Sperian Biosystems NXP Reference Manual

Biosystems
NXP
Fixed Gas
Detector Head
Reference
Manual
Sperian Instrumentation
651 South Main Street Middletown, CT 06457 800-711-6776 860-344-1079 Fax 860-344-1068 Part Number 13-345 Version 1.01 10SEP2009 http://www.biosystems.com
2
THE BIOSYSTEMS NXP FIXED GAS DETECTOR HEAD HAS BEEN DESIGNED FOR THE DETECTION AND MEASUREMENT OF POTENTIALLY HAZARDOUS ATMOSPHERIC CONDITIONS.
IN ORDER TO ASSURE THAT THE USER IS PROPERLY WARNED OF POTENTIALLY DANGEROUS ATMOSPHERIC CONDITIONS, IT IS ESSENTIAL THAT THE INSTRUCTIONS IN THIS REFERENCE MANUAL BE READ, FULLY UNDERSTOOD, AND FOLLOWED.
Biosystems NXP
Reference Manual
Sperian Instrumentation Part Number 13-345
Version 1.01
Copyright 2009
by
Sperian Protection Instrumentation, LLC
Middletown, Connecticut 06457
All rights reserved.
No page or part of this operation manual may be reproduced in any form without
written permission of the copyright owner shown above.
Sperian reserves the right to correct typographical errors.
Specifications are subject to change without notice.
1
Table of Contents
Signal Words .......................................................................................................................5
Modification of Sperian Instrumentation Products ...............................................................5
Operating Temperature and Humidity Limits .......................................................................5
Warnings and Cautions .......................................................................................................6
1. Overview.......................................................................................................................7
1.1 Capabilities ..........................................................................................................7
1.2 Method of sampling..............................................................................................7
1.3 Sensor options .....................................................................................................7
1.4 Verification of accuracy........................................................................................7
1.5 Displays and controls...........................................................................................8
1.6 Power requirements.............................................................................................8
1.7 Alarms and Alarm Logic.......................................................................................8
1.7.1 Atmospheric Hazard Alarms ............................................................................8
1.7.2 Sensor Over-range Alarms ..............................................................................9
1.7.3 Alarm Relays ..................................................................................................10
1.8 Fault Conditions .................................................................................................10
1.9 Remote Sensors ................................................................................................10
1.10 4-20 mA Current Loop .......................................................................................11
1.11 Event Logger......................................................................................................11
1.12 Firmware Upgrades............................................................................................11
2. Installation ..................................................................................................................12
2.1 Installation overview...........................................................................................12
2.2 Cautions.............................................................................................................13
2.3 NXP Detector Head Environmental Ratings ......................................................13
2.4 Siting of detector heads .....................................................................................13
2.5 Wiring requirements...........................................................................................13
2.5.1 Grounding ......................................................................................................14
2.6 Output Specifications .........................................................................................14
2.7 External Wiring...................................................................................................14
2.8 Remote Sensor Installation ................................................................................16
2.8.1 Disconnecting the sensor ...............................................................................16
2.8.2 Connecting the sensor cable..........................................................................16
2.8.3 Mounting the remote sensor ..........................................................................16
2.8.4 Connecting the cable (non-explosion-proof sensor) ......................................16
2.8.5 Connecting the cable (explosion-proof sensor)..............................................16
3. Field Usage ................................................................................................................18
2
3.1 Modes of operation ............................................................................................18
3.2 Clearing the MAX reading..................................................................................19
3.3 Effect of contaminants on NXP sensors.............................................................20
3.3.1 Effects of contaminants on oxygen sensors...................................................20
3.3.2 Effects of contaminants on combustible sensors ...........................................20
3.3.3 Effects of contaminants on toxic gas sensors ................................................20
3.4 Sensor Replacement .........................................................................................21
4. Verification of Accuracy (Calibration) .........................................................................22
4.1 Warnings and cautions concerning verification of accuracy ..............................22
4.2 Overview of verification of accuracy...................................................................22
4.2.1 Calibration Interval .........................................................................................22
