Sorvall RC-28S User manual

OPERATING INSTRUCTIONS
SORVALU RC-28S
SUPRASPEED
CENTRIFUGE
Du Pont Company Sorvall® Instruments Wilmington, Delaware 19898 U.S.A.
Sorvall Centrifuges
PN 59154-1 Issued August 1989
Sorvall® Centrifuges
This manual is a guide for use of the
SORVALL® RC-28S Supraspeed® Centrifuge
Data herein has been verified and validated and is believed adequate for the intended use of the centrifuge. If the centrifuge or procedures are used for purposes over and above the capabilities specified herein, confirmation of their validity and suitability should be obtained, otherwise Du Pont does not guarantee results and assumes no obligation or liability. This publication is not a license to operate under, nor a recommendation to infringe upon, any process patents.
NOTES,
CAUTIONS, and WARNINGS within the text of this manual are used to
emphasize important and critical instructions:
WARNING: A Warning informs the operator of a hazard or an unsafe practice that
could result in personal injury, affect the operator's health, or contaminate the environment.
CAUTION: A Caution informs the operator of an unsafe practice that could result in
damage of equipment.
NOTE:
1988 by E.I. du Pont de Nemours & Co.
A Note highlights essential information.
ii
Sorvall® Centrifuges Safety
Important Safety Information
Certain potentially dangerous conditions are inherent to the use of all centrifuges. To ensure safe operation of this centrifuge, anyone using it should be aware of all safe practices and take all precautions described below and throughout this manual.
WARNING
Use Sorvall® rotors only. Use of another manufacturer's rotor can cause rotor failure
which could result in personal injury and/or centrifuge damage. When using radioactive, toxic, or pathogenic materials, be aware of all characteristics of
the materials and the hazards associated with them in the event leakage or rotor failure occurs during centrifugation, in which case neither the centrifuge nor the rotor can protect you from particles dispersed in the air. To protect additional precautions be taken to prevent exposure to these materials, e.g., use of controlled ventilation or isolation areas.
Always be aware of the possibility of contamination when using radioactive, toxic, or pathogenic materials. Take all necessary precautions and use appropriate decontamination procedures if exposure occurs.
yourself,
we recommend
Never use any material capable of producing flammable or explosive vapors. Never exceed the maximum rated speed of the installed rotor; to do so can cause rotor
failure.
Always reduce (derate) rotor speed whenever:
• the rotor speed/temperature combination exceeds the solubility of the gradient material and would cause it to precipitate;
• rotor compartment load exceeds the maximum allowable compartment load specified (check this whenever average fluid density is greater than 1.2 g/ml). See Chapter 4, Operation.
Failure to reduce rotor speed under these conditions can cause rotor failure.
CAUTION
Do not run or precool a rotor at the critical speed, as this will have a detrimental effect on centrifuge component life. See Table 4-1, Rotor Information.
Do not operate centrifuge with a rotor out of balance. To do so can cause damage to the centrifuge drive assembly.
Do not operate centrifuge without rotor properly installed: rotor cover must be on and locked in place, and the rotor locked to the centrifuge drive. See rotor instruction manual.
Locate the centrifuge on a level floor to avoid rotor imbalance during operation.
The centrifuge can be damaged if connected to the wrong voltage. Check the voltage before plugging the centrifuge into a power source. Du Pont is not responsible for incorrect installation.
Always maintain the centrifuge in the recommended manner. See Chapter 7, Maintenance.
iii/iv
Sorvall® Centrifuges Table of Contents
Table of Contents
Page
Safety Information iii
Chapter 1. DESCRIPTION
General Description 1-1 Centrifuge Specifications 1-2 Centrifuge Features 1-3 System Descriptions 1-5
Chapter 2. INSTALLATION
Inspection 2-1 Location 2-1
Ambient Temperature 2-2
Leveling 2-2 Electrical 2-2
Chapter 3. CONTROLS
Control Panel 3-1 PROGRAM Key Functions & Display Screens 3-3 CALCULATE Key Functions & Display Screens 3-9 DIAGNOSTICS Key Functions & Display Screens 3-9
Chapter 4. OPERATION
Understanding the Centrifuge Controls 4-1 Turning the Centrifuge On 4-1 Normal Run 4-1 Slow Start/Slow Stop 4-3 RCF Run 4-5 Key#Code 4-6
Zonal Run 4-7 Hold Run 4-8 fctfdt Run 4-9 Programmed Run 4-10 Control Panel Lock 4-11 Precool/Preheat 4-12 Reducing Speed for Rotor Compartment Loads in
Excess of Design Mass 4-12
Table of Contents Sorvall® Centrifuges
Table of Contents (continued)
Page
Chapter 5. CALCULATOR
Calculator Display Screens 5-1 Calculate Speed/RCF for Rotor Time/fo^dt 5-1 Calculate Speed/RCF at a Radius 5-2 Calculate K Factor 5-2 Calculate Rotor Conversion/Time 5-2
Chapter 6. TROUBLESHOOTING & DIAGNOSTICS
Display Messages 6-1 Diagnostics:
General 6-5 Screen 1, Faults 6-6 Screen 2 6-7
Screen 3 6-9 How to Assign or Change a Key# 6-9 How to Change Slow Start/Slow Stop Transition Speed 6-12
Chapter 7. MAINTENANCE
Maintenance Schedule 7-1 Cleaning:
Rotor Chamber 7-2
Tapered Drive Spindle 7-2
Cabinet 7-2
Control Panel 7-2 Door Seal 7-2 Refrigeration System Condenser 7-4
Air Louvers 7-4 Lubrication 7-4 Changing the Vacuum Pump Oil & Mist Eliminator Filter 7-4 Emergency Sample Recovery 7-8
Test of RUN Display Lights & Control Panel Indicators 7-10 Parts Ordering Information 7-11 Service Decontamination Policy 7-11
APPENDIX
Vacuum Pump Oil Specifications Warranty
Glossary Index Program Log Cards Decontamination Certificates
VI
Sorvair Centrifuges
Table of Contents
List of Illustrations
Figure
1-1. SORVALL® RC-28S Supraspeed® Centrifuge viii
2-1.
Centrifuge Dimensions 2-1
2-2.
Leveling Adjustment 2-2
3-1.
Control Panel 3-2
7-1.
Door Seal Removal 7-3 7-2. Parts Identification, Vacuum Pump Oil Change 7-6 7-3.
Door Interlock Override 7-9
7-4. Vacuum Release Solenoid Valve Location 7-9
List of Tables
Table
4-1.
Rotor Information 4-4
6-1.
DIAGNOSTICS, Screen 2 Fields 6-7 6-2. DIAGNOSTICS, Screen 2 Operations 6-7 6-3.
DIAGNOSTICS, Screen 3 Fields 6-8
7-1.
Maintenance Schedule 7-1
Page
Page
vii
Sorvall® Centrifuges
viii
Figure 1-1. SORVALL® RC-28S Supraspeed® Centrifuge
Sorvall®Centrifuges Description
Chapter 1
DESCRIPTION
This manual provides you with the information you need to install, operate and maintain your SORVALL® RC-28S Supraspeed® Centrifuge. If you encounter any problem concerning either operation or maintenance that is not covered in the manual, please telephone Du Pont for assistance. In the United States, call toll free
(800)551-2121.
SORVALL® Products.
General Description
The Sorvall® RC-28S Supraspeed® Centrifuge defines a new class of centrifuge that brings an unparalleled level of performance to centrifugation. Created from the latest
design technology, it incorporates usability, dependability, and safety with an extremely
versatile speed range to set a new standard in value. When used with a Sorvall®
Supraspeed® rotor, the RC-28S can reach a top speed of 28 000 rpm with a maximum g force of 100 095. At 100 095 x g you can perform applications in the RC-28S that previously required an ultracentrifuge. In addition to the Supraspeed® rotors designed for use with the RC-28S, the RC-28S accepts all Sorvall® superspeed rotors, increasing its usefulness by making it capable of a wide range of applications.
Outside the United States, contact the distributor or agent for
The centrifuge is microcomputer controlled and designed to be user friendly. It can be operated using basic parameters only or you can decide to use its special features. Because of this versatility and its user-friendliness, it can easily be operated by inexperienced persons learning centrifugation methods or by experienced persons experimenting with new methods.
The brushless dc drive which offers the optimum in centrifuge drive performance is one distinct feature of the RC-28S. Other features of the centrifuge are: automatic rotor recognition; programmability; RCF and for sample reorientation; overtemperature control; a control panel lock to prevent run parameters from being changed; and a built-in calculator that can be used to perform
centrifugation calculations, including conversions between RCF and speed, time and fco2dt or the calculation of K factors and run times when conversion of published run parameters is desired.
The centrifuge also has extensive diagnostic routines that allow the user to identify system malfunctions and perform maintenance and repair procedures.
For more on the features of the RC-28S, plus more on the centrifuge drive and other systems, read the table on the next two pages and the descriptive paragraphs that follow the table.
fco2dt
control; slow start and slow stop control
1-1
Description
Centrifuge Specifications
Son/all® Centrifuges
Rotor Management
SPIN-RIGHT® Rotor Management System
automatically recognizes installed Sorvall® rotor when chamber door is closed. Pre-
vents operation of rotors above their maxi-
mum speed. Primary function of the system
is to provide an added margin of safety during operation; however it is effective only when it is operated as intended. Any mis­use of the system, as with any other centri­fuge system or feature, could result in unsafe operation.
Rotor Speed
Set Speed:
from 100 rpm to 28 000 rpm in increments of 10
Speed Control:
±10 rpm from 0-2000 rpm;
±0.5 % above 2000 rpm.
Speed Displayed:
digital readout during a run of actual speed
Rotor Temperature
Set Temperature:
from -20°C to +40°C
Overtemperature:
automatic or settable from 0°C to 45°C
Temperature Control:
±10Cfrom2°Cto40°C
Temperature Displayed:
digital readout during a run of actual temperature in 1 ° increments
1
Time
Set Time:
one minute to 99 hours and 99 minutes; can select HOLD for
continuous operation
Time (continued)
Time Displayed:
digital readout indicates remaining time for a timed run or elapsed time for continuous run
Vacuum
Rotary vane pump system.
Chamber pressure:
< 15 millibars (1.5 kPa)
Ambient Temperature Range
from 15°Cto38°C
2
Dimensions:
Width: Height: 93 cm (36.5 in)
Depth:
Clearances:
15 cm (6 in) required on left side of centrifuge
70 cm (27.5 in)
to top deck
123 cm (48.5 in)
to top of control panel
103 cm (40.5 in)
Net Mass
385 kg (850 lbs)
Electrical Requirements
Can be supplied for use on the following single phase power sources
200 V, 60 Hz, 30 A 200 V, 50 Hz, 30 A 208 V, 60 Hz, 30 A3 220 V, 50 Hz, 30 A 230 V, 60 Hz, 30 A3 240 V, 50 Hz, 30 A
Receptacle:
power cord fits Hubbell receptacle No.
9330 or equivalent
1
Temperature range with power ON but no run in process (standby) is from -10°C to +30°C.
2
The centrifuge will operate at ambient temperatures up to 38°C, but refrigeration system performance will be less than optimal above 25°C.
3
CSA Certification applied for.
1-2
Sorvall® Centrifuges
Description
Centrifuge Features
High Performance Drive
A brushless dc direct drive system de-
signed for longer motor life and greater reliability; provides high torque; fast ac­celeration; and smooth, precise speed
control. No brushes means less mainte­nance. Controlled by dual microproces­sors with accurate speed control during acceleration and a Positive Action Brake system that provides controlled braking from set speed all the way to 0 rpm. Includes Sorvall® Gyro-Action Centering System that permits "eye bal­ancing" of tubes. Imbalance detector protects against damage from gross im­balance conditions. Accepts all Sorvall® Superspeed and Supraspeed® rotors.
Self-
Automatic Rotor Recognition
SPIN-RIGHT® Rotor Management System automatically recognizes installed
Sorvall® rotor when chamber door is closed. Provides optimal instrument
control. Prohibits operation of a rotor above its maximum speed. Primary function of the system is to provide an added margin of safety during operation; however, it is effective only when it is operated as intended. Any misuse of the system, as with any other centrifuge system or feature, can result in unsafe operation.
Vacuum
Allows Sorvall® Supraspeed® rotors to be
operated at speeds up to 28 000 rpm.
Automatically activated when required.
RCF Run
RCF can be chosen in place of a speed to control run. Allows RCF to be repro­duced from run to run and rotor to rotor.
Accepts values up to 100 095 x g, de-
pending on the rotor used.
Slow Start / Slow Stop
Automatic control of acceleration and
deceleration to and from transition speed; provides optimal reorientation conditions which prevent disturbance of gradient or pellet. Optimal times to and from transition speed are automatically determined for identified rotor or can be manually selected; transition speed can even be changed if desired. Flexibility of control means you can duplicate a slow start/slow stop run done in any other centrifuge.
Zonal Run
Allows operation with the chamber door
open to dynamically load a Sorvall®
TZ-28 Zonal Rotor. Loading speed selec­tion from 1300 rpm to 2500 rpm. A key number is required for zonal operation.
Precise Temperature Control
A high-capacity, highly efficient refrigera­tion system that combines heating and cooling. The combined heating and cooling provides temperature control from 2°C to 40°C and eliminates the need to defrost the chamber. Tempera­ture sensor is mounted away from the chamber wall to further improve accu­racy of rotor temperature measurement.
Automatic sample temperature protec-
tion; maximum temperature will be set automatically or can be entered manu­ally. Convenient precooling of rotors.
fco2dt Run
A built-in integrator that provides exact run-to-run reproducibility by controlling the accumulating ing value is displayed during the run.
fo)2dt.
The accumulat-
Control Panel Lock
Locks the control panel to prevent run conditions from being changed without
the operator knowing. A key number is required to lock and unlock the control panel.
