Spectrum 3440 Product Manual

Temp / RH / CO2
Hand-Held Meter
PRODUCT MANUAL
Item # 3440
CONTENTS
Specifications 2 Battery 3 LCD Display 3 Keypad 3 Operation 4 Troubleshooting 7 Warranty 8 CE Declaration of Conformity 8
Thank you for purchasing the Temp/RH/CO2 Hand-Held Meter. With an NDIR (non-dispersive infrared) CO this unit is stable for long-term monitoring. Please read this manual thoroughly before using your meter.
sensor,
2
Specifications
CO2
Range 0 to 5000 ppm Accuracy ±50ppm, ±5% of reading
Pressure Dependence +1.8% reading / 10hPa deviation from 1000hPa
Temperature
Range -14 to 140°F (-10 to 60°C) Accuracy ±0.9°F (±0.6°C)
Humidity
Range 0 to 99.9% RH Accuracy ±3% (10-90%), ±5%(other)
Warm up 30 seconds Power 4 AA batteries (or 9VDC, 1A adaptor, not incl.) Battery life 24 hours (alkaline)
Battery
Install the 4 AA batteries into the battery compartment on the rear and make sure they are in correct polarity.
When the battery voltage gets low, the battery icon and “Lob” will appear on the LCD, and the beeper will sound (press any key to stop the beeps).
LCD Display
Symbols HOLD Readings are frozen TWA Time weighted average (8 hours) STEL Short-term exposure limit (15 min wt avg) MIN/MAX Minimum/Maximum readings DP Dew point temperature AIR Air temperature WBT Wet bulb temperature
Keypad
Operation
Power On/Off
Press the On/Off/Set key to turn the meter on or off. During warm-up, it will perform a 30 second countdown, then will dis­play current CO turns off automatically after 20 minutes of inactivity. To bring the meter to non-sleep mode, press and hold SET and HOLD for 2 to 3 seconds. You should see a character (n) appear on the LCD display. At this point, let go of the buttons to cancel the automatic power-off function.
Taking a Measurement
The meter measures continuously and updates the LCD every second. With a significant environmental change, the
sensor requires 30 seconds, and the RH sensor 30 min-
CO
2
utes to respond. Note face because exhalation affects the CO
Switch Temperature Measurement
Press the DP/WBT key to cycle the temperature display be­tween air, dew point, and wet bulb temperatures.
Data Hold
Press the HOLD button to freeze the readings. HOLD will display on the LCD. Press HOLD again to cancel the hold function.
Backlight
Hold the MODE button for more than 1 second to activate and cancel the backlight function.
MIN, MAX, STEL, TWA
Press the M mum, maximum, and weighted average readings.
In MIN and MAX modes, minimum and maximum readings for CO
, air/dew point/wet bulb temperature, RH will display.
2
In STEL and TWA modes, the display shows the weighted average of the CO or 8 hours (TWA), as well as the current readings for air/dew point/wet bulb temperature, and RH.
, temperature, and humidity. The meter
2
: Do not hold the meter close to the
reading.
2
N
/AV button to switch between the normal, mini-
X
readings for the past 15 minutes (STEL)
2
Data Hold
The data hold function works in both IR and probe modes. In IR measurement, just release MEAS button, the meter stops measuring and holds the last reading until next measurement is taken or power off. To hold the data in probe measure­ment, release MEAS button and wait until it gets the stable reading. Then the “HOLD” icon will display on the top of LCD along with the held reading.
Setup
In normal mode, hold down the SET button for at least 1 sec­ond to enter setup mode. To exit setup, press the CAL/ESC button.
P1.0 Set CO
Alarm
2
When entering setup mode, “AL” and “P1.0” are displayed on the LCD. Press the M CO
alarm threshold. The current value will be blinking on
2
N
/AV button to enter “P1.1” to set the
X
the LCD. Press MODE (up-arrow) to increase the value, or DP/WBT (down-arrow) to decrease it. Press the M
N
/AV but-
X
ton to save the value, or CAL/ESC to exit without saving. The meter features audible alarm to warn by emitting 80dB
beeps when the CO
concentration exceeds the limit. The
2
alarm can be stopped by pressing any button except SET. It will also stop when readings fall below the set value.
