Spellman uX50P50, uX65P65, uX50P75 Instruction Manual

Page 1
R
Instruction Manual
uX SERIES
SPELLMAN HIGH VOLTAGE ELECTRONICS CORPORATION
475 Wireless Blvd. Hauppauge, New York, 11788
+1(631) 630-3000*FAX: +1(631) 435-1620* E-mail: sales@spellmanhv.com Website: www.spellmanhv.com
High Voltage Power Supply
uX MANUAL 118152-001 Rev A
Page 2
The uX Series is the result of Spellman’s exceptional high voltage packaging and surface mount fabrication techniques coupled with proprietary encapsulation technology producing this ultra-compact X-Ray generator module. The uX powers grounded cathode X-Ray tubes from a variety of well-known manufacturers, featuring a 0 to 50kV/65kV high voltage output @ 2mA limited to 50, 65 or 75 Watts. The uX uses closed loop filament control circuitry providing highly regulated beam current. The low noise dc filament supply operates between 0.3 and 3.5 amps. Offering tight regulation, high stability and low ripple, the uX provides users local and remote analog control to set beam voltage, emission current and filament current limit. USB, RS-232 and Ethernet interface is standard.
TYPICAL APPLICATIONS
Powering grounded cathode X-Ray tubes from Kevex, Oxford, RTW, Superior, Varian and Trufocus.
OPTIONS
XCC XRM Compatible HV Cable (50kV only) 5VPM 0 to 5 Volt Programming and Monitor Scaling
GB Grid Bias Option K5302 Mammoflex HV cable for uX K2001 Mammoflex HV cable for uX w/XCC option
SPECIFICATIONS
Input:
+24Vdc ±10%, 5.0A maximum for either 50 Watts or 75 Watts. +24Vdc ±1V, 5.0A maximum for 65kV/65W units.
Efficiency:
75%, typical
Output:
0 to 50kV at 0 to 2mA, limited to a maximum of 50 wattsr or 75 Watts. 0-65kV at 2mA limited to 65 Watts.
Voltage Control:
Local: Internal multi-turn potentiometer to set voltage
from 0 to full output voltage.
Remote: 0 to +10Vdc proportional from 0 to full output voltage.
Accuracy: ±1%. Z
IN
: 10Mohm.
Emission Control:
Local: Internal potentiometer to set beam current
between 0 and full output current.
Remote: 0 to +10Vdc proportional from 0 to full output current.
Accuracy: ±1%. Z
IN
: 10Mohm. Filament limit and
filament preheat control capability is also provided.
DC Filament Supply:
Islolated filament power supply generates emission current feedback signal for accurate low X-Ray tube current performance.
Current: 3.5A, adjustable limit Voltage: 5.0 volt limit
Environmental:
Operational: 0°C to +50°C Storage: -40°C to +85°C Humidity: 0% to 90%, non-condensing
Temperature Coefficient:
0.01% per °C, voltage and current.
Stability:
0.05% per 8 hours after 1/2 hour warm-up.
Voltage and Current Monitors:
0 to +10Vdc proportional from 0 to rated output. Accuracy ±1%.
Redundant Voltage Monitor:
A redundant high voltage feedback divider with proportional 0 to +10Vdc = 0 to 100% output voltage signal can be provided on a custom basis.
Dimensions:
50kV Unit: 4.00”H x 2.87”W x 8.00”D
(101.6mm x 72.95mm x 202.20mm).
65kV Unit: 4.00”H x 2.87”W x 9.00”D
(101.6mm x 72.95mm x 228.60mm).
XCC Option: 4.00”H x 2.87”W x 9.00”D
(101.6mm x 72.95mm x 228.60mm).
Weight:
4.5 lbs. (2.1kg) typical
Regulatory Approvals:
Compliant to 2004/108/EC, the EMC Directive and 2006/95/EC, the Low Voltage Directive (approval pending). UL/CUL recognized, File E227588 (approval pending). Compliant to 2002/95/EC, RoHS.
50kV at 2 mA. 50 or 75 Watt Max.
65kV at 2 mA. 65 Watt Max.
Adjustable Isolated Filament Supply
Overvoltage & Short Circuit Protection
Voltage & Current Programming
Local and Remote Emission Control
Safety Interlock
RS-232, Ethernet, & USB Standard
Redundant HV Monitor Signal Available
OEM Customization Available
SPELLMAN HIGH VOLTAGE ELECTRONICS CORPORATION
uX
50W/65W/75W X-RAY GENERATOR
PAGE 1 OF 5
30% Smaller
than MNX
Corporate Headquarters
Hauppauge, New York USA
+1-631-630-3000 FAX: +1-631-435-1620
e-mail: sales@spellmanhv.com
www.spellmanhv.com
128108-001 REV. A
Spellman High Voltage is an ISO 9001:2008 and ISO 14001:2004 registered company
Copyright © 2014 Spellman High Voltage Electronics Corp.
Page 3
PAGE 2 OF 5
SPELLMAN HIGH VOLTAGE ELECTRONICS CORPORATION
uX
50W/65W/75W X-RAY GENERATOR
Grid Bias Option (GB):
Plug-n-Play compatibility for Oxford’s Apogee X-Ray Tube Spellman’s Grid Bias Option for the uX Series is specifically designed for popular commercially available grid bias X-Ray tubes. The Grid Bias voltage is developed via the use of sepa­rate integrated high frequency switching circuit, providing maxi­mum flexibility and control. The Grid Bias output is a voltage regulated, current compliant topology ideally suited for Wehnelt electrode applications. Arc and short circuit protection of the Grid Bias output prevents any damage due to transient events or installation errors.
Tracking Mode Operation
Functioning in tracking mode the voltage monitor (0-10Vdc = 0 to 50kV) of the main high voltage output is internally connected to the Grid Bias programming input (0-10Vdc = 0 to -300Vdc of Grid Bias). Connected in this manner the Grid Bias output will track in a linearly proportional fashion the setting of the main kV output.
A multiturn potentiometer limits the maximum magnitude of Grid Bias output applied to the X-Ray tube, providing unparalleled flexibility.
The output of the Grid Bias option is provided via an auxiliary two position Phoenix Contact terminal block, the mating connecter is provided.
GRID BIAS SPECIFICATIONS
Output Voltage: 0 to -300Vdc Output Current: 0.25mA, maximum Load Regulation: 1% of output voltage, no load to full load Line Regulation: 1% for a ±10% change in input voltage Ripple: 1% of maximum rated voltage
uX with Grid Bias Option.
