Spellman PMX Instruction Manual

R
PMX
Instruction Manual
MODEL : SERIAL# : DATE :
SPELLMAN HIGH VOLTAGE ELECTRONICS CORPORATION
475 Wireless Blvd. Hauppauge, New York, 11788
+1(631) 630-3000*FAX: +1(631) 435-1620* E-mail: sales@spellmanhv.com Website: www.spellmanhv.com
PMX MANUAL 118163-001 Rev A
IMPORTANT SAFETY PRECAUTIONS
SAFETY
THIS POWER SUPPLY GENERATES VOLTAGES THAT ARE DANGEROUS AND MAY BE FATAL.
OBSERVE EXTREME CAUTION WHEN WORKING WITH THIS EQUIPMENT.
High voltage power supplies must always be grounded.
Do not touch connections unless the equipment is off and the
Capacitance of both the load and power supply is discharged.
Allow five minutes for discharge of internal capacitance of the power supply.
Do not ground yourself or work under wet or damp conditions.
SERVICING SAFETY
.
Maintenance may require removing the instrument cover with the power on.
Servicing should be done by qualified personnel aware of the electrical hazards.
WARNING note in the text call attention to hazards in operation of these units
that could lead to possible injury or death.
CAUTION notes in the text indicate procedures to be followed to avoid possible
damage to equipment.
Copyright © 2000, Spellman High Voltage Electronics Corporation. All Rights Reserved.
This information contained in this publication is derived in part from proprietary and patent data. This information has
been prepared for the express purpose of assisting operating and maintenance personnel in the efficient use of the
model described herein, and publication of this information does not convey any right to reproduce it or to use it for
any purpose other than in connection with installation, operation, and maintenance of the equipment described.
118004-001 REV. B
WICHTIGE SICHERHEITSHINWEISE
SICHERHEIT
DIESES HOCHSPANNUNGSNETZTEIL ERZEUGT LEBENSGEFÄHRLICHE HOCHSPANNUNG.
SEIN SIE SEHR VORSICHTIG BEI DER ARBEIT MIT DIESEM GERÄT.
Das Hochspannungsnetzteil muß immer geerdet sein.
Berühren Sie die Stecker des Netzteiles nur, wenn das Gerät ausgeschaltet ist und die elektrischen
Kapazitäten des Netzteiles und der angeschlossenen Last entladen sind.
Die internen Kapazitäten des Hochspannungsnetzteiles benötigen ca. 5 Minuten, um sich zu entladen.
Erden Sie sich nicht, und arbeiten Sie nicht in feuchter oder nasser Umgebung.
SERVICESICHERHEIT
Notwendige Reparaturen können es erforderlich machen, den Gehäusedeckel während des Betriebes zu
entfernen.
Reparaturen dürfen nur von qualifiziertem, eingewiesenem Personal ausgeführt werden.
“WARNING” im folgenden Text weist auf gefährliche Operationen hin, die zu Verletzungen oder zum Tod
führen können.
“CAUTION” im folgenden Text weist auf Prozeduren hin, die genauestens befolgt werden müssen, um
eventuelle Beschädigungen des Gerätes zu vermeiden.
118004-001 REV. B
PRECAUTIONS IMPORTANTES POUR VOTRE SECURITE
CONSIGNES DE SÉCURITÉ
CETTE ALIMENTATION GÉNÈRE DES TENSIONS QUI SONT DANGEUREUSES ET PEUVENT ÊTRE FATALES.
OYEZ EXTRÊMENT VIGILANTS LORSQUE VOUS UTILISEZ CET ÉQUIPEMENT.
S
Les alimentations haute tension doivent toujours être mises à la masse.
Ne touchez pas les connectiques sans que l’équipement soit éteint et que la capacité à la fois de la charge et de
l’alimentation soient déchargées.
Prévoyez 5 minutes pour la décharge de la capacité interne de l’alimentation.
Ne vous mettez pas à la masse, ou ne travaillez pas sous conditions mouillées ou humides.
CONSIGNES DE SÉCURITÉ EN CAS DE REPARATION
La maintenance peut nécessiter l’enlèvement du couvercle lorsque l’alimentation est encore allumée.
Les réparations doivent être effectuées par une personne qualifiée et connaissant les risques électriques.
Dans le manuel, les notes marquées « WARNING » attire l’attention sur les risques lors de la manipulation de ces
équipements, qui peuvent entrainer de possibles blessures voire la mort.
Dans le manuel, les notes marquées « CAUTION » indiquent les procédures qui doivent être suivies afin d’éviter
d’éventuels dommages sur l’équipement.
118004-001 REV. B
IMPORTANTI PRECAUZIONI DI SICUREZZA
SICUREZZA
QUESTO ALIMENTATORE GENERA TENSIONI CHE SONO PERICOLOSE E
POTREBBERO ESSERE MORTALI.
PONI ESTREMA CAUTELA QUANDO OPERI CON QUESO APPARECCHIO.
Gli alimentatori ad alta tensione devono sempre essere collegati ad un impianto di terra.
Non toccare le connessioni a meno che l’apparecchio sia stato spento e la capacità interna
del carico e dell’alimentatore stesso siano scariche.
Attendere cinque minuti per permettere la scarica della capacità interna dell’alimentatore
ad alta tensione.
Non mettere a terra il proprio corpo oppure operare in ambienti bagnati o saturi d’umidità.
SICUREZZA NELLA MANUTENZIONE.
Manutenzione potrebbe essere richiesta, rimuovendo la copertura con apparecchio
acceso.
La manutenzione deve essere svolta da personale qualificato, coscio dei rischi elettrici.
Attenzione alle AVVERTENZE contenute nel manuale, che richiamano all’attenzione ai
rischi quando si opera con tali unità e che potrebbero causare possibili ferite o morte.
Le note di CAUTELA contenute nel manuale, indicano le procedure da seguire per evitare
possibili danni all’apparecchio.
