Middletown, CT 06457
800 711-6776 860 344-1079
Fax 860 344 – 1068
6 May 2011
Part Number 13-322
Version 2.08
www.honeywellanalytics.com
1
PHD6 PERSONAL PORTABLE GAS DETECTORS HAVE BEEN
DESIGNED FOR THE DETECTION AND MEASUREMENT OF
POTENTIALLY HAZARDOUS ATMOSPHERIC CONDITIONS
IN ORDER TO ASSURE THAT THE USER IS PROPERLY WARNED OF
POTENTIALLY DANGEROUS ATMOSPHERIC CONDITIONS, IT IS
ESSENTIAL THAT THE INSTRUCTIONS IN THIS REFERENCE MANUAL
BE READ, FULLY UNDERSTOOD, AND FOLLOWED.
PHD6
Reference Manual
Part Number 13-322
Version 2.08
Copyright 2011
by
Honeywell Analytics, Inc.
Middletown, Connecticut 06457
All rights reserved.
No page or part of this operation manual may be reproduced in any form without
written permission of the copyright owner shown above.
Honeywell Analytics reserves the right to correct typographical errors.
Specifications are subject to change without notice.
1
Table of Contents
ERTIFICATION INFORMATION 4
C
OPERATING TEMPERATURE AND HUMIDITY LIMITS 4
SIGNAL WORDS4
WARNINGS AND CAUTIONS 4
1. DESCRIPTION7
1.1 Methods of sampling 7
1.2 Multi-sensor capability 7
1.3 Calibration 7
1.4 Alarm logic 7
1.4.1 Atmospheric hazard alarms 8
1.4.2 Low battery alarms 8
1.4.3 Sensor over range alarms 8
1.4.4 PID lamp out alarm 8
1.4.5 LEL response failure due to lack of O2 alarm 8
1.4.6 Security beep/flash 8
1.4.7 Latching alarms 8
1.4.8 Fault detection 9
1.5 Other electronic safeguards 9
1.6 Sensors 9
1.7 Optional sample draw pump 9
1.7.1 Special precautions when using the PHD6 pump 9
1.8 Data storage 9
1.8.1 Black box data recorder 10
1.8.2 Event logger 10
1.9 PHD6 design components 10
1.10 PHD6 standard accessories 10
1.10.1 Alkaline PHD6 detectors 11
1.10.2 Li-Ion PHD6 detectors 11
1.11 PHD6 kits 11
1.11.1 PHD6 Confined Space Kits 11
1.11.2 PHD6 Value Packs 11
2. BASIC OPERATIONS 11
2.1 Turning the PHD6 On 11
2.1.1 Start up with pump 12
2.1.2 Start up with PID or IR sensor 12
2.2 Operating Logic 12
2.2.1 Status Bar 12
Battery Status Icon 12
Heartbeat Symbol 13
Pump Status Icon 13
Calibration and Bump Due Warnings 13
Time 13
2.2.2 Screen Flip 13
2.3 Turning the PHD6 Off 13
2.4 Atmospheric Hazard Alarms 13
2.4.1 O2 Alarms 13
2.4.2 Combustible Gas Alarms 13
2.4.3 Toxic and VOC sensor alarms 13
2.4.4 Alarm Descriptions 13
Warning Alarms 13
Danger Alarms 14
STEL Alarms 14
TWA Alarms 14
2.5 Other Alarms 14
2.5.1 Missing Sensor Alarms 14
2.5.2 Sensor Overrange alarm 14
2.5.3 PID Lamp Out Alarm 14
2.5.4 O2 Too Low for LEL Alarms 14
2.5.5 Low Battery Alarms 14
2.5.6 Calibration Due Warning 15
2.5.7 Out of Temperature Range 15
2
PC Connection via Infrared Port 15
2.6
2.7 PID sensor reactivity ratios 15
2.7.1 Displayed VOC 16
2.7.2 Specified VOC Calibration Gas 16
2.8 Special Instructions for NDIR sensors 16
2.8.1 Special Calibration Requirement for NDIR CO2 (Carbon Dioxide) Sensor 16
2.8.2 Special Consideration for IR CH4 Methane sensor gas calibration 16
2.8.3 Hydrogen Warning for IR CH4 Methane Sensor 16
3. SAMPLING16
3.1 Manual sample draw kit 17
3.1.1 Manual sample draw kit usage 18
3.2 Motorized sample draw pump 18
3.2.1 Starting the motorized sample pump 18
3.2.2 Turning off the pump 19
3.2.3 Pump low flow alarm 19
3.3 Sample draw probe 19
4. CALIBRATION20
4.1 Functional (Bump) testing 20
4.2 Fresh Air/Zero Calibration 20
4.2.1 Fresh air calibration failure 21
4.2.2 Forced fresh air calibration 21
4.2.3 Fresh air calibration in a contaminated atmosphere 21
4.3 Gas Calibration 22
4.3.1 Gas calibration failure: All sensors except oxygen 22
4.3.2 Gas calibration failure: Oxygen sensors 22
4.4 Special Calibration Instruction for NDIR CO2 sensor 23
4.4.1 CO2 Sensor True Zero 23
4.5 Special Calibration Instructions for NDIR-CH4 Sensor 23
5. MENU OPTIONS23
5.1 Basic Menu 23
5.1.1 Entering the Basic Menu 23
5.2 Main Menu 24
5.2.1 Entering the Main Menu 24
5.2.2 Using the submenus. 25
5.2.3 Alarms Menu 25
5.2.4 Calibration Menu 25
5.2.5 Configuration Menu 26
5.2.6 Screen Menu 26
5.2.7 Information Menu 26
5.2.8 Datalogger Menu 26
6. MAINTENANCE27
6.1 Batteries 27
6.2 Replacing alkaline batteries 27
6.3 Maintaining Li-Ion battery packs 28
6.3.1 Storage guidelines for the Li-Ion battery 28
6.3.2 Charging guidelines for Li-Ion battery 28
6.3.3 Charging procedure for Li-Ion battery 28
6.3.4 Charging with the pump attached 28
6.3.5 Battery troubleshooting 28
6.4 Sensors 28
6.4.1 Sensor replacement 28
6.4.2 Care and maintenance of PID sensors 29
6.4.2.1 Troubleshooting the PID 29
6.4.2.2 Cleaning and replacing PID components 29
6.5 Sample probe assembly 30
6.5.1 Changing sample probe filters 30
6.5.2 Changing sample probe tubes (wands) 30
6.6 PHD6 Pump Maintenance 30
6.6.1 Replacing pump filters 30
APPENDICES32
Appendix A Toxic gas measurement – Warning, Danger, STEL and TWA alarms 32
1. Warning and Danger Alarms 32
2. Time Weighted Average (TWA) 32
3. Short Term Exposure Limits (STEL) 32
3
Appendix B Calibration Frequency Recommendation 33
Appendix C PHD6 Sensor Information 34
*The CO2 sensor has an internal resolution of 0.025% but displays readings rounded to the
nearest 0.01%. It will, therefore, display steps of 0.03%, 0.05%, 0.08%, 0.10%, etc. 34
Appendix D Electrochemical Toxic Sensor Cross-Sensitivity 34
**SENSOR MANUFACTURER RATES CROSS SENSITIVITY FOR (54-54-23)HCN SENSOR TO H2S AS
FOLLOWS FOR
SATURATION
N/D = NO DATA HONEYWELL ANALYTICS WARRANTY GAS DETECTION PRODUCTS 34
HONEYWELL ANALYTICS WARRANTY GAS DETECTION PRODUCTS 35
20PPM EXPOSURE AT 20C:“SHORT GAS EXPOSURE IN MINUTE RANGE; AFTER FILTER
: CA.40PPM READING”. 34
Certification Information
The PHD6 carries the following certifications:
SGS USTC Class I Division 1 Groups A,B,C,D Temp Code T3C (Approved to UL 913)
SGS USTC Class II Division 1 Groups E,F,G (Approved to UL 913)
SGS USTC Class III (Approved to UL 913)
CSA Class I, Division 1 Groups A,B,C,D Temp Code T3C
ONLY THE COMBUSTIBLE GAS DETECTION PORTION OF THIS INSTRUMENT HAS BEEN ASSESSED
FOR PERFORMANCE.
ATEX: Ex d ia IIC 170 °C (T3)
IECEx: Ex d ia IIC 170 °C (T3)
CE Mark
Operating Temperature and Humidity Limits
The PHD6’s operating temperature range is printed on the label on the
back of the instrument. Use of Honeywell Gas Detectors outside of the instrument’s
specified operating temperature range may result in inaccurate and potentially dangerous
readings.
Signal Words
The following signal words, as defined by ANSI Z535.4-1998, are used in the PHD6
Reference Manual.
indicates an imminently hazardous situation which, if not avoided, will
result in death or serious injury.
indicates a potentially hazardous situation which, if not avoided, could
result in death or serious injury.
indicates a potentially hazardous situation, which if not avoided, may
result in moderate or minor injury.
CAUTION used without the safety alert symbol indicates a potentially hazardous
situation which, if not avoided, may result in property damage.
Warnings and Cautions
1. The PHD6 personal, portable gas detector has been designed for
the detection of dangerous atmospheric conditions. An alarm condition indicates the
presence of a potentially life-threatening hazard and should be taken very seriously.
Failure to immediately leave the area may result in serious injury or death.
2.
established procedures. The safest course of action is to immediately leave the affected
area, and to return only after further testing determines that the area is once again safe
for entry. Failure to immediately leave the area may result in serious injury or death.
3.
alkaline batteries are removed from the alkaline battery pack. Removal of the alkaline
batteries from the battery pack in a hazardous area may impair intrinsic safety.
In the event of an alarm condition it is important to follow
The PHD6 must be located in a non-hazardous location whenever
4
4.
Use only Duracell MN1500 or Ultra MX1500, Eveready Energizer
E91-LR6, Eveready EN91 batteries in the alkaline battery pack. Substitution of batteries
may impair intrinsic safety.
5.
To reduce the risk of explosion, do not mix old or used batteries
with new batteries and do not mix batteries from different manufacturers.
6.
Do not charge the PHD6 with any charger other than the
appropriate PHD6 charger. Standard versions of the PHD6 must be charged with the
UL/CSA-approved charger, which is part number 54-49-103-1. European versions of the
PHD6 must be charged with the ATEX-approved charger, which is part number 54-49103-5.
7.
The PHD6 must be located in a non-hazardous location during the
charging cycle. Charging the PHD6 in a hazardous location may impair intrinsic safety.
8.
PHD6 rechargeable battery packs are supplied with Panasonic
CGR18650D Lithium-Ion batteries. The Li-Ion batteries in the battery packs may not be
replaced by the user. The rechargeable pack must be obtained from Honeywell Analytics
and replaced as a complete assembly to maintain intrinsic safety.
9.
The accuracy of the PHD6 should be checked periodically with
known concentration calibration gas. Failure to check accuracy can lead to inaccurate
and potentially dangerous readings. (The Canadian Standards Association (CSA)
requires an accuracy check using known concentration calibration gas prior to each
day’s use.)
10.
Fresh air/zero calibrations may only be performed in an
atmosphere that is known to contain 20.9% oxygen, 0.0% LEL and 0 PPM toxic gas.
11.
The accuracy of the PHD6 should be checked immediately
following any known exposure to contaminants by testing with known concentration test
gas before further use. Failure to check accuracy can lead to inaccurate and potentially
dangerous readings.
12.
A sensor that cannot be calibrated or is found to be out of
tolerance should be replaced immediately. An instrument that fails calibration may not
be used until testing with known concentration test gas determines that accuracy has
been restored, and the instrument is once again fit for use.
13.
Do not reset the calibration gas concentration unless you are using
a calibration gas concentration that differs from the one that is normally supplied by
Honeywell Analytics for use in calibrating the PHD6.
Customers are strongly urged to use only Honeywell calibration materials when
calibrating the PHD6. Use of non-standard calibration gas and/or calibration kit
components can lead to dangerously inaccurate readings and may void the standard
Honeywell Analytics warranty.
14.
Use of non-standard calibration gas and/or calibration kit
components when calibrating the PHD6 can lead to inaccurate and potentially
dangerous readings and may void the standard Honeywell Analytics warranty.
Honeywell Analytics offers calibration kits and long-lasting cylinders of test gas
specifically developed for easy PHD6 calibration. Customers are strongly urged to use
only Honeywell calibration materials when calibrating the PHD6.
15.
16.
Substitution of components may impair intrinsic safety.
For safety reasons this equipment must be operated and serviced
by qualified personnel only. Read and understand this reference manual before
operating or servicing the PHD6.
17.
A rapid up-scale reading followed by a declining or erratic reading
may indicate a hazardous combustible gas concentration that exceeds the PHD6’s zero
to 100 percent LEL detection range.
18.
The PHD6 is not designed for use in oxygen enriched
atmospheres.
5
19.
Do not use the PHD6 pump for prolonged periods in an
atmosphere containing a concentration of solvent or fuel that may be greater than 50%
LEL.
20.
Do not unplug the NDIR-CH4 or NDIR-CO2 sensors in an explosive
atmosphere. Unplugging IR sensors in an explosive atmosphere may impair intrinsic
safety.