Shown with Oxford Apogee X-Ray Tube (not included)
Note: Units ordered with the GB Option will be provided with the XCC Option for proper high voltage cable compatibility.
Main Control Screen
Communication Screen
Filament Status Screen
Digital Interface
The uX features a standard USB, RS-232 and Ethernet digital interface. Utilizing these standard digital interfaces can dramatically simplify power supply interfacing requirements saving the user both time and money, while enhancing func­tionality and overall capabiity. Spellman provides a GUI with the uX that allows the customer to both customize operational features of the uX while also providing basic power supply operational features. Details of the uX’s digital interface capability are described in detail in the uX manual.
Closeup showing digital
interface connectors
Corporate Headquarters
Hauppauge, New York USA
+1-631-630-3000 FAX: +1-631-435-1620
e-mail: sales@spellmanhv.com
www.spellmanhv.com
128108-001 REV. A
Spellman High Voltage is an ISO 9001:2008 and ISO 14001:2004 registered company
Copyright © 2014 Spellman High Voltage Electronics Corp.
Page 4
PAGE 3 OF 5
SPELLMAN HIGH VOLTAGE ELECTRONICS CORPORATION
uX
50W/65W/75W X-RAY GENERATOR
ANALOG INTERFACE CONNECTOR MALE 15 PIN MINI “D”
PIN SIGNAL PARAMETER
1 Monitor Return Signal Ground
2 Voltage Monitor 0-10 volts = 0 to full scale, Zout=1KΩ
3 Current Monitor 0-10 volts = 0 to full scale, Zout=1KΩ
4 Interlock Output Connect 12V HVON bulb to pin 15 to enable
5 +10 Volt Reference +10 Volts at 1mA, maximum
6 Filament Monitor 1 volt = 1 amp, Zout=1KΩ
7 Voltage Program Input 0-10 volts = 0 to full scale, Zin=10MΩ
8 Local Voltage Program* 0-10 volts, screwdriver adjust
9 Filament Limit Setpoint* 1 volt = 1 amp, screwdriver adjust
10 Current Program Input 0-10 volts = 0 to full scale, Zin=10MΩ
11 Local Current Program* 10 turn pot, screwdriver adjust
12 Not used (+24V Out for Interlock) (Optional Interlock configuration)
13 Not used (Interlock Coil) (Optional Interlock configuration)
14 Filament Preheat Setpoint* 1 volt = 1 amp, screwdriver adjust
15 Interlock Return Interlock Ground
POWER INPUT/FILAMENT CONNECTOR 4 PIN PHOENIX CONTACT
PIN SIGNAL PARAMETER
1 +24V Input +24 volts @ 5A, max. 2 24V Return (Gnd.) Power Ground 3 Filament Out 0.3A to 3.5A, 5 volt, max. 4 Filament Return Filament Return
HIGH VOLTAGE OUTPUT CONNECTOR
Spellman drywell type detachable connector. Standard: A one meter (39.4”) long polyethylene mating high
voltage cable is provided.
K5302: A one meter (39.4”) long Mammoflex mating high voltage
cable is provided, SHV p/n 201946-007
K2001: A one meter (39.4”) long Mammoflex mating high voltage cable
is provided, compatible with the XCC Option SHV p/n 201946-002
GRID BIAS CONNECTOR 2 PIN PHOENIX CONTACT
PIN SIGNAL PARAMETER
1 Ground Chassis Ground 2 Grid Bias 0 to -300Vdc
How To Order:
Sample model number:
50 Watt unit: uX50P50 65 Watt unit: uX65P65 75 Watt unit: uX50P75
Options are added to the model number as follows:
uX50P50/XCC or uX50P75/GB
ETHERNET DIGITAL INTERFACE— 8 PIN RJ45 CONNECTOR
PIN SIGNAL SIGNAL PARAMETERS
1 TX+ Transmit Data + 2 TX- Transmit Data ­3 RX+ Receive Data + 4 NC No Connection 5 NC No Connection 6 RX- Receive Data ­7 NC No Connection 8 NC No Connection
RS-232 DIGITAL INTERFACE— 9 PIN FEMALE D CONNECTOR
PIN SIGNAL SIGNAL PARAMETERS
1 NC No Connection 2 TX out Transmit Data 3 RX in Receive Data 4 NC No Connection 5 SGND Ground 6 NC No Connection 7 NC No Connection 8 NC No Connection 9 NC No Connection
USB DIGITAL INTERFACE— 4 PIN USB “B” CONNECTOR
PIN SIGNAL SIGNAL PARAMETERS
1 VBUS +5 Vdc 2 D- Data ­3
D+ Data +
4 GND Ground
*Denotes 10 turn potentiometer accessable through holes in cover
Corporate Headquarters
Hauppauge, New York USA
+1-631-630-3000 FAX: +1-631-435-1620
e-mail: sales@spellmanhv.com
www.spellmanhv.com
128108-001 REV. A
Spellman High Voltage is an ISO 9001:2008 and ISO 14001:2004 registered company
Copyright © 2014 Spellman High Voltage Electronics Corp.
Note: The filament return wire cannot be grounded as this would short circuit the tube return current monitoring to the uX. If grounding of the filament is required, please consult the factory.
Page 5
PAGE 4 OF 5
DIMENSIONS: in.[mm]
SPELLMAN HIGH VOLTAGE ELECTRONICS CORPORATION
uX
50W/65W/75W X-RAY GENERATOR
50KV
50KV WITH XCC OPTION
Corporate Headquarters
Hauppauge, New York USA
+1-631-630-3000 FAX: +1-631-435-1620
e-mail: sales@spellmanhv.com
www.spellmanhv.com
128108-001 REV. A
Spellman High Voltage is an ISO 9001:2008 and ISO 14001:2004 registered company
Copyright © 2014 Spellman High Voltage Electronics Corp.
Page 6
PAGE 5 OF 5
DIMENSIONS: in.[mm]DIMENSIONS: in.[mm]
SPELLMAN HIGH VOLTAGE ELECTRONICS CORPORATION
uX
50W/65W/75W X-RAY GENERATOR
65KV
Corporate Headquarters
Hauppauge, New York USA
+1-631-630-3000 FAX: +1-631-435-1620
e-mail: sales@spellmanhv.com
www.spellmanhv.com
128108-001 REV. A
Spellman High Voltage is an ISO 9001:2008 and ISO 14001:2004 registered company
Copyright © 2014 Spellman High Voltage Electronics Corp.
