![](/html/7e/7e91/7e91a261ea33c915c590318f15d5c0db62d85f9d9a376499355a25ea8215b253/bg1.png)
OWNER’S MANUAL
Model: F500S
SIT-TO-STAND PATIENT LIFT
Ensure the product has been assembled according to the instructions in this manual.
All operators should read and understand the instructions for safe and proper operation of the
patient lift.
P11584
![](/html/7e/7e91/7e91a261ea33c915c590318f15d5c0db62d85f9d9a376499355a25ea8215b253/bg2.png)
Thank you for choosing Span
To better serve you, please record
the following information for future use:
Supplier Name:
Supplier Telephone:
Product Serial Number:
Date of Purchase:
Date of Manufacture
The manufacture date can be found on the serial
number. The YY on the serial number is the
manufacture year, and the MM is the manufacture
month.
INSPECT YOUR MERCHANDISE
Upon receipt of your lift, verify that it is complete and free from any shipping damage.
Refuse delivery if the packaging appears to be badly damaged. If the merchandise is
damaged or any component is missing, contact the shipping company immediately
and le a claim. For further assistance, contact your local dealer or Span-America.
DEFINITIONS & SYMBOLS
In this manual the user refers to the patient or resident and may be used
interchangeably at different times. Caregiver refers to the operator or person who is
assisting with the transfer.
Symbols used in this manual and their meanings:
Warning! Failure to heed this warning may result in damage to the product
or serious injury to the operator and/or user.
Important instructions follow. Read and understand the instructions in the
manual before using the product.
i
Note! Important information concerning the product and/or its correct and
proper usage follows.
1
![](/html/7e/7e91/7e91a261ea33c915c590318f15d5c0db62d85f9d9a376499355a25ea8215b253/bg3.png)
Safety Instructions
The Span-America series of patient lifts refl ect innovative state of the
art design to increase user mobility. The product will provide years of
service if it is properly maintained as any electric and/or mechanical
equipment requires. Please pay careful attention to the following important
information regarding the care, maintenance, and operation of the patient lift.
Carefully read these instructions before assembling the lift, or attempting to lift a user
with the device.
Please Note
• Special care must be taken with users/patients who cannot
themselves provide assistance while being lifted. (i.e. patients who are
i
not be used for transporting or moving any patient from one location to another
location.
• During lifting or lowering, whenever possible, always keep the base of the lift in
the widest position.
• The base of the lift should be closed before moving the lift.
• Do not roll casters over any object while the user/patient is in the sling.
• Do not lock casters during lifting.
• While being lifted in a sling, always keep the user/patient centered over the base
and facing the caregiver operating the lifter.
• Never leave the user/patient unattended during lifting.
comatose, spastic, agitated, or otherwise severely handicapped.)
• The patient lift should be used solely for transferring a user/patient from
one utility (beds, chairs, toilets, etc.) to another. The patient lift should
The F500S can assist a person to be lifted and transferred safely, with minimum
physical effort provided by the caregiver. Before attempting to lift anyone, please
practice operating the lift. Also prior to actual lifting, explain the lifting procedure to
the user/patient.
Service personnel may contact Span for any information, instructions or certain parts
for servicing purpose.
Class II Equipment
2
![](/html/7e/7e91/7e91a261ea33c915c590318f15d5c0db62d85f9d9a376499355a25ea8215b253/bg4.png)
3
WARNING!
• NO modi cation of this equipment is allowed. It is unsafe to modify the
equipment from its original state. This product is deemed safe to use with
only the original parts as delivered. Unauthorized use of any modi ed
equipment may result in serious injury or death.
• DO NOT replace any components of the lift without consulting with Span,
and follow proper instructions when replacing any components.
• DO NOT service any parts of the lift while in use with a patient.
• It is NOT safe to load the equipment more than the indicated maximum safe
working load.
• IF any accessories with lower load capacity than the equipment itself is used,
DO NOT load the equipment more than the lowest maximum safe
working load.
• REMOVE the battery pack if the lift will not be used for some time, to prevent
battery leakage. Periodically check the battery pack as well as the charger if
necessary
• This equipment DOES NOT consist of any materials or ingredients to
which patients are exposed, that may cause bodily harm or injury.
• Prolonged exposure or physical contact with any part of the equipment will
not cause skin irritation, allergic reaction, etc.
• Using accessories, detachable parts, or materials not described in the
instruction manual MAY RESULT IN SERIOUS INJURIES.
• Keep the equipment away from children and pets when in use. IT MAY
CAUSE SERIOUS INJURY OR DEATH.
• Keep the equipment away from pests or pest infested area. Diseases
transferred from pests to the equipment MAY CAUSE SERIOUS ILLNESS
to patients upon contact.
• If the lift is left under the sun for long period of time, be CAUTIOUS as the
surface of the lift may be hot.
• There are no risk if the lift is exposed to reasonably foreseeable
environmental conditions
• Caution! The controller contains parts and assemblies that could be
susceptible to damage by electrostatic discharge (ESD). To reduce
the likelihood of this from happening, one should be grounded prior to
operate the controller.
• DO NOT use the lift on a inclined surface that that greater than 5°
![](/html/7e/7e91/7e91a261ea33c915c590318f15d5c0db62d85f9d9a376499355a25ea8215b253/bg5.png)
FEATURES & OVERVIEW
6 Rear Caster 11 Emergency Lowering
2 Footplate
7 Front Caster 12 Knee Pad
Emergency Stop Button
5 Leg Opening Lever
Prior to assembly, unpack all parts from the shipping carton and check for any
i
missing parts. Contact you dealer immediately if a part is missing.
Easy Tools
Screw Driver
1
11
10
12
8
13
3
4
5
9
** Heaviest detachable part
2
6
7
4
![](/html/7e/7e91/7e91a261ea33c915c590318f15d5c0db62d85f9d9a376499355a25ea8215b253/bg6.png)
J
B
K
L
A
C
600
M
D
E
N
F500S
mm (in)
A
B 559.7 (22.0)
C 940.7 (37.0)
D 19.8 (0.8)
E 386.8 (15.2)
F
700
G
H
F 816.1 (32.1)
G 749.8 (29.5)
H 547 (21.5)
I 291.9 (11.5)
J 646.3 (25.4)
K 200.8 (7.9)
I
L 201.2 (7.9)
M 123.4 (4.9)
N 33.4 (1.3)
5
![](/html/7e/7e91/7e91a261ea33c915c590318f15d5c0db62d85f9d9a376499355a25ea8215b253/bg7.png)
Additional Dimensions & Specifications
Model F500S
4" Rear Caster with Brake
Single Piece Molded Kneepad
Depth Adjustable Kneepad
Position
Lift Weight Unloaded (lb)
Lift Weight Unloaded (kg)
SPECIFICATIONS & OPTIONS
Specifi cations of Control Unit
Input Voltage 220-240VAC 50Hz / 110-120VAC 60Hz
Output Voltage 24 VDC
Battery Pack 24 VDC 5AH
Lift Control Handset and Panel
Protection Class IP54 (Protection against dust and splash of water)
Sound Level Under Full Load 16db
Sound Level Under No Load 11db
Cycles Per Charge 25 full cycles under full load
Overload Warning by LCD
Battery Status by LCD
Duty Cycle 10% Duty cycle (Max 2 minute ON,18 minute OFF)
* Span-America is committed to continuous improvements of our products therefore
the specs, dimensions, and features listed above are for guidance only and subject
to change without prior notice.
6
F500S
107.1
6