Du Pont Company
Sorvall® Instruments
Wilmington, Delaware 19898
U.S.A.
Sorvall Centrifuges
PN 59154-1
Issued August 1989
Sorvall® Centrifuges
This manual is a guide for use of the
SORVALL® RC-28S Supraspeed® Centrifuge
Data herein has been verified and validated and is believed adequate for the intended
use of the centrifuge. If the centrifuge or procedures are used for purposes over and
above the capabilities specified herein, confirmation of their validity and suitability
should be obtained, otherwise Du Pont does not guarantee results and assumes no
obligation or liability. This publication is not a license to operate under, nor a
recommendation to infringe upon, any process patents.
NOTES,
CAUTIONS, and WARNINGS within the text of this manual are used to
emphasize important and critical instructions:
WARNING: A Warning informs the operator of a hazard or an unsafe practice that
could result in personal injury, affect the operator's health, or contaminate
the environment.
CAUTION: A Caution informs the operator of an unsafe practice that could result in
damage of equipment.
NOTE:
1988 by E.I. du Pont de Nemours & Co.
A Note highlights essential information.
ii
Sorvall® Centrifuges Safety
Important Safety Information
Certain potentially dangerous conditions are inherent to the use of all centrifuges. To
ensure safe operation of this centrifuge, anyone using it should be aware of all safe
practices and take all precautions described below and throughout this manual.
WARNING
Use Sorvall® rotors only. Use of another manufacturer's rotor can cause rotor failure
which could result in personal injury and/or centrifuge damage.
When using radioactive, toxic, or pathogenic materials, be aware of all characteristics of
the materials and the hazards associated with them in the event leakage or rotor failure
occurs during centrifugation, in which case neither the centrifuge nor the rotor can
protect you from particles dispersed in the air. To protect
additional precautions be taken to prevent exposure to these materials, e.g., use of
controlled ventilation or isolation areas.
Always be aware of the possibility of contamination when using radioactive, toxic, or
pathogenic materials. Take all necessary precautions and use appropriate
decontamination procedures if exposure occurs.
yourself,
we recommend
Never use any material capable of producing flammable or explosive vapors.
Never exceed the maximum rated speed of the installed rotor; to do so can cause rotor
failure.
Always reduce (derate) rotor speed whenever:
• the rotor speed/temperature combination exceeds the solubility of the gradient
material and would cause it to precipitate;
• rotor compartment load exceeds the maximum allowable compartment load
specified (check this whenever average fluid density is greater than 1.2 g/ml). See
Chapter 4, Operation.
Failure to reduce rotor speed under these conditions can cause rotor failure.
CAUTION
Do not run or precool a rotor at the critical speed, as this will have a detrimental effect
on centrifuge component life. See Table 4-1, Rotor Information.
Do not operate centrifuge with a rotor out of balance. To do so can cause damage to the
centrifuge drive assembly.
Do not operate centrifuge without rotor properly installed: rotor cover must be on and
locked in place, and the rotor locked to the centrifuge drive. See rotor instruction
manual.
Locate the centrifuge on a level floor to avoid rotor imbalance during operation.
The centrifuge can be damaged if connected to the wrong voltage. Check the voltage
before plugging the centrifuge into a power source. Du Pont is not responsible for
incorrect installation.
Always maintain the centrifuge in the recommended manner. See Chapter 7,
Maintenance.