DANGER
VOLTAGE
HIGH
Spellman
J1
0-50KV
Output
See Wiring Diagrams for Recommended Analog Interface Connections
uX Power Supply
HIGH VOLTAGE ELECTRONICS CORP.
Typical uX Operating Setup
4 PIN
1
2
FIL I LIM ADJ
KV ADJ
mA ADJ
POWER ON
I/O
PH ADJ
HV ON
FIL RET
FIL OUT
GND
+24V
DANGER
HIGH
VOLTAGE
+24V Input Connections
+24V @ 4A
+24V Input Return
+24V @ 4A Input Power
ANODE
X-RAY TUBE
FILAMENT
+
FIL. P.S.
5.0V/3.5A
-
+10.00V
+10.00V
VOLTAGE PROGRAM IN
+10.00V
LOCAL VOLTAGE CONTROL
25K, 20T
CURRENT PROGRAM IN
+10.00V
LOCAL CURRENT CONTROL
25K, 20T
uX Power Supply
+24VDC
J1
+24VDC
+24V, Power To HV Circuits
1K
1K
1K
10K
10K
+10.00V
14V
Filament Connections
3
4
1
2
3
6
9
14
5
15 PIN
VOLTAGE MONITOR
CURRENT MONITOR
FILAMENT MONITOR
FILAMENT LIMIT SETPOINT
FILAMENT PREHEAT SETPOINT
+10.00V REFERENCE
X-Ray Tube Filament
Analog Interface
-
M
+
-
M
+
7
Jumper J4-7 to J4-8
8
to Use Local Voltage Control
10
Jumper J4-10 to J4-11
11
12
to Use Local Current Control
15 PIN
13
12V LAMP
EXTERNAL
INTERLOCK
.5W-2W
4
15
Fail Safe Interlock / HV ON Connections
-
M
+
-
M
+
-
M
+
-
EXTERNAL VOLTAGE
V
CONTROL
+
-
EXTERNAL CURRENT
V
CONTROL
+
Alternate Interlock Configurations
uX Power Supply
+24VDC
+24VDC
J1
14V
12
13
15 pin
Fail Safe Interlock/HV ON:
Remove internal jumper J1 to isolate the interlock relay coil + from the internal +24V. Connect Pin 13 to an external +24V source or to Pin 12 to re-connect internal +24V.
4
+24V, Power to HV Circuits
Alternate Interlock Configuration: Fail Safe Lamp Replaced With a 270 Ohm Resistor
270, 1W
15
uX Power Supply
+24VDC
+24VDC
+24V, Power to HV Circuits
Alternate Interlock Configuration: Customer Provided +12V For HV ON Relay, Relay Return Grounded
15 pin
12
J1
13
14V
EXTERNAL INTERLOCK
4
15
+12V, User Provided
Page 7
IMPORTANT SAFETY PRECAUTIONS
SAFETY
THIS POWER SUPPLY GENERATES VOLTAGES THAT ARE DANGEROUS AND MAY BE FATAL.
OBSERVE EXTREME CAUTION WHEN WORKING WITH THIS EQUIPMENT.
High voltage power supplies must always be grounded.
Do not touch connections unless the equipment is off and the
Capacitance of both the load and power supply is discharged.
Allow five minutes for discharge of internal capacitance of the power supply.
Do not ground yourself or work under wet or damp conditions.
SERVICING SAFETY
.
Maintenance may require removing the instrument cover with the power on.
Servicing should be done by qualified personnel aware of the electrical hazards.
WARNING note in the text call attention to hazards in operation of these units
that could lead to possible injury or death.
CAUTION notes in the text indicate procedures to be followed to avoid possible
damage to equipment.
Copyright © 2000, Spellman High Voltage Electronics Corporation. All Rights Reserved.
This information contained in this publication is derived in part from proprietary and patent data. This information has
been prepared for the express purpose of assisting operating and maintenance personnel in the efficient use of the
model described herein, and publication of this information does not convey any right to reproduce it or to use it for
any purpose other than in connection with installation, operation, and maintenance of the equipment described.
118004-001 REV. B
Page 8
WICHTIGE SICHERHEITSHINWEISE
SICHERHEIT
DIESES HOCHSPANNUNGSNETZTEIL ERZEUGT LEBENSGEFÄHRLICHE HOCHSPANNUNG.
SEIN SIE SEHR VORSICHTIG BEI DER ARBEIT MIT DIESEM GERÄT.
Das Hochspannungsnetzteil muß immer geerdet sein.
Berühren Sie die Stecker des Netzteiles nur, wenn das Gerät ausgeschaltet ist und die elektrischen
Kapazitäten des Netzteiles und der angeschlossenen Last entladen sind.
Die internen Kapazitäten des Hochspannungsnetzteiles benötigen ca. 5 Minuten, um sich zu entladen.
Erden Sie sich nicht, und arbeiten Sie nicht in feuchter oder nasser Umgebung.
SERVICESICHERHEIT
Notwendige Reparaturen können es erforderlich machen, den Gehäusedeckel während des Betriebes zu
entfernen.
Reparaturen dürfen nur von qualifiziertem, eingewiesenem Personal ausgeführt werden.
“WARNING” im folgenden Text weist auf gefährliche Operationen hin, die zu Verletzungen oder zum Tod
führen können.
“CAUTION” im folgenden Text weist auf Prozeduren hin, die genauestens befolgt werden müssen, um
eventuelle Beschädigungen des Gerätes zu vermeiden.
118004-001 REV. B
Page 9
PRECAUTIONS IMPORTANTES POUR VOTRE SECURITE
CONSIGNES DE SÉCURITÉ
CETTE ALIMENTATION GÉNÈRE DES TENSIONS QUI SONT DANGEUREUSES ET PEUVENT ÊTRE FATALES.
OYEZ EXTRÊMENT VIGILANTS LORSQUE VOUS UTILISEZ CET ÉQUIPEMENT.
S
Les alimentations haute tension doivent toujours être mises à la masse.
Ne touchez pas les connectiques sans que l’équipement soit éteint et que la capacité à la fois de la charge et de
l’alimentation soient déchargées.
Prévoyez 5 minutes pour la décharge de la capacité interne de l’alimentation.
Ne vous mettez pas à la masse, ou ne travaillez pas sous conditions mouillées ou humides.
CONSIGNES DE SÉCURITÉ EN CAS DE REPARATION
La maintenance peut nécessiter l’enlèvement du couvercle lorsque l’alimentation est encore allumée.
Les réparations doivent être effectuées par une personne qualifiée et connaissant les risques électriques.