P3.0 Set Temperature Units From the P1.0 display, press MODE (up-arrow) or DP/WBT (down-arrow) to change to the P3.0 screen (“Unit” will display on the LCD). Press the M
N
/AV button to enter “P3.1” to set
X
the temperature units. The current units, “°C” or “°F”, will be displayed on the lower left area of the LCD. Press MODE (up-arrow) or DP/WBT (down-arrow) to switch between “°C” and “°F”. Press the M
N
/AV button to save the setting, or
X
CAL/ESC to exit without saving. To exit setup, press the CAL/ESC button.
CO
Calibration
2
The meter is calibrated at the factory to a standard 400ppm
CO2 concentration. It is suggested to do a manual calibration regularly to maintain good accuracy. The calibration should be done during sunny weather in fresh outdoor air that is well-ventilated.
CAUTION: Do not calibrate the meter in the air with an un­known CO
concentration. It will be calibrated as 400ppm,
2
which can lead to inaccurate measurements. Place the meter in the calibration site. Turn on the meter and
hold down CAL/ESC and MODE simultaneously to enter CO
calibration mode. “400ppm” and “CAL” will blink on the
2
LCD while performing the calibration. Wait about 5 minutes for the calibration to complete, which is indicated by the blink­ing stops and LCD returns to normal mode. To abort the cali­bration, turn off the meter at any time.
NOTE: Ensure the batteries are at full power during the cali­bration to prevent interruption or a failed calibration.
RH Calibration
Humidity calculations on the meter can be recalibrated using 33% and 75% salt solutions. Ambient conditions should be at 25°C or 77°F, with stable humidity.
Plug the sensor probe into a 33% salt bottle. Hold down the CAL/ESC and DP/WBT buttons until “CAL” is blinking on the LCD display. In approximately 60 minutes “CAL” will stop blinking, indicating the end of this phase. To skip the 75% recalibration, press CAL/ESC to exit.
Plug the sensor probe into the 75% salt bottle, then press the
N
M
/AV button. “CAL” will blink on the LCD until the calibra-
X
tion is completed in approximately 60 minutes. The 33% calibration can be skipped by pressing MODE (up-
arrow) or DP/WBT (down-arrow) during the first 5 minutes of the 33% calibration step above.
Troubleshooting
Unable to power on Press the On/Off/Set key for more than 0.3 seconds and try again. Check whether the batteries have good contact and the correct polarity.
Readings do not change Check whether data hold function was activated (HOLD icon at the left top).
Slow sensor response Check whether the air flow channels on the back of the meter are blocked.
Error messages E01: CO
2
sensor damaged. E02: The value is under range. E03: The value is over range. E04: Error code for dew point or wet bulb accompanying a primary sensor failure code. E07: Too low voltage to measure CO Replace batteries. E11: Retry humidity calibration. E17: Retry CO
calibration.
2
E31: Temperature sensor damaged. E34: Humidity sensor damaged.
.
2
Warr anty
This product is warranted to be free from defects in material or work­manship for one year from the date of purchase. During the warranty period Spectrum will, at its option, either repair or replace products that prove to be defective. This warranty does not cover damage due to improper installation or use, lightning, negligence, accident, or unau­thorized modifications, or to incidental or consequential damages be­yond the Spectrum product. Before returning a failed unit, you must obtain a Returned Materials Authorization (RMA) from Spec­trum. Spectrum is not responsible for any package that is returned without a valid RMA number or for the loss of the package by any ship­ping company.
This equipment has been manufactured for
Spectrum Technologies, Inc.
12360 S. Industrial Dr. East
Plainfield, IL 60585 USA
The Manufacturer’s DECLARATION OF CONFORMITY is on file at the above address, and certifies conformity to the following:
Model Number: 3440 Description: Temp/RH/CO2 Hand-Held Meter Type: Electrical Equipment for Measurement, Control, and Laboratory Use
Directive: 2004/108/EC Standards: EN 61326-1 (2006) EN 61000-3-2 (2006) EN 61000-3-3 (1995/A1:2001/A2:2005) EN 61000-4-2 (1995/A1:1998/A2:2001) EN 61000-4-3 (2006) /-4(2004) /-5(2006 ) EN 61000-4-6 (1996/A1:2001) /-11(2004)
(800) 248-8873 or (815) 436-4440
E-Mail: info@specmeters.com
www.specmeters.com
12360 S. Industrial Dr. E
Plainfield IL 60585
Fax (815) 436-4460
R 10/12
Loading...