Sperian PHD6 Quick Start Guide

PHD6 Quick Reference Card
Operation Procedure Turn on
Turn off
Auto Cal
Press MODE button. Hold MODE button until display says “release.” Turn PHD6 on and allow sensors to stabilize. Take
to fresh air. Press MODE 3 times to enter “Auto Cal Mode.” At prompt press MODE to Auto-Zero. At next prompt press MODE to begin Auto-Span. Attach calibration adaptor and fittings. Apply span gas. Remove fittings when complete. If Auto-Zero or Auto-Span fails see Reference Manual.
01JUNE2011 – Rev. 2.00 – 13-323.doc
Test hand aspirated sample draw kit
Use hand aspirated sample draw kit
Test continuous sample pump
Use continuous pump
Attach sample draw kit to instrument. Test for leakage by holding finger over end of probe. Then squeeze bulb. Bulb should stay deflated until finger is removed.
Allow one squeeze of bulb for every one foot of hose for sample to first reach sensors. Continue to squeeze bulb once per second for 1 additional minute to complete sampling.
Slip pump onto instrument with tubing and probe attached. PHD6 will prompt for leak test. Hold finger over end of probe. Pump should shut down. Remove finger and press MODE to reset pump.
Pump must be tested prior to use. Allow one second for each foot of hose plus 1 additional minute to sample the environment.