118004-001 REV. B
Contents
CHAPTER 1 INTRODUCTION ....................................................2
1.1 PMX DESCRIPTION...............................................2
1.1.1 Power Supply Requirements ................................3
1.1.2 Environment Requirements ..................................4
1.2 THEORY OF OPERATION ....................................4
1.2.1 Function Overview ...............................................4
1.2.2 Line Rectification and Filterin g ...........................4
1.2.3 Inverter .................................................................4
1.2.4 High Voltage Transformer ...................................4
1.2.5 High Voltage Assembly ........................................4
1.2.6 System Control PWB ............................................4
1.2.7 Filament Power Supply ........................................5
1.2.8 High Speed Starter ...............................................6
1.3 SAFETY ...................................................................7
1.3.1 Safety and Warning Symbols ................................7
CHAPTER 2 INSTALLATION .....................................................8
2.1 INTRODUCTION ....................................................8
2.2 UNPACKING ..........................................................8
2.3 OVERALL CONNECTIONS ............................................8
2.4 POWER LINE MAINS .......................................... 10
2.5 Cable connection illustration ............................. 10
2.5.1 HV Cable Connection ........................................ 10
2.5.2 Stator Cable Connection .................................... 11
2.5.3 Filament Driver Cable Connection .................... 11
2.6 MULTI INTERFACE CABLE CONNECTIONS .. 12
2.6.1 Pin Layout of the Multi interface ....................... 12
2.7 INTERLOCK CONNECTIONS ...................................... 14
2.8 TUBE CURRENT MONITOR CONNECTION................. 14
2.9 GROUND CONNECTION .................................... 14
2.9.1 X-Ray Tube Housing Ground ............................. 14
2.9.2 Chassis Ground .................................................. 15
2.10 FINAL CHECKS .................................................... 15
2.11 GUI SOFTWARE .................................................. 15
CHAPTER 3 ........................................................................... 17
INTERFACING ........................................................................ 17
3.2 MULTI INTERFACE............................................. 17
3.2.1 Connector Style and Pin Layout ........................ 17
3.2.2 Serial Communication Interface ........................ 18
3.2.3 Digital I/O Control Inte rface ............................. 19
3.2.4 Remote Status Interface ..................................... 20
3.2.5 Remote Monitoring Interface ............................. 22
3.3 ETHERNET INTE RFACE ............................................. 22
CHAPTER 4 CONFIGURATION, CAL IBRATION AND
OPERATION .......................................................................... 23
4.1 INTRODUCTION ........................................................ 23
4.2 TUBE PREHEAT TABLE CONFIGURATION .................. 23
4.3 FILAMENT AU TO -CALIBRATION ............................... 25
4.4 OPERATING X-RAY EXPOSURES ............................... 26
4.4.1 Normal Exposure ............................................... 26
4.4.2 AEC Exposure .................................................... 28
4.4.3 Smart AEC Exposure ......................................... 29
CHAPTER 5 ........................................................................... 32
TROUBLE SHOOTING ............................................................ 32
5.1 INTRODUCTION ................................................. 32
5.2 STATUS AND ERROR CODES ........................... 32
5.2.1 Status Messages ................................................. 32
5.2.2 Faults Messages ................................................ 33
CHAPTER 6 ....................................................................... 39
MAINTENANCE ............................................................... 39
6.1 INTRODUCTION ................................................. 39
6.2 PERFORMANCE TEST ............................................... 39
6.3 HIGH VOLTAGE D IVIDER S ........................................ 39
CHAPTER 7 ....................................................................... 40
REPLACEMENT PARTS .................................................. 40
CHAPTER 8 ....................................................................... 41
FACTORY SERVICE ........................................................ 41
PMX MANUAL 1 118164-001 REV. A
CHAPTER 1 INTRODUCTION
1.1 PMX DESCRIPTION
Spellman’s PMX redefines the standard for high performance Mam mography X-Ray generator s including FFDM and DBT. It features 2 and 3 point exposure modes as well as smart AEC with pre exposure.
The 50kV @ 5kW (peak) high voltage output coupled with a dual focal spot, DC current source filament power supply provides stable and accurate X-Ray tube emission currents. The solid encapsulated high voltage output section eliminates oil concerns while reducing effects of environmental humidity and contamination.
PMX’s flexible interfacing and controlling capabilities as well as powerful application features represent a radically new advanced approach to mammography X-Ray generators.
PMX MANUAL 2 118164-001 REV. A
1.1.1 Power Supply Requirements Single Phase
Line Voltage 200Vac-240Vac, +/-10%, 1 Phase, 35 Amps Line Frequency 50/60 Hz. In Rush Current 50 Amps, Maximum Standby Current 1 Amp
WARNING: The PMX generator can provide power that can damage the X-Ray tube. The user is
responsible for limiting the maximum exposure values to within the ratings of the X-Ray tube. Spellman is not liable for any damage to the X-Ray tube by the misuse of the PMX X-Ray generator.
The following table defines the power line requirements for the generators.
NOTE: THE FOLLOWING TABLE CONTAINS RECOMMENDED VALUES FOR THE WIRE SIZES
BETWEEN THE MAINS DISCONNECT AND THE GENERATOR. THE ACTUAL VALUES USED AT AN INSTALLATION ARE DEPENDENT ON THE QUALITY OF THE INPUT LINE (VOLTAGE LEVEL) THE CURRENT REQUIREMENTS AND THE LENGTH OF THE CABLE RUN AND MUST BE CONFIRMED BY THE INSTALLER.
ALL THE RATINGS LISTED CONSIDER THE GENERATOR REQUIREMENTS ONLY. THE INSTALLER MUST MAKE THE NECESSARY COMPENSATION FOR ADDITIONAL LOAD REQUIREMENTS
A POOR QUALITY INPUT LINE MAY RESULT IN THE INSTALLER HAVING TO DERATE THE GENERATOR'S MAXIMUM POWER
Mains
Voltage
220 Vac
Minimum
Recommended
Mains
Disconnect
to Generator
15 ft./5 m max)
#10
2
(6 mm
)
External EMC Filter (required to meet CE/EMC specifications) – Not provided with the generator. Custom er is responsible f or providing the external filter as needed; Schaffner FN2070-08-36A type EMI filter or equivalent one recommended.
