SpectraPure MC-RODI-UHE-400 Instructions Manual

Page 1
MaxCap-400 Ultra-High Efficiency RO/DI
(MC-RODI-UHE-400)
Please read carefully before proceeding with installation. Failure to follow any
attached instructions or operating parameter may lead to the product’s failure and
REV. 02-17-2010
INSTRUCTIONS
WARNING
possible damage to property.
Page 2
SpectraPure
®
2
Thank You for your purchase of a
provide you with high quality water for years to come. All SpectraPure® products are rigorously tested by us for safety and reliability. If you have any questions or concerns, please contact our customer service department at
1.800.685.2783 or refer to our online troubleshooting at www.spectrapure.com.
SpectraPure® System. With proper installation and maintenance, this system will
TABLE OF CONTENTS
Operational Specifications
Contents of Drinking Water Systems .... ...................................... ................................................ ............................................. 3
Operation Description (how the system functions) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Preparation
Choosing the best Mounting Location.................................................................................................................................... 5
Tools recommended for installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ............ 5
Introduction to push fittings . . . . . . . .. . . . . . . . . . . . . .. . . . . . . . . . . . . . .. . . . . . . . . . . . . . .. . . . . . . . . . . . . . .. . . . . . . . . . . . . . .. . . . . . . . . . . . . . .. . . . . . . . . . . . . . .. . . . . . . . . . . . . . .......................... 5
Getting to know your system . .. . . .. . .. . . .. . .. . .. .. .. .. . . .. . .. . . .. . .. . .. .. .. .. . . .. . .. . . .. . .. . . .. . .. . . .. . .. . . .. . .. . .. .. .. .. . . .. . .. . . .. . .. . . .. . .. . . .. . .. . .. .. ........................ 6
Set-Up Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .............................................. 7
Installation of floats for the Pure Water Flush Cycle Tank . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Pressure adjustment with regulator . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . . .. . . . ............ 8
Adjusting Concentrate to Purified Water ratio . .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. . .. .. .. .. .. .. .. .. 8
Re-installing DI resin cartridges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..................... 8
Connecting Bulk Head fittings ... .. .. ... .. .. .. ... .. .. .. ... .. .. .. ... .. .. ... .. .. .. .. ... .. .. .. ... .. .. .. ... .. .. .. ... .. .. ... .. .. .. ... .. .. .. .. ... .. .. .. ... .. .. .. ........................ 9
Final Connection............................................................................................................................................................................. 1 0
Maintenance Regimen
Metering & Diagnostics . .. . .. .. . .. .. . .. .. .. . .. .. . .. .. . .. .. . .. .. .. . .. .. . .. .. .. . .. .. . .. .. . .. .. .. . .. .. . .. .. . .. .. .. . .. .. . .. .. . .. .. .. . .. .. . .. .. .. . .. .. . .. ................................... 1 1
Shelf life for filters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .............................................. 1 1
Sediment Pre-Filter replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .................. 1 1
Carbon Block Pre-Filter replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .......... 1 2
Reverse Osmosis Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ........................ 1 2
Reverse Osmosis Membrane Performance Testing ... .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. 1 3
Reverse Osmosis Membrane replacement . . . . . . . . .. . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . .. . 1 4
Tips for long membrane life . .. . .. . . .. . . .. . . .. . . .. . . .. . .. . . .. . . .. . . .. . . .. . . .. . .. . . .. . . .. . . .. . . .. . .. . . .. . . .. . . .. . . .. . . .. .. . .. . . .. . . .. . . .. .. . .. . . .. . . .. . . .. . .......................... 1 4
De-Ionization Cartridge maintenance and replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 5
Troubleshooting Guide................................................................................................................................................................ 1 6
Warranty Information .... .... .... ... .... .... ... .... .... ... .... .... ... ... .... .... ... .... .... ... .... .... ... .... .... ... .... ... .... .... ... .... .... ... ....................................... 1 7
Replacement Parts List ... ... ... ... .. ... ... ... .. ... ... ... .. ... ... ... ... .. ... .. ... ... ... ... .. ... ... ... .. ... ... ... .. ... ... ... .. ... ... .. ... ... ... ... .. ...................................... 1 8
Copyright © 2009 by SpectraPure Inc.
ALL RIGHTS RESERVED
No part of this publication may be reproduced, stored in a retrieval system, or transmitted in any
form or by any means, electronic, mechanical, photocopying, recording or otherwise without
the prior written permission of SpectraPure Inc.
SpectraPure®Inc.
