Sorvall WX 100 Ultra, WX 90 Ultra, WX 80 Ultra Instruction Manual

Page 1
50105750-4 October 2010
Thermo Fisher Scientific
WX Ultra Series WX 100, WX 90, WX 80
Instruction Manual
Page 2
© 2010 Thermo Fisher Scientific Inc. All rights reserved.
is either a registered trademark or a trademark of Thermo Fisher Scientific.
Microsoft, Microsoft Excel, Windows, Windows XP, and Windows Vista are registered trademarks of Microsoft Corporation in the United States and/or other countries.
All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries.
Thermo Fisher Scientific Robert-Bosch-Straße 1 D - 63505 Langenselbold
Germany
Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to use in the product operation. This document is copyright protected and any reproduction of the whole or any part of this document is strictly prohibited, except with the written authorization of Thermo Fisher Scientific Inc.
The contents of this document are subject to change without notice. All technical information in this document is for reference purposes only. System configurations and specifications in this document supersede all previous information received by the purchaser.
Thermo Fisher Scientific Inc. makes no representations that this document is complete, accurate or error­free and assumes no responsibility and will not be liable for any errors, omissions, damage or loss that might result from any use of this document, even if the information in the document is followed properly.
This document is not part of any sales contract between Thermo Fisher Scientific Inc. and a purchaser. This document shall in no way govern or modify any Terms and Conditions of Sale, which Terms and Conditions of Sale shall govern all conflicting information between the two documents.
Release history: 50105750-4 printed in October 2010.
For Research Use Only. Not for use in diagnostic procedures.
Page 3
Thermo Scientific WX Ultra Series i
C
Chapter 0 Preface - Intended use and safety definitions ...............................................................................................iii
Safety Definitions..................................................................................................................................iii
Important safety reminder .....................................................................................................................iv
Chapter 1 Description...........................................................................................................................................................1-1
General description............................................................................................................................. 1-1
External view of ultracentrifuge ...................................................................................................... 1-2
Design................................................................................................................................................ 1-3
Control panel ................................................................................................................................. 1-3
Rotor chamber................................................................................................................................ 1-6
Safety devices.................................................................................................................................. 1-7
Rotor adapter ................................................................................................................................. 1-8
Chapter 2 Operation..............................................................................................................................................................2-1
Run preparation ................................................................................................................................. 2-2
Starting up the ultracentrifuge ........................................................................................................ 2-3
Preparing tubes/bottles and rotor.................................................................................................... 2-4
Basic operation ................................................................................................................................... 2-4
Setting run conditions .................................................................................................................... 2-4
Setting user ID code ....................................................................................................................... 2-9
Using Rotor Life Management .....................................................................................................2-10
Basic operating procedure................................................................................................................. 2-11
Normal operation......................................................................................................................... 2-13
Zonal operation............................................................................................................................ 2-15
Acceleration and deceleration rates ................................................................................................... 2-17
How to use the FUNCTION field...................................................................................................2-19
Programmed operation ................................................................................................................. 2-21
Step-mode operation .................................................................................................................... 2-32
RTC (real-time control) feature.................................................................................................... 2-40
Displaying and setting RCF.......................................................................................................... 2-48
Setting w2T.................................................................................................................................. 2-56
Defrost function ........................................................................................................................... 2-58
Setting and resetting zonal operation ............................................................................................ 2-59
Rotor management........................................................................................................................... 2-60
Rotor life management ................................................................................................................. 2-60
Automatic rotor deration.............................................................................................................. 2-61
Automatic rotor exclusion............................................................................................................. 2-61
Registering a rotor ........................................................................................................................ 2-61
Deleting a registered rotor ............................................................................................................ 2-64
Updating the existing data for a rotor ........................................................................................... 2-65
When power failure occurs ............................................................................................................... 2-66
Features of the menu screen.............................................................................................................. 2-69
Contents
Page 4
Contents
ii WX Ultra Series Thermo Scientific
Centrifuge scheduler..................................................................................................................... 2-69
User list ........................................................................................................................................ 2-71
Alarm information........................................................................................................................ 2-76
Rotor catalog ................................................................................................................................ 2-77
User Customizations..................................................................................................................... 2-77
Print utilities (option)....................................................................................................................... 2-81
Operating procedure..................................................................................................................... 2-82
Contents of print utilities ............................................................................................................. 2-84
Automatic run result printing....................................................................................................... 2-86
Printer fault handling ................................................................................................................... 2-87
Lockout (optional)............................................................................................................................ 2-88
Chapter 3 Maintenance........................................................................................................................................................3-1
Rotor chamber.................................................................................................................................... 3-1
Drive spindle ...................................................................................................................................... 3-2
Cabinet............................................................................................................................................... 3-2
Replacement parts .............................................................................................................................. 3-2
Service decontamination policy .......................................................................................................... 3-2
Chapter 4 Troubleshooting ..................................................................................................................................................4-1
Alarm indicators ................................................................................................................................ 4-1
Diagnosed problems - requiring maintenance..................................................................................... 4-3
Non-diagnosed problems ................................................................................................................... 4-3
Chapter 5 Preinstallation.....................................................................................................................................................5-1
Location Requirements....................................................................................................................... 5-1
Electrical requirements ....................................................................................................................... 5-2
Leveling.............................................................................................................................................. 5-2
Moving the ultracentrifuge ................................................................................................................. 5-3
Chapter 6 Specifications .....................................................................................................................................................6-1
Chapter 7 Supply list.............................................................................................................................................................7-1
Appendix A Service Decontamination Policy....................................................................................................................A-1
Appendix B Condensed operating instructions .................................................................................................................B-1
Appendix C Chemical Compatibility Chart .........................................................................................................................C-1
Appendix D WARRANTY.........................................................................................................................................................D-1
Appendix E Contact Information ...........................................................................................................................................E-1
Page 5
Thermo Scientific WX Ultra Series iii
P
Preface - Intended use and safety definitions
This manual is a guide to the use of the Sorvall WX Ultra series ultracentrifuge.
Information herein has been verified and is believed adequate for the intended use of the centrifuge. Because failure to follow the recommendations set forth in this manual could produce personal injury or property damage, always follow the recommendations set forth herein. Thermo does not guarantee results and assumes no obligation for the performance of centrifuges or other products that are not used in accordance with the instructions provided. This publication is not a license to operate under, nor a recommendation to infringe upon, any process patents.
Safety Definitions
DANGER, WARNING, CAUTION, and NOTE within the text of this manual are used to emphasize
important and critical instructions.
DANGER, WARNING and CAUTION information is accompanied by a hazard symbol and appears throughout the manual, both on the Important Safety Reminder pages and near the information it corresponds to.
Before you operate the centrifuge, we recommend that you read this instruction manual thoroughly, particularly all DANGERS, WARNINGS and CAUTIONS. Never operate the centrifuge without first considering all items on the Important Safety Reminder pages, and never operate the centrifuge in any manner not described in this instruction manual.
DANGER Informs the operator of an extreme hazard or an unsafe practice that will result in death or serious injury.
WARNING Informs the operator of a hazard or an unsafe practice that could result in serious injury or death, affect the operator's health, or contaminate the environment.
CAUTION Informs the operator of a hazard or an unsafe practice that could cause minor injury, or result in damage of equipment or property.
Note Highlights essential information that is not hazard-related.
Page 6
Preface - Intended use and safety definitions
Important safety reminder
iv WX Ultra Series Thermo Scientific
Important safety reminder
Certain potentially dangerous conditions are inherent to the use of all centrifuges. To ensure safe operation of this centrifuge, anyone using it should be aware of all safe practices and take all precautions described below and throughout these operating instructions.
DANGER
• Before removing any cabinet panel (such as for maintenance or repair), always turn the main Power Switch OFF and unplug the centrifuge, then wait at least three minutes to eliminate the potential for severe electric shock.
WARNING
• Never exceed the maximum rated speed of the installed rotor; to do so can cause rotor failure.
• Always reduce (derate) rotor speed whenever:
1. The rotor speed/temperature combination exceeds the solubility of the gradient material and causes it to precipitate.
2. The compartment load exceeds the maximum allowable compartment load (or design mass) specified for your rotor.
Failure to reduce speed under these conditions can cause rotor failure.
• Always inspect the rotor as specified in your rotor manual. Do not use a rotor that shows signs of damage or corrosion.
• Centrifuges routinely deal with high energy levels and could move suddenly in the unlikely event of rotor failure. During centrifuge operation, never lean on or move the centrifuge, keep the surrounding area clear of objects (including all hazardous and flammable materials), and do not work on top of or next to the centrifuge (as a rule, keep people and objects at least 300 mm [12 inch] away).
• Never unlock the chamber door, attempt to open the door, or otherwise attempt to touch the rotor while it is rotating. In the event of a power outage, it can take more than three hours for the rotor to stop - be sure to wait at least three hours before opening the door.
• Never attempt to override or otherwise disable any safety features.
• When using radioactive, toxic, or pathogenic materials, be aware of all characteristics of the materials and the hazards associated with them in the event leakage occurs during centrifugation. If leakage does occur, neither the centrifuge nor the rotor can protect you from particles dispersed in the air. To protect yourself, we recommend additional precautions be taken to prevent exposure to these materials, for example, use of controlled ventilation or isolation areas.
• Always be aware of the possibility of contamination when using radioactive, toxic, or pathogenic materials. Take all necessary precautions and use appropriate decontamination procedures if exposure occurs.
Page 7
Preface - Intended use and safety definitions
Important safety reminder
Thermo Scientific WX Ultra Series v
WARNING
• The use of sealed rotors, buckets and/or sample containers will provide increased protection from contamination during routine operation. However, these items will not guarantee contamination protection from accidents resulting in damage to the rotor or buckets. Do not run hazardous materials in the centrifuge unless placed in a biohazard enclosure and operated using all appropriate safety precautions.
• Never use any materials capable of producing flammable or explosive vapors, or extreme exothermic reactions.
• The centrifuge is equipped with a three-wire power cord that has one wire for connection to ground. The centrifuge must be correctly grounded to guard against shock hazards.
CAUTION
• Do not operate the control keys using a ball-point pen or a sharp object.
• Do not run or precool the rotor at the critical speed as this will have a detrimental effect on centrifuge component life (see rotor manual).
• Do not operate the centrifuge with a rotor that is not balanced within specification. To do so can cause damage to the centrifuge drive assembly.
• Always make sure the rotor is correctly assembled and properly seated on the centrifuge drive spindle before operation. If using a swinging-bucket rotor, always make sure that a bucket is installed at each position and that each is properly seated and can swing freely after installing the rotor on the centrifuge drive spindle (see rotor manual).
• Do not exceed the maximum speed of the tubes, bottles, or adapters being used. Check manufacturer's published specifications or see Sorvall Product Guide; if unsure, to avoid loss of valuable sample, we recommend performing a test run.
• Do not continue to operate the centrifuge if abnormal sounds occur during operation. Immediately discontinue use of the centrifuge and contact Thermo Service.
• Supply power must be checked before the centrifuge is connected to power because the centrifuge can be damaged if connected to the wrong voltage. Thermo is not responsible for incorrect installation and warranty is void if an initial installation or electrical modification of the centrifuge is not performed by Thermo or a Thermo representative.
• Before moving the centrifuge, always remove the rotor from the rotor chamber to avoid damage to the drive assembly.
• Do not place fluids inside, on top of, or close to the centrifuge --- spillage can result in electrical or mechanical failure.
• Always operate and maintain the centrifuge and all rotors as instructed in this manual and in the rotor manual(s).
Page 8
Preface - Intended use and safety definitions
Important safety reminder
vi WX Ultra Series Thermo Scientific
Page 9
Thermo Scientific WX Ultra Series 1-1
1
Description
General description
The Sorvall WX Ultra series ultracentrifuges are designed and manufactured based on our long experience in the development of centrifuges; they are easy to use and highly reliable. Features include the following:
1. Maximum speed of 100,000 rpm (802,006 x g)
2. Control panel is simple with easy key operation and easy-to-see liquid crystal screen.
3. The displayed language can be switched over between English and Japanese.
4. The real-time control (RTC) feature enables setting a start time or a finish time, thus letting you run your machine at a desired date and time.
5. Centrifugal force (RCFmax and RCFavg) can be displayed and set.
6. Twenty varieties of nine stepped modes can be programmed for a wide range of applications.
7. Various alarms notify users of causes and necessary actions when fault conditions are detected allowing easier and quicker troubleshooting.
8.
Space saving design. The installation area required is 0.81 m2 (90 x 90 cm) pr 9.0 ft2 (3.0 x 3.0 ft). Lower
top deck makes it easy to install and remove the rotor.
9. Low noise level makes the ultracentrifuge suitable for use in any laboratory.
10. Samples can be easily balanced visually.
11. CFC-free thermoelectric cooling system with greater cooling capacity.
Advanced features (options)
The options of the Sorvall WX Ultra series ultracentrifuges are: Compass Software, Lockout Kit, and a Printer.
The Compass Software is a calculation/simulation system for centrifuge support that is compatible
with
Windows
R 2000 / XP Professional. Compass Centrifugation software helps determine appropriate run
conditions prior to starting centrifugal separation, and calculates optimum centrifuge and rotor run conditions for materials in which proper run conditions are unknown. For more information on this please consult the Compass Operator's Manual.