PMX MANUAL 3 118164-001 REV. A
Generator
In Rush Line
Current
50 A 35 A #10
Minimum
Recommended
Generator
Service Rating
Mains Contactor - Not provided within the generator. Customer is responsible for mains safety disconnection.
Minimum
Recommended
Ground
Wire
Size
2
(6 mm
)
Apparent
Mains
Resistance
0.06
Operating Temperature
10 to 40 °C (50 to 104 °F).
Relative Humidity
20 to 85%, non-condensing.
Atmospheric pressure range
500 to 1060 hPa (375 to 795 mm Hg).
TRANSPORT AND STORAGE
Ambient temperature range
-40 to 85 °C (-40 to 185 °F).
Relative humidity
5 to 95%, non-condensing.
Atmospheric pressure range
500 to 1060 hPa (375 to 795 mm Hg).
1.1.2 Environment Requirements Operating Environment
1.2 THEORY OF OPERATION
1.2.1 Function Overview
The PMX X-Ray generator utilizes sophisticated power conversion technology. A variety of analog, digital and switching pow er conversion techniques are used throughout.
The PMX generator is basically an AC to DC power converter. Within the generator, conversions of AC to DC, then to high frequency AC, then to high voltage DC take place. By reviewing further the sub-assemblies, a basic understanding of the process can be gained.
1.2.2 Line Rectification and Filtering
Basic AC line rectification provides the DC buss voltage for the high voltage inverter, Anode rotor drive, and the filament supply.
The input line voltage can vary from 180V up to 264V within the series.
The line voltage is connected directly to the EMI filter. The EMI filter reduces conducted HF noise on the AC mains. The output of the EMI f ilter is connected to an off line auxiliary power inverter circuit that provides 2 4V power for gener ating other auxiliary supply DC voltages.
1.2.3 Inverter
The inverter is a “half -bridg e” topolog y. Cur rent mode control is used for driving the inverter. An IGBT power module employed. The IGBT provides high frequency switching to control the
primary current flow to the high voltage transformer.
Circuits on the Power PCB board provide the gate control of the IGBT switches. The system control board generates gate drive control signals. The circuits on t he Po wer PC B pro vi de the required gate voltages and off s ets .
1.2.4 High Voltage Transformer
The output of the High Frequency Quasi­resonant Inverter is connected to the primary of the High Voltage Transformer. The High Voltage Transform er is a step up type. T ypical secondary output voltage is in the range of
12.5kV depending upon output voltage ratings.
1.2.5 High Voltage Assembly
The High Voltage Assembly will vary depending upon the model ord ered. The circu itry typically consists of an arrangement of half wave voltage doubler.
A high bandwidth resistive/capacitive divider provides voltage feedback for regulation and monitoring. A sense resistor connected at the low voltage end of the High Voltage Multiplier circuit pro vides current feedback for regulation and monitoring.
1.2.6 System Control PWB
Auxiliary DC voltages are generated in the low voltage power supply section of the System Control PWB.
PMX MANUAL 4 118164-001 REV. A
LEDs provide status diagnostics needed for recognizing the operation/shutdown mode of the generator.
A precision +10Vdc reference is generated on the control PCB for programming kV, mA and other references.
Control of the generator utilizes sophisticated analog and digital circuitr y resulting in fast and accurate control, protection and communication interfacing.
This generator is based on advanced quasi­resonant PWM control. Analog signals are digitized in A/D conver ter and pr ocess ed with in FPGA and DSP circuits to provide maximum accuracy and reliability.
All feedback signals are sent to the user interface through digital an d D/A circ uits where switching is possible between feedback and program signals. This allows the us er to pres et the desired output before energizing high voltage.
All program voltages ar e typically ramped up to set level by the digital ramp generator.
FPGA and accompanying A-D and D-A converters and drivers provide system Fault
Control Indicat ion. User interface is proces sed on this PWB as well, providing isolated relay coils, opto couplers and open collector contacts.
1.2.7 Filament Power Supply
The filament inverter provides the power to drive the filament winding of the X-Ray tube. The filament inverter is a high frequency, series resonant inverter. The inverter provides high frequency ac current to the primary of the filament isolation transformer. The filament isolation transformer secondary is connected to a full wave rectifier and capacitor f ilter creating a DC filament. Rela y K4 connects the f ilament output to TB3 terminals 1 or 3 for small or large filament respectively as required by the operator. The filam ent circuitry also provides a variety of control, diagnostic and protection functions.
See Figure 1.2 for a simplified diagram of the X­Ray tube connection and current sensing circuits. The filament circuitry also provides a variety of control, diagnostic and protection functions.
If any abnormal c ondition appears, monitoring circuitry will shut down the unit indicating the fault.
Figure 1.2 Simplified schematic of X-Ray tube connection
PMX 5 118164-001 REV. A
1.2.8 High Speed Starter
The rotating anode is powered and controlled by the action of a split-phase t ype of induction motor circuits in the PMX generator. The rotating anode has considerable mass to dissipate the heat generated by the beam current, and subse quently requires some tim e to accelerate to its norm al operat ing sp eed. T o reduce this time, the “h igh s peed s tart er ” cir cuit in the PMX applies a series of “boost” steps consisting of different frequencies and increased voltage during the boost cycle to
increase the accelerati on of the anode rotat ion to operating speed within several seconds. Following the boost cycle, the applied voltage is reduced to proper vo ltage to maintain ro tation. There are two operati ng modes for the anode rotor: low speed (60Hz stator voltage), and high speed (180Hz stator voltage). These are selected by the operator depending on the requirements of the application. The PMX is equipped with safety interlock s for user personnel and equipment protection. An open interlock circuit inhibits operation of the PMX generator.
PMX MANUAL 6 118164-001 REV. A
High voltage symbol use d to indic ate the pres ence of
Warning symbol used to indicate a potential hazard to
1.3 SAFETY
1.3.1 Safety and Warning Symbols
WARNING: THIS X-Ray UNIT MAY BE DANGEROUS TO OPERATOR UNLESS SAFE EXPOSURE
FACTORS AND OPERATING INSTRUCTIONS ARE OBSERVED.