2167 East Fifth St, Tempe, Arizona 85281
480.894.5437 Call us toll-free 1.800.685.2783
Page 3
SpectraPure
®
3
OPERATIONAL SPECIFICATIONS
RO Feed water requirements
Tap Water Pressure: 30 PSI. pH Range: 3-11. Max. Temperature: 100°F (38°C). Max. Feed Turbidity: 1.0 NTU. Max. Silt Density Index: 5.0 (based on 15 min. test time). Maximum Chlorine: < 0.1 ppm Maximum TDS: 2000 ppm. Maximum Hardness: 10 grains (170 ppm as CaCO3). Maximum Iron: < 0.1 ppm. Maximum Manganese: < 0.1 ppm Maximum Hydrogen Sulfide: 0 ppm. Langlier Saturation Index (LSI): must be negative.
NOTE: MOST MUNICIPAL WATER SUPPLIES MEET THE ABOVE REQUIREMENTS. IF WELL WATER IS USED, PLEASE MAKE SURE THAT YOU OBTAIN A WATER TEST BEFORE INSTALLATION.
CONTENTS OF UHE WATER SYSTEM
UHE System Components
(1) 5-Stage System
- (1) Membrane (300 GPD)
- (1) 20”, 0.2 Micron Zeta-Zorb Sediment Filter
- (1) 20”, 0.5 Micron Carbon Block
- (1) 20”, MaxCap DI Cartridge
- (1) 20”, Silica-Buster DI Cartridge
- (1) High Flow Re-Circ Pump
- (1) UHE Module
- (1) Pressure Regulator
TDS METER #1
Recirc Pump
Sediment Filter (SF-ZZ-0.2-ABS-20)
- (1) Liquid Filled Pressure Gauge
- (2) Dual Probe In-Line TDS Meters (1) High & Low Float Switches (with 6ft extension cable &
transformer) (1) Filter Wrench (2) Bulk-Head Fittings
Solenoid Valve Controlled
TDS METER #2
300 GPD RO MEMBRANE
Silica-Buster DI Cartridge (DI-MC-20)
3/8” BLACK LINE: Tap Water input 1/4” BLUE LINE: Product Water 3/8” YELLOW LINE: Waste Water 3/8” BLUE LINE: Pure Water Flush
Float Switches
Carbon Filter (CF-0.5-20)
IF ANY OF THE ITEMS LISTED ABOVE ARE MISSING PLEASE CONTACT SPECTRAPURE PRIOR TO INSTALLATION.
ALL RETURNS WITHOUT RMA# WILL BE REFUSED. CLAIMS MUST BE WITHIN 10 DAYS FROM RECEIPT.
SpectraPure®Inc.
UHE SYSTEM DIMENSIONS: 24 W x 34 H x 8 D
MaxCap DI Cartridged (DI-MC-20)
480.894.5437 Call us toll-free 1.800.685.2783
2167 East Fifth St, Tempe, Arizona 85281
Page 4
SpectraPure
®
4
OPERATION DESCRIPTION (how the system functions)
1. First, the incoming feed water (3/8” Black line) is passed through a 0.2 micron Zeta-Zorb sediment pre-filter. This filter is required to remove excess turbidity (particulate matter) that may cause the carbon block filter to plug.
2. The second stage of filtration is a 0.5 micron carbon block pre-filter. This filter removes organics and chlorine from the feed water that can damage the membrane.
3. The third filtration stage of the system is a high rejection thin film composite (TFC) reverse osmosis membrane. It removes over 98% of most inorganic salts, all micro-organisms and organics above 100 diatoms molecular weight.
4. The fourth and fifth stage filter are our These two cartridges remove the remaining silicates, phosphates, and nitrates left by an RO membrane.
High and Low Float Operation:
1. Floats are to be mounted horizontal when dry and float up when wet. (NOTE: Pressure switches could also be used with pressure tanks. Please call for more details)
2. When the bottom float goes dry, the input solenoid (INSOL) is opened and one second later, the Re-Circ pump (PUMP) turns on.
MAXCAP DI
™ cartridge followed by our SilicaBuster ™ cartridge.
Flush Cycle (flush cycle consumes only a couple of quarts):
1. As the system is running, a periodic sequence of flushes is performed.
2. When the top float gets wet, or every 8 hours of uninterrupted operation, the flush cycle is initiated.
3. After the flush cycle, the system shuts down if the tank is full or pauses for an hour to let the pure water soak the membrane before continuing.
SpectraPure®Inc. assumes no responsibility for water damage due to leaks. It is
the user’s responsibility to determine that the system is leak-free.
SpectraPure®Inc.
480.894.5437 Call us toll-free 1.800.685.2783
2167 East Fifth St, Tempe, Arizona 85281
Page 5
SpectraPure
®
5
PREPARATION
1. CHOOSING THE BEST MOUNTING LOCATION:
LIGHT SOURCE: Algae is more likely to thrive inside clear housings when exposed to ultraviolet light and other
sources like metal halide lighting. Avoid installing the unit in bright light or direct sunlight.
