Spectra Watermakers Ventura 150, Ventura 200T Installation And Operating Manual

Spectra Watermakers Katadyn Desalination LLC. 2220 S. McDowell Blvd. Petaluma, CA 94954
Phone: 415-526-2780 Fax: 415-526-2787
techsupport@spectrawatermakers.com
Installaon and Operang Manual
2
3
Fresh Water Flush
Pressing the Fresh Water Flushbutton flushes the watermaker with fresh water from the ves­sels domestic water tanks. After completing the flush, Spectra Connect will automatically enter the Auto Store mode.
Stop
Pressing the Stopbutton from the Home Screen will end the Auto Store mode count­down timer and leave your watermaker in Standby mode.
When you first power up the system, you will get a warning message, asking if the system has
been stored with chemicals.
If the system has been pickled, winterized, this is the first startup, or the condition of the sys­tem is unknown, go to COMMISSIONING on page 23 or serious damage may occur.
Start
Pressing the Startbutton once advances to
the Run Mode selection screen.
Spectra Connect Quick Start Guide
Run Mode
Select your desired Run Mode to start making
water and filling your tanks. See details on Run Mode options on p. 37
Spectra Connect Modes and Definitions
Auto Store: After the watermaker fresh water flushes, it will start a countdown timer that can be seen on the Home screen. The timer indicates the next programmed fresh water flush if the water­maker is not started again, or the Stopbutton is not pressed.
Fill Tank: The watermaker will automatically run until the Tank Full switch in the water storage tanks closes. Once the Tank Full switch closes, the watermaker automatically fresh water flushes, then reverts to the Auto Store mode.
Auto Run: The watermaker can be set to run for a number of hours, or for a quantity of water to be produced. When the desired quantity of water is produced or the run timer expires, the watermaker
will Fresh Water Flush and enter the Auto Store mode.
4
5
Operation
Table of Contents
Installaon……………………………………………………………………………………………...
Geng Started ................................................................................................................ 7
Installaon Basics ............................................................................................................ 8
Introducon .................................................................................................................... 9
Components .................................................................................................................. 10
Ventura Plumbing Schemac ......................................................................................... 12
Installaon Notes .......................................................................................................... 13
Product Water plumbing ............................................................................................... 14
Tube Fing Assembly Procedures ................................................................................. 15
Spectra High Pressure Fing Instrucons ...................................................................... 16
Electrical ........................................................................................................................ 17
Wiring ........................................................................................................................... 19
Tank Switch Wiring and operaon ................................................................................. 20
Oponal Z-Ion ............................................................................................................... 22
Operaon……………………………………………………………………………………………………………….…25
New Systems Start Up and Tesng ................................................................................ 27
Dry Tesng with an Arcial Ocean ............................................................................... 29
Sensor Calibraon ......................................................................................................... 30
Networking.................................................................................................................... 33
Normal Operaon ......................................................................................................... 37
Maintenance, Storage, and Troubleshoong………………………….………………………..
Maintenance ................................................................................................................. 42
Suggested Spares………………………………………………………………………………………………………….49
Introducon to Spectra Chemicals ................................................................................. 44
Long Term Storage Procedures ...................................................................................... 46
Winterizing with Propylene Glycol ................................................................................. 47
Membrane Cleaning Procedure ..................................................................................... 48
Troubleshoong ............................................................................................................ 50
Manual Operaon ......................................................................................................... 52
Ventura Flow Test ......................................................................................................... 53
Poor Water Quality ........................................................................................................ 55
Technical Bullens ......................................................................................................... 56
Spectra Sengs ........................................................................................................ 61
Fault Handling ............................................................................................................... 66
Wiring Schemac........................................................................................................... 67
Spectra Connect PCB Fuses and Electrical Specicaons................................................ 68
Exploded Views and Part Numbers………………………………………………………….………...69
6
7
Unpack the system and inspect it for damage during shipping. Freight damage must be report­ed to the carrier within 24 hours.
Refer to the shipping list for your system to ensure you received all of the components listed.
Do not discard any packaging unl you have found and idened all of the parts. The small
installaon parts are listed on the kit list.
Warning! We will not be held responsible for shortages that are not reported within thirty days of the ship date.
Study the system layout diagram, component photos, and descripons before beginning in-
stallaon.
Lay out the system. Before starng the installaon idenfy where each module and compo­nent will be placed. Ensure that there is enough clearance around the components for remov-
al of lters and system service. Make sure you have adequate tubing and hose before starng.
Addional parts may be ordered.
THE VENTURA 200T IS DESIGNED FOR WARM WATER USE. OPERATION IN WATERS BELOW 500 F (100 C) MAY CAUSE HIGH OPERATING PRESSURES AND INCREASED WEAR ON THE FEED PUMP.
Geng Started
Ventura 150-200T Connect Shipping List:
Ventura Connect kit (includes control box and display)
Accumulator Assembly with analogue gauge
High pressure Clark Pump with membrane pressure vessel
Black high pressure lter housings
Feed pump assembly with fresh water ush module
• Ventura Connect install kit with black product water tubing
Service kit
Two lengths 5/8” Hose (25’)
Oponal: Z-Ion system
Installaon
8
Installaon Basics
Read the direcons!
Avoid ght hose bends and excessive runs.
Use heavy gauge wire.
Install feed pump module as low as possible.
Use a dedicated thru-hull with scoop type strainer.
Do not mount components over electrical devices.
Avoid geng dirt or debris into the piping or hoses during
assembly. A small bit of debris can stop the system!
Thru-hull Locaon: The system must be connected to a dedicated 1/2” to 3/4” forward facing
scoop-type intake thru-hull and seacock.
Install the thru-hull intake as far below the waterline and as close to centerline as possible to
avoid contaminaon and air entering the system. Do not install the intake close to, or down­stream of, a head discharge, behind the keel, stabilizer ns, or other underwater xtures.
Thru-hulls in the bow area are suscepble to air intake in rough condions. Sharing a thru-hull can introduce unforeseen problems such as intermient ow restricons, air bubbles, contami­nants, and will void the warranty. For racing boats and high speed boats traveling above 15 knots, a retractable snorkel-type thru-hull ng is preferred because it picks up water away from the hull.
The brine discharge thru-hull should be mounted above the waterline, along or just above the
boot stripe, to minimize water li and back pressure.
Double clamp all hose connecons below the waterline.
Avoid restricons or long runs on the enre inlet side of the plumbing from the thru-hull to the feed pump module.
Secure the piping away from moving objects such as engine belts and hatches. Prevent chafe on the tubing as required. Test and inspect all piping and hose clamps aer several hours of opera­on.
Pipe Fing Instrucons: To seal plasc-to-plasc ngs, wrap 6 to 8 layers of Teon tape over their threads. Hold the ng in your le hand and ghtly wrap the threads clockwise. For smoother assembly, do not tape the rst (starng) threads.
Wiring
Pay aenon to wire size or system performance will be impaired
Perform wiring to UL, ABYC, CE or applicable standards
Seawater Flow
Thru-hull Not Supplied.
9
Introducon to the Ventura Connect
The Ventura is the finest watermaker for small and midsized yachts. Properly installed and maintained it will provide years of reliable service. Prudent operation is required with any ma-
rine equipment. Always maintain enough reserve water to get safely into your next port.
The Spectra Intensifier, known as the Clark Pump, was introduced in 1997 and has been con­tinually improved since. It is built of modern non-corrosive composites and comes with a 20”
high rejection membrane.
Front View
Pressure Relief Valve
Double rubber mounts
to absorb vibration
The Clark Pump Membrane Module. Pre-mounted and plumbed together as a single unit. Saves time and adds reliability.
Ventura Feed Pump Module
Includes the feed pump, cooling fan, charcoal filter, flush valve, service valve, and service port. The module has compact and streamlined plumbing. The cooling fan is included for longevity.
Fresh Water Solenoid
Service
Port
Service Valve
Cooling Fan
Shurflo
Pump
Note: If your system came with the optional Z-Ion, the Z-Ion unit will replace the char­coal filter housing. The photo above, and all subsequent photos of the Feed Pump Mod­ule, will look slightly different.
Quick disconnect fitting to facilitate maintenance.
10
Components
Feed Pump Module
Mount the feed pump module on a vertical surface, up to 3-feet (1.0M) above the waterline. It is preferable to mount as low as possible. Locate in an area that allows easy access to the char­coal filter, and the service valve. Keep future maintenance in
mind when choosing a location, and do not mount above water
-sensitive equipment.
IF INSTALLING THE Z±ION, SEE Z±ION INSTALL INSTRUCTIONS FOR MORE DETAILS.
Sea Strainer Mount close to the intake through-hull, in a
location that can handle water spillage during service.
Dual Prefilter Filter Bowls Locate in an area that al­lows easy access as these will be the most frequently serviced module while cruising. Mount vertically and leave room below bowls for filter changes. Do not mount above water-sensitive equipment.
11
Double rubber mounts
to absorb vibration
This module must be installed in an area that maintains a temperature below 113°F (45°C). It
may be placed as high in the boat as you desire, and mounted in any position, even upside down. Make sure that the area around and under the pump does not have any water sensitive equipment, as water will be spilled during any repairs or if a leak occurs. Allow for easy access to the pressure relief valve.
The Clark pump and membrane module comes complete with a mounting system. Be sure to
use the supplied washers on the rubber feet.
Pressure Relief Valve
Components continued...
Clark Pump and Membrane
Remote Touchscreen
The remote control panel can be mounted anywhere dry and convenient. Cut a 5-5/8” (12.7 cm) wide by 3-1/4” (7.62 cm) high opening for the panel. The display needs minimum 2
1/2” deep clearance for the cable. Take care not to damage the
plugs on the ends of the cable when routing. Use only a Spec-
tra-approved cable.
12
Ventura Connect Plumbing
Sea Strainer: Mount with included Quick
Block
Feed Pump Module: Mount vertically
as low as practical, no more than 3’ (1M)
above waterline and not over electrical equipment. Leave clearance below for filter changes.
Accumulator:
Factory pre-
charged
Fresh Water Flush inlet to charcoal filter: Plumb to the pressurized side fresh water system.
Ships Freshwater Pressure must be between 20-50psi
(1.4 - 3.5bar)
Spectra Clark Pump and Pressure Vessel/Membrane: Mount in a cool location (below 120 deg. F/49 deg. C). May be oriented in any position and can be well above waterline. Leave access to the pressure relief valve. Do not mount over electrical equipment. Use supplied spacers and washers for the vibration mounts.
1/2” or 3/4 scoop strainer thru-hull and Seacock: Mount low, in a clear flow and
away from head discharge.
Brine discharge thru hull (not in­cluded): place above waterline or tee
into another visible drain.
Note: Brine discharge may be connected to ei­ther side of Clark Pump
Use the supplied 5/8-inch (15.9mm) clear braided vinyl hose for all runs. More hose may be or-
dered from Spectra, or
bought at a hardware or marine store.