6
1. Description
The PHD6 is a multi-sensor gas detector that can
be configured to meet a wide variety of user
requirements. This chapter provides an overview
of many of the features of the PHD6. More
detailed descriptions of the specific features of
the PHD6 are contained in the subsequent
chapters of this manual.
1.1 Methods of sampling
The PHD6 may be used in either diffusion or
sample-draw mode. In either mode, the gas
sample must reach the sensors for the instrument
to register a gas reading. The sensors are
located at the lower front of the instrument.
The sensor ports must be
kept free of obstruction. Blocked sensor
ports can lead to inaccurate and potentially
dangerous readings.
In diffusion mode, the atmosphere being
measured reaches the sensors by diffusing
through the sensor ports at the front of the
instrument. Normal air movements are enough to
carry the sample to the sensors. The sensors
react quickly to changes in the concentrations of
the gases being measured. Diffusion-style
operation monitors only the atmosphere that
immediately surrounds the detector.
The PHD6 can also be used to sample remote
locations with its hand-aspirated sample-draw kit
or with its motorized, continuous sample draw
pump. During remote sampling, the gas sample
is drawn into the sensor compartment through the
probe assembly and a length of tubing. Remote
sampling operations only monitor the atmosphere
at the end of the sample draw probe.
Use of the hand-aspirated sample draw kits is
covered in section 3.1.
Use of the motorized sample draw pump is
covered in section 3.2.
A detailed description of the PHD6 probe
assembly is given in section 6.5
1.2 Multi-sensor capability
The PHD6 can be configured to simultaneously
monitor oxygen, combustible gases and vapors,
volatile organic compounds (VOCs), and a wide
variety of toxic gases. All sensors are
replaceable in the field.
Note: The accuracy of the PHD6 must be
verified by calibration with known
concentration test gas whenever a change is
made to the sensors installed in the
instrument.
Calibration procedures are discussed in detail
in Chapter 4.
The PHD6 can utilize a variety of sensor types to
detect atmospheric contaminants including
electrochemical sensors, PID (Photo Ionization
Detector) sensors, NDIR (Non-Dispersive InfraRed Absorbance) sensors and catalytic hot-bead
LEL sensors.
Different measurement units are used depending
on the gas being measured.
Type of Hazard Measurement unit
Oxygen (O2)
Combustible gas
(LEL Sensor)
Hydrocarbon-specific
combustible gas
sensor
(NDIR – CH
)
4
Volatile Organic
Compounds (VOCs)
(PID Sensor)
Percentage by
volume
Percentage of lower
explosive limit
(%LEL) or %/Vol CH
Percentage of lower
explosive limit
(%LEL) or %/Vol CH
Parts-per-million
(PPM) or tenths of a
part-per-million
(0.1PPM)
The PHD6 detector features fully automatic fresh
air and gas calibration.
The accuracy of the
PHD6 should be checked periodically with
known concentration calibration gas. Failure
to check accuracy can lead to inaccurate and
potentially dangerous readings. (The
Canadian Standards Association (CSA)
requires an accuracy check using known
concentration calibration gas prior to each
day’s use.)
Calibration procedures are discussed in detail
in Chapter 4.
Recommended calibration frequency is
discussed in Appendix B.
1.4 Alarm logic
PHD6 gas alarms can be adjusted manually
using the PHD6’s built in menu functions, with
7
BioTrak II software via IrDA interface, or with the
IQ Database Manager Program through the IQ6
Dock. (See Chapter 6 for direct menu
programming instructions). Alarms may be set
anywhere within the nominal range of the specific
sensor. When an alarm set point is exceeded a
loud audible alarm sounds, and the bright red
LED alarm lights flash.
1.4.1 Atmospheric hazard alarms
PHD6 portable gas
detectors have been designed for the
detection of deficiencies of oxygen,
accumulations of flammable gases and
vapors, and accumulations of specific toxic
gases. An alarm condition indicating the
presence of one or more of these potentially
life-threatening hazards should be taken very
seriously. Failure to immediately leave the
area may result in serious injury or death.
In the event of an alarm
condition it is important to follow established
procedures. The safest course of action is to
immediately leave the affected area, and to
return only after further testing determines
that the area is once again safe for entry.
Failure to immediately leave the area may
result in serious injury or death.
A rapid up-scale reading
followed by a declining or erratic reading may
indicate a hazardous combustible gas
concentration that exceeds the PHD6’s zero
to 100 percent LEL detection range. Failure to
immediately leave the area may result in
serious injury or death.
The combustible gas alarms are activated when
the reading for combustible gases exceeds one
of the alarm setpoints. Combustible gas readings
are typically given in terms of percent of LEL
(Lower Explosive Limit), but may also be shown
in terms of percent-by-volume methane (CH
4
).
The PHD6 includes Warning and Danger alarms
for the both the LEL sensor and the NDIR-CH
4
sensor.
Two oxygen alarm set points have been
provided; a low alarm for oxygen deficiency and a
high alarm for oxygen enrichment.
Up to four alarm set points are provided for the
PID sensor and for each toxic gas sensor:
Warning, Danger, STEL (Short Term Exposure
Limit) and TWA (Time Weighted Average).
Appendix A discusses Warning, Danger,
STEL and TWA alarms.
1.4.2 Low battery alarms
The PHD6 includes multi-staged alarms for both
the Li-Ion and alkaline battery packs to let the
user know that the battery is running low.
For detailed information concerning the low
battery alarms, see section 2.5.5.
Use only Duracell MN1500
or Ultra MX1500, Eveready Energizer E91-LR6,
Eveready EN91 batteries. Substitution of
batteries may impair intrinsic safety.
1.4.3 Sensor over range alarms
The PHD6 will go into alarm if a sensor is
exposed to a concentration of gas that exceeds
its established range. In the case of an LEL or
NDIR-CH
sensor reading that exceeds 100%
4
LEL, the sensor channel will be automatically
disabled by the instrument and the instrument will
remain in constant alarm until it is turned off,
brought to an area that is known to be safe, and
then turned back on. The display will show a
vertical arrow with two heads in place of the
sensor reading for any channel that has gone into
over range alarm.
See section 2.5.2 for further details.
In the event of an LEL
overrange alarm the PHD6 must be turned off,
brought to an area that is known to be safe
and then turned on again to reset the alarm.
1.4.4 PID lamp out alarm
The PHD6 monitors the status of the PID lamp to
ensure that it is functioning properly. Alarms are
generated if the PHD6 determines that the lamp
is out. See section 2.5.3 for further details
1.4.5 LEL response failure due to lack of O
2
alarm
The PHD6 features automatic warning against
LEL sensor response failure due to lack of
oxygen. See section 2.5.4 for details.
1.4.6 Security beep/flash
The PHD6 includes a security beep function that
is designed to notify the user that the instrument
is powered up and running. Once enabled the
PHD6 will emit a short audible beep and give a
short flash on the LED at a user-defined interval.
The security beep/flash can be enabled manually
through the Main Menu (see chapter 5), with
BioTrak II software or through the IQ6 Dock.
1.4.7 Latching alarms
The PHD6’s alarms are self-resetting unless the
alarm latch is enabled. With the PHD6’s alarm
latch enabled, the audible and visible alarms will
continue to sound after the atmospheric hazard
has cleared. To reset the alarms, simply press
the MODE button. If the alarm latch is disabled
and the alarm condition is no longer present, the
instrument will automatically return to normal
operation, and the visible and audible alarms
cease without further input from the user.
Latching alarms can be enabled manually
through the Main Menu (see chapter 5), with
BioTrak II software or through the IQ6 Dock.
8
1.4.8 Fault detection
PHD6 software includes a number of additional
alarms designed to ensure the proper operation
of the instrument. When the PHD6 detects that
an electronic fault or failure condition has
occurred, the proper audible and visible alarms
are activated and an explanatory message is
displayed.
Faults and other electronic safeguards are
discussed in detail in section 2.5.
The PHD6 is designed to
detect potentially life threatening atmospheric
conditions. Any alarm condition should be
taken seriously. The safest course of action
is to immediately leave the affected area, and
return only after further testing determines
that the area is once again safe for entry.
1.5 Other electronic safeguards
Several automatic programs prevent tampering
and misuse of the PHD6 by unauthorized
persons. Each time the detector is turned on, the
PHD6 automatically tests the LED alarm lights,
audible alarm, internal memory and pump status
(if so equipped). The battery is monitored
continuously for proper voltage. The PHD6 also
monitors the connection of sensors that are
currently installed. The detection of any
electronic faults causes the activation of the
audible and visible alarms and causes the display
of the appropriate explanatory message.
1.6 Sensors
The PHD6 can be configured to simultaneously
monitor oxygen, combustible gases and vapors,
volatile organic compounds (VOCs) and a
number of toxic gases. The sensor configuration
of the PHD6 may be specified at the time of
purchase, or changed in the field by appropriately
trained personnel.
The PHD6 must be calibrated following any
sensor replacement.
Replacement sensor part numbers and
sensor ranges are given in Appendix C.
A sensor that cannot be
calibrated or is found to be out of tolerance
must be replaced immediately. An instrument
that fails calibration may not be used until
testing with known concentration test gas
determines that accuracy has been restored,
and the instrument is once again fit for use.
Calibration procedures are discussed in detail
in Chapter 4.
1.6.1 Cross Sensitivity
Sensor cross-sensitivity figures are given in
Appendix D.
The CO channel in the Duo-Tox sensor in the
PHD6 may exhibit high levels of cross sensitivity
to organic vapors (VOCs). For best performance
in an atmosphere known to contain VOCs, use a
dedicated CO sensor.
1.7 Optional sample draw pump
A motorized sample-draw pump is available for
the PHD6 for situations requiring continuous
"hands free" remote monitoring.
The PHD6 continuous
sample draw pump (part number 54-54-102) is
the only pump that can be used with the
PHD6.
The pump contains a pressure
sensor that detects restrictions in
airflow caused by water or other
obstructions being drawn into the
unit and immediately acts to turn
the pump off in order to protect
the sensors, pump, and other
PHD6 components from damage.
Pump status is continuously
monitored by the PHD6
microcontroller. When the pump
is active and functioning properly,
the spinning pump icon is displayed in the status
bar at the bottom of the display. Low flow or
other pump fault conditions activate audible and
visible alarms and cause the display of the
appropriate explanatory message.
1.7.1 Special precautions when using the
PHD6 pump
The rubber material used in the PHD6 diaphragm
pump is susceptible to temporary compromise by
exposure to high levels of flammable fuel and
solvent vapors. If the PHD6 is being used to
sample atmospheres that exceed 50% LEL, test
the pump frequently to ensure that pump function
has not been compromised.
To test the pump, block the end of the sampling
line (probe) inlet with a finger. The pump should
quickly go into alarm, which indicates proper
function. If the pump fails to go into alarm while
the inlet is blocked, it is not working properly; and
the PHD6 may not be providing an accurate
reading. If the pump test fails, the safest course
of action is to immediately leave the affected area
and to return only after further testing with known,
functional detection equipment confirms that the
area is once again safe for entry.
Do not use the pump to
sample for prolonged periods in conditions
where the concentration of solvent or fuel
vapors may be greater than 50% LEL.
1.8 Data storage
The PHD6 includes a black box data recorder
and an event logger as standard features. A full
datalogger is available as an upgrade at any
time.
9
1.8.1 Black box data recorder
A black box data recorder is a standard feature in
the PHD6. The “black box” is continually in
operation whether the user is aware of it or not.
The black box stores important information such
as gas readings, turn-on times, turn-off times,
temperatures, battery conditions, the most recent
calibration date and settings, types of sensors
currently installed, sensor serial numbers,
warranty expiration and service due dates, and
current alarm settings.
There is a finite amount of memory storage
available in the black box data recorder. Once
the memory is “full”, the PHD6 will begin to write
the new data over the oldest data. The black box
data recorder will store a minimum of 63 hours of
data in one-minute increments before it begins to
write new data over the oldest data. In this way,
the newest data is always conserved.
To extract the information from the black box data
recorder, the PHD6 must be returned to
Honeywell Analytics. Once the data is
downloaded from the instrument, a report will be
generated. The unit and the report will then be
returned to the user. Simply call Honeywell
Analytics’ Instrument Service Department to
obtain a return authorization number. There is no
charge for the downloading service, but the user
is responsible for any freight charges incurred.
The “black box” data recorder in the PHD6 can
be upgraded to a fully enabled datalogger at any
time. All that is required is the activation code
that corresponds to the serial number of the
PHD6 and the PHD6 Upgrade Utility Program.
1.8.2 Event logger
The event logger in the PHD6 stores data
associated with alarm conditions. Each (alarm)
event includes the following data for each of the
installed sensors:
Sensor type
Max reading
Average reading
Start time
End time
Duration of the event.
The PHD6 stores the data from the 20 most
recent alarm events. Once 20 events have been
stored, the PHD6 will begin to systematically
overwrite the data from the oldest event in
memory with data from new events. One event
may be a combination of different alarms
occurring simultaneously or in immediate
succession. The event logger may be
downloaded using BioTrak II software. The PC
must be equipped with IrDA to provide a
connection.