EXTREME TEMPERATURES: The unit MUST be kept out of areas that are subject to extreme temperatures like
freezing or temperatures greater than 100° F (38° C)
FRESH WATER RESERVOIR: Mount system in an area that will have enough room for a small trash can or reservoir.
This Fresh Water Reservoir is to be used for the Flush Cycle. (Reservoir needs to hold at least 10 gallons)
ELEVATION FOR SYSTEM: When mounting your UHE, make sure it is elevated above your fresh water reservoir:
USE DELIVERY PUMP TO DELIVER WATER WHERE NEEDED FOR SALT MIXING OR TO SUMP
GRAVITY DRAIN WATER FROM ELEVATED TANK TO TRASH CAN FOR SALT MIXING OR TO SUMP
2. TOOLS RECOMMENDED FOR INSTALLATION:
- Adjustable Wrench
- Sharp Exacto Knife
- Open End Wrench(s)
- Electric Drill
-5/8” Drill bit
-13/16” Drill bit
- Phillips Screw Driver
- Trash Barrel or Reservoir
- Optional Ball Valve or Float Valve. (as shown above)
- Optional Delivery Pump (as shown above)
3. INTRODUCTION TO PUSH FITTINGS:
Push fittings are similar to a Chinese finger trap toy. In order to release its grip from tubing,..
- Firmly depress and hold the push-fitting collar down with your thumbnail.
- While the push-fitting collar is depressed, pull the tubing straight out of the push-fitting. Once the tubing is removed, release the collar.
SpectraPure®Inc.
480.894.5437 Call us toll-free 1.800.685.2783
2167 East Fifth St, Tempe, Arizona 85281
Page 6
SpectraPure
®
6
4. GETTING TO KNOW YOUR SYSTEM: I
it is very important that you rinse the MaxCap DI cartridge with the downstream DI removed. There is a large ionic load present in a new MaxCap DI cartridge. Failure to flush this cartridge, will shorten the life of the down stream DI.
f you are setting up this system for the first time, or replacing either DI cartridges,
fresh water reservoir.
NOTE: When mounting your UHE,
make sure it is elevated above your
SpectraPure®Inc.
2167 East Fifth St, Tempe, Arizona 85281
TOP VIEW
480.894.5437 Call us toll-free 1.800.685.2783
Page 7
SpectraPure
®
7
SET-UP PROCEDURES
STEP 1: Unscrew the third position housing (MaxCap DI) and the fourth position housing (Silica-Buster DI) from the system and remove the DI Cartridges. Then, re-install the empty housings back onto the system.
STEP 2: Connect the 3/8” Black Line to the cold water line by using either the garden hose adapter, which is provided with the system, or a Feed Water Adapter.
Feed Water Adapter
Garden Hose Adapter
STEP 3: PURE WATER FLUSH CYCLE INSTALLATION USING THE HIGH & LOW FLOAT SWITCHES:
A. Install both the High and Low Floats onto a open container like a Rubber MaidTM trash can. The floats can be mounted by drilling a 5/8” hole. (The High Float shuts down the system, while the Low Float turns the system on). B. Connect the floats to the “Y” adapter. C. Attach the 3-pin connector on the float switches to the HI-REC MODULE.
NOTE: Floats are to be mounted horizontal when dry and float up when wet. In most instances we recommend locating the low float approx. half way down your reservoir. DO NOT OVERTIGHTEN when installing the floats.
“Y” adapter
NOTE: Floats have to be installed in a FRESH WATER storage container.
NOTE: This system MUST be connected to the floats to operate the Flush Cycle. Failure to install the pure water flush connection for this unit will void any warranty.
STEP 4: For now, direct both the 3/8” blue (product) line and the yellow (waste) line to drain.
STEP 5: Open the cold water line, where the black line is connected to. Plug in the 24 V transformer to the UHE Module and connect to power.
SpectraPure®Inc.
FRESH WATER
480.894.5437 Call us toll-free 1.800.685.2783
2167 East Fifth St, Tempe, Arizona 85281
Page 8
SpectraPure
®
STEP 6: At this stage, the system will start producing water out of the 1/4” blue and 3/8” yellow lines. Discard that water for approx. 15-20 min.
membrane. This is normal.)
(The recirc pump produces a loud noise due to the pump size and flow rate for
8
STEP 7: PRESSURE ADJUSTMENT USING THE PRESSURE REGULATOR:
(show in picture)
A. The Pressure Regulator, when pushed down, is in the lock position. Pull out on it to release lock. B. While the system is producing water, turn the Pressure Regulator clockwise to increase pressure or
counter clockwise to decrease pressure. Set the pressure to 100 PSI. Then, re-lock the pressure knob.