Connect Control Box:
Remember to mount in
central location, where various cables can reach all components and the power supply.
20 and 5 micron filters. Do not mount over electrical equipment. Leave clearance underneath for filter changes.
Red In
Green Out
13
Installation Notes
14
Product to tank: Route the product water
into the top of a vented tank, or a tee in the
water tank ll hose. DO NOT feed into a vent line, manifold, or into the boom of the tank. Bladder tanks will create too much back pressure. There must be no restricon or back pressure whatsoever in this piping.
Salinity Probe
Diversion Valve
The 1/4”(6mm) product water tubing is pre-plumbed from the membrane into the electric di-
version valve. The diversion valve will reject product water into the overboard brine stream
unl Connect controller determines the product water is below 750 PPM salinity. Once below 750 PPM, the Spectra Connect control opens the diversion valve and product water is diverted into the fresh water tank.
Product Diversion Valve Manifold
RotoFlow Meter
Accepted Product
Water Outlet
Product from membrane
Check Valve
Product Water Plumbing
Manual Override Buon
Rejected product outlet
15
16
Spectra High Pressure Fitting Instructions
The Ventura has eight high pressure fittings, two on each cylinder on the Clark Pump, two on the pressure vessel end caps, and two 90-degree elbows on the back of the Clark Pump. As the
compression fitting is tightened, it compresses a ferrule onto the stainless tubing, fixing the fer-
rule permanently to the tube and holding the compression nut captive.
The body of the fitting seals to the underlying component with an O-ring. On the Clark Pump cylinders and the end caps this O-ring is compressed by tightening the entire fitting. The O­rings on the 90-degree fittings on the back of the Clark Pump have captive nuts and washers, which compress the O-rings without turning the entire fitting.
If a tube fitting leaks it can sometimes be resealed by just tightening. You must use two wrenches, a 13/16-inch wrench to hold the base, and a 7/8-inch wrench to turn the compression nut. The 13/16-inch wrench will need to be thin so as not to interfere with the compression nut. If this doesnt work, disassemble the fitting, grease liberally with silicone grease (the ferrule and the threads) and re-tighten firmly.
The base O-rings should be gently compressed to achieve a good seal, and may be damaged by overtightening.
Stainless Fitting Hex Nut
Connector O-RING
Nickel-Bronze High Pressure Elbow
Nickel-Bronze High Pressure Straight Fitting
Ferrule
17
Ventura Connect systems are pre-fitted with waterproof connectors for electrical and signal
cables. Each connector is different, so it is impossible to connect them incorrectly, and the con-
nectors are color-coded. The system has a power inlet harness with a terminal block and cover, Pump power cable, pressure transducer red and green cables, a 5-conductor diversion valve and Rotoflow sensor cable, a 2–conductor fresh water flush solenoid cable, a Z-Ion power and sig­nal cable, and a 50-foot (15M) cable for the Connect display.
Do not install it in hot or poorly ventilated locations.
Power inlet harness with terminal block
Spectra Connect Circuit Board (Inside control box)
Run Auto/Manual/Service toggle switch
Electrical
Z-ION Connector (with green band). Left unused if your system was not ordered with the optional Z-ION.
Tank Monitoring Terminal Block
Standard Fresh Water Flush connector (Blue)
Product manifold assembly (Red-connects left of Clark Pump)
Pressure transducer cable. Connect to like red/green sensors atop filter bowl
Feed Pump Power Cable
Tank Full Switch Green + Orange
Tank Level Sensor
5vdc: Red
Signal: White Ground: Black
18
The Ventura Connect has three external ports; one for the display, one for networking, and one for the salinity probe. The Salinity probe is a 5 pin cable with RJ-12 connectors (phone jack) on both ends this cable runs in parallel. The
Display cable is a standard Cat 5e ethernet cable.
Electrical continued
Wire Size Guide for the Ventura 12V: Protect with 15 Amp Fuse or Circuit Breaker
10 Gauge (5mm²) up to 15 feet (4.5M)
8 Gauge (8mm²) up to 25 feet (7.5M)
6 Gauge (13mm²) up to 35 feet (10.6M)
Wire Size Guide for the Ventura 24V: Protect with 7 Amp Fuse or Circuit Breaker
10 Gauge (5mm²) up to 25 feet (7.6M)
8 Gauge (8mm²) up to 35 feet (10.6M)
Note: If the specified circuit breaker sizes are unavailable, use the next higher rating but do not exceed the specification by more then 10%. All wiring to be done to applicable ABYC, Marine UL, or CE standards.
Distances at left represent the total
ROUND TRIP wire length (DC pos-
itive length plus DC negative length), NOT the length of the pair of wires together. Size cables ac­cordingly.
Mount the main power terminal block in a junction box or on a bulkhead adjacent to the feed pump module. Make sure that this is a dry location
well above bilge level and not subject to water spray. Be sure to install
the terminal block cover.
Check the wire size chart for appropriate wire sizes. DC power feeds should be uninterruptible to insure proper operation of the auto store feature. Avoid house breaker panels that could be accidentally tripped.
Diversion Valve and Rotoflow Cable (5­conductor, Red).
Z-Ion power and signal cable (3-conductor, Green)
Fresh water flush sole­noid cable (2-conductor,
Blue)
19
Wiring
Mount the control box with 4 appropriate fasteners on a vercal surface, above components containing water, with the wire grommets down, and central to the other system components.
Make sure the cables will reach all of the modules to avoid splices. The main power feed
should come from an appropriately-sized fuse or breaker on the main DC panel. See wire size and fuse/breaker tables below.
Warning! Do not connect the main power feed unl all other connecons are made.
Filter Sensor harness 6’ (2M)
Clark Pump harness, Rotoow meter, and diversion valve cables 15’ (4.5M). See next page.
Spectra Connect Display. Mount in dry, central locaon. Connects to control cable.
Main power feed. Install cover aer terminang wires.
15’ (4.5M) fresh water ush sole­noid cable. Crimp onto to ush sole­noid leads and heat shrink the connectors. The wires have no po­larity.
Connect Control Box
50’ Control Cable (15.2M): Protect
connector while roung through vessel.
Auto/Manual toggle switch
Pump power feed: 4’(1.2M). May be extended as required. Count addional length in wire sizing.
20
Factory Supplied Tank Full Switch
Installing the supplied tank full float switch at the top of your water tank al­lows the Spectra Connect to fill your water tank then automatically stop, fresh water flush and return to Auto Store mode with no additional user commands.
Drill and tap a 1/8” npt port into the top of the fresh water tank that is being
filled by the watermaker. This can be installed on a tank access cover, or di-
rectly into the water tank. If installing on an access cover, be sure to leave a service loop on the float switch wiring to allow removal of the tank access cover.
For Instructions on Auto Fillmode and installing a Tank Low switch, see Auto Fillmode instructions later in this manual.
Watermakers should never be run unattended.
Your watermaker comes with the top-mounted Tank Full Switch (EL-SWT-LV) and is de­signed to be mounted to the top of the water tank. The side-mounted float switch (EL-SWT­SMLV) can be sourced from your local dealer or Full Service Provider if you cannot access the top of your tank.
If you do not install the Tank Full switch in your tank, you MUST DISABLE the Tank Full Switch in the system settings. See System Settings instructions later in this manual.
Tank Switch Installation
The supplied float switch must be connected to the Green and Orange conductors at the Feed Pump Module, labeled Tank Full Switch. There is no polarity.
Use 18/2 tinned wire or larger. Wire is not included in the installation kit.
Drill and Tap
Connect to Gr/Or wires
21
Optional Tank Low Switch
Installing the tank low float switch about 2/3rds down from the top of your water tank allows the Spectra Connect to run in Auto Fill
mode; automatically filling your water tank, stopping itself, fresh water flushing, returning to Auto
Store mode, and then turning itself on again to fill the tank when the water level drops below the Tank Low Switch with no additional user commands.
Drill 5/8” hole into the side of the fresh water tank that is being filled by the watermaker. This can be installed wherever there in convenient access to the tank, approximately 2/3rds of the way down from the top of the tank.
Drill 5/8” hole
Tank Switch Installation
2/3
The float switch must be connected to the Tank Switch 1 terminals on the Spectra Connect
control board at the Feed Pump Module. There is no polarity.
For Instructions on enabling the Auto Fillmode see the System Settings instructions later in this manual.
Watermakers should never be run unattended!!
Tank Switch 1
22
Z-Ion Installation
The Z-Ion replaces the fresh water ush module on the feed pump module. The Z-Ion lter
bowl, like the normal fresh water ush module, houses the charcoal lter, which must be re-
placed every six months.
The Z-Ion control box comes with four-foot cables for exibility in mounng on the bulkhead above or adjacent to the feed pump module.
Plug the connector from the Z-Ion generator bowl into the connector from the Z-Ion control box.
Connect the power/signal cable, marked green, from the Z-Ion control box to the Z-Ion control cable, also marked green, coming from the Spectra Connect box.
Solenoid valve (opens during fresh water ush/Z-Ion treatment)
Z-Ion control box
Connector to generator bowl (2-pin)
Connector to Spectra Connect control box (3-pin, labeled green)
23
This revolutionary adaptation of an ancient technology effectively and safely protects the membrane
and filters on your Spectra Watermaker from biological growth. Your system will be kept ready to op­erate without any additional flushing, external power sources, pickling chemicals, or complex proce­dures. The controller has been specifically designed to batch process (produce) metallic ions for system disinfection.
The process has been around for many years, however most systems produce ions on a continuous slow
basis
rather than a short intermittent process. We found that the most stable way to produce ions in a batch process is to maintain a constant current to the electrodes and vary the voltage. By being able to boost the volt­age the amount of electrode surface area can be smaller to produce the needed ions in a brief period of time.
The Z-Ion should be energized at all times, but will only consume power when water is running
through it. Upon initial power-up the LED will flash red/green and then will turn solid green.
Follow the instructions for Normal Operation and Fresh Water Flush (for treatment with the Z-Ion, the process is identical, only the Z-Ion will release silver and copper ions into the flush water.)
When fresh water flows, the operation cycle begins and the LED turn off, and only quick flashers of
green and amber. The cycle will continue until either the flush cycle stops or the adjustable timer times out (factory set for 15 minutes).
If the voltage is out of range, below 10V or above 56V, the LED will flash red every two seconds and the unit will shut down.
Each fresh water flush with the Z-Ion will protect your watermaker for up to 30 days, after which the process must be repeated. MPC units with the A40 board have the ability to change the flush interval form 5 day to 30 days. Other MPC boards and chips will need to be upgraded. Contact Spectra about upgrading to the appropriate microchip or board.
After 720 cycles the service light on the front of the control box will light up, indicating that the probes
on your
Z-Ion may be wearing down, and should be tested. The service light is just a reminder that your Z-Ion rods need to be inspected. Before resetting check that rods have not dissolved and are not touching.
To reset the service counter, touch two magnets, at the same time, to the two red reed switches on the Z
-Ion
circuit board, labeled Switch 1 and Switch 2 on the following page.