iii/iv
Sorvall® Centrifuges Table of Contents
Table of Contents
Page
Safety Information iii
Chapter 1. DESCRIPTION
General Description 1-1
Centrifuge Specifications 1-2
Centrifuge Features 1-3
System Descriptions 1-5
Understanding the Centrifuge Controls 4-1
Turning the Centrifuge On 4-1
Normal Run 4-1
Slow Start/Slow Stop 4-3
RCF Run 4-5
Key#Code 4-6
Zonal Run 4-7
Hold Run 4-8
fctfdt Run 4-9
Programmed Run 4-10
Control Panel Lock 4-11
Precool/Preheat 4-12
Reducing Speed for Rotor Compartment Loads in
Excess of Design Mass 4-12
Table of Contents Sorvall® Centrifuges
Table of Contents (continued)
Page
Chapter 5. CALCULATOR
Calculator Display Screens 5-1
Calculate Speed/RCF for Rotor Time/fo^dt 5-1
Calculate Speed/RCF at a Radius 5-2
Calculate K Factor 5-2
Calculate Rotor Conversion/Time 5-2
Chapter 6. TROUBLESHOOTING & DIAGNOSTICS
Display Messages 6-1
Diagnostics:
General 6-5
Screen 1, Faults 6-6
Screen 2 6-7
Screen 3 6-9
How to Assign or Change a Key# 6-9
How to Change Slow Start/Slow Stop Transition Speed 6-12
Chapter 7. MAINTENANCE
Maintenance Schedule 7-1
Cleaning:
Rotor Chamber 7-2
Tapered Drive Spindle 7-2
Cabinet 7-2
Control Panel 7-2
Door Seal 7-2
Refrigeration System Condenser 7-4
Air Louvers 7-4
Lubrication 7-4
Changing the Vacuum Pump Oil & Mist Eliminator Filter 7-4
Emergency Sample Recovery 7-8
Test of RUN Display Lights & Control Panel Indicators 7-10
Parts Ordering Information 7-11
Service Decontamination Policy 7-11
APPENDIX
Vacuum Pump Oil Specifications
Warranty
Glossary
Index
Program Log Cards
Decontamination Certificates
VI
Sorvair Centrifuges
Table of Contents
List of Illustrations
Figure
1-1. SORVALL® RC-28S Supraspeed® Centrifuge viii
2-1.
Centrifuge Dimensions 2-1
2-2.
Leveling Adjustment 2-2
3-1.
Control Panel 3-2
7-1.
Door Seal Removal 7-3
7-2. Parts Identification, Vacuum Pump Oil Change 7-6
7-3.
This manual provides you with the information you need to install, operate and
maintain your SORVALL® RC-28S Supraspeed® Centrifuge. If you encounter any
problem concerning either operation or maintenance that is not covered in the manual,
please telephone Du Pont for assistance. In the United States, call toll free
(800)551-2121.
SORVALL® Products.
General Description
The Sorvall® RC-28S Supraspeed® Centrifuge defines a new class of centrifuge that
brings an unparalleled level of performance to centrifugation. Created from the latest
design technology, it incorporates usability, dependability, and safety with an extremely
versatile speed range to set a new standard in value. When used with a Sorvall®
Supraspeed® rotor, the RC-28S can reach a top speed of 28 000 rpm with a maximum
g force of 100 095. At 100 095 x g you can perform applications in the RC-28S that
previously required an ultracentrifuge. In addition to the Supraspeed® rotors designed
for use with the RC-28S, the RC-28S accepts all Sorvall® superspeed rotors, increasing
its usefulness by making it capable of a wide range of applications.
Outside the United States, contact the distributor or agent for
The centrifuge is microcomputer controlled and designed to be user friendly. It can be
operated using basic parameters only or you can decide to use its special features.
Because of this versatility and its user-friendliness, it can easily be operated by
inexperienced persons learning centrifugation methods or by experienced persons
experimenting with new methods.
The brushless dc drive which offers the optimum in centrifuge drive performance is one
distinct feature of the RC-28S. Other features of the centrifuge are: automatic rotor
recognition; programmability; RCF and
for sample reorientation; overtemperature control; a control panel lock to prevent run
parameters from being changed; and a built-in calculator that can be used to perform
centrifugation calculations, including conversions between RCF and speed, time and
fco2dt or the calculation of K factors and run times when conversion of published run
parameters is desired.
The centrifuge also has extensive diagnostic routines that allow the user to identify
system malfunctions and perform maintenance and repair procedures.
For more on the features of the RC-28S, plus more on the centrifuge drive and other
systems, read the table on the next two pages and the descriptive paragraphs that follow
the table.
fco2dt
control; slow start and slow stop control
1-1
Description
Centrifuge Specifications
Son/all® Centrifuges
Rotor Management
SPIN-RIGHT® Rotor Management System
automatically recognizes installed Sorvall®
rotor when chamber door is closed. Pre-
vents operation of rotors above their maxi-
mum speed. Primary function of the system
is to provide an added margin of safety
during operation; however it is effective only
when it is operated as intended. Any misuse of the system, as with any other centrifuge system or feature, could result in
unsafe operation.