1-3
Description
Centrifuge Features
Sorvall® Centrifuges
Programmability
Records run settings in memory. Once recorded, programs can be recalled to allow duplication of run. Up to 20 pro­grams can be recorded. A key number is required to save a program to pre-
vent unauthorized changes; however,
programs can be recalled without a key number.
Calculator
A built-in calculator that performs
basic centrifugation calculations: RCF, Speed, Time, leting time. Can be used to convert run conditions from rotor to rotor.
S(o2dU
K Factor, and pel-
Security Key Number
A four digit security number that is
required to access protected data and some diagnostics. Also required to activate Zonal, Program Save, and the Lock. Centrifuge is supplied with one factory-set number, but you can pro­gram multiple numbers for greater se­curity and multiple users.
Advanced Diagnostics
Built-in diagnostic routines include several levels of electronic trouble­shooting aids that allow you to diag­nose problems if they occur. Also used to perform limited maintenance proce­dures and centrifuge updates.
Rotor Chamber
The stainless steel rotor chamber is
epoxy coated and free of wall sensors
making it durable and easy to clean.
Chamber seal is located in the door rather than in the top of the chamber; minimizes accumulation of dirt on the seal which improves the seal. The door seal can easily be removed for cleaning. The sliding chamber door retracts which makes the chamber easy to access and solves door clearance prob­lems.
Size
The RC-28S is narrower than most superspeed and ultraspeed centrifuges. Requires less floor space.
Operator's Guide
Readily accessible operating instruc­tions and aids that pull out from under the control panel. Includes convenient Program Log records and rotor specifi­cations.
Serviceability & Technical Support
Extensive diagnostics and easy to access components make the RC-28S easy to service. Complete product literature available plus a toll free
number to call for technical assistance.
1-4
Sorvall®Centrifuges Description
System Descriptions
Drive System
The RC-28S drive system uses state-of-the-art brushless dc motor technology that
offers the optimum in centrifuge drive performance.
A brushless dc motor uses an electronic commutator in place of a mechanical
commutator, which eliminates brushes and the maintenance problems associated with them, plus it eliminates the speed and current limitations of a mechanical commutator. The high torque of the dc motor results in faster acceleration and braking, while the high efficiency means less power consumption and less heat. Less heat in the motor means that the drive runs cooler which eliminates the need for any auxiliary means of motor cooling, such as Freon® or water.
The motor used is inherently non-cogging which ensures the smoothest acceleration
and deceleration possible. Cogging is typical of induction and brush type motors and can cause sample disturbance during acceleration and deceleration.
The RC-28S drive system is controlled by dual microprocessors allowing extremely accurate speed control during acceleration and deceleration and eliminating overshoot of set speed. The accuracy of the speed control results in smoother slow start and slow stop,
plus positive action braking from set speed to 0 rpm. The microprocessors operate independent of each other to constantly monitor various centrifuge functions and, as an added safety factor, they maintain constant internal communication with each other to test and evaluate the performance of the other. The many tests that are performed routinely throughout operation determine if any preset limitations are exceeded; any failure detected will cause immediate shutdown by the processor detecting the problem.
In addition, the centrifuge will shut down if communication fails between the processors
or in the unlikely event both processors fail.
SPIN-RIGHT® Rotor Management System
The Sorvall® RC-28S SPIN-RIGHT® Rotor Management System automatically insures
optimal control of the selected speed and temperature parameters and prevents the
inadvertent setting of a speed beyond the capability of a given Sorvall® rotor. The Sorvall® rotor installed in the chamber is recognized through an ultrasonic scanning device that is activated when the chamber door is closed. The rotor identity is then displayed in the SET display of the control panel.
The primary function of the rotor management system is to provide an added margin of safety during operation. This system is effective only when it is operated as intended.
Any misuse of the system, as with any other centrifuge system or feature,can result in
unsafe operation
Refrigeration System
The refrigeration system consists of: a stainless steel evaporator that acts as the rotor
chamber, a 2-hp hermetically sealed compressor, an air-cooled condenser, and the
suction-capillary tube assembly. A hot gas bypass circuit allows for quiet, non-cycling
1-5
Description Sorvall®Centrifuges
compressor operation as well as improved temperature control and frost-free operation.
The refrigeration system is factory charged with Freon® 502. The desired run temperature is entered on the control panel and is maintained by the
system electronics which cycles the hot gas circuit as required. If desired, a maximum temperature can be entered on the control panel; if a value is not entered manually, the maximum temperature will default to +5°C above set temperature. Any failure causing the temperature to exceed the maximum setting will result in the termination of the run and a corresponding message will be displayed on the control panel to alert the operator.
Vacuum System
The vacuum system includes a direct drive vacuum pump, an air-inlet solenoid (bleed) valve, mist eliminator, the evaporator chamber, and interconnecting tubes and fittings. Reduced rotor chamber pressure is sensed by a Pirani convection gauge located at the pump inlet manifold and monitored by the machine control microprocessor. When activated, this system provides a vacuum level between 1 and 15 millibars.
The vacuum system is automatically activated only when the set speed is higher than the top air speed of one of the Sorvall® Supraspeed® rotors designed specifically for use with the RC-28S.
In the event of a vacuum loss or insufficient vacuum, a message will appear in the control panel SET display to notify the operator. In an emergency, such as a power failure, when sample recovery is desirable, the chamber door can be opened manually
(see Chapter 7 for procedure).
1-6
Sorvall® Centrifuges
Installation
Chapter 2
INSTALLATION
This chapter contains instructions to prepare your Sorvall® RC-28S for operation.
Inspection
When you receive your centrifuge, carefully inspect it for any signs of shipping
damage. If you find damage, please report it immediately to the transportation company and file a damage claim, then notify Du Pont.
Check the parts received with the centrifuge against the shipping list; if any parts
are missing, contact Du Pont (see back cover).
Location
Locate the centrifuge on a level floor. To permit free air circulation, the location must allow space for the size of the centrifuge (figure 2-1), plus an additional 15 cm (6 inch) space on the left side of the centrifuge. Free air circulation is very important for the centrifuge to function properly.
LEFTSIDE OF
CENTRIFUGE
POWER
CORD
200cm -
(78.5 in)
123 cm
(48.5 in)
Figure 2-1. Centrifuge Dimensions
2-1
Installation
Son/all® Centrifuges
Ambient Temperature
An ambient temperature range of 15°C to 38°C must be maintained. If the inlet air
temperature is above 25°C, the centrifuge will not maintain low temperatures at
high speeds.
Leveling
WARNING
The front stabilizers must be used as explained below to lift the casters off the
floor; this will help minimize centrifuge movement in the event of a rotor
failure.
When the centrifuge is situated at its operating location, lower and secure the front locking stabilizers. Lower the two locking stabilizers by hand until they are finger tight against the floor. Then use the 9/16 inch (14 mm) wrench supplied with the centrifuge to rotate each stabilizer an additional 2 1/2 turns. This will raise the front casters about 3 mm (1/8 inch) off the floor and level the centrifuge (based on level floor).
LOCKING STABILIZERS
(9/16 in [14 mm] HEX)
APPROX. 3mm (1/8 in)
OFF THE FLOOR
Figure 2-2. Leveling Adjustment
Electrical
Your centrifuge was ordered for a specific single phase power source (preferably regulated). The nameplate on the back panel of the centrifuge tells which one of the following classifications you ordered:
200 V, 60 Hz, 30 A 208 V, 60 Hz, 30 A* 230 V, 60 Hz, 30 A*
200 V, 50 Hz, 30 A 220 V, 50 Hz, 30 A 240 V, 50 Hz, 30 A
If the line voltage varies ±10% of its nominal value, you may notice variations in the performance of the centrifuge. If the line voltage exceeds this tolerance, it may damage the centrifuge.
* CSA Certification applied for.
2-2
Sorvall® Centrifuges installation
CAUTION
The centrifuge can be damaged if connected to the wrong voltage. Check the voltage before plugging the centrifuge into any power source. DuPont is not responsible for improper installation.
The centrifuge is equipped with a 3-wire power cord with a 3-prong cap, including ground connection, to fit Hubbell receptacle No. 9330 or equivalent.
To connect the centrifuge to a voltage other than what is specified on the nameplate, the centrifuge will have to be rewired and the power cord may have to be replaced. In this case,
contact DuPont to have a Service Representative do the rewiring.
The ON/OFF main power switch is a 30 A circuit breaker, therefore a separate line disconnect switch is not needed unless required by local codes.
2-3/2-4
Sorvall® Centrifuges Controls
Chapter 3
CONTROLS
This chapter provides:
• a detailed description of the RC-28S controls;
• a brief description of the built-in calculator; and
• a brief description of the diagnostic routines.
For step-by-step operating instructions see Chapter 4, and for further information on the calculator and the diagnostics, see Chapters 5 and 6.
NOTE
A convenient Operator's Guide containing simplified instructions is located under the centrifuge control console.
Control Panel
The RC-28S is easy to use. All commands are entered from the convenient keypad on the control panel and the set and run parameters are visible in the displays. All selected parameters can be locked to prevent changes to run settings. Figure 3-1 on the next page shows the RC-28S control panel and gives a brief description of the keys and displays. A complete description follows the illustration.
Set Display
The SET display is the lower display that changes by pressing the PROGRAM, CALCULATE or DIAGNOSTICS key to show various screens of parameters or operations. A cursor in the SET display indicates the field where the next input character will appear. Lights below the display indicate the position of the cursor.
Run Display
The RUN display includes the three upper displays that show actual speed, temperature, and time or Jco2dt during a run.
3-1
CO
SCREEN INDICATOR
Indicates which screen has been selected and appears in SET display.
CURSOR INDICATOR
Indicates cursor location in SET display.
2
1
SET DISPLAY
Displays parameters to be selected, calculator modes, and diagnostic conditions. Change dis­play by pressing one of
the three control keys be-
low the display.
RUN DISPLAY
Speed
Displays current rotor speed (rpm) during a run.
Temperature
Displays current rotor temperature (°C) during a run.
Time
TIMED Run: displays re-
maining run time. HOLD Run: displays ac-
cumulating run time.
displays accu-
mulating integral value.
Ja/WflNDICATOR
Indicator light is ON when lco2dt run is selected.
START
Starts the centrifuge run.
O
O
a
PROGRAM
Primary control key. Se­lects one screen of basic and three screens of op-
B
tional run parameters.
Display changes each
I
time key is pressed and a
corresponding indicator light above the key comes on to indicate which screen is in the display.
CALCULATE
Selects calculator mode to
perform basic centrifuga­tion calculations. Four screens of calculations available. Display chan­ges each time key is pressed and a corre­sponding indicator light above the key comes on to indicate which screen is in display.
DIAGNOSTICS
Selects diagnostic mode to help isolate malfunc­tions,
provide status of
various systems, and to
perform maintenance pro­cedures.
ENTER
Moves the cursor right or left through SET display. Also causes an input value to be accepted.
ON/OFF +/-
The +/- changes set tem­perature values from plus to minus or from minus to plus.
The On/Off selects brake mode and other options.
STOP
Stops the centrifuge run.
NUMERIC KEYPAD
Used to input numbers to select parameters.
HOLD
Selects a continuous run.
CLEAR
Cancels an input or a pre­viously entered value.
CO
o
5
(D
CO
Sorvall® Centrifuges
Controls
Description of Display & Key Functions
v
j v j V j
Cancels the entry again,
Starts
Stops
the
the
the
The numeric keypad
select parameter values.
The ENTER keys cause accepted, plus they cause the right
On/Off word
On or Off
of
the SET the temperature value from positive negative
an
input
is set to
previously entered value
centrifuge
centrifuge
or a
previously entered value. When
zero
(0) and a new
run.
run.
is
or to the
is
used
to
is
used
to
left through
to
select options. When pressed,
will appear
display.
positive.
value
restored.
input numbers from
an
input value
the
cursor
the SET
in the
The +/- is
can be
input. When pressed
appropriate field
used only
to
the key is
0 to 9 to
to
move
display.
to
negative
pressed,
to be
to
the
change
or
Selects a continuous the
SET
display is pressed, show accumulating
Selects four screens screens
come
the
The four PROGRAM parameters
are
labeled
on to
SET
indicate which screen
display will change
and
1
2
the
1, 2, 3, and 4 and
operations. A description
Rotor Speed Hr:Min Tenp/Max Brake
RCF Slow-Start Slow-Stop Zonal/Speed
3
run. The
and all
timer will start
run
zeros will
time.
Program
of
standard
you
key
to the
screens
next screen.
word Hold will appear
be in the RUN
and the
Key
and
optional centrifugation conditions.
have corresponding lights above
have accessed. Each time
are
illustrated below showing
of the
Time field
Functions & Display Screens
fields
in
each screen follows.
in the
display.
of the RUN
As
the key is
PROGRAM DISPLAY SCREENS
w2dt Tine-out Total Lock
HnMin field
soon
as
START
display will
PROGRAM
The
the key
the
available
that
pressed,
of
4
PROGRflM> Recal1# Save#
3-3
Controls Sorvall® Centrifuges
PROGRAM
Program Key Functions & Display Screens
Rotor Speed HrsMin Temp^Max Brake
Screen
1
F-28^13 23000
0:90
+4+9
On
When the "1" indicator light above the PROGRAM key will include temperature,
changed during a run.
Following is a description of each field in this screen.
the and
basic
brake.
run
All
parameters: rotor identification, speed, time,
parameters, except rotor identification,
is
on, the SET display screen
can be
Rotor Speed HrsMin TeMp/Max Brake
The SPIN-RIGHT® Rotor Management System automatically recognizes the installed Sorvall® rotor and RCF calculation. When the chamber door in the display. Whenever the chamber door the display unless the Lock display.
The rotor identification cannot be changed after START has been pressed.
Iliiiiiil Speed
Centrifugation speed increments of the installed rotor erroneous entry.
to
of
provide temperature compensation, top speed limit, speed control,
is
closed, the rotor name will appear
is
open, the word NONE will appear
is
on,
in
is
10 rpm.
is
In
which case the rotor name will stay
selected
If a
entered, all digits
this case, clear the entry and make a valid entry.
by
entering a value from 100
speed value greater than the maximum allowable speed
in
the Speed field will flash indicating an
to
28 000 rpm
in
in
the
in
NOTE
If an RCF value the chosen RCF value will automatically be calculated and displayed in this field.
