SpectrumLabs Tangential Flow Depth Filtration Product Instructions For Use

Product Instructions for Use
Tangential Flow Depth Filtration
®
(TFDF®)
Introduction
SpectrumLabs.com offers a unique TFDF®filter that combines the benefits of the non-fouling nature of tangential flow filters
ith the benefits of high dirt capacity depth filters. Please refer
w to the following instructions for the care and use.
A. Materials of Construction
Housing: Polysulfone
Filter Media: HDPE 1-3um Pore Rating
HDPE 2-5um Pore Rating
Potting Material: Urethane
-All materials USP Class VI compatible
-Pore Ratings for reference only: Retention rating may vary based on application
B. Filter Preparation
TFDF®filters are shipped ready-to-use with no surfactant or storage agent. Filters wet readily with water and/or buffer.
C. Filter Sterilization
SpectrumLabs.com recommends using the filters with Pro­Connex gle-use MagLev pump and pressure sensors for perfusion applications. When needed, Sampler and MiniKros ters part numbers beginning with S and N) filters can be auto­claved one time using the conditions below:
NOTE: KrosFlo
1. Remove and discard all pouches, as these are not auto-
2. Loosen all module connections to avoid damaging the
3. Wrap all open ports of the connecting tubing with auto-
®
Ready-to-Use, Sterile Flow Paths that include the sin-
®
filters (fil-
®
and KrosFlo®Max filters (filters beginning with K or X) have not been validated for autoclave cycles.
clavable; wet the membrane (optional).
module during the autoclave cycle.
clave paper, making sure permeate cavity is open to steam penetration along with the inlet/retentate side.
4. Ramp up the autoclave by increasing the temperature by 1-2°C/min up to 121°C (15 psig)
TANGENTIAL FLOW DEPTH FILTRATION®(TFDF®) p 1
Tangential Flow Depth Filtration®(TFDF®)
. Hold the autoclave at or above 121°C for 50 min, not to
5
exceed 124.1°C.
6. Slowly ramp down the autoclave temperature by 1­2°C/min until room temperature.
. Filter Operation
E
SpectrumLabs.com recommends the following operating parameters:
Pressure: Not to exceed 30psig inlet pressure.
7. Tighten all module connections.
8. Perform a module integrity test.
D. Integrity Test
Gross Leak Integrity Test
®
TFDF extremely robust. When needed, an integrity test can be per­formed prior to use. The following pressure hold test is recom­mended to verify the integrity of a wetted filter.
1. With the bottom permeate port closed fully, wet the filter
2. Close retentate line and introduce air into the filter via the
3. Do not exceed 1.0 psi.
4. Wait. If rapid bubbles do not appear in the permeate cav-
filters are integrity tested prior to shipment and are
with water and ensure that permeate side outside of the fibers is flooded. Direct the upper permeate line to drain.
inlet line with either pump or air source.
ity within 2-3 seconds, then module is integral. Small bubbles will slowly start to form after a few seconds and will increase in strength to rapid bubbles over time. The first few seconds are key to see the gross failure.
NOTE:
Flow: For mammalian cell perfusion, shear rates from
As process pressures change, retention charac­teristics of the TFDF®filter might be affected.
1500s-1 to 3000s-1 are recommended. For protein or virus clarification, shear rates of up to 8000s-1 may be required for better performance.
F. Storage
To optimize shelf life, we recommend store in original packag­ing at ambient conditions protected from moisture, extreme temperatures, and UV light.
NOTE: The TFDF
tives are a high probability. If rapid bubbles are seen in the first few seconds, redo wet-out proce­dure and integrity test to confirm gross integrity fail­ure.
®
media readily passes air so false nega-
SpectrumLabs.com - Americas
voice 310-885-4600 (world-wide) • 800-634-3300 (toll-free US and Canada)
fax 310-885-4666 (world-wide) • 800-445-7330 (toll-free US and Canada)
e-mail customerservice@spectrumlabs.com
web www.spectrumlabs.com
SpectrumLabs.com China
voice
e-mail
(+86) 21 68810228
fax
(+86) 21 60919246 spectrum.cn@spectrumlabs.com
web www.spectrumlabs.cn
400-13770-000 Rev. 01 DCO 6536 • 170108
SpectrumLabs.com Japan
+81-77-552-7820
voice
fax
+81-77-552-7826
e-mail
spectrum.jp@spectrumlabs.com
web www.spectrumlabs.jp
SpectrumLabs.com Europe
voice +31 (0) 76 5719 419 fax +31 (0) 76 5719 772 e-mail info@spectrumlabs.eu
web www.spectrumlabs.eu
SpectrumLabs.com India
voice
+91.80.4207.7396
fax
+91.80.4207.7396
e-mail
info-India@spectrumlabs.com
web
www.spectrumlabs.in
TANGENTIAL FLOW DEPTH FILTRATION®(TFDF®) p 2
Loading...