Rotor Speed
Set Speed:
from 100 rpm to 28 000 rpm in
increments of 10
Speed Control:
±10 rpm from 0-2000 rpm;
±0.5 % above 2000 rpm.
Speed Displayed:
digital readout during a run of actual
speed
Rotor Temperature
Set Temperature:
from -20°C to +40°C
Overtemperature:
automatic or settable from 0°C to 45°C
Temperature Control:
±10Cfrom2°Cto40°C
Temperature Displayed:
digital readout during a run of actual
temperature in 1 ° increments
1
Time
Set Time:
one minute to 99 hours and 99
minutes; can select HOLD for
continuous operation
Time (continued)
Time Displayed:
digital readout indicates remaining
time for a timed run or elapsed time
for continuous run
Vacuum
Rotary vane pump system.
Chamber pressure:
< 15 millibars (1.5 kPa)
Ambient Temperature
Range
from 15°Cto38°C
2
Dimensions:
Width:
Height: 93 cm (36.5 in)
Depth:
Clearances:
15 cm (6 in) required on left side of
centrifuge
70 cm (27.5 in)
to top deck
123 cm (48.5 in)
to top of control panel
103 cm (40.5 in)
Net Mass
385 kg (850 lbs)
Electrical Requirements
Can be supplied for use on the following
single phase power sources
200 V, 60 Hz, 30 A 200 V, 50 Hz, 30 A
208 V, 60 Hz, 30 A3 220 V, 50 Hz, 30 A
230 V, 60 Hz, 30 A3 240 V, 50 Hz, 30 A
Receptacle:
power cord fits Hubbell receptacle
No.
9330 or equivalent
1
Temperature range with power ON but no run in process (standby) is from -10°C to +30°C.
2
The centrifuge will operate at ambient temperatures up to 38°C, but refrigeration system performance will be less than optimal above 25°C.
3
CSA Certification applied for.
1-2
Sorvall® Centrifuges
Description
Centrifuge Features
High Performance Drive
A brushless dc direct drive system de-
signed for longer motor life and greater
reliability; provides high torque; fast acceleration; and smooth, precise speed
control. No brushes means less maintenance. Controlled by dual microprocessors with accurate speed control during
acceleration and a Positive Action Brake
system that provides controlled braking
from set speed all the way to 0 rpm.
Includes Sorvall® Gyro-Action
Centering System that permits "eye balancing" of tubes. Imbalance detector
protects against damage from gross imbalance conditions. Accepts all Sorvall®
Superspeed and Supraspeed® rotors.
Self-
Automatic Rotor Recognition
SPIN-RIGHT® Rotor Management System
automatically recognizes installed
Sorvall® rotor when chamber door is
closed. Provides optimal instrument
control. Prohibits operation of a rotor
above its maximum speed. Primary
function of the system is to provide an
added margin of safety during operation;
however, it is effective only when it is
operated as intended. Any misuse of the
system, as with any other centrifuge
system or feature, can result in unsafe
operation.
Vacuum
Allows Sorvall® Supraspeed® rotors to be
operated at speeds up to 28 000 rpm.
Automatically activated when required.
RCF Run
RCF can be chosen in place of a speed to
control run. Allows RCF to be reproduced from run to run and rotor to rotor.
Accepts values up to 100 095 x g, de-
pending on the rotor used.
Slow Start / Slow Stop
Automatic control of acceleration and
deceleration to and from transition
speed; provides optimal reorientation
conditions which prevent disturbance of
gradient or pellet. Optimal times to and
from transition speed are automatically
determined for identified rotor or can be
manually selected; transition speed can
even be changed if desired. Flexibility of
control means you can duplicate a slow
start/slow stop run done in any other
centrifuge.
Zonal Run
Allows operation with the chamber door
open to dynamically load a Sorvall®
TZ-28 Zonal Rotor. Loading speed selection from 1300 rpm to 2500 rpm. A key
number is required for zonal operation.