The Lockout Kit limits centrifuge access to a group of registered users (up to 40) which is controlled by lab managers/supervisors (up to 3). For more information on this see page 2-95, Lockout.
The printer provides hard copy reports of information maintained in the centrifuge including: Set and actual run conditions, rotor use, programmed memory contents, and others. For more
information on
this see page 2 - 89, Print Utilities.
Page 10
Description
General description
1-2 WX Ultra Series Thermo Scientific
External view of ultracentrifuge
The WX Series ultracentrifuge have the same external view, except for the model name printed on front cover. The following is the external view of the WX ultracentrifuge.
External view of ultracentrifuge
Figure 1-1
External view of the Sorvall WX Ultra series
Note * This height is measured from level floor surface.
Top d eck
Door
Safety cover
Display panel
690 mm (27.2 in) 790 mm (31.1in) 850 mm (33.5 in) 1000 mm (39.4 in)
Operation key
Rotor rubber mat
Power switch
Front cover
Page 11
Description
Design
Thermo Scientific WX Ultra Series 1-3
Design
Control panel
The control panel for the Sorvall WX Ultra series consists of a color screen, a touch sensitive display panel and a keyboard.
The display panel incorporates an easy-to-read liquid crystal display. The display panel (field display) displays running conditions and running status (this screen is called the Run Screen), along with features such as Programmed Run, Rotors List, and User Customization Screens. Fig. 1-2 shows the display panel, and Fig.1-3 represents the keyboard.
Figure 1-2 Display panel
Figure 1-3 Keyboard
(3) Message indicator
{
(1) Field display
(5) Vacuum indicator
(4) RUN mode indicator
(2) FUNCTION field
}
}
Operation state display field
Setting value display field
(12) Key pad
(11) Cursor key
(6) START key
(7) STOP key
(8) VACUUM key
(10) MENU key
(9) ESC key
Page 12
Description
Design
1-4 WX Ultra Series Thermo Scientific
Table 1-1 [Functions of the display panel-keyed by item no. to fig. 1-2]
No. Name and Symbol Functions and actions
(1) Field display
SPEED
TIME
TEMP
Displays various fields. The SPEED, TIME, and TEMP fields give the current status indicator in the top row and the setting indicator in the bottom row. (For setting, see Section “Setting run conditions”.)
• SPEED (Speed indicator) (Top row) Displays speeds in increments of 10 rpm at lower than 5,000 rpm, and in steps of 100 rpm at 5,000 rpm or more. (Bottom row) Sets speeds from 1,000 to maximum speed in increments of 100 rpm. The lower two digits (one, ten positions) display zeros.
• TIME (running time indicator) (Top row) Displays the remaining operation time or the time elapsed during operation if settings are performed on the User Customization screen. If the running time is set to HOLD, this field displays time elapsed. (Bottom row) Specifies a setting in the range from 1 minute to 99 hours 59 minutes in steps of minutes and hours.
• TEMP (temperature indicator) (Top row) Displays in steps of 0.1 °C. (Bottom row) Sets a setting in the range from 0 °C to 40 °C in increments of 0.1 °C
• ACCEL (acceleration mode indicator). Displays acceleration modes 1 through 9.
• DECEL (deceleration mode indicator). Displays deceleration modes 1 through 9, along with free coast (F).
(2) Function field • ID CODE Sets an ID code.
• RLM Switches to the Rotor Management screen.
• PROG • RTC Sets, recalls programmed runs or sets the time and the end time for a programmed run.
• RCF • ω
2
T Sets and displays the centrifugal force or sets an ω2T
• DEF Sets and resets the defrost function.
• ZONAL Sets the zonal operation mode.
(3) Message indicator Displays an alarm message and various suggestions for operation.
(4)
Run mode indicator
The following terms are displayed: STOP, ACCEL, RUN (running at the set speed) DECEL, WAIT (waiting for vacuum during acceleration) ZONAL (for zonal operation)
DELAY (until the start time in an RTC run)
(5) VACUUM indicator Displays the following four stages according to the vacuum of the rotor chamber.
(1) Atmospheric state. The vacuum pump is not activated.
(2) Low vacuum. The rotor waits at 4,000 rpm until the vacuum reaches an inter-
mediate level
(3) Intermediate vacuum.
(4) High vacuum.
Note If the sample is sensitive to a temperature rise, do not press the START key until the
chamber is at high vacuum level.
Page 13
Description
Design
Thermo Scientific WX Ultra Series 1-5
Table 1-2 [Functions of keyboard-keyed by item no. to fig. 1-3]
No. Name and Symbol Functions and actions
(6) START key Starts rotor rotation. If VACUUM is off, this key activates the vacuum pump and starts
temperature control.
(7) STOP key Stops rotor rotation.
(8) VACUUM key Starts up the vacuum pump and activates air vent (As soon as vacuum pump is on, tem-
perature control starts.) Air vent for vacuum chamber after a run cannot be opened as long as the rotor is spinning.
(9)
ESC key Moves the display back to the screen at the preceding level (for example, to switch
back from the Menu Screen to the Run Screen).
(10) MENU key Displays the Menu Screen.
The Menu Screen offers the choice of Centrifuge Scheduler, User List, Alarm Informa­tion, Rotor Catalog, and User Customization Routines.
(11) Cursor key a. Displays the cursor on the Run Screen, putting the display into input wait status.
b. Move the cursor on the screen.
1. Moves the cursor up
2. Moves the cursor down
3. Moves the cursor to the right
4. Moves the cursor to the left
(12) Numeric key Used to type numbers for setting run conditions.
During time and temperature entry: Moves cursor from hours to minutes.
Acts as decimal point for data entry.
a. During operation time entry: sets continuous run.
b. When entering deceleration conditions: sets a free coast.
Use this when you have entered the wrong value while entering an
operating condition or entering a number or when the alarm device is activated. Functions of this key a. This key clears the cursor-carrying input field and returns you to the preinput state. b. Use this key to clear an alarm signal. If more than one alarm signal is on, this key will clear them one by one.
Registers the entered value.
Page 14
Description
Design
1-6 WX Ultra Series Thermo Scientific
Rotor chamber
The structure of the rotor chamber (vacuum chamber) is shown in Fig. 1-4.
Figure 1-4 Rotor chamber
Note If sample or water drops to the window of the temperature sensor, it may cause an incorrect
detection. Whenever the sensor is wet, wipe it with a clean, dry cloth. Take care not to scratch the surface of the sensor.
Drive spindle
Speed sensor
Chamber door seal
Handle
Chamber door
Temperature sensor (measuring the temperature of the rotor)
Door lock system
Vacuum chamber
Bowl (Rotor Chamber)
Thermoelectric cooling element (cooling the rotor)
Guard
Overspeed detector (detecting any instance exceeding the maximum allowable speed of the rotor)
Page 15
Description
Design
Thermo Scientific WX Ultra Series 1-7
Safety devices
(1) Containment
Should a rotor failure occur, the guard ring will contain the contents of the rotor inside the centrifuge (Fig. 1-4).
(2) Imbalance detector
If during operation the vibration of the rotor becomes excessive due to serious imbalance or improper bucket setting, the imbalance detector detects the situation and decelerates the rotor immediately. However, the ultracentrifuge is designed to tolerate imbalance associated with visual balancing-it is equipped with an imbalance tolerant drive. (For more information on the balancing of rotors, see Section 2-1-2, “Preparing tubes/bottles and rotor”.)
(3) Door lock system
The chamber door automatically locks for safety while the rotor is spinning. When the power supply is off, the door remains locked. The door can only be opened and closed when the rotor is at rest and the rotor chamber is vented. Unless the door is closed, the rotor will not start rotating except in zonal mode. To open the door in the event of a power failure, see Section 2-7, “When power failure occurs”.
(4) Speed sensor and overspeed detector
For protection in the event of entry errors the ultracentrifuge is provided with an automatic system to stop the rotor when its speed exceeds the maximum allowable speed. If a speed higher than the maximum permitted speed is set, the ultracentrifuge will detect the mistake before the speed reaches 3000 rpm, and then will display an alert message and decelerate the rotor to a stop.
Page 16
Description
Design
1-8 WX Ultra Series Thermo Scientific
Rotor adapter
Overspeed Decal
The overspeed decal located on the rotor base has alternating black and white bands. The number of bands corresponds to the maximum permitted speed of the rotor. (See Fig. 1-5.)
Figure 1-5 Standard rotor
To protect the overspeed decal, be sure to store the rotor on the rotor stand provided. (See Fig. 1-6.)
Figure 1-6 Rotor stand
Overspeed decal
Standard rotor with overspeed decal
Page 17
Thermo Scientific WX Ultra Series 2-1
2
Operation
The Sorvall WX Ultra series ultracentrifuges are capable of operation in more than one mode to meet a wide range of applications. The outline of each available mode is given below:
Outline Reference
Normal operation
Section Normal opera-
tion
Optional features
Programmed opera­tion
You can store set run conditions in memory for later use in repeated operation.
Section How to use the
FUNCTION field
Section Programmed
operation
Step-mode operation
More than one normal operation can be combined into a sequence of operations or step for successive centrifuga­tion.
Section Step-mode
operation
RTC (real-time con­trol) operation
Run starts or completes at a required date and time.
Section RTC (real-time
control) feature
RCF (centrifugal force) value display setting
This feature calculates centrifugal force (RCF) values from set speed. It can also calculate reversely, i.e., finding speed from such values.
Section Displaying and
setting RCF
Speed
Time
Speed
Speed
Time
Time
Store
Recall
Speed
Time
Speed
Time
A finish date and time re­quired
START key ON
Speed
Time
RCF
value setting
Page 18
Operation
Run preparation
2-2 WX Ultra Series Thermo Scientific
Run preparation
Optional features
ω2T setting
Calculates and displays the running time from the set val-
ues of speed and ω
2
T.
Section Setting w2T.
Zonal operation
Zonal operation is a mode of operation using a zonal rotor.
Section Zonal opera-
tion.
Speed
Time
Speed
Time
Zonal speed
WARNING
1. Never use any materials capable of producing flammable or explosive vapors, or extreme
exothermic reactions.
2. When using radioactive, toxic, or pathogenic materials, be aware of all characteristics of the materials and hazards associated with them in the event leakage occurs during centrifugation. If leakage does occur, neither the centrifuge nor the rotor can protect you from particles dispersed in the air. To protect yourself, we recommend additional precautions be taken to prevent exposure to these materials, for example, use of controlled ventilation or isolation areas.
CAUTION Do not place fluids inside, on top of, or close to the centrifuge---spillage can result
in electrical or mechanical failure.
Page 19
Operation
Run preparation
Thermo Scientific WX Ultra Series 2-3
Starting up the ultracentrifuge
Before setting run conditions, display the run screen (screen for setting run conditions)
(1) Displaying the Run Screen (screen for setting run conditions)
Figure 2-1
Initial screen, and run screen
Initial screen
1. Turn on the POWER switch.
2. The initial screen appears.
3. The Run Screen appears.
Page 20
Operation
Basic operation
2-4 WX Ultra Series Thermo Scientific
Preparing tubes/bottles and rotor
The Sorvall WX Ultra series ultracentrifuges allow you to balance, by eye, tubes or bottles containing a sample solution and then centrifuge them. Make sure that the difference between meniscus levels of sample solution in tubes or bottles in within 5 mm (0.197 in) (See fig. 2-2).
Figure 2-2
Balancing tubes/bottles containing a sample solution
To prevent tube or bottle failure, some tube and rotor combinations cannot be run to the maximum speed of the rotor when partially filled. The tube or bottle must be full in the following cases:
1. When a thin tube or seal tube is used.
2. When a thick tube is used for swinging rotor.
3. When a bottle is used 100,000 x g or more.
Basic operation
Setting run conditions
This section will first describe the screen for basic operation (the run screen) and the cursor keys.
Within 5 mm
(0.197 in)
Within 5 mm
(0.197 in)
(1) Tubes
(2) Bottles
WARNING Do not lean on, tilt or place anything on the centrifuge while in operation.
CAUTION
1. Do not operate the keyboard with a ball-point pen or a sharp object.
2. Do not continue to operate the centrifuge if abnormal sounds occur during operation. Immediately discontinue use of the centrifuge and contact Thermo Service.
Page 21
Operation
Basic operation
Thermo Scientific WX Ultra Series 2-5
[run screen]
The screen for displaying run conditions and operational status is called the run screen. Speed, time, and temperature are displayed in two rows: the top row displays the current actual run conditions, while the bottom row displays the set run conditions. The acceleration (ACCEL) and the deceleration (DECEL) fields display set conditions.
Figure 2-3
Run screen
[Cursor key]
Pressing the cursor key will highlight the field where changes can be made. (This blinking/highlighted object is referred to as the cursor in this manual.) The screen setting field is in either of the following states depending on whether the cursor is there.
1. Determined input state: This is a normal state and the cursor does not appear.
2. Input wait state: Press a cursor key (either the top, bottom, right, or left) while in the determined input state, and the numerical part of the setting field will blink a 0 (or numerical value) and display the cursor. In this state, the system accepts a numerical input. Press cursor keys to move the cursor.