The following advisory symbols are used on the safety warning labels, and/or on circuit boards.
high voltage.
operators, service personnel or to the equipment.
PMX MANUAL 7 118164-001 REV. A
CHAPTER 2 INSTALLATION
2.1 INTRODUCTION
This chapter contains instructions for unpacking, positionin g, and cabl ing the PMX generator, allowing for initial power-up and exposures.
2.2 UNPACKING
WARNING: THE GENERATOR
WEIGHS APPROXIMATELY 23 POUNDS (10.5KG).
1. Inspect the package exterior for evidence of dam age du e to handling in transit. Notify the carrier and Spellman immediately if damage is evident. Do not destroy or remove any of the packing material used in a damaged shipment.
2. Remove the cardboard outer packaging. See the cautionary note below before removing the packaging.
CAUTION: OPEN THE CARDBOARD PACKAGING CAREFULLY. SHARP TOOLS MAY DAMAGE THE CONTENTS.
3. Set aside the cardboard packaging.
4. After unpacking, inspect for visible damage.
5. Keep the shipping containers. In case of shipping dam age, place the unit(s) back in its s hipping pack and notify the carrier and the Customer Support Department as indicated in this manual.
2.3 OVERALL CONNECTIONS
NOTE: THIS IS BASIC CONNECTION ILLUSTRATION FOR TESTING. MORE COMPLEX AND DEDICATED CIRCUITRY IS NEEDED IN THE FINAL APPLICATION.
All cables should be route d away from high voltage areas, and dressed and secured neatly in place. Cables s hould be cut to the correct length if possible as excess cabling may contribute to EMI/RFI problems. For those cables that cannot be cut to the correct length (HV cables and console cables for example), try to minimize the area inside any loops of excess cable, as t hese loops are in effect an antenna.
PMX MANUAL 8 118164-001 REV. A
PMX MANUAL 9 118164-001 REV. A
Figure 2-1: PMX I/O and indicator locations
TB1
1.GROUND
2.AC-1
3.AC-2
JB1
CONTROL
INTERFACE
TB3
TB2
ANODE DRIVE
FILAMENT DRIVE
HIGH VOLTAGE OUTPUT
JB3
ETHERNET
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
1
2
3
4
GROUND STUD
SYSTEM INTERLOCK
FILTER
N
L
PE
mA TEST PINS
FUSE CONTACTOR
INTERFACE CABLE
SYSTEM INTERFACE
(OPTIONAL)
X-RAY TUBE
HV CABLE
VMX GENERATOR
POWER LINE
LV CABLE
INTERLOCK
SHORT THE WIREPAIR IF SYSTEM INTERLOCK NOT AVAILABLE
INTERLOCK WIRE
USER SUPPLIED COMPONENT
NOTE:
2.4 POWER LINE MAINS
WARNING: MAIN CONTACTOR IS NOT PROVIDED WITHI N TH E G EN ER A TO R , CUSTOMER IS RESPONSIBLE FOR MAINS SAFETY DISCONNECTION.
PMX MANUAL 10 118164-001 REV. A
Figure 2-2: Overall connection
The PMX supply does not contain power line current limiting or fuses. It is strongly recommended that the user provide fuses or circuit breaker protection on the input power lines to limit fault current that may occur due to component failure, or operator error. Fuses or circuit breake rs with current rating of 35 Amps are recommended
Refer to Chapter 1 for generator po wer and generator power line requirements.
1. Check the input voltage rating on the nameplate of the supply and make certain that this is the rating of the power source to be connected.
2. Units operate on 200V-240VAC, +/-
3. A three-wire terminal block (TB1) is
4. DO NOT SWITCH ON MAINS
2.5 Cable connection illustration
2.5.1 HV Cable Connection
The X-Ray tube should be mounted in its
1. Verify that the HV cable terminations
2. Be sure that the rubber expansion
10%, 50/60Hz, single phase 35A.
provided for connecting to the AC supply (refer to Figure 2-2).
POWER AT THIS TIME.
normal fixture i.e. tube stand or other device.
are clean and in good condition.
ring is located at the bottom of the HV cable plug.
3. Connect HV cable to the CA3 type high voltage output terminal at rear of the unit.
4. Connect the other side of HV c able to the corresponding HV te rminal on tube housing.
5. Connect the cable shield wire accompanying HV c able to the PMX ground stud and tube housing separately.
2.5.2 Stator Cable Connection
Refer to Table 2-1 for the X-Ray tube stator connections.
1. Tube low voltage cable is a hybrid one with filament and stator drive wire bundled together. Tube side should be standard terminal style, while generator side is unterminated. Customer should term inate the wire with suitable terminal.
2. Connect wires from the X-Ray tube housing low voltage connector stator pins to PMX terminal block TB2 terminals 1 (stator PHASE), 2 (stator RUN), 3 (stator COM), and 4 (GROUND).
3. Connect tube thermal switch wires to TB3-5 and TB3-6 if present. If not, short the two pins. Normally, there are no such separate wires for MAMMO tubes; it is always integrated into the three phase line. Refer to tube hous ing dat a sheet for specific information.
2.5.3 Filament Driver Cable Connection
1. Filament driver cable has three wires and bundled together with stator wires within same cable. Customer should terminate the wire with suitable terminals and lea ve enough length to reach the terminal TB3.
2. Connect wires from the X-Ray tube housing low voltage connector filament pins to PMX terminal block TB3 terminals 1 (FIL Small), 2 (FIL COM), and 3 (FIL Large).
PMX MANUAL 11 118164-001 REV. A
Pin #
Definition
Description
TB3-1
SMALL FIL
To tube small filament.
TB3-2
COMMON
To tube filament common.
TB3-3
LARGE FIL
To tube large filament.
Pin #
Definition
Description
JB1-1
GND
Generator signal groun d for referenc e of interface
JB1-2
(+) 5VDC Out
5 volts DC from generator for use in powering
such as optocouplers. 100mA
maximum.