1.9 PHD6 design components
1. Case: The instrument is enclosed in a solid
PC (polycarbonate) case with TPE (rubber)
overmold.
2. Front face: The front face of the instrument
houses the MODE button, navigation keys,
LCD (liquid crystal display), LED alarm lights,
and audible alarm ports.
3. Display: A liquid crystal display (LCD)
shows readings, messages, and other
information.
4. Alarm LEDs: Top, front and side-mounted
LED (light emitting diode) alarm lights provide
a visual indication of alarm state.
5. Infrared Port: The infrared port is located at
the bottom of the instrument and is used for
communication between the PHD6 and a PC.
6. On / Off "MODE" button: The large black
push-button on the front of the instrument is
the "MODE" button. The MODE button is
used to turn the PHD6 on and off as well as
to control most other operations, including
the initiation of the automatic calibration
adjustment.
7.Navigation Keys: The up and down
navigation keys are located between the
MODE button and the display.
8. Sensor compartment cover: The sensors
are located in a vented compartment at the
bottom of the instrument.
9. Audible alarm ports: Two cylindrical ports
extending through the front of the instrument
on opposing sides of the MODE button house
the loud audible alarms. The waterproof
audible alarms seat directly to the rubber
inner-liner to protect the instrument against
leakage or exposure to liquids.
10. Battery pack: Two types of interchangeable
battery packs (rechargeable Lithium Ion (LiIon) and disposable alkaline) are available for
use. Li-Ion battery packs are recharged with
the pack installed on the PHD6.
11. Battery charger connector: A water-
resistant connector at the bottom of the case
assembly is used to connect the PHD6 to the
“drop in” style charger.
12. Battery Compartment / Clip: The battery
inserts from the back of the instrument. A
sturdy clip attached to the battery allows the
user to wear the PHD6 on a belt or other
article of clothing.
1.10 PHD6 standard accessories
Standard accessories included with every PHD6
include calibration adapter, additional tubing for
use during calibration, manual sample draw kit,
reference manual and quick reference card. The
manual sample draw kit consists of a sample
draw / calibration adapter, squeeze bulb,
replacement sample probe filters, ten feet/three
meters of tubing and a sample probe.
10
Standard configurations of the PHD6 are
delivered in a cardboard box with cardboard
inserts.
1.10.1 Alkaline PHD6 detectors
If the PHD6 has been purchased as an alkaline
instrument, the standard accessories include an
alkaline battery pack and a set of 3 disposable
AA alkaline batteries.
1.10.2 Li-Ion PHD6 detectors
If the PHD6 has been purchased as a Li-Ion
rechargeable instrument, the standard
accessories include Li-Ion battery pack and a
slip-in PHD6 charger.
1.11 PHD6 kits
PHD6 detectors may also be purchased as part
of a complete kit that includes calibration gas,
fixed-flow regulator and a hard-shell carrying
case.
1.11.1 PHD6 Confined Space Kits
In addition to the standard accessories listed
above, Confined Space Kits also include
calibration fittings, fixed-flow regulator with
pressure gauge, and appropriate large cylinder(s)
of calibration gas in a foam-lined, waterproof
hard-shell carrying case.
1.11.2 PHD6 Value Packs
PHD6 Value Packs include an alkaline PHD6, all
standard accessories, calibration fittings, small
cylinder of calibration gas, and fixed flow
regulator in a foam-lined non-waterproof hardshell carrying case.
2. Basic Operations
The PHD6 is a three-button gas detector. Most
day-to-day functions are initiated solely with the
MODE button. The MODE button controls:
Turning the PHD6 on and off
Turning on the backlight
Viewing the MAX, STEL and TWA reading
screens
Initiating the calibration sequence
2.1 Turning the PHD6 On
To turn the PHD6 on, press and hold the MODE
button for one second. The introduction screen is
followed by a screen showing a list of installed
sensors and the sensor ports they occupy. The
PHD6 has 5 sensor ports, but can display
readings for as many as 6 distinct gases.
The serial number will then
be shown. If the detector
has a fully enabled
datalogger, the interval and
memory capacity will be
shown.
The sampling interval is
given in minutes and
seconds. The datalogger
samples continuously, so the data stream must
be broken into intervals to be recorded. The
datalogging interval defines the frequency of the
breaks in the data stream. The capacity is the
number of hours and minutes it will take to
completely fill the datalogger’s memory. Once
the memory is filled, the PHD6 will start to write
new data over the oldest data in order to
conserve that most recent data.
The sampling interval in the fully enabled
datalogger may be modified using BioTrak II
Software, the IQ Systems or manually through
the Main Menu.
If the PHD6 is equipped with
the standard black box
datalogger, it will show Black
Box.
In the PHD6, a one-minute
sampling interval will result in
the ability to store a minimum
of 63 hours of readings
before the oldest data is overwritten by new data.
If fewer than 5 sensors are used, the capacity will
increase.
As the instrument performs a
basic electronic self test, the
date, time, temperature and
battery type will be
displayed. During the selftest, the PHD6 performs a
system memory check and
tests to see if a motorized
pump is attached to the
instrument. If a pump is detected, it will be briefly
activated during the self-test. For details on start
up procedures for pump-equipped PHD6
instruments see section 2.1.1 below.
The PHD6 will then display each installed sensor
along with any associated alarms levels.
→
→
→
11
→
For more information concerning atmospheric
hazard alarms, see section 2.4.
After the alarm screens, the PHD6 will show that
“Starting Session, Resetting Averages” followed
by the calibrations status screen. Whenever the
PHD6 is turned on, it automatically starts a new
operating session and resets STEL and TWA
calculations. The MAX reading is also reset for
the new session.
PID and IR readings that
are displayed during the sensor warm up
period should not be considered accurate.
The use of the PHD6 to monitor for
compounds detected by the PID or IR sensor
during the warm up period may lead to
inaccurate and potentially dangerous
readings.
2.2 Operating Logic
Once the PHD6 has completed the start up
sequence, the current gas readings screen will be
shown. The status bar at the bottom of the
display shows time plus calibration, pump and
battery status.
To turn on the backlight press the MODE button
once. To view the peak readings screen, press
the MODE button a second time. Press the
MODE button a third time to view the Short Term
Exposure Limit (STEL) and Time Weighted
Averages (TWA) for the operating session.
→
If calibration is due and the calibration due
warning is enabled, the user will need to
acknowledge the calibration due status by
pressing the MODE button. Once the MODE
button is pressed, the PHD6 will continue to the
current gas readings screen
and the appropriate
calibration due icons will flash
to remind the user that the
instrument is past due for
calibration. If calibration is
not due, the number of days
until the next calibration will
be shown before the
instrument proceeds to the current gas readings
screen.
2.1.1 Start up with pump
PHD6 instruments that are equipped with a builtin motorized sample draw pump will have a
slightly longer start up sequence. After the
calibration status screens, the PHD6 will prompt
you to leak test the pump.
See section 3.2 for further instructions on
using the PHD6 pump.
2.1.2 Start up with PID or IR sensor
When a PID or IR sensor is
installed in the PHD6, there will
be a warm-up period during
which the hourglass icon and
either “PID” or “IR” will be shown.
The VOC gas type and reading are
shown in reverse text.
→
Screens that are accessible with the MODE
button (including the Peak and STEL/TWA
screens) are selectable by the user. See section
5.2.6 for details.
Note: The PHD6 must be in continuous
operation for at least 15 minutes before it will
be able to calculate STEL or TWA values. For
the first 15 minutes of any operating session,
the screen will show the length of time that
the instrument has been operating instead of
the STEL and TWA values.
2.2.1 Status Bar
The status bar at the bottom of the current gas
readings shows general information including:
Battery Status
Heartbeat (instrument status)
Pump Status
PID Hourglass (PID warmup period)
PID Lamp Status (shows “Check Sen.”)
Bump Due Warning
Calibration Due Warning
Time
Battery Status Icon
The battery status icon is located at the far lower
left of the screen. The battery icon gives an
indication of how much power is left in the
battery.
12
When the battery icon is empty, it is
considered a low battery condition and the
user should take the appropriate steps to
either recharge the Li-Ion battery or
replace the alkaline batteries.
For more information on the low
battery alarms, see section 2.5.5.
IR Hourglass Symbol
The hourglass symbol along with IR
are shown in the status bar during
the IR sensor’s 1-minute warm-up
period. Once the warm-up period is over, the
hourglass will no longer be shown.
PID Hourglass Symbol
The hourglass symbol along with PID are shown
in the status bar during the PID sensor’s 5-minute
warm-up period. Once the
warm-up period is over, the
hourglass will no longer be
shown.
When a PHD6 is equipped with both an IR and a
PID sensor, the PID hourglass is shown since the
PID sensor takes longer to warm up than the IR
sensor.
Heartbeat Symbol
When the instrument is properly
charged, calibrated and functioning
normally, the heartbeat symbol will
flash in the status bar.
Pump Status Icon
If the pump is attached and
functioning, the moving fan icon
will appear in the status bar.
Calibration and Bump Due Warnings
If the PHD6 is due for
calibration the calibration
bottle icon and triangular
warning symbol will be flash
in the status bar.
Time
The time is shown on the
current gas readings screen
at the lower right.
2.2.2 Screen Flip
The screen orientation of the PHD6 may be
flipped (so that it can be read looking down from
above instead of up from below) by pressing the
up and down arrows simultaneously at the
Current Gas Readings screen.
2.3 Turning the PHD6 Off
To turn the PHD6 off, press
and hold the MODE button
until the display reads
“Release MODE to shut
down”. Then release the
MODE button. The display
will briefly show “Shutting Down” and “Saving
Sensors” before the display goes blank.
→
2.4 Atmospheric Hazard Alarms
The PHD6 is configured with a series of alarms
that are designed to warn the user of hazardous
atmospheric conditions.
The PHD6 is designed to
detect potentially life threatening atmospheric
conditions. Any alarm condition should be
taken seriously. The safest course of action
is to immediately leave the affected area, and
return only after further testing determines
that the area is once again safe for entry.
2.4.1 O2 Alarms
The PHD6 is equipped with both high and low
alarms for oxygen. Fresh air contains 20.9%
oxygen.
The low oxygen alarm indicates oxygen
deficiency and is normally set at 19.5% at the
factory.
The high alarm indicates oxygen enrichment and
is normally set at 23.5% at the factory.
2.4.2 Combustible Gas Alarms
The PHD6 is equipped with a 2-stage alarm for
concentrations of combustible gas.
The default LEL warning alarm setting is 10%
LEL. The default LEL danger alarm setting is
20% LEL.
The default warning alarm for NDIR-CH
is 10% LEL or 0.5%/vol CH
alarm is 20% LEL or 1.0%/vol CH
. The default danger
4
4
2.4.3 Toxic and VOC sensor alarms
The PHD6 is equipped with up to four different
alarms for toxic gases and volatile organic
compounds (VOCs). The combination of alarms
is designed to protect the user from both chronic
and acute toxic hazards.
Current alarm settings are shown during the
startup sequence, and can also be accessed
through the Alarms Menu.
2.4.4 Alarm Descriptions
Warning Alarms
Warning alarms indicate a hazardous
atmospheric condition that has not yet risen to
the level necessary to initiate the danger alarms.
Warning alarms can be temporarily silenced by
pressing the MODE button.
.
sensors
4
13
Danger Alarms
Danger alarms indicate a significantly hazardous
condition. The danger alarms cannot be silenced
by the user.
STEL Alarms
STEL (Short Term Exposure Limit) alarm values
represent the average concentration of
instrument readings for the target gas for the
most recently completed 15 minutes of operation.
TWA Alarms
TWA (Time Weighted Average) values are
calculated by taking the sum of exposure to a
particular toxic gas in the current operating
session in terms of parts-per-million-hours and
dividing by an eight-hour period.
2.5 Other Alarms
The PHD6 will display warnings or error
messages when it detects problems during
operation.
2.5.1 Missing Sensor Alarms
During startup, if the PHD6 fails to detect a
sensor that was present when the instrument was
last turned off, it will show the sensor channel
with “None” and the triangular warning symbol at
the Loading Sensors screen.
↔
Press MODE to acknowledge the missing sensor
If the PHD6 loses connection with a sensor
during an operating session, it
will immediately go into alarm
and show an “X” in the space
on the display allotted for the
sensor reading. The PHD6
must be turned off to reset the
missing sensor alarm.
2.5.2 Sensor Overrange alarm
The PHD6 will show a vertical
double-headed arrow and go into
alarm if a sensor is exposed to a
concentration of gas that
exceeds its established range.
In the case of an LEL reading
that exceeds 100% LEL, the LEL channel will be
automatically disabled by the instrument and the
alarm will latch (remain on) until the instrument is
turned off. The PHD6 must be turned off, brought
to an area that is known to be safe (containing
20.9% oxygen, 0% LEL and 0 PPM toxic gases),
and then turned back on. The display will show a
vertical arrow with two heads in place of the
sensor reading for any channel that has gone into
over range alarm.
A combustible sensor
overrange alarm indicates a potentially
explosive atmosphere. Failure to leave the
area immediately may result in serious injury
or death!