The speed setting can Speed field and entering a new value.
is
be
selected
to
control the run, the speed required
changed after a run has been started by returning
to
achieve
Rotor Speed HrsMin TerwMax Brake
Centrifugation time can
Values are displayed exactly
displayed 0:90 which represents one hour and 30 minutes run time. The same run time
of
one hour and
displayed as 1:30.
3-4
be set
as
30
minutes can
from one minute entered;
for
be
example,
selected
to 99
if
by
hours and
you input 90,
entering 130; this will
99
minutes.
it
to
will
the
be be
Sorvall® Centrifuges Controls
PROGRAM
As soon
timer will begin when START from HOLD will appear
(0) timer will begin after Time entering a new value.
as an
set
value
and
will show accumulating
can be
entry to
changed after
is
zero in the SET
made,
to
show remaining
the
rotor
the run has
Program
the
is
pressed
display
run
Key
entered value will appear
and the RUN
time. When operating
has
been loaded
Functions & Display Screens
and the RUN
run
time.
started
by
display value will count down
If
display value will begin
and the
returning
HOLD
NOTE
If
the
integrator accumulated. During integral value
HnMin field. started.
During a timed
value will
be
displayed
is On, the run
a run, the RUN
and the SET
The
run
integrator cannot
with
no
in the
will
end
when
display will show accumulating
be
turned
preselected integral value,
G&dt
Total field
of
the
selected integral value
display will show
Off (or On)
Screen
3.
the
Rotor 5r*eed HrsMin Tenp/Max Ilillfil
. i
in the
in the
to the
after
RUN display.
is
selected,
Zonal mode,
chamber door closed.
Hr:Min field
the
accumulating
run
time
a run has
accumulating integral
the
at
has
in the
been
The word
zero
the
and
There
centrifugation temperature; protection limit.
Centrifugation temperature Acceptable value Temp value
to 0°C).
Max temperature
desired value, however,
operation, value until
are two
is not
If
the
temperature shown
value (shown
indicated temperature
Temp value
the Max
rotor temperature settings.
Max
the
temperature values
entered,
is -6°C or
or at
rotor
the
lower
in the SET
0°C.
can be
temperature setting should
has
and
changed
it cooled
the
other.
can be set
value will default
in
which case
Max,
from -20°C
are
from
NOTE
in RUN
display),
will decrease
must
to the
display
Max
as
by
moving
be at
least
selected temperature.
The
to 5°C
the Max
is 5°C or
temperature will remain
first
one,
selects
0°C to
rotor cools until
the
not be
the
sample overtemperature
to
+40°C
+45°C. above temperature will always default
cursor
above
the
more above entered Temp
to Max and
set
Temp value.
changed from
Brak
Temp, selects desired
in
increments
If a Max
Temp value (unless
it is 5°
e
of 1°C.
temperature
3°C
above
above
entering
the
set
For
default
the
best
Press On/Off Brake field
to
turn
the
to
indicate choice.
brake
on or off. The
word
On or Off
will appear
in the
3-5
Controls Sorvall® Centrifuges
PROGRAM
Program Key Functions & Display Screens
When the brake is Off, the rotor will coast from set speed to zero rpm at the end of the run. If a slow stop has been selected, the rotor will coast to the transition speed, then continue to decelerate at the controlled rate. When the Brake is On, the rotor will brake from set speed to zero rpm at the end of a run; if a slow stop has also been selected, the rotor will brake to transition speed, then continue to decelerate at the controlled rate.
C
o *> "CF
screen
To access, press PROGRAM until the "2" indicator light above the key is lit. The
SET display will include the following fields: RCF, Slow Start, Slow Stop and Zonal.
RCF
z
100095 On fluto On fluio Off 1500
Slow-Start Slow-Stop Zonal/Speed
The RCF field is used to: (1) display calculated RCF value for the identified rotor and the set run speed prior to a run and to display actual RCF value during a run; or
(2) to set the centrifuge to operate at a given g force. When an RCF value is
selected to control the run, the speed required to achieve it will automatically be
calculated and displayed in the Speed field of Screen 1.
During a run, the RCF (at rmax of the identified rotor) at actual speed is always
displayed in this field whether the run is programmed for speed or g force. At the end of a run, the value zero will remain in this field until the chamber door is opened. When the door is opened, the display will change back to show the RCF
value for the set speed.
RCF Slow-Start
To select a slow start, move the cursor to Slow-Start and press will show "On Auto". Auto is the optimal rate automatically calculated to have the
installed rotor accelerate to transition speed in approximately 2 minutes. When the run is started, a message will appear in the SET display to indicate that a slow start is in progress. When the rotor reaches transition speed, the message in the
display will go out and the rotor will accelerate to set speed at the maximum rate. The length of time for acceleration from 0 rpm to transition speed can be changed by positioning the cursor under "Auto" and entering the number of minutes (from 1
to 30) that you want for acceleration. Once Slow-Start has been turned On, it will
remain On until it is turned Off. When Slow-Start is Off, the rotor will accelerate
from 0 rpm to set speed at the maximum rate.
On/Off;
the display
3-6
NOTE
The preset transition speed (approximately 300 rpm for S-20 rotor and approximately 800 rpm for all other rotors) can be changed. See Chapter 6.
Sorvall® Centrifuges Controls
PROGRAM
Program Key Functions & Display Screens
RCF Slow-Start Slow-Stop Zonal/Speed
To select a slow stop, move the cursor to Slow-Stop and press the On/Off key; the
display will show "On Auto". When On, the rotor will decelerate from set speed to transition speed at the rate controlled by the brake setting. When the rotor reaches transition speed, a message will appear in the SET display to indicate that a slow stop has been initiated and the rotor will decelerate at a rate that will take it approximately 3 minutes to reach 0 rpm. This rate of deceleration from transition speed to 0 rpm is automatically calculated for the rotor in use. The rate of
deceleration can be changed by positioning the cursor under "Auto" and entering the number of minutes (from 1 to 30) desired for deceleration. Once Slow-Stop has been turned On, it will remain On until turned Off. When Slow-Stop is Off, the rotor will decelerate at the rate controlled by the brake setting from set speed to 0 rpm.
NOTE
The preset transition speed (approximately 300 rpm for S-20 rotor and approximately 800 rpm for all other rotors) can be changed. See Chapter 6.
RCF Slow-Start Slow-Stop Zonal/Speed
Used to select the Zonal mode of operation when dynamically loading a sample and/or gradient into the Sorvall® TZ-28 Rotor. A Key# is required to activate the Zonal mode.
When the Key# is entered and accepted, the word On appears in the display and
the cursor moves to Speed. Enter the Zonal loading speed from 1300 to 2500 rpm
(in increments of 10 rpm). When the chamber door is closed and START is
pressed, the rotor will accelerate to the Zonal speed, at which time the chamber door can be opened and the sample/gradient loaded. After the sample and gradient are loaded, close the chamber door and the rotor will accelerate to the run speed selected in Screen 1 (the maximum Zonal run speed is 20 000 rpm).
At the end of the run, Zonal will turn Off automatically; therefore, each Zonal run
requires this option be turned On and the Key# entered. Zonal cannot be stored in program memory; the zonal speed can be saved, but the
option must be turned On before the run.
w2dt Tine-out Total Lock
On
1.97e07
This screen includes w2dt and Lock. To access, press PROGRAM until the "3" indicator light above the key is lit.
5.2QeQ7
On
3-7
Controls Sorvall® Centrifuges
PROGRAM
Program Key Functions & Display Screens
w2dt Tine-out Total Lock
The Ja)2dt option than a time value. Press On/Off previous
deceleration,
(ENTER
Time-out value, first enter a three digit (X.XX) coefficient value will automatically
the
During and
accumulating value. During a timed
value The integrator cannot
run
O) or
run.
an
integrator
the SET
and the
is
used
will automatically
and
Total
(START)
be
display will show selected Time-out integral value
Time-out value
to
program
to
be
is the
to
accept
placed), then
run, the RUN
be
turned
value
the
at the
On or Off
a run to end at a
turn
the |
displayed: Time-out
at 0 rpm,
Time-out value
a two
beginning
digit exponent value. Press (START)
display will show accumulating integral value
run, the SET
after
u2dt 1 on. The
which includes deceleration. Press
and
of
deceleration.
a run has
desired integral value rather
J(o2dt values from
is the
begin
display shows
the run. To
been started.
value
w2dt Tine-out Total Lock
This feature locks
changed until unlocked. proper Key#. Once in
the
Lock field
enter
the
Key#.
all set run
The
run
conditions have been locked
and in all
conditions (including options)
Lock
calculator screens.
is
activated
by
pressing On/Off
in, the
To
unlock, press On/Off again
cursor will only appear
at the
(the
decimal point
and the
the
accumulating
so
they cannot
and
start
change
to
begin
Total
entering
the
of
the
be
the
and
The lock cannot
~
A
screen
Run conditions control centrifuge operation. however, a program
be stored
Chapter
To access, press PROGRAM until
and a
6).
be
stored
in
program memory.
PROQRflM> Recall* Sauett
4
1
are
stored
can be
different Key#
in
program memory
A
Key# must
recalled without
can be
the "4"
PROGRFlM> Recall** Save*
Used
to
recall the assigned program number (from will automatically appear run.
an
established program. Position
in
Screen
1 of the SET
^
and
be
entered
the
Key#. A total
assigned
indicator light above
1 to 20). The
to
the display. Press START
recalled from memory
to
save
each program,
the key is lit.
cursor recorded program parameters
in
a new
of 20
this field
programs
if
desired
to
program;
and
begin
to
can
(see
enter
the
3-8
Sorvall® Centrifuges
Controls
PROGRAM
Program
Key
PROGRRM> Recall* Sauett
Used
to
save
all set this field Key#. When
and
enter a number from
the assigned program number appears that
was
the program.
which case,
input becomes
If
the program numbers have been used, a new program.
old program
In
this case,
is
automatically deleted from memory.
The CALCULATE RCF,
speed,
below.
See
run
Chapter 5 for
run
conditions except Zonal
1 to
Key#
is
entered
and
in the
the
assigned program number that will
the
incorrect Key#
is
program number must
the
number assigned
Calculate
key
is
used
to
perform basic centrifugation calculations, including
time, Ju)2dt
and K Factor.
instructions
on
accepted,
entered,
be
Key
how
Functions & Display Screens
and
20. The the
Save#
and
Lock. Position
display will prompt
display shows "Saved"
Recall# fields.
the
you
The
be
the
display will show
reentered, then
program
is
can
stored
for
the
be
stored over
the
new
"Not
correct Key#.
program
Functions & Display Screens
The
four calculator screens
to
use the
calculator.
used
Saved";
an
are
cursor
to
enter
and
in
the
number
to
recall
in
If
all
existing
and the
CALCULATE
pictured
1 2 3 4
CflLO CRLO CflLO CRLO
Rotor
r(om)
Speed
Speed Speed K-Faciorl
Rrcax
Tinel
Diagnostics Key Functions & Display Screens
DIAGNOSTICS
is
used
to
monitor system operation, view system accumulators,
perform user diagnostic routines, install
calibrate replacement subassemblies.
screens
and
instructions
1 2 3
Fault
User-1 Maint-2
on
the use
Door- Sysien
User-2 Calib
of
See
this
new
key.
RCF
Tine w2dt
RCF
Rnin K-Factor2
rotors
and
Chapter
Maint-1 Serv-1
K-Factor
Tine2
software, troubleshoot,
6 for
examples
Instal
of
1
Serv-2
DIAGNOSTICS
and
display
3-9/3-10
Son/all® Centrifuges Operation
Chapter 4
OPERATION
This chapter provides:
• step-by-step operating instructions for the normal and zonal modes of operation; RCF and stop;
and HOLD runs;
• a description of the control panel lock and the Key# code;
• a precool/preheat procedure to equilibrate rotor and chamber temperature;
and
• a Rotor Information Chart.
Jco2dt
operation; programmed runs; slow start and slow-
Understanding the Centrifuge Controls
It is recommended that you read the description of the controls in Chapter 3
before using the centrifuge for the first time. You probably won't use all of the
controls right away, but reading Chapter 3 will give you information that will be helpful in understanding the displays and how the control keys function, and it will help you become familiar with all features of the centrifuge. As you use the centrifuge, continue to refer to Chapter 3 whenever additional information is
needed.
Turning the Centrifuge On
The power ON/OFF switch is placed in the upper right hand corner of the front cabinet panel. The microprocessor automatically runs a diagnostic test everytime the centrifuge is turned on. If an error is found, a display message will appear in the SET display on the control panel. If no error is found, the run parameters will appear in the display indicating that the centrifuge is ready for operation.
Normal Run
A normal run is one that is controlled by time with selected run temperature and
run speed parameters and the brake set on or off. These parameters are selected in the first screen of the (PROGRAM) key.
Prior to performing the run, prepare and install the rotor according to the
instructions in the rotor manual. To ensure that the sample, rotor, and chamber temperatures are at equilibrium at the start of the run, follow the precool/preheat procedure on page 4-12.
4-1
Operation Sorvall® Centrifuges
WARNING
When loading the rotor, be sure not to exceed the maximum compartment mass of the rotor (see Table 4-1, Rotor Information). If maximum compart­ment mass is exceeded, the maximum operating speed must be lowered
(see Reducing Speed for Rotor Compartment Loads In Excess of Design Mass,
page 4-12). Failure to do so can cause rotor failure which could result in
personal injury and/or centrifuge damage.
To perform the run:
1.
Turn the centrifuge power ON.
NOTE
For best run conditions and prolonged centrifuge life, the rotor chamber should be dry at the start of a run. Before installing the rotor, wipe chamber
dry with a clean cloth or papertowel.
2.