PRESSURE REGULATOR
STEP 8: ADJUSTING THE CONCENTRATE TO PURIFIED WATER RATIO:
If not enough concentrate is allowed to leave the system during operation, impurities will precipitate out on the membrane surface, clogging the RO Membrane. To keep this from happening, the Concentrate to Purified Water Ratio must be checked and adjusted in order to compensate for pressure and temperature variations that exist in all water supplies.
to the product flow rate (1:1). NOTE: If your tap water TDS reading is below 500 PPM, you can adjust the waste: product ratio to 1:2.
a. Hold both 1/4” blue and 3/8” yellow lines side-by-side and observe the flow rates.
b. Slowly adjust the waste flow adjustment valve (W.F. Adj, shown on page 6) until the flows appear to be equal. NOTE:
c. Get two measuring cups and a stop watch to measure the waste and product water. Make sure you have equal measurements and adjust accordingly.
any waste flow. This will damage the membrane and void any warranty.
Note: NEVER close the W.F. Adj valve, completely blocking
The waste flow can be slightly less than the product.
The flow rate of the concentrate should be equal
This procedure will assure you of maximum life and reliability of your SpectraPure System. Failure to perform this procedure can permanently damage the membrane and will void the pro-rated Membrane Warranty.
STEP 9: After you adjust the flow rate, follow STEP 7 again, if needed, to assure that the system is operating with 100 PSI.
STEP 10: RE-INSTALLING AND FLUSHING DI CARTRIDGES:
A. Disconnect power to the system and shut down the water supply line.
B. Unscrew the third clear housing and install the MaxCap Cartridge.
NOTE: at this time leave the fourth clear housing empty.
C. Turn on supply line and plug power back to system. Allow water to run through the MaxCap Cartridge,
and down the drain, until the second meter (POSITIONED “IN”) reads “0”.
D. Then, follow step A and place the Silica-Buster in the fourth clear housing.
E. Finally, turn the supply line back on, plug the power back in and allow the water to run through the
Silica-Buster Cartridge, and down the drain, until the second meter (POSITIONED “OUT”) reads “0”.
SpectraPure®Inc.
480.894.5437 Call us toll-free 1.800.685.2783
2167 East Fifth St, Tempe, Arizona 85281
Page 9
SpectraPure
®
Note: At least 2 gallons (7.57 liters) of purified water will be discarded during the flushing before you can collect purified water for use.
Upon start-up, air may be trapped in the DI cartridges (housing may not appear full), this is a normal condition and it will not affect the operation of the DI system.
9
METER #1 IN: Tap Water OUT: RO Water
METER #2 IN: MaxCap Water OUT: Silica-Buster Water
STEP 11: CONNECT BULKHEAD FITTINGS:
(Bulkhead Fitting MUST be installed on the same container that the floats are on, SEE DIAGRAM ON PAGE 7)
1/4” BULK HEAD FITTING: Use a paddle bit and carefully drill a 5/8” hole for installation at the top of the reservoir (above the high level float).
3/8” BULK HEAD FITTING: Use a paddle bit and carefully drill a 13/16” hole for installation at the bottom of the reservoir (below the low level float).
SpectraPure®Inc.
480.894.5437 Call us toll-free 1.800.685.2783
2167 East Fifth St, Tempe, Arizona 85281
Page 10
SpectraPure
®
10
STEP 13: FINAL CONNECTION:
1. Direct the waste line (yellow ) to a drain or connect it permanently to a drain pipe via a drain saddle. (Drain Saddle is not included with system) Look for leaks and do not leave the system alone until you are sure there are no
leaks. Monitor the system for 72 hrs.
OPTIONAL INSTALLATION FOR WASTE WATER LINE (USING A DRAIN SADDLE)
Yellow tubing from system
CHECK BEFORE LEAVING SYSTEM UNATTENDED:
- Ensure that all fittings are tight and leak-free before leaving the system unattended.
- The concentrate line (yellow) includes a smaller capillary tube (flow restrictor) that is located “inside” of the tubing. Do not remove or discard this restrictor; the system will not produce permeate water without the flow restrictor.
- Downstream of the flow restrictor is a Waste Flow Adjustment Valve fine-tune the waste water. NEVER close it off completely, as that will damage the membrane.
- The following conditions must be met before water will flow through the system: The water source must be turned “ON” and the water level in the reservoir tank must be below the low level float, if present.
SpectraPure®Inc.