Dans le manuel, les notes marquées « WARNING » attire l’attention sur les risques lors de la manipulation de ces
équipements, qui peuvent entrainer de possibles blessures voire la mort.
Dans le manuel, les notes marquées « CAUTION » indiquent les procédures qui doivent être suivies afin d’éviter
d’éventuels dommages sur l’équipement.
118004-001 REV. B
Page 10
IMPORTANTI PRECAUZIONI DI SICUREZZA
SICUREZZA
QUESTO ALIMENTATORE GENERA TENSIONI CHE SONO PERICOLOSE E
POTREBBERO ESSERE MORTALI.
PONI ESTREMA CAUTELA QUANDO OPERI CON QUESO APPARECCHIO.
Gli alimentatori ad alta tensione devono sempre essere collegati ad un impianto di terra.
Non toccare le connessioni a meno che l’apparecchio sia stato spento e la capacità interna
del carico e dell’alimentatore stesso siano scariche.
Attendere cinque minuti per permettere la scarica della capacità interna dell’alimentatore
ad alta tensione.
Non mettere a terra il proprio corpo oppure operare in ambienti bagnati o saturi d’umidità.
SICUREZZA NELLA MANUTENZIONE.
Manutenzione potrebbe essere richiesta, rimuovendo la copertura con apparecchio
acceso.
La manutenzione deve essere svolta da personale qualificato, coscio dei rischi elettrici.
Attenzione alle AVVERTENZE contenute nel manuale, che richiamano all’attenzione ai
rischi quando si opera con tali unità e che potrebbero causare possibili ferite o morte.
Le note di CAUTELA contenute nel manuale, indicano le procedure da seguire per evitare
possibili danni all’apparecchio.
118004-001 REV. B
Page 11
Table of Contents
PAGE
1. INTRODUCTION
1.1 Description of the uX Series ................................................................................1
1.2 Standard Features .................................................................................................2
1.2.1 Remote Operating Features .....................................................................2
1.3 Options .................................................................................................................2
1.4 Interpreting the Model Number ...........................................................................2
2. INSPECTION & INSTALLATION
2.1 Initial Inspection ..................................................................................................3
2.2 Mechanical Installation ........................................................................................3
3. OPERATING INSTRUCTIONS
3.1 Operation..............................................................................................................6
3.2 Standard Features .................................................................................................7
4. PRINCIPLES OF OPERATION
4.1 Chassis .................................................................................................................12
4.2 Inverter .................................................................................................................12
4.3 High Voltage Transformer ...................................................................................13
4.4 High Voltage Assembly .......................................................................................13
4.5 Control PWB ........................................................................................................13
4.6 Filament Supply ...................................................................................................13
4.7 Options .................................................................................................................13
5. OPTIONS
5.1 XCC Option .........................................................................................................14
5.2 5VPM ...................................................................................................................14
5.3 GB ........................................................................................................................14
5.4 K5302 ...................................................................................................................14
5.5 K2001 ...................................................................................................................14
563 Custom Designed Models ....................................................................................14
6. MAINTENANCE
6.1 Periodic Servicing ................................................................................................15
6.2 Performance Test .................................................................................................15
6.3 High Voltage Dividers .........................................................................................15
7. REPLACEMENT PARTS
7.1 Replacement Parts ................................................................................................16
7.2 Correspondence and Ordering Spare Parts ..........................................................16
8. FACTORY SERVICE
8.1 Warranty Repairs .................................................................................................17
8.2 Factory Service Procedures ..................................................................................17
uX MANUAL i 118151-001 REV A
Page 12
8.3 Ordering Options and Modifications ...................................................................17
8.4 Shipping Instructions ...........................................................................................17
APPENDIX
A. Specification Controls (Custom Models Only)
LIST OF FIGURES
Figure 2.1 Dimensions 50kV uX .....................................................................................4
Figure 2.2 Dimensions 65kV uX .....................................................................................5
Figure 3.1 Typical Operating Setup uX ..........................................................................8
Figure 3.2 DC Input & Filament Connections ................................................................9
Figure 3.3 Monitors & Voltage/Current Control Connections........................................10
Figure 3.4 Recommended Interlock / HV on Connections .............................................11
LIST OF TABLES
Table 1.1 Options ...........................................................................................................2
uX MANUAL ii 118151-001 REV A
Page 13
Chapter 1
INTRODUCTION
1.1 Description of the uX Series
he uX series of high voltage power supplies represent an advanced approach to X-ray generator power
T
requirements. These power supplies provide all the power, control, and support functions required for practically all X-ray applications. The uX series provide high voltage, high current outputs with very low ripple. Extremely stable voltage and current outputs result in significant performance improvements over previously available technology. Low output ripple provides higher intensity levels, with no increase in tube loading.
All these advancements are possible only by Spellman’s long history in X-ray power systems. This series of power supplies utilize extremely advanced resonant conversion techniques, along with sophisticated digital technology.
The uX series is specifically designed for X-ray tube application where the high voltage is a positive polarity, and the filament circuits are referenced to the cathode ground potential, (grounded filament).
The X-ray tube voltage and tube emission current are all continuously adjustable.
The power supplies operate from +24Vdc and are convection cooled for 0 to 75W models. Custom designed units for single use or OEM applications are available.
uX MANUAL 1 118151-001 REV A
Page 14
1.2 Standard Features
y
r
r
The uX series incorporates several standard features designed to optimize user satisfaction and safety.
KV AND MA/FILAMENT CURRENT RAMP CIRCUITS: This feature provides for a gradual rise for
kV, mA and filament current. This feature is designed to limit voltage shock and filament shock to the X-ray tube. The kV ramp rate is approximately 4 seconds. The filament current is slowly increased until the desired mA level is achieved. This time is typically 4 seconds for full mA output. These ramp conditions are started at the initial INTERLOCK CLOSED control signal. Prior to closing the Interlock the filament operates at a user determined preheat current level. Preheat levels are selected for the desired X-ray tube to minimize mA overshoot.
INDICATOR LEDS: HIGH VOLTAGE ON and
POWER ON indicators.
OUTPUT CABLE: Standard units are provided with a 1
meter shielded high voltage output cable. The cables are designed with a plug arrangement so that they can be easily removed from the mating receptacle located on the front of the chassis. For non-standard units, see Spec. Control drawing.
LOCAL AND REMOTE PROGRAMMING:
Potentiometers accessible through the top cover are provided for controlling tube voltage, tube emission current, tube filament limit set point and tube filament preheat set point.
