctures.
2. Tears or nicks in the sealing area.
3. Deterioration from age, heat, or contamination.
FACEPIECE HEADSTRAP AND BUCKLE STRAPS
FACEPIECE BUCKLE Crushed, bent, broken, or corroded.
FACEPIECE INLET NOZZLE AND NOZZLE ADAPTER
O-RINGS/TWIN SEALS Cuts, nicks, abrasions, or excess stretching.
CARTRIDGE CONNECTORS 1. Cracks, heat, or impact damage.
HEADNET 1. Signs of deterioration, weakening, and wear in the the Headnet and straps.
1. Abrasions or nicks.
2. Deterioration from age, heat, or contamination.
1. Heat or impact damage.
2. Nicks, cracks, tears, or creases in the exhalation valve. Remove the valve cover to inspect the exhalation valve.
3. Nicks, cracks or dents in the exhalation valve seat.
4. Sticking exhalation valve. Exhale a few times to test. The valve must close after each exhalation. Valves that fail to close must be replaced.
5. Cuts, nicks, abrasions, or excess stretching to the Nozzle Adapter O-rings.
2. Cuts, nicks, or abrasions on connector sealing ange.
2. Unraveled or loose stitching.
Additional Inspection Table for Supplied Air Respirators (SAR)
For supplied air respirators (SAR), the following inspection steps should be followed in addition to the general inspection steps listed in the General Inspection Table above.
COMPONENT LOOK FOR
BREATHING TUBE (ALL CF-SAR CONFIGURATIONS)
Air Line Hoses (all CF-SAR congurations) 1. Fittings corroded, cracked, or leaking.
Belt (all CF-SAR congurations) 1. Webbing color change, excessive wear, fraying, cuts, nicks, or broken
1. Cuts, nicks, cracks, punctures, or tears that may reduce the level of protection.
2. Age or heat induced cracking, checking, or hardening.
3. Crushed, broken, or cracked connectors.
4. Tube is securely attached to the connectors and the hose clamps are tight.
2. Cuts, nicks, cracks, punctures, age or heat induced cracking, checking, abrasions, blistering, holes, crushing, kinking, or hardening.
3. Tightness of connectors and quick couplers.
4. Swivel nut gaskets in place and undamaged.
stitching.
2. Buckle damaged or corroded.
3. Loose hardware.
34
Page 38
SURVIVAIR Opti-Fit™ Convertible Facepiece
XIV. RESPIRATOR INSPECTION TABLESContinued
Additional Inspection Table for Powered Air Purifying Respirators (PAPR)
For powered air purifying respirators (PAPR), the following inspection steps should be followed in addition to the general inspection steps listed in the General Inspection Table above.
COMPONENT LOOK FOR
Blower (mask-mounted and belt-mounted PAPR)
Battery and Cord (mask-mounted and belt-mounted PAPR)
Belt (mask-mounted and belt-mounted PAPR)
Breathing Tube (belt-mounted PAPR)
1. Cracked or nicked gasket
2. Missing gaskets.
3. Loose cover screws.
4. Test for leaks and cracks in the unit by equipping the blower with three (3) OV cartridges, P/N 150100, donning the unit and facepiece, turning on
the blower, and passing a t test ampule, P/N 140098, near the facepiece,
breathing tube, and blower. If the wearer detects the banana oil odor, a leak is detected.
1. Cord or connector cracked or broken.
2. Loose battery case lid or damaged connector.
3. Properly working battery switch.
1. Webbing color change, fraying, cuts, nicks, or broken stitching.
2. Buckle damaged or corroded.
3. Loose hardware.
1. Cuts, nicks, cracks, punctures, or tears that may reduce the level of protection.
2. Age or heat-induced cracking, checking, or hardening.
3. Crushed, broken, or cracked connectors.
4. The tube is securely attached to the connectors and the hose clamps are tight.
35
Page 39
SURVIVAIR Opti-Fit™ Convertible Facepiece
XV. RESPIRATOR REPAIR TABLES
General Repair Table
The following repair table lists general repair steps to be followed for all respirator congurations (APR, SAR, PAPR).
COMPONENT INSTRUCTIONS
Head Strap/Headnet Remove the old head strap from the facepiece, noting the routing of the straps.
Install new head strap/Headnet.
Twin Seal/Connector 1. To remove, rotate the connector clockwise. Lift the connector hook and
Nozzle Cover 1. To remove, pinch the two lower latches while lifting the lower tab.
Nozzle 1. Remove the nozzle cover as described above.
Nozzle Adapter 1. Slightly rotate the Adapter clockwise (tightening) to allow the latch to spring
rotate the connector counterclockwise, and pull outward.
2. Remove the twin seal o-ring from the groove in lens.
3. To install, gently separate the o-ring and gasket halves of the new twin seal. A small part number tab should remain attached to each half. Do not use the o-ring if it is damaged.
4. Place the o-ring into the groove on the lens with the part number tab aligned with the slot.
5. Align the three lugs on the underside of the connector with the three slots in the lens. Insert the connector into the lens port and rotate clockwise. An audible click should be heard.
6. Install the new gasket onto the connector. The gasket must be seated in the groove in the connector.
2. Disengage the upper hook by lifting the cover upward.
3. To install, engage the cover on the nozzle upper hook.
4. Rotate cover downward until both lower latches engage the cover. An au­dible click should be heard.
2. Grasp the outside of the lens with both hands so that both thumbs can push the latches inward and downward.
3. To install, verify that the nozzle o-ring is free from debris and install the o­ring onto the nozzle.
4. Apply Christolube lubricant, SPERIAN P/N 910452, or equivalent, to o-ring.
5. From the inside of the lens, align the top and bottom ribs on the front of the nozzle with the notches in the lens.
6. Push the nozzle inward until both latches engage the lens. An audible click should be heard.
7. Inspect the installed nozzle o-ring for proper installation and seal. If the o­ring appears twisted, pinched, or extruding out of the o-ring groove, remove the nozzle, relubricate the o-ring, and reinstall the nozzle.
8. Install the nozzle cover as described above.
up and down freely.
2. From the inside of the mask, locate the word “PUSH” on the Adapter latch. Push the latch down and unscrew the Adapter from the Nozzle.
3. Install a new front (small) and rear (large) o-ring onto the Adapter.
4. Apply Christolube lubricant, SPERIAN P/N 910452, or equivalent, to the larger rear o-ring.
5. Screw the Adapter back into the nozzle until an audible click is heard.
36
Page 40
www.sperianprotection.com
Loading...