Open chamber door all the way and install rotor. Close chamber door.
WARNING
Use Sorvall® rotors only. Use of another manufacturer's rotor can cause rotor failure which could result in personal injury and/or centrifuge damage.
3.
The
"1"
light above the (PROGRAM) key should be on. If not, press (PROGRAM) until it is.
The name of the installed rotor will appear in the display and the cursor will be positioned in the | Speed I field.
NOTE
If using an to the 1 Rotor 1 field and input 18 for rotor will be recognized as an S-20/36. Press (ENTER O).
4.
Input desired run parameters: | speed|, lHr:tiinl, and jTer>p/riax|. Press (ENTER E>) after
each input. Max temperature will automatically default to 5°C above set run temperature; if
desired, enter a different value. For a continuous run, press (HOLD).
5.
Press (onoft) to make 1 Brake 1 selection. The word |0n| or [off | will appear in the
display to indicate selection made.
S-20/17
or
S-20/20
rotor ONLY: Press (ENTER <J) to move the cursor
S-20/17
or 17 for S-20/20; otherwise the
After you have selected the above parameters, the SET display should look similar
to the example below:
Rotor Speed HrsMin Tenp/Max Brake
F-28/13 28090
4-2
0:90
+4+9
On
Sorvall® Centrifuges Operation
6. Select [siou-start] and/or [siou-SioH if desired (see below).
7.
Press (START).
Any of the run parameters, except rotor identification, can be changed while a run
is in process.
The rotor will accelerate to the set run speed and remain there for the selected
length of time then decelerate to a stop. To end a run in HOLD or before the selected run time has elapsed, press (STOP).
NOTE
Integral value can be checked anytime during a timed run by pressing (PROGRAM) until | u2dt | is in the display ("3" indicator light on). The accumulating fa)2dt value will be in the | Total 1 field.
To do another run with the same parameters: prepare rotor and install it in the
centrifuge, then close chamber door and press (START) . All entered parameters
remain until a change is made.
Slow Start/Slow Stop
Slow Start and Slow Stop are used to prevent gradient disturbance during acceleration and deceleration. When they are on, the rotor will accelerate or
decelerate to or from transition speed at a controlled rate that is slower than normal (transition speed is preset at approximately 800 rpm for all rotors except the S-20, which is preset at approximately 300 rpm; preset transition speed can be changed, if desired - see Chapter 6). The length of time for the acceleration/ deceleration phase will automatically default to the optimal time for identified rotor; or, if desired, time (from 1 to 30 minutes) can be entered manually.
NOTE
If Slow Start or Slow Stop are on, they will remain on for future runs until
turned off.
1.
Follow Steps 1 through 5 for a Normal run (page 4-2).
2.
Press (PROGRAM). The "2" indicator light above the key will light.
To select Slow Start:
1.
Press (ENTER O) to move cursor to |siou>-stari|.
2.
Press (on off).
3.
To accept the "Auto" time for acceleration and begin the run, press (START).
RCF blow-Start blow-Stop Zonal/Speed
100995
To change the length of time for acceleration from 0 to transition speed, press
(ENTER c>) to move cursor to | Ruto
30).
If a slow stop is desired (see page 4-5); if not, press (START) to begin the run.
On
fluto
Off Off 1509
I
and input desired number of minutes (from 1 to
4-3
Operation
Use Sorvall® rotors only. Use of another manufacturer's rotor can cause rotor failure which could result in personal injury and/or centrifuge damage.
Sorvall® Centrifuges
Table 4-1. Rotor Information
WARNING
ROTOR
SV-288
SV-80
GS-3
SA-600
SS-34
SE-12
HB-4
HS-4
SM-24
GSA
TZ-28 SH-MT SH-80
CODE
#
1
2 3
4
5
6
7
8
9
10
11 12 13
MAX
SPEED
(rpm)
20 000
19.000 9000
16 500
20 000
21000
13 000
7000
20 000
13 000
20 000
13 730
20 000
CRITICAL
SPEED
(rpm)
1000
$
1600 1650
1200 1050
750
1200
1350 1700 1200
1000
1300
900
MAX
RCF
40 301
40 96$
13 689
39 411
47 807
45 959
27 579
9425
49 460
27 504
42 535
19140
45 394
FACTOR
4201
1678
4487
2
2019
s
K
210
97
792
750
513
457
605 656 399
COMPART-
MENT MASS
250.0 g
1035,0 0
2
680.0 g
1620.0 g
3
MAX
50.0 g
6.4 g
780.0 g
115,0$
115.0 g
30,0 $
27.0 g
36.4 g
78.0 g
r
MAX (cm)
9.02
10.16
15.13
12.96
10.70
9.33
14.61
17.22
11.07
14J7
9.52
9.Q9
10.16
r
MIN
(cm)
6.47
8.84
3.93 5,52
3.26
3.81
4.75
7.21
2
5.37
2
3.77
3.65
3
5.57
5.40
3
F-28/13
F-28/36
3-20/36
S-20/20
S-20/17
SS-34/KSB-R
TZ-28/GK
1
With maximum allowable volume at maximum speed.
2
Outer row
3
Maximum speed of rotor is 20 000 rpm; actual maximum speed depends on tubes. Values given are typical for 1.5 ml microtubes.
14
15
16
17
18
19
20
28 000
28 000
20 000
20 000
20 000
20 000
19 000
1300
1300
400
400
400
1350
1000
100 09$
100 095
71934 58 083
7416S
47 807
38 38$
143
198
466
299
573
750
671
23,0 8
66.0 g
. 156.6 g
115.1 g
130.1 $
115.0g
1620.0g
11x43
11.43
16.10
13.00
16*60
10.70
9.52
4-4
7,34
6.18
7.70
8.10
6.70
3.26
3,65
Sorvall® Centrifuges Operation
As soon as (START) is pressed, the |*** sioui start in Progress ***[ message will appear in
the SET display and will remain there until the rotor has accelerated to approximately 20 rpm.
To select Slow Stop:
1.
Press (ENTER E>) to move cursor to [Sloui-stop].
2.
Press (on off).
3.
To accept the "Auto" rate of deceleration and begin the run, press (START).
RCF Slow-Start Slow-Stop Zonal/Speed
190095 On Ruto On Ruto Off 1500
To change the length of time for deceleration from transition speed to 0 rpm,
press (ENTER O) to move cursor to I nuto | and input desired number of minutes (from
1 to 30). To begin run, press (START).
During deceleration, the | *** siou stop in Progress *** [ message will appear in the SET display as soon as the rotor reaches transition speed; the message will remain
in the display until the rotor stops spinning.
RCF Run
The centrifuge can be set to operate at a given g force rather than speed. When you know the RCF (g force) you want applied to the particles during centrifugation, you can use this feature to select the value. The run speed required for the set RCF value will automatically be calculated.
To perform an RCF Run:
1.
Load and balance the rotor according to the instructions in the rotor manual. To
ensure that the rotor and chamber temperatures are at equilibrium, follow the
precool/preheat procedure on page 4-12.
WARNING
Use Sorvall® rotors only. Use of another manufacturer's rotor can cause rotor failure which could result in personal injury and/or centrifuge damage.
Do not exceed maximum compartment mass of the rotor (see Table 4-1, Rotor Information). If maximum compartment mass is exceeded, the maximum operating speed must be lowered (see Reducing Speed for Rotor Compartment Loads In Excess of Design Mass, page 4-12). Failure to do so can cause rotor failure which could result in personal injury and/or centrifuge
damage.
2.
Turn the centrifuge power ON.
4-5
Operation Sorvall® Centrifuges
NOTE
For best run conditions, the rotor chamber should be dry at the start of a run.
Before installing a rotor, wipe the chamber dry with a clean cloth or
papertowel.
3.
Open chamber door all the way and install rotor. Close chamber door.
4.
The
"1"
light above the (PROGRAM) key should be on. If not, press (PROGRAM) until it is.
NOTE
If using an S-20/17 or S-20/20 rotor ONLY: Press (ENTER O) to move the cursor to the 1 Rotor 1 field and input 18 for an S-20/17 or 17 for an S-20/20; otherwise
the rotor will be recognized as an S-20/36. Press (ENTER O).
5. Input desired lHr:Minl and | Tenp/Max | values. Press (ENTER O) after each input. (Speed is not entered because run speed will automatically be calculated and set when
the RCF value is entered.) Max temperature will automatically default to 5°C above set run temperature; if
desired, enter a different value. For a continuous run, press (HOLD).
6. Press (on off) to make 1 Brake I selection. The word |0n | or joff | will appear in the display to indicate selection made.
7. Press (PROGRAM) again. I
RCF I will appear in the display with the cursor positioned in the I RCF 1 field and the
"2"
light above the (PROGRAM) key will be on.
8. Input desired |RCF| value (for rmax of the installed rotor). Press (ENTER O).
The display should look similar to the example below.
RCF Slow-Start Slow-Stop Zonal/Speed
93180 Off Off Off 1500
9. Press (START).
The rotor will accelerate to the speed required to achieve the set RCF value and remain there for the selected length of time or until (STOP) is pressed, then
decelerate to a stop.
Key# Code
The Key# is a security number that must be entered in order to access Zonal, Program Save, the control panel Lock, and some Diagnostic routines. Whenever you attempt to use these features, the display will prompt you to enter the Key#.
4-6
Sorvall® Centrifuges Operation
A different number can be assigned to each of the features that require a Key#,
including a different number for each of the 20 programs saved. There is also a System Access Key# that allows access to any of the fields that require a Key#, except for some Diagnostic routines. The centrifuge is shipped with the System
Access Key# 1234. This is the only number assigned when you receive the
centrifuge. For the procedures to change the System Access number or to assign
Key#s to the various features, see Chapter 6.
Zonal Run
Zonal operation allows you to operate the centrifuge with the door open at speeds from 1300 rpm to 2500 rpm. This is used with a Sorvall® TZ-28 Zonal Rotor to dynamically load gradient and sample. A Key# is required to access Zonal mode.
Assemble the rotor according to the procedure in the rotor instruction manual. To
ensure that rotor and chamber temperatures are at equilibrium at the start of the
run, follow the precool/preheat procedure on page 4-12.
To perform the run:
1.
Turn the centrifuge power ON.
2.
Open chamber door all the way and install rotor. Close chamber door.
3.
The
"1"
light above the (PROGRAM) key should be on. If not, press (PROGRAM) until it is.
4.
Input desired run parameters: I Speed I, |Hr:Minl and | Tenp/Max]. Press (ENTER O) after
each input.
NOTE
Maximum speed in the Zonal Mode is 20 000 rpm.
Max temperature will automatically default to 5°C above set run temperature; if desired, enter a different value. For a continuous run, press (HOLD).
5.
Press (on off) to make | Brake I selection. The word
display to indicate selection made.
6. Press (PROGRAM) until the "2" light above the key is on and [ Zonal /Speed | appears in the display.
[ On
| or |off [ will appear in the
NOTE
It is best to use a slow start and slow stop when doing gradient separations in
the TZ-28 Rotor (see Slow Start/Slow Stop, page 4-3).
7.
Select Isiou-start] and/or lsiou-stop] if desired.
8. Press (ENTER t>) to move the cursor to [ Zonal /Speed [. Press (on off). The display will
prompt you to enter | Ke^tt | (page 4-6).
9. Input | Ke^tt I and press (ENTER O). When the Key# has been accepted, the word
will appear in the display indicating that the centrifuge will operate in the Zonal mode.
I On I
4-7
Operation Sorvall® Centrifuges
10.
Input the desired loading speed (from 1300 rpm to 2500 rpm in increments
of 10). Press (ENTER D>).
At this point, the display should look similar to the following:
RCF Slow-Start Sloui-Stop Zonal/Speed
42535
When rotor reaches selected loading speed, the I*** Rotor at Zonal Speed> Load ***| message will appear in the display indicating that the door can be opened.
When operating the centrifuge with the chamber door open and the rotor
spinning, remove any articles that could fall into the chamber (e.g., jewelry,
necktie, objects in shirt pockets). Tie back long hair and roll up shirt sleeves.
12.
Open chamber door and load the sample/gradient according to the instructions
in the rotor manual. When loading is complete, disconnect and remove any
equipment that was used to load it.
Off Off On 1500
WARNING
WARNING
When loading the sample and gradient, be sure not to exceed maximum
compartment mass specified for the rotor (see Table 4-1, Rotor Information). If maximum compartment mass is exceeded, operating speed must be lowered (see Reducing Speed for Rotor Compartment Loads In Excess of Design Mass, page 4-12). Failure to do so can cause rotor failure which could
result in personal injury and/or centrifuge damage.
13.
Close the chamber door. The timer will begin and the rotor will accelerate to
the set run speed and remain there for the selected run time. At the end of the run, the rotor will decelerate to 0 rpm and the Zonal function will turn off
automatically.
Hold Run
A Hold run is a run that will continue until (STOP) is pressed. All parameters except
time are entered as described for the type of run you will perform.
To select Hold:
Press (HOLD). Hold will appear in the iHrtliinl field of the SET display. While the
run is in process, the RUN display will show the accumulating time.
To end a HOLD run:
Press (STOP). The rotor will begin to decelerate and the run will end.
4-8
Sorvall® Centrifuges Operation
To change a HOLD run to a timed run:
Press (HOLD). The key is a toggle switch so pressing it while in HOLD will return the centrifuge to a timed condition. The last entered time value will appear in
the iHr:Hinl field of the SET display and the RUN display will show the remaining time.
NOTE
When the centrifuge returns to a timed condition, the computer will account for the length of time the centrifuge was in HOLD. Therefore, if the last entered time value had already elapsed when the HOLD key was pressed again,
the RUN display will show zeros (0.00) and the rotor will begin to decelerate.
To enter a new time value and cancel the accumulated HOLD time:
Enter the new time value. The |Hr:Min| field of the SET display will show the new set time, and the RUN display will change to countdown from the newly entered value. The length of time the centrifuge was in HOLD has no affect on
the new set time.
fw2dt Run
An fa)2dt run allows you to set the centrifuge to end a run when a desired integral value is reached.