In the event of an LEL
overrange alarm the PHD6 must be turned off,
brought to an area that is known to be safe
(containing 20.9% oxygen, 0% combustible
gases and 0 PPM toxic gases), and then
turned on again to reset the alarm.
2.5.3 PID Lamp Out Alarm
The PID sensor in the PHD6
uses a lamp to ionize the gas
sample and generate a reading.
If the lamp fails to light during
instrument startup, the PHD6 will attempt to start
it for the duration of the warm-up cycle. If the
lamp lights, the PHD6 will complete the warm-up
cycle and then enter standard operating mode. If
the lamp fails to light by the end of the 5-minute
warm-up cycle, the instrument will display an X in
the PID channel and the instrument will go into
alarm. The status bar at the bottom of the screen
will also show “Check Sen.” to let the user know
that the PID sensor is not functioning.
The PHD6 also tests the lamp in the PID sensor
at regular intervals during normal operation. If the
PHD6 determines that the lamp has gone out, the
X will appear on the display in the PID channel,
the instrument will go into alarm and the status
bar will show “Check Sen.”
2.5.4 O
Too Low for LEL Alarms
2
The LEL sensor in the PHD6 requires a certain
amount of oxygen to function properly. When
oxygen levels fall below 11% by volume, the
PHD6 will show “X“ in place of the LEL reading
and will indicate the oxygen levels are too low.
2.5.5 Low Battery Alarms
When the battery icon in the LCD appears
empty, it means that a low battery
condition exists. Leave the area
immediately.
If the PHD6 is equipped with an alkaline
battery pack, proceed to an area that is known to
be safe area (containing 20.9% oxygen, 0%
combustible gases and 0 PPM toxic gases) and
change the batteries.
The PHD6 mustbe located
in a non-hazardous location whenever
alkaline batteries are removed from the
14
alkaline battery pack. Removal of the alkaline
batteries from the battery pack in a hazardous
area may impair intrinsic safety.
CAUTION Always turn the PHD6 off prior to
removing the battery pack. Removal of the
battery pack with the instrument turned on
may cause corruption of stored data in the
PHD6.
If the PHD6 is equipped with a Li-Ion battery
pack, proceed to an area that is known to be safe
and recharge the battery pack.
If the PHD6 continues to be used during a low
battery condition, it will eventually go into a low
battery alarm, and the warning alarm will sound
and the screen will display the low battery
warning. To silence the alarms, the user will
need to acknowledge the low battery condition by
pressing the MODE button before the instrument
will resume monitoring. Once the MODE button
is pressed, the empty battery cell and the caution
icon will flash. After 5 minutes the warning will
sound again. This cycle will continue until the
battery reaches a “very low battery” condition,
when the instrument will go into alarm for the last
time, notify the user that it is shutting itself and
proceed to turn itself off.
Alkaline battery replacement and Li-Ion
battery charging instructions are contained in
sections 6.2 and 6.3.
The PHD6 must be located
in a non-hazardous location during the
charging cycle. Charging the PHD6 in a
hazardous location may impair intrinsic
safety.
2.5.6 Calibration Due Warning
If the PHD6 is due for calibration, the triangular
warning symbol and span
bottle icons will flash in the
status bar at the bottom of the
LCD once per second as a
reminder.
2.5.7 Out of Temperature Range
If the operating temperature
falls outside of the normal
operating range of a sensor in
the PHD6, the instrument will go
into alarm and the thermometer
icon will be shown on the display at the sensor.
2.6 PC Connection via Infrared Port
PHD6 instruments that
are equipped with a fully
enabled datalogger can
be downloaded to a PC
using BioTrak II or IQ
software through the
PHD6’s infrared port. The
IrDA port is located on the
bottom of the instrument
towards the back.
1. If the PHD6 is turned off, hold the MODE
button down for about 5 seconds until
“Communication Mode” is shown. If the
PHD6 is on already, proceed to step 2.
2. Align the infrared port on
the PHD6 with the PC’s
infrared port to complete
the connection.
Note: For further
instructions concerning the
download procedure for the
PHD6, see the BioTrak II or
IQ System manual as appropriate.
2.7 PID Sensor Correction Factors
The PHD6 may be equipped with a PID (Photo
Ionization Detector) sensor designed to detect
Volatile Organic Compounds. The PID sensor
employs an ultraviolet lamp to ionize the VOCs in
the sample. The detector is then able to measure
the level of the VOCs and generate a reading.
.
PID sensors are broadband in nature. This
means that they are inherently non-specific. Any
gas or vapor that is ionized at the UV lamp
energy will give a response.
It must be understood that the
selection of a particular VOC or gas from the
onboard PID library in the PHD6 does not imply
that the detector will only respond to that
material. It only means that the sensitivity scale
(and default alarms) has been set to approximate
the target material.
Regardless of the library material selected, the
PID sensor always remains broadband in nature
and therefore will respond to any gases/vapors in
the ambient environment that are present and are
ionized at the UV lamp energy. This
consideration is particularly important when trace
or hard to detect materials (higher correction
factor (CF)) are present in highly contaminated
backgrounds. In this case the PID would be a
poor choice for detection of the target gas/vapor.
Correction factors in the PHD6
onboard PID library for various, common VOCs
and gases should be considered as approximate.
The PHD6 with PID has been fully tested and
validated only for performance with isobutylene.
For other materials requiring verified accuracy it
is necessary to calibrate the detector to the
gas/vapor to be monitored directly. Further if
using remote sample draw and/or physical
conditions in the field that differ from ambient, to
perform calibrations as close to the physical and
actual setup conditions as possible.
15
The convention in the gas detection industry
is to calibrate the PID sensor to a known
concentration of isobutylene and (as
required) to use response factors or to select
the scale of target gas from a preprogrammed menu. Sensitivity scale is
displayed on the channel with 7 character
designation whether it is isobutylene or
another material.
2.7.1 Displayed VOC
To change the displayed
VOC, first enter the Basic
Menu by holding the MODE
button to turn the PHD6 off.
When “Release MODE to
Shut Down” is shown,
continue to hold the MODE
Button until the Basic Menu
is shown.
At the Basic Menu press the down arrow once to
select “Displayed VOC”. A list of Volatile Organic
Compounds will be shown. Use the navigation
arrows to highlight the appropriate VOC and
press MODE to select it. The new VOC will be
shown when the PHD6 is restarted.
2.7.2 Specified VOC Calibration Gas
To change the calibration gas for PID sensor,
follow the instruction in section 5.2.1 to reach the
Main Menu. Then access the Calibration Menu
followed by the Gas Values submenu. Once in
the Gas Values submenu, select the VOC
sensor. Then select Cal Gas Type and specify
the appropriate compound and amount for
calibration.
2.8 Special Instructions for NDIR
sensors
Two NDIR sensors are available for the PHD6:
One for the detection of carbon dioxide (CO
and one for the detection of methane (CH
2.8.1 Special Calibration Requirement for
NDIR CO
Unlike most sensors the Infrared CO
(Carbon Dioxide) Sensor
2
sensor
2
requires two different gas sources to fully
calibrate the instrument. The reason for this is
that it is effectively impossible to zero calibrate a
CO
detector in ambient air because there is an
2
unknown and varying amount of background CO
present in the atmosphere.
See section 4.4 for more details.
),
2
).
4
2.8.2 Special Consideration for IR CH
Methane sensor gas calibration
The NDIR-CH
sensor is designed specifically for
4
the detection of methane. Gas calibration should
always be done with methane calibration gas at
the actual amount of methane shown on the
cylinder. See section 4.5 for details.
2.8.3 Hydrogen Warning for IR CH
Sensor
Unlike other types of sensors used to measure
combustible gases and vapors, the IR CH
used in the PHD6 does not respond to hydrogen.
Do not use the NDIR CH4
sensor for the detection of hydrogen. Unlike
catalytic hot-bead LEL sensors, the NDIR CH
sensor in the PHD6 does not respond to
hydrogen. Use the of the NDIR CH
4
detection hydrogen may lead to property
damage, personal injury or death.
3. Sampling
The PHD6 may be used in
either diffusion or sample-draw
mode. In either mode, the gas
sample must reach the sensors
for the instrument to register a
gas reading. The sensors are
located on the front of the
instrument near the bottom in a
vented compartment.
The sensor
ports must be kept free of
obstruction. Blocked sensor
ports can lead to inaccurate
and potentially dangerous
readings.
In diffusion mode, the atmosphere being
measured reaches the sensors by diffusing
through vents in the instrument. Normal air
movements are enough to carry the sample to
the sensors. The sensors react quickly to
changes in the concentrations of the gases being
measured. Diffusion-style operation monitors
only the atmosphere that immediately surrounds
the detector.
The PHD6 can also be used to sample remote
locations with either the hand-aspirated sampledraw kit, or with the motorized sample draw
pump. During remote sampling, the gas sample
is drawn into the sensor compartment through the
probe assembly and a length of tubing.
The PHD6 is delivered
with a sample draw kit that contains 10 feet/3
meters of polyester urethane (fuel-resistant)
tubing part number 53-001. This material is
completely compatible with common
combustible gases/vapors, and the toxic
2
gases CO and H
with a sample draw pump or kit to sample
with any of the gas types and tubing lengths
S. When using the PHD6
2
4
Methane
4
sensor
4
for the
4
16
listed in the chart below, FEP-Lined Tubing
(part number 53-036) should be used.
Gas Type Tubing Length
CL2, CLO2 Up to 10 ft/3m Max.
HCN Up to 100 ft/30m
Max.
PID, SO2, NO, NO2,
PH
, NH3
3
> than 10 ft/3m up
to 100 ft/30m Max.
Standard polyester urethane (fuel-resistant)
tubing (part number 53-001) can be used
otherwise. Use of other types of tubing may
cause inaccurate and potentially dangerous
readings that could result in serious injury or
death.
For sampling using a PID sensor please refer
to the Application Note titled “Usage and
Applications of PID sensors version B1”
included with your PhD documentation or
contact customer service at 800-711-6776 to
request a copy.
Do not use the NDIR CH4
sensor for the detection of hydrogen. Unlike
catalytic hot-bead LEL sensors, the NDIR CH
sensor in the PHD6 does not respond to
hydrogen. Use of the NDIR CH
for the
4
detection of hydrogen may lead to property
damage, personal injury or even death.
4
3.1 Manual sample draw kit
The manual sample draw kit is comprised of a
sample draw probe, 2 sections of tubing, a
squeeze bulb and an adapter that is used to
connect the sample draw accessories system to
the PHD6.
Note: The maximum amount of tubing that
can be used with the manual sample draw kit
is 50 feet/15 meters.
17
3.1.1 Manual sample draw kit usage
The PHD6’s manual sample
draw kit may not be used
for the detection of
chlorine (Cl
dioxide (ClO
) or chlorine
2
) due to the
2
reactive properties of
these gases.
To use the manual sample
draw kit:
1. Connect the short
section of hose that
comes off the squeeze
bulb to the sample draw
adapter.
2. To test the seals in the sample draw system,
cover the end of the sample draw probe with
a finger, and squeeze the aspirator bulb. If
there are no leaks in the sample draw kit
components, the bulb should stay deflated for
a few seconds.
3. Secure the calibration adapter (with the
sample draw assembly attached) to the
PHD6 by inserting the tab and tightening the
knurled screw into the brass nut at the
bottom of the adapter.
4. Insert the end of the sample probe into the
location to be sampled.
5. Squeeze the aspirator bulb to draw the
sample from the remote location to the
sensor compartment.
To ensure accurate readings while using
the manual sample draw kit, it is
necessary to squeeze the bulb once for
every one foot of sampling hose for the
sample to first reach the sensors, and
then to continue squeezing the bulb once
per second for an additional 45 seconds
or until readings stabilize. As an example,
if 10 feet/3 meters of tubing is used, it will
be necessary to draw the sample in by
squeezing the bulb continuously for a
minimum of 55 seconds or until readings
stabilize.
6. Note the gas measurement readings.
CAUTION: Hand-aspirated remote sampling
only provides continuous gas
readings for the area in which
the probe is located while the
bulb is being continuously
squeezed. Each time a reading
is desired, it is necessary to
squeeze the bulb a sufficient
number of times to bring a
fresh sample to the sensor
compartment.
(part number 54-54-102) is the only pump that
can be used with the PHD6.
A motorized sample-draw pump is available for
the PHD6 for situations requiring continuous
"hands free" remote monitoring. The pump is
powered by the PHD6 battery. When the pump is
attached to the instrument, the spinning fan icon
will be shown on the display in the current gas
readings screen.
Note: The maximum amount of
tubing that can be used with the
motorized sample draw pump is
100 feet/30 meters.
To ensure accurate readings while using the
continuous sample pump, it is necessary to
allow the pump to draw the sample for one
second for every one foot of sampling hose
plus an additional 45 seconds or until
readings stabilize. For example, with 10 ft/3m
of tubing, it will be necessary to allow a
minimum of 55 seconds for the sample to be
drawn into the sensor chamber and for the
readings to stabilize.
PHD6 instruments are designed to automatically
recognize the pump whenever it is attached to
the instrument. If the pump is attached when the
PHD6 is turned off, the instrument will
automatically initiate the pump start up sequence
when the instrument is turned on. If the pump is
attached while the instrument is running, the
PHD6 will automatically initiate the pump test
sequence before returning to the current gas
readings screen.