480.894.5437 Call us toll-free 1.800.685.2783
2167 East Fifth St, Tempe, Arizona 85281
(W.F. Adj shown on page 6)
used to
Page 11
SpectraPure
®
11
MAINTENANCE REGIMEN
Maintenance and trouble-shooting procedures have been made easy and effective with a combination of the (2) Dual TDS meters and a built-in pressure gauge. It is recommended that you keep replacement cartridges on hand, ready to install as soon as the monitor indicates that the cartridge in use is exhausted.
METERING AND DIAGNOSTICS:
The digital TDS meter (left side) will provide a reliable means of evaluating the efficiency of the RO membrane. This meter will indicate the tap water conductivity (IN PROBE) and the RO water conductivity (OUT PROBE). The difference between the two meter readings will be used to calculate the percentage rejection of the (TFC) RO membrane. (as shown on page 12)
The Digital TDS meter (right side) will be used to determine the condition of the two DI stages. When the reading on the (IN PROBE) reads 50% of the first meter’s OUT reading, then the MaxCap Cartridge needs to be replaced. When a reading on the (OUT PROBE) is higher than 1 PPM, replace the Silica-Buster Cartridge
SHELF LIFE FOR FILTERS:
PRE-FILTERS (Sediment & Carbon Block Filters): Pre-Filters have an infinite shelf life.
REVERSE OSMOSIS MEMBRANE (RO:) Purchase an RO membrane when needed. If you would like to keep one on hand, you can store it in the refrigerator for up to 1 year. (Keep in its sealed, original packaging). If the membrane dries out, it will not produce water and any warranty will be voided).
DE-IONIZING CARTRIDGES (DI) Replacement DI cartridges come shipped in a Mylar® Bag. Keep them in their sealed, original packaging until use. You can store it for up to 1 year.
SEDIMENT PRE-FILTER REPLACEMENT:
For maximum contaminant removal and long membrane life, the sediment pre-filter must be changed when the system fails to maintain the pre-set pressure of 100 PSI. If your water contains a great deal of sediment the pre-filter may require more frequent changes to maintain adequate production rate and extended membrane life.
Materials Required: 0.2-micron Zeta-Zorb™ Sediment Filter (SF-ZZ-0.2-20), Filter Wrench.
Zeta-Zorb sediment filter for replacement will void the system’s warranty.)
Procedure: (FIRST FILTER ON LEFT)
1. Turn off water supply to the system.
2. Using the provided filter wrench, remove the first housing on the left. Unscrew it counterclockwise as viewed from the bottom.
3. Remove the old filter and discard.
4. Thoroughly wash the housing with a mixture of hot soapy water and a few teaspoons of household bleach. Rinse well with clean hot water.
5. Install the new pre-filter into the housing, Screw the housing back onto the assembly, and hand tighten
only. NOTE: Do not use filter wrench to tighten housings. Over-tightening will damage housings and void your warranty.
6. Proceed with carbon block filter replacement.
(Failure to use a
SpectraPure®Inc.
480.894.5437 Call us toll-free 1.800.685.2783
2167 East Fifth St, Tempe, Arizona 85281
Page 12
SpectraPure
®
CARBON BLOCK PRE-FILTER REPLACEMENT: For maximum contaminant removal and long membrane life, the Carbon pre-filter must be changed when the system fails to maintain the pre-set pressure of 100 PSI.
Materials Required: 0.5 micron Carbon Block Filter (CF-0.5-20), Filter Wrench, Chlorine Test Kit (TK-CL-10)
Procedure: (SECOND FILTER ON LEFT WHEN FACING SYSTEM)
1. Perform steps 1-5 on page 11.
2. Turn on system’s water supply and check for leaks.
REVERSE OSMOSIS MEMBRANE DIAGNOSTICS: Although RO membranes are capable of maintaining high water quality over extended periods of time they eventually will begin to deteriorate. Normally, the conductivity of the permeate water will increase as the membrane ages. By comparing the difference in TDS readings between the Tap water conductivity and the RO water conductivity, the percentage of rejection of the RO membrane may be calculated and the resultant number may then be used to determine the condition of the membrane. Thus, the operator will know when the membrane needs to be replaced. Membrane fail­ure will be indicated by a reduction in the percentage of rejection which will be determined by calculating the differential between the input and output numbers.
In order to accurately determine the condition of the RO Membrane, a conductivity tester (TDS meter) capable of reading the tap water conductivity and the permeate water conductivity has been provided with this system. With the assistance of the first (meter #1) Dual TDS meter, you will be able to easily determine the RO membrane’s condition.
Before performing the diagnostic test on the RO membrane, make sure that the RO system has been “ON” and producing pure water for a minimum of 10 minutes. Also check the brine (yellow) line to make sure that water is flowing and that the flow ratio between the brine water and the permeate water is at a ratio that is between 1:1 and 1:2. (NOTE: The pressure gauge should indicate a pressure reading of 100 PSI during this 15-20 minute test period.