To perform an fo>2dt run:
1.
Load and balance rotor according to instructions in the rotor manual. To
ensure that the rotor and chamber temperatures are at equilibrium, follow the
precool/preheat procedure on page 4-12.
WARNING
Use Sorvall® rotors only. Use of another manufacturer's rotor can cause
rotor failure which could result in personal injury and/or centrifuge damage.
Do not exceed maximum compartment mass of the rotor (see Table 4-1, Rotor Information). If maximum compartment mass is exceeded, the maximum operating speed must be lowered (see Reducing Speed for Rotor Compartment Loads In Excess of Design Mass, page 4-12. Failure to do so can cause rotor failure which could result in personal injury and/or centrifuge damage.
2.
Turn the centrifuge power ON.
NOTE
For best run conditions, the rotor chamber should be dry at the start of a run. Before installing a rotor, wipe the chamber dry with a clean cloth or
papertowel.
4-9
Operation
3.
Open chamber door all the way and install rotor. Close chamber door.
4.
The
"1"
light above the (PROGRAM) key should be on. If it is not, press (PROGRAM) until
it is.
NOTE
Sorvall® Centrifuges
If usingan_S-20/17 or S-20/20 rotor ONLY: Press to the [Rotor| field and input 18 for an S-20/17 or 17 for an S-20/20; otherwise
the rotor will be recognized as an S-20/36. Press (ENTERED).
5. Input desired I Speed! and press (ENTER >) twice to move the cursor to | Tenp/Max I. (Time is not entered since the run will end when the integral value has
accumulated.)
6. Input desired 1 Tenp/Max 1 values. Press (ENTER PQ after each input.
7. Press (on off) to make [Brake! selection. The word [On] or | Off | will appear in the
display to indicate the selection made.
8. Press (PROGRAM) until the "3" light above the key is lit and |
display. The word [on] or [off] will appear in the | u2dt | field with the integral values
from the previous run displayed next to it. See example below.
CENTER<Q
to move the cursor
u>2dt I
appears in the
NOTE
Tine-out | is the Total | is the fafdt value from the previous run at 0 rpm, which includes the
deceleration time.
J(o2dt
value from the previous run at the start of deceleration.
u2dt Tine-out Total Lock
On
1.94e07
5.29e07
Off
9. If Off, press (on off) to turn On.
10.
To use the |Tine-out) value, press (START). TO change the [ Tine-out 1 value, press
(ENTER
a two digit exponent value. Press (START).
The rotor will accelerate to the set speed and remain there until the integral value
has accumulated.
To end a run before the selected integral value has accumulated, press (STOP).
t>) and enter a new value: first a three digit (X.XX) coefficient value, then
Programmed Run
The RC-28S memory will store up to 20 programs. Each program saves a set of run
parameters that can be recalled by entering the assigned program number (from 1 to 20) and pressing (START).
4-10
Sorvall® Centrifuges Operation
A Key# must be entered in order to store a new program or to make a change to an existing program; however, a stored program can be recalled without the Key#. A different Key# can be assigned to each program, if desired; see Chapter 6 for procedure.
Program Save
1.
Enter all desired run parameters.
NOTE
Program Save will not store if the control panel Lock is on; also, Program
Save will not turn Zonal On.
2.
Press (PROGRAM) until the "4" indicator light above the key is lit.
3.
Press (ENTERO) to move the cursor to fsavettl and input a number from 1 to 20.
Press CENTER
4.
Input 1 Ke'd# | (page 4-6). Press (ENTER O) to store program. Press (START) to begin the
run.
PROGRflN > Reca11# Save#
1
1
5.
Record program number and run parameters in the Program Log Record
(located in back of the Operator's Guide under the centrifuge control panel).
NOTE
Program Recall will not turn Zonal On, but it will set loading speed and all
other parameters.
Program Recall
To recall a program:
1.
Press (PROGRAM) until the "4" indicator light above the key is lit.
2.
Input assigned program number (from 1 to 20). Press (START).
All prerecorded run parameters will automatically be established to control centri-
fuge operation. The displays will change to show the set run parameters for the selected program.
Saved
Control Panel Lock
The control panel Lock will lock all set run conditions (including options) so they
cannot be changed until unlocked.
To lock or unlock set run conditions:
1.
Press (PROGRAM) until the
"3"
indicator light above the key is lit.
4-11
Operation Sorvall® Centrifuges
2.
Press (ENTER oj to move the cursor to I Lock].
3.
Press (onoff) and input 1 Ke*#l (page 4-6). Press (ENTER O).
After run conditions have been locked in, the cursor will only appear in the [Lock field and in all calculator screens.
Precool/Preheat
The following precool/preheat procedure is recommended to ensure that the rotor and the chamber temperatures are at equilibrium at the start of a run.
1.
Open chamber door and install rotor. Close chamber door.
2.
Press (PROGRAM) until the
3.
Input I speed I of 2000 rpm. Press (ENTER O).
4.
Press (HOLD), then press (ENTER O).
5.
Input desired iTenpl. Press (ENTER O).
"1"
light above the key is lit.
6. Press (START) . The rotor/chamber should achieve temperature in approximately 30 minutes.
Reducing Speed for Rotor Compartment Loads
in Excess of Design Mass
There is a maximum allowable compartment mass established for each centrifuge rotor (see Table 4-1, Rotor Information). To prevent rotor failure, the total
contents of any compartment, including specimen, tubes, covers, and adapters (if used), proportionately.
Strict adherence to the maximum allowable compartment mass or reduced speed
(see below) is required to prevent rotor failure. Observe WARNING on the Safety
Information Page in front of this manual.
The rotor speed is reduced in proportion to the square root of the ratio for the
maximum allowable compartment mass to the actual compartment mass (including
specimen, tubes, covers, and adapters). If the compartment mass is more than
that specified for the rotor, the reduced speed can be determined by using the
formula given below.
must not exceed the figure given in Table 4-1 unless rotor speed is reduced
4-12
„_,_,„ Reduced
^ ... „ « _, * I Maximum Compartment Mass
Rotor Speed = Max.mum Rotor Speed x \ Actual Compartment Mass (grams)
Sorvall® Centrifuges
Calculator
Chapter 5
CALCULATOR
The RC-28S calculator allows you to perform basic centrifugation calculations to
help determine run conditions necessary to provide desired separation results. Four screens of calculator functions are available. When you enter the CALCULATE
mode, the values displayed in Screens 1, 2 and 3 are calculated from the data entered in the first three screens of the PROGRAM mode. To do calculations for a different rotor, enter the rotor code number (see Table 4-1, Rotor Information) in j
Rotor 1 field of CALCULATE Screen 1. The values displayed in Screen 4 (if any) are
from the previous calculation in that screen.
CALCULATOR DISPLAY SCREENS
2 3 4
1
CRLO CflLO CRLO CflLO
Rotor Ken) Speed
Speed Speed
Rroax
K-Factorl Tinel
RCF Tine
RCF
Rnin
K-Factor
K-Factor-2
w2dt
Tine2
liBIllillll CflLO Rotor Speed RCF Tine w2dt
Use to calculate SPEED, RCF (at rmax), TIME, and fo)2dt. To access, press
(CALCULATE) until the "1" indicator light above the key is on. Enter rotor code
number (see Rotor Information Chart) in the j Rotor 1 field.
To calculate SPEED from RCF:
1.
Position cursor in
2.
Press (ENTER O»). The calculated speed will be displayed.
To calculate RCF (at rmax) from SPEED:
1.
Position cursor in 1 Speed 1 field and input speed.
2.
Press (ENTER o). The calculated RCF value will be displayed.
|RCF 1 field and input RCF value.
To calculate estimated run TIME from
1.
Position cursor in | Speed 1 field and input speed.
2.
Press (ENTER t>) to move cursor to | u>2dt | and input Jafdt value.
3.
Press (ENTER O). The calculated run time will be displayed.
To calculate estimated
1.
Position cursor in | Speed [field and input speed.
2.
Press (ENTER O) to move cursor to [Tinel and input time.
3.
Press (ENTER t>). The estimated integral value will be displayed.
* It is not possible to exactly calculate Time from
approximate value to help you save time.
fa>2dt
from SPEED and TIME: *
j<o2dt
J<a2dt
or\w2dt from Time. These calculations are meant to give an
and SPEED: *
5-1
Calculator Sorvall® Centrifuges
Calculator Use
Screen 2 ICflLO r(cn) Speed RCF
Use to calculate RCF or SPEED at a rotor radius other than the maximum. To
access, press (CALCULATE) until the "2" indicator light is on. The values in the display are from Screen 1 of the (CALCULATE) key.
To calculate SPEED from RCF at a radius other than rmax:
1.
Position cursor in iKcrOl field and input radius.
2.
Press (ENTER O) to move cursor to |RCF 1 and input RCF.
3.
Press (ENTER P). The calculated speed will be displayed.
To calculate RCF from SPEED at a radius other than rmax:
1.
Position cursor in lr<cn)| field and input the radius.
2.
Press (ENTER O) and input [SpeedI.
3.
Press (ENTER P). The calculated RCF value will be displayed.
CRLO Speed Rmax Rnin K-Factor
Use to calculate K Factor. To access, press (CALCULATE) until the "3" indicator light above the key is on. The values in the display are from Screen 1 of the (CALCULATE) key.
To calculate K Factor at a particular speed using rmin and rmax diplayed:
1.
Position cursor in 1 Speed | field and input desired speed.
2.
Press (ENTER P). The calculated K Factor will be displayed.
To calculate K Factor using an rmin or rmax other than what is displayed:
1.
Verify speed and change if desired.
2.
Move cursor to | Rrcax | and/or [Rninl field and input radius value.
3.
Press (ENTER P). K Factor will be calculated and displayed.
Screen 4 CfiLO K-Factorl Tinel K-Factor-2 Tine2
Use primarily for pelleting runs to convert centrifugation time from one rotor (Rotor-1) to another rotor (Rotor-2). To do so, you need to know the K Factor at desired run speed for both rotors.
To calculate:
1.
Press (CALCULATE) until the "4" indicator light above the key is on.
2.
Input K Factor of
3.
Input run time from Rotor-1 in iTineil field. Press (ENTER P).
4.
Input K Factor of Rotor-2 in 1K-Factor2l field. Press (ENTER P). The run
time calculated for Rotor-2 will be displayed in jTine2| field.
Rotor-1.
Press (ENTER P).
* K Factor (the rotor clearing factor) is a function of sample path length (distance between rmin and rmax) and rotor speed. The
lower the K Factor value, the faster a given centrifugation protocol can be completed. Protocols can be converted from one rotor to another if K Factors are known for both rotors.
5-2
Sorvall® Centrifuges Troubleshooting
& Diagnostics
Chapter 6
TROUBLESHOOTING
and DIAGNOSTICS
This chapter provides:
• a description of SET display messages;
• DIAGNOSTICS information;
• the procedure to assign or change Key#'s; and
• the procedure to change the Slow Start/Slow Stop transition speed.
Display Messages
The RC-28S has a number of display messages that can appear in the SET display
on the control panel. These messages notify you of an operating condition or a centrifuge fault that, in most cases, requires attention or demands action.
Following is a list of the SET display messages, along with the probable cause (in bold print) and the corrective action.
***
Check Installed Rotor
The rotor management system did not recognize the installed rotor. Possible causes are:
1.
the rotor management system was not properly activated;
2.
the installed rotor is not intended for use in the RC-28S;
3.
a rotor code was manually input that is incorrect for the installed rotor; or
4.
rotor/rotor lid not properly installed.
1.
To reactivate rotor management system, open the chamber door all the way and
close it again slowly.
2.
If the installed rotor is not meant for use in this centrifuge, remove it and install
one that is (see Table 4-1, Rotor Information).
Use Sorvall® rotors only. Use of another manufacturer's rotor can cause rotor failure which could result in personal injury and/or centrifuge damage.
*#*
WARNING
3.
Install rotor to match selected rotor code; or, if the rotor you want is installed,
clear the message by opening and closing the chamber door. The rotor
management system will automatically recognize the installed rotor.
4.
Properly install rotor/rotor lid; refer to rotor instruction manual.
6-1
Troubleshooting Sorvall® Centrifuges & Diagnostics
*## Close Chanber Door
The chamber door
Close
the
chamber door
the door
Centrifuge malfunction.
Press (DIAGNOSTICS)
(page
all the way
***
Fault> Press DIRGNOSTICS ***
6-5) for
more information.
***
There
pressed.
Open
was no
the
chamber door
was not
first, then close
to
shut
Install Rotor
rotor installed when
***
completely closed when
all the way and
it.
off
alarm
press (START);
and
select DIAGNOSTIC mode.
***
the
chamber door
and
install a rotor. Close
the
the
(START)
it may be
was
door
and
key was
necessary
See
closed
press (START).
and
pressed.
to
open
Diagnostics
(START)
was
**# Open Chamber Door
Run
has
ended.
Open chamber door
***
Power Outage> Run Interrupted ***
A power outage occurred during restored.
No corrective action required. Press (PROGRAM)
**# Pouer Outage> Run Terminated **#
A power outage occurred during temperature significantly varied from
Restart
run.
and
remove rotor.
*##
the run. The run
the run. Run
set
parameters.
resumed when power
to
clear message.
terminated because speed and/or
was
6-2
Sorvall® Centrifuges Troubleshooting
& Diagnostics
*** Recalled Rotor Hot Installed ***
A program was recalled that requires a rotor different than the one installed in the
chamber.
Install the correct rotor for the program recalled, or verify that you used the correct program number.
*#*
Release Door Handle ***
Indicates that you tried to open the door before the door latch had released.
Release the door handle and wait until the message clears, then open the door. If the problem does not clear, the door latch solenoid may be defective.
*#*
Replace NOURflM Battery Soon *#*
Battery for the centrifuge memory is low and must be replaced. (Average battery life is from 5 to 10 years.)
