SpectraDynamics HPDA-100RM-A Operating Manual

SDI
SPECTRADYNAMICS, INC.
HIGH PERFORMANCE DISTRIBUTION AMPLIFIER
HPDA-100RM-A
OPERATING MANUAL
SPECTRADYNAMICS, INC 1849 Cherry St. Unit 2. Louisville, CO 80027
www.spectradynamics.com
1
Description
SDI
SPECTRADYNAMICS, INC 1849 Cherry St. Unit 2. Louisville, CO 80027 Phone: (303) 665-1852 Fax: (303) 604-6088
Se habla español
The HPDA-100RM-A is a high performance distribution and isolation amplifier. It contains one module that provides five unity gain outputs. Typical cross channel isolation and reverse isolation is typically greater than 100 dB. The phase noise of the modules is exceptionally low, typically –145 dBc/Hz @ Fourier frequency of 1 Hz and –173 dBc/Hz @ Fourier frequency greater than 10 kHz. Both the input and output are matched to 50 ohms to obtain better than 25 dB return loss. All outputs are AC coupled and the grounds are DC isolated to reduce the effect of ground loops.
Copyright © SpectraDynamics, Inc. 2010
2
Safety and Preparation for Use
SDI
CAUTION!
Voltages capable of causing injury or death are present in this instrument. Use extreme caution whenever the instrument cover is removed.
Line Voltage
This instrument may be setup to operate on 100-120 or 220-240 VAC and a line frequency of 50 to 60 Hz. The setup voltage for this HPDA-100RM-A is specified on
page 4. Fuse
A 2.0 Ampere 250V slow-blow fuse is used for 100-120 VAC operation. A 1.0 Ampere 250V slow-blow fuse is used for 220-240 VAC operation. Only replace fuses with the same type and specifications.
Line Cord
The instrument has a detachable, three wire power cord for connection to a grounded power source. The enclosure of the unit is directly connected to the outlet ground to protect against electrical shock. Always use an outlet with a protective ground and do not disable this safety mechanism.
Service
Do not attempt to service or adjust the instrument unless another person, capable of providing first aid or resuscitation, is present. Contact SDI for any questions or repairs.
SpectraDynamics, Inc. 1849 Cherry Street Unit 2. Louisville, CO 80027 USA
Tel: (303) 665-1852 Fax: (303) 604-6088 support@spectradynamics.com www.spectradynamics.com
Copyright © SpectraDynamics, Inc. 2010
3
The Front Panel
SDI
AC Power LED
The LED labeled POWER is on when power is applied to unit and the unit is operating properly.
Signal LED
The LED labeled SIGNAL will turn on when you apply an input signal to the SMA input connector greater than +7 dBm.
Copyright © SpectraDynamics, Inc. 2010
4
The Back Panel
SDI
AC POWER ENTRY MODULE
The HPDA-100RM-A is configured to operate on:
100-120 VAC 220-240 VAC
SMA INPUT
The signal to be distributed should be connected to the SMA jack labeled INPUT.
SMA OUTPUTS
The five outputs of each amplifier module are DC isolated from the chassis ground to prevent ground loops. Make sure that the amplifier ground does not float to a potential greater than 50 VDC from the chassis ground. An output ground potential greater than 50 VDC will damage the amplifier and could cause injury or death to personnel.
Copyright © SpectraDynamics, Inc. 2010
5
Operation
SDI
This unit is designed to operate only with the specified voltage on page 4. For conversion to a different voltage of operation contact SDI.
To operate the unit, locate the power entry module on the rear of the enclosure and connect the power cord. Plug the unit into an appropriate power outlet. The green LED on the front panel labeled POWER will turn on. Attach the signal to be distributed to the SMA input connector on the front panel. Any HPDA-100RM-A output may be used to drive the input of another distribution module.
Absolute Maximum Ratings
Input power +20dBm Maximum Reverse Power +20dBm Maximum Input DC Voltage 10 VDC Maximum
Output DC Voltage 50 VDC Maximum Storage Temperature -10 to +75 ºC Operation Environment 0 to +50 ºC
Rack Mount Chassis 1U H, 19“ W, 14” D
Copyright © SpectraDynamics, Inc. 2010
6
PARAMETER
CONDITIONS
MIN
TYP
MAX
UNITS
Max. Input Power
1 dB compression
15
16 - dBm
Bandwidth
+/- 1 dB
80 - 120
- - MHz
Gain
@ 100 MHz - -
+/- 0.5
dB
Impedance
Input Output - -
50 50 - -
Ohms
Return Loss
Input (S11) 5 MHz Output (S22) 5 MHz
- - -22
-30
-20
-25
dB
Distortion
+13 dBm - -45
-40
dBc
Isolation
Output to output Output to input
95
105
100 110 - -
dB
Phase Noise
10 Hz 100 Hz 1 kHz 10 kHz
-
-
-
-
-158
-165
-171
-174
-155
-162
-168
-171
dBc/Hz
Specifications
SDI
All tests done at 100 MHz and +13 dBm input unless otherwise specified.
Copyright © SpectraDynamics, Inc. 2010
7
SDI
Warranty
SDI
SPECTRADYNAMICS, INC 1849 Cherry St. Unit 2. Louisville, CO 80027 Phone: (303) 665-1852 Fax: (303) 604-6088
Se habla español
The HPDA-100RM-A is warranted to be free of defects under normal operating conditions, as specified, for one year from date of original shipment from
SpectraDynamics, Inc (SDI). SDI’s obligation and liability under this warranty is
expressly limited to repairing or replacing, at SDI’s option, any product not meeting the
said specifications. This warranty shall be in effect for one (1) year from the date a HPDA-100RM-A is sold by SDI. SDI makes no other warranty, express or implied, and makes no warranty of the fitness for any particular purpose. SDI’s obligation under this warranty shall not include any transportation charges or costs of installation or any liability for direct, indirect, or consequential damages or delay. Any improper use, operation beyond capacity, substitution of parts not approved by SDI, or any alteration or
repair by others in such manner as in SDI’s reasonable judgement affects the product
materially and adversely shall void this warranty. No employee or representative of SDI is authorized to change this warranty in any way or grant any other warranty.
Copyright © SpectraDynamics, Inc. 2010
HPDA-100RM-A:1I-5O/R02