If fail-safe interlocking of the X-ray On is not required the lamp can be replaced with a 270 Ohm, 1 W resistor. Provisions are provided for isolating the interlock relay coil from the internal +24V, allowing the user to connect the coil to an external +12V source. This eliminates the need for either the lamp or the 270-Ohm resistor allowing pin 4 to be grounded for HIGH VOLTAGE ON control. See Figure 3.4 for alternate Interlock Configurations.
1.4 Options
CODE DISCRIPTION
XCC XRM Compatible HV Cable (50kV only)
5VPM 0 to 5 Volt Programming and Monitor
Scaling
GB Grid Bias Option.
K5302 Mammoflex HV cable for uX
K2001 Mammoflex HV cable for uX w/XCC
option
Table 1.1 uX Options
The options available are listed in Table 1.1. See Section 5 for more information on these options along with operating and set-up instructions. With few exceptions, these options and modifications can be retrofitted to your power supply at the factory in a short time. For price and retrofit arrangements, contact Spellman’s Sales Department.
Tube voltage and tube emission current can also be controlled remotely via the Analog Interface connector (J4).
1.2.1 Remote Operating Features
REMOTE MONITOR: Allows remote monitoring of
the tube voltage, tube emission current, tube filament current, tube filament limit set point and tube filament preheat set point via the Analog Interface connector.
EXTERNAL INTERLOCK: Interlock connections are
provided on the Analog Interface connector on the front of the chassis for connection to a safety switch. The unit will not operate unless the interlock circuit is closed. The recommended configuration is to close the interlock circuit through a 12V lamp rated for 0.5W to 0.8W and a safety switch. This configuration provides fail safe interlocking. During high voltage operation, opening the safety switch or failure of the 12V lamp will cause the High Voltage to shut OFF. This option should be used for safety interlock circuits.
1.4 Interpreting the Model Number:
The model number of the power supply describes its capabilities. After the series name is:
(1) the maximum voltage (in kV)
(2) the maximum output (in watts)
(3) the option codes for all options that are included.
Custom units have an X number after the option codes.
uX 30 P 50 /X CC/ X(#)
eries
ame
Maximum Voltage
Polarit
Maximum Powe
Option
Custom
"X" Numbe
uX MANUAL 2 118151-001 REV A
Page 15
Chapter 2
Inspection and Installation
nitial inspection and preliminary checkout procedures are recommended. For safe
I
operation, please follow the step-by-step procedures described in Chapter 3, Operating Instructions.
2.1 Initial Inspection
Inspect the package exterior for evidence of damage due to handling in transit. Notify the carrier and Spellman immediately if damage is evident. Do not destroy or remove any of the packing material used in a damaged shipment.
After unpacking, inspect the panel and chassis for visible damage.
Standard Spellman uX X-Ray generators are covered by warranty. Custom and special order models (with an X suffix in the model number) are also covered by warranty.
2.2 Mechanical Installation
The uX Series of modular X-ray generators are designed for installation into existing or newly developed OEM equipment. The X-ray generator can also easily fit into bench top applications or test set requirements. Standard unit dimensions are shown in Figure 2.1.
For custom mounting requirements or specific package size requirements consult Spellman’s Sales Department. Spellman has many package designs available, or can design a specific enclosure for your requirements.
The uX Series utilizes solid encapsulation for corona free operation. No periodic maintenance is required.
uX MANUAL 3 118151-001 REV A
Page 16
uX MANUAL 4 118151-001 REV A
Figure 2.1 50kV uX DIMENSIONS
Page 17
uX MANUAL 5 118151-001 REV A
Figure 2.2 65kVuX DIMENSIONS
Page 18
Chapter 3
Operating Instructions
3.1 Operation
WARNING
THIS EQUIPMENT GENERATES
DANGEROUS VOLTAGES THAT MAY BE
FATAL. PROPER GROUNDING OF ALL HIGH
VOLTAGE EQUIPMENT IS ESSENTIAL.
WARNUNG
DIESES GERÄT ERZEUGT
LEBENSGEFÄHRLICHE HOCHSPANNUNG.
ALLE HOCHSPANNUNGSGERÄTE MÜSSEN
ÜBER EINE GEEIGNETE ERDUNG
VERFÜGEN.
IMPORTANT
Before connecting the power supply to the
AC line, follow this step-by-step procedure.
Do not connect the power to the X-ray
generator until Step G is reached.
Failure to follow these procedures may void
the warranty.
ACHTUNG
BEVOR SIE DAS
HOCHSPANNUNGSNETZEIL AN DIE
STROMVERSORGUNG ANSCHLIESSEN,
MÜSSEN FOLGENDE PUNKTE GEPRÜFT
WERDEN.
SCHLIESSEN SIE DAS
HOCHSPANNUNGSNETZEIL NICHT AN DIE
SPANNUNGSVERSORGUNG BEVOR PUNKT
G ERREICHT IST. EVENTUELLE
AUFTRETENDE BESCHÄDIGUNG DES
GERÄTES DURCH NICHT BEFOLGEN
DIESER ANWEISEN KANN ZUM VERLUST
DES GARANTIEANSPRUCHES FÜHREN.
A) Check the input voltage rating on the nameplate
of the supply and make certain that this is the rating of the available power source. Spellman uX units operate on 24Vdc unless ordered with a different input voltage.
B) PROPER GROUNDING TECHNIQUE: The
chassis of high voltage power supplies must be grounded, preferably to a water system ground using copper pipe or other earth ground using the connection terminal at the rear of the unit. See Figure 3.1, for a typical operating setup.
The return line from the load should be connected to the terminal on the rear of the power supply. Using a separate external ground at the load is not recommended.
C) Attach the output cable to the load. D) Plug the high voltage output cable into the front
of the supply and hand tighten the knurled collar.
E) Options Note: See section 5 for hook up and
operating instructions for the options on your unit. Custom models may also require set up changes.
F) For initial turn-on, set the programming voltage
to the zero voltage position.
G) The input power cable may now be connected
and power applied. The POWER ON LED should light up. No high voltage will be generated at this time.
H) Close the INTERLOCK. The HIGH VOLTAGE
ON LED should light up and the output will slow start to the preset level output voltage and/or output current.
NOTE: The uX series is equipped with a slow start circuit that ramps the output up to its maximum setting in approximately 4 seconds after the INTERLOCK is closed.
I) To terminate the generation of output power,
open the INTERLOCK. In the HIGH VOLTAGE OFF mode the power supply’s fault and interface circuits are still active and the filament operates at the Preheat level.
J) To turn off the power supply, disconnect the
power.
uX MANUAL 6 118151-001 REV A
Page 19
WARNING
AFTER TURNOFF, DO NOT HANDLE
THE LOAD UNTIL THE CAPACITANCE
HAS BEEN DISCHARGED!