Call for Service to replace battery.
#** Rotor at Zonal Speed> Load #*#
Zonal run is in process and rotor is at selected loading speed.
Open chamber door and load sample/gradient. Close chamber door.
*** Rotor Imbalance ***
Excessive rotor imbalance; run terminated.
Remove the rotor and balance it according to the instructions in the rotor manual, then install rotor and restart the run.
*** Rotor Overtenperature *#*
Run terminated because display temperature is higher than selected 1 Max temperature. Possible causes are: (1) selected 1 Max 1 temperature set below displayed temperature; (2) rotor not precooled; (3) selected 1 Max [ low for the rotor/speed combination; (4) high ambient temperature; or (5) condenser inlet blocked.
temperature too
1
Check possible causes: (1) if 1 Max 1 temperature was manually set lower than display temperature, run will terminate when (START) is pressed; wait until actual temperature cools to 1 Max 1 temperature setting and restart run or set a higher | temperature and press (START) to restart run; (2) if rotor was not precooled, precool
Max
6-3
1
Troubleshooting Sorvall® Centrifuges
& Diagnostics
it according to procedure on page 4-12; (3) acceptable rotor/speed/temperature
combinations vary depending on rotor used; in some cases, if temperature is set lower than +2°C, rotor temperature may not be attainable at set speed; to achieve temperature, set a lower speed; (4) check ambient temperature: it should be between 15°C and 25°C for optimal performance; will operate at temperatures up to 38°C but performance will be less than optimal; (5) check for 6 inch clearance on left side of centrifuge; clean condenser (to clean the condenser, see Chapter 7). If problem does not clear, there may be a refrigeration system malfunction. See DIAGNOSTICS.
*#*
Secure Rotor Properly
The rotor is not properly locked on the centrifuge drive.
Open the chamber door and tighten rotor locking knob(s); if the rotor has two knobs, tighten the larger one first. Close chamber door and press (START).
***
Self Test
in
Progress
*#*
***
A diagnostic test of all centrifuge systems. This is done once every 24 hours that
the centrifuge power is ON.
No corrective action required.
##* Slow Start in Progress ***
Indicates that a Slow-Start was selected and initiated.
Message will clear automatically when rotor has accelerated to transition speed; no
corrective action.
*## Slow Stop
Indicates that a Slow-Stop was selected and initiated.
Message appears when the rotor has decelerated to transition speed. It will clear as soon as the rotor has stopped spinning; no corrective action.
Tach Failure> Rotor Spinning
in
Progress
*#*
***
Tachometer has failed. Run is terminating, but the rotor is still spinning, even though RUN display may indicate 0 rpm. The chamber door cannot be opened until the rotor has come to a complete stop.
Wait until the
chamber door and remove the rotor. Call for Service.
6-4
I***
Open chamber Door ***[rnessage appears in the display, then open the
Sorvall® Centrifuges
Troubleshooting
& Diagnostics
*** Tine
Twenty
Change vacuum pump
or
** Turn DIRGNOSTICS
The DIAGNOSTIC mode
to
Service Uacuun Punp
***
less vacuum runs remain before vacuum pump
oil as
Off
was
soon
as
possible (see Chapter
Before Start
exited with
one of
not start.
Press (DIAGNOSTICS) twice. Press (on Press (PROGRAM)
***
Uacuun Punp Must Be Serviced ***
to
exit DIAGNOSTICS
off) to
and
turn
start
the
run.
Vacuum will not operate until vacuum pump oil if set speed
is
higher than
the top
air speed*
of
rotor.
Change vacuum pump
oil
(see Chapter
7).
oil
must
be
changed.
7).
**
the fields turned on; centrifuge will
field
off.
Input
is
changed. Centrifuge will not start
the installed Sorvall® Supraspeed®
Q.
and
press (ENTER
O).
* Top
air
speed
is
equal
to
maximum speed
of
the rotor in an RC-5C Centrifuge. See rotor instruction manual.
Diagnostics
General
The DIAGNOSTIC information provided understand tasks that require
the
basics
of the
the use of
DIAGNOSTIC mode
DIAGNOSTICS. Common tasks done using
DIAGNOSTIC mode include assigning Key#s, changing slow start/slow stop transition speed, identifying system faults, shooting. Other routines
are
available qualified service personnel only, therefore, instructions included
best The illustrations below
(Screen
to
in
this manual.
call
Du
4 is
currently
1
Pont
For
for
assistance.
are
not
Fault
1
of 2
information regarding
representative
used.)
Door Systerc
in
this manual
and to
in
DIAGNOSTICS that
use of the
of the
is
intended
teach
the
you to
vacuum pump
are
designed
for
their
service routines,
three DIAGNOSTIC screens.
to
help
you
perform routine
the
oil and the
and
trouble-
for use by
use are not
it is
2
3
User-l
Off
Maint-2
Off
User-2
Off
Calib
Off
Maint-l
Off
Serv-1
Off
Install
Off
Serv-2
Off
6-5
Troubleshooting & Diagnostics
Sorvall® Centrifuges
Following
Screen
Screen 1 identifies faults
1***
Fault> Press DIRGNQSTICS *** [ will appear the cursor indicator lights beneath alarm will sound every minute until (DIAGNOSTICS)
When (DIAGNOSTICS) The fault information
System, Vacuum System); second, specific cause(s) Failure, Vacuum Release Failure). As soon fault description will appear faults have occurred (e.g., Fault two general faults occurred, the first being
Press (ENTER O). When you do, the fault information a specific cause for the fault along with a string of binary numbers. Each" 1" of binary numbers represents a separate cause for the fault. If there is more than one press (ENTER another cause
is an
1,
explanation of how to enter and move around
Faults
and
is
pressed, descriptive fault information will appear
is
arranged
Fault
1 of 2
O)
again and the cursor will move
for
the fault.
gives specific causes
the
display will move from left
in
two groups: first, general faults (e.g., Refrigeration
as
in the
1 of
Door System
display with
1, Fault 1 of 2, etc.). The example below shows that
in
in
the SET display, an alarm will beep, and
is
(DIAGNOSTICS)
an
the Door System.
to
the next" 1" and the display will show
in
the DIAGNOSTIC mode.
for
them. When a fault occurs,
pressed and the fault
for the
indication
in
the display will change
fault (e.g., Compressor
is
pressed, the first general
of
how many general
to
right.
is
cleared.
in the
in
display.
to
the string
The
show
"1",
Continue returned
Fault
1 of
change
to
Fault
1
of 2
to
press (ENTER
to the
show the next general fault;
|Fault| field.
3), when the cursor returns
Door
O)
If
Not
until
there was more than one general fault (i.e., Fault
Locked
all
causes have been displayed
to
the Fault field,
for
example:
Fault Uacuun Systen
2
of 2
Continue screen again includes the first general fault (e.g., Fault
As
displayed; this information will
troubleshoot the problem.
After |»»* not clear when (STOP) to clear
faults were displayed.
to
press (ENTER
O)
until
all
faults have been displayed
NOTE
you
move through
all
faults have been identified
Fault > Press DlRGNOSTics ***| message from the display.
the
message. Call
the
fault screens,
is
pressed, turn
for
be
sure
be
required
and
Service; when you call,
written down, press
the
centrifuge power OFF then back ON
and
the
display will automatically
and the
1 of
2, Door System).
to
make a note
to
help Service personnel
If
let
Service know which
of
each fault
the message does
cursor
SET display
has
1 of 2,
clear
6-6
Sorvall® Centrifuges
Troubleshooting
& Diagnostics
Screen
Screen
perform some maintenance routines, such
2
2 is
used
to
monitor system operation; install
as
assigning Key#'s (procedure
new
changing slow start/slow stop transition speed (procedure
the vacuum pump Table
6-1
describes what each field
to access
the
field.
oil
(page
Table
7-4).
6-1.
is
used
for and
what authorization (Key#)
Screen 2 Fields, DIAGNOSTICS
Field Purpose
Screen
User-1
2
To
User-1 User-2
monitor state
such
as
vacuum level
of
system operation,
and air
Maint-1 Install
temperature.
User-2
To
such vacuum
perform user diagnostic routines,
as
entering Key#s
and
changing
oil.
software;
on
add new
page 6-12),
rotors;
on
and
is
page
changing
required
Authorization Required
None
System Key#
and
6-9),
Maint-1
Instal 1 To
To enter Screen
Position
necessary, then enter
operation (Table
To
the
cursor under
view system accumulators. install software
2,
press (DIAGNOSTICS) until
the
the
6-2) and
press (ENTER
To exit DIAGNOSTICS, press and press (ENTER
Selection
USER-1
#
1
2 3
4
5
6 7
8
9 10 11 12 13
14 15
O-).
Table
Vacuum Level (mbar)
Ctrlr Samp Temp
6-2.
Operation
Samp Temp Offset
Air Temp
Base Temp Jumper Temp Winding Temp
(deg
(deg
C)
(deg
(deg
Measured Rotor Inertia Door Home Microswitch Door X Position
Door Closed Switch Door Lock Sensor URRID lst/2nd URR Y Value
Motor Direction
None
field
and add new
the "2"
you
want
to use and
rotors.
System Key#
indicator light above
press
(on off).
selection number that corresponds
O).
Press
to turn selection
DIAGNOSTICS, Screen
(deg
C)
(deg
C)
C)
C)
C)
(on off) to
Selection
USER-1
turn selection
off,
enter selection number
2,
Operations
#
Operation
(continued)
16
Tachometer Frequency-
17
Slow Transition Speed
18
Vacuum Transition Speed
19
Imbalance Threshold
20
Drive Relay-
21
Rail Enable
22
Rail Setting
23
Rail Volt Sense
24
Brake Enable Brake Setting
25 26
Drive Enable (Ctrlr)
27
Drive Enable (Host)
28
Bridge Enabled (Yes/No)
29
Bridge Chop (Half/Full)
30
Bridge Setting (Amps)
the key is lit.
Enter Key#,
to the
on.
(V)
(%)
if
desired
0
continued
6-7
Troubleshooting & Diagnostics
Sorvair Centrifuges
Selection
USER-1
USER-2
Table
#
6-2.
DIAGNOSTICS,
Operation
(continued)
31
Bridge Current Sense
32
HES Offset
33
Host State
34
NovRAM
35
NovRAM
36
Min
37
Max
38
Raw
39
Min
40
Max
41
Raw Base Temp
42
Max Base Temp
43
Raw Jumper Temp
44
Min Jumper Temp
45
Max Jumper Temp
46
Raw Winding Temp
47
Max Winding Temp
1
System Access Key#
2
Zonal Access Key#
3
Lock Access Key#
4
Program (1-20) Key#
5
Slow Transition Speed
6
Press
7
LCID Display Test
8
Vacuum System Override
9
Remote Baud Rate
10
Clear Faults
(U2) (U4)
Vac
Level (mbar)
Vac
Level (mbar)
Air
Temp
Air
Temp
Air
Temp
On
for
Oil
Battery Battery
(deg C)
(deg C)
(deg C)
(deg C) (deg C)
(deg
(deg C)
(deg C)
(deg C) (deg C)
Change
(A)
OK? OK?
C)
Screen
Selection
MAINT-1
INSTALL
2,
Operations (continued)
#
1
Serial Number (Year)
2
Serial Number (Unit)
3
Hours
4
Hours
5
Number
6
Hours
7
Hours Since
8
Number
9
Cycles Since
10
Drive Revs
11
w2dt (mantissa)
12
w2dt (exponent)
13
Number
14
Successful
15
No Match
16
Failed
17
Loose
18
Mismatched
19
Runs Ending with Faults
20
Last General Faults
21
Prior General Faults
22
Controller ROM Version
23
Host
24
Host ROM
1
Install Rotor
2
Install NovRAM
3
Install ROM
Operation
of
System
on
Drive
of
Drive Cycles
on
Vacuum Pump
Oil
of
Vacuum Cycles
Oil
(*10A6)
of
URR Cycles
URR
URR
URR
Cycles
Lid URR
URR
ROM (U5)
(U8)
(U5)
on
Time
Change
Change
on
Drive
on
Drive
Cycles
Cycles
Cycles
Cycles
Version Version
6-8
Screen
Screen
lists
the
for
use by
Field Purpose
Screen
Maint-2
Calib Seru-1
Serv-2
3
3 is
used
fields
DuPont
3
To
To
To
To
for
in
calibration
Screen
Table
3 and
or
other Service Personnel only.
6-3.
and
advanced troubleshooting
describes their purpose.
All of
and
the fields
Screen 3 Fields, DIAGNOSTICS
Authorization Required
Maint-2 Calib Seru-2
access diagnostic routines available
DuPont personnel only.
calibrate replacement subassemblies.
monitor system status
for
advanced
troubleshooting.
access diagnostic routines
advanced troubleshooting
and
for
repair.
to
repair. Table
in
this screen
No access
Service Personnel Service Personnel
Service Personnel
6-3
are
Sorvall® Centrifuges
Troubleshooting
& Diagnostics
How to Assign or Change a Key#
The centrifuge is shipped with a factory-set System Access Key# of "1234" that will
allow access to most fields that require a Key# (Screen 3 Service Diagnostics cannot be accessed by the System Access Key#).
If desired, you can change the System Access Key# and you can set different Key#s for various features, such as: the control panel Lock, Zonal, and up to twenty different Program Save Key#s. All Key#s must consist of four digits or less.
NOTE
Record all Key#s and keep them in a place where they will not be lost. If
Key#s are lost, call Du Pont for assistance.
System Access Key#
To change the System Access Key#:
1.
Press (DIAGNOSTICS) until the "2" indicator light above the key is on and the display
looks as follows:
User-1 User-2 Maint-1 Install
Off Off Off Off
2.
Press (ENTER O) to move the cursor to 1 User-2
3.
Press (on off). The display will prompt you to enter a Key#:
User-1 User-2 Maint-1 Install
Off Key# Off Off
4.
Input current |s-=isterc Recess Ke*#1 and press (ENTER O) ; display will look as follows:
User-2 <Ouride> Press Q or On to Exit
9 Off
5.