4.2.2 Special Instructions for Fixed Gas Detectors .................................................22
4.2.3 Accuracy of Oxygen Sensors.........................................................................22
4.2.4 Accuracy of Toxic and LEL Sensors ..............................................................23
4.3 Fresh air calibration ...........................................................................................23
4.4 Span calibration of Toxic and LEL sensors........................................................24
4.5 Test Mode ..........................................................................................................25
4.6 Calibration Failures ............................................................................................26
4.6.1 Fresh Air Calibration Failure ..........................................................................26
4.6.2 Span Calibration Failure.................................................................................26
5. Configuration ..............................................................................................................28
5.1 Accessing the Setup menu and Submenus .......................................................28
5.2 Setup Submenu Flow Chart...............................................................................29
5.3 Set Date and Time .............................................................................................30
5.4 Set Loop Range .................................................................................................30
5.5 Set Alarm Levels................................................................................................30
5.6 Set Alarm Latch .................................................................................................31
5.7 Set Combustible Gas Sensor Display................................................................31
5.8 Set Cal Gas Concentration ................................................................................31
5.9 Set Cal Interval ..................................................................................................31
5.10 Set Bump Test Interval ......................................................................................31
6. Diagnostics .................................................................................................................32
6.1 Accessing the Diagnostics menu and Options...................................................32
6.2 Diagnostic Menu Flow Chart ..............................................................................33
6.3 Serial Number / Software Version......................................................................34
6.4 Sensor Count Display ........................................................................................34
6.5 Sensor Temperature Display .............................................................................34
3
6.6 Current Loop Output Test ..................................................................................34
6.7 Relay Test..........................................................................................................34
6.8 12-Volt Output Test............................................................................................35
6.9 LCD and Backlight Test .....................................................................................35
6.10 LED Test ............................................................................................................35
6.11 Audible Alarm Test.............................................................................................35
7. Service........................................................................................................................36
Returning your NXP to Sperian Instrumentation for service or repair ............................36
Appendices ........................................................................................................................37
Appendix A: NXP Toxic Sensor Cross Sensitivity Data .................................................37
Appendix B: Fault Conditions.........................................................................................38
Appendix C: Calibration Frequency Recommendation ..................................................39
Appendix D: Sperian Instrumentation Warranty Gas Detection Products......................40
4
Signal Words
The following signal words, as defined by ANSI Z535.4-1998, are used in the NXP Reference Manual.
indicates an imminently hazardous situation which, if not
avoided, will result in death or serious injury.
indicates a potentially hazardous situation which, if not
avoided, could result in death or serious injury.
indicates a potentially hazardous situation, which if not avoided,
may result in moderate or minor injury.
CAUTION used without the safety alert symbol indicates a potentially
hazardous situation, which, if not avoided, may result in property damage.
The NXP MUST be installed in a non-hazardous area.
Modification of Sperian Instrumentation Products
Any unauthorized modification of any Sperian Instrumentation product may compromise the certification and the safety of the product, and may lead to serious personal injury or death. Modification of any part or component of a Sperian Instrumentation product requires the express written approval of both the appropriate certification agency and Sperian Instrumentation.
Operating Temperature and Humidity Limits
The Biosystems NXP’s operating temperature range is printed on the label on the inside of the instrument. The Biosystems XPR’s operating temperature range is printed on the housing. Use of Sperian Gas Detection products outside of their specified operating temperature ranges may result in inaccurate and potentially dangerous readings.
5
Warnings and Cautions
1. The Biosystems NXP gas detector has been designed for
the detection of dangerous atmospheric conditions. An alarm condition indicates the presence of a potentially life-threatening hazard and should be taken very seriously.
2.
In the event of an alarm condition it is important to follow established procedures. The safest course of action is to immediately leave the affected area, and to return only after further testing determines that the area is once again safe for entry. Failure to immediately leave the area may result in serious injury or death.