12
Procedure:
1. Turn on the left side meter by depressing the on button.
2. Locate the meter slide switch on the front of the DM-1 TDS meter.
3. Slide the switch to the left “IN” and read the Tap water conductivity then record the reading ________ .
4. Next, slide the switch the to right “OUT” and read the permeate water conductivity then record the reading ________.
5. Take the two reading and apply them to the calculation below:
(X - Y) = Z TAP WATER TDS = X
_______________
X
EXAMPLE: (500 - 20) = 480 / 500 = 0.96 which is 96% rejection
6. The general rule to knowing when to replace your Reverse Osmosis Membrane (RO) is if your % rejection is below 95%
RO WATER TDS = Y
= % REJECTION
BY MEMBRANE
SpectraPure®Inc.
480.894.5437 Call us toll-free 1.800.685.2783
2167 East Fifth St, Tempe, Arizona 85281
Page 13
SpectraPure
®
13
REVERSE OSMOSIS MEMBRANE PERFORMANCE TESTING: Actual performance of a membrane is dependent on three factors: Osmotic Pressure, Tap Water Pressure and Tap Water Temperature.
What is Osmotic Pressure: The flow of water through a membrane in response to differing concentrations of solutes on either side generates an additional pressure across the membrane called osmotic pressure. In other words, as the hardness of the tap water increases, greater tap water pressure is required to force the water through the membrane.
Pressure & Temperature Compensation: In order to characterize a membrane, there must be common testing data. Industry standards to testing membrane performance is: 60 PSI tap pressure with 77 degree water temperature.
Most locations will not have exactly 60 PSI or 77 degree water. Use the calculation below to determine what the membrane will do with your water conditions.
OSMOTIC PRESSURE CORRECTION:
DOW FilmTec’s Standard for Total Dissolved Solids in the water is 250. Your first step is to subtract what your water is from DOW’s standard.
Next Step, subtract your current house pressure with the osmotic pressure. This is your corrected tap water pressure.
PRESSURE COMPENSATION:
TEMPERATURE COMPENSATION:
Temperature Correction Factor Table (TCF)
F / C TCF
41.0/5 0.521
42.8/6 0.540
44.6/7 0.560
46.4/8 0.578
48.2/9 0.598
50.0/10 0.620
51.8/11 0.640
53.6/12 0.661
55.4/13 0.684
57.2/14 0.707
F / C TCF
59.0/15 0.730
60.8/16 0.754
62.6/17 0.779
64.4/18 0.804
66.2/19 0.830
68.0/20 0.857
69.8/21 0.884
71.6/22 0.912
73.4/23 0.941
75.2/24 0.970
Now, take you corrected pressure and divide it by the standard PSI set by the industry. (Which is 60 PSI)
Multiply the rated membrane size by your corrected compensation value.
Look up the water temperature on the chart below to determine TCF value.
F / C TCF
77.0/25 1.000
78.8/26 1.031
80.6/27 1.063
82.4/28 1.094
84.2/29 1.127
86.0/30 1.161
87.8/31 1.196
89.6/32 1.232
91.4/33 1.267
93.2/34 1.304
Finally, multiply the Pressure Compensation value by the TCF value to give you your actual gallons per day with your specific water conditions.
SpectraPure®Inc.
CALCULATING EXAMPLE
480.894.5437 Call us toll-free 1.800.685.2783
2167 East Fifth St, Tempe, Arizona 85281
Page 14
SpectraPure
®
REVERSE OSMOSIS MEMBRANE REPLACEMENT:
1. Turn off the water supply to the RO system and unplug the transformer. Place the system where the membrane housing is easily accessible.
2. Remove the black tubing from the membrane feed push-fitting by depressing the collar on the fitting with your thumb and pulling the tubing from the push-fitting.
3. Lift the membrane housing from the retention clips.
4. Unscrew the membrane housing lid. This may require two people.
5. Use a pair of pliers to grasp the membrane stem and pull the membrane from the housing.
6. clean water.
7. Insert new membrane into the housing, with the double O-ring end first. The tube must fit into the recess at the bottom of the membrane housing. When the membrane is aligned with the hole, firmly push the membrane into the hole until it bottoms out.
8. Place the black housing O-ring on the housing rim and carefully screw the lid back on to the base.
9. Reconnect the black tubing to the membrane feed push-fitting.
Remove the black housing O-ring. Wash the empty housing
with soapy water. Rinse thoroughly with hot,
14
Removing the
Membrane Element
Black membrane housing O-ring
Black
membrane
brine seal
TIPS FOR LONG MEMBRANE LIFE:
1. Replacement of 0.2 micron sediment filter at least once every 6 months. This will prevent membrane fouling due to silt or sediment depositing on the membrane.