Precise Temperature Control
A high-capacity, highly efficient refrigeration system that combines heating and
cooling. The combined heating and
cooling provides temperature control
from 2°C to 40°C and eliminates the
need to defrost the chamber. Temperature sensor is mounted away from the
chamber wall to further improve accuracy of rotor temperature measurement.
Automatic sample temperature protec-
tion; maximum temperature will be set
automatically or can be entered manually. Convenient precooling of rotors.
fco2dt Run
A built-in integrator that provides exact
run-to-run reproducibility by controlling
the accumulating
ing value is displayed during the run.
fo)2dt.
The accumulat-
Control Panel Lock
Locks the control panel to prevent run
conditions from being changed without
the operator knowing. A key number is
required to lock and unlock the control
panel.
1-3
Description
Centrifuge Features
Sorvall® Centrifuges
Programmability
Records run settings in memory. Once
recorded, programs can be recalled to
allow duplication of run. Up to 20 programs can be recorded. A key number
is required to save a program to pre-
vent unauthorized changes; however,
programs can be recalled without a key
number.
Calculator
A built-in calculator that performs
basic centrifugation calculations: RCF,
Speed, Time,
leting time. Can be used to convert
run conditions from rotor to rotor.
S(o2dU
K Factor, and pel-
Security Key Number
A four digit security number that is
required to access protected data and
some diagnostics. Also required to
activate Zonal, Program Save, and the
Lock. Centrifuge is supplied with one
factory-set number, but you can program multiple numbers for greater security and multiple users.
Advanced Diagnostics
Built-in diagnostic routines include
several levels of electronic troubleshooting aids that allow you to diagnose problems if they occur. Also used
to perform limited maintenance procedures and centrifuge updates.
Rotor Chamber
The stainless steel rotor chamber is
epoxy coated and free of wall sensors
making it durable and easy to clean.
Chamber seal is located in the door
rather than in the top of the chamber;
minimizes accumulation of dirt on the
seal which improves the seal. The door
seal can easily be removed for cleaning.
The sliding chamber door retracts
which makes the chamber easy to
access and solves door clearance problems.
Size
The RC-28S is narrower than most
superspeed and ultraspeed centrifuges.
Requires less floor space.
Operator's Guide
Readily accessible operating instructions and aids that pull out from under
the control panel. Includes convenient
Program Log records and rotor specifications.
Serviceability & Technical
Support
Extensive diagnostics and easy to
access components make the RC-28S
easy to service. Complete product
literature available plus a toll free
number to call for technical assistance.
1-4
Sorvall®Centrifuges Description
System Descriptions
Drive System
The RC-28S drive system uses state-of-the-art brushless dc motor technology that
offers the optimum in centrifuge drive performance.
A brushless dc motor uses an electronic commutator in place of a mechanical
commutator, which eliminates brushes and the maintenance problems associated with
them, plus it eliminates the speed and current limitations of a mechanical commutator.
The high torque of the dc motor results in faster acceleration and braking, while the
high efficiency means less power consumption and less heat. Less heat in the motor
means that the drive runs cooler which eliminates the need for any auxiliary means of
motor cooling, such as Freon® or water.
The motor used is inherently non-cogging which ensures the smoothest acceleration
and deceleration possible. Cogging is typical of induction and brush type motors and
can cause sample disturbance during acceleration and deceleration.
The RC-28S drive system is controlled by dual microprocessors allowing extremely
accurate speed control during acceleration and deceleration and eliminating overshoot
of set speed. The accuracy of the speed control results in smoother slow start and slow
stop,
plus positive action braking from set speed to 0 rpm. The microprocessors operate
independent of each other to constantly monitor various centrifuge functions and, as an
added safety factor, they maintain constant internal communication with each other to
test and evaluate the performance of the other. The many tests that are performed
routinely throughout operation determine if any preset limitations are exceeded; any
failure detected will cause immediate shutdown by the processor detecting the problem.
In addition, the centrifuge will shut down if communication fails between the processors
or in the unlikely event both processors fail.
SPIN-RIGHT® Rotor Management System
The Sorvall® RC-28S SPIN-RIGHT® Rotor Management System automatically insures
optimal control of the selected speed and temperature parameters and prevents the
inadvertent setting of a speed beyond the capability of a given Sorvall® rotor. The
Sorvall® rotor installed in the chamber is recognized through an ultrasonic scanning
device that is activated when the chamber door is closed. The rotor identity is then
displayed in the SET display of the control panel.