LOAD CAPACITANCE MAY BE DISCHARGED BY
SHORTING TO GROUND.
WARNUNG
Nach drm Ausschalten des Gerätes die Last erst
berühren wenn diese vollständig entladen ist.
Die elektrische Kapazität der Last kann durch
einen Kurzschluß zur Erde entladen werden.
WARNING
THE VOLTAGE MONITOR DOES NOT READ THE
OUTPUT VOLTAGE WHEN THE POWER IS
TURNED OFF, EVEN IF A CHARGE STILL EXISTS
ON THE LOAD.
WARNUNG
Der Spannungsmonitor arbeitet nicht bei
abgeschalteter Versorgungspannung, auch nicht
wenn die Last noch aufgeladen ist.
CAUTION
ALWAYS OPERATE THE UNIT WITH THE COVER
ON. DO NOT ATTEMPT TO ACCESS OR REPAIR
ANY INTERNAL CIRCUITS. DANGEROUS AND LETHAL VOLTAGES ARE GENERATED INSIDE
THE MODULE.
CAUTION
Betreiben Sie das Hochspannungsnetzteil
ausschließlich mit geschlossenem Gehäuse.
Versuchen Sie nicht die internen Schaltkreise zu
berühren oder zu reparieren, da
lebensgefährliche Hochspannungen in Innern
erzeugt werden.
3.2 Standard Features
A note on remote interface circuitry and remote signal grounding. Whenever possible, electrical isolation should be provided when interfacing with any high voltage power supply. For power control signals such as
EXTERNAL INTERLOCK, HIGH VOLTAGE OFF and HIGH VOLTAGE ON isolated relay contacts should be used. If possible, analog programming and monitoring signals should be isolated via analog isolation amplifiers. Spellman application engineers are available to assist in interface circuitry design. All interface cables should be properly shielded. All power supply signals should be referenced to the power supplies signal ground.
LOCAL AND REMOTE PROGRAMMING: Allows
the operator to select local or remote adjustment of the output voltage and current. Adjustments are made using the screwdriver adjustable potentiometers accessible through the top cover of the power supply or via an external voltage source provided by the operator. In local control jumpers are installed on the 15 pin interface connector between pin 7 and pin 8 for voltage control and between pin 10 and pin 11 for current control.
For remote programming, the jumpers are removed and a positive voltage source, from 0 to 10 volts, is applied to the appropriate terminals. Programming signals should be referenced to signal ground. By adjusting the voltage source from 0 volts (zero output) to 10 volts (full rated output) the desired output can be selected. 0 to 5 volts equals 0-100% of output on units with 5VPM option. See Figure 3.4 for wiring diagram and specifications.
An alternate method of controlling the output remotely is by using external resistance such as a potentiometer or a resistor network. For remote control the jumpers are removed and the desired resistor configuration is installed. See Figure 3.4 for wiring diagram.
REMOTE MONITOR: Test points are made available
on the 15 pin D connector for monitoring the voltage, current and filament outputs and for reading the filament current limit and preheat set points. The test points are always positive regardless of the output polarity, where 0 to 10 volts equals 0-100% of output on standard units. 0 to 5 volts equals 0-100% of output on units with 5VPM option. Test points have an output impedance of 1K ohms. See Figure 3.3 for test point designation.
EXTERNAL INTERLOCK: Interlock connections are
provided on the 15 pin D connector for connection to a safety switch. The unit will not operate unless the interlock circuit is closed. During high voltage operation, opening the interlock circuit will cause the unit to revert to the HIGH VOLTAGE OFF mode. See Figure 3.3 for the recommended interface circuit.
uX MANUAL 7 118151-001 REV A
Page 20
FILAMENT LIMIT: The maximum current that the
filament power supply can operate is user adjustable from
1.0A to 3.5A. The Limit adjustment is made when the power supply is in the high voltage OFF mode (Interlock Open) via the screwdriver adjustable potentiometer labeled “FIL I LIM ADJ” that is accessible through the top cover.
A test point is provided on pin 9 on the 15 pin D connector for monitoring the Filament Limit set point. The test point scaling is 1VDC = 1amp. See Figure 3.3 for the recommended interface circuit.
the +24VDC input power is applied to the unit and the Interlock is Open (High Voltage OFF). The Preheat Level is adjustable from between .8A to 2.5A.
Selecting the correct preheat current level can greatly reduce or eliminate the overshoot that typically occurs on the output emission current when operating into a “cold filament”.
The Preheat adjustment is made when the power supply is in the high voltage OFF mode (Interlock Open) via the screwdriver adjustable potentiometer labeled “PH ADJ” that is accessible through the top cover.
Due to the wide variety of X-ray tubes available, uX power supplies are shipped with the Filament Limit set for minimum. The operator must set filament limit at the time of installation in accordance with the X-ray tube manufacturers recommendations.
FILAMENT PREHEAT: uX filament power supplies
operate at a user selectable preheat current level whenever
DANGER
DANGER
FOR CONTINUED PROTECTION
AGAINST RISK OF FIRE
REPLACE ONLY WITH SAME
TYPE AND RATING FUSE.
A test point is provided on pin 14 on the 15 pin D connector for monitoring the Filament Preheat set point. The test point scaling is 1VDC = 1ADC. See Figure 3.3 for the recommended interface circuit.
Due to the wide variety of X-ray tubes available, uX power supplies are shipped with the Filament Preheat set for minimum. The operator must set filament Preheat at
the time of installation
FIL I LIM ADJ
kV ADJ
mA ADJ
HV ON
POWER ON
FIL RET
FIL OUT
GND
+24V
I/O
.
ANODE
X-RAY TUBE
FILAMENT
Figure 3.1 Typical Operating Setup uX
uX MANUAL 8 118151-001 REV A
Page 21
Table 3.2 DC Input & Filament Connections
uX MANUAL 9 118151-001 REV A
Page 22
Figure 3.3 Monitors & Voltage/Current Control Connections
uX MANUAL 10 118151-001 REV A
Page 23
Figure 3.4 Recommended Interlock/HV on Configurations
uX MANUAL 11 118151-001 REV A
Page 24
Chapter 4
Principles of Operation
he uX series of modular X-ray generators utilizes sophisticated power conversion technology. A
T
variety of analog, digital and power conversion techniques are used throughout. The intention of the Principles of Operation is to introduce the basic function blocks that comprise the uX power supply. For details on a specific circuit, consult Spellman’s Engineering Department.