To set a run condition, enter the cursor into an input wait state, move the cursor to the item you want to set, then enter a value. If you have made no keystroke (such as a numerical input) for more than 30 seconds, the system will automatically enter a determined input state.
Note When the screen is in a determined input state without a cursor (when the Run Screen is on after power-up), if you wish to enter a numerical value in a specified parameter field, press a cursor key (either of the upward, downward, right, and left arrow keys), move the cursor to the specified parameter field, then enter the value. The cursor keys display and move a cursor. Once a cursor appears, pressing a cursor key moves the cursor to the corresponding direction (upward, downward, rightward, or leftward).
Operation state display field
Setting value display field
Run mode indicator
} }
}
Page 22
Operation
Basic operation
2-6 WX Ultra Series Thermo Scientific
Figure 2-4 Setting indicator
How to set speed, running time, temperature, and other parameters
Here are some examples and descriptions:
Note
1. If you enter the wrong value, press CE key to return to the input wait state. If you have pressed ENTER key, press a cursor key, enter the device into an input wait state, then enter the correct value.
2. When setting two or more run conditions, you do not have to press ENTER key after each setting. Pressing the cursor key will enter the setting, thus making the system wait for a new input.
3. If the system is running in (HOLD) and you want to set it to shut down at a future time, enter a new time setting while the instrument is in operation; enter the sum of the time elapsed plus the time remaining. If, for example, this machine has run continuously for five hours and you want to stop it one and a half hours later, use cursor keys to enter TIME into an
input wait state,
then enter
Setting item RPM (SPEED Running time (TIME)
Typical setting 100,000 rpm 2 hours 30 minutes
Cursor key
Cursor (blinking)
The system does not accept a numerical input.
(1) Determined input state
The system accepts a numerical input. Use cursor keys to move to the next setting item.
(2) Input wait state
Page 23
Operation
Basic operation
Thermo Scientific WX Ultra Series 2-7
Operation procedure
1 Press cursor keys to enter the system into an
input wait state.
The system enters an input wait state.
2 hours 30 minutes
2 Use cursor keys to move the cursor to the
status indicator. (The arrows indicate the directions the cursor can be moved..
The system enters an input wait state.
3 The cursor in the setting item field blinks for
about 30 seconds. Blinking means that the system enters an input wait state.
4 Use numeric keys to enter a setting.
Entered numbers are moved to the left every time a new number is entered.
The last two digits are fixed.
Press the “ ” key to
move the cursor to the “minutes” position. For a continuous run, press HOLD/FREE.
5 Make a check, then press ENTER.
After pressing a cursor key, you can still enter a setting similarly to the ENTER key. Use CE to cancel an input.
Set it to 100,000 rpm. Set it to 2:30
(2 hours 30 minutes).
Setting range and units Can be set to any value in
the range from 1,000 rpm to maximum speed in increments of 100 rpm.
Can be set to any value up to 99 hours 59 minutes in increments of 1 min­ute.
Arrow
Temperature (TEMP) Acceleration (ACCEL) Deceleration (DECEL)
4.5 °C 9 7
Enters an input wait state. Enters an input wait state. Enters an input wait state.
Page 24
Operation
Basic operation
2-8 WX Ultra Series Thermo Scientific
The cursor blinks at one place.
When one decimal place is not required, you do not need to enter
. If you press , it becomes a “decimal place” input and the machine waits for an input of deci­mal places.
For free coast, press HOLD/FREE. Display: F
Set it to 4.5 °C. Set it to 9 Set it to 7.
Can be set to any value in the range from 0 to 40 °C in increments of 0.1 °C.
1 to 9 1 to 9
+ free coast (F)
FUNCTION field
• The FUNCTION field switches to the ID CODE field.
• The number of units of ID CODE blinks.
• The entered number is not displayed on the screen. * is displayed instead of the number.
>> After the registered ID CODE is entered, the name (USER NAME) corresponding to
the ID CODE is displayed.
Page 25
Operation
Basic operation
Thermo Scientific WX Ultra Series 2-9
Setting user ID code
The USER ID CODE is a number to identify each user and can be set in up to 4 digits. When the user ID CODE is entered, the user record will be stored in the memory of the centrifuge and can be printed if a printout operation is done (optional). Entering a user ID CODE may not be always required for operation (Lockout feature optional). If the centrifuge is not equipped with the optional lockout feature and the user does not need to be identified, the centrifuge can be operated without an ID CODE.
• To use a name (user name) corresponding to the user ID code or ID code, it must be registered. For the registration method, see Section User list.
Step Key operation Screen display and considerations
1 On the Run Screen, use cursor keys to
move the cursor to ID CODE and press the ENTER key.
2 Use numeric keys to enter ID CODE
(4 digits). (When entering 0123)
ID code
3 Press the ESC key on the keyboard to
return to the FUNCTION field.
• The display switches to the rotor control field, which displays the Rotor List screen.
• The display switches back to the Run Screen and “T-8100-1234” is displayed in the message field.
Page 26
Operation
Basic operation
2-10 WX Ultra Series Thermo Scientific
Using Rotor Life Management
The rotor type and serial number may be set in rotor management to allow for rotor logging within the centrifuge. Whether the rotor type and serial number for a particular rotor have been set or not, it does not affect the normal operation of the ultracentrifuge. However, you are advised to set the rotor type and serial number of the rotor for each run to allow effective rotor life management. Any registered rotor that does not have its rotor type and serial number set will appear as an undefined rotor in the displayed list (“Undefined rotor 1”). You can use a particular rotor as undefined if you need to manage the rotor separately from the other defined rotors in terms of the number of runs made and accumulated run time (“Undefined rotor 2”). For more information on rotor life management, see Section Rotor management.
Step Key operation Screen display and considerations
1 On the Run Screen, use cursor keys to
move the cursor to RLM and press the ENTER key.)
2 On the rotor control field, use the left
and right arrow keys
and to select each page.
3 Use the up and down arrow keys to
move the cursor to the row of the rotor type to be used and press the ENTER key.
In this example, the cursor moves to “T-8100-1234”.
Select “Running”.
Page 27
Operation
Basic operating procedure
Thermo Scientific WX Ultra Series 2-11
Basic operating procedure
There are two basic modes of operation, normal and zonal. The procedures for these two modes are shown in Fig. 2-5.
START
Turn on the POWER.
Open the chamber door and
install the rotor.
Normal run
Close the door and set the run conditions.
Press VACUUM.
For information on how to set the run conditions, see Section Basic operation
Vacuum pump is ON. Temperature control starts. Check the set run conditions for accuracy.
Press START.
Rotor accelerates to the set speed, and then
high-speed centrifugation begins.
: Indicates panel key.
Zonal run
Reaches the set time,
or press STOP.
Rotor comes to a stop.
Press VACUUM.
Temperature control is OFF. Vacuum pump is OFF. Air valve is open and door lock is released.
Remove the rotor and turn off
the POWER switch.
END
Page 28
Operation
Basic operating procedure
2-12 WX Ultra Series Thermo Scientific
Figure 2-5 Basic operating procedure
Move the cursor to ZONAL in the Run Screen.
Press 3, 0, ENTER
Set run conditions.
For information on how to set the run conditions, see Section Basic operation.
Press START.
Temperature control starts. The rotor rotates at the zonal speed. Load the sample, then cap the rotor.
Close the chamber door and press VACUUM.
Vacuum pump is ON. Check the set run conditions for accuracy.
* This step may be skipped. Just press START to run the ultracentrifuge. When START is pressed, the vacuum pump turns on and the rotor begins to accelerate. Then the rotor stays at 4000 rpm until the rotor chamber reaches predetermined vacuum level.
Press START.
Rotor accelerates to the set speed.
Reaches the set time, or press STOP.
Rotor decelerates to the zonal speed. The buzzer sounds.
Press VACUUM and open the chamber door.
Move the cursor to ZONAL in the Run Screen.
Unload the sample.
Press 0, 0, ENTER
Rotor comes to a stop. Temperature control is OFF.
Remove the rotor, then turn off
the POWER switch.
END
Page 29
Operation
Basic operating procedure
Thermo Scientific WX Ultra Series 2-13
Normal operation
Given below is a description of the operational procedure for a normal run.
Figure 2-6
Normal operation mode
Note Before starting up this machine, carefully read the operation manual for your rotor and make sure that you have selected the correct type of tubes and entered the
correct amount of sample.
Speed
Tim e
Step Operation and keystrokes Unit operation and considerations
1 Turn on the POWER switch on this machine. >> The panel display lights up.
>> The door is unlocked.
2 Open chamber door and make sure chamber is
dry.
3 Install the rotor. >> Before installation, read the rotor instruction manual carefully.
4 Set run conditions. >> See Section Setting run conditions and set run conditions.
5 Press VACUUM key. (optional) >> The machine starts evacuating the rotor chamber.
>> Temperature control starts. >> The degree of vacuum in the rotor chamber is displayed on the vacuum indicator on the display panel.
>> If the rotor compartment has moisture or frost on it, it takes a long time to reach a high vacuum. In that case, wipe it off with a clean, dry cloth or sponge. >> If the sample is sensitive to a temperature rise, do not press the START key until the chamber is at high vacuum level.
6 Press START key >> The rotor starts spinning.
>> The timer begins operating. >> The rotor accelerates to the set speed. >> This ultracentrifuge waits at 4,000 rpm until an intermediate vacuum is reached.
(1) In a low vacuum
(2) In an intermediate vacuum
(3) In a high vacuum
(1 indicator)
(2 indicators)
(3 indicators)
Page 30
Operation
Basic operating procedure
2-14 WX Ultra Series Thermo Scientific
The run mode indicator on the display panel displays the following:
Figure 2-7
Displays of run modes
7 The specified centrifugation time elapses
(time-out). Or press STOP key.
>> The rotor decelerates and stops.
8 The rotor brakes or coasts to a stop. >> Beeps to indicate that the rotor has stopped.
9 Press VACUUM key. >> The vacuum stops, the air leak valve gets activates, and the rotor chamber
reaches atmospheric pressure. >> The door unlocks, and is able to be opened and closed.
10 Take out the rotor.
Note Use the vacuum key to avoid prolonged vacuum wait time at 4,000 rpm when the ambient temperature is low.
Speed (rpm)
Time
Page 31
Operation
Basic operating procedure
Thermo Scientific WX Ultra Series 2-15
Zonal operation
Zonal operation is a mode of operation using a zonal rotor or continuous flow rotor for density gradient centrifugation with large amounts of sample or continuous sedimentation of particles from a large volume sample.
The following zonal rotors are available for operation in the Sorvall WX Ultra series ultracentrifuges:
TZ-28 sealed zonal rotor TCF-32 continuous flow zonal rotor
The TZ-28 can only be operated in a sealed zonal mode. Dynamic loading/unloading or continuous flow operation IS NOT allowed. The TCF-32 zonal rotor can be operated in a continuous flow mode requiring dynamic loading and unloading. The rotor comes with a kit to modify the Sorvall WX Ultra series.
The continuous flow zonal operation consists of the following three stages:
1. Centrifugation at low speed, called the zonal speed *, the gradient or sample being loaded in this stage.
2. Acceleration to set speed and separation of the sample, or
3. Centrifugation at zonal speed, the sample and or gradient being unloaded in this stage.
Figure 2-8
Zonal operation mode
* Zonal speed: The zonal speed is one that is required for loading and unloading the sample. The zonal speed is normally set at 3,000 rpm, but it can be set in the range from 2,000 to 3,000 rpm in increments of 100 rpm to obtain the required speed. For details of how to change zonal speed, see Section User Customizations.
The following explains how to perform a zonal run.
* Zonal speed
Tim e
Speed
WARNING When operating the centrifuge with the door open and the rotor spinning, eliminate all possibility of contact with the rotor. Remove any objects not essential to operating procedures which could either catch on or drop into the spinning rotor (that is, neckties, rings, bracelets, watches, long sleeves, or any loose articles in breast pockets).
Step Operation and keystrokes Unit operation and considerations
1 Install the zonal rotor on the drive spindle. >> See zonal rotor manual.
•The characters ZONAL are displayed in the message display field on the screen and the zonal operation mode is set.
Page 32
Operation
Basic operating procedure
2-16 WX Ultra Series Thermo Scientific
2 On the Run Screen, move the cursor to ZONAL
and enter the following by operating numeric keys to set the zonal operation mode.
3 Set the run conditions. >> For details, see Section Basic operation.
4 Press the START key.
Note The time being consumed at zonal speed is not counted as part of the run.
5 Install the seal assembly on the rotor. >> For details, see zonal rotor manual.
6 Load the sample and the gradient solution.
7 Install the cap assembly on the rotor.
8 Close the chamber door.
9 Press the VACUUM key. (This step may be omit-
ted.)
>> Vacuum pump begins to work. >> Chamber door is locked.
10 Press the START key again. >> Rotor begins to accelerate to the set speed.
Note The run time is counted from when the START key is pressed. The
instrument can also count only the time elapsing while the rotor is spinning at high speed. This is possible by changing the run time setting range. For details, see Section “Run time range setting”
on page 2-80.
11 If you need to stop the run before the set time
elapses, press the STOP key.