The primary function of the rotor management system is to provide an added margin of
safety during operation. This system is effective only when it is operated as intended.
Any misuse of the system, as with any other centrifuge system or feature,can result in
unsafe operation
Refrigeration System
The refrigeration system consists of: a stainless steel evaporator that acts as the rotor
chamber, a 2-hp hermetically sealed compressor, an air-cooled condenser, and the
suction-capillary tube assembly. A hot gas bypass circuit allows for quiet, non-cycling
1-5
Description Sorvall®Centrifuges
compressor operation as well as improved temperature control and frost-free operation.
The refrigeration system is factory charged with Freon® 502.
The desired run temperature is entered on the control panel and is maintained by the
system electronics which cycles the hot gas circuit as required. If desired, a maximum
temperature can be entered on the control panel; if a value is not entered manually, the
maximum temperature will default to +5°C above set temperature. Any failure causing
the temperature to exceed the maximum setting will result in the termination of the run
and a corresponding message will be displayed on the control panel to alert the
operator.
Vacuum System
The vacuum system includes a direct drive vacuum pump, an air-inlet solenoid (bleed)
valve, mist eliminator, the evaporator chamber, and interconnecting tubes and fittings.
Reduced rotor chamber pressure is sensed by a Pirani convection gauge located at the
pump inlet manifold and monitored by the machine control microprocessor. When
activated, this system provides a vacuum level between 1 and 15 millibars.
The vacuum system is automatically activated only when the set speed is higher than
the top air speed of one of the Sorvall® Supraspeed® rotors designed specifically for use
with the RC-28S.
In the event of a vacuum loss or insufficient vacuum, a message will appear in the
control panel SET display to notify the operator. In an emergency, such as a power
failure, when sample recovery is desirable, the chamber door can be opened manually
(see Chapter 7 for procedure).
1-6
Sorvall® Centrifuges
Installation
Chapter 2
INSTALLATION
This chapter contains instructions to prepare your Sorvall® RC-28S for operation.
Inspection
When you receive your centrifuge, carefully inspect it for any signs of shipping
damage. If you find damage, please report it immediately to the transportation
company and file a damage claim, then notify Du Pont.
Check the parts received with the centrifuge against the shipping list; if any parts
are missing, contact Du Pont (see back cover).
Location
Locate the centrifuge on a level floor. To permit free air circulation, the location must
allow space for the size of the centrifuge (figure 2-1), plus an additional 15 cm (6 inch)
space on the left side of the centrifuge. Free air circulation is very important for the
centrifuge to function properly.
LEFTSIDE OF
CENTRIFUGE
POWER
CORD
200cm -
(78.5 in)
123 cm
(48.5 in)
Figure 2-1. Centrifuge Dimensions
2-1
Installation
Son/all® Centrifuges
Ambient Temperature
An ambient temperature range of 15°C to 38°C must be maintained. If the inlet air
temperature is above 25°C, the centrifuge will not maintain low temperatures at
high speeds.
Leveling
WARNING
The front stabilizers must be used as explained below to lift the casters off the
floor; this will help minimize centrifuge movement in the event of a rotor
failure.
When the centrifuge is situated at its operating location, lower and secure the front
locking stabilizers. Lower the two locking stabilizers by hand until they are finger tight
against the floor. Then use the 9/16 inch (14 mm) wrench supplied with the centrifuge
to rotate each stabilizer an additional 2 1/2 turns. This will raise the front casters
about 3 mm (1/8 inch) off the floor and level the centrifuge (based on level floor).
LOCKING STABILIZERS
(9/16 in [14 mm] HEX)
APPROX. 3mm (1/8 in)
OFF THE FLOOR
Figure 2-2. Leveling Adjustment
Electrical
Your centrifuge was ordered for a specific single phase power source (preferably
regulated). The nameplate on the back panel of the centrifuge tells which one of
the following classifications you ordered:
200 V, 50 Hz, 30 A
220 V, 50 Hz, 30 A
240 V, 50 Hz, 30 A
If the line voltage varies ±10% of its nominal value, you may notice variations in the
performance of the centrifuge. If the line voltage exceeds this tolerance, it may
damage the centrifuge.