3.
The accuracy of the NXP should be checked periodically with known concentration calibration gas. Failure to check accuracy can lead to inaccurate and potentially dangerous readings.
4.
Fresh air/zero calibrations may only be performed in an atmosphere that is known to contain 20.9% oxygen, 0% LEL and 0 PPM toxic gas. If fresh air is unavailable, a cylinder of “Zero Air” must be used during the fresh air/zero calibration procedure. Calibration of the NXP in a contaminated atmosphere may lead to inaccurate and potentially dangerous readings.
5.
The accuracy of the NXP should be checked immediately following any known exposure to contaminants by testing with known concentration test gas before further use. Failure to check accuracy can lead to inaccurate and potentially dangerous readings.
6.
A sensor that cannot be calibrated or is found to be out of tolerance should be replaced immediately. An instrument that fails calibration may not be used until testing with known concentration test gas determines that accuracy has been restored, and the instrument is once again fit for use.
7.
Do not reset the calibration gas concentration unless you are using a calibration gas concentration that differs from the one that is normally supplied by Sperian for use in calibrating the NXP. Customers are strongly urged to use only Sperian calibration materials when calibrating the NXP. Use of non­standard calibration gas and/or calibration kit components when calibrating the NXP can lead to inaccurate and potentially dangerous readings and may void the standard Sperian warranty.
Sperian offers calibration kits and long-lasting cylinders of test gas specifically developed for easy NXP calibration. Customers are strongly urged to use only Sperian calibration materials when calibrating the NXP.
8.
For safety reasons this equipment must be operated and
serviced by qualified personnel only. Read and understand this reference manual before operating or servicing the NXP.
9.
A rapid up-scale reading followed by a declining or erratic
reading may indicate a hazardous combustible gas concentration that exceeds the XPR’s zero to 100 percent LEL detection range.
6
1. Overview
1.1 Capabilities
The NXP Gas Detector Head is a fixed gas detection system from Sperian Instrumentation. The NXP can function on its own as a stand-alone unit, and can also be used as part of a 4-20 mA current loop system with a PLC or other type of controller.
Standard features include user interface, built-in digital readout to allow for direct calibration at the head, and relays to allow for additional alarms and control.
Each NXP Gas Detector Head includes a sensor housing and gas-specific sensor. Sensor housings are available in non­explosion-proof and remote explosion­proof (XPR) versions. Sensors can be mounted directly to the NXP housing, or can be placed remotely (up to 50 feet away) using a special remote sensor cable.
Note: Conduit, wire and other components necessary to connect the NXP Gas Detector Head to a controller are not included and must be ordered separately.
Installation of explosion-proof housings must be performed in accordance with local regulations.
1.2 Method of sampling
The atmosphere being measured reaches the sensor in the NXP/XPR by diffusing through a protective filter directly into the sensor. Normal air movements are usually enough to carry the sample to the sensor. The sensor reacts to changes in the concentration of the hazard being measured. Values are constantly updated and displayed on the NXP Gas Detector head’s LCD readout. If the head is connected to a controller, the appropriate level 4-20 mA signal is simultaneously sent to the controller.
The NXP Gas Detector head may be installed at a substantial distance from the controller. The maximum distance between the head and the controller is 2000 feet when using the proper cable.
Wiring requirements are discussed in detail in section 2.5.
1.3 Sensor options
NXP Gas Detector Heads can be configured to detect a number of different atmospheric hazards. Each NXP uses a single substance-specific sensor that determines the hazard that the system is configured to detect. The sensors that are currently available, along with their ranges and resolutions, are shown in table 1.3.
Note: Changing the sensor range will require calibration of the instrument.