Z-Ion Operation and Installation
24
Switch 1
There are 7 pins on the phoenix connector from left to right:
Pin 1 Supply Voltage-Battery ( + ) Pin 2 Ground Pin 3 Trigger Pin 4 Auxiliary Output 1 Pin 5 Auxiliary Output 2 Pin 6 To ion generator (bowl). No polarity. Pin 7 To ion generator (bowl). No polarity.
Auxiliary Outputs 1 and 2 are switched to the supply voltage when turned on. These outputs are protected by 100mA self resetting fuses.
Output 1 – ON during generating cycle – intended to drive a pump relay on some systems Output 2 – ON when cycle counter reaches pre-programmed number (usually 720), when generator element may
need replacement and should be tested and/or inspected.
Operation – LEDS
Power-up indication – fast red/green flash for a few seconds Ready, Idle – solid green Generating – the LED will quickly flash at a programmed interval (factory set to 17 seconds between flashes.) The
flash color with alternate between green and amber and will be off in between voltage spikes. The color change
symbolizes the polarity alternating from positive to negative. Bad power – fast red flash followed by shut down High temperature – fast red flash Cycle counter reached limit—slow red flash
Switch 2
Pin 1
Pin 7
Run Light
Will be green when while Z-Ion is
powered and ready
Service Light
Figure 1
Voltage Spike Counter
Z-Ion Layout and Specs.
25
Operaon
26
27
Avoid running the Ventura system if the vessel is in contaminated water. The system should be
fully tested before leaving port. If the locaon or weather prevents proper tesng, refer to Dry
Tesng with an Arcial Ocean on page 34.
Warning! Damage may occur if the purge sequence is bypassed and the membrane is pressurized with storage chemical in it.
First check:
Thru-hull inlet and the brine discharge valves are open.
All of your hose connecons are ght.
The washer with the green tag under the pressure relief valve has been removed.
The pressure relief valve is open 1/2 turn.
The toggle switch on the Connect control box is set to the Auto posion.
There are at least several gallons of water in your fresh water tank for priming.
New System Start-Up and Tesng
Open 1/2 Turn when purging chemicals! Remove Tag and Washer!
3. Turn on the power to the system and the Spectra Connect screen will display, Has the sys­tem been stored with chemicals?Press Yes’, to start the Purge sequence. Note: The wa­termaker will shut down if the pressure relief valve is left closed during the Purge mode.
Alarm will sound
Display will read Has the sys-
tem been stored with chemi­cals?
Confirm the Pressure Relief
Valve is open
Press Yesto start the Purge se-
quence
28
Note: If you must stop the purge sequence for any reason, the control will default back to the
beginning of the purge mode to protect your system.
5. After the purge sequence the display will alarm with the message Close pressure re­lief valve.Close the valve and proceed by pressing Ok to resume the Purge Cycle.
If the system is new from the factory, or stored with Propylene Glycol, additional purging time
may be required.
6. The system will now run under pressure and desalinate water. This mode diverts the product water overboard in case there is any residual chemicals in the membrane. Carefully inspect for leaks over the en­tire system! Shut down the system and repair any leaks
you find.
Check the strainer and the brine discharge for water flow. The system should fully prime with­in 60-90 seconds and all air should be out of the feed water hoses. The feed pump will sound smooth.
There should be no bubbles anywhere in the intake hoses. If the feed pump continues to sound rough, find the reason before continuing! Inspect the system for leaks.
4. The system will start purging and the display will
show the progress and time remaining
New System Start-Up and Testing continued...
7. After the Product Purge cycle completes, the system will prompt to Restart, then advance to the Main Menu. If this is a NEW INSTALLATION, continue to the Calibration In- structions to finalize the installation. If you are putting your watermaker back into ser­vice, your system is now ready for use.
29
Dry Testing with an Artificial Ocean
If it is not possible to test run the system with the boat in the water, you may test the system with an artificial ocean. You will need 1.3 lbs. of non-iodized salt (rock salt, sea salt, or aquarium salt) to make a 5 gallons (33 grams of salt per liter) of water that is about 33,000 PPM salinity (average seawater salinity). Make sure the domestic water system is powered up and the boats tank has at
least 60 gallons (230 Liters) of water to purge the storage chemicals from the system. Confirm
that the charcoal filter is installed in the feed pump module, and the domestic water line is con­nected.
1. Open the pressure relief valve on the Clark Pump. Remove the green tag and spacer.
2. Power up the system. When the Spectra Connect displays “Has the system been stored with chemicals?, select Noto bypass the purge mode.
3. Press the Fresh Water Flush button to run a full flush cycle. Do this six times to purge the
storage chemicals, a 36 minute process.
4. Connect the inlet service hose to the service port on feed pump module, then connect the brine discharge service hose to the quick disconnect fitting on Clark Pump. Refer to the photo be­low. Route both hoses into the 5 gallon (20 Liter) container. Turn the product sample valve to the sample position, and route the product into the bucket.
5. Using the display select fresh water flush, do this until the bucket is filled with fresh water.
6. Mix salt with the fresh water to the proper proportion or use an aquarium hydrometer to ad­just the salinity level.
7. Push the Start button, then press Auto Run and program the Connect to make 10 gal (38 L) of water, roughly 40 minutes of operation.
8. Allow the system to prime and then close the pressure relief valve. The system should build pressure shortly and start making water, with the brine and product water recombining in the
bucket to be cycled again. This will gradually heat the water. Do not let the water temperature
exceed 120 deg. F (49 deg. C).
9. Run the system under pressure, checking for proper operation and leaks. After testing the sys­tem, re-install the brine discharge hose, product tube, and fresh water hose from the strainer. You can now flush the system by pressing the Fresh Water Flush button.
Remove tag
and washer
Open the pressure relief valve
Connect the brine discharge service hose:
30
Sensor Calibration
3. When the Calibration Sequence is complete, press the
button in the upper left corner to return to the Main Menu.
When prompted by the display, Click Save to make
sure that the Calibration is stored in the system memory.
4. The Filter Condition has now been calibrated.
1. During the calibration sequence the system will automatically start, begin to make water for
several minutes and then shut itself down. Make sure that the filters have been replaced for
new, the strainer is clean, and all thru-hulls are open before proceeding.
2. Follow the steps in Figures 14 below to initiate the Calibration Sequence.
Sensor Calibration
Many of the settings on your system have been pre-calibrated during standard factory testing, however, there are a few settings that will vary based on the installation, vessel, and other onboard conditions. If the system has just been installed you must calibrate the Prefilter
Condition before proceeding.
This procedure does not need to be done with each filter change under normal operation, it should ONLY BE DONE IF THE FILTER CONDITION WONT RESET TO 100%
WITH NEW FILTERS AND A CLEAN/NEW STRAINER SCREEN.
Prefilter Condition Calibration
Fig. 1 Fig. 2
Fig. 3 Fig. 4
31
Sensor Calibration
Flush Adjustments
The Ventura Flush Duraon is set to a factory default of 3 minutes, which is usually the right
amount of me to ensure that sea water is thoroughly ushed out of the watermaker using
the least amount of fresh water. However, due to dierent lengths of hose runs, dierent rates of ow, and dierent pressures in shipboard fresh water systems, the ush duraon can be opmized for your boat.
Detailed instrucons about how to access the programming funcon and set the ush dura-
on can be found on page 70, Programming From the Display. Set the Flush Duraon so that the fresh water ush comes to an end just as the salinity of the brine discharge drops below 1000 PPM, or no longer tastes brackish. Since the ush duraon can only be adjusted in round minutes, you may want to lessen the duraon to 2 minutes, to save water, or increase to 4 minutes to ensure a thorough ush.
Also, the charcoal lter is rated for 1.5 GPM (6 LPM): If the system pushes more than 1.5 GPM through the charcoal lter (4.5 gallons in 3 minutes), a ow regulator can be added.
Tank Level Sensor Calibration
1. Press the Menu Button 2. Press the Calibrate Sensors Button
3. Press the Tank Level 1 button
4. Fill the fresh water tank to be monitored.
Fig. 1 Fig. 2
Fig. 3 Fig. 4
Follow the steps in Figures 14 laid out below to enter the calibration sequence for the option-
al Tank Level Sensor(s).
32
Tank Level Sensor Calibration—Contd
5. Measure the approximate height of the water tank in feet and inches (or in meters and cm’s).
6. Press the Feet (Meter) field to enter the tank height in feet (meters). Press the Inch (cm) field to enter the height in inches.
Ex: If the Tank height is 150cm: Enter ‘1’ in the field labeled Meter Enter ‘50’ in the field labeled cm
7. Press Submitand confirm your entry matches your meas-
urement.
8. You must save all changes when prompted after exiting the settings menu.
33
3. Press the Salinity—Product Button 4. Press Continue to acknowledge the warning
5. The system will run for a minimum of 5 minutes to stabilize the product water salinity.
6. Using a calibrated handheld TDS meter measure the salinity of the product water at a sam­pling port, or a convenient location if no sampling valve was installed.
7. Touch the PPMfield and enter the reading taken above.
8. Press Proceedto store the new calibration value. You must save all changes when prompt-
ed after exiting the settings menu
Fig. 1 Fig. 2
Fig. 3 Fig. 4
Salinity Calibration
1. Press the Menu Button 2. Press the Calibrate Sensors Button
The Salinity probe has been calibrated at the factory during tesng and is not normally required during commissioning. If the product quality is not reading accurately, follow calibraon steps.
34
4. Once the system has started and a steady stream of water is owing from the product wa- ter tubing, me in minutes and seconds, how long it takes to ll a container of a known volume.
5. Touch the Gal’ (’Liter’) eld to enter the volume of the container used.
6. Touch the Minuteseld to enter the minutes it took to ll the container. Only enter the
minutes, ex: 3 min 15 sec should be entered as 3.
7. Touch the seconds eld to enter the seconds it took to ll the container. Only enter the seconds, ex: 3 min 15 sec should be entered as 15.
8. Press Proceed’. You must save all changes when prompted aer exing the sengs menu
3. Press the Product Flow Button
The Product Flow sensor has been calibrated at the factory during tesng and isnt normally required during commissioning. If the product ow is not reading accurately, conrm the product ow rate by following the Product Flow calibraon steps.
Product Flow Calibration
1. Press the Menu Button
2. Press the Calibrate Sensors Button
35
Networking
Your Spectra Connect is equipped with state of the art networking options to allow the maxi­mum user control in a wide variety of installations. The instructions below will help you get
the most out of your Spectra Connect.
Connecting to the existing Network
1. Turn power to the system off.
2. Connect a standard Cat5e or Cat6 ethernet cable from jack labeled network to your ships router or networking switch.
3. Turn power to the system back on.
4. Follow the screen prompts below:
5. Note the line Device IPshown in Fig. 3 and record the 10 digit numerical address on the
front of this manual.