JB1-3
RS232 Tx Out
RS232 Transmit (out) from generator
JB1-4
RS232 Rx In
RS232 Receive (in) to generator
JB1-5
PREP
User Signal (Contact Closure) to alert the this signal is active, exposure parameters are enables the starter to boost the rotor. Contact
JB1-6
READY
Generator Signal to user to indicate the rotor r uns
JB1-7
ROTOR SHUTDOWN
User Signal to generator to brake the rotor drive off.
JB1-8
EXPOSURE
User Signal (Contact Closure) to generator to Voltage is generated after the boost time. pin 24. Closed=Exposure.
Table 2-2. Pin layout of filament driver connector.
2.6 MULTI INTERFACE CABLE CONNECTIONS
The PMX has multi interface connections include digital I/O, serial communication, and interlock. Operator m ust verify and connect every signal properly though some of them are optional, before initiating power-up and basic test.
The interlock, serial communication and exposure buttons interface shall be made
available via a multi signal cable. Refer to *-* for a schematic of isolation and signal direction.
2.6.1 Pin Layout of the Multi interface
The multi signal interface shall have a female 25 pin D-Sub (JB1) on the PMX. Twisted pairs shall be used where applicable. The pin layout is giv en in Table 2-3.
circuits
interface circuitry
generator that exposure sequence will begin. Once locked in and cann ot be changed. The gener ator connection to pin 24. Closed = PREP.
to speed and the generator is ready for X-Ray exposure. Transistor output. Low/Active=Ready
Low brakes Rotor. (Open wil l not restart rot or until a new PREP cycle is started). Reference to GND.
generate X-Rays. Filament is boosted, and High
PMX MANUAL 12 118164-001 REV. A
(Approximately 1 second). Contact connection to
JB1-9
X-Ray ON 75% Status
Transistor output to indicate X-Ray ON status. Status Synchronize with 75% kV setting point.
JB1-10
X-Ray ON Status
Transistor output to indicate HV ON status. Synchronize with HV start-up.
JB1-11
N/C
Not Used
JB1-12
X-Ray SHUTDOWN/AEC
User Signal to generator to rapidly turn HV OFF
Reference to
GND. LOW Shuts down HV.
JB1-13
RS232 ISO Ground
Isolated ground from the isolated RS232 transceiver I.C.
JB1-14
HVG FAULT Status
Generator signal indicating generator fault. Open collector transistor output. Lo w/Ac ti ve =Fau lt
JB1-15
Status Bit 1
3 bit status lines f or up to 6 status m essages . See functionality. Open Collector transistor output.
JB1-16
Status Bit 2
JB1-17
Status Bit 3
JB1-18
N/C
N/C
JB1-19
N/C
N/C
JB1-20
kV Monitor
Signal from generator. 0-10V=0-50kV. Zout=1K ohm
JB1-21
Emission Monitor
Signal from generator. 0-10V=0-200m A. Zout=1K ohm
JB1-22
Filament Current Monitor
Signal from generator. 0-10V=0-6A. Zout=1K ohm
JB1-23
Program/Monitor Return
Ground for reference of program and monitor signals
JB1-24
+24VDC Out
For connection to P REP and EXPOSURE control relay coils.
JB1-25
SHIELD/GND
For connection of interface cable shield to generator chassis ground.
(Reference to GND). High/Active=X-Ray ON
(Reference to GND). High/Active=HV ON Status.
and ON during Exposure. (AEC).
separate matrix (Table 2-4) describing Low/Active=Message.
Table 2-3 Pin layout of multi signal connector
Refer to chapter 3 for detailed requirements and function descriptions.
PMX MANUAL 13 118164-001 REV. A
Pin #
Definition
Description
TB3-5
Interlock 2 +
Used if tube has separate thermostat switch. TB3-6
Interlock 2 -
TB3-7
Interlock 3 +
Used if tube has cooling circulator flow switch. TB3-8
Interlock 3 -
TB3-9
Safety Interlock+
User signal (Contact Closure) f or safety interlock s
HV OFF, or
TB3-10
Safety Interlock-
Pin #
Definition
Description
TB3-15
Tube current +
Tube current flows out from this pin.
TB3-16
Tube current -
Tube current flows into this pin.
2.7 Interlock Connections
Interlock connections are provided for a variety of safety functions such as room door interlock, separate thermal switch interlock, tube flow switch interlock. The unit will not operate unless the interlocks are closed.
2.8 Tube Current Monitor Connection
Tube current test pins are provided for monitoring the actual tube current (mA) if needed. Normally, a jumper is installed to short the two pins (refer to Figure 2-2).
The pin definition shows below:
Opening the interlock circuit will cause the unit to shut off.
Three safety interlocks are located in TB3 for system interlock purpose. Short the terminals if not used.
The pin definition shows below:
Open=over temp. (Short terminals if not used).
Open=no flow. (Short terminals if not used).
such as door interlocks. Open turns inhibits HV from being generated.
Remove the jumper and install the current meter in series to monitor the actual tube current.
Two test pins are located in TB3 (TB3-15, TB3-16). Short the two pins if not used.
2.9 GROUND CONNECTION
NOTE: THE INSTALLER SHOULD ENSURE THAT ALL CABLE CONNECTIONS TO THE GENERATOR
ARE SECURE, AND ALL CABLES EXTERNAL TO THE GENERATOR ARE ADEQUATELY PROTECTED AGAINST ACCIDENTAL DISCONNECTION.
2.9.1 X-Ray Tube Housing Ground
A separate ground wire (10 AWG, 6mm must be connected from the X-Ray tube housing to the ground stud on the r ear of the PMX chassis. These ground locations may have other ground wires already connected;
PMX MANUAL 14 118164-001 REV. A
ensure that these existing ground wires are not disconnected when making the X-Ray
2
)
tube ground connection. Failure to make this groun d connection may
result in intermittent operation and/or exposure errors or even worse IGBT failure.