* CSA Certification applied for.
2-2
Sorvall® Centrifuges installation
CAUTION
The centrifuge can be damaged if connected to the wrong voltage. Check
the voltage before plugging the centrifuge into any power source. DuPont
is not responsible for improper installation.
The centrifuge is equipped with a 3-wire power cord with a 3-prong cap, including
ground connection, to fit Hubbell receptacle No. 9330 or equivalent.
To connect the centrifuge to a voltage other than what is specified on the nameplate, the
centrifuge will have to be rewired and the power cord may have to be replaced. In this
case,
contact DuPont to have a Service Representative do the rewiring.
The ON/OFF main power switch is a 30 A circuit breaker, therefore a separate line
disconnect switch is not needed unless required by local codes.
2-3/2-4
Sorvall® Centrifuges Controls
Chapter 3
CONTROLS
This chapter provides:
• a detailed description of the RC-28S controls;
• a brief description of the built-in calculator; and
• a brief description of the diagnostic routines.
For step-by-step operating instructions see Chapter 4, and for further information
on the calculator and the diagnostics, see Chapters 5 and 6.
NOTE
A convenient Operator's Guide containing simplified instructions is located
under the centrifuge control console.
Control Panel
The RC-28S is easy to use. All commands are entered from the convenient keypad
on the control panel and the set and run parameters are visible in the displays. All
selected parameters can be locked to prevent changes to run settings. Figure 3-1
on the next page shows the RC-28S control panel and gives a brief description of
the keys and displays. A complete description follows the illustration.
Set Display
The SET display is the lower display that changes by pressing the PROGRAM,
CALCULATE or DIAGNOSTICS key to show various screens of parameters or
operations. A cursor in the SET display indicates the field where the next input
character will appear. Lights below the display indicate the position of the cursor.
Run Display
The RUN display includes the three upper displays that show actual speed,
temperature, and time or Jco2dt during a run.
3-1
CO
SCREEN INDICATOR
Indicates which screen
has been selected and
appears in SET display.
CURSOR INDICATOR
Indicates cursor location in
SET display.
2
1
SET DISPLAY
Displays parameters to
be selected, calculator
modes, and diagnostic
conditions. Change display by pressing one of
the three control keys be-
low the display.
RUN DISPLAY
Speed
Displays current rotor
speed (rpm) during a run.
Temperature
Displays current rotor
temperature (°C) during a
run.
Time
TIMED Run: displays re-
maining run time.
HOLD Run: displays ac-
cumulating run time.
displays accu-
mulating integral value.
Ja/WflNDICATOR
Indicator light is ON when
lco2dt run is selected.
START
Starts the centrifuge run.
O
O
a
PROGRAM
Primary control key. Selects one screen of basic
and three screens of op-
B
tional run parameters.
Display changes each
I
time key is pressed and a
corresponding indicator
light above the key comes
on to indicate which
screen is in the display.
CALCULATE
Selects calculator mode to
perform basic centrifugation calculations. Four
screens of calculations
available. Display changes each time key is
pressed and a corresponding indicator light
above the key comes on to
indicate which screen is in
display.
DIAGNOSTICS
Selects diagnostic mode
to help isolate malfunctions,
provide status of
various systems, and to
perform maintenance procedures.
ENTER
Moves the cursor right or
left through SET display.
Also causes an input
value to be accepted.
ON/OFF +/-
The +/- changes set temperature values from plus
to minus or from minus to
plus.
The On/Off selects
brake mode and other
options.
STOP
Stops the centrifuge run.
NUMERIC KEYPAD
Used to input numbers to
select parameters.
HOLD
Selects a continuous run.
CLEAR
Cancels an input or a previously entered value.
CO
o
5
(D
CO
Sorvall® Centrifuges
Controls
Description of Display & Key Functions
v
j v j V j
Cancels
the entry
again,
Starts
Stops
the
the
the
The numeric keypad
select parameter values.