Input 1 in the 1 User-21 field and press (ENTER O); cursor moves to |<Ovride)l.
User-2 (Quride) Systen Recess Key#
1 Off 1234-
CURRENT
ACCESS
KEY#
SYSTEM
6. Press (onoff) to turn | <ovride)[ on, then press (ENTERE»»).
7.
Input the new | systen Recess Ke^#| (four digits or less). Press (ENTER D>).
User-2 <Ovride> Systen Recess Key#
1 On
8. Press (on off) to turn l«3vride)| off and press (ENTER O).
4444-
EXAMPLE NEW
SYSTEM
ACCESS
OF
KEY#
6-9
Troubleshooting & Diagnostics
Sor vail® Centrifuges
9. Record
the
new System Access Key#.
To change other Key#s (i.e., Zonal, Lock, appropriate procedure that follows; System Access Key# procedure.
If
you are not
performing steps
10.
Input
Q.
in 1 User-21 field
going
10 &
to
assign
11 below.
and
or
press (ENTER D=>). The display will change back
User-1 User-2 Maint-1 Install
Off
Off Off Off
11.
Press
Zonal and Lock Access Key#s
Follow Steps 1 through
(PROGRAM)
to
exit
DIAGNOSTICS.
4 of
System Access Key# procedure, then:
NOTE
or
Program),
do not
repeat Steps 1 through
change other Key#s, exit DIAGNOSTICS
go to
Step
1 of the 4 of the
to:
by
1.
To
change
Key# input
2.
Press (ENTER
3.
Press (on off)
4.
Press (ENTER
5.
Press (on off)
6. Press (ENTER
the
Zonal Key#, input
2 in the
| User-21 field. The display will change:
2 in the I user-21 field:
User-2 CQuride) Zonal Recess Key#
2
Off 1234
or
User—2 (Ouride) Lock
3 Off
O); the
to
O) and
cursor will move
turn kovride)
input
the
to
|<ovride>l.
on.
new Key# (four digits
or
less). Press (ENTER
User-2 <0vride) Lock Recess Key#
3
On
to
turn kovride)!
>) to
move cursor
2222
off.
to 1 User-21 field.
or to
change
O).
the
CURRENT ACCESS KEY#"S
EXAMPLE
NEW
LOCK KEY#
Lock
OF
6-10
Sorvall® Centrifuges
NOTE
To assign or change Program Key#s, go to Step 1 of the following procedure,
but do not repeat Steps 1 through 4 of the System Access Key# procedure. If you are not going to assign Program Key#s, do steps 7 and 8 to exit DIAGNOSTICS.
7.
Input £ in 1 User-21 and press (ENTER O). The display will change back to:
Troubleshooting
& Diagnostics
8.
Press
(PROGRAM)
User-1
Off
to
exit
User-2
Off
DIAGNOSTICS.
Maint-1
Off
Install
Off
9. Record new Key#.
Program Key#(s)
A Key# is required to save a program and to make changes to an existing program,
but it is not required to recall a program.
A different Key# can be assigned to each of the 20 possible programs, if desired.
Follow Steps 1 through 4 of the System Access Key# procedure, then:
1.
Input 4 in the 1 User-21 field and press (ENTER O).
User-2 < Task
Pro9ran<l-20) Key#
4 Off
2.
Press (on off) to turn |< Task >1 on; display changes:
User-2 ( Task ) Enter Program Nunber
4 On 9
3.
Press (ENTER
again.
User-2 < Task
4.
Press (ENTER O) and input the Key# (four digits or less) that you want assigned to
the program (Program #01 is in the example above and below).
User-2 < Task > Key# for Progran #91
5.
Press (ENTER o); cursor moves to |< Task )[. Press (on off) to turn [< Task )| off.
To assign or change Key#s for other Programs, repeat steps 2 through 5.
and input the desired program number (1-20). Press (ENTER O)
Key# for Pro9ran #91
4 On
1234
4 On 3333
CURRENT FOR
PROGRAM 1
EXAMPLE NEW
KEY# FOR
PROGRAM
NOTE
KEY#
OF
1
6-11
Troubleshooting
Sorvall® Centrifuges
& Diagnostics
6. Press (ENTER O) to move cursor to | User-21 field. Input Q, and press (ENTER O). The display will change back to:
User-1
Off
7.
Press
8. Record new Key#s.
(PROGRAM) to
exit
User-2
Off
DIAGNOSTICS.
Maint-1
Off
Install
Off
How to Change Slow Start/Slow Stop Transition Speed
The slow start/slow stop transition speed is approximately 800 rpm for all rotors except
the S-20 rotor, which has a transition speed of approximately 300 rpm; and the GS-3 rotor, which has a transition speed of approximately 650 rpm.
Transition speed can be changed to any speed from 100 rpm to 1000 rpm (the selected
speed will be rounded to the nearest number that is divisible by 4). When the transition speed is changed, it will affect the next run only (unless stored in a program); at the end of the run, it will automatically change back to the default speed of 800 rpm (300 rpm for the S-20 or 650 rpm for the GS-3). Anytime you want a run with a transition speed other than the default speed, you must change it prior to the run or store the transition speed as part of a saved program.
Prepare and install the rotor according to the instructions in the rotor manual. Close chamber door and select other desired run parameters, then change transition speed:
1.
Press (DIAGNOSTICS) until the "2" indicator light above the key is on and the display
looks as follows:
User-1
Off
2.
Press (ENTER O) to move the cursor to |user-2|.
3.
Press (on off) The display will prompt you to enter a lKe*# |.
4.
Input the Is^sten Recess Ke*#| and press (ENTER O).
User-2
Off
User-2 <0uride) Press
Q
Off
5. Input §. in the [user-21 field and press (ENTER
Maint-1
Off
Q or On to
Install
Off
Exit
User-2 <0uride) Slow Transition Speed
5 Off 8Q0
6. Press (on off) to turn l«x»ride)l on, then press (ENTER
I
Slow
Transition
Speed
I.
to move cursor to
6-12
Sorvall® Centrifuges Troubleshooting
& Diagnostics
7.
Input desired transition speed (from 100 rpm to 1000 rpm).
CAUTION
Do not set transition speed at the critical speed (±100 rpm) of the rotor (see Table 4-1, Rotor Information, for critical speed). Slow start in the critical speed range can have a detrimental effect on centrifuge components.
8. Press (ENTER O) ; when you do, the cursor will move to 1 <0vride)|. Press (on off) to turn kOvride)! off.
9. Press (ENTER O) and the cursor will move to
10.
Input Q in|user-2land press (ENTER I>). The display will change back to:
|Us.er-21.
User-1 User-2 Maint-1 Install
Off
Off Off Off
11.
Press
12.
Press
(PROGRAM)
(START)
to exit DIAGNOSTICS.
to begin run.
NOTE
If the chamber door is opened after the transition speed has been changed, the transition speed will automatically change back to the default speed.
6-13/6-14
Sorvall® Centrifuges
Maintenance
Chapter 7
MAINTENANCE
This chapter describes routine maintenance procedures that are to be performed by the
user. Any further maintenance should be performed by a DuPont Service Representative or other qualified service personnel.
WARNING
Because the high voltage electrical circuits in this centrifuge can cause severe
electrical shock, untrained personnel must not attempt procedures that are not included in this manual. Failure to comply with this warning can result in personal injury.
If hazardous materials have been processed in the centrifuge, take all necessary precautions when cleaning or servicing the centrifuge to avoid personal exposure to contaminants.
To gain the most in safety, performance, and useful life from your centrifuge, it is necessary to maintain it as described in this chapter. Proper maintenance is a simple and inexpensive way to help prevent major failures. In addition to the maintenance described in this chapter, a Du Pont service contract, which includes a yearly Preventive Maintenance visit, is recommended to keep your centrifuge in top condition.
The following is a suggested maintenance schedule.
Table 7-1. Maintenance Schedule
Maintenance
Clean the rotor chamber
Wash the tapered spindle
Clean cabinet panels Clean the control panel Clean condenser coils Clean air inlet louvers Clean and inspect door seal
Daily or immediately after a spill Once a week Once a month Once a month Check monthly; clean as required Check monthly; clean as required Once a month or whenever vacuum
performance is poor
Frequency
Change vacuum pump oil and mist eliminator filter
Test RUN display lights and control panel indicators
Once a year or when the message
|+**
Tine to Service Uacuuri Punp ***|
appears in the SET display. Once a year or whenever numbers
displayed seem erroneous.
7-1
Maintenance Sorvall® Centrifuges
Cleaning
Rotor Chamber
The rotor chamber should be kept clean and wiped dry routinely to maintain
efficient cooling. Wash the chamber with a mild dishwashing liquid, then rinse and dry with a soft absorbent cloth.
Use 70% ethanol to disinfect the rotor chamber or a 2% glutaraldehyde solution to
sterilize it. For general radioactive decontamination, use a solution of equal parts of ethanol, 10% SDS, and water. Follow this with ethanol rinses then deionized water rinses.
radioactive waste containers.
Tapered Drive Spindle
Wipe the drive spindle with a soft cloth each time a rotor is to be installed to reduce the chance of the rotor sticking. Once a week, wash the drive spindle with warm water.
Dry with a soft absorbent cloth. Dispose of all wash solutions in proper
CAUTION
Chlorides are extremely harmful to aluminum alloy rotors and can cause stress corrosion cracking. Therefore, do not use chlorides to disinfect the chamber.
Cabinet
Clean the Noryl® cabinet panels with a household wax cleaner. Use a mild
noncdkaline
detergent and water to clean the top deck and the chamber door.
Control Panel
To clean the control panel, spray a small amount of a mild cleaning liquid (such as window cleaner) on a clean cloth and wipe the panel clean. DO NOT spray cleaners directly on the control panel.
Door Seal
The chamber door seal should be cleaned once a month. If the seal is not kept clean, it can result in an insufficient vacuum level during a vacuum run.
To clean the seal:
1.
Open the chamber door all the way. Using a #1 phillips (or Reed-Prince)
screwdriver, loosen the quarter-turn fastener (turn counterclockwise) on the
front edge of the door (figure 7-1, view 1).
2.
Close the chamber door halfway, then locate the hole in the center of the tray
assembly, underneath the door (figure 7-1, view 2). Insert your fingers approxi­mately 13mm (1/2 inch) into the hole and pull forward on the silver seal tray until the hole is visible. Continue pulling tray out while pushing back on the door
handle until the door is fully open and the seal tray assembly is separated from
the door. Remove the sealing assembly.
7-2
Sorvall® Centrifuges
3.
Inspect the top and bottom perimeter of the seal for signs of cracks, tears, or
abrasions. If damaged, replace the door seal assembly, Catalog No. 59235.
4.
Wash the centrifuge deck and both sides of the door seal and tray with a mild
noncdkaline laboratory detergent and rinse well.
5.
With the door fully open, carefully engage both corners of the tray in the door
tracks, then slide the seal tray assembly back into the door. Secure it in place
with the quarter-turn fastener.
Do not operate the centrifuge without the seal tray assembly in place. The assembly is part of the centrifuge containment system, which provides
protection in the event of a rotor failure.
1.
Maintenance
WARNING
3.
Figure 7-1. Door Seal Removal
7-3
Maintenance Sorvall® Centrifuges
Refrigeration System Condenser
To maintain the efficiency of the refrigeration system, keep the condenser fins (located
on the left side of the centrifuge) free of dust and dirt. Clean the fins with a brush or
vacuum cleaner at one to three month intervals. The fins will bend easily; be very
careful when cleaning them.
Air Louvers
To allow proper air flow through the centrifuge, the air louvers on the front cabinet panel should be kept free of dust and dirt. Clean the louvers with a brush or a vacuum
cleaner as needed.
Lubrication
All components are prelubricated and require no further lubrication. The refrigeration
unit is hermetically sealed and the ball bearings in the centrifuge motor are
permanently lubricated. Vacuum pump oil must be changed whenever the | »** Tirte to Service Uacuurc
message appears in the SET display or once a year, whichever occurs first.
Pump
*** |
Changing the Vacuum Pump Oil & Mist Eliminator Filter
The most common cause of vacuum pump failure or poor performance is contaminated
oil.
To ensure the efficiency of the vacuum system operation, the vacuum pump oil must be changed whenever the 1 *** Tine to Service Uacuurc Punt* ***! message appears in the SET display or once a year, whichever occurs first.
The | *** Tine to Service Uacuurc Punp *** 1 message will appear in the SET display on
the control panel to warn you that twenty or less vacuum runs* remain before the
vacuum will not operate. The oil should be changed as soon as possible after this
message appears. If the oil is not changed before the twenty vacuum runs are up, the {*** Uacuun Furcp Must Be Serviced »»*! message will appear in the display when you attempt to start the twenty-first vacuum run and an interlock will prevent further
vacuum operation until the oil is changed. A DuPont Oil Change Kit, Catalog No. 12870, is required for this procedure. The kit
includes the items needed to change the oil: a bottle of vacuum pump oil, a mist eliminator filter, a plastic exhaust port blocking tube, a piece of tape, and a funnel.
NOTE
If you are not using the DuPont Oil Change Kit, you must use oil that meets the same specifications as the oil supplied in the kit. The specifications for the oil are given in the Appendix of this manual with a list of additional items required.
* A vacuum run occurs when a Sorvall9 Si/praspeed® rotor is operated above its maximum air speed. Maximum air speed is equal
to maximum speed of the rotor in an RC-5C Centrifuge. See rotor instruction manual.
7-4
Sorvall® Centrifuges Maintenance
Refer
to
Figure
7-2 for
parts location
and
change
the
vacuum pump
oil:
NOTE
It
is
best
to
change
warm
1.
Turn
2.
Peel
over
3.
Turn
of panel
4.
Remove right access panel from behind control console
panel
and
will pump
the
centrifuge power
the
back
off the
the
left-hand inlet port
the two
and
fasteners
out,
then down,
lifting
up.
the oil
out
while
easier.