Fig. 1 Fig. 2
Fig. 3
Note: Your Spectra Connect is only available when your device is connected to the same local network as the Spectra Connect control board. If you have difficulty connecting to your water­maker control application, double check the that your device network is the same as your Spec­tra Connect
36
8. Your computer should now show the same image as shown on your local Spectra Connect
7. On the computer, tablet or smartphone, open a web browser such as Firefox, Chrome, or Safari. In the web address bar at the top, type the Device IPaddress previously recorded. Press Enter’.
Note: Internet Explorer may not be compatible with your Spectra Connect web app. If
formatting issues occur, use another browser such as Firefox, Safari, or Chrome.
Ex: Address Bar—Firefox
9. Your web browser is now synced with your Spectra Connect. Any buttons you press on your web browser will be controlling your watermaker.
Caution! If operating your watermaker from a computer, phone, or tablet, you must
keep the tab open while the system is in operation and the volume turned up on your de-
vice in order to hear any audible alarm faults.
Connecting to the existing Network—Contd
Ships Router or
6. Connect your computer, tablet or smart phone to the local network your Spectra Connect is plugged into.
Wired Connection: simply plug your computers ethernet port directly into the router or switch where you connected the watermaker. Wireless Connection: make sure your device is connected to the same local wireless network as the Spectra Connect
Wired Connection
Wireless Connections
37
If the system has been pickled or stored with chemicals, use the New System Startup pro­cedure on page 23.
Your watermaker will fresh water flush after every use. Remember that you need to run the system approximately half an hour to make enough fresh water for one flush.
You may notice that the system output is higher when charging your batteries. This is normal.
1. Check to see that the inlet and brine discharge seacocks are open and the domestic pressur-
ized water system is turned on.
2. Press the Startbutton, then select the desired operating mode.
Runs your watermaker until the Tank Full switch closes, fresh water flushes the
system, then goes into Auto Storemode and the Flush Interval timer starts.
This is the default mode of operation.
Gives you the option to run for a preset amount of time, or a preset volume of water to be produced. If no tank switches are installed, and they have been
disabled in the system settings, this is the only Operating Mode available.
Normal Operation
Standard Operating Modes
Enter Volume or Time
Select Gallons/Liters or Hours
Starts the Wa­termaker
38
3. The system will now begin the start sequence. Pressing Stopwill bring you back to the
Main Menu.
4. Once the Boost Pressure reaches the minimum threshold, the system starts operating and you will be taken directly to the Main Dashboard.
Page
Left
Page
Righ
Run Mode (High)
Operating Mode
Current Screen
Diversion Valve Status (CLOSED)
Diversion Valve Status
(OPEN)
6. When the Product Water Quality is below the programmed threshold, the Diversion Valve opens, allowing water to enter the tanks and the screen image changes.
7. Pressing the < (Page Left) or > (Page Right) arrows will show a different screen with differ-
ent operating information for your watermaker.
Verify that the system is operating according to the factory specifications detailed on p.36. See the Troubleshooting section to identify any anomalies.
Normal Operation—Contd
39
To access this information about your watermaker while it is running press the < or > buttons to scroll through the pages.
Nominal Operating Parameters
Parameter Minimum Maximum
Product Quantity 5.7 GPH (21.5 LPH)/Ventura 150
7.7 GPH (29.1 LPH)/Ventura200
6.5 GPH (24.6 LPH)/Ventura 150
8.3 GPH (31.4 LPH)/Ventura 200
Product Quality <300 ppm 750 ppm
Feed Pressure N/A 70 psi (5 bar)/Ventura 150
90 psi (6.3 bar)/Ventura 200
Amp Draw Ventura 150 9.0A @ 14VDC; 4.5A @ 26VDC
Amp Draw Ventura 200 10.0A @ 14VDC; 5A @ 26VDC
Total Flow (Brine + Product) 1.65 GPM (6.2 LPM)
8. When the Run Cycle completes, or by pressing Stop’, the system will start to Fresh Water
Flush. You MUST FRESH WATER FLUSH the watermaker AFTER EACH USE, or se-
rious damage can occur
8. After Fresh Water Flushing the system will enter the Auto Storemode, and the interval timer until the next scheduled flush will show on the screen.
Normal Operation—Contd
40
Other Operating Modes
Run Low Mode
You can toggle back and forth between Run High Mode and Run Low Mode by tapping the Hightoggle button.
Run Low Mode may be selected to reduce power consumption or to lower the feed pres­sure.
Note: The system will automatically drop to Low Mode when it senses high feed pressure,
or dirty pre-filters.
Normal Operation—Contd
Auto Fill Mode
If your vessel is equipped with a Tank Low and a Tank Full switch, and both are enabled in
the system settings, then your Start Menu will allow the system to be operated in Auto Fill
WARNING: NEVER RUN YOUR WATERMAKER UNATTENDED. EQUIMENT CAN BE SERIOUSLY DAMAGED, AND IT IS POSSIBLE TO SINK THE VESSEL!!
Use extreme caution when operating in Auto Fill mode!
In Auto Fill Mode the Spectra Connect will automatically fill your water tank, stop itself, fresh water flush, return to Auto Store mode with the flush interval timer running, and then turn itself on again to fill the tank as soon as the water level drops below the Tank Low Switch with no additional user commands.
Additionally, if power is interrupted at any stage of operation, the Spectra Connect will return
to Auto Fill mode, ensuring that your tanks will always have water.
41
Warning! Proper understanding of the Spectra
flush system and the vessels fresh water system is mandatory for extended use of Auto Store. The
flush cycles must not be allowed to drain all the fresh water from the tank or damage to the vessels systems and the watermaker may occur.
The Auto Store function flushes the watermaker at programmed intervals. As long as the watermak­er is flushed with fresh water every 5 days (30 days with the Z-Ion) you need not store the system
with chemicals.
Make sure there is enough water in the fresh water tanks to supply the watermaker for
more than the expected time of operation in the Auto Store mode. If there isn't enough fresh water in your tank, seawater will be drawn in and the system will not be flushed with fresh wa­ter. The Catalina 340c requires about 7 gallons (26 liters) for each flush. The boats pressure
water supply must be on and stay on while the system is in Auto Store mode. If these condi-
tions cannot be met, then pickling with SC-1 storage chemical or propylene glycol is preferable.
Make sure the pressure relief valve on the Clark Pump is closed.
The system must be continually powered on during the Auto Store mode. Turning off the
power will disable the automatic fresh water flush and damage may occur.
Auto Store
Flushes the system with fresh water, then
activates the flush interval timer:
Normal Operation—Contd
42
Maintenance
General
Periodically inspect the entire system for leakage and chafing. Repair any leaks as soon as you find them. Some crystal formation around the Clark Pump blocks is normal. Wipe down any salt encrusted areas with a damp cloth.
Watermakers are at their best when run regularly. Biological fouling in the membrane is more likely when a watermaker sits idle. A warm environment will cause more growth than a cold envi­ronment. A fresh water flush every five days (30 days with the Z-Ion) will greatly reduce biologi-
cal growth, but may not stop it completely. The Z-Ion system protects the membrane from bio-
fouling without the use of storage chemicals.
The Seawater Strainer
The seawater strainers stainless steel element should be inspected, removed, and cleaned as need­ed. Ensure that the thru-hull is closed before disassembly and the gasket is in place before reas-
sembly. When the system is put into storage, remove the strainer, rinse with fresh water, and reas-
semble dry to impede corrosion. Check frequently during operation.
The Pre-filters
Service the pre-filters as soon as possible after the pre-filter condition graph begins to rise. If the
filter condition graph gets all the way to Replacethe machine will slow down. When display
reaches Replacea second time, the alarm will sound and the system will shut down to prevent damage.
To service the filters close the thru-hull, open the housings, remove the old filters, clean out the housing bowls, and reassemble the housings with new 20 and 5 micron filter elements. The 5 mi­cron filter goes downstream from the 20 micron. Leave dry until next startup.
Use only Spectra-approved filters or you may void your warranty. Occasionally, lightly lubricate the O-rings with silicone grease.
Oil/Water Separator (Optional)
To install oil water separator capability, add a second filter housing UPSTREAM of the 20 and 5
micron housings. Service as you would per the instructions above.
The Charcoal Fresh Water Flush Filter
Replace the charcoal filter element at least every 6 months. This filter protects the membrane by removing chlorine from the flush water. Use only a Spectra-approved replacement.
43
The Feed Pump and Clark Pump
The feed pump and the Clark Pump require no routine maintenance except inspection for leaks. Tighten any hose clamps or fittings that show signs of leakage. The high pressure fittings threaded into the Clark Pump have O-ring seals with a straight thread. These should never leak and should never be over-tightened. If one of the tube nuts starts to leak, it can be un-threaded, sealed with a bit of silicone grease or oil, and tightened with two wrenches very tightly.
The Membrane
Membranes are susceptible to mineral scaling, biofouling and oxidation damage. The leading cause of fouling is biological growth that forms when the system is left unused without flushing or pickling. Fouling from mineral scaling can happen under certain seawater conditions, or from rust. Oxidation damage can occur if the membrane comes into contact with any strong oxidant, such as Ozone, Chlorine, etc. Monitor the product salinity and feed pressure for higher than normal readings, take environmental conditions into consideration.
Note that:
Cold feed water or a higher salinity seawater source can cause high pressure.
Low product flow is usually due to low voltage, a worn feed pump, or worn Clark Pump.
Due to the unique design of your Spectra system, low product water volume is typically not a
membrane problem, but frequently related to low voltage, a worn feed pump head, or a worn
Clark Pump. Always perform a flow test before cleaning your membrane.
Test to see if biological growth has occurred: Before running the system, remove the pre­filters and examine their condition. If the filter housings are full of smelly, discolored water, the system was not properly stored. Install clean pre-filters.
Next check the membrane. Detach the brine discharge hose, attach the brine service hose, and lead it to a bucket. Open the pressure relief valve 1/2 turn, and manually run the system for 30 seconds (metal toggle switch on feed pump module). Examine the brine water: If it is discol­ored and smells bad, perform an SC-2 cleaning with unchlorinated water before running the system pressurized. If the brine is fairly clean, follow the New System Startup procedure on page 24 and run normally. Check for performance. Clean the membranes only if performance is
reduced.
See the Cleaning Procedure for complete instructions.
Maintenance Contd
44
Introducon to Spectra Chemicals
We use four types of chemicals: SC-1, SC-2, SC-3, and propylene glycol anfreeze. SC-1 and propylene glycol are for system storage, while SC-2 and SC-3 are for membrane cleaning. Do not use sodium-bisulfate, citric acid, or any other storage chemical not supplied by Spectra. These chemicals, used to store other watermaker brands, will damage the Clark Pump, mem-
brane end plugs, manifolds, and other components. Using non-Spectra chemicals will void
the warranty.
Note: Never use any chemicals with the system pressurized! Always open the pressure relief
valve 1/2 turn. Always follow the instrucons for purging the chemicals as shown in the New System Startup secon (page 39) of your owners manual.