2.9.2 Chassis Ground
The chassis of the P MX generator m ust be grounded to the local earth grou nd and also to the tube housing groun d. See Figure 2-2 for typical operating setup. A three-wire terminal block is provided for connecting to the AC supply.
2.10 FINAL CHECKS
NOTE: THE INSTALLER SHOULD ENSURE THAT ALL CABLE CONNECTIONS TO THE
GENERATOR ARE SECURE, AND ALL CABLES EXTERNAL TO THE GENERATOR ARE ADEQUATELY PROTECTED AGAINST ACCIDENTAL DISCONNECTION.
2.11 GUI SOFTWARE
The optional GUI utility software allows for data communication between the generator and an external computer. This provides a “virtual console” with which user can release all the command defined in the protocol and verifies all the functions independently before
The room interface c onnect ions m ay now be completed. Before power on, user needs to check the items as below finally.
When finished all wiring, check that all connections are tight and secure.
Double check the breaker, EMC filter, fuses, etc. User is res ponsible f or on t he main input power lines.
Check that all cables are dressed neatl y outside the cabinet, and secured as necessary.
Check the ground connection again.
integrating the generator into their ultimate system.
The GUI consists of the Analog Monitor Section, the System Status Section, the Fault Section, the Command Tab Section, the Com Monitor Display, and the Menu.
PMX MANUAL 15 118164-001 REV. A
The Command Tab section, along with the Monitor / Status / Fault sections provide access to the various com mands defined in the PMX Protocol Spec. The Analog Monitor, System Status, and Fault sections are always visible and can be accessed at any time. The Command Tab section provides access to
function. Clicking on the tabs at the top of the Command Tab section selects a particular group of commands. Only one group (tab) can be selected at any one time.
See separate docum ent for t he detaile d GUI instruction.
other commands grouped together by
PMX MANUAL 16 118164-001 REV. A
Pin #
Definition
Description
JB1-1
GND
Generator signal groun d for referenc e of interface
JB1-2
+5VDC Out
5 volts DC from generator for use in powering
. 100mA
JB1-3
RS232 Tx Out
RS232 Transmit (out) from generator
JB1-4
RS232 Rx In
RS232 Receive (in) to generator
JB1-5
PREP
User Signal (Contact Closure) to alert the this signal is active, exposure parameters are enables the starter to boost the rotor. Contact
CHAPTER 3 INTERFACING
3.1 introduction This chapter describes the interfacing of the PMX generator to the customer s ystem control side, especially with th e serial communication,
3.2 MULTI INTERFACE
NOTE: THE INSTALLER MUST PROVIDE THE NECESSARY INTERFACING CABLES FOR WIRING TO THE
GENERATOR INPUTS AND OUTPUTS DESCRIBED IN THIS SECTION.
WARNING: LINE VOLTAGE IS PRESENT INSIDE THE GENERATOR AT ALL TIMES THAT THE MAIN
DISCONNECT IS SWITCHED ON. FOR SAFETY, THE MAIN DISCONNECT SHOULD BE SWITCHED OFF AND LOCKED OUT WHILE CONNECTING ROOM EQUIPMENT.
3.2.1 Connector Style and Pin Layout
exposure control switch. Also, exposure mode is introduced with the timing sequence described as well.
Figure 3-1 multi signal interface connector
PMX MANUAL 17 118164-001 REV. A
circuits
interface circuitry such as opto couplers maximum.
generator that exposure sequence will begin. Once locked in and cann ot be changed. The gener ator connection to pin 24. Closed = PREP.
JB1-6
READY
Generator Signal to user to indicate the rotor r uns
JB1-7
ROTOR SHUTDOWN
User Signal to generator to brake the rotor drive off.
JB1-8
EXPOSURE
User Signal (Contact Closure) to generator to
pin 24. Closed=Exposure.
JB1-9
X-Ray ON 75% Status
Transistor output to indicate X-Ray ON status.
Ray ON
Status Synchronize with 75% kV setting point.
JB1-10
X-Ray ON Status
Transistor output to indicate HV ON status. Synchronize with HV start-up.
JB1-11
N/C
Not Used
JB1-12
X-Ray SHUTDOWN/AEC
User Signal to generator to rapidly turn HV OFF
Reference to
GND. LOW Shuts down HV.
JB1-13
RS232 ISO Ground
Isolated ground from the isolated RS232 transceiver I.C.
JB1-14
HVG FAULT Status
Generator signal indicating generator fault. Open collector transistor output. Lo w/Ac ti ve =Fau lt
JB1-15
Status Bit 1
3 bit status lines f or up to 6 status m essages . See functionality. Open Collector transistor output.
JB1-16
Status Bit 2
JB1-17
Status Bit 3
JB1-18
N/C
N/C
JB1-19
N/C
N/C
JB1-20
kV Monitor
Signal from generator. 0-10V=0-50kV. Zout=1K ohm
JB1-21
Emission Monitor
Signal from generator. 0-10V=0-200m A. Zout=1K ohm
JB1-22
Filament Current Monitor
Signal from generator. 0-10V=0-6A. Zout=1K ohm
JB1-23
Program/Monitor return
Ground for reference of program and monitor signals.
JB1-24
+24Vdc Out
For connection to P REP and EXPOSURE control relay coils.
JB1-25
SHIELD/GND
For connection of interface cable shield to generator chassis ground
to speed and the generator is ready for X-Ray exposure. Transistor output. Low/Active=Ready
Low brakes Rotor. (Open wil l not restart rot or until a new PREP cycle is started). Reference to GND.
generate X-Rays. Filament is boosted, and High Voltage is generated after the boost time. (Approximately 1 second). Contact connection to
(Reference to GND). High/Active=X-
(Reference to GND). High/Active=HV ON Status.
and ON during Exposure. (AEC).
Table 3-2 Pin layout of multi signal interface
3.2.2 Serial Communication Interface
separate matrix (Table 2-4) describing Low/Active=Message.