>> Rotor is decelerated to the zonal speed and then the buzzer sounds.
12 Press the VACUUM key. >> Vacuum pump stops working and air enters the rotor chamber.
>> Door lock is released.
• Rotor begins to accelerate to the zonal speed.
• Temperature control begins.
Page 33
Operation
Acceleration and deceleration rates
Thermo Scientific WX Ultra Series 2-17
Acceleration and deceleration rates
In order to meet various experimental protocols, the acceleration and deceleration rates can be adjusted between 0 to 500 rpm. The figure and table below show the relationship between ACCEL/DECEL code numbers selected and resulting approximate acceleration/deceleration times.
13 Open the chamber door. >> For details, see zonal rotor manual.
14 Remove the cap assembly
15 Install the seal assembly and unload the sam-
ple.
16 On the Run Screen, move the cursor to ZONAL
in the FUNCTION field and enter the following by operating numeric keys to set the normal operation mode.
>> Rotor decelerates to a stop.
Note During the zonal mode, the rotor will not decelerate even if the
STOP key is pressed during the rotor's spinning at the zonal sped.
17 Be sure that the rotor is at rest, then remove the
rotor.
>> If the rotor is still spinning, do not remove it. Wait until it comes to a complete stop.
Note By changing to the NORMAL mode when loading and unloading the sample, you can decelerate the rotor to a stop.
ACCEL/DECEL code numbers
Tim e
Speed (rpm)
Page 34
Operation
Acceleration and deceleration rates
2-18 WX Ultra Series Thermo Scientific
* Minimum time is the time for accelerating or decelerating by the driving motor with maximum torque. this time depends on the type of rotor in use.
The following are typical examples of application of acceleration and deceleration rates
Code No.
Acceleration time (minutes) from rest to 500 rpm
Deceleration time (minutes) from 500 rpm to rest
9 Minimum time* Minimum time*
81 1
72 2
63 3
54 4
45 5
36 6
27 7
18 8
F - Coasting deceleration
Suggested code nos.
Characteristic of separation
ACCEL DECEL
Density gradient centrifugat ion using a vertical rotor
5 7 The sample and gradient in tubes reorient during accel-
eration and deceleration. Therefore, the sample and gradient can become mixed, especially in wide tubes, if you use rapid acceleration or deceleration.
DNA separation by CsCl iso­pycnic separation (self-forming gradients)
9 7 You can operate at maximum acceleration because the
density gradient is not formed during the run. As for the deceleration, it is better to decelerate slowly to obtain sharp bands.
Pelleting using a fixed angle rotor
9 9 Rapid pelleting of samples is possible (the run time
decreases).
Density gradient centrifugat ion using a swinging bucket rotor
8 8 The sample and gradient do not reorient. Therefore,
mixing of the layers is less than that in the case of using a vertical rotor.
Note For a swinging bucket, there is no difference with regard to turbulence if ACCEL/DECEL is less than or equal to 8. However, when the mode for long acceleration time is selected, (1, 2, 3), an imbalance alarm indicator may light.
Page 35
Operation
How to use the FUNCTION field
Thermo Scientific WX Ultra Series 2-19
Figure 2-9 Operation with setting ACCEL = 0 and DECEL = 0
How to use the FUNCTION field
This ultracentrifuge incorporates a number of features. These features are displayed and specified in the FUNCTION field. The FUNCTION field is extended as shown in the following figure.
Figure 2-10
FUNCTION field
Allowable maximum speed
500 rpm
2 minutes
15 minutes
15 minutes 2 minutes
Speed
ENTER key
ENTER key
Page 36
Operation
How to use the FUNCTION field
2-20 WX Ultra Series Thermo Scientific
The above features can be used in combination. When all settings are entered, press ESC to move back to the Run Screen. Then enter a setting for another feature to form a combination.
ID CODE:
Sets an ID code.
Figure 2-11 Rotor radius
RLM:
Switches the Run Screen to the rotor management field.
PROG:
Programs, stores, and recalls run conditions. This feature also offers a step-mode operation: a continuous run of multiple run conditions.
RTC:
Sets a start time or a finish time and runs the ultracentrifuge at a desired date and time.
RCF
max
: Maximum centrifugal force for the maximum radius r
max
of the rotor used. This feature is used to cause the system to automatically calculate and display RCF
max
. It also sets an
RCF
max
value and calculates the speed.
RCF
avg
: Causes the system to automatically calculate and display the
average centrifugal force RCF
avg
for the average radius r
avg
of the rotor used. It also sets an RCF
avg
value and calculates
the speed.
ω2T: Performs an rω2T run and arithmetic operations.
DEF:
Turns on and off the defroster.
ZONAL:
Selects between zonal operation and normal operation.
Note To perform a combination of PROG and RTC, first set PROG and then set RTC. Once RTC is activated, you cannot change the run time. You therefore cannot activate PROG.
Page 37
Operation
How to use the FUNCTION field
Thermo Scientific WX Ultra Series 2-21
Programmed operation
This ultracentrifuge has a programmed operation feature that stores run conditions. Storing run conditions which you often use allows you to recall those conditions as often as you wish, thus saving time in setting. (The saved run conditions will remain in effect even after the power switch is turned off.) This centrifuge incorporates the program areas indicated below. It has twenty memory areas and nine steps in each memory number. Accordingly, twenty patterns of run conditions can be stored and each memory can store nine steps. Running this machine with each memory number retaining multiple steps will allow you to change the speed, run time, temperature, and others while in operation. (Step-mode operation).
Figure 2-12
Program areas
Page 38
Operation
How to use the FUNCTION field
2-22 WX Ultra Series Thermo Scientific
Basic operation of the programmed operation feature
Use cursor keys to
select and
press ENTER key.
ENTER key
ENTER key
ENTER key
Memory unit number ENTER key
Next page
Press ESC several times to go back to the Run Screen.
Programmed operation possible
Enter a memory number.
If you know which memory number stores the run condition.
Use cursor keys to
select and
press ENTER key.
To store, change or delete a program, or if you have forgotten the memory number where the run conditions are stored.
Press key to cause
to blink and
press ENTER key.
Enter a desired memory number and press ENTER key.
Page 39
Operation
How to use the FUNCTION field
Thermo Scientific WX Ultra Series 2-23
Note You cannot create, change, or delete a program while in running. Perform these operations while not in running. However, you can search the memory screen every time.
Page 40
Operation
How to use the FUNCTION field
2-24 WX Ultra Series Thermo Scientific
1. Programming procedure for run conditions (creating or changing)
Shown below is the procedure for storing (creating) or changing a run condition.
Step Key operation Screen display and considerations
1 While in the Run Screen, use the cur-
sor keys to move the cursor to PROG • RTC and press ENTER key. Then, move the cursor to PROG and press ENTER key.
2
Press the upward cursor key to cause PROGRAM to blink, then press
ENTER key.
>> If you wish to store a new condition, enter it into an empty memory number.
3 Use numeric keys to enter a memory
number you wish to store (or change). Example: To store a condition in mem­ory number 3, press
4 Select “Creating or Changing.”
• The FUNCTION field switches to the PROGRAM field.
• The MEMORY No. setting field blinks.
• The display switches to the Program Map screen and the MEMORY No. selection field blinks.
• The memory screen you have just specified appears.
• The message field in the bottom row switches to what is shown in the figure to the left.
Page 41
Operation
How to use the FUNCTION field
Thermo Scientific WX Ultra Series 2-25
5
Enter a desired condition. For example, select “SPEED”.
6 Select a desired rotor.
To select a desired rotor from the “Used Rotor List”:
To select a desired rotor from the “Rotor Catalog”: (1) Select “Rotor Catalog”.
(2) Select Fixed Angle Rotor in the rotor types. (When using fixed angle rotor T-880)
(3) Select fixed angle rotor T-880.
(4) Select “Yes”.
• The display switches to the Choose Rotor screen.
• The display switches to the Used Rotor List screen.
• You can search for the rotor number you need in the order starting with the latest date/time.
• The display switches to the Select Rotors screen.
• The display switches to the Fixed Angle Rotor list screen.
• The specifications of the T-880 are displayed.
Page 42
Operation
How to use the FUNCTION field
2-26 WX Ultra Series Thermo Scientific
6 To select a desired rotor from the
“Rotor Management”: (1') Select “Rotor Management”.
(2') Select a desired rotor using the upward or downward cursor key
.
(When using fixed angle rotor T-8100)
(3') Select “Yes”.
To select no rotor: Select “None”.
7 Enter values into SPEED (RCF
max
or
RCF
avg
or ω2T), TIME, TEMP, A
(ACCEL), and D (DECEL).
>> To enter run conditions in multiple steps in a step-mode operation, continue the operation from the above. (See Section Step-mode operation.) >> You cannot skip a step. (You cannot enter anything in step 2 with nothing entered into step 1.)
• The display switches to the Rotor Management screen and the Rotor List is displayed.
• The rotor type you have just entered appears in the top row in the screen.
• The SPEED field in the SPEED priority screen blinks.
• The SPEED field in the SPEED priority screen blinks.
Example
(Downward cursor key)
• When you finish entering run conditions for step 1, step 2 enters into an input wait state.
Page 43
Operation
How to use the FUNCTION field
Thermo Scientific WX Ultra Series 2-27
To change a run condition, use cursor keys to move to the item you want to change and then enter a value.
8
Press the downward cursor key to blink the value part of the message
field in the bottom row. Then select “Yes”.
Yes
No
>> If you select “Yes”:
>> If you select “No”, the message field in the bottom row switches to what you see in process 4. Select a new setting.
9 After entering memory run conditions,
perform an entry operations for User Name. (1) Press the ESC key.
(2) When attaching User Name to the memory run conditions, select “Yes”.
>> When you press ESC key, the bottom row message is displayed. (1) A selection field as to whether User Name should be input or deleted appears.
(2) A selection field for selecting either inputting or deleting appears.
(3) Select “Inputting User Name”.
(4) Enter 4 digits of the ID CODE that is already registered.
(5) Check User name and select “Yes”.
(3) The display switches to the ID CODE input field.
(4) The display switches to the User Name check field.
(5) The display switches to the Registration completion message field.
• This diagram shows what you will get if you move the cursor and enter the TIME field into an input wait.
• The message field in the bottom row switches to what is shown in the diagram to the left, and the storing of the run conditions in memory number 3 is completed.
ID code
Page 44
Operation
How to use the FUNCTION field
2-28 WX Ultra Series Thermo Scientific
10 Add Remark to the run conditions in
the memory. (1) Press the ESC key.
(2) To add Remark to the run conditions in the memory, select “Yes”.
(3) Select “Inputting Remark”.
(4) Enter Remark in 8 characters or less.
(5) After entering the characters for Remark, move the cursor to SET and press the ENTER key to register the character string. (6) Check the contents of Remark and select “Yes”.
>> When you press ESC key, the bottom row message is displayed. (1) A selection field to input or delete Remark appears.
(2) A selection field to select either inputting or deleting appears.
(3) The display switches to the Remark registration field.
(4) Remark is entered from the left.
(5) The display switches to the Remark check field.
(6) The display switches to the registration completion message field.
Page 45
Operation
How to use the FUNCTION field
Thermo Scientific WX Ultra Series 2-29
11 When the storing is over, press ESC
key several times.
>> The first push on ESC key switches the message display in the bottom row to what you see in process 3 as shown below. To run this centrifuge, select Run­ning.
>> Pushing ESC key a second time displays the Program Map screen shown in process 3.
>> The third push on ESC key switches you back to the Run Screen (program screen).
>> The fifth push on ESC key switches you back to the Run Screen (FUNCTION field).
• Make sure that the condition is stored in memory number 3.
Note
1. If you make and store changes in a memory area that already stores run conditions, the previous conditions are replaced by the new conditions.
2. You cannot store a run condition while running (while the rotor is spinning). Always perform this function while not running.
Page 46
Operation
How to use the FUNCTION field
2-30 WX Ultra Series Thermo Scientific
2. How to perform a programmed operation
Shown below is how to perform a “programmed operation”, that is, how to recall a stored set of run conditions and run this centrifuge accordingly.
(a) If you know which memory unit number you need
Step Key operation Screen display and considerations
1 Turn on the POWER switch on the cen-
trifuge.
>> The panel display appears. >> The door unlocks.
2 Install a rotor. >> Install the rotor securely on the shaft.
3 While in the Run Screen, use cursor
keys to move the cursor to PROG • RTC and then press ENTER key. Next, move the cursor to PROG and press ENTER key.
4 Enter the memory number you wish to
operate.
Example: to call memory number 3, press
>> The message field in the bottom row displays PROG. >> If the STEP field displays what is shown below, this means that the memory unit you have just called stores multiple run conditions (stepmode feature). For details, see Section Step-mode operation.
>> The system only accepts numbers 1-20.
5 Run the centrifuge under normal oper-
ation without making changes to the run conditions.
>> Run this machine according to Normal operation. >> If you make changes to the run conditions (such as SPEED and TIME) after call­ing a program, the program you have just called is canceled. You must call it again to use it.
• The FUNCTION field switches to the PROGRAM field.
• The MEMORY No. setting field blinks.
• The run conditions in memory appear in the respective fields.