Storage
SC-1 prevents biological growth when your system is idle. It should not be used as a cleaning chemical, nor will it protect your system from freezing. A jar of SC-1 is mixed with 1 to 2 gal-
lons of product or dechlorinated fresh water in a bucket and circulated through the system for
10 minutes. This treatment will protect the system for six months, aer which the SC-1 treat­ment must be repeated. To use SC-1, follow the instrucons for Storage Procedure.
Spectra systems should be stored with propylene glycol if freezing is likely to occur. Propylene glycol can be used instead of Spectra SC-1 storage chemical for storage in any climate, and treatment is eecve for one year. Propylene glycol is a food-grade anfreeze used to winter­ize RVs, boats, and cabins. Do not use ethylene glycol automove anfreeze, which is toxic and will damage the system.
The propylene glycol formulaons sold in marine and RV stores are usually diluted with water.
The water remaining in the watermaker before the storage procedure will further dilute the
anfreeze, reducing the microbial protecon and increasing the temperature at which the mix­ture will freeze.
Anfreeze labeled Minus Fiyis a 25% soluon and will begin to form an icy slush at about +15Degrees F (-10C) and will only provide burst protecon to about Zero F (-18C). Aer a fur-
ther 50% percent diluon by water remaining in the watermaker, Minus Fiyanfreeze will
only protect from bursng down to about +25F (-4C). Therefore if low temperature freezing protecon is required a 60% or stronger anfreeze should be used. 60% soluons are labeled Minus 100” and will provide burst protecon to -15F (-27C) even aer a y percent diluon with residual water. Minus 200” formulaons are pure propylene glycol.
Maintenance Contd
45
Introducon to Spectra Chemicals—Contd
Complete microbial protecon requires a 25% soluon of propylene glycol, so care must be taken that the soluon remaining in the watermaker during long term storage is at least 25%, even if freeze protecon is not required. For these reasons Spectra recommends that all pick­ling be carried out with a 60% or greater concentraon.
See Winterizing with Propylene Glycol.
Propylene glycol can be dicult to ush from a membrane, especially aer extended storage periods. This results in high salinity water (high PPM) and residual avor in the product water. We recommend ushing the system WITH THE PRESSURE RELIEF VALVE OPEN for 4-6 hours aer storage with propylene glycol—the longer the beer. If, aer extended ushing, you sll
experience low product water quality, cleaning with SC-2 usually removes all traces of propyl-
ene glycol and returns the salinity to the level it was before storage with propylene glycol. See the Membrane Cleaning Procedure.
Cleaners
Avoid unnecessary cleaning, and avoid cleaning as a diagnosc tool.
SC-2 is an alkaline cleaner used to remove light oil, grime and biological growth. It is most eecve if heated to 120 deg. F (49 deg. C). In most cases the water quality will increase in PPM (salinity) aer an SC-2 cleaning. Aer a few hours it should recover to near the level it
produced before the cleaning.
SC-3 is an acid cleaner used to remove mineral and scale deposits. In most cases this is used rst and if there is no improvement, go on to the SC-2 cleaning. SC-3 will in most cases lower the product PPM and overall pressures. Scaling is a slow process that may take several months or years.
For cleaning with either SC-2 or SC-3, see Membrane Cleaning Procedure.
Maintenance Contd
46
Ventura Connect Storage Procedure
1. Turn the yellow service valve on the feed pump module to OFF (horizontal, see photo).
2. Push Auto Store to fresh water ush the system. Repeat, to ush the system twice.
3. Remove the cap on the service port on the feed pump module and install the inlet service
hose from the service kit. Remove the quick disconnect ng from the brine discharge outlet of the Clark Pump, and replace it with the quick disconnect brine discharge service hose. Lead both hoses to a 5 gallon bucket.
4. Push the Auto Store buon and run the feed pump unl you have one gallon of fresh wa-
ter in the bucket from the brine discharge service hose. Stop the system.
5. Mix 1 container of SC-1 storage compound with the water in the bucket.
6. Make sure the pressure relief valve on the Clark Pump is OPEN (un-pressurized) by turning 1/2 turn counterclockwise
7. Place the yellow service valve in the Service posion (pictured below).
8. Turn on the feed pump using the manual con-
trol switch on the Spectra Connect Control Box. The soluon will be drawn from the buck­et with the service hose, and returned to the bucket from the brine discharge service hose. Circulate the storage chemical in the system
for approximately 10 minutes. Turn o the
feed pump.
Clean Up
Remove the brine discharge service hose from the Clark Pump, and replace the brine dis-
charge hose that leads to the thru-hull. You may now pump the bucket dry by using the manu-
al control switch on the Spectra Connect Control Box.
Turn the service valve 180° back to the RUN posion, and remove the inlet service hose. Close the seacock, drain and clean the strainer and any lters in the system. Reassemble dry. Leave the pressure relief valve open, since the next me you run the system you will need to purge
the storage chemicals with the system unpressurized.
Service valve OFF Service hose connected and
service valve to SERVICE
Connecng brine dis-
charge service hose
47
Winterizing with Propylene Glycol
See descripon of propylene glycol formulaons, and ushing from system, on page 44-45.
1. Turn the yellow service valve on the feed pump module to OFF (horizontal, see photos).
2. Push Auto Store to fresh water ush the system. Repeat, to ush the system twice.
3. Connect the inlet service hose to the feed pump module (above the yellow service valve) and lead the hose to the boom of a bucket. Connect the brine service hose, and run it into a sec­ond container.
4. Turn the yellow service valve on the feed pump module to the SERVICE posion.
5. Pour 1 gallon (4 liters) of propylene glycol of appropriate concentraon (see pages 46-47) into
the bucket with the intake service hose.
6. Make sure the pressure relief valve on the Clark Pump is OPEN 1/2 turn (un-pressurized).
7. Run the feed pump using the manual switch on the Connect control box unl about a gallon of water has owed from the brine discharge service hose, or anfreeze appears. Propylene gly-
col will look slightly dierent, and feel more slippery, than water. Stop the pump. Add more
propylene glycol to the intake bucket if necessary.
8. Lead the brine discharge service hose into the intake bucket of propylene glycol. The service hose will now draw propylene glycol from the bucket, and the brine discharge service hose will return it. Run the feed pump and circulate the anfreeze for 10 minutes.
9. Stop the feed pump. Close the seawater intake and turn the yellow service valve to OFF. Drain
the seawater strainer and the hose leading to the feed pump module. Disconnect the product
tubing from the membrane housing and blow residual water out of the tubing. Empty the charcoal lter housing and ush water lines.
Your watermaker is now protected from biological growth and freezing for one year.
Service hose connected and service valve to SERVICE
Connecng brine discharge service­hose
Service Valve OFF
Open the Pressure Relief Valve
48
Membrane Cleaning Procedures
Spectra cleaning compound (SC-2 or SC-3) must be mixed with fresh water at a rao of 1 container of compound to 3 gallons (12L) of unchlorinated water. An average of two gallons (8L) of water is already present inside a Ventura system, so this water must be gured into the mixture. A Ventura system re-
quires one container of compound per cleaning.
1. Turn the yellow service valve on the feed pump module to OFF (horizontal, see photo).
2. Push Auto Store to fresh water ush the system. Repeat, to ush the system twice.
3. Remove the cap on the service port on the feed pump module and install the inlet service hose from the service kit. Remove the quick disconnect ng from the brine discharge outlet of the
Clark Pump, and replace it with the quick disconnect brine discharge service hose. Lead both
hoses into a 5 gallon (20 liter) bucket.
4. Push the Auto Store buon and run the feed pump unl one gallon of fresh water runs into the bucket from the brine discharge service hose. Stop the system.
5. Turn the yellow service valve to the SERVICE posion.
6. Make sure that the pressure relief valve on the Clark Pump is open (un-pressurized).
7. Mix the cleaning chemical in the bucket. If possible, heat the soluon to 120 deg. F (49 deg. C).
8. Start the system using the manual switch on the Spectra Connect control box. The intake ser- vice hose will draw soluon from the bucket and the brine discharge service hose will return it. Circulate the soluon through the system in this manner for 45 minutes.
9. Stop the pump. Replace the brine discharge overboard hose and run the pump unl the bucket
is empty. Stop the pump and turn the yellow service valve back to the RUN posion. Restart
the pump and run for 20 minutes to ush the chemicals out of the system (DO NOT CLOSE the
pressure relief valve!)
10. Move the switch on the Spectra Connect Control Box to AUTO. The system may now be restart- ed, ushed, or stored.
Service valve in OFF posion
Open the pressure relief valve!
Service hose connected and service valve to SERVICE
Aaching brine discharge service hose
49
Suggested Spares for Ventura Connect
Short term cruising, weekends etc.
A basic cruise kit B. This kit consists of three 5 micron lters, three 20 micron lters, and SC­1 storage chemical.
Cruising 2 to 6 months at a me.
Two basic cruise kits, one replacement charcoal lter, and one replacement feed pump
head.
Longer than 6 months
Addional lters, oshore cruising kit consisng of Clark Pump seals, O-rings, tools and
membrane cleaning chemicals. One replacement strainer screen, replacement O-ring for
strainer screen, and replacement O-rings for the lter housings. Spare feed pump or feed pump diaphragm.
Common Parts:
Item Part Number
SC-1 STORAGE CHEMICAL KIT-CHEM-SC1 SC-2 CLEANER KIT-CHEM-SC2 SC-3 CLEANER KIT-CHEM-SC3 BASIC CRUISE B KIT-BCK-B
OFFSHORE REBUILD KIT KIT-OFFSH
5 MICRON FILTER FT-FTC-5 20 MICRON FILTER FT-FTC-20 CHARCOAL FILTER FT-FTC-CC 5” STRAINER SCREEN FT-STN-5S OIL/WATER FILTER FT-FTC-OW
FEED PUMP EL-FP-12V or 24V
FEED PUMP HEAD PL-PMP-SFPH FEED PUMP DIAPHRAGM EL-FP-DP 5” STRAINER O-RING SO-STN-5SS FILTER HOUSING O-RING SO-FHS-10H SALINITY PROBE EL-MPC-SP4 CHARCOAL FILTER HOUSING O-RING SO-FHS-3PCS10
50
Pump runs intermiently, cycling on/o
Overpressure switch on ShurFlo pump opening
Adjust or replace switch (see page 58)
Feed pump runs constant­ly, will not turn o
Turn o manual switch on control box
Manual switch in ON posion on control box
Display acvates, but pump will not run
Loose or broken pump wire connecon
Tanks are full (if equipped
with tank switch)
Check wiring at terminal block inside Spectra Connect
Check tanks– system cannot
be started if tanks are full.