PMX MANUAL 18 118164-001 REV. A
Pin #
Definition
Description
JB1-3
RS232 Tx
RS232 Transmit (out) from
JB1-4
RS232 Rx
RS232 Receive (in) to
JB1-13
RS232 ISO Ground
Isolated ground from the
The serial communication is part of the Multi Signal Interface (JB1).
The pin definition shows below:
User should connect the R S232 cable properl y with system side such as PC. Twisted wires are preferable to enhance the EMC performance.
GUI software can be used temporarily to test the PMX provided installed on the user PC. (Ref er to separate document for detailed information of GUI). But user need to develop their own ultimate software based on open protocol (Refer to separate document) and design system control board to integrated PMX into their system with proper method (Refer to 2.6.1 NOTE the isolation requirement).
3.2.3 Digital I/O Control Interface
Remote digital control of the PREP, ROTOR SHUTDOWN, EXPOSURE, X-Ray SHUTDOWN, are available on the 25-pin I/O connector for maintenance. All of these controls except PREP and EXPOSURE are toggle on/off type signals. PREP and EXPOSURE are
The RS232 communic ation interface has been isolated inside t he HVG with specific IC.
generator
generator
isolated RS232 transceiver I.C .
controlled by a momentary normally open contact. If remote control is required, the remote interface dummy plug must be removed and replaced with the appropriate remote circuits. See Figure 3-2 for recommended interface circuits.
In order to eliminate sig nal bounce caused by mechanical contact, an el ectron ic sw itch such as opto coupler device is recommended for these signals.
Warning: Max current that the +24Vdc supply can provide is 100mA.
PMX MANUAL 19 118164-001 REV. A
JB1
Customer Interface
High Voltage power supply
1
7
12
5
8
+5V
10K
Transistor On = Rotor shutdown
+5V
10K
Transistor On = X-ray OFF
+24V
Switch closed
for PREP
+24V
Switch closed
for EXPOSURE
NOTES: Relay coil Z=178
Ω
+5V
24
+24V
3.2.4 Remote Status Interface
Remote status signals f or KV MIN tube curren t limit, tube power limit, overvoltage, and filament
Figure 3-3 for recommended interface circuits.
Figure 3-2 Digital I/O Control Interface
Current limit, over temperature, and power supply fault is provided. These signals are open collector type. (Normally off), and are intended to drive the diode of an opto-coupler.
PMX MANUAL 20 118141-002 REV
JB1
Customer Interface
Generator interface
6
9
NOTES: 50mA max collector current on internal power supply transistors
+5V
330
100
+5V
1K
100
+5V
1K
100
10
+5V
330
100
+5V
330
100
+5V
330
100
+5V
330
100
READY
X-RAY ON Status
HV ON Status
14
HVG FAULT Status
15
BIT1 Status
16
BIT2 Status
17
BIT3 Status
LOW=READY
HIGH=X-RAY 75% ON
HIGH=X-RAY ON
HIGH=FAULT
LOW=ACTIVE
LOW=ACTIVE
LOW=ACTIVE
BIT1
BIT2
BIT3
MESSAGE
high
high
high
No Messages
low
high
high
Tube Temp/Flow Fault/HSS Fault
high
low
high
Arc Detected/Shutdown
high
high
low
Over Load: Power/MAS/Time/Duty
low
low
high
Over Voltage/Over Current/Regulation error
low
low
low
Open Filament/Filament Fault
Figure 3-3 Remote status interface
If a fault occurs, the power supply will shut down and a 3 bit binary code will be gen erate d at the user interface connector J B1. T o res et, use the reset function on remote user interface.
Table 3-3 status messages matrix
The fault inf orm ation is provided with three bits code. Below is the information list. More than one fault is defined with one code combination. Use can get specific fault information by serial command.
PMX MANUAL 21 118164-001 REV. A
JB1
Generator side
22
Customer side
20
21
1K
0.1
1K
0.1
1K
0.1
23
M
M
M
kV monitor
mA monitor
Filament
monitor
Monitor return
NOTES:
1) Metering circuit should be high Z type(>10Meg)for accurat e measurement.
2) All signals 0-10Vdc = 0-100 % of specified maximum outputs .
Pin #
Definition
Description
JB3-1
TX+ White/Orange
JB3-2
TX- Orange
JB3-3
RX+ White/Green
JB3-4
POE V+ blue
JB3-5
POE V+ blue/white
JB3-6
RX- Green
JB3-7
POE V- brown/white
JB3-8
POE V- brown
3.2.5 Rem ote Monitoring Interface
Test points are made available at the 25-pin I/O connector from monitoring tube voltage, tube current and tube filament current. The test
Figure 3-4 for wiring diagram and specifications for remote monitor interface.
points are always positive regardless of the output polarity, where 0 to 10V = 0-100% of output. Accuracy is within ±0.5%. All test points have output impedance of 1Kohm.
Figure 3-4 Remote monitoring interface
3.3 ETHERNET INTERFACE
Ethernet is an optional serial communication interface with separate standard RJ45 style
RJ45 pin definition shows below:
connector. JB3 is terminal number.
10/100 Base-T Ethernet Interface
PMX MANUAL 22 118164-001 REV. A
CHAPTER 4 CONFIGURATION, CALIBRATION AND OPERATION
4.1 INTRODUCTION
This chapter describes the procedure for configuring the tube type, configuring user
WARNING: USE EXTREME CARE IN MEASURING HIGH VOLTAGES. ACCIDENTAL CO NTACT
MAY CAUSE SERIOUS INJURY OR DEATH.
MAINS VOLTAGE WILL BE PRESENT INS IDE THE GENERATOR CHASSIS, EVEN
WITH THE POWER SWITCHED OFF. THIS VOLTAGE IS EXTREMELY DANGEROUS, USE EXTREME CAUTION.
THE DC BUS CAPACITORS, LOCATED ON THE DC BUS ASSEMBLY, PRE SENT A
HAZARD FOR A MINIMUM OF 5 MINUTES AFTER THE GENERATOR HAS BEEN SWITCHED OFF. CHECK THAT THESE CAPACITORS ARE DISCHARGED BEFORE TOUCHING ANY PARTS IN THE GENERATOR.
allowed dual speed starter data and operating filament auto-calibration.