The ENTER keys cause
accepted, plus they cause
the right
On/Off
word
On or Off
of
the SET
the temperature value from positive
negative
an
input
is set to
previously entered value
centrifuge
centrifuge
or a
previously entered value. When
zero
(0) and a new
run.
run.
is
or to the
is
used
to
is
used
to
left through
to
select options. When pressed,
will appear
display.
positive.
value
restored.
input numbers from
an
input value
the
cursor
the SET
in the
The +/- is
can be
input. When pressed
appropriate field
used only
to
the key is
0 to 9 to
to
move
display.
to
negative
pressed,
to be
to
the
change
or
Selects a continuous
the
SET
display
is pressed,
show accumulating
Selects four screens
screens
come
the
The four PROGRAM
parameters
are
labeled
on to
SET
indicate which screen
display will change
and
1
2
the
1, 2, 3, and 4 and
operations. A description
Rotor Speed Hr:Min Tenp/Max Brake
RCF Slow-Start Slow-Stop Zonal/Speed
3
run. The
and all
timer will start
run
zeros will
time.
Program
of
standard
you
key
to the
screens
next screen.
word Hold will appear
be in the RUN
and the
Key
and
optional centrifugation conditions.
have corresponding lights above
have accessed. Each time
are
illustrated below showing
of the
Time field
Functions & Display Screens
fields
in
each screen follows.
in the
display.
of the RUN
As
the key is
PROGRAM DISPLAY SCREENS
w2dt Tine-out TotalLock
HnMin field
soon
as
START
display will
PROGRAM
The
the key
the
available
that
pressed,
of
4
PROGRflM> Recal1# Save#
3-3
Controls Sorvall® Centrifuges
PROGRAM
Program Key Functions & Display Screens
Rotor Speed HrsMin Temp^Max Brake
Screen
1
F-28^13 23000
0:90
+4+9
On
When the "1" indicator light above the PROGRAM key
will include
temperature,
changed during a run.
Following is a description of each field in this screen.
the
and
basic
brake.
run
All
parameters: rotor identification, speed, time,
parameters, except rotor identification,
is
on, the SET display screen
can be
Rotor Speed HrsMin TeMp/Max Brake
The SPIN-RIGHT® Rotor Management System automatically recognizes the installed
Sorvall® rotor
and RCF calculation. When the chamber door
in the display. Whenever the chamber door
the display unless the Lock
display.
The rotor identification cannot be changed after START has been pressed.
Iliiiiiil Speed
Centrifugation speed
increments
of the installed rotor
erroneous entry.
to
of
provide temperature compensation, top speed limit, speed control,
is
closed, the rotor name will appear
is
open, the word NONE will appear
is
on,
in
is
10 rpm.
is
In
which case the rotor name will stay
selected
If a
entered, all digits
this case, clear the entry and make a valid entry.
by
entering a value from 100
speed value greater than the maximum allowable speed
in
the Speed field will flash indicating an
to
28 000 rpm
in
in
the
in
NOTE
If an RCF value
the chosen RCF value will automatically be calculated and displayed in this field.
The speed setting can
Speed field and entering a new value.
is
be
selected
to
control the run, the speed required
changed after a run has been started by returning
to
achieve
Rotor Speed HrsMin TerwMax Brake
Centrifugation time can
Values are displayed exactly
displayed 0:90 which represents one hour and 30 minutes run time. The same run
time
of
one hour and
displayed as 1:30.
3-4
be set
as
30
minutes can
from one minute
entered;
for
be
example,
selected
to 99
if
by
hours and
you input 90,
entering 130; this will
99
minutes.
it
to
will
the
be
be
Sorvall® Centrifuges Controls
PROGRAM
As soon
timer will begin when START
from
HOLD will appear
(0)
timer will begin after
Time
entering a new value.
as an
set
value
and
will show accumulating
can be
entry
to
changed after
is
zero
in the SET
made,
to
show remaining
the
rotor
the run has
Program
the
is
pressed
display
run
Key
entered value will appear
and the RUN
time. When operating
has
been loaded
Functions & Display Screens
and the RUN
run
time.
started
by
display value will count down
If
display value will begin
and the
returning
HOLD
NOTE
If
the
integrator
accumulated. During
integral value
HnMin field.
started.