System runs, no product water delivered to water tanks, GPH bar graph shows OK, rejectLED acvated
High salinity of product water, causing system to reject water
Salinity probe out of cali­braon or defecve, bad cable
Chlorine damage to mem-
branes
Pressure relief valve open
Check for low feed pressure
Check for leaks at high pressure
hoses
Test product water with hand-held tester– if over 500 PPM for 1 hour, contact factory
Close pressure relief valve
No lights or display, system does not operate
Remote display not con­nected
Check display cable connecons at back of display and at control box
Check and reset main DC supply breaker
Check for voltage (12 or 24 VDC) at
control box power input studs
Try manual switch on Spectra Connect control box: If pump runs, then con­trol or display is defecve
System runs, no product
water delivered to water
tanks, GPH bar graph shows OK, GoodLED ac­vated
Diversion valve inoperave or wiring fault.
Disconnected or broken product tubing
Diversion valve plunger stuck
Check wiring at diversion valve and inside control box
Check product tubing
Disassemble and clean diversion
valve plunger or replace valve (contact factory)
Feed pump runs with loud noise
Intake blocked
Air in system
Check thru-hull valve
Check sea strainer for leaks
Check fresh water ush module for
leaks
Re-prime system (restart)
Troubleshoong Ventura Connect Systems
SYMPTOMS
PROBABLE CAUSE
REMEDY
51
Salinity High
High product water salinity
Chlorine damage to mem-
branes
Defecve salinity probe or cable, cable disconnected
Check for low feed pressure
Check for leaks at high pres-
sure hoses
Remove and clean probe con­tacts. Check calibraon
Check cable connecons
Clean membrane
SYMPTOMS PROBABLE CAUSE REMEDY
System Stalled
(“system stalledmay alarm
when using the control pan­el to run system for servicing with the pressure relief valve open– use manual override switch instead)
Pressure relief valve open
Intake thru-hull closed
Airlocked system
No signal from Rotoow
meter
Close pressure relief valve
Check thru-hull
Purge air
Clean or replace Rotoow
meter
Check brine discharge
Clean membrane
High Pressure
Blocked brine discharge
Fouled membrane
Voltage Too High” “Voltage Too Low
Baery voltage too high or low
Loose wires or poor connec-
ons
Charge baeries
Check charging voltage
Check power connecons
Re-starng
No signal from Rotoow meter at startup.
System stalled
See remedy above for system
stalled
Check Fuse(followed by
fuse number)
Blown fuse at circuit board
Replace fuse (mini automove type
ATM). See page 82.
Look for cause
Service Prelter
Clogged lters
Loose or defecve pres-
sure sensor wires
Ventura MPC Fault Alarms
Install new filters
Check sensor wiring
With clean filters, recalibrate
Clean Pressure
52
In the event of a sensor failure resulng in a shut down due to a false alarm, the failed sensor can be overridden. High Pressure, Service Prelter, System Stalled (airlock), and Salinity Probe issues can
all be overridden and the system will sll run automacally with all other funcons intact. Be abso-
lutely certain that the alarm is false before overriding the automac controls.
In the event of complete Spectra Connect control failure, the system may be operated manually as follows:
Switch on the feed pump by turning the toggle switch to RUN MAN. The automatic safety con-
trols are disabled in manual mode. Shut the unit down if the Clark pump does not cycle, if air is
continuously present in the intake line, or if the feed pump is excessively noisy.
Always discard the product water for the first few minutes of operation as the initial product
water may not be potable. Take a water sample by loosening the 1/4-inch product tube fitting at either the feed pump module or the membrane outlet. Check it with a handheld salinity meter or taste it.
The diversion valve is an electrically operated three-way valve, which is normally energized by
the Spectra Connect controls in order to send water to the tank. It will not open automatically in manual mode, and must be opened using the mechanical override button. The button is located on the side closest to the Clark pump cylinder. Push the manual override button in and rotate 1/4 turn clockwise to open the valve.
Alarm Override and Manual Operaon
Figure 1: Manual Operaon
Diversion valve manual override button
53
Ventura Flow Test
The ow test is the most useful diagnosc test for system performance, and should be done be­fore replacing or cleaning your membrane. Changes in producon or water quality are normally caused by something other than the membrane, unless the system has been le unused for a
long me.
Before the ow test, change all lters and clean the sea strainer. Carefully check for water or air leaks, as air in the system will cause low producon and errac salinity. Look for air bubbles in the product ow meter, feed water hoses, and brine overboard hose.
Run the system and watch the feed pressure very closely. If the feed pressure to the Clark Pump
is asymmetrical from one stroke to another, this could be part of the problem. A dierence of a few PSI is acceptable, but anything over that is an issue. If the pump is asymmetrical, Clark Pump repairs should be done before connuing with these tests.
If no asymmetry is noted, connue with this test.
Make sure the ShurFlo overpressure cutout switch (PL-PMP-SFPH) is set to 125 PSI. With the pump running, close the brine discharge thru-hull or kink the brine discharge hose. The feed pressure should rise to 125 PSI, then the pump should shut o. If the pump shuts o at a lower pressure see Adjust ShurFlo Pressure Switch on page 45.
You will need a graduated bucket, either a graduated pitcher or large measuring cup, and a stop­watch. Before the flow test, change all filters and clean the sea strainer. Log the voltage at the feed pump at the same time. Confirm at least 12.5 Volts at the feed pump on 12-Volt DC sys­tems; 25 volts on 24-Volt DC systems.
Take two measurements and compare them with the table on the following page. The first
measurement is the product flow alone. The second is the product flow combined with the brine discharge flow to get the total flow or feed flow. You may take these measurements by two methods:
1. Time the product flow into a graduated pitcher, then divert both the product flow and brine
discharge together into a bucket to measure total flow.
OR
2. Divert the product flow into the pitcher while diverting the brine discharge into the bucket. Time the flow of both. After calculating the product flow, pour the pitcher of product into the bucket of brine to measure total flow.
The ratio of product flow to total flow gives us our recovery rate, as a percentage. If the percent­age is below the minimum it indicates an internal leak in the Clark Pump.
54
1. Product Flow: Product flow is expressed in Gallons Per Hour (GPH) or Liters Per Hour (LPH), by this equation:
3600 ÷ time in seconds x quantity of water in gallons or liters=GPH or LPH There are 3600 seconds in an hour.
Example: It took 3 minutes and 35 seconds to collect 1 gallon of product water. 3600 ÷ 215 x 1 = 16.74 GPH (3 minutes, 35 seconds is 215 seconds)
Example: It took 2 minutes and 25 seconds to collect 2.5 liters of product water.
3600 ÷ 145 x 2.5 = 62.07 LPH (2 minutes, 25 seconds is 145 seconds)
2. Total Flow or Feed Flow: Feed flow or total flow (brine + product) is expressed in Gallons Per Mi-
nute (GPM) or Liters Per Minute (LPM) , by this equation:
60 ÷ time in seconds x quantity of water in gallons or liters = GPM or LPM
Example: It took 1 minute and thirty-seven seconds to collect 5 gallons of total flow. 60 ÷ 97 x 5 = 3.09 GPM (1 minute, 37 seconds is 97 seconds) Example: It took 53 seconds to collect 12 liters of total flow. 60 ÷ 53 x 12 = 13.58 LPM
3. Recovery Rate: Product Flow ÷ Total Flow = Recovery Rate % Example: 6.5 GPH product flow = .063 or 6.3%
1.7 GPM total flow x 60 (you must first multiply total flow by 60 to convert from GPM to GPH)
*pressure relief valve open ½ turn
In order to make good quality product water, you need the proper amount of feed water ow, as in the table above. Compare the product ow to the total feed ow. Product ow should be 6.5% mini­mum 5.5%) of total ow for a Ventura 150, and 9% (minimum 8%) of total ow for a Ventura 200T. If product percentage is low, you may have an internal leak in the Clark Pump.
For every 1/
10
th
of a GPM feed water ow loss, we will lose about 1/2 gallon per hour of product ow
and the salinity will go up 100 PPM.
Low feed ow combined with low system pressures is most frequently caused by a worn Shuro
pump head (PL-PMP-SFPH).
System
Feed Static * Feed Flow Product Flow
Pres-
sure
Pres-
sure Flow MIN MIN Flow Flow MIN MIN
psi bar psi gpm lpm gpm lpm gph lph gph lph
Ventura 60-70 4.2-5 10-15 1.7 6.4 1.65 6.2 6.5 24.6 5.7 21.5
VT 200 80-90
5.6-
6.3 20-25 1.7 6.4 1.6 6.0 8.3 31.4 7.7 29.1
55
Poor Product Water Quality
With any product water quality issue, you must ensure accurate calibraon if you are using a salinity meter. For general quality evaluaon, your taste is always good enough.
Membranes are not an exact science and two idencal systems can have dierent product quality. World health standards deem water of up to 1000 PPM of total dissolved solids ac­ceptable for drinking. We consider any thing below 750 PPM acceptable but not ideal, and anything below 500 PPM excellent. Factors that could aect water quality are addressed be­low.
LOW SYSTEM FLOW OR PRESSURE will equate to lower product quality (higher PPM).
Ventura systems, which have a higher feed to output pressure rao (See nominal pressures under Flow Test, page 55), as well as a higher feed ow/membrane area rao, will produce water in the 150-200 PPM range.
DAMAGE TO THE MEMBRANE by chlorine contaminaon. Flushing the system with chlo-
rinated water will irreparably damage the membrane. Charcoal lters are used to ab­sorb any chlorine which might be present in ush water. They must be of proper specicaon to be suitable. See page 56. There is no test for chlorine damage except the process of eliminaon of other causes.
DIRTY OR SCALED membranes. A dirty (foreign material), scaled (mineral deposits), or
contaminated (bacterial growth) membrane can result in poor water quality and ab­normal operang pressures. If operang pressures are above normal, then cleaning is indicated. If the system pressures are within operang normal range, cleaning may have lile result. Avoid cleaning as a diagnosc tool. Low water quality aer storage with propylene glycol can usually be remedied by extended ushing or an SC-2 clean­ing. (See pages 44-45 and 48.)
MECHANICAL LEAKAGE within the membrane pressure vessel. This is an unlikely but pos-
sible cause of poor water quality. A pinched or damaged O-ring within the pressure
vessel, a scratch on the product tube on the membrane, a scratch within one of the
end caps, or a seal fouled by contaminaon could allow sea water into the product water.
If system ow (product plus brine) is 1.5 GPM or above, the membrane is clean, the product ows are consistent with the system ow and the water quality is sll not acceptable, then
replacement of the membrane is indicated.
56
Technical Bullens
The following pages include Spectras most commonly used technical bullens, covering tests, adjustments, troubleshoong, and common points of confusion. Many more technical bulle­ns are available on the Spectra website, www.spectrawatermakers.com.
ACCUMULATOR PRESSURE
Your Spectra watermaker is supplied with a pressure accumulator tank (PL-ACC-TK), which should be installed in the feed water line between the pre-lters and the Clark Pump.