In the Run Screen, the run conditions of this step are displayed (the first step
in this case).
Final step
Page 47
Operation
How to use the FUNCTION field
Thermo Scientific WX Ultra Series 2-31
(b) If you do not know which memory number you need
Step Key operation Screen display and considerations
1 Turn on the POWER switch of this
machine.
>> The panel display appears. >> The door unlocks.
2 Install a rotor. >> Install the rotor securely on the shaft.
3 While in the Run Screen, use cursor
keys to move the cursor to PROG • RTC and then press ENTER key. Then, move the cursor to PROG and press ENTER key.
4
Press the upward cursor key to blink PROGRAM. Then press ENTER
key.
5 Use numeric keys to enter the memory
number where you have stored data.
To call memory number 3, press
>>If the contents of the memory number you have just recalled is not what you want, press ESC to switch back to the Program Map screen and then call another memory unit number.
6 If you get the memory number you
want, select Running.
• The FUNCTION field switches to the PROGRAM field.
• The MEMORY No. setting field blinks.
• The display switches to the Program Map screen and the MEMORY No. selection field blinks.
• The image of the specified memory number appears.
• The message field in the bottom row switches to what you see below.
Page 48
Operation
How to use the FUNCTION field
2-32 WX Ultra Series Thermo Scientific
Step-mode operation
This ultracentrifuge incorporates a step-mode operation feature, which stores two or more run conditions in one program memory area and switches between different values of speed, running time, temperature, and other parameters while in operation. This centrifuge can store up to nine steps. This section explains how to make settings by showing some examples.
(1) How to activate a step-mode operation
[Typical settings] Shown below is the example of a three-step run and how to activate a step-mode operation.
7 To run the centrifuge, select “Yes”.
Yes
No
>> If you select “Yes”
>> The message field in the bottom row displays PROG. >> If the STEP field displays what is shown in the figure below, that means that the memory number you have just called stores multiple run conditions (step-mode feature). For details, see Section Step-mode operation.
>> If you select “No”, the message field in the bottom row displays what you see in process 5. Select a new condition.
8 Start running the centrifuge under nor-
mal operation with no changes made to the run conditions.
>> Run this machine according to Section Normal operation. >> If you make changes to the run conditions (such as SPEED and TIME) after recalling a program, the program you have just recalled will be canceled. You must recall the program again to use it.
• The display switches back to the Run Screen.
• The run conditions in memory appear in the respective fields.
In the Run Screen, the run conditions of this step are displayed (the first step in this case).
Final step
Note
1. To check the contents (run conditions) of the memory number for the program while in running, follow processes 3 through 5 in (b) “If you do not know which memory unit number you need.” After the check, press ESC to get back to the Run Screen.
2. To perform a combination of a programmed run with RTC (real-time control) (see Section
RTC (real-time control) feature), first recall a programmed memory number, then set RTC.
The system will then calculate the total of the running times of all steps of the programmed run and calculate the start time for RTC. Therefore, you cannot recall the program memory after setting RTC.
Page 49
Operation
How to use the FUNCTION field
Thermo Scientific WX Ultra Series 2-33
Figure 2-13 A typical step-mode run
Run time
Step 1
Step 2
Step 3
Speed (rpm)
Page 50
Operation
How to use the FUNCTION field
2-34 WX Ultra Series Thermo Scientific
Step Key operation Screen display and considerations
1 While in the Run Screen, use the cur-
sor keys to move the cursor to PROG • RTC and press ENTER key. Then move the cursor to PROG and press ENTER key.
2
Press the upward cursor key to cause PROGRAM to blink, then press
ENTER key.
3 Use numeric keys to enter the memory
unit number you want to store.
To store data in memory number 3, press
• The FUNCTION field switches to the PROGRAM field.
• The MEMORY No. setting field blinks.
• The display switches to the Program Map screen and the MEMORY No. selection field blinks.
• To store a new condition, enter it into an empty memory number.
• The image of the specified memory number appears.
Page 51
Operation
How to use the FUNCTION field
Thermo Scientific WX Ultra Series 2-35
4 Select “Creating or Changing.”
5
Enter a desired condition. To select SPEED, press
To select no rotor, press.
>> To select and enter RCF
max
or RCF
avg
, see Section 1. Programming procedure for
run conditions (creating or changing)
6 Specify run conditions for step 1.
>> If you press the cursor key (downward) after setting numeric value for DECEL (D) in step 1, setting the run conditions in step 1 finishes and step 2 enters
into an input wait state.
7 Then specify run conditions for step 2.
• The message field in the bottom row switches to what you see on your left-hand side.
• The display switches to the
Choose Rotor screen.
• The display switches to the
SPEED priority setting field and the SPEED area blinks.
Downward cursor key
• The entry of run conditions for
step 1 is complete.
Downward cursor key
• The entry of run conditions for
step 2 is complete.
Page 52
Operation
How to use the FUNCTION field
2-36 WX Ultra Series Thermo Scientific
8 Lastly, specify run conditions for step
3.
9
Press the downward cursor key to blink the value part of the message
field in the bottom row. Then select “Yes”.
Yes
No
>> If you select “Yes”:
If you select “No”, the message field in the bottom row switches to what you see in process 4. Input a new setting.
10 After entering step-mode run condi-
tions, perform an entry operations for User Name.
(1) Press the ESC key.
(2) When attaching User Name to the memory run conditions, select “Yes”.
(3) Select “Inputting User Name”.
(4) Enter 4 digits of the ID CODE that is already registered.
(5) Check User Name and select “Yes”.
>> When you press ESC key, the bottom row message is displayed. (1) A selection field as to whether User Name should be input or delete appears.
(2) A selection field for selecting either inputting or deleting appears.
(3) The display switches to the ID CODE input field.
(4) The display switches to the User Name check field.
(5) The display switches to the Registration completion message field.
Downward cursor key
• The entry of run conditions for step 3 is complete and step 4 is ready for input.
ID code
Page 53
Operation
How to use the FUNCTION field
Thermo Scientific WX Ultra Series 2-37
11 Add Remark to the run conditions in
the memory. (1) Press the ESC key.
(2) To add Remark to the run conditions in the memory, select “Yes”.
(3) Select “Inputting Remark”.
(4) Enter Remark in 8 characters or less.
(5) After entering the characters of Remark, move the cursor to SET and press the ENTER key to register the character string.
(6) Check the contents of Remark and select “Yes”.
>> When you press ESC key, the bottom row message is displayed. (1) A selection field to input or delete Remark appears.
(2) A selection field to select either inputting or deleting appears.
(3) The display switches to the Remark registration field.
(4) Remark is entered from the left.
(5) The display switches to the Remark check field.
(6) The display switches to the registration completion message field.
12 Press ESC key when the storing is over.
13 Select “Running”.
• The message field in the bottom row switches to what is shown in the diagram on your lefthand side, and the storing of the run conditions in the appropriate memory number is completed.
Page 54
Operation
How to use the FUNCTION field
2-38 WX Ultra Series Thermo Scientific
1. Step display
The PROGRAM field displays steps as follows:
When the system finishes separating step 1 and moves to step 2,
When the system moves from step 2 to the final step 3,
Thus, you can see at a glance how many steps are stored in the specified memory and which step the centrifuge is following.
2. While the instrument is running (while the rotor is spinning), you cannot store a stepmode run condition. Always perform this function while not in running.
3. To check the contents (run conditions) of a step-mode program memory number, follow processes 1 through 3. To get back to the Run Screen after the check, press ESC key.
4. To perform a combination of a step mode run with an RTC (real-time control) run (see Section
RTC (real-time control) feature), call a program memory number, then set RTC.
The system then calculates the total of running times of all steps of the programmed run and calculates the startup time for RTC. You therefore cannot call a program memory number after setting RTC.
14 To run the centrifuge, select Yes.
Yes
No
>> If you select “Yes”:
>> The message field in the bottom row displays PROG.
>> If you select “No”, the message field in the bottom row displays what you see in process 10. Make the selection again.
15 Start running the centrifuge under nor-
mal operation with no changes made to the operational conditions.
>> Run the centrifuge according to Normal operation. >> If you make changes to the run conditions (such as SPEED and TIME) after call­ing a program, the program you have just called gets canceled. You must call the program again. >> Pressing START makes the rotor rotate under the run condition of step 1. >> When step 1 ends, the system moves to steps 2 and 3 automatically.
The display switches back to the
Run Screen.
• The run conditions of step 1 in memory appear in the respective fields.
In the Run Screen, the run conditions of this step are displayed (step 1 in this case).
Final step
Page 55
Operation
How to use the FUNCTION field
Thermo Scientific WX Ultra Series 2-39
(2) Deleting a program
This section explains how to delete a set of run conditions stored in program memory. To delete a memory number, delete all steps in that memory number.
Note While the instrument is running, you cannot delete a program memory number.
Step Key operation Screen display and considerations
1 While in the Run Screen, use cursor
keys to move the cursor to PROG • RTC and press ENTER key. Then move the cursor to PROG and press ENTER key.
2
Press the upward cursor key to cause PROGRAM to blink, then press
ENTER key.
3 Use numeric keys to enter the memory
unit number you want to delete.
To delete memory unit 3, press
4 Select “Deleting”.
• The FUNCTION field switches to the PROGRAM field.
• The MEMORY No. setting field blinks.
• The display switches to the Program Map screen and the MEMORY No. field blinks.
• The image of the memory unit you have just specified appears.
• The message field in the bottom row switches to what you see on your left.
Page 56
Operation
How to use the FUNCTION field
2-40 WX Ultra Series Thermo Scientific
(3) Other procedures
1. Making changes to the run conditions Call the memory number you want to make changes to and make the changes. Alternatively, delete the memory number, then store a new set of run conditions. To make a change that will result in fewer steps, delete the memory number and then enter a new set of run conditions and store them. While in a programmed run, modification of a run condition is limited to the step which is currently running.
2. Running the centrifuge starting from an intermediate step You cannot run the centrifuge starting from an intermediate step in a memory number that contains multiple steps. Store (register) run conditions for the intermediate step and later steps in another memory number. Then recall the memory number and run it.
3. What if a “SPEED” alarm goes on? If a step stores a speed exceeding the maximum allowable speed of your rotor, the system will detect it in the step 1 run and display the “SPEED” alarm. Double-check the speed of all steps and correct any wrong ones.
4. Stopping the centrifuge in operation Press STOP. The rotor stops and the system does not move to the next step.
RTC (real-time control) feature
The Sorvall WX Ultra series ultracentrifuges contain an internal clock, allowing you to run the machine at a specified start or finish time for centrifugation. This feature for running the machine at a specified time is called the RTC (real-time control) feature. Below is an example of how to perform an RTC run.
Example: If you wish to install your rotor on the centrifuge under the run conditions listed below on the evening of October 1 and take the samples out around 9:30 a.m. the next morning;
1. Rotor: T-8100
2. RPM: 100,000 rpm
3. Separation time: 4 hours
5 To delete this memory number, select
Yes.
Yes
No
>> If you select “Yes”.
>> If you select “No”, the screen switches to what you see in process 3. Make the selection again.
6 Press ESC several times to move back
to the Run Screen.
>> If you delete a memory number, you can no longer run the centrifuge according to that number.
• The system deletes all run conditions in memory number 3. (While in step mode, all steps get deleted.)
Page 57
Operation
How to use the FUNCTION field
Thermo Scientific WX Ultra Series 2-41
4. Control temperature: 4 °C
5. Acceleration mode: 9
6. Deceleration mode: 7
Figure 2-14 A typical RTC run
In this example, you set the above run conditions (2) through (6), set the finish time for RTC run to 9:30 a.m., October 1 and start the centrifuge. (You can make an identical setting by setting the start time to 5:30 a.m. instead of setting the finish time to 9:30 a.m.)
100,000
Speed (rpm)
START key ON
4:30 p.m, October 1
Automatic start at 5:30 a.m., October 2
4 h
Stop around 9:34 a.m.
Finish time
9:30 a.m.
Page 58
Operation
How to use the FUNCTION field
2-42 WX Ultra Series Thermo Scientific
1. How to perform an RTC run Step Key operation Screen display and considerations
1 Set operational conditions while in the
Run Screen. Use cursor keys to enter the SPEED part into an input wait, then press
Move the cursor to PROG • RTC and press ENTER key.
2 Move the cursor to RTC and press
ENTER key.
3 Select the item you wish to set. To
select the “Setting stop time”, press
>> Setting the “Setting stop time”
>> Setting the “Setting start time”
• The display switches to the RTC Setup screen.
• The “dd” item blinks. “mm” and “dd” indicate the current date as a default.
• If you do not make any input for 30 seconds, the cursor disappears, and you cannot make an input. At that time, press a cursor key.
Page 59
Operation
How to use the FUNCTION field
Thermo Scientific WX Ultra Series 2-43
4 Use cursor keys and numeric keys to
enter month, day, hours, and minutes.
Then, press ENTER key.
>> To set the hours, use a number between 0 and 23 (in the 24-hour system). >> Enter a time later than the current time. When setting the “Setting stop time”, allow for centrifugation time and set the centrifugation start time to a time later than the current time. >> You cannot set the centrifugation start time to a date more than 20 days ahead.
