S PE L L M A N H I G H V O LTA G E E L E C T R O N I C S C O R P O R AT I O N
Spellman’s Bertan brand of 230 Series high voltage power supplies provide regulated high voltage outputs from 1 to 30kV. The low noise, linear topology employed results in extremely low output ripple specifications. These 12 to 15 watt units are inherently reversible by design, providing either positive or negative output polarity. The 230 Series is fully arc and short circuit protected. Excellent regulation specifications are featured along with outstanding stability performance.
TYPICAL APPLICATIONS
HiPot Testing Electrostatics General Laboratory Usage
SPECIFICATIONS
Input Voltage:
115Vac, ±10%, 50/60 Hertz @ 0.5 amp 230Vac, ±10%, 50/60 Hertz @ 0.25 amps Input voltage is switch selectable
Output Voltage:
See “model selection” table
Output Polarity:
All units are reversible polarity by design
Output Current:
See “model selection” table
Voltage Regulation:
Line: 0.002% of rated output voltage over specified
input voltage range
Load:0.005% of rated output voltage for a full load
change
Current Regulation:
Internally set to limit at less than 125% of rated current. A rear panel switch allows limiting at 25% of rated full current.
Ripple:
See “model selection” table
Temperature Coefficient:
100ppm/°C
Stability:
0.01%/hour, 0.02% per 8 hours after a 1/2 hour warm up
Accuracy:
Front panel control: ±(0.2% of setting + 0.2% of maximum) Front panel Meter: Voltage ±(0.5% of setting + 0.5% of
maximum), Current ±(2% of setting + 0.5% of maximum) Remote Programming: ±(0.1% of setting + 0.1% of maximum) Voltage Monitor: ±(0.1% of reading + 0.1% of maximum) Current Monitor: ±(2% of reading + 1% of maximum)
Front Panel Metering and Controls:
Power ON/OFF switch
3.5 digit metering for voltage and current, switch selectable Polarity indicator 10 turn locking potentiometer to set output voltage HV output connector Ground stud
Operating Temperature
0°C to +50°C
Storage Temperature:
-40°C to +85°C
Humidity:
20% to 85% RH, non-condensing
Input Line Connector:
IEC320 EMI filter/input connector, a detachable line cord is provided
Interface Connector:
9 pin “D” connector, a mating connector is provided
Output Connector:
A detachable 10 foot (3 meter) HV cable is provided for units up to 5kV; 10kV through 20kV: 59” (1.5 meter); 30kV: 10 foot (3 meter)
Cooling:
Convection cooled
Dimensions
7.63˝ W X 5.03˝ H X 8.91˝ D (194mm X 128mm X 226mm)
Weight:
10 pounds (4.5kg)
Regulatory Approvals:
Compliant to 2004/108/EC, the EMC Directive and 2006/95/EC, the Low Voltage Directive.
230
HIGH VOLTAGE POWER SUPPLY
MODULAR BENCH TOP DESIGN
LOW RIPPLE AND NOISE
3.5 DI GIT FRONT PANEL DIGITAL METERING
REVERSIBLE OUTPUT POLARITY
USA +1-631-630-3000 FAX: +1-631-435-1620 UK +44 (0)1798 877000 FAX: +44 (0)1798 872479 JAPAN +81 (0)48-447-6500 FAX: +81 (0)48-447-6501 CHINA +86 (0)512-67630010 FAX: +86 (0)512-67630030
e-mail: sales@spellmanhv.com www.spellmanhv.com
128051-001 REV.D
Spellman High Voltage is an ISO 9001:2000 and ISO 14001:2004 registered company
www.spellmanhv.com/manuals/230
FRONT VIEW
BACK VIEW
SIDE VIEW
HV CONN
POS NEG
METER
OFF
POWER
ON
PCS
VFS
GROUND
HIGHVOLTAGE
ADJUST
7.63 194]
5,03 [128]
1.0 [25]
.66 [17]
8.50 [216]
8.25 [210]
PAGE 2 OF 2
MODEL SELECTION TABLE
230 Voltage Current Ripple Series
230-01R 0 to 1kV 0 to 15mA 10mV 230-03R 0 to 3kV 0 to 5mA 30mV 230-05R 0 to 5kV 0 to 3mA 50mV 230-10R 0 to 10kV 0 to 1.5mA 500mV 230-20R 0 to 20kV 0 to 0.5mA 2 volts 230-30R 0 to 30kV 0 to 0.4mA 5 volts
INTERFACE CONNECTOR
PIN SIGNAL PARAMETERS
1 Voltage Monitor 0 to 5Vdc = 0 to 100% rated voltage, Zout = 10K 2 n/c none 3 Enable TTL “0” disables HV, TTL “1” or open enables HV 4 +5Vdc Reference +5.0Vdc @ 10mA, maximum 5 Current Monitor 0 to 5Vdc = 0 to 100% rated current, Zout = 10K 6 Voltage Program Input 0 to 5Vdc = 0 to 100% rated voltage, Zin = 1M 7 Analog Ground Ground 8 Digital Ground Ground 9 Polarity Indicator Open collector, 30V @ 25mA, positive = ON
DIMENSIONS: in.[mm]
S P E L L M A N H I G H V O LTA G E E L E C T R O N I C S C O R P O R AT I O N
230
HIGH VOLTAGE POWER SUPPLY
OPTIONS: Isolated (Floating) Output-Option F
Units up to and including 5kV can be provided with differential outputs capable of floating up to ±2kV from ground. Voltage programming and monitoring functions normally referenced to ground. Current monitoring and metering is eliminated. Replace “R” suffix with “F” for this option.