During a timed
value will
be
displayed
is On, the run
a run, the RUN
and the SET
The
run
integrator cannot
with
no
in the
will
end
when
display will show accumulating
be
turned
preselected integral value,
G&dt
Total field
of
the
selected integral value
display will show
Off (or On)
Screen
3.
the
Rotor 5r*eed HrsMin Tenp/Max Ilillfil
. i
in the
in the
to the
after
RUN display.
is
selected,
Zonal mode,
chamber door closed.
Hr:Min field
the
accumulating
run
time
a run has
accumulating integral
the
at
has
in the
been
The
word
zero
the
and
There
centrifugation temperature;
protection limit.
Centrifugation temperature
Acceptable
value
Temp value
to 0°C).
Max temperature
desired value, however,
operation,
value until
are two
is not
If
the
temperature shown
value (shown
indicated temperature
Temp value
the Max
rotor temperature settings.
Max
the
temperature values
entered,
is -6°C or
or at
rotor
the
lower
in the SET
0°C.
can be
temperature setting should
has
and
changed
it
cooled
the
other.
can be set
value will default
in
which case
Max,
from -20°C
are
from
NOTE
in RUN
display),
will decrease
must
to the
display
Max
as
by
moving
be at
least
selected temperature.
The
to 5°C
the Max
is 5°C or
temperature will remain
2°
first
one,
selects
0°C to
rotor cools until
the
not be
the
sample overtemperature
to
+40°C
+45°C.
above
temperature will always default
cursor
above
the
more above entered Temp
to Max and
set
Temp value.
changed from
Brak
Temp, selects desired
in
increments
If a Max
Temp value (unless
it is 5°
e
of 1°C.
temperature
3°C
above
above
entering
the
set
For
default
the
best
Press On/Off
Brake field
to
turn
the
to
indicate choice.
brake
on or off. The
word
On or Off
will appear
in the
3-5
Controls Sorvall® Centrifuges
PROGRAM
Program Key Functions & Display Screens
When the brake is Off, the rotor will coast from set speed to zero rpm at the end of
the run. If a slow stop has been selected, the rotor will coast to the transition
speed, then continue to decelerate at the controlled rate. When the Brake is On,
the rotor will brake from set speed to zero rpm at the end of a run; if a slow stop
has also been selected, the rotor will brake to transition speed, then continue to
decelerate at the controlled rate.
C
o *> "CF
screen
To access, press PROGRAM until the "2" indicator light above the key is lit. The
SET display will include the following fields: RCF, Slow Start, Slow Stop and Zonal.
RCF
z
100095 On fluto On fluio Off 1500
Slow-Start Slow-Stop Zonal/Speed
The RCF field is used to: (1) display calculated RCF value for the identified rotor
and the set run speed prior to a run and to display actual RCF value during a run; or
(2) to set the centrifuge to operate at a given g force. When an RCF value is
selected to control the run, the speed required to achieve it will automatically be
calculated and displayed in the Speed field of Screen 1.
During a run, the RCF (at rmax of the identified rotor) at actual speed is always
displayed in this field whether the run is programmed for speed or g force. At the
end of a run, the value zero will remain in this field until the chamber door is
opened. When the door is opened, the display will change back to show the RCF
value for the set speed.
RCF Slow-Start
To select a slow start, move the cursor to Slow-Start and press
will show "On Auto". Auto is the optimal rate automatically calculated to have the
installed rotor accelerate to transition speed in approximately 2 minutes. When
the run is started, a message will appear in the SET display to indicate that a slow
start is in progress. When the rotor reaches transition speed, the message in the
display will go out and the rotor will accelerate to set speed at the maximum rate.
The length of time for acceleration from 0 rpm to transition speed can be changed
by positioning the cursor under "Auto" and entering the number of minutes (from 1
to 30) that you want for acceleration. Once Slow-Start has been turned On, it will
remain On until it is turned Off. When Slow-Start is Off, the rotor will accelerate
from 0 rpm to set speed at the maximum rate.
On/Off;
the display
3-6
NOTE
The preset transition speed (approximately 300 rpm for S-20 rotor and
approximately 800 rpm for all other rotors) can be changed. See Chapter 6.
Loading...
+ 49 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.