The uX power supply is basically a DC-to-DC converter. Within the power supply, conversions of DC to high frequency AC, then to high voltage DC takes place. By reviewing further the sub-assemblies, a basic understanding of the process can be gained.
WARNING
To reduce the risk of fire, replace fuse with same type and
rating.
WARNING
Um die Brandgefahr zu verringern, muss die Sicherung
durch eine neue gleichen Typs ersetzt werden.
4.1 Chassis
The uX is a compact, high efficiency, X-ray generator. The power supply can supply up to 75 watts of DC power. (Output power capability may be higher or lower depending upon model ordered). Output voltages of up to 65kV can be generated.
WARNING
The energy levels used and generated by the
power supply can be lethal! Do not attempt to
operate the power supply unless the user has a
sufficient knowledge of the dangers and hazards
of working with high voltage. Do not attempt to
approach or touch any internal or external
circuits or components that are connected or
have been connected to the power supply. Be
certain to discharge any stored energy that may
be present before and after the power supply is
used. Consult IEEE recommended practices for
safety in high voltage testing #510-1983.
WARNUNG
Die verwendete und erzeugte Engergie des
Hochspannungsnetzteiles kann tötlich sein! Betreiben Sie
das Hochspannungsnetzeil nur wenn Sie ausreichendes
Kenntnis über mögliche Gefahren beim Ungang mit der
Hochspannung haben. Versuchen Sie nicht interne
Schaltkreise oder elektrische Bauelemente, die an das
Hochspannungsnetzteil angeschlossen sind, zu berühren.
Vergewissern Sie sich, daß alle Restladungen, die vor
oder nacj dem Betreib des Hochspannungsnetzteils
vrhanden sind, entladen werden. Weitere Informationen
finden Sie der IEEE-Vorschrift Nr. 510-1983.
4.2 Inverter
The inverter is a series resonant, parallel loaded topology. A PWM scheme is used for regulating the power generated from the inverter. Q1 is a high speed MOSFET. This device provides high frequency switching to control the resonant current flow. The typical resonant operating period is approximately 15μ seconds.
uX MANUAL 12 118151-001 REV A
Page 25
4.3 High Voltage Transformer
The output of the High Frequency Resonant Inverter is connected to the primary of the High Voltage Transformer. The High Voltage Transformer is a step-up type. Typically secondary voltages are in the range of 5kV depending upon output voltage ratings.
4.4 High Voltage Assembly
The High Voltage Assembly will vary depending upon the model ordered. The circuitry typically consists of series arrangements of a half wave voltage multiplier. The higher voltage ranges utilize various series arrangements of a voltage doubler.
Output filtering is typically provided by an R-C type filter. A high bandwidth resistive/capacitive divider provides voltage feedback for regulation and monitoring. Current feedback for regulation and monitoring is provided by a sense resistor connected at the low voltage end of the High Voltage Rectifier/Multiplier Circuit.
4.5 Control PWB
The majority of control circuits for power supply controls are located on the CONTROL/POWER PWB.
+15VDC, -15VDC, and +10VDC are generated on the CONTROL/POWER PWB. High Voltage On/Off control is accomplished by K2, and its associated circuitry. Interlock control is provided by K2.
Voltage feedback from a high voltage divider, located on the High Voltage Assembly, is sent to U16. Gain adjustment is provided by R98. The KV feedback signal is sent to the 15 pin D connector for remote monitoring.
Program voltages are typically ramped up to set level by the slow start circuits of U16.
Current feedback from the high voltage rectifier is sent to sense resistors located on the High Voltage Assembly. Feedback is then sent to U15.
The resonant control circuitry consists of a voltage to pulse width converter. U17 generates all pulse width control signals.
4.6 Filament Supply
The filament inverter provides the power for the X-ray tube filament. The filament inverter is a high frequency inverter. The inverter provides regulated current to the primary of the filament transformer. The filament isolation transformer secondary is then connected to the output connector.
4.7 Options
Due to the many variations of models and options provided in the uX series, details of actual circuits used may differ slightly from above descriptions. Consult Spellman’s Engineering Department for questions regarding the principles of operations for the uX series.
uX MANUAL 13 118151-001 REV A
Page 26
Chapter 5
OPTIONS
He options available for this power supply are
described in this section. Interface diagrams are
T
shown where required. Options are specified by including the option code in the model number as described in Section 1.5.
5.1 XRM Compatible Cable XCC
The XCC option allows the uX power supply to except the same high voltage output cable that is used on the XRM series power supply.
5.2 5 Volt Programming and Monitor 5VPM
5VPM changes the voltage and current programming inputs and monitors to 0-5V = 0-full rated output.
5.3 Grid Bias Option GB
Plug-n-play compatibility for Oxfords Apogee X-ray tubes that utilize grid bias control.
5.4 Mammoflex HV Cable K5302
Mammoflex HV cable for uX.
5.5 Mammoflex HV Cable K2001
Mammoflex HV cable for uX with XCC option.
5.6 Custom Designed Models X(#)
Units built to customer specifications are assigned an X number by the factory. If this unit is an X model, a specification control sheet is added at the end of this instruction manual.
Spellman welcomes the opportunity to tailor units to fit your requirements or to develop new products for your applications. Contact Spellman Sales Department with your needs.
uX MANUAL 14 118151-001 REV A
Page 27
Chapter 6
MAINTENANCE
his section describes periodic servicing and performance testing procedures.
T
WARNING
THIS POWER SUPPLY GENERATES VOLTAGES
THAT ARE DANGEROUS AND MAY BE FATAL.
OBSERVE EXTREME CAUTION WHEN WORKING
WITH HIGH VOLTAGE.
WARNUNG
DIESES HOCHSPANNUNGSNETZTEIL ERZEUGT
LEBENSGEFÄHRLICHE HOCHSPANNUNG!
SEIEN SIE SEHR VORSICHTIG BEIM ARBEITEN
MIT HOCHSPANNUNG!
6.1 Periodic Servicing
No periodic servicing is required on this module.
6.2 Performance Test
WARNING
HIGH VOLTAGE IS DANGEROUS.
ONLY QUALIFIED PERSONNEL SHOULD
PERFORM THESE TESTS.
WARNING
HOCHSPANNUNG IST GEFÄHRLICH!
NUR QUALIFIZIERTES PERSONAL DARF DIESE
PRÜFUNGEN DURCHFÜHREN.