5 If you are sure that you have made the
correct settings, select Yes.
Yes
No
>> If you select “Yes”
>> If you select “No”, you will get the display you see in process 2. Make the selection again.
6 Check the RTC setting in the Run
Screen, then press START key.
The RTC run does not get activated unless you press START key.
>> Run this machine according to Section Normal operation. >> Once the RTC setting is made, you cannot change the run time (centrifugation time). If you wish to make a change to the run time, cancel RTC first. >> Pressing START enters the system into a “DELAY” (and vacuum comes on), condition causing the centrifuge to wait until the set time comes. This centrifuge will turn the rotor automatically at the specified time and run it for the specified period.
• The display switches to the Run Screen and the specified RTC time appears.
• The message field displays RTC.
• The run mode in the message display field switches to “DELAY”.
• Blink the PROG • RTC in the FUNCTION field.
• To specify the run time (centrifugation time), do not set HOLD. Use ten keys to enter the run time instead.
Page 60
Operation
How to use the FUNCTION field
2-44 WX Ultra Series Thermo Scientific
1. For an RTC run, the run mode indicator on the display panel displays the following:
Figure 2-15
Run mode display (RTC)
2. You cannot make an RTC setting in any of the following cases:
a. When the Run Screen is set to HOLD (continuous run)
Set the run time (centrifugation time) not to HOLD but to a numerical value.
b. When it is past the start time
Set the start time to a time later than the current time.
c. When the start time is more than 20 days after the current time
Set the time to a time no more than 20 days afterwards.
3. To change the run time (centrifugation time) after making an RTC setting, cancel RTC and then set a new run time.
4. To perform a combination of a programmed operation (including a step-mode operation) with an RTC run, call a program memory number, then set RTC. The system calculates the total run times of all steps of the programmed operation and calculates the start time for RTC. Therefore, you cannot recall the program memory number after setting RTC.
5. To stop this ultracentrifuge in RTC operation, press STOP key. The system then stops RTC and stops the rotor.
Speed (rpm)
START key ON
Automatic start
Tim e
Page 61
Operation
How to use the FUNCTION field
Thermo Scientific WX Ultra Series 2-45
Making changes to the RTC settings
This section shows how to cancel or make changes to the RTC settings you have made. If you have already started an RTC run, press STOP to abort the RTC first. Press STOP to cancel the RTC run.
Step Key operation Screen display and considerations
1 (1) While in the Run Screen, use cursor
keys to move the cursor to PROG • RTC and then press ENTER key.
(2) Use cursor keys to move the cursor to RTC and then press ENTER key.
>> The message indicator displays RTC.
2 Press cursor keys to blink RTC and then
press ENTER key.
• The FUNCTION field switches to the RTC field.
Page 62
Operation
How to use the FUNCTION field
2-46 WX Ultra Series Thermo Scientific
Canceling the RTC settings
The canceling procedure varies depending on whether the RTC run is started (START is pressed) or
not.
3 Making a change to date and time set-
tings (1) Select “Setting stop time” or “Set­ting start time”. To change the “Setting stop time”, press
(2) Perform process 4 and later ones in
1. How to perform an RTC run men-
tioned above and change the date and time settings.
See Section 1. How to perform an RTC run.
• The screen switches to a screen where you can enter a stop time (or start time).
Step Key operation Screen display and considerations
1 (1) While in the Run Screen, use cursor
keys to move the cursor to PROG • RTC and then press ENTER key.
(2) Use cursor keys to move the cursor to RTC and then press ENTER key.
>> The message indicator displays RTC.
• The FUNCTION field switches to the RTC field.
Page 63
Operation
How to use the FUNCTION field
Thermo Scientific WX Ultra Series 2-47
2 Press cursor keys to blink RTC and then
press ENTER key.
3 [Before starting the RTC run]
(1) Select “Deleting the set RTC”.
(2) Select “Yes”.
>>If you select “Yes”
>> If you select “No”, you will get the display you see in process 2. Make a selec­tion again.
[After starting the RTC run] (1') Select “Deleting the set RTC”.
(2') i) Select “Cancel RTC-NO run”.
• The message indicator in the bottom row changes.
• The RTC setting has been canceled.
• The message indicator in the bottom row changes.
Page 64
Operation
How to use the FUNCTION field
2-48 WX Ultra Series Thermo Scientific
Displaying and setting RCF
This ultracentrifuge stores the maximum and average radii of each rotor in internal memory. Setting a speed causes this centrifuge to automatically calculate and display the RCF (relative centrifugal force) value, while setting an RCF value causes the centrifuge to automatically calculate and display the speed. Given below is a description of how to display and set RCF.
1. How to display an RCF value
• When the rotor to be used is already identified
(when rotor type is displayed in the message display
field)
ii) Select “Cancel RTC-Start run”.
iii) Select “Continue RTC”.
(3) Press ESC after process (2') i) or ii).
>> The RTC in the message indicator in the bottom row disappears.
• The display switches back to the Run Screen.
Page 65
Operation
How to use the FUNCTION field
Thermo Scientific WX Ultra Series 2-49
Step Key operation Screen display and considerations
1 While in the Run Screen, set a speed. >> See Section Setting run conditions.
2
Use cursor keys to move the cursor to
RCF • ω
2
T and press ENTER key.
3 Move the cursor to RCF
max
(or RCF
avg
)
and press ENTER key.
>> The RCF value associated with the set speed you have just entered in process 1 appears in the RCF setting field.
•The FUNCTION field switches to what is shown in the diagram to the left, and the RCF
max
field blinks.
• The RCF display field displays the RCF value of the rotor.
RCF display field
RCF setting field
Note Press ESC key to cancel the RCF field.
Page 66
Operation
How to use the FUNCTION field
2-50 WX Ultra Series Thermo Scientific
• When the rotor to be used is not identified (when rotor type is displayed in the message display field)
Step Key operation Screen display and considerations
1 Set the speed on the Run Screen. >> See Section Setting run conditions.
2 (1) Use cursor keys to move the cursor
to RCF • ω
2
T and press the ENTER
key.
(2) Use cursor keys to move the cursor to RCF
max
(or RC F
avg
) and press the ENTER
key.
3 Check that ROTOR LIST blinks and
press the ENTER key.
• The contents of the display in the FUNCTION field change.
•The RCF
max
area blinks.
• The FUNCTION field switches to the RCF field.
• ROTOR LIST blinks.
Page 67
Operation
How to use the FUNCTION field
Thermo Scientific WX Ultra Series 2-51
4 Select a desired rotor. To select a
desired rotor from the “Used Rotor List”:
To select a desired rotor from the “Rotor Catalog”: (1) Select “Rotor Catalog”.
(2) Select Fixed Angle Rotors in the rotor types. (When using fixed angle rotor T-890)
(3) Select fixed angle rotor T-890.
4 (4) After checking the specifications of
the rotor, select “Yes”.
>> When you select “Yes”
• The RCF value of the running rotor is displayed in the RCF display field.
>> When you select “No”, the display goes to the same as the display of process 4 (2).
• The display switches to the Used Rotor List screen.
• You can search for the rotor number you need in the order starting with the latest date/time.
• The display switches to the Rotor List screen.
• The fixed angle rotor list is displayed.
• The specifications of the T-890 rotor are displayed.
• The display switches back to the Run Screen and the selected rotor type is displayed in the message display field of the bottom row.
RFC display field
The display switches to the Choose Rotor screen.
Page 68
Operation
How to use the FUNCTION field
2-52 WX Ultra Series Thermo Scientific
4 To select a desired rotor from the
“Rotor Management”: (1') Select “Rotor Management”.
(2') Select a desired rotor using the upward or downward cursor key
.
(When using fixed angle rotor T-8100)
(3’) Select “Yes”.
>> The RCF value of the running rotor is displayed in the RCF display field.
• The display switches to the Rotor Management screen and the Rotor List is displayed.
• The display switches back to the Run Screen and the selected rotor type is displayed in the message display field of the bottom row.
RFC display field
Page 69
Operation
How to use the FUNCTION field
Thermo Scientific WX Ultra Series 2-53
How to set an RCF value
Enter a rotor number and an RCF value, and the machine will calculate, set, and display the speed. Shown below is the procedure.
Step Key operation Screen display and considerations
1 While in the Run Screen, set a speed. >> See Section Setting run conditions.
2 (1) While in the Run Screen, use cursor
keys to move the cursor to RCF • ω2T and press ENTER key.
(2) Move the cursor to RCF
max
(or
RCF
avg
) and press ENTER key.
3 Check that ROTOR LIST blinks and
press the ENTER key.
• The FUNCTION field switches to what is shown in the diagram to the left.
•The RCF
max
field blinks.
•The FUNCTION field switches to the RCF field.
• The ROTOR LIST field blinks.
• The display switches to the Choose Rotor screen.
Page 70
Operation
How to use the FUNCTION field
2-54 WX Ultra Series Thermo Scientific
4 Select a desired rotor.
To select a desired rotor from the “Used Rotor List”: (1) Enter No. that is found in the Rotor List.
To select a desired rotor from the “Rotor Catalog”: (1)' Select “Rotor Catalog”.
(2)' Select Fixed Angle Rotors in the rotor types. (When using fixed angle rotor T-890)
(3)' Select fixed angle rotor T-890.
(4') After checking the specifications of the rotor, select “Yes”.
>> When you select “Yes”
>> The RCF value corresponding to the currently set speed is displayed in the RCF setting field. (--- --- --- --- --- ---: Displayed at the initial time of no setting.) >> When you select “No”, the display goes to the same as the display of process 4 (2'). Perform setting again.
• The display switches to the Used Rotor List screen.
• You can search for the rotor number you need in the order starting with the latest date/time.
The display switches to the Rotor
List screen.
The angle rotor list is displayed.
• The specifications of the T-890 rotor are displayed.
• The display switches back to the Run Screen and the selected rotor type is displayed in the message display field of the bottom row.
RCF display field
RCF setting field
Page 71
Operation
How to use the FUNCTION field
Thermo Scientific WX Ultra Series 2-55
4 To select a desired rotor from the
“Rotor Management”: (1) Select “Rotor Management”.
(2) Select a desired rotor using the upward or downward cursor key
.
(When using fixed angle rotor T-8100)
(3) Select “Yes”.
>> The RCF value corresponding to the currently set speed is displayed in the RCF setting field. (--- --- --- --- --- ---: Displayed at the initial time of no setting.)
5 Use cursor keys to move the cursor to
the RCF setting field, and enter a desired centrifugal force. Example: RCFmaX = 5000 x 100g
• Specify the RCF value to hundreds. If you make no input for 30 seconds, the cursor will disappear, and you will be unable to make an input. At that time, press a cursor key.
• The display switches to the Rotor Management screen and the Rotor List is displayed.
• The display switches back to the Run Screen and the selected rotor type is displayed in the message display field of the bottom row.
RCF display field
RCF setting field
• The value you have just entered appears in the RCF setting field.
• Based on the rotor and RCF value, the system automatically calculates the speed and displays it in the SPEED display field.
Page 72
Operation
How to use the FUNCTION field
2-56 WX Ultra Series Thermo Scientific
Setting ω2T
This machine is provided with a function to perform an ω2T run. To perform an ω2T run, set the ω2T value instead of the run time.
Note
1. Press ESC key to cancel the RCF field.
2. When you run this centrifuge at a speed determined based on an RCF value, a slight rounding may occur (of up to 2%) between the set RCF value and the actual value, because the speed setting is set in increments of 100 rpm.
Speed (rpm)
Tim e
ω2T
2
T run]
Page 73
Operation
How to use the FUNCTION field
Thermo Scientific WX Ultra Series 2-57
Step Key operation Screen display and considerations
1 Set the Speed in the Run Screen. >> See Section Setting run conditions.
2 (1) Use cursor keys to move the cursor
to RCF • w
2
T on the Run Screen and
press the ENTER key.
(2) Use cursor keys to move the cursor
to w
2
T and press the ENTER key.
3 Use cursor keys to move the cursor to
the w
2
T setting field and enter a
desired w
2
T value.
Example: 9.87 x 1011
4 When you press START key, the
machine starts to perform an w
2
T run.
• The contents of the display in the FUNCTION field change.
ω
2
T blinks.
• The FUNCTION field switches to the ω
2
T field.
ω2T display field
ω2T setting field
• The value you entered is displayed in the ω
2
T setting
field.
• The time value is calculated from the speed and the ω
2
T
value and the run time is displayed in the Time Setting field.
• Arithmetic and integral operations are performed in the ω
2
T display field.
ω2T display field
Page 74
Operation
How to use the FUNCTION field
2-58 WX Ultra Series Thermo Scientific
Defrost function
If the chamber door is left open between low temperature runs, moisture from the ambient air humidity can condense (and even freeze) on the cold evaporator chamber walls. The potential for condensation increases as the ambient air temperature and humidity increases. If a run is performed with a significant amount of moisture in the evaporator chamber, the length of time required to reach a high vacuum will be increased.
To minimize condensation, when the evaporator chamber is cold, we recommend that you keep the chamber door closed as much as possible. If moisture does accumulate, before performing the next run, we recommend that you wipe the chamber dry. For situations where condensation has frozen to the chamber walls, a defrost function is provided as a convenient means to melt ice buildup so that the moisture can then be removed.