Do not use the PHD6
pump for prolonged periods in an atmosphere
containing a concentration of solvent or fuel
that may be greater than 50% LEL.
3.2.1 Starting the motorized sample pump
First attach the probe and tubing to the pump,
then secure the pump (with the sample draw
assembly attached) to the PHD6 by hooking the
tabs on the pump into the corresponding slots on
the back of the PHD6. Once the pump is in
position over the sensors, tighten the knurled
screw on the adapter into receptor at the center
of the sensor cover.
Note: The sample probe assembly must be
attached to the pump when the pump is
attached to the instrument.
Once the pump is recognized, the pump test
sequence will be initiated automatically. The
instrument will instruct you to block the sample
inlet.
3.2 Motorized sample
draw pump
The PHD6
continuous sample draw pump
18
→
Block the sampling inlet by placing a finger over
the end of the sample probe assembly. Once the
blockage is detected, the PHD6 will indicate that
the test has been passed and instruct you to
remove the blockage. Once the blockage is
removed, it will proceed to the current gas
readings screen and the pump icon will be shown
in the status bar.
in place, contaminants may cause damage to
the pump, sensors and internal components
of the PHD6
When the pump is active and
functioning properly, the moving
pump icon is shown on the lower
status bar on the display. Low
flow or other pump fault
conditions activate audible and visible alarms and
cause the display of the appropriate explanatory
message.
→
→
If the instrument is unable to detect the vacuum
resulting from the pump blockage within 30
seconds, the test will fail, the instrument will go
into alarm and you will be directed to remove the
pump.
Remove the pump and press the MODE button to
resume diffusion operation.
3.2.2 Turning off the pump
To turn off the pump, simply remove the pump
from the bottom of the instrument. The screen
will show “Pump Fault” followed by “Pump
Disconnected”. Press MODE to continue without
the pump.
→
Press MODE once the blockage has been
cleared to restart the pump.
The pressure sensor in the sample draw pump is
designed to detect pressure changes while the
sample-draw probe is being held in a vertical
position. If the probe is held horizontally or at a
low angle while inserted into a fluid, a pressure
drop sufficient to cause the pump to shut down
may not be generated, and water could be drawn
into the pump assembly causing damage to the
pump, sensors and internal components of the
PHD6.
CAUTION: Insertion of the sample draw tube
into a fluid horizontally or at a low angle may
lead to water ingress and may cause damage
to the sensors and internal components of the
PHD6.
If the PHD6 determines that a significant increase
in pressure has occurred, it will go into alarm and
notify the user that there is a blockage of the
pump. The display will alternate between the
following two screens.
Remove the blockage and press the MODE
button to acknowledge the alarm and resume
sampling.
3.2.3 Pump low flow alarm
The PHD6 Pump contains a pressure sensor that
continuously monitors for restrictions in airflow
caused by water or other fluids being drawn into
the unit and immediately acts to turn the pump off
in order to protect the sensors, pump, and other
PHD6 components from damage.
CAUTION: Never perform remote sampling
with the PHD6 without the sample probe
assembly. The sample probe handle contains
replaceable filters designed to block moisture
and remove particulate contaminants. If the
pump is operated without the probe assembly
3.3 Sample draw probe
The PHD6’s sample draw probe is the standard
probe assembly from Honeywell Analytics. The
sample probe handle contains moisture barrier
and particulate filters designed to remove
contaminants that might otherwise harm the
instrument.
Particulate contaminants are removed by means
of a cellulose filter. The hydrophobic filter
includes a Teflon barrier which blocks the flow
of moisture as well as any remaining particulate
contaminants.
Sample probe filters should be replaced
whenever visibly discolored due to contamination.
19
See section 6.5 for a probe diagram and a list
of available sample probe filter replacement
kits.
4. Calibration
The accuracy of the PHD6 should be verified on
a regular basis. Verification can be as simple as
performing a bump test, which is described below
in section 4.1. If the instrument fails the fresh air
test, then it must be fresh air calibrated before
use. If the instrument fails the bump test with
calibration gas, it must be successfully span
calibrated before use.
Note: The NDIR-CO
cannot be zero calibrated in fresh air. For
specific instructions on calibrating the CO
sensor, proceed to section 4.4.
Note: The NDIR-CH
must be calibrated with methane calibration
scale to the actual amount of methane in the
cylinder in terms of percent volume methane.
See section 4.5 for details.
*
Association (CSA) requires combustible gas
sensors to be bump tested prior to each day’s
use with calibration gas containing between
25% and 50% LEL. The functional (bump) test
procedure is covered in section 4.1.
**
Association (CSA) requires combustible gas
sensors to undergo calibration when the
displayed value during a bump test fails to fall
between 100% and 120% of the expected
value for the gas.
For Honeywell Analytics’ official
recommendations concerning calibration
frequency, see Appendix B.
4.1 Functional (Bump) testing
The accuracy of the PHD6 may be verified at any
time by a simple functional (bump) test.
To perform a functional (bump) test, do the
following:
1. Turn the PHD6 on and wait at least three
minutes to allow the readings to fully
stabilize. If an IR or PID sensor is in use,
wait until the stabilization period ends before
proceeding. If any of the sensors have just
been replaced, the new sensor(s) must be
allowed to stabilize prior to use. See section
6.4 for further details on sensor stabilization
requirements.
2. Make sure the instrument is located in fresh
air.
sensor used in the PHD6
2
sensor used in the PHD6
4
The Canadian Standards
The Canadian Standards
2
Figure 4.1 Bump Test / Gas calibration set up
3. Verify that the current gas readings match
the concentrations present in fresh air. The
oxygen (O
) sensor should read 20.9%/vol.
2
(+/-0.2%/vol.). The readings for the LEL
sensor should be 0% LEL. The PID, NDIRCH
and toxic sensors should read 0 parts-
4
per-million (PPM) in fresh air. For the NDIRCO
sensor, a carbon dioxide level between
2
0.03% and 0.10% is considered normal in
fresh air. If the readings deviate from the
expected levels in a fresh air environment,
proceed to section 4.2 and perform the fresh
air calibration adjustment then proceed to
step 4.
4. Attach the calibration adapter and connect
the calibration cylinder to the PHD6 as shown
in figure 4.1. Flow gas to the sensors.
5. Wait for the readings to stabilize. (Forty-five
seconds to one minute is usually sufficient.)
6. Note the readings. Toxic, VOC and
combustible gas sensor readings are
considered accurate in a bump test if they
are between 90%* and 120% of the expected
reading as given on the calibration cylinder.
If the readings are considered accurate, then
the instrument may be used without further
adjustment. If the readings do not fall within
90%* and 120% of the expected reading as
given on the calibration cylinder, then
readings are considered inaccurate. If
readings are considered inaccurate, proceed
to section 4.3 and perform the gas
calibration.
*
Note: The Canadian Standards Association
(CSA) requires combustible gas sensors to
undergo calibration when the displayed value
during a bump test fails to fall between 100%
and 120% of the expected value for the gas.
Honeywell multi-calibration gas mixtures
contain approximately 18% oxygen. During
the bump test the oxygen sensor should read
within +/-0.5% of the level given on the
calibration cylinder.
4.2 Fresh Air/Zero Calibration
Note: The NDIR-CO2 sensor in the PHD6 may
not be zero calibrated in fresh air. See
section 4.4 for further instructions.
20
Fresh air/zero calibrations
may only be performed in an atmosphere that
is known to contain 20.9% oxygen, 0.0% LEL
and 0 PPM toxic gas.
To initiate the fresh air/zero calibration:
1. Press the MODE button three times within
two seconds to begin the fresh air/zero
calibration sequence. The PHD6 will briefly
display AUTO CAL and then begin a 5second countdown.
2. Press the MODE button before the end of the
5-second countdown to begin the fresh
air/zero calibration. The fresh air/zero
calibration is initiated when the PHD6 shows
“Calibrating” on the screen.
→
3. The PHD6 will indicate when the fresh
air/zero calibration is complete. It will then
proceed to a second 5-second countdown for
the gas calibration. If gas calibration is not
required, allow the countdown to reach 0
without pressing the MODE button.
→
For instructions on the Gas Calibration,
proceed to section 4.3.
4.2.1 Fresh air calibration failure
In the event of a fresh air
calibration failure, the
alarms will be activated and
the instrument will display
the following screen. Note
that the sensor(s) that fail
the zero calibration are
shown (in this case, CO)
After 3 seconds, the PHD6
will return to the current gas readings screen and
the visual and audible alarms will cease.
When calibration is due, the triangular warning
symbol along with the span bottle icon the
PHD6’s status bar will show
If a successful fresh air calibration is not
performed prior to instrument shut down, the
PHD6 will note that Fresh Air Calibration is due
during instrument start up.
Possible causes and solutions
1. The atmosphere in which the instrument is
located is contaminated (or was
contaminated at the time the instrument was
last fresh air calibrated.
2. A new sensor has just been installed.
3. Instrument has been dropped or banged
since last turned on.
4. There has been a significant change in
temperature since the instrument was last
used.
Recommended action:
Take the instrument to fresh air and allow
readings to stabilize. Perform the fresh air/zero
adjustment again. If the manual fresh air/zero
procedure fails to correct the problem, perform
the manual fresh air / zero calibration procedure
as described in section 4.2.2 below.
4.2.2 Forced fresh air calibration
The PHD6 includes safeguards to prevent fresh
air calibration in contaminated environments. If
the standard fresh air calibration fails a second
time, the instrument may be “forced” to accept
the fresh air calibration by performing the manual
fresh air calibration.
Fresh air calibrations may
only be performed in an atmosphere that is
known to contain 20.9% oxygen, 0.0% LEL
and 0 PPM toxic gas. Performing a fresh air
calibration in a contaminated atmosphere
may lead to inaccurate and potentially
dangerous readings.
1. Initiate the standard fresh air / zero
calibration sequence by pressing the MODE
button three times in rapid succession. The
5-second countdown will begin.
2. Press and hold the down arrow key and then
press the MODE button before the end of the
5-second countdown. Continue to hold the
down arrow.
3. The fresh air/zero calibration is complete
when the instrument begins another 5second countdown for the gas calibration. If
gas calibration is not required, allow the
countdown to reach 0 without pressing the
MODE button.
If the PHD6 still fails to calibrate after this
procedure is attempted, contact Honeywell
Analytics.
4.2.3 Fresh air calibration in a contaminated
atmosphere
To fresh air calibrate the PHD6 in a contaminated
atmosphere, connect a cylinder of “zero air”
containing 20.9% oxygen and no contaminants to
the PHD6 and flow gas to the instrument. Then
perform the fresh air calibration. See figure 4.1
above for setup.
21
4.3 Gas Calibration
Once the fresh air / zero calibration has been
successfully completed, the PHD6 will
automatically proceed to the automatic gas
calibration countdown screen.
Press the MODE button before the countdown is
complete to initiate the gas calibration. The
screen will immediately show “APPLY GAS” and
then list the sensors for calibration and the
expected levels of calibration gas.
→
Note: Honeywell
Analytics recommends
the use of multicomponent calibration
gas for calibrating the
PHD6.
Apply calibration gas. The
readout will change to a
numerical display almost
immediately and show the current readings along
with the expected calibration gas value.
If multiple cylinders are required to complete the
calibration, the PHD6 will prompt the user to
apply the next cylinder as needed.
As sensors are calibrated,
the PHD6 will briefly show
the reserve values for each
sensor. The reserve values
give an indication of the
remaining sensitivity of the
sensors. When the reserve
value for a specific sensor
reaches 0%, it is time to
replace the sensor.
The oxygen sensor is tested
for response to diminished
oxygen levels during gas
calibration. Honeywell multigas calibration cylinders
contain approximately
18.0% oxygen. In order to
pass the gas calibration, the
PHD6 must register an
oxygen reading below
19.5% during gas calibration. If the detector fails
to register the reduced oxygen levels during the
gas calibration, it will show “Check O2 Sensor
Response”. Press MODE to acknowledge.
See section 4.3.2 below if the oxygen sensor
does not detect the drop in oxygen level and
fails the gas calibration.
Note: Disconnect the calibration assembly as
soon as the calibration is complete.
4.3.1 Gas calibration failure: All sensors
except oxygen
When there is a gas calibration failure, the
display will show CAL Error and display the
sensor whose calibration has failed.
If the instrument fails to recognize the correct
type or concentration of calibration gas, it will
show “no GAS”.
When gas calibration is due, the PHD6’s display
will show the warning symbol while intermittently
displaying the calibration bottle in the gas
readings screen.
The PHD6 will also display a “Needs Cal”
message for any sensors that are currently due
for calibration during instrument start-up.
Possible causes of gas calibration failure and
remedies:
1. Empty calibration gas cylinder. Verify that
there is calibration gas in the cylinder.
2. Expired calibration gas cylinder. Verify that
the expiration date on the cylinder has not
passed.
3. Calibration gas setting does not correspond
to calibration gas concentration. If the values
on the calibration cylinder differ from the
calibration gas settings in the PHD6, the
PHD6’s calibration gas settings must be
changed to match the new values. Changing
the calibration gas settings can be done
manually through the MODE button or
through BioTrak II using an IrDA link to the
instrument.