WARNING: THE PROCEDURES IN THIS CHAPTER REQUIRE THE PRODUCTION OF X-Rays.
TAKE ALL APPROPRIATE SAFETY PRECAUTIONS TO PROTECT PERSONNEL FROM X-Ray RADIATION.
WARNING: ALWAYS ENSURE THAT THE EQUIPMENT UNDER TEST AND ALL ASSOCIATED
TEST EQUIPMENT IS PROPERLY GROUNDED
ENSURE THAT THE HIGH VOLTAGE C ABLES ARE INTACT / UNDAMAGED AND
PROPERLY CONNECTED BEFORE ATTEMPTING EXPOSURES
ENSURE THAT THE PMX IS PROPERLY INSTALLED AS PER CHAPTER 2 AND CHAPTER 3 BEFO RE OPER AT I NG W H AT THIS CHAPTER DESCRIBES.
4.2 TUBE PREHEAT TABLE CONFIGURATION
The PMX contains up to 2 tube types power limitation an d filament value tables. Consult Spellman representati ves when ordering for
available tube tables pre-installed or tube type field configuration possibility.
Reloading the default preheat tables in the PMX u nit requires use of the DIP switc hes. The DIP switches are located near the edge of the board, to the right of the connector. The DIP switch identifiers are stenciled on the PC Board just above the switches, and the individual switches are numbered at the bottom of the DIP switch. The figure below illustrates the DIP switch placement.
PMX MANUAL 23 118164-001 REV. A
Table 1 -Tube Table Selection Matrix
Tube Selected
Tube number
Switch #1 Position
Switch #2 position
0
(Varian) M113T
3
Down
Up
Invalid Setting (do
not use)
Step
Action
1.
Power down the system, if it is not already powered down. Switch #8 on DIP Switch 1 should be in the up position as show above. This is the normal working position.
2.
Set Switch #8 on Dip Switc h 1 to the down position, th is is with the switch s et toward #2 on DIP SW1 to the desired tube table as specified in Table 1 above.
1 2 3 4 5 6 7 8
1 2 3 4 5 6 7 8
SW1
SW2
DIP SW1 – Switch
DIP SW1 – Switch #1
DIP SW1 – Switch #2
(Up Position Shown)
Figure 4-1: DIP Switches for Configuration
Switch 8 on DIP SW 1 controls the loading of the default Tube Table in to FRAM. When the unit is powered up with this switch in the down position the sel ected tube def ault data is loaded in to FRAM, overwriting any current data, including learned pr eheat values. This switch only causes tab le up date when it is in the down position at power up. Setting the switch to the down position after power up does not change the tube table.
(Down Position Shown)
#8 (Up Position Shown)
Switch 1 and Switch 2 o n DIP SW 1 select which tube table wi ll be loaded in to FRAM and become the active T ube. The setting of these switches only matters when Switch 8 is in the Down position at power u p. Changing the Switch 1 and/or Switch 2 does not change or reload the tube table if the unit is po wer ed up with Switch 8 in the up position.
(IAE)XM1016T
Follow the below steps to select the desired table from the pre-installed tube table:
the center of the board (opposite to the above illustration). Also set switch #1 and switch
PMX MANUAL 24 118164-001 REV. A
Up Up
Down Down
3.
Turn on the power to the PMX unit, and wait for 10 seconds. This will reload the selected default table in to the FRAM.
4.
After the 10 seconds has elapsed, turn the power off, and return Switch 8 t o the Up leave Switch 8 in the Up position.
Step
Action
1.
Entering calibration mode
2.
Enable Prep and Exposure to start calibration
filament will be affected. For example if small filament is calibrated and you attempt to
position. The tables have been reloaded and the system is ready to use. Make sure to
4.3 FILAMENT AUTO-CALIBRATION
The calibration function automatically establishes calibration curves relating filament current to tub e cur r ent (mA) at three or four kV settings, dep ending on generator type. This allows the g enerator to predic t t he filament temperature (filament current) that will be required for any mA value over the full kV operating range.
It is recommended that the X-Ray tube be seasoned before att empting auto cal ibrat ion , particularly if the tube has not been used for
Sending a “Set Calibration Mode” command (command number 28) with an argument of 1 or 2 or 3 with Prep and Exposure control in the OFF state. Gener ator w ill enter calibration mode but in ‘pause’ state as the Prep and Exposure is inactive.
Calibration will be starting if both Prep and Exposure are activated. Once the sequence has started the generator will automatically cycle through the exposure settings as long as the Prep and Exposure control are ON until it either finishes calibration or there is an exposure fault.
An exposure fault will cause the calibration routine to abort, and the unit to exit calibration mode. If the user removes Prep and / or Exposure control during the sequence, the calibration routine will enter a “Paused” state. It will not exit calibration mode, but it will stop doing exposures and wait until Prep and Exposure control are both returned to the ON state. If the user returns Prep and Exposure control ON, the PMX will continue the calibration. Usually it will continue from where it was when the sequence was paused, but depending on what it was doing when the calibration sequence was interrupted it may go back and repeat a previous step.
The user may cancel calibration at any time by first setting Prep and / or exposure control off, and then sending a “Set Calibration Mode” Command with an argument of 0. Sending this command resets the calibration sequence. The next time calibration is started it will go back to the beginning. Once calibration is started previous values for preheat will be overwritten. Stopping calibration in the middle will result in some points being calibrated and others not being calibrated. This may result in poor or inconsistent preheat adjustment.
If user chooses to calibrate large and small filaments separately, only the chosen
some time. Refer to tube datasheet for seasoning operation.
There are two different modes for user selection to calibrate filament. User can calibrate the large filament and small filament separately or calibrate both large filament and small filam ent at one t im e. Sendi ng “Set Calibration Mode” command (command number 28) with different argument will enter different calibratio n mode (Refer to protoc ol specification document for details).
Use these steps to per form the X-Ray tube auto calibration function:
PMX MANUAL 25 118164-001 REV. A
Loading...
+ 108 hidden pages