How to use the defrost function:
Defrost operation will heat the evaporator gradually and after the lapse of 10 minutes, the defrost operation and vacuum ends automatically. (After the end of the defrost operation, the DEF display in the message field disappears.) During the defrost operation, DEF in the message display field in the bottom row on the Run Screen blinks. To abort or end the Defrost Function, highlight DEF and press ENTER. To turn off vacuum, press vacuum key.
Note When the ω
2
T display field reaches the set value of ω2T, the machine decelerates and stops.
During deceleration, the
ω2T display field continues to accumulate value until the rotor stops and
the run ends.
Step Key operation Screen display and considerations
1 Use cursor keys to move the cursor to
DEF on the Run Screen.
2 With DEF highlighted, press the ENTER
key.
• Use cursor keys to highlight the DEF function.
• DEF blinks in the message field to notify that a defrost operation and vacuum has started.
• Vacuum indicators will also be displayed.
Page 75
Operation
How to use the FUNCTION field
Thermo Scientific WX Ultra Series 2-59
Setting and resetting zonal operation
Selection between normal operation and zonal operation is performed by special operations using the FUNCTION field on the Run Screen.
1. Setting zonal operation
Note
1. Before starting defrost operations, wipe away any moisture and close the door.
2. The temperature rise of the rotor due to defrost operations (for 10 minutes) should be 1 C or less.
3. During rotor acceleration or at set speed, defrost can be set but the defrost function is not available.
4. If DEF is selected during accel or set speed, it will not function until the run times out or stop is pressed.
Step Key operation Screen display and considerations
1 On the Run Screen, use cursor keys to
move the cursor to ZONAL in the FUNCTION field, and press
or
• ZONAL is displayed in the message field so that the machine is ready to start zonal operation.
Page 76
Operation
Rotor management
2-60 WX Ultra Series Thermo Scientific
2. Resetting zonal operation
Rotor management
It is very important to manage the life of each rotor you use with the ultracentrifuge. The warranted life of the rotor is different depending on the type of rotor, the material the rotor is fabricated from and the policy of the manufacturer of the rotor. The warranted life of a Thermo Ultracentrifuge rotor is defined by a maximum number of runs. Other manufacturers use a combination of either the maximum number of runs or a maximum number of hours to limit the warranted life of the rotor.
For some rotors, there exists a primary and a secondary life. The primary life is defined as the initial usage at speeds up to the maximum rated speed of the rotor. Once this primary life is expended, the rotor has a secondary life at speeds up to a derated maximum speed of the rotor-typically 90% of the initial maximum rated speed.
Rotor life management
In order to comply with warranty requirements, rotor use must be documented in a rotor log book provided with the ultracentrifuge.
Additionally, the Sorvall WX Ultra series ultracentrifuges have the capability to keep track of the number of runs and hours of use for each rotor. This information can be used to automatically notify the user of the need for deration at the end of the primary life and to preclude the use of rotors once their warranted life as defined by these parameters have come to an end.
Step Key operation Screen display and considerations
1 On the Run Screen, use cursor keys to
move the cursor to ZONAL in the FUNCTION field, and press
or
• ZONAL in the message field disappears and the machine status switches to normal operation.
Note If no key operation is performed for 30 seconds, the cursor disappears and any entry is disabled. At that time, press the cursor key to cause the cursor to blink once again.
Page 77
Operation
Rotor management
Thermo Scientific WX Ultra Series 2-61
To utilize this feature, each time a rotor is used it must be recorded by the ultracentrifuge. To do this, a rotor must be initially registered in the Rotor Management software. Each time the rotor is used in the Sorvall WX Ultra series ultracentrifuge, it must be selected from the Rotor Management Screen prior to starting the run. Additionally, should the rotor be used in other ultracentrifuges, the Rotor Management information can be edited to record this usage. Rotors may also be deleted from the Rotor Management Screen if they are no longer to be used.
If you use the rotor beyond its maximum permitted life, a serious accident may occur. (For more information, see rotor instruction manual.)
Automatic rotor deration
You can register a rotor in the ultracentrifuge before use. After the rotor has been registered, you can utilize the rotor management feature of the ultracentrifuge each time a run is made using the rotor. When the registered rotor comes close to the end of its life, the ultracentrifuge indicates it by displaying a message on the screen. When the rotor reaches its primary life, the ultracentrifuge derates its maximum permitted speed and inserts [] at the right side of the rotor S/N. Once the maximum permitted speed is derated, the actual speed of the rotor is automatically restricted within the new limits.
Automatic rotor exclusion
When a registered rotor reaches the end of its life, the ultracentrifuge indicates it by displaying a message on the screen. Subsequent runs of this registered rotor are precluded.
Registering a rotor
A standard rotor can be registered when you use the rotor management feature of the ultracentrifuge. If the rotor has not been registered, the feature does not operate for the rotor. To register a rotor, use the procedure described below.
As an example, suppose that you need to register a fixed angle rotor with the following data, and that it has been used with an ultracentrifuge of a different model from the one mentioned below: Model: T-890 Serial number: 9910876 Total number of runs thus far made: 500 Total number of hours thus far spent in the runs: 1200 hours, 24 minutes
Note The rotor life management data stored in the ultracentrifuge is not covered by the warranty. When a rotor is used with the ultracentrifuge, be sure to record the rotor management data in the rotor log book. (You can also paste a printout in the log book.) If the rotor log book is not maintained correctly, the rotor will not be warranted.
disk
Page 78
Operation
Rotor management
2-62 WX Ultra Series Thermo Scientific
Step Key operation Screen display and considerations
1 While in the Run Screen, use the cur-
sor keys to move the cursor to RLM and press ENTER key.
Switch each page of the Rotor List by using the left and right arrow keys
().
2
Use the cursor keys ( and ) to
move the cursor to the line that is not yet registered (New Rotor), and press the ENTER key.
3 Select “Creating or Changing”.
4 Select “Fixed Angle Rotor” in the Rotor
List screen.
and
• The display switches to the Rotor Management screen and the Rotor List is displayed.
• The display switches to the Rotor List screen.
• The display switches to the Fixed Angle Rotor List screen
Page 79
Operation
Rotor management
Thermo Scientific WX Ultra Series 2-63
5 Select “T-890” rotor.
6 After checking the specifications of the
rotor, enter “Yes”.
7 As the cursor has been moved to the
S/N field, enter serial No. of the rotor by using numeric keys. (Do not press the ENTER key.)
Registration of an unused rotor is fin­ished by pressing the ENTER key.
8 Enter the total number of runs in the
Runs field. (Do not press the ENTER key.)
9 Enter the total number of hours in the
Hours field by using numeric keys. (Do not press the ENTER key.) Note: Enter this running time up to one decimal place in units of hour. 1200 hours 24 minutes ---> 1200.4
10 Press the ENTER key.
• Specifications of the T-890 rotor are displayed.
• The display switches to the Rotor Management screen.
• You can enter a serial number of the rotor.
• The cursor moves to the Runs field. You can enter the total number of runs.
• The cursor moves to the Hours field. You can enter the number of hours.
Page 80
Operation
Rotor management
2-64 WX Ultra Series Thermo Scientific
Thus the rotor has been registered.
Deleting a registered rotor
If a registered rotor becomes unnecessary, you can delete it from ultracentrifuge memory.
As an example, suppose that you need to delete the T-890 fixed angle rotor registered in the previous section.
Note When the ENTER key is pressed in the above steps “7” to “9”, “0” is registered as a numeric value in each step. When the serial number must be corrected, the rotor must be deleted once and the value must be reentered from step “1”. (See the next page.) In the steps “7” to “9”, it is possible to change the value repeatedly by moving the cursor with the arrow keys unless the ENTER key is pressed.
Note Once a registered rotor is deleted, all of the life management data for the rotor will be lost.
Step Key operation Screen display and considerations
1 While in the Run Screen, use the cur-
sor keys to move the cursor to RLM and press ENTER key.
2 Use cursor keys to move the cursor to
the T-890 line to be deleted and press the ENTER key.
3 Select “Deleting” by pressing the fol-
lowing keys.
4 Select “Yes” by pressing the following
keys.
>> If you select “No”, you will get the display you see in process 2.
• The display switches to the Rotor Management screen.
Page 81
Operation
Rotor management
Thermo Scientific WX Ultra Series 2-65
Updating the existing data for a rotor
When a rotor is registered in the Sorvall WX Ultra series ultracentrifuges without a Rotor Management function, both Total number of runs and Total number of hours can be updated. The updated value is accepted only when it is greater than the registered value. It must not be smaller than the registered value.
To update the existing life management data for a standard rotor, perform the following procedure:
Step Key operation Screen display and considerations
1 While in the Run Screen, use the cur-
sor keys to move the cursor to RLM and press ENTER key.
2 Use cursor keys to move the cursor to
the T-890 line to be updated and press the ENTER key.
3 Select “Creating and Changing” by
pressing the following keys.
4 Enter the updated total number of runs
and the total number of hours by using numeric keys. Total number of runs
500.0 ---> 505.0 Total number of hours
1200.4 ---> 1210.4
• The display switches to the Rotor Management screen.
• In the Runs field, the cursor waits for an input.
Page 82
Operation
When power failure occurs
2-66 WX Ultra Series Thermo Scientific
Thus, the current total number of runs and hours have been updated.
When power failure occurs
1. Rotation of the rotor
The rotor coasts to a stop. If the power is restored when the current speed is more than 500 rpm, the rotor will be automatically accelerated to the set speed. If it is restored when the current speed is less than 500 rpm, the rotor will decelerate to a stop.
2. Displays on the panel
All displays on the panel go out. The set values are all backed up by battery during the power failure. When the power is restored, the displays will come on as before the power failure; the set values will be restored; and the message “Power failure” will appear in the message display.
3. Removing the rotor from the ultracentrifuge
First open the chamber door, then remove the rotor according to the following procedure: (1) Check that the rotor is at rest. Listen carefully for any sounds coming from the drive.
(2) Then unplug the power cord from the wall outlet, and wait at least three minutes to avoid the risk of electrical shock.
(3) Remove the four M5 hexagon head bolts fixing the front cover from both sides and pull down the front cover by pulling its lower side forward. Then, remove the front cover.
The upper side of the front cover is an insertion type but not a screw fixing type.
DANGER Before removing the cover, top deck, or other components. for maintenance, be sure to turn off the POWER switch of the centrifuge. Then unplug the power cord from the wall outlet, and wait at least three minutes to avoid the risk of electrical shock.
WARNING
1. Never open the door during rotation.
2. Never touch the rotor during rotation.
CAUTION Never conduct operations in a manner other than as described in this operation manual.
WARNING It can take more than three hours for the rotor to come to a complete stop because the rotor chamber is under vacuum. Before opening the door, wait until the rotor comes to a stop.
Page 83
Operation
When power failure occurs
Thermo Scientific WX Ultra Series 2-67
Figure 2-16 Removing the front cover
(4) Open the air vent (by turning the vacuum release screw on the left of the vacuum chamber counterclockwise) to let air into the rotor chamber. When the pressure in the rotor chamber reaches atmospheric pressure, do not forget to tighten the vacuum release screw as it was before. (See Fig. 2-16.)
(5) Pull the door unlock wire on the right of the vacuum chamber and, at the same time, push the door handle. Then the door opens. When opening the door, confirm that the rotor is not rotating. If it is still rotating, immediately close the door.
(6) Remove the rotor. After the removal, close the air vent and put the front cover back to the ultracentrifuge. Insert the top surface of the front cover in the bottom surface of the front side of the table and put the bottom surface of the front cover on the support plate of the front cover by reversing the removing procedure. Attach it on the frame with four M5 hexagon head bolts.
Tab le
Front cover
Cabinet cover
M5 hexagon head bolt for fixing the front cover (4 pcs.)
WARNING Never touch the rotor while it is rotating.
Page 84
Operation
When power failure occurs
2-68 WX Ultra Series Thermo Scientific
Figure 2-17 Ultracentrifuge with the front cover removed
Figure 2-18 Installing the front cover
(7) Turn on the power breaker.
Handle
Vacuum release screw
Vacuum chamber
Door unlock wire
Power breaker
M5 hexagon head bolt for fixing the front cover (4 pcs.)
WARNING When the centrifuge will not be used for a long time, keep the power breaker
off.
Page 85
Operation
Features of the menu screen
Thermo Scientific WX Ultra Series 2-69
Features of the menu screen
Press MENU key on the key board, and the Menu Screen will appear. This feature is designed to allow you to use the Sorvall WX Ultra series with additional options including: (1) Centrifuge scheduler (2) User list (3) Alarm information (4) Rotor catalog (5) User customization routines Key in the number of the item you wish to use and press the ENTER key, and the respective screen will appear.
Figure 2-19
Menu screen
Centrifuge scheduler
Scheduled centrifuge operation can be reserved.
1. Reservation of scheduled centrifuge operation
Step Key operation Screen display and considerations
1 On the Menu Screen, select a centri-
fuge operation schedule.
2 Use cursor keys to move the cursor to
No. you desire. (In this example, No.3 is selected as new reservation.)
• The Menu Screen switches to the Centrifuge Reservation screen.