4. LEL only: Type of calibration gas (standard)
has changed significantly. LEL calibration
gas may be based on several different
response standards. Methane, propane and
pentane are the most common. If using a
new cylinder of calibration gas, make sure
that the type and amount of combustible gas
is identical to that of the previous bottle.
Honeywell Analytics offers calibration gases
in Methane, Propane Equivalent and Pentane
Equivalent.
5. Dead sensor. Replace sensor.
6. Instrument problem. Return the instrument to
Honeywell Analytics. Call the phone number
on the front of this manual.
4.3.2 Gas calibration failure: Oxygen
sensors
Honeywell multi
calibration gas cylinders
contains approximately
18.0% oxygen. The
reduced oxygen level in
the calibration gas
cylinder allows the
oxygen sensor’s response
to be tested in the same
22
manner as the toxic and LEL sensors.
If the O2 sensor fails to register a reading below
19.5% during the gas calibration, the display will
show “Check O2 Sensor Response”. Press
MODE to continue.
If the oxygen sensor fails to register the drop in
oxygen during the gas calibration while being
challenged with calibration gas containing less
than 19.0% oxygen, it should be considered out
of tolerance and retired from service immediately.
See section 5.2.4 under Gas Values for more
information on the O2 sensor check.
A sensor that cannot be
calibrated or is found to be out of tolerance
should be replaced immediately. An
instrument that fails calibration may not be
used until testing with known concentration
test gas determines that accuracy has been
restored, and the instrument is once again fit
for use.
Possible causes and remedies for oxygen
sensor failure:
1. Calibration gas cylinder does not contain a
reduced level of oxygen. Verify that the
cylinder contains less than 19.0% oxygen.
To challenge the oxygen sensor without
calibration gas, hold you breath of about 10
seconds and then slowly exhale directly onto
the face of the sensor (in the same way you
would attempt to fog up a piece of glass). If
the descending oxygen alarm is set to 19.5%,
the instrument should go into alarm after a
few seconds. If the oxygen sensor fails to go
into alarm during the exhalation test, the
oxygen sensor must be replaced.
2. Oxygen sensor has just been replaced and
has not had time to stabilize.
3. Oxygen sensor failure.
4.4 Special Calibration Instruction
for NDIR CO
The Infrared CO2 sensor requires two different
gas sources for full calibration. The reason for
this is that it is effectively impossible to zero
calibrate a CO
there is an unknown and varying amount of
background CO
4.4.1 CO
To determine if the CO
calibration, connect the PHD6 to a cylinder of
calibration gas that contains 0.00% CO
instrument is in normal operation.
If the reading shows 0.00% CO
sensor does not require zero calibration.
Disconnect the cylinder from the PHD6.
If the reading shows anything other than 0.00%
CO
, leave the calibration gas on and press the
2
MODE button three times within two seconds to
initiate the zero calibration sequence. Press
sensor in ambient air because
2
present in the atmosphere.
2
Sensor True Zero
2
sensor
2
sensor requires zero
2
while the
2
, then the CO2
2
MODE again when prompted to begin the zero
calibration. Instruments equipped with a CO
2
sensor will automatically show the message
“Press MODE if applying Zero Air” with another 5second countdown. Press MODE again to begin
the true zero calibration and follow the
instructions given on the screen. Once the zero
calibration is complete, remove the zero air
cylinder from the instrument and proceed to the
gas calibration (if necessary).
The gas calibration of the CO
sensor is
2
performed during the standard gas calibration
that is described above in section 4.3. The PHD6
will automatically prompt the user to apply the
CO
calibration gas during the standard gas
2
calibration sequence.
4.5 Special Calibration Instructions
for NDIR-CH4 Sensor
In many ways, the NDIR-CH4 sensor used in the
PHD6 is similar to a hot bead LEL sensor. For
the purpose of calibration, they are very different.
While LEL sensors can be calibrated with a
number of other gases when properly configured,
The NDIR-CH
sensor must be calibrated with
4
methane to the exact amount shown on the
calibration gas cylinder. (This is different from
LEL sensors, where methane may be used for
calibration, but is often done at a scale that
makes the readings mimic those given by a
specific amount of propane or pentane).
The NDIR CH
sensor in
4
the PHD6 must be calibrated using methane
(CH
) calibration gas at the actual amount
4
shown on the cylinder. The default
calibration gas value for the NDIR-CH
sensor
4
is 50% LEL. The appropriate calibration gas
level for the 50% LEL default calibration gas
setting is 2.50%/vol. CH
. Use of
4
inappropriate calibration gas may lead to
inaccurate and potentially dangerous
readings.
5. Menu Options
The PHD6 operating firmware includes two menu
options: the Basic Menu and the Main Menu.
5.1 Basic Menu
The Basic Menu is a shortened version of the
Main Menu that offers immediate access to a few
key functions including:
PID On/Off (enable or disable the PID
sensor)
Displayed VOC (select the target
compound for the VOC sensor)
Contrast (display’s light vs. dark setting)
Main Menu access
5.1.1 Entering the Basic Menu
To access the Basic Menu, with the PHD6 on and
the current gas readings screen shown, hold the
23
MODE button down until the PHD6 beeps four
times and the “Release MODE to Shut Down” is
shown. Then continue to hold the MODE Button
until the Basic Menu is shown.
To navigate through the menu options, use the
up and down navigation arrows to highlight the
desired submenu and press MODE to enter the
submenu.
5.2 Main Menu
The PHD6 is fully configurable through the Main
Menu. The Main Menu contains 6 sub menus
that lead to controls for the individual instrument
functions.
To navigate through the menu options, use the
up and down navigation arrows to highlight the
desired submenu and press MODE to enter the
submenu.
To navigate through the menu options, use the
up and down navigation arrows to highlight the
desired submenu and press MODE to enter the
submenu.
Main Menu Options Diagram
5.2.1 Entering the Main Menu
There are two paths into the main menu.
If the instrument is on, press
and hold the MODE button
down for three seconds until
“Shutting Down” is shown,
then release the MODE
button. The next screen will
show “shutting down…” along
with two black blocks at the
bottom of the screen. Press
and hold the two arrow keys
while the two blocks are
shown to enter the main
menu.
If the instrument is off, press
the MODE button to start the
instrument. When “Starting
Session, Resetting Averages”
is shown along with two black
blocks, press and hold the
two arrow keys while the two blocks are shown to
enter the main menu.
24
The Main Menu is the access point to 6
submenus that control virtually every aspect of
the PHD6’s functionality.
NOTE: Changes made in the Main Menu can
have a direct affect on the PHD6’s
functionality and should only be made by
those who are trained in proper gas detection
and monitoring techniques.
5.2.2 Using the submenus.
In the Main Menu and the sub-menus, use the up
and down arrows to navigate between the options
and press MODE to enter. Three buttons will
appear on the display to show the
functions of the MODE button and the two
navigation keys on any screen that allows
instrument setup changes.
5.2.3 Alarms Menu
The Alarms Menu contains the following 6
submenus (options in parenthesis). Description
follows (as needed).
Current Alarms (select any sensor to view
current sensor alarm settings, then select any
current sensor alarm to make changes)
Default Alarms (scroll to view default sensor
alarms for each recognized sensor plus
option to Set Default Alarms for all sensors
Alarm Latch (set on or off)
The PHD6’s alarms are self-resetting
unless the alarm latch is enabled. With the
PHD6’s alarm latch enabled, the audible
and visible alarms will continue to sound
after the atmospheric hazard has cleared.
Press the MODE button to reset the
alarms. If the alarm latch is disabled and
the alarm condition is no longer present,
the instrument will automatically return to
normal operation, and the visible and
audible alarms cease without further input
from the user.
Temp Alarms (enable or disable high and
low temperature alarms)
If the operating temperature falls outside of
the operating range of the PHD6, the
instrument will go into alarm and the
thermometer icon will be shown on the
display.
Event History (use up and down arrows to
scroll through saved alarm events – includes
time, duration and peak and average sensor
readings during the event)
Vibrator (if equipped) (enable or disable the
vibrating alarm)
5.2.4 Calibration Menu
Fresh Air Cal (initiates Fresh Air Calibration
sequence)
Fresh air/zero calibrations
may only be performed in an atmosphere that
is known to contain 20.9% oxygen, 0.0% LEL
and 0 PPM toxic gas.
Gas Calibration (initiates Gas Calibration
sequence (calibration gas required))
Gas Cal (initiates true O2 Zero Calibration
O
2
sequence)
Note that this procedure requires a
cylinder of calibration gas that contains
0.0% oxygen.
Gas Values (select any sensor to view or
change current calibration gas values).
Note: The selection of the calibration gas
for the PID sensor is NOT linked to the
displayed substance. A ratio is used to
calculate readings for various VOCs
against the calibration standard. See
section 2.7 for more details on the PID gas
values.
Note: In the case of the oxygen sensor,
the O
gas setting can be used to enable
2
or disable the oxygen sensor check that
takes place during gas calibration with
multi calibration gas. To disable the
oxygen sensor check, select “No”.
Disabling
the oxygen sensor check
may result in the failure to
recognize an oxygendeficient atmosphere.
Always use a multi cal
gas cylinder containing
18% oxygen to
calibrate the PHD6.
Reminders/Lock (access to submenus
below)
Cal on Startup (enable or disable)
When enabled, calibration is
automatically initiated whenever the
instrument is turned on. The calibration
can be bypassed (unless Cal Due Lock is
enabled) by letting the clock run out and
proceeding to the current gas readings
screen. Cal on Startup is usually
disabled on new instruments and must
be enabled by the user.
Cal Reminder: (adjust between every day
and every 180 days). The default setting
for standard instruments leaving the factory
is 30 days.
To disable the cal reminder, set the value
to 0.
Cal Lock: (enable or disable)
Enable to require calibration when the
Cal Reminder is on. PHD6 automatically
shuts down if Cal Lock is enabled, and
calibration is due but not performed. Cal
Lock is usually disabled on new
instruments and must be enabled by the
user.
25
Bump Reminder: (enable, disable and
adjust between every day and every 30
days)
Used exclusively with the IQ6 Dock.
Reminds the user to process the
instrument in the dock. To disable set
the value to 0. The Bump Reminder is
usually disabled on new instruments and
must be enabled by the user.
Service Interval (enable, disable and
adjust between every day and every 730
days (2 years))
The service interval is a reminder that
tells the user when the instrument is due
for service. The Service Interval is
usually disabled on new instruments and
must be enabled by the user.
Service Done (reset service date)
Used to reset the service interval
following instrument service.
Cal History (scroll through recent
calibrations, includes span reserve listing –
which allows for predictive maintenance)
5.2.5 Configuration Menu
Security Beep (enable or disable)
Once enabled the PHD6 will emit a short
audible beep and give a short flash on the
LEDs at a user-defined interval to notify the
user that the instrument is powered up and
running. The Security Beep is usually
disabled on new instruments and must be
enabled by the owner.
Basic Passcode (enable, disable and
change passcode)
Enable to require the entry of a passcode
to access the Basic Menu. The Basic
Passcode is usually disabled on new
instruments and must be enabled by the
owner. To permit access to the Basic
Menu, and restrict it from the Main Menu,
the Basic Passcode must differ from the
Main Passcode.
Main Passcode (enable, disable and change
passcode)
Enable to require the entry of a passcode
to enter the Main Menu. The Main
Passcode is usually disabled on new
instruments and must be enabled by the
owner. The Main Passcode can be used to
enter both the Main Menu and the Basic
Menu.
Display Formats (contains submenus for
sensor readings, sensor clamping and
temperature)
Sensor readings (for toxic gases select
PPM (XX) or tenths-of-a-PPM (X.X) for
sensors with this capability (such as H
For NDIR-CH
CH
(the CH4 reading will display in
4
choose between LEL and
4
2
S).
%/Vol.)). Sensors that cannot be adjusted
will show “Fixed”.
Temperature (select display in Celsius or
Fahrenheit) Most PHDs leave the factory
configured to read temperature in
Fahrenheit unless the customer requests
otherwise.
Language (select English, French or
Spanish). Most PHDs leave the factory
configured in English unless the customer
requests otherwise.
Date/Time (set time and date)
5.2.6 Screen Menu
Contrast (screen contrast setting)
Orientation (shifts display to be viewable
from top or bottom of the instrument)
Backlight Mode (select continuous, Timed
Off or Time Auto)
Select Continuous to have the backlight
on at all times,
Select Timed Off to require a MODE press
or an alarm condition to activate the
backlight. The default setting for most new
PHD6 instruments when leaving the factory
is to turn the backlight off after 20 seconds.
Select Time Auto to enable the automatic
backlight for low light conditions.
Backlight Time (set the time before the
backlight turns off in Time Off Mode)
Enable Screens (select the screens that are
accessible by sequentially pressing the
MODE button including: Peak, Average,
STEL and TWA screens.
5.2.7 Information Menu
Versions (view instrument serial number,
software version, and time and date of
instrument manufacture)
Service Info (view Honeywell Analytics’
phone contact numbers).