The purpose of the feed line accumulator is to reduce the spikes in the feed pressure caused by the cycling of the Clark Pump. If the accumulator is not properly charged it can lead to problems with the Shuro Pump pressure cutout switch (see page 58). The accumulator has a Schrader air valve, like a car re, which allows the internal air bladder of the accumulator to be pre-charged. The accumulator should be pumped up to about 65 psi (4.5 bar) for best re-
sults. Add air using a re pump or air compressor. You can experiment with the exact pressure
that will give the best pulsaon dampening on your installaon.
Schrader valve
Connecons to feed
water hose, between Shuro pump and Clark Pump
Analogue feed pressure
gauge
57
PREFILTERS
During normal operaon, the feed water is ltered in two stages. First it passes through a ne mesh metal sea strainer, which protects the feed pump from foreign materials and sea crea­tures. Aer passing through the feed pump, the feed water passes the lter housings contain­ing 20 and 5 micron elements, removing very ne parcles that could damage the Clark Pump and shorten membrane life. An addional carbon lter prevents the entrance of chlorine dur­ing fresh water ushing (see next page).
Pre-lter maintenance schedules will vary widely depending on how and where the system is used. If large amounts of feed water are run through the system over a relavely short period of me in biologically ferle near-shore waters, the prelters will plug up, water producon
and quality will drop, and the system pressure will change dramacally. In blue water condi-
ons the pre-lters may only need to be changed every week or two.
When operated for only an hour or two a day in inland or near-shore waters, the trapped plankton will begin to decay in the lters long before the elements plug up. The decaying plankton and bacteria will cause a roen egg smell in the product water. This decay will set in
overnight in tropical waters, or aer a week or two in higher latudes. If handled gently and
changed regularly before they get too smelly, lters can be cleaned several mes. (See Maintenance, page 42.)
Our lter element part numbers are FT-FTC-XX, where the last digits indicate the micron rang. FT-FTC-5 is for a 5 micron element, FT-FTC-20 is a 20 micron element. The oponal oil/ water separator is FT-FTC-OW.
5 Micron
20 Micron Oponal Oil/Water Separator
58
CHARCOAL FILTERS
The charcoal lter element (FT-FTC-CC) removes chlorine from the fresh water ush water supply, as the RO membrane can only handle small amounts of chlorine without permanent damage.
The charcoal lter used for the fresh water ush system will not plug up unless you have very dirty domesc water in your boats supply tank.
The charcoal lter we supply removes 99.7% of the chlorine. Beware when buying other char­coal lters. If they dont specify the percentage of chlorine removed, dont use them. Cheap
ones may remove only 60% or 70%. Also, there are aermarket lters which are very close to,
but not exactly, the right dimensions, and they will not seal in the housing. If you skimp on the charcoal lter you risk damaging a $600.00 membrane on the rst ush. The other factor is the ow rate that the lter can handle. Because the chlorine is adsorbed by the charcoal, it must remain in contact with the charcoal for a sucient period of me for the all of the chlo­rine molecules to be captured. The lters we use can handle 1.5 gallons (6 liters) per minute
ow, and are good for 3000 gallons (12,000 liters) at 1.5 GPM, or six months, whichever
comes rst. Regardless of the amount of water treated, the charcoal loses its eecveness aer six months.
Charcoal lter, Spectra part number FT-FTC-CC
59
SHURFLO PUMP WONT RUN
If the pump has power to it but the pump wont run, rst check the pressure switch. The pres­sure switch (EL-FP-PS) is located on the wet end of the pump and has two red wires plugged into it (see photo, next page). Jump the two red wires together and see if the pump runs. You
can safely run the system with the pressure switch jumped, just keep an eye on the feed pres-
sure and dont let system pressure exceed 110 PSI. Replace the switch when a spare is availa­ble. The pressure switch should never open unless there is a problem with the system or it is incorrectly adjusted. See Adjust Shuro Pressure Switch on next page.
If the pump will not run with the pressure switch jumped then it is most likely a problem with
the brushes or overheat protecon switch inside the motor. The motor will come completely
apart by removing the two screws on the end of the motor. Remove the rear cover and paper insulator. Pull out the plasc brush holder. The thermal switch is located on one of the brush leads. With an ohmmeter, check for connuity through the switch. If it is open, you can make temporary repairs by wiring around it, being careful that your new wiring doesnt chafe on the moving parts, nor resist the springs that push the brushes on to the commutator. The overheat switch is unlikely to fail unless the motor has overheated. Consider relocang the pump or im­proving venlaon if the overheat protecon has failed.
If any corrosion is apparent the brushes may be scking. Once apart clean all the carbon dust from all the parts. Clean the commutator with light sandpaper. Make sure to clean the small
grooves on the commutator with a small sharp tool to remove the carbon in between the seg-
ments. Adjust the springs on the brush holders so the brushes slide smoothly in and out. If the bearings are rough and binding, remove the rubber dust cover to clean them, grease them, and work them free by hand. Don't service the bearing unless absolutely necessary. Reassem­ble in reverse order. You can hold the carbon brushes back with papers clips inserted through the slots in the brush holder so they don't hang up on the bearing during assembly. Make sure
the corrugated bearing shim doesn't push out. If it does, push it back into place.
60
ADJUST SHURFLO PRESSURE SWITCH
Shuro feed pumps are equipped with a high pressure cutout switch (EL-FP-PS). This is the small black unit on the end of the wet end of the pump head (PL-PMP-SFPH) where the two red wires connect. If the pressure switch is not properly adjusted the pump may cut out each
me the Clark pump cycles and the feed pressure spikes. When this happens the producon
will drop and salinity will increase. The points in the switch will fail quickly if set too low be­cause of the constant arcing each me the Clark Pump shis.
On the very center of the switch is a small 5/64” Allen screw. While running the system close
the brine discharge seacock or kink the discharge hose, to block the ow. Watch the pressure
gauge and adjust the pressure switch to shut o at 125 psi. Turn the Allen screw clockwise to increase the cut o set point.
Pressure Switch Adjusng Screw
61
Spectra Connect Settings
Your new Spectra Connect is designed to make your watermaker easier than ever to operate, maintain and enjoy. This section will guide you through some of the more advanced settings options available.
Always use caution when changing any factory default settings, as serious damage can oc-
cur.
The Spectra Connect automatically monitors the operation of the system to ensure a long and trouble-free service life. If an operating parameter changes, the Connect can switch operating modes, shut itself down, or automatically store itself in order to protect your watermaker.
It includes advanced calibration sequences to make proper setup and maintenance of your wa-
termaker easier than ever.
The onboard time clock feature allows for temporary power interruptions without detrimental effects on the system. In some cases your watermaker will continue to function in its last known operating state.
The Spectra Connect has built in data logging, allowing for easy access to historical operating data—which can indicate a wearing component or spares to be carried along before a failure occurs.
Built in warnings for preventative maintenance automatically alert a user of pending mainte­nance items, helping to keep your watermakers up-time to 100%! Advance warnings are pre­programmed for Prefilter Life, Clark Pump rebuilds, feed pump heads, membranes, Z-Ion reac-
tor rod life, and carbon filter life. These warnings are resettable, allowing you to perform the
maintenance before a catastrophic failure, then reset the interval—so youre always on top of the maintenance cycle!
62
LCD Brightness: Set brightness of the hardwired display(s) from 10100% System Units: Change from US Standard units to Metric Factory Reset: Resets any changed parameters a user has made back to the factory defaults for that configuration.
CAUTION!! Never disable a Fault Alarm without being certain that the issue is with a bad sen-
sor. Disabling a fault and running the system can cause serious damage or injury.
High Pressure Fault: Disables the High Pressureshutdown fault in the event of a feed pressure sensor failure.
System Stall Fault: Disables the System Stalledshutdown fault in the event of a failed rotoflow sensor. System
stalled alarms occur when the control board does not sense any product water being produced, and shuts down to protect the pump from running dry.
Poor Quality Fault: Disables the High Salinity’ shut down fault in the event that the salinity probe has failed or cannot be calibrated within range. NOTE: The diversion valve will always be active when this fault is disa-
bled. ALWAYS VERIFY PRODUCT QUALITY BEFORE DRINKING. Serious health risks may occur.
Prefilter Fault: Disables the Service Prefiltershut down fault in the event that the boost pressure sensor has failed or cannot be calibrate within range. CAUTION: Permanent damage to the feed pump can occur if this fault is disabled, use caution when operating this system with this fault disabled.
User Settings
Spectra Connect Settings, Contd
Fault Alarms
63
Settings
Dealer Access Settings
It is highly recommended that users consult with a factory trained technician before altering any settings behind the
Dealer Access Point’. Changing this settings without understanding the full effect of each change can void the warranty of your system, and cause irreparable damage.
If any settings are inadvertently changed, they can be reverted back to the defaults by using the Factory Resetfeature.
Default Login: admin Default Password: admin
System Model: Configures the Spectra Connect for a different system model from a preset list of options. Clear Statistics: Resets all of the Estimated Maintenance Intervals back to 100%. This feature should only be
used on a brand new system.
Change Username/password: Changes the default username and password. If you forget your changed
username and password, a Factory Reset will revert back to the default username and password. Set MFD: Changes the Manufactured Date on the system. This should only be adjusted if a control board is being replaced on an older system. Set Serial ID: Changes the Serial Number recorded in the Spectra Connect. This should only be adjusted if a control board is being replaced on an older system. Limit Runtime: Limits the maximum run time for the system before shutting down and fresh water flushing. Disabling this setting allows the watermaker to be operated 24/7.
64
Settings
Dealer Access Settings
It is highly recommended that users consult with a factory trained technician before altering any settings behind the
Dealer Access Point’. Changing this settings without understanding the full effect of each change can void the warranty of your system, and cause irreparable damage.
If any settings are inadvertently changed, they can be reverted back to the defaults by using the Factory Resetfeature.
Purge Setup: Adjusts the time and maximum feed pressure allowed for the Purge Mode. CAUTION: Perma- nent damage to the membrane can occur if this setting is adjusted. Consult the factory before making any adjustments. Fresh Water Flush Settings: Allows adjustment of the fresh water flush duration and the interval between flush-
es. If the Z-Ion is installed, the Flush Interval should be changed to 30 days.
Conductivity Setup: Allows for enabling or disabling conductivity sensors on the feed water and product water.
Set the threshold for the diversion valve to divert water to the tanks.
Flow Setup: Allows the user to adjust the flow sensor settings, or disable a flow sensor circuit altogether. DO NOT USE THIS SETTING TO CALIBRATE THE PRODUCT FLOW. Follow instructions on calibrating the
flow sensor in this manual. Tank Level Monitors: Enable and disable the Tank Level Sensors, which read the % remaining in the tank, and the tank switches, which allow the system to turn on/off automatically. Boost Pressure Setup: Enable alternate Boost Pressure sensors, change the Low Vacuum Limit, or Boost Pres-
sure Setpoint. CAUTION: Permanent damage to the pump can occur if this setting is adjusted. Consult the
factory before making any adjustments. Low Vacuum Limit: The minimum boost pressure required at the inlet to the pump. This setting pre vents the pump from getting damaged by running under high vacuum. Adjusting it to a lower number in creases the risk that the pump will suffer damage during normal operation. Boost Pressure Setpoint: During startup the controller turns on the boost pump and waits for the Boost Pressure to reach the Boost Pressure Setpoint. If the boost pressure fails to reach this setpoint, then the main pump wont turn on. Reducing the Boost Pressure Setpoint may cause the system to start, then
immediately shut down due to low boost pressure.