2. Replacement of .5 micron carbon block filter at least once every 6 months or when chlorine breakthrough occurs. This will ensure good membrane life and protect the membrane from chlorine damage.
3. Membrane should not be operated at lower than the recommended concentrate to purified water ratios.
4. Operating reverse osmosis systems on softened feed water greatly reduces the chances of membrane fouling.
Membrane product water tube
Inserting the New
Membrane Element
Double O-ring end of membrane product water stem tube
SpectraPure®Inc.
480.894.5437 Call us toll-free 1.800.685.2783
2167 East Fifth St, Tempe, Arizona 85281
Page 15
SpectraPure
®
DE-IONIZATION CARTRIDGE MAINTENANCE AND REPLACEMENT
Maintenance for MaxCap DI Cartridge:
Procedure:
1. When the reading on the right-hand TDS meter (set to “IN”) displays 50% (or greater) of the reading on the left-hand TDS meter (set to “OUT”), it is time to replace the
2. Follow the directions on page 8 to replace the
3. Make sure the DI cartridge is installed in the correct direction as marked on the cartridge shell and be sure that the top seal is securely attached to the top of the cartridge.
4. Tighten the cartridge housing by rotating it clockwise and hand tighten.
5. Turn on system and check for leaks.
Maintenance for Silica-Buster DI Cartridge:
1. When the reading on the right-hand TDS meter (set to “OUT”) displays “001” (or greater), it is time to replace the SilicaBuster ™
2. Follow the directions on page 8 to replace the SilicaBuster ™
3. Make sure the DI cartridge is installed in the correct direction as marked on the cartridge shell and be sure that the top seal is securely attached to the top of the cartridge.
4. Tighten the cartridge housing by rotating it clockwise and hand tighten.
5. Turn on system and check for leaks.
Procedure:
DI cartridge.
MAXCAP DI
MAXCAP DI
™ cartridge.
DI cartridge.
™ cartridge.
15
REPLACEMENT PROCEDURE FOR DI CARTRIDGES:
Materials Needed: Filter wrench and DI cartridges.
Procedure:
When replacing the MaxCap Cartridge:
A. Disconnect power to the system and shut down the water supply line.
B. Unscrew the third clear housing and place in a new Cartridge. Discard the old one in the trash. NOTE: at this time leave the fourth clear housing empty.
C. Turn on supply line and plug power back to system. Allow water to run through the MaxCap Cartridge until you get a reading coming out of the second meter (POSITIONED “IN”) to “0”.
When replacing the MaxCap Cartridge:
D. Follow step a-b and place the Silica-Buster in the fourth clear housing.
E. Turn the supply line back on, plug the power back in and allow the water to run through the Silica-Buster Cartridge until the second meter (POSITIONED “OUT”) reads “0”.
NOTE: When replacing DI cartridges, ALWAYS flush the MaxCap with the down stream cartridge removed.
SpectraPure®Inc.
480.894.5437 Call us toll-free 1.800.685.2783
2167 East Fifth St, Tempe, Arizona 85281
Page 16
SpectraPure
®
TROUBLE SHOOTING GUIDE
Problem Cause Corrective Action
1. Low production rate. a. plugged pre-filters. i. Replace pre-filters.
16
b. low water
temperature.
e. fouled membrane. v. Replace membrane.
f. plugged flow
restrictor.
2. Zero production rate. a. Dried membrane. i. Replace membrane.
b. Power supply not
plugged in
c. Both floats need to
be dry
3. Extremely high production rate.
4. Milky colored water. a. Air in system. i. Air in the system is a
5. The water in the Pure­Water container drains out when not used.
6. Cannot adjust waste water flow
a. Ruptured membrane. i. Replace membrane.
a. The system is not
elevated higher than the Pure-Water container
a. W.F adjustment valve
in lock position
ii. Use higher GPD
membrane.
vi. Replace flow restrictor &
membrane.
ii. Plug power in.
v. Both floats need to be
dry and horizontal, not drooping
normal occurrence with initial start-up of the RO/DI system. This milky appearance will disappear during normal use within 1-2 weeks. If condition reoccurs after filter changes, drain tank 1 to 2 times.
i. Elevate the system
higher than the Pure­Water container to avoid siphoning.
i. Unsrew lock nut on W.F
valve
SpectraPure®Inc.
b. W.F ajustment valve
stripped
480.894.5437 Call us toll-free 1.800.685.2783
2167 East Fifth St, Tempe, Arizona 85281
Replace
Page 17
SpectraPure
®
17
THREE YEAR MANUFACTURERS WARRANTY
Effective on products purchased after March 10, 2005.