Sensor Range Resolution
Oxygen (O2)
Combustible
(LEL sensor)
Carbon
Monoxide
(CO)
Carbon
Monoxide
(CO-H)
Hydrogen
Sulfide (H
Hydrogen
Sulfide (H
Sulfur
Dioxide
(SO
)
2
0-25%/Vol. 0-30%/Vol. 0-100% LEL 0-5.0% CH 0-100 PPM 0-500 PPM 0-1000 PPM 0-100 PPM 0-500 PPM 0-1000 PPM
0-10 PPM 0.1 PPM
S)
2
0-50 PPM 0-100 PPM
S)
2
0-250 PPM
0-25 PPM 0-50 PPM
4
0.1% Vol
1% LEL
0.05% CH
1 PPM
1 PPM
1 PPM
0.1 PPM
Table 1.3 – Sensor Ranges
1.4 Verification of accuracy
NXP Gas Detector Heads have been designed for easy verification of accuracy.
The accuracy of the NXP
should be checked periodically with known concentration calibration gas. Failure to check accuracy can lead to inaccurate and potentially dangerous readings.
Please see Sperian’s calibration recommendations in Appendix C.
Verification of detector accuracy is a two­step procedure for toxic and LEL (XPR only) sensor-equipped detectors, and a one-step procedure for oxygen sensor­equipped detectors.
Verification normally begins by exposing the NXP’s sensor to known “fresh air” and
7
4
checking the readings. If the readings differ from those expected in fresh air (20.9% oxygen, 0 PPM toxic gas, 0% LEL combustible gas), a "fresh air zero" adjustment must be made. For most fixed applications, where the freshness of the ambient air is in question, a cylinder of “Zero Air” must be used during the fresh air/zero calibration.
The second step, which only applies to toxic and LEL sensors, is to verify sensor accuracy by exposing it to known concentration test gas and noting the response. Toxic and LEL readings are considered accurate if the readings are between 90% and 120% of the expected value as given on the gas cylinder. If the reading is accurate, then the instrument requires no further adjustment. Toxic and LEL readings that fall outside of this range are considered inaccurate and indicate that the sensor must be calibrated before further use.
Calibration procedures are discussed in detail in Chapter 4.
1.5 Displays and controls
NXP Gas Detector Heads include a 3-digit LCD located on the face of the detector housing. The heads also include a MODE button that allows the user to initiate and control the calibration of the detector.
1.6 Power requirements
Power to the NXP can be provided either from a controller or via a standard wall cube available from Sperian.
The NXP requires a power supply of 13-30 VDC. Power is applied to connector J1 (see Figure 2.5). Once powered up and running, the NXP is intended for continuous monitoring. If it is necessary to power down the NXP, power should be removed from the wall cube or controller that is supplying power to the NXP.
1.7 Alarms and Alarm Logic NXP Gas Detector
Heads have been designed for the detection of dangerous atmospheric conditions. An alarm condition indicates the presence of a potentially life-threatening hazard and should be taken very seriously.
In the event of an alarm condition it is important to follow established procedures. The safest course of action is to immediately leave the affected area, and return only after further testing together with other appropriate safety procedures determine that the area is once again safe for entry.
1.7.1 Atmospheric Hazard Alarms
NXP Fixed Gas Detectors with toxic or combustible gas sensors include two levels of alarms. One alarm serves as the warning alarm; the second alarm serves as the danger alarm. Detectors with oxygen sensors have four alarm levels, as described below.
Figure 1.5 – Main NXP Components
Toxic and combustible gas sensors have “ascending” alarms, meaning that the alarms are activated when gas readings rise above a pre-set alarm threshold.
Oxygen sensors have both “ascending” and “descending” alarms, meaning that the alarms activate when gas readings rise above a pre-set alarm threshold or fall below a pre-set alarm threshold. There is a warning alarm and a danger alarm in each direction.
Table 1.7 lists the available alarm levels and span gas values for each sensor type. Custom alarm levels and calibration gas values may be set using BioTrak II
8
software. Please see the BioTrak II operation manual for instructions. The alarm levels apply to both warning and danger alarms, with the following restrictions:
Alarms can not be set to a value above the full-scale range. If a range
is selected that is below current alarm levels, the alarms will be adjusted to default values for that range.