Outlet Pressure Setup: Set High Pressure Limit, enable alternate high pressure sensors, select pressure sensor scaling. CAUTION: Permanent damage to the pump can occur if this setting is adjusted. Consult the facto-
ry before making any adjustments. Network Setup: Enabling the Spectra Connect Wireless access turns on a Power Over Ethernet feature on the wired connection. ENABLING THIS FEATURE CAN CAUSE SERIOUS DAMAGE TO YOUR SHIP’S NETWORK. DO NOT ENABLE THIS FEATURE WITHOUT CONSULTING A QUALIFIED TECHNI­CIAN OR THE FACTORY.
65
Settings
Dealer Access Settings
It is highly recommended that users consult with a factory trained technician before altering any settings behind the
Dealer Access Point’. Changing this settings without understanding the full effect of each change can void the warranty of your system, and cause irreparable damage.
If any settings are inadvertently changed, they can be reverted back to the defaults by using the Factory Resetfeature.
Automatic PRV: Enables an optional Automatic Pressure Relief Valve, after it is installed. This setting should
remain off unless you are certain that you have this feature installed on your system.
Power Sensor: Enables or disables an optional power sensor, after it is installed. PH/ORP: Enables an optional pH or ORP meter, after it is installed. Z-Ion: Enables or Disables the optional Z-Ion system, after it is installed. If the Z-Ion is enabled, you should also
adjust the Flush Interval to 30 days.
Tank Level Monitors
Tank Setup - Enable/disable tank sensors.
Enable Tank Switch High - Enable/disable tank switch high. If this is disabled Auto Fill and Fill Tank run modes will not be available. Enable Tank Switch Low - Enable/disable tank switch low. If this is disabled Auto Fill mode will not be available. Both High and Low tank switches must be enabled for Auto Fill mode.
Enable Tank Level 1 - Enable/disable tank level sensor 1. If this is disabled there will be no tank level
reading and tank level gauge will read “!”. Enable Tank Level 2 - Enable/disable tank level sensor 2..
66
Fault Handling
Faults are hazardous conditions that might occur during running of your watermaker. The control board has the
ability to monitor these faults in real time and take necessary actions to prevent damaging your equipment.
HIGH PRESSURE FAULT
High pressure fault is triggered if
Outlet pressure (Feed/Membrane pressure) > Pressure Limit
If a high pressure fault is triggered, the system goes to low production mode if it is running in high production
mode, or stops the operation if it is already running in low production mode. Then the system will begin the Auto Store mode.
Resolutions
Check for kinked or blocked hoses. Confirm “#3 Sensor PSI High limitand “#3 Sensor PSI Offsetoptions on Outlet Settings. Clean membrane.
SYSTEM STALL FAULT
System stall fault is triggered if
There is no product flow for 1 minute continuously.
If system stall fault is triggered, machine will stop the current run cycle and will prompt to restart. If restarted it
will retry the previous running mode. If the stall condition persists even after restart, the system will begin the Auto Store mode.
Resolutions
Confirm product water at membrane endcap. Check intake line for restrictions, blockages or air leaks. Close Pressure Relief Valve on Pump.
Confirm controller settings correct.
SERVICE PREFILTER FAULT
The Service Prefilter fault is triggered if
Filter reading between the two sensors has a difference greater than 10 psi.
Resolutions
Change prefilters. Check sensor for proper operation
Unplug and unscrew sensor. Gently check for gunk or debris with paper clip or wire.
POOR QUALITY FAULT
The High Salinity fault triggers if
The Salinity of the product water is above the threshold (measured salinity > Salinity 1 threshold) for
more than 8 minutes.
If the High Salinity fault is triggered, the machine will stop the current run cycle and will prompt to restart. If
restarted it will retry the previous running mode. If the High Salinity fault condition persists even after
restart, the system will begin the Auto Store mode.
Resolutions
Check pump operation - Clark Pump (pressure relief valve closed), Feed Pump (moving water). Confirm product water quality. Membrane damage - clean or replace. Salinity probe out of calibration. Clean or replace salinity probe.
67
Tank 1
Tank 1 Tank 2
Tank 2
68
Electrical Specifications
Operating voltage limits: 1032vdc
Outputs: 5VDC:– 2000mA Total for all outputs.
VLV 1&2: 1 Amp DC priming valve solenoid. AUX 1-3: 1 Amp DC auxiliary output.
Never connected anything to the Spectra Connect without factory approval. Serious dam­age can occur!
69
Exploded Views and Part Numbers
70
71
Top View
Rear View
Plate Bracket
FM-PVB-PB
Rubber Mount
HD-RBP-RM Plasc Spacer
HD-SPN-MKINS
Plate Bracket End Cap
FM-PVB-PBE
Pressure Vessel End Ring
FT-PV-ER
Part Numbers
Nickel-Bronze High Pressure
Elbow
PL-MTE-3/4S1/2B
Pump to Pressure Vessel Upper SS Tube:
PL-TB-HP-HPTVUS
Pump to Pressure Vessel Upper SS Tube:
PL-TB-HP-HPTVUS
3/8”NPT x 3/8” Tube Fing JG
PL-MTS-3/8x3/8J
Pressure Vessel End Cap
KIT-PV-25EP1
72
Sea Strainer (Lid & Bowl& Screen)
FT-STN-5
Sea Strainer Screen
FT-STN-5S
Sea Strainer O-ring
SO-STN-5SS
1-Way Solenoid Valve 12V, 24V
PL-SLN-1/4O12SV, O24
1/4”NPT X 5/8” Hose Barb El.
PL-HBE-1/4X5/8
3/4”NPT X 5/8 Hose Barb El.
PL-HBE-3/4X5/8
Pressure Sensor
EL-SSR-150
Filter Housing
Bracket Duplex
FT-FHB-10HD
3/4” Nipple
PL-NP-3/4N
High Pressure Filter Housing (Lid & Bowl)
FT-FTH-10H
Filter Housing O-ring
SO-FHS-10
Accumulator Tank
PL-ACC-TK
Pressure Gauge
PL-PSG-LP2.5
1/2” X 1/8” Bushing Reducer
PL-BSH-1/2X1/8N
1/2” Nipple
PL-NP-1/2N
1/2” Female Tee
PL-TEE-1/2FN
1/2”NPT X 5/8 Hose Barb
PL-HBS-1/2X5/8
3/8”NPT X 5/8” Hose Barb El.
PL-HBE-3/8X5/8
Part Numbers
SUB-CT-Z-HPILHA CT 340 Z Quick Disconnect
73
3/8” NPT x 3/8” Tube ST JG
PL-MTS-3/8X3/8J
1/8” Hex Plug
PL-HP-1/8
Part Numbers
EL-MPC-SP4
SALINITY PROBE
PL-MNF-DVMBK1P
DIVERSION VALVE MANIFOLD
EL-SSR-IFMLR
INLINE FLOW METER SENSOR
PL-CKV-1/4M-F
1/4" PLASTIC CK VALVE (M-F)
PL-SLN-1/4D12SP 1/4" DIVERSION VALVE 12V/24V
REDUCING COUPLING
PL-BSH-3/8X1/4
Pump to Pressure Vessel
Lower SS Tube: Not shown
PL-TB-HP-HPTPVLS
PL-MTE-1/4x3/8J
1/4”x3/8TUBE FITT. ELBOW
PL-FTE-1/4X1/4P
1/4"FPT X 1/4"TUBE FITT ST.
PL-MTE-1/4x1/4J
1/4”x1/4TUBE FITT. ELBOW
PL-SLN-1/4D12M
DIVERSION VALVE SOLENOID
1/4” Product Tube
PL-NLT-1/4LP
KIT-SC-RMPNL SPECTRA CONNECT REMOTE
CNTRL.
DISP.4.3" (REMOTE, comes with 100’ of cabling)
74
Pump Head Assembly W/Press. Switch
PL-PMP-SFPH
1/4” Product Tube
PL-NLT-1/4LP
3/8” Low Pressure Tube
PL-NLT-3/8LP
Part Numbers
Stainless Fing Hex Nut
PL-HWR-1/2HN
1/2” Stainless Ferrule
PL-HWR-1/2FR
Connector O-RING
SO-HPP-CT
3/4”-16 Straight Thread
O-RING
SO-FT-STF
3/4”X 1/2” Tube Fing Nickel-Bronze Elbow
PL-MTE-3/4S1/2B
75
Clear Filter Housing (Bowl & Cap) FT-FTH-10L3PCS Clear Filter Housing O-ring
SO-FHS-3PCS10
3/4”NPT X 1/2” Hose Barb Elbow PL-HBE-3/4X1/2
3/4”NPTX5/8Hose Barb PL-HBS-3/4X5/8
3/4” 3-Way Valve
PL-VLV-3W3/4
Cooling Fan 12V, 24V KIT-FK-12,24
Feed Pump Heat Sink
EL-FP-FPHS
Ventura Feed Pump Bracket FM-VT-ITM
Feed Pump 12V, 24V EL-FP-12V, 24V
3/8”NPT X 5/8” Hose Barb El. PL-HBE-3/8X5/8
Feed Pump Pressure Switch EL-FP-PS
Part Numbers
1-Way Solenoid Valve 12V, 24V PL-SLN-1/4O12, 024
EL-TS-5000-CTRL
KATADYN-5000
CONTROL BOARD
EL-SWT-TG-DPDT
TOGGLE SWITCH DP/DT
FM-MPC-MPCB
MPC BOX
EL-SWT-TSBHC
TOGGLE SWITCH BOOT SEAL
EL-RL-40A12 12V, 40A CUBE RELAY or EL-RL-40A24 24V, 40A CUBE RELAY
-Voltage specific, one per system, not shown
76
HP-TB-VEB-B
HP-TB-VB
HP-TB-VEB-A
HP-TB-BV
HP-CB-CB7
HP-CYL-SST
HP-CYL-R
HP-CYL-EC
HP-CYL-CCA
Not used
77
HD-CPS-5/16X3
KIT-HP-10VSA
HP-TB-ARP
SO-HPP-AR1
SO-HPP-RV
HP-TB-BV
HD-CPS-5/162.75
SO-HPP-SP
HP-TB-VSP
HP-TB-SR
SO-HPP-VB
SO-HPP-PLP
SO-HPP-VP
78
Parts
HP-CYL-3/4R,
79
PL-MTS-3/8X1/2B
HP-CYL-SST
HP-CYL-CCA SO-HPP-ECCB
HP-CYL-EC
HP-CYL-R
HP-CYL-PT
Parts
80
Loading...