SpectraPure, Inc.® warrants the product to the original owner only to be free of defects in material and workmanship for a period of three years from the date of receipt. SpectraPure’s liability under this warranty shall be limited to repairing or replacing at SpectraPure’s option, without charge, F.O.B. SpectraPure’s factory, any product of SpectraPure’s manufacture. SpectraPure will not be liable for any cost of removal, installation, transportation or any other charges which may arise in connection with a warranty claim. Products which are sold but not manufactured by SpectraPure are subject to the warranty provided by the manufacturer of said products and not by SpectraPure’s warranty. SpectraPure will not be liable for damage or wear to products caused by abnormal operating conditions, accident, abuse, misuse, unauthorized alteration or repair or, if the product was not installed in accordance with SpectraPure’s or other manufacture’s printed installation and operating conditions, or damage caused by hot water, freezing, flood, fire or acts of God.
SpectraPure will not be responsible for any consequential damages arising from installation or use of the product, including any water or mold damage due to flooding which may occur due to malfunction or faulty installation, including, but not limited to failure by installer to over- or under-tighten fittings, housings, and/ or push-style fittings, or improper installation of push-style fittings. Consumable items such as pre filters and membranes are not covered under the two year warranty.
SpectraPure warrants (pro-rated) the performance of tested SpectraSelect™ RO membrane elements only, for one year from date of receipt by the buyer, providing that the loss of performance was not caused by fouling , neglect or water conditions exceeding the feed water parameters listed in the applicable product manual (refer to detailed membrane warranty information). SpectraPure will, on confirmation of loss of performance during the warranty period, credit the pro-rated amount of the current catalog price of the element. The disposable filters and cartridges are not covered under the warranty.
To obtain service under this warranty, the defective system or components must be returned to SpectraPure with proof of purchase, installation date, failure date and supporting installation data. Any defective product to be returned to the factory must be sent freight prepaid; documentation supporting the warranty claim and a Return Goods Authorization (RGA) number must be included. SpectraPure will not be liable for shipping damages due to the improper packaging of the returned equipment and all returned goods must also have adequate insurance coverage and a tracking number.
SpectraPure will not pay for loss or damage caused directly or indirectly by the presence, growth, proliferation, spread or any activity of “fungus”, wet or dry rot or bacteria. Such loss or damage is excluded regardless of any other cause or event that contributes concurrently or in any sequence to the loss. We will not pay for loss or damage caused by or resulting from continuous or repeated seepage or leakage of water, or the presence or condensation of humidity, moisture or vapor, that occurs over a period of 14 days or more. “Fungus” and “fungi” mean any type or form of fungus or Mycota or any by-product or type of infestation produced by such fungus or Mycota, including but not limited to, mold, mildew, mycotoxins, spores, scents or any biogenic aerosols.
SpectraPure will not be liable for any incidental or consequential damages, losses or expenses arising from installation, use, or any other causes. There are no expressed or implied warranties, including merchantability or fitness for a particular purpose, which extend beyond those warranties described or referred to above.
* The three year limited warranty does not apply to consumable items, including but not limited to,
filters and cartridges unless specifically stated above
SpectraPure®Inc.
480.894.5437 Call us toll-free 1.800.685.2783
2167 East Fifth St, Tempe, Arizona 85281
Page 18
SpectraPure
®
REPLACEMENT & OPTIONAL PARTS
Replacement Parts
No. Catalog No. Replacement Parts Description
18
1. SF-ZZ-0.2-20 Sediment Filter 0.2 Micron Sediment Filter (1st Stage)
Zeta-Zorb sediment filter for replacement will void the system’s warranty.)
2. CF-0.5-20 Carbon Filters 0.5 Micron Carbon Pre-Filter (2nd Stage)
3. MEM (300) RO Membrane Replacement Membrane (3rd Stage)
4. DI-MC-20 MaxCap DI MaxCap De-ionizaiton Cartridge (4th Stage)
5. DI-SB-20 Silica-Buster DI Silica-Buster De-ionizaiton Cartridge (5th Stage)
6. FR (25-90) Flow Restrictor 90 gallon a day Flow Restrictor
7. VA-CK-IL-1LB-4GJ Check Valve 1 LB Check Valve
8. PUMP-ROT-IM-140-SS High-Flow Re-Circ Pump Re-Circluation Pump
Optional Accessories
No. Catalog No. Description
1. PUMP-DEL-5800-1 Delivery Pump, GPM flow rate at 20 PSI = 1.45
2. V-FLOAT-4-SS-PP Ball Float with SS arm and built in bulk-head fitting
(Failure to use a
SpectraPure®Inc.