The warning alarm can not be set beyond the danger alarm setting.
Sensor
Available Alarm Levels
Type
(Process) High: OFF, 1.0, 2.0, 2.5, 5%
Oxygen (O2)
Combustible
(LEL display)
Combustible
display)
(CH
4
Carbon
monoxide
(CO, CO-H)
Hydrogen
sulfide (H
Sulfur Dioxide
(SO
S)
2
)
2
(Safety) High: 22.0, 22.5, 23.0, 23.5%
(Safety) Low: 20.0, 19.5, 19.0, 18.5, 18.0%
OFF, 0.25, 0.5, 0.75, 1.0, 1.25, 2.5% CH
OFF, 10, 25, 35, 50, 100, 150, 200 PPM
(Process) Low: OFF
OFF, 5, 10, 15, 20, 25, 50% LEL
OFF, 2, 5, 10, 15, 20 PPM
OFF, 2, 5, 10, 15, 20 PPM
Table 1.7 – Available alarm levels and calibration span gas settings
When an alarm set point is exceeded the LEDs on the front panel and in the strobe housing will flash (Yellow for Warning, Red for Danger), and the audible alarm will sound. Any external devices controlled by means of the NXP’s alarm relay contacts will also be activated.
Latching Alarms
NXP alarms are normally self-resetting. When readings drop back below the pre­set alarm levels, the instrument returns to normal operation, and the alarms and relay contacts are turned off.
It is possible to set NXP’s alarms so that they "latch". During latched operation, once an alarm condition occurs, the visual and audible alarms and relays will remain active even after the atmospheric hazard has cleared. They must be manually reset using the MODE button. See section 5.6 for information on setting latching alarms.
Available
Span Gas
Settings
N/A
25, 30, 40, 50,
60% LEL
4
1.25, 1.5, 2.0, 2.5,
3.0% CH
35, 50, 75, 100,
150, 200 PPM
10, 20, 25, 30
PPM
10, 20, 25, 30
PPM
4
Alarm Acknowledgement
During an alarm condition, the audible alarm and the strobe LEDs can be acknowledged (temporarily deactivated) by pressing the MODE button. The LEDs on the display board and the alarm relays will remain active. Once acknowledged, the audible alarm and strobe LEDs will be reactivated under the following conditions:
Gas level remains above the alarm level for 5 minutes
Gas level decreases below the alarm level, and then rises above it again
Warning alarm was acknowledged, and the gas level then rises above the Danger alarm level
1.7.2 Sensor Over-range Alarms
If the concentration of the gas being detected goes above the selected sensor range, a sensor over-range occurs.
9
The instrument will act as if a danger alarm condition exists, even if the alarms have been disabled.
Oxygen and Toxic Gas Over-range
When the gas concentration falls back within the sensor’s range, the over-range alarm will be reset, unless latching alarms have been selected.
Combustible Gas Over-range
In the case of an LEL over-range condition (LEL over 100%), power to the LEL sensor will be disconnected to protect the sensor from overheating and premature failure. The instrument will remain in alarm with the sensor powered down until the MODE button is pressed or the instrument’s power is cycled.
If the MODE button is pressed during this time, power to the sensor will be turned on briefly. If the gas level is below 100% LEL, the sensor will remain on and the instrument will resume gas detection mode, along with any applicable alarms. If the gas level is still over 100% LEL, the sensor will be shut off again and the instrument will remain in over-range alarm.
1.7.3 Alarm Relays
The NXP incorporates relays that are activated during alarm conditions. There are separate relays for Warning and Danger alarms, as well as a Fault condition relay (see section 1.8).
The Warning and Danger relays are activated any time that the gas level exceeds the current alarm levels. Each relay has a normally-open and a normally­closed contact. Normally-open contacts close (make contact) during an alarm condition, while normally-closed contacts open (break contact).