5.2.8 Datalogger Menu
Interval (set datalogger interval between 1
second and 1 hour) (menu option only not
available in Black Box Datalogger versions)
The datalogger samples continuously, so the
data stream must be broken into intervals to
be recorded. The datalogging interval defines
the frequency of the breaks in the data
stream. The interval may be set anywhere
between one second and one hour by using
the navigation arrows as detailed below. The
default datalogging interval is 1 minute. At a
one-minute interval, the PHD6 will log a
minimum of 63 hours of data before the
oldest data is overwritten by newer data.
Sessions (view datalogger session data
including date, time, interval, temperature
26
and sensor minimum and maximum
readings)
Clear Datalog (clears all information from the
datalogger)
Select User (User name will be saved in the
session data)
Users’ names must be entered in BioTrak II
to appear in the user list.
Select Location (Location name will be
saved in the session data)
Location names must be entered in
BioTrak II to appear in the location list.
User on Startup (enable or disable a prompt
to select user and location at startup)
User and location names must be entered
into the instrument via BioTrak II before this
option can be enabled.
6. Maintenance
To prevent ignition of
flammable or combustible atmospheres,
disconnect power before servicing any parts
in the PHD6.
6.1 Batteries
The PHD6 is powered by
interchangeable alkaline
and Li-Ion rechargeable
battery packs.
To remove the battery pack
first loosen the top center
screw on the back of the
instrument, then gently pull
the top of the battery away
from the instrument. The battery is hinged from
below. Remove the battery once the top clears
the upper housing by pulling up and away.
CAUTION Always turn the PHD6 off prior to
removing the battery pack. Removal of the
battery pack with the instrument turned on
may cause corruption of stored data in the
PHD6.
Note: Center screw on ATEX / European
version may be slightly different.
6.2 Replacing alkaline batteries
The alkaline battery pack contains three AA
alkaline batteries.
The PHD6 must be located
in a non-hazardous location whenever
alkaline batteries are removed from the
alkaline battery pack. Removal of the alkaline
batteries from the battery pack in a hazardous
area may impair intrinsic safety.
Use only Duracell MN1500
or Ultra MX1500, Eveready Energizer E91-LR6,
Eveready EN91 batteries. Substitution of
batteries may impair intrinsic safety.
27
To replace the alkaline batteries:
1. Remove the battery pack from the PHD6 as
discussed in above in section 6.1.
2. Loosen the two screws at the top of the
battery pack by turning each ¼ turn
counterclockwise.
3. Remove the three alkaline batteries and
replace them. Be sure to align the positive
and negative ends in accordance with the
diagram under
each battery.
4. Reinstall the back
cover plate that
was removed in
step 2.
5. Return the battery
pack to the PHD6
and re-tighten the top center screw. The
PHD6 will automatically turn itself on once
the battery pack is reinstalled.
6.3 Maintaining Li-Ion battery packs
The PHD6 may be equipped with a rechargeable
Li-Ion (Lithium Ion) battery pack.
6.3.1 Storage guidelines for the Li-Ion
battery
Never store Li-Ion -version PHD6 instruments at
temperatures above 30 degrees Celsius (86
degrees Fahrenheit). Li-Ion batteries may suffer
deterioration resulting in damage to the internal
components when stored at high temperatures.
The battery may be irretrievably damaged
resulting in reduced battery capacity and voltage.
Honeywell Analytics recommends leaving PHD6
instruments with Li-Ion rechargeable batteries on
the charger when not in use.
6.3.2 Charging guidelines for Li-Ion battery
The Li-Ion battery in the PHD6 should never be
charged at temperatures lower than 5 degrees
Celsius (40 degrees Fahrenheit) or higher than
30 degrees Celsius (86 degrees Fahrenheit.
Charging at temperature extremes can
permanently damage the PHD6 Li-Ion battery.
The PHD6 must be located
in a non-hazardous location during the
charging cycle. Charging the PHD6 in a
hazardous location may impair intrinsic
safety.
6.3.3 Charging procedure for Li-Ion battery
Do not charge the PHD6
with any charger other than the appropriate
PHD6 charger manufactured by Honeywell
Analytics. Standard versions of the PHD6
must be charged with the UL/CSA-approved
charger, which is part number 54-54-001.
European versions of the PHD6 must be
charged with the ATEX-approved PHD6
charger.
1. Verify that the instrument is turned off. (If it is
not, press the MODE button for three
seconds until the message "Release Button"
appears.)
2. Plug the power supply in. The red LED is
labeled “Power” and will be lit whenever the
charger is plugged into a power source.
3. Insert the PHD6 into the charging cradle
bottom side down with the display facing
forward. The green LED on the charger is
labeled “Charge” and will blink while the
battery is charging.
4. When the battery is fully charged, the green
“Charge” LED will be lit and not blinking.
See section 5.3.4 for battery troubleshooting
guidelines.
6.3.4 Charging with the pump attached
The PHD6 with pump attached may be charged
according to the instruction given in section 6.3.3
above.
6.3.5 Battery troubleshooting
If the green “Charge” LED on the charger fails to
light when the PHD6 with Li-Ion battery pack is
placed in the charger, remove the instrument
from the charger and press the MODE button to
attempt to start the instrument.
If the battery has been inserted into the charger
without the instrument, return it to the instrument
prior to attempting the restart.
1. If the PHD6 starts and the battery icon if full,
then the battery is fully charged and may be
used as is. In this case, the charger has
recognized that the battery is charged and
will not charge it any further.
2. If the PHD6 fails to turn on, then the battery
may be severely discharged and should be
returned to the charger. The charger will
then begin a very slow recharge in order to
protect the battery. The green “Charge” LED
may not be lit during the first four hours of the
slow recharge. If the “Charge” LED has still
not been lit after four hours, the battery pack
or charger is probably damaged.
3. If the PHD6 starts and any battery level other
than full is indicated, then either the battery is
damaged or the charger is damaged. Call
Honeywell Analytics for further instructions.
6.4 Sensors
6.4.1 Sensor replacement
The sensors in the PHD6 are located in a vented
compartment at the bottom of the instrument.
To install a sensor:
1. Turn the PHD6 off.
2. Remove the battery pack as described in
section 6.1. This will automatically
disconnect power from the instrument.
28
3. Remove the four screws that are located
below the battery pack insertion from the
back face of the PHD6.
4. Turn the instrument over to reveal the front
face and gently remove the sensor cover.
5. Remove the sensor that is to be replaced.
6. Insert the new sensor into the appropriate
location on the sensor board.
7. Reinstall the sensor cover by aligning it
properly over the sensors and securing it with
the four screws that were removed in step 3.
8. Reattach the battery pack and re-tighten the
top center screw.
9. New sensors must be allowed to stabilize
prior to use according to the following
schedule. The detector must be powered off
and a functional battery pack must be
installed for the sensor to stabilize.
Sensor Stabilization Period
Oxygen (O2)
LEL
PID
NDIR-CH4 or
NDIR-CO
2
All Toxic Sensors
except NO
NO (nitric oxide)
1 hour
none
5 minutes
1 minute
15 minutes
24 hours
Note: Steps 9 and 10 assume that the sensor
stabilization period has passed.
10. Perform the Fresh Air/Zero calibration and
the Gas calibration as discussed in sections
4.2 and 4.3.
6.4.2 Care and maintenance of PID sensors
The two critical
components of a PID
sensor are the
electrode stack and
the lamp. The
electrode stack can be
replaced in the field. The lamp can be cleaned or
replaced in the field. The frequency of
maintenance to both items will vary with the type
of usage and the nature of the contaminants to
which the sensor is exposed.
As a general rule, baseline shifts tend to be
caused by the electrode stack and losses of
sensitivity tend to be caused by the lamp.
6.4.2.1 Troubleshooting the PID
When to replace the electrode stack:
1. Baseline reading climbs following fresh air
zeroing of the sensor.
2. PID sensor becomes sensitive to humidity.
3. Baseline becomes generally unstable.
4. Baseline shifts when the instrument is in
motion.
When to clean the PID lamp
Loss of sensitivity in the sensors as shown
during bump-testing (reading will be low).
When to replace the PID lamp
If the cleaning of the lamp fails to correct a loss
of sensitivity, the lamp should be replaced.
6.4.2.2 Cleaning and replacing PID
components
To remove the lamp and stack
1. Wash your hands thoroughly.
2. On a clean surface, remove the PID sensor
from the PHD6 as described above (section
6.4.1 steps 1-5).
3. Place one finger on top of the sensor and
insert the stack removal tool into the two slots
at the top side of the sensor body. Squeeze
gently until the spring releases and the stack
can be removed from the top of the sensor.
The lamp is spring-loaded against the stack,
so keeping a finger on top of the stack
prevents their ejection from the sensor body.
4. Gently remove the stack and pull the lamp
and spring out of the sensor body. Do not
touch the top of the lamp window with bare
fingers.
5. Set the spring aside.
To replace the stack or lamp
1. Discard the used lamp, stack or both as
needed and rebuild with replacement part(s).
2. Drop the spring into the center of the sensor
body.
3. When reinserting the lamp and electrode
stack, it is essential to make sure that the
lamp is fit snugly into the o-ring slot on the
electrode stack – NOTE PICTURE BELOW.
When inserting the lamp into the o-ring slot, it
is recommended that a twisting motion is
used. When properly assembled, the lamp
should then be flush against the stack, and
should be fully supported.
4. Snap the stack with lamp attached on to the
sensor body so that the sensor is whole
again and the stack cannot be removed
without the removal tool.
5. The sensor should have a gasket and a filter
on it. If necessary, install a sensor filter and
gasket on top of the sensor.
6. Reinstall the sensor into the PHD6.
7. Reassemble the PHD6.
8. Calibrate the PID prior to use after the 5
minute warm up periods ends.
29
To clean the lamp
1. Follow the directions above to remove the
lamp from the instrument.
2. Make sure your hands are clean.
3. Coat the cotton swab in a thin layer of lamp
cleaning powder of 0.1 to 0.25
m -
alumina.
4. Pick up the lamp with the other hand. Do not
touch the top of the lamp window with bare
fingers.
5. Using the cotton swab dipped in the cleaning
powder, polish the top of the lamp with a
swirling motion. Cleaning typically takes
about 30 seconds and is finished when the
swab starts to squeak.
6. Reassemble the sensor and the PHD6. See
steps 2-8 above in the directions to replace
the stack or lamp.
6.5 Sample probe assembly
The PHD6’s sample draw probe is the standard
probe assembly from Honeywell Analytics. The
illustration below gives a breakdown of all parts in
the sample draw probe with part numbers. The
sample probe handle contains moisture barrier
and particulate filters designed to remove
contaminants that might otherwise harm the
instrument.
without the probe assembly in place,
contaminants may cause damage to the
pump, sensors and internal components of
the PHD6.
Particulate contaminants are removed by means
of a cellulose filter. The hydrophobic filter
includes a Teflon barrier which blocks the flow
of moisture as well as any remaining particulate
contaminants.
6.5.1 Changing sample probe filters
The threaded sample probe handle is accesses
the filters. The particulate filter is held in place by
means of a clear filter cup. To replace the
particulate filter, remove the old filter and cup,
insert a new filter into the cup, and slide the cup
back into place in the probe handle. The
hydrophobic barrier filter fits into a socket in the
rear section of the probe handle. (The narrow
end of the hydrophobic barrier filter is inserted
towards the rear of the handle.)
To avoid accidentally introducing particulate
contaminants into the system, turn the sample
probe upside-down prior to removing either the
hydrophobic filter or the particulate filter.
The following replacement filter kits are currently
available from Honeywell Analytics:
Sample probe filters should be replaced
whenever visibly discolored due to
contamination.
CAUTION: Never perform remote sampling
without the sample probe and hose
assembly. The sample probe handle
contains replaceable filters designed to
block moisture and remove particulate
contaminants. If the pump is operated
6.5.2 Changing sample probe tubes (wands)
The standard 11.5” long butyrate probe tube is
held in place with a hex-nut compression fitting
and compression sleeve. The standard probe
tube can be interchanged with other custom
length sections of 1/4” OD tubing, or probe tubes
made of other materials (such as stainless steel).
Probe tubes are exchanged by loosening the
hex-nut compression fitting, removing the old
tube, sliding the compression sleeve into place
around the new tube, inserting the new tube into
the probe handle, then replacing and tightening
the hex-nut.
Note: The sample probe must be checked
for leakage (as discussed in Section 3.1.1)
whenever filters or probe tubes are
exchanged or replaced before being
returned to service.
6.6 PHD6 Pump Maintenance
PHD6 pumps are fairly maintenance free with the
exception of the replacement of the pump filters
on a regular basis.
6.6.1 Replacing pump filters
1. Remove the two screws that hold the inlet
port to the pump.
30
2. Gently pull the dust filter holder free of the
pump.
3. Remove and replace the dust filter that is
located in the holder.
4. The hydrophobic filter is located beneath the
inlet port in the pump housing. Use a small
screwdriver or other object to punch through
the filter and remove it. The gasket that sits
between the inlet port and the filter should
come out with the filter.
5. Place the new hydrophobic filter with the filter
side down in place of the one removed in
step 4. The gasket should be located on top
of the filter and should sit against the dust
filter holder, which will be reinstalled in step
6.