USA +1-631-630-3000 FAX: +1-631-435-1620 UK +44 (0)1798 877000 FAX: +44 (0)1798 872479 JAPAN +81 (0)48-447-6500 FAX: +81 (0)48-447-6501 CHINA +86 (0)512-67630010 FAX: +86 (0)512-67630030
e-mail: sales@spellmanhv.com www.spellmanhv.com
128051-001 REV.D
Spellman High Voltage is an ISO 9001:2000 and ISO 14001:2004 registered company
IMPORTANT SAFETY PRECAUTIONS
SAFETY
THIS POWER SUPPLY GENERATES VOLTAGES THAT ARE DANGEROUS AND MAY BE FATAL.
OBSERVE EXTREME CAUTION WHEN WORKING WITH THIS EQUIPMENT.
High voltage power supplies must always be grounded.
Do not touch connections unless the equipment is off and the
Capacitance of both the load and power supply is discharged.
Allow five minutes for discharge of internal capacitance of the power supply.
Do not ground yourself or work under wet or damp conditions.
SERVICING SAFETY
.
Maintenance may require removing the instrument cover with the power on.
Servicing should be done by qualified personnel aware of the electrical hazards.
WARNING note in the text call attention to hazards in operation of these units
that could lead to possible injury or death.
CAUTION notes in the text indicate procedures to be followed to avoid possible
damage to equipment.
Copyright © 2000, Spellman High Voltage Electronics Corporation. All Rights Reserved.
This information contained in this publication is derived in part from proprietary and patent data. This information has
been prepared for the express purpose of assisting operating and maintenance personnel in the efficient use of the
model described herein, and publication of this information does not convey any right to reproduce it or to use it for
any purpose other than in connection with installation, operation, and maintenance of the equipment described.
118004-001 REV. B
WICHTIGE SICHERHEITSHINWEISE
SICHERHEIT
DIESES HOCHSPANNUNGSNETZTEIL ERZEUGT LEBENSGEFÄHRLICHE HOCHSPANNUNG.
SEIN SIE SEHR VORSICHTIG BEI DER ARBEIT MIT DIESEM GERÄT.
Das Hochspannungsnetzteil muß immer geerdet sein.
Berühren Sie die Stecker des Netzteiles nur, wenn das Gerät ausgeschaltet ist und die elektrischen
Kapazitäten des Netzteiles und der angeschlossenen Last entladen sind.
Die internen Kapazitäten des Hochspannungsnetzteiles benötigen ca. 5 Minuten, um sich zu entladen.
Erden Sie sich nicht, und arbeiten Sie nicht in feuchter oder nasser Umgebung.
SERVICESICHERHEIT
Notwendige Reparaturen können es erforderlich machen, den Gehäusedeckel während des Betriebes zu
entfernen.
Reparaturen dürfen nur von qualifiziertem, eingewiesenem Personal ausgeführt werden.
“WARNING” im folgenden Text weist auf gefährliche Operationen hin, die zu Verletzungen oder zum Tod
führen können.
“CAUTION” im folgenden Text weist auf Prozeduren hin, die genauestens befolgt werden müssen, um
eventuelle Beschädigungen des Gerätes zu vermeiden.
118004-001 REV. B
Loading...
+ 11 hidden pages