High voltage test procedures are described in Bulletin STP-783, Standard Test Procedures for High Voltage Power Supplies. Copies can be obtained from the Spellman Customer Service Department. Test equipment, including an oscilloscope, a high impedance voltmeter, and a high voltage divider such as the Spellman HVD-100 or HVD-200, is needed for performance tests. All test components must be rated for operating voltage.
6.3 High Voltage Dividers
High voltage dividers for precise measurements of output voltage with an accuracy up to 0.1% are available from Spellman. The HVD-100 is used for voltages up to 100KV. The HVD-200 measures up to 200KV. The Spellman divider is designed for use with differential voltmeters or high impedance digital voltmeters. The high input impedance is ideal for measuring high voltage low current sources, which would be overloaded by traditional lower impedance dividers.
uX MANUAL 15 118151-001 REV A
Page 28
Chapter 7
REPLACEMENT PARTS
7.1 Replacement Parts
otact the Spellman Customer Service Department for parts lists for specific models.
C
Spellman provides parts and subassemblies for its high voltage power supplies but recommends that only qualified personnel perform the repair. High voltage is dangerous; even minor mistakes in repairs can have serious consequences.
When requesting parts please give the model number and serial number of the power supply
7.2 Correspondence and
Ordering Spare Parts
Each Spellman power supply has an identification label on the rear of the chassis that bears its model and serial number.
When requesting engineering or applications information, please state the model and serial number of the power supply. If specific components or circuit sections are involved in the inquiry, it is helpful to indicate the component symbol number(s) shown on the applicable schematic diagram.
When ordering spare parts, please specify the part’s description, the part’s reference designation or part number, and the model and serial number of the unit.
uX MANUAL 16 118151-001 REV A
Page 29
Chapter 8
FACTORY SERVICE
8.1 Warranty Repairs
During the Warranty period, Spellman will repair all units free of charge. The Warranty is void if the unit is worked on by other than Spellman personnel. See the Warranty in the rear of this manual for more information. Follow the return procedures described in Section 8.2. The customer shall pay for shipping to and from Spellman.
8.2 Factory Service Procedures
Spellman has a well-equipped factory repair department. If a unit is returned to the factory for calibration or repair, a detailed description of the specific problem should be attached.
For all units returned for repair, please obtain an authorization to ship from the Customer Service Department, either by phone or mail prior to shipping. When you call, please state the model and serial numbers, which are on the plate on the rear of the power supply, and the purchase order number for the repair. A Return Material Authorization Code Number (RMA Number) is needed for all returns. This RMA Number should be marked clearly on the outside of the shipping container. Packages received without an RMA Number will be returned to the customer. The Customer shall pay for shipping to and from Spellman.
A preliminary estimate for repairs will be given by phone by Customer Service. A purchase order for this amount is requested upon issuance of the RMA Number. A more detailed estimate will be made when the power supply is received at the Spellman Repair Center. In the event that repair work is extensive, Spellman will call to seek additional authorization from your company before completing the repairs.
8.3 Ordering Options and
Modifications
Many of the options listed in Chapter 5 can be retrofitted into Spellman power supplies by our factory. For prices and arrangements, contact our Sales Department.
8.4 Shipping Instructions
All power supplies returned to Spellman must be sent shipping prepaid. Pack the units carefully and securely in a suitable container, preferably in the original container, if available. The power supply should be surrounded by at least four inches of shock absorbing material. Please return all associated materials, i.e. high voltage output cables, interconnection cables, etc., so that we can examine and test the entire system.
All correspondence and phone calls should be directed to:
Spellman High Voltage Electronics Corp.
475 Wireless Boulevard Hauppauge, New York 11788 TEL: (631) 630-3000 FAX: (631) 435-1620 E-Mail: sales@Spellmanhv.com http://www.spellmanhv.com
uX MANUAL 17 118XXX-001 REV A
Page 30
SPELLMAN HIGH VOLTAGE ELECTRONICS
WARRANTY
Spellman High Voltage Electronics (“Spellman”) warrants that all power supplies it manufactures will be
free from defects in materials and factory workmanship, and agrees to repair or replace, without charge, any power supply that under normal use, operating conditions and maintenance reveals during the warranty period a defect in materials or factory workmanship. The warranty period is twelve (12) months from the date of shipment of the power supply. With respect to standard SL power supplies (not customized) the warranty period is thirty-six (36) months from the date of shipment of the power supply.
This warranty does not apply to any power supply that has been:
Disassembled, altered, tampered, repaired or worked on by persons unauthorized by Spellman;
subjected to misuse, negligent handling, or accident not caused by the power supply;
installed, connected, adjusted, or used other than in accordance with the original intended application and/or
instructions furnished by Spellman.
THE FOREGOING WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING THOSE OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
The buyer’s sole remedy for a claimed breach of this warranty, and Spellman’s sole liability is limited, at
Spellman’s discretion, to a refund of the purchase price or the repair or replacement of the power supply at Spellman’s cost. The buyer will be responsible for shipping charges to and from Spellman’s plant. The
buyer will not be entitled to make claim for, or recover, any anticipatory profits, or incidental, special or consequential damages resulting from, or in any way relating to, an alleged breach of this warranty.
No modification, amendment, supplement, addition, or other variation of this warranty will be binding unless it is set forth in a written instrument signed by an authorized officer of Spellman.
For an authorization to ship contact Spellman’s Customer Service Department. Please state the model and serial numbers, which are on the plate on the rear panel of the power supply and the reason for return. A Return Material Authorization Code Number (RMA number) is needed from Spellman for all returns. The RMA number should be marked clearly on the outside of the shipping container. Packages received without an RMA Number may delay return of the product. The buyer shall pay shipping costs to and from Spellman. Customer Service will provide the Standard Cost for out-of-warranty repairs. A purchase order for this amount is requested upon issuance of the RMA Number (in-warranty returns must also be accompanied by a “zero-value” purchase order). A more detailed estimate may be made when the power supply is received at Spellman. In the event that the cost of the actual repair exceeds the estimate, Spellman will contact the customer to authorize the repair.
Spellman will warrant for three (3) months or balance of product warranty, whichever is longer, the repaired assembly/part/unit. If the same problem shall occur within this warranty period Spellman shall undertake all the work to rectify the problem with no charge and/or cost to the buyer. Should the cause of the problem be proven to have a source different from the one that has caused the previous problem and/or negligence of the buyer, Spellman will be entitled to be paid for the repair.
For a complete listing of Spellman’s Global Service facilities please go to: http://www.spellmanhv.com/customerservice/service.asp
Factory Service Procedures
Factory Service Warranty
Spellman Worldwide Service Centers
101520-007 REV D