6. Replace the dust filter holder (which now has
a new filter in it) and secure it with the two
screws removed in step 1.
31
Appendices
Appendix A Toxic gas measurement –
Warning, Danger, STEL and TWA
alarms
Many toxic substances are commonly encountered in
industry. The presence of toxic substances may be
due to materials being stored or used, the work being
performed, or may be generated by natural processes.
Exposure to toxic substances can produce disease,
bodily injury, or death in unprotected workers.
It is important to determine the amounts of any toxic
materials potentially present in the workplace. The
amounts of toxic materials potentially present will affect
the procedures and personal protective equipment that
must be used. The safest course of action is to
eliminate or permanently control hazards through
engineering, workplace controls, ventilation, or other
safety procedures. Unprotected workers may not be
exposed to levels of toxic contaminants that exceed
Permissible Exposure Limit (PEL) concentrations.
Ongoing monitoring is necessary to insure that
exposure levels have not changed in a way that
requires the use of different or more rigorous
procedures or equipment.
Airborne toxic substances are typically classified on the
basis of their ability to produce physiological effects on
exposed workers. Toxic substances tend to produce
symptoms in two time frames.
Higher levels of exposure tend to produce immediate
(acute) effects, while lower levels of long-term (chronic)
exposure may not produce physiological symptoms for
years.
Hydrogen sulfide (H
acutely toxic substance which is immediately lethal at
relatively low concentrations. Exposure to a 1,000
ppm (parts per million) concentration of H2S in air
produces rapid paralysis of the respiratory system,
cardiac arrest, and death within minutes.
Carbon monoxide (CO) is a good example of a
chronically toxic gas. Carbon monoxide bonds to the
hemoglobin molecules in red blood cells. Red blood
cells contaminated with CO are unable to transport
oxygen. Although very high concentrations of carbon
monoxide may be acutely toxic, and lead to immediate
respiratory arrest or death, it is the long term
physiological effects due to chronic exposure at lower
levels that take the greatest toll of affected workers.
This is the situation with regards to smokers, parking
garage attendants, or others chronically exposed to
carbon monoxide in the workplace. Exposure levels
are too low to produce immediate symptoms, but small
repeated doses reduce the oxygen carrying capacity of
the blood over time to dangerously low levels. This
partial impairment of the blood supply may lead over
time to serious physiological consequences.
Because prudent monitoring programs must take both
time frames into account, there are two independent
exposure measurements and alarm types built into the
PHD6 design.
S) is a good example of an
2
alarm levels in the PHD6 are less than or equal to the
OSHA-assigned ceiling levels for both CO and H
2
S.
Never enter an environment even momentarily
when concentrations of toxic substances exceed
the level of either the Warning or the Danger Alarm.
Time History Graph
Ceiling
2. Time Weighted Average (TWA)
The maximum average concentration to which an
unprotected worker may be exposed over an eight
hour working day is called the Time Weighted Average
or TWA value. TWA values are calculated by taking
the sum of exposure to a particular toxic gas in the
current operating session in terms of parts-per-millionhours and dividing by an eight-hour period.
Time History Graph
(8 hour)
Ceiling
TWA
3. Short Term Exposure Limits (STEL)
Toxic substances may have short term exposure
limits which are higher than the eight hour TWA. The
STEL is the maximum average concentration to which
an unprotected worker may be exposed in any fifteen
minute interval during the day. During this time,
neither the eight hour TWA or the ceiling
concentration may be exceeded.
Any fifteen minute periods in which the average STEL
concentration exceeds the permissible eight hour
TWA must be separated from each other by at least
one hour. A maximum of four of these periods are
allowed per eight hour shift.
Time History Graph
15 Minutes
Ceiling
STEL
TWA
1. Warning and Danger Alarms
OSHA has assigned some, but not all, toxic
substances with a ceiling level which represents the
highest concentration of a toxic substance to which an
unprotected worker should ever be exposed, even for a
very short time. The default Warning and Danger
32
Appendix B Calibration
Frequency
Recommendation
One of the most common questions
that we are asked at Honeywell
Analytics is: “How often should I
calibrate my gas detector?”
Sensor Reliability and Accuracy
Today’s sensors are designed to
provide years of reliable service. In
fact, many sensors are designed so
that with normal use they will only
lose 5% of their sensitivity per year
or 10% over a two-year period.
Given this, it should be possible to
use a sensor for up to two full years
without significant loss of
sensitivity.
Verification of Accuracy
With so many reasons why a
sensor can lose sensitivity and
given the fact that dependable
sensors can be key to survival in a
hazardous environment, frequent
verification of sensor performance
is paramount.
There is only one sure way to verify
that a sensor can respond to the
gas for which it is designed. That is
to expose it to a known
concentration of target gas and
compare the reading with the
concentration of the gas. This is
referred to as a “bump” test. This
test is very simple and takes only a
few seconds to accomplish. The
safest course of action is to do a
“bump” test prior to each day’s
use. It is not necessary to make a
calibration adjustment if the
readings fall between 90%* and
120% of the expected value. As an
example, if a CO sensor is checked
using a gas concentration of 50
PPM it is not necessary to perform
a calibration unless the readings
are either below 45 PPM or above
60 PPM.
*The Canadian Standards
Association (CSA) requires the
instrument to undergo
calibration when the displayed
value during a bump test fails to
fall between 100% and 120% of
the expected value for the gas.
Lengthening the Intervals
between Verification of Accuracy
We are often asked whether there
are any circumstances in which the
period between accuracy checks
may be lengthened.
Honeywell Analytics is not the only
manufacturer to be asked this
question! One of the professional
organizations to which Honeywell
Analytics belongs is the Industrial
Safety Equipment Association
(ISEA). The “Instrument Products”
group of this organization has been
very active in developing a protocol
to clarify the minimum conditions
under which the interval between
accuracy checks may be
lengthened.
A number of leading gas detection
equipment manufacturers have
participated in the development of
the ISEA guidelines concerning
calibration frequency. Honeywell
Analytics procedures closely follow
these guidelines.
If your operating procedures do not
permit daily checking of the
sensors, Honeywell Analytics
recommends the following
procedure to establish a safe and
prudent accuracy check schedule
for your Honeywell instruments:
1. During a period of initial use of
at least 10 days in the
intended atmosphere, check
the sensor response daily to
be sure there is nothing in the
atmosphere that is poisoning
the sensor(s). The period of
initial use must be of sufficient
duration to ensure that the
sensors are exposed to all
conditions that might have an
adverse effect on the sensors.
2. If these tests demonstrate that
it is not necessary to make
adjustments, the time between
checks may be lengthened.
The interval between accuracy
checking should not exceed 30
days.
3. When the interval has been
extended the toxic and
combustible gas sensors
should be replaced
immediately upon warranty
expiration. This will minimize
the risk of failure during the
interval between sensor
checks.
4. The history of the instrument
response between verifications
should be kept. Any
conditions, incidents,
experiences, or exposure to
contaminants that might have
an adverse effect on the
calibration state of the sensors
should trigger immediate reverification of accuracy before
further use.
5. Any changes in the
environment in which the
instrument is being used, or
33
changes in the work that is
being performed, should
trigger a resumption of daily
checking.
6. If there is any doubt at any
time as to the accuracy of the
sensors, verify the accuracy of
the sensors by exposing them
to known concentration test
gas before further use.
Gas detectors used for the
detection of oxygen deficiencies,
flammable gases and vapors, or
toxic contaminants must be
maintained and operated properly
to do the job they were designed to
do. Always follow the guidelines
provided by the manufacturer for
any gas detection equipment you
use!
If there is any doubt regarding your
gas detector's accuracy, do an
accuracy check! All it takes is a few
moments to verify whether or not
your instruments are safe to use.
One Button Auto Calibration
While it is only necessary to do a
“bump” test to ensure that the
sensors are working properly, all
current gas detectors offer a onebutton auto calibration feature. This
feature allows you to calibrate a
Honeywell gas detector in about
the same time as it takes to
complete a “bump” test. The use of
automatic bump test and calibration
stations can further simplify the
tasks, while automatically
maintaining records.
Don't take a chance
with your life.
Verify accuracy frequently!
Please read also Honeywell
Analytics’ application note:
AN20010808 “Use of ‘equivalent’
calibration gas mixtures”. This
application note provides
procedures to ensure safe
calibration of LEL sensors that are
subject to silicone poisoning.
Honeywell Analytics’ website is
located at:
www.honeywellanalytics.com
Appendix C PHD6 Sensor Information
Part No. Description Range Resolution
54-54-80 LEL Combustible Gas 0 – 100% LEL 1% LEL
54-54-90 O
Oxygen 0 – 30% by Volume 0.1%
2
54-54-01 CO Carbon Monoxide 0 – 1000 PPM 1 PPM
54-54-19 CO-H CO Minus, reduced sensitivity to H2 0 – 1000 PPM 1 PPM
54-54-05
CO+ CO Plus dual purpose CO / H
Provides a non-specific readout for CO and H2S)
(
S
2
CO: 0 – 1000 PPM
H2S: 0 – 200 PPM
1 PPM
54-54-02 H2S Hydrogen Sulfide 0 – 200 PPM 1 PPM
54-54-14
Duo-Tox Dual Channel CO/H
Provides substance specific readouts for CO & H
*The CO2 sensor has an internal resolution of 0.025% but displays readings rounded to the nearest 0.01%. It will, therefore, display steps of
0.03%, 0.05%, 0.08%, 0.10%, etc.
Appendix D Electrochemical Toxic Sensor Cross-Sensitivity
The table below provides the cross-sensitivity response of the PHD6 electrochemical toxic gas sensors to common
interference gases. The values are expressed as a percentage of the primary sensitivity, or the reading of the sensor
when exposed to 100ppm of the interfering gas at 20ºC. These values are approximate. The actual values depend on
the age and condition of the sensor. Sensors should always be calibrated to the primary gas type. Cross-sensitive
gases should not be used as sensor calibration surrogates without the express written consent of Honeywell Analytics.
** Sensor manufacturer rates Cross Sensitivity for (54-54-23) HCN sensor to H2S as follows for 20 PPM exposure at 20C: “Short gas exposure
in minute range; after filter saturation: ca. 40 PPM reading”.
n/d = no data
Honeywell Analytics Warranty Gas Detection Products
General
Honeywell Analytics warrants gas detectors, sensors and accessories manufactured and sold by
Honeywell Analytics - Middletown, to be free from defects in materials and workmanship for the periods
listed in the tables below.
Damages to any Honeywell Analytics products that result from abuse, alteration, power fluctuations
including surges and lightning strikes, incorrect voltage settings, incorrect batteries, or repair procedures
not made in accordance with the Instrument’s Reference Manual are not covered by the Honeywell
Analytics warranty.
The obligation of Honeywell Analytics under this warranty is limited to the repair or replacement of
components deemed by the Honeywell Analytics Instrument Service Department to have been defective
under the scope of this standard warranty. To receive consideration for warranty repair or replacement
procedures, products must be returned with transportation and shipping charges prepaid to Honeywell
Analytics at its manufacturing location in Middletown, Connecticut, or to a Honeywell Analytics Authorized
Warranty Service Center. It is necessary to obtain a return authorization number from Honeywell Analytics
prior to shipment.
THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND
REPRESENTATIONS, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO, THE WARRANTY
OF FITNESS FOR A PARTICULAR PURPOSE. HONEYWELL ANALYTICS WILL NOT BE LIABLE FOR
LOSS OR DAMAGE OF ANY KIND CONNECTED TO THE USE OF ITS PRODUCTS OR FAILURE OF
ITS PRODUCTS TO FUNCTION OR OPERATE PROPERLY.
Instrument & Accessory Warranty Periods
Product(s) Warranty Period
PHD6
ToxiPro®, MultiPro
ToxiLtd®
Battery packs and chargers, sampling pumps and
other components, which by their design are
consumed or depleted during normal operation, or
which may require periodic replacement
As long as the instrument is in service
2 years from date of purchase
2 years after activation or 2 years after the
“Must Be Activated By” date, whichever
comes first
One year from the date of purchase
Sensor Warranty Periods
Instrument(s) Sensor Type(s) Warranty Period
PHD6, Cannonball3, Multi Vision,MultiPro, Toxi
Vision, ToxiPro
®
All Others All Sensors 1 Year
** Damage to combustible gas sensors by acute or chronic exposure to known sensor poisons such as
volatile lead (aviation gasoline additive), hydride gases such as phosphine, and volatile silicone gases
emitted from silicone caulks/sealants, silicone rubber molded products, laboratory glassware greases,
spray lubricants, heat transfer fluids, waxes & polishing compounds (neat or spray aerosols), mold
release agents for plastics injection molding operations, waterproofing formulations, vinyl & leather
preservatives, and hand lotions which may contain ingredients listed as cyclomethicone, dimethicone
and polymethicone (at the discretion of Honeywell Analytics’ Instrument Service department) void
Honeywell Analytics’ Standard Warranty as it applies to the replacement of combustible gas sensors.
O2, LEL**, CO, CO+,
S & Duo-Tox
H
2
2 Years
All Other Sensors 1 Year
35
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.