ON and
piece
of
tape supplied
to
completely seal
at
the
bottom
and
remove panel from centrifuge.
If
necessary, strike
it is
open
of
warm, since
the
chamber door.
in
the Oil
the
the
lower front cabinet panel. Pull bottom
end
of
it is
less viscous when
Change
port.
by
panel lightly
Kit,
then stick
pressing
to
pop
in on
panel
off.
the
tape
end
of
WARNING
Leave chamber door open while access panel avoid
the
and
possibility
the
door track behind
of
pinching your hand
the
console.
is
off.
This
and
fingers between
is
necessary
the
to
door
5.
Loosen knob
6. Remove mist eliminator filter
7.
Install
the eliminator. Reinstall full turns after
8. Remove entire length
on
cap
of
mist eliminator. Then, remove
and
screen. Discard filter
exhaust port blocking tube (supplied
the cap and
you
first feel resistance.
of
vacuum pump drain line from channel
knob assembly
in
and
the kit)
panel.
9. Remove drain line tubing clamp, then securely position bottle having a capacity
of
one
liter
or
more.
NOTE
10.
Make sure the access panel
Turn
the
a.
Press (DIAGNOSTICS) until
appears
the
drain line tubing
and
vacuum pump
in
the
display.
User-1
Off
where
on:
the
User-2
is
not
kinked, particularly near
it
was
crushed
"2"
indicator light above
Maint-1
Off
by
Off
the
tube clamp.
cap and
and
on
tighten
end
of
the key
Install
Off
knob assembly.
keep
the
screen.
the
stem
of
it:
turn
the
in
front access
drain line
in a waste
the top
is on
and 1 User-2
the
knob
of
mist
two
b.
Press
prompt
(ENTER
you
User-1
E>)
to
Off
to
enter
move
the
a Key#:
User-2 Key#
cursor
to User-2
Maint-1
Off
.
Press (on off);
Install
th
Off
7-5
Maintenance
Sorvall® Centrifuges
LEFT-HAND
INLET PORT
KNOB-
MIST ELIMINATOR CAP
MIST ELIMINATOR SCREEN
EXHAUST PORT BLOCKING
TUBE
MIST ELIMINATOR FILTER
STEM OF
MIST
ELIMINATOR
RIGHT ACCESS PANEL
QUARTER-TURN
FASTENERS
7-6
FRONT
PANEL
Figure
7-2. Oil
DRAIN LINE
TUBING CLAMP
DRAIN
LINE
CLAMP
IN
PLACE
Change Parts Identification
RIGHT SIDE INLET PORT PLUG
ONE
LITER
WASTE
BOTTLE
CLAMP REMOVED
Sorvall® Centrifuges
c. Input the System Access Key# and press (ENTER D>).
User-2 (Quride) Press 9 or On to Exit
d. Input £ and press (ENTER O).
User-2 < Task ) Press On for Oil Change
e. Press (on off) and press (ENTER O).
Usei—2 < Task ) Enter 1 to Run PUNP
f. Input 1 and press (ENTER O).
User-2 < Task ) Enter Q Nhen Enpty
Maintenance
9 Off
6 Off
6 On 9
6 On
NOTE
If vacuum pump turns off within 30 seconds after it is first turned on, check to make sure the piece of tape inside chamber completely seals inlet port and that the inlet port plug on left side of centrifuge is firmly in place.
11.
Run the pump until air comes through the drain line and froth begins to appear in
the waste container; this indicates that the pump is empty. Immediately press
(ENTER O) , input Q. and press (ENTER O) again to turn pump OFF and avoid
damaging the pump.
CAUTION
Do not run the pump after it is empty. To do so can damage the pump.
NOTE
The pump will turn off automatically after 4 minutes. If the pump is not
empty after the 4 minutes, press (ENTER O), input 1 and press (ENTER E>)
again to continue running the pump.
12.
Press (onoff) to turn |< Task )| off.
13.
Press (ENTER O) to move the cursor under | User-21 • Input 0 and press (ENTER O») .
14.
Press (PROGRAM) to exit DIAGNOSTICS.
15.
Wipe oil from drain line tubing and clamp.
16.
Position tubing clamp in place on drain line. Close clamp and tighten it securely to
seal the drain line. Store the line back in the front access panel.
17.
Unscrew the knob on the mist eliminator cap and remove cap and knob assembly.
7-7
Maintenance Sorvall® Centrifuges
18.
Remove exhaust port blocking tube from the stem of the mist eliminator. Do not
install mist eliminator filter yet.
19.
Unscrew the nut from the oil fill line fitting (located in front of the mist eliminator)
and remove the fill line plug assembly.
20.
Insert the funnel (supplied in the kit) in the oil fill line fitting. Make sure the funnel
is in as far as it will go, then secure it in place by tightening the locking nut.
21.
Open the bottle of oil and slowly pour entire contents (675 ml) into funnel, being
careful not to overflow the oil in the funnel.
22.
Properly dispose of the collected waste oil.
WARNING
Disposal of the waste oil must conform to all local, state, and federal standards.
23.
Loosen the locking nut on the funnel and remove the funnel.
24.
Reinstall the fill line plug assembly in the oil fill line fitting as far as it will go and
tighten the nut; the nut should be finger tight.
25.
Position the new mist eliminator filter in place, then install the mist eliminator
screen over the filter.
26.
Install mist eliminator cap and knob assembly over the mist eliminator screen,
making sure screen is fully seated inside cap; tighten the assembly in place by turning the knob on the cap.
27.
Reinstall the access panel behind the control console.
28.
Reinstall the front cabinet panel.
29.
Remove piece of tape from inside the chamber. The centrifuge is ready for use.
Emergency Sample Recovery
WARNING
When the main power shuts off, the brake will not operate. Wait until the rotor stops before using the mechanical override. Reaching into the rotor
chamber while the rotor is spinning can cause personal injury.
a. Nonvacuum Runs
Wait until the rotor has coasted to a stop. When the rotor stops, insert a straightened paperclip in the door interlock override hole and push down (figure 7-3). Open chamber door.
b.
Vacuum Runs
If a power failure occurs, vacuum will not release automatically and you will not be able to open the chamber door. The chamber will maintain a partial vacuum and the rotor
will safely coast to a stop. The vacuum must then be released manually.
7-8
Sorvall® Centrifuges
To release vacuum:
Maintenance
STRAIGHTENED PAPER CLIP
DOOR INTERLOCK OVERRIDE HOLE
CHAMBER DOOR
Figure 7-3. Door Interlock Override
1.
Wait until the rotor has coasted to a stop, then turn the power switch OFF.
2.
When the rotor stops, unplug the centrifuge.
3.
Locate the vacuum release solenoid valve (figure 7-4). Insert a screwdriver
through the grille on left side of back panel, and turn vacuum release screw fully
clockwise (approximately 4 to 5 turns). Air will begin to enter chamber.
VACUUM
RELEASE
SOLENOID
VALVE
VACUUM
RELEASE
SCREW
Figure 7-4. Vacuum Release Solenoid Valve Location
7-9
Maintenance
4.
When the air
stop),
5. Insert a straightened paperclip in the door interlock override hole (figure 7-3). Open
turn the vacuum release solenoid valve screw fully counterclockwise.
door.
in
the chamber has reached atmospheric pressure (hissing noise will
So r
vail® Centrifuges
Test
Periodically test the control panel RUN display and indicators to verify that the lights are functional:
1.
2.
of
RUN Display Lights
Turn the centrifuge power ON.
Press (DIAGNOSTICS) until the
looks
as
follows:
"2"
&
Control Panel Indicators
indicator light above the key
is
on and the display
User-1 User-2 Maint-1 Install
Off
Off Off Off
3.
Press (ENTER
4.
Press (on off). The display will prompt you
5. Input the System Access Key# and press (ENTER O).
6. Input
7 in
O) to
User-2 <Qvride> Press
the I User-21 field and press (ENTER t>).
move the cursor
Q
Off
to 1 User-21.
to
enter
9 or On to
a
Exit
User-2 <Ouride) LCID Display Test
7 Off S
7. Press (onoff). When (on off)
• the number
all
• the minus temperature symbol will light
• the cursor indicator lights will If a number other than "8" appears
come on, call for Service.
8. Press (on off).
9. Press (ENTER
indicator lights on the control panel will come on;
is
pressed:
"8"
O) .
will light
in
each space across the RUN display;
all
come on.
in
the display or
in
the RUN display; and
if
any of the other lights do not
7-10
Sorvall® Centrifuges Maintenance
10.
Input 0 under | User-21 and press
(ENTER
O). The display will change back to:
User-1 User-2 Maini-1 Install
Off
Off Off Off
11.
Press (PROGRAM) to exit DIAGNOSTICS.
Parts Ordering Information
To order replacement parts: in the United States, telephone toll-free (800) 551-2121. Outside the United States, contact local distributor or agent. Be sure to provide a description of the part, centrifuge model, and centrifuge serial number.
Service Decontamination Policy
WARNING
Because of the characteristics of the samples likely to be processed, biological or radioactive contamination may occur. Always be aware of this possibility and take normal precautions. Use appropriate decontamination procedures should exposure occur.
If a centrifuge or rotor that has been used with radioactive or pathogenic material requires servicing by DuPont personnel, either at the customer's laboratory or at a DuPont facility, comply with the following procedure to ensure the safety of all personnel:
• Clean the centrifuge and/or rotor to be serviced of all encrusted material and decontaminate it prior to servicing by the DuPont representative. There must be
no radioactivity detectable by survey equipment.
• Complete and attach Decontamination Information Certificate (Sorvall® Products
Form No. IPDP-59 or E53603) to the centrifuge or rotor.
If a centrifuge or rotor to be serviced does not have a Decontamination Information Certificate attached and, in DuPont's opinion presents a potential radioactive or biological hazard, the DuPont representative will not service the equipment until proper decontamination and certification is complete. If Du Pont receives a centrifuge or rotor at its Service facilities which, in its opinion, is a radioactive or biological hazard, the sender will be contacted for instructions as to the dispositon of equipment. Disposition costs will be borne by the sender.
Decontamination Information Certificates are included with these instructions.
Additional certificates are available from the local Technical or Service Representative.
In the event these certificates are not available, a written statement certifying that the
7-11
Maintenance Sorvall® Centrifuges
unit has been properly decontaminated and outlining the procedures used will be acceptable.
NOTE
The Service Representative will note on a Customer Service Repair Report if
decontamination was required and, if so, what the contaminant was and what
procedure was used. If no decontamination was required, it will be so stated.
7-12
Vacuum Pump Oil Specifications
If you are not using DuPont Oil Change Kit (Cat. No. 12870) when changing the vacuum pump oil, you must use oil that meets the specifications below.
Oil Specifications:
Molecular distilled general purpose hydrocarbon mechanical pump fluid,
675 ml.
Ultimate pressure @ 25°C: Viscosity <g> 38°C: Specific Gravity @ 22°C: Flash Point: Fire Point: Auto Ignition:
These properties are typical of Cambridge Mill Products #VB-X.
Also Required:
Mist Eliminator Filter, Technolab #ED5-IL Plastic Tape, minimum width 50 mm (2 inch); 3M #471 or equivalent
6x105Torr 375 SUS 80 CST
0.88 251 °C 279°C
449°C
Warranty
WARRANTY
E. I. du Pont de Nemours and Company makes no warranty of any kind, expressed or implied, except as stated in this warranty policy.
The Sorvall® RC-28S Supraspeed® Centrifuge is warranted to be free from defects in material and workmanship for a period of one year from the date of delivery. Du Pont will repair or replace and return free of charge any part which is returned to its factory within said period, transportation prepaid by user, and which is found upon inspection to have been defective in materials or workmanship. This
warranty does not apply to any damage to any instrument resulting from: normal
wear and tear; misuse; abuse; use of electrical currents or circuits other than those specified on the plate affixed to the instrument; accident; negligence; failure to follow operating instructions; or use of any rotor other than a Sorvall® rotor intended for use in this instrument.
Du Pont reserves the right to change, alter, modify or improve any of its instruments without any obligation whatsoever to make corresponding changes to any instrument previously sold or shipped.
THE FOREGOING OBLIGATIONS ARE IN LIEU OF ALL OTHER OBLIGATIONS AND LIABILITIES INCLUDING NEGLIGENCE AND ALL WARRANTIES, OF MERCHANT-
ABILITY OR OTHERWISE, EXPRESSED OR IMPLIED IN FACT OR BY LAW, AND
STATE OUR ENTIRE AND EXCLUSIVE LIABILITY AND BUYER'S EXCLUSIVE REMEDY FOR ANY CLAIM OR DAMAGES IN CONNECTION WITH THE SALE OR FURNISHING OF GOODS OR PARTS, THEIR DESIGN, SUITABILITY FOR USE,
INSTALLATION OR OPERATION. DU PONT WILL IN NO EVENT BE LIABLE FOR
ANY SPECIAL OR CONSEQUENTIAL DAMAGES WHATSOEVER, AND OUR
LIABILITY UNDER NO CIRCUMSTANCES WILL EXCEED THE CONTRACT PRICE FOR THE GOODS FOR WHICH LIABILITY IS CLAIMED.
Glossary
Cursor
Default
Value
Display
Message
Enter
Entry
Fault Fault
Message
Field
Input
A movable indicator in the SET display screens that shows
position for data entry.
The choice made by the computer system when no explicit
choice is made by the operator.
A message that appears in the display that prompts the operator
to perform a specific operation or informs the operator of an
operating condition.
To transfer an input to the computer by pressing either the
right (ENTER O) or the left (ENTER <Q cursor key. An input that has been transferred to the computer. A centrifuge malfunction.
Message produced by the computer program that indicates a fault
has occurred.
One item of information within a screen; cursor must be positioned in field to make an entry.
The data to be transferred to the computer by pressing the
keypad.
Integrator
K Factor
Key Number
RCF
Run Display
Screen
Set Display
Keeps track of the accumulating centrifugal effect during a run, including