These contacts can be used to operate external alarms or larger relays that drive other equipment, such as ventilation systems. See Figure 2.5 for relay connection locations.
The relay contacts in the NXP do not supply power to the external accessories. They act as switch contacts only. External alarms, relays and contactors require their own power supplies.
The maximum rating of the NXP’s relay contacts is 30 VDC at 5 amps.
Under no circumstances should AC line voltage be connected directly to the relay contacts.
1.8 Fault Conditions
The NXP is capable of detecting certain internal fault conditions. During one of these conditions the loop current will fall to 2 mA, the Fault relay will be activated, and the red LEDs will flash every 5 seconds to alert the user to the problem. Fault codes are listed below. See Appendix B for explanations of fault codes.
F1 - Sensor Fault
F2 – EEPROM Read Fault
F5 – EEPROM Write Fault
F11 – Power Low Fault
F12 – Power High Fault
Power Failure Fault
The Fault relay is normally energized. This is called Fail-Safe mode. In the case of a total loss of power to the instrument, the Fault relay will trip, and any external devices connected to it will be activated. Loop current will fall to 0 mA.
Note: The Fault relay does not supply power to the device(s) connected to it. It acts as a switch contact only. External power must be supplied to any devices connected to this relay.
Under no circumstances should AC line voltage be connected directly to the relay contacts.
1.9 Remote Sensors
The NXP normally comes with the sensor housing attached to the instrument. When necessary, the sensor can be installed remotely, up to 50 feet from the instrument.
10
The standard sensor can be used for non­hazardous (non-explosive) environments.
For use in hazardous locations, the XPR explosion-proof sensor housing must be used. The XPR is built into a heavy-duty stainless steel explosion-proof housing. See section 2.8 for remote sensor installation.
The XPR remote
sensor housing must not
be opened in a hazardous location unless power to the NXP has been disconnected. Power must not be restored to the NXP until the XPR housing has been completely closed.
Installation of explosion-proof housings must be performed in accordance with local regulations.
1.10 4-20 mA Current Loop
The NXP incorporates a connection for an industry-standard 4-20 mA current loop. This circuit generates a current based on the gas level detected. The NXP is normally set to “Source” mode, meaning that it generates a current output. It can also be set to “Sink” mode if needed, meaning that it sinks (inputs) the current from an external source.
A 4 mA signal indicates a reading of zero, and a 20 mA signal indicates a full-scale reading. Full-scale is determined by the Loop Range setting in the Setup menu. See Table 1.3 for available ranges and section 5.4 for instructions on setting the Loop Range. See Figure 2.5 for wiring of the loop signal.
1.11 Event Logger
The event logger in the NXP stores data associated with alarm conditions. Each event (alarm) includes the following data:
Event Start time
Event End time
Duration of the event.
Minimum reading during event
Maximum reading during event
Average reading during event
Temperature at end of event
Sensor type
Flags for ‘In calibration’, warning alarm,
danger alarm and temperature alarm during event
Time event was acknowledged
The NXP stores the data from the 128 most recent alarm events. Once 128 events have been stored, the NXP will begin to overwrite the data from the oldest event in memory with data from new events. One event may be a combination of different alarms occurring simultaneously or in immediate succession.
The event logger may be downloaded using BioTrak II software. The PC must be equipped with an IrDA port to provide a connection.
While the NXP is communicating through its IrDA connection, it IS NOT
sensing the gas level of the atmosphere, and WILL NOT activate any alarms. The area must be made safe or another form of atmospheric monitoring must be used during IrDA communication.
1.12 Firmware Upgrades
As improvements are made and features are added to the NXP’s firmware, new versions of the firmware will be made available to customers. Normally, this firmware can be downloaded and installed by the customer using a computer with IrDA communication capability and the FX2 Flash Upgrade utility software.
Instrument firmware and upgrade utility software can be downloaded from
http://www.biodownloads.com
in the
Download Section.
11
Loading...
+ 29 hidden pages