The ID Code input field is displayed in the message
field of the bottom row.
Page 86
Operation
Features of the menu screen
2-70 WX Ultra Series Thermo Scientific
2. Deleting the centrifuge reservation
You can cancel the operation schedule that is already reserved. This cancellation of reservation is limited to the reserving person proper (with the same ID Code). The reserved operation schedule will be deleted from the centrifuge reservation list after the lapse of the scheduled date/time regardless of operation/non-operation of the centrifuge.
3 Enter an ID Code that is already regis-
tered. (4-digit number)
4 Select “Reserve”.
5 Use cursor keys to move the cursor to
the position at which you want to change the date/time of reservation, and enter a number by using numeric keys.
Note Enter a time later than the current time.
6 At the desired date/time of reserva-
tion, press the ENTER key.
ID code
•The User Name corresponding to the ID Code is displayed. (In this example, ABC lab. is displayed.)
• The current date/time is displayed at the reservation time of the specified reservation No.3.
Step Key operation Screen display and considerations
1 On the Menu Screen, select a centri-
fuge operation schedule.
• The Menu Screen switches to the Centrifuge Reservation screen.
Page 87
Operation
Features of the menu screen
Thermo Scientific WX Ultra Series 2-71
User list
The Centrifuge User List screen is displayed. This screen allows you to register, change, and delete any ID code and user name.
2 Use cursor keys to move the cursor to
reservation No. you want to delete. (In this example, No. 3 is selected.)
3 Enter the ID Code of the reserving per-
son proper. (4-digit number)
4 Select “Delete”.
5 Select “Yes”.
• The ID Code input field is displayed in the message field in the bottom row.
ID code
•When the ID Code is accepted, the reserve/delete selection message is displayed in the message field of the bottom row.
• The confirmation message is displayed in the bottom row.
• The No.3 reservation display disappears from the centrifuge reservation list and the reservation delete completion message is displayed in the bottom row.
Page 88
Operation
Features of the menu screen
2-72 WX Ultra Series Thermo Scientific
• ID code and user name
• Using example of ID code and user name
If ID Code is used when the centrifuge is operated:
1. The user can be known on the screen when the centrifuge is operated.
2. The user record is left in the centrifuge and the usage status for each user can be printed (printer: optional).
3. The User Name can be attached to the user's original programmed operation and the user's own program memory can be checked.
4. Reservation can be attained by the operation schedule table with the User Name attached.
1. Registration of user name
ID Code Password comprised of a 4-digit number. This number is not displayed
when the centrifuge is operated. The mark * is displayed instead of the number.
User Name Character string of 8 digits or less. The user name corresponding to the
ID Code is displayed on the screen. Alphabetical upper-case and lower-case characters and several types of special characters are available.
Note Registration of ID Code and User Name is not always required for operation. If the user does not need to be identified, the centrifuge can be operated without entering ID Code and User Name.
Step Key operation Screen display and considerations
1
Select “User List” on the Menu Screen.
2 Use cursor keys to move the cursor and
press the ENTER key. (For registration, select No.9.)
• The Menu Screen switches to the User List Screen.
• For registration select unregistered number.
• The inquiry message to ask what to do in the message field is displayed.
Page 89
Operation
Features of the menu screen
Thermo Scientific WX Ultra Series 2-73
3 Select “Creating”.
4 Use cursor keys to highlight a charac-
ter and press the ENTER key to select it.
5 After completion of the entry of char-
acters, use cursor keys to move the cursor to SET and press the ENTER key to register the character string.
6 Select “Yes”.
7 Enter the ID Code (4 digits) by using
numeric keys and press the ENTER key.
v
• The User List screen switches to the User Name screen.
•Screen for entry of characters
• The cursor is positioned at SET.
• The confirmation message is displayed.
• The ID Code input field is displayed in the message field of the bottom row.
ID code
• Enter the ID Code and press the ENTER key, and the date of registration will be displayed.
• The mark * is displayed instead of the ID Code on the screen.
Page 90
Operation
Features of the menu screen
2-74 WX Ultra Series Thermo Scientific
2. Change of User Name
Note Changing a User Name is limited to the person (with the same ID Code) who registered it and wants to change it.
Step Key operation Screen display and considerations
1 On the Menu Screen, select “User
List”.
2 Use cursor keys to move the cursor and
press the ENTER key. (Select No.9.)
3
Select “Changing”.
4 Enter the ID Code used for registering
No.9 by using numeric keys.
5 Change the User Name according to
the indication given on the screen.
• The Menu Screen switches to the User List screen.
• For a change, select a registered No.
• The inquiry message to ask what to do in the message field of the bottom row is displayed.
• The ID Code input field is displayed in the User List screen.
• The User List screen switches to the User Name Registration screen.
• Character string input status
Page 91
Operation
Features of the menu screen
Thermo Scientific WX Ultra Series 2-75
3. Deleting a user name setting
6 After entering characters, use cursor
keys to move the cursor to SET and press the ENTER key to register the character string.
7 Select “Yes”.
• The cursor is positioned at SET.
• The confirmation message is displayed.
• The changed User Name is displayed together with its date of registration on the User List screen.
Note Deleting a User Name is limited to the person (with the same ID Code) who registered it.
Step Key operation Screen display and considerations
1
On the Menu Screen, select “User List”.
2 Use cursor keys to move the cursor to
No.9 and press the ENTER key. (The User Name and ID Code of No.9 are deleted.)
3 Select “Deleting”.
• The Menu Screen switches to the User List screen.
• The inquiry message to ask what to do is displayed in the message field.
• The ID code input field is displayed.
Page 92
Operation
Features of the menu screen
2-76 WX Ultra Series Thermo Scientific
Alarm information
The Alarm Information screen displays the contents of the alarm signals and what to do when such signals occur. If an alarm signal occurs while you are using this centrifuge, you can use this screen to take immediate action.
For details, see Section 4 Troubleshooting.
Figure 2-20 Alarm Information screen
4 Enter the ID Code used for registering
No.9 by using numeric keys.
5 Select “Yes”.
>> When you select “No.” the display switches back to the screen display of Step
2.
6 Press the ESC key. >> The User List screen switches to the Menu Screen.
• The confirmation message is displayed.
• The User Name and ID Code of No.9 disappear.
Page 93
Operation
Features of the menu screen
Thermo Scientific WX Ultra Series 2-77
Rotor catalog
You can refer to rotor types and rotor specifications that are available.
User Customizations
You can specify the language in which displays are given, set the current time, adjust screen contrast, and make other settings, all of which enable you to easily run the centrifuge.
1. Instrument identification number
This function is intended to enter a centrifuge identification number (ID number). You can enter a number of up to 8 digits. The centrifuge identification number is printed when the machine log is output to the printer (printer: optional). The same identification number is also used as the centrifuge identification number for managing log data by using the operation log management program (option) on a PC. The identification number is not factory-set. Unless you need it specially, use it as it is.
Step Key operation Screen display and considerations
1 On the Menu Screen, select “Rotor
Catalog”.
2 To refer to the fixed angle rotor
P100AT2, select “Fixed Angle Rotors” on the Rotor List screen.
3 Select the “T-8100” rotor.
4 Press the ESC key several times to go
back to the Run Screen.
>> The Rotor Specification screen switches to the Run Screen.
• The display switches to the Rotor List screen.
• The display switches to the Fixed Angle Rotor List screen.
• Specifications of the T-8100 rotor is displayed.
Page 94
Operation
Features of the menu screen
2-78 WX Ultra Series Thermo Scientific
Figure 2-21 User Customizations screen
Figure 2-22 Screen Utilities screen
2. Screen Utilities settings
1. Display language and select Japanese or English. Key in the number of the desired language and press ENTER key.
2. Changing from the Run Screen >> Normal: Displays the Run Screen. >> Graphics and zoom: Twenty seconds after the actual speed reaches the set speed while in the Run Screen, the display will automatically switch to the screen shown in Fig. 2-23 or 2-24. The display will then return to the Run Screen if you press any key on the keyboard or while a deceleration shift is in progress.
Figure 2-23
Graphics screen
Page 95
Operation
Features of the menu screen
Thermo Scientific WX Ultra Series 2-79
Figure 2-24 Zoom screen
3. Changing date and time Use this option to precisely set the current time setting of the internal clock. Set an exact date and time for RTC operation. Use cursor keys and numeric keys and then press ENTER to make a setting.
4. Screen contrast Use cursor keys to adjust the contrast.
Figure 2-25
Contrast adjustment
3. Setting run parameters
(1) Run Time display This function selects either “Time elapsed” or “Remaining time” as the time display in the run status. When you select “Time elapsed”, “lapse” is displayed in the time display frame on the screen. When you select “Remaining time”, “rem.” is displayed in the same frame.
Figure 2-26
Run Condition setting screen
Note
>> To get back to the Run Screen, press ESC several times. >> If you make no entry for 30 seconds, the cursor will disappear, and you will be unable to make a further input. If this happens, press a cursor key.
and
Low contrast
High contrast
Page 96
Operation
Features of the menu screen
2-80 WX Ultra Series Thermo Scientific
(2) Run Time range This function selects one of the following two items as the running time setting range in the normal operation mode or zonal operation mode: >> Accel + Run Time from pressing the START key to a start of deceleration: T1 (ordinary running time).
Figure 2-27
Run time range setting
(3) Start mode select This function allows you to select one of the following values as the degree of vacuum in the process from the vacuum wait to accelerationto the set speed: >> Medium vacuum: Approx. 133 Pa (ordinary) >> High vacuum: Approx. 13 Pa
Figure 2-28
Vacuum waiting operation mode
Note The vacuum waiting time is not counted in the normal operation. To eliminate the run time difference in the high vacuum start mode, vacuum the rotor chamber up to a high degree during the rotor stop and then start the operation.
>> Run Time from reaching to the set speed to a start of deceleration: T2
Speed
Speed
T1
T2
T1
T2
Tim e
Tim e
[Normal operation]
[Zonal operation]
Speed (rpm)
4,000
Page 97
Operation
Print utilities (option)
Thermo Scientific WX Ultra Series 2-81
(4) Zonal speed This function can change (in units of 100 rpm) the zonal speed in the range of 2,000 to 3,000 rpm. For the zonal speed, see Section Zonal operation)
Figure 2-29
Run condition setting screen
4. Select option unit
This function allows you to set one of the following optional units through the RS232C.
1) Printer
2) Compass Centrifugation Software
3) Unused
Figure 2-30
Optional Unit connection screen
Print utilities (option)
This submenu option is used to print the various information maintained in the ultracentrifuge on the optional printer. The information includes run records, rotor usages, memory contents, etc. When you select “5. PRINT UTILITIES” from the “user customization routines” submenu, this submenu option starts working and displays the PRINT UTILITIES screen.
Note When “Unused” is selected in the optional unit setting, machine log data will be deleted in the order starting with the oldest data and new data will be stored in the memory if the machine log data exceeds the memory area capacity of the centrifuge. For managing such machine log data, be sure to set an optional unit. When such an optional unit has been set, the printer printout or PC data input request message is displayed if the machine log data exceeds the memory area capacity.
Page 98
Operation
Print utilities (option)
2-82 WX Ultra Series Thermo Scientific
Figure 2-31 PRINT UTILITIES Screen
Operating procedure
Note
• Select “1. Printer” of “4. Select Option Unit” in the User Customization Routines screen. Refer to Section User Customizations.
No. Item Operating procedure
1 Machine Log
(1) Select the corresponding item number on the Print Utilities Menu screen.
2 Rotor Management
3 Rotor Log (1) On the Print Utilities Menu screen, select “3. Rotor Log”.
(2) On the Rotor Management screen, move the cursor to select a rotor and press the ENTER key.
(3) Select “1: Yes” as print permission.
4 Set run parameters (1) Select the corresponding item number on the Print Utilities menu screen.
5 Run result
6 Set run parameters and result
The display switches to the Rotor
Management screen.
Page 99
Operation
Print utilities (option)
Thermo Scientific WX Ultra Series 2-83
7 Print program (1) On the Print Utilities menu screen, select “7. Print Program”.
(2) Select a program No. to be printed.
8 User log (1) On the Print Utilities Menu screen, select “8. User Log”.
(2) Move the cursor to select a User Name and perform printing.
9 Machine Management (1) On the Print Utilities menu screen, select “9. Machine Management”.
10 Record after each run (1) On the Print Utilities menu screen, select the corresponding item number.
(2) Move the cursor and select a setting condition.
11 Printout after each run
• The display switches to the Program Map screen.
The display switches to the User List screen.
Page 100
Operation
Print utilities (option)
2-84 WX Ultra Series Thermo Scientific
Contents of print utilities
No. Item Example printout
1 CENTRIFUGE RUN RECORDS
>> Ultracentrifuge run records for a maximum of 40 runs can be printed out at a time. >> Once the ultracentrifuge run records are printed out, they are gone. Therefore, if necessary, save the produced printout for future reference.
2 RUN RECORDS OF ALL ROTORS
>> Run records of all the rotors registered are printed out. >> When a rotor reaches its primary life, an asterisk ([*]) is dis­played at the right of the maximum speed and the maximum rated speed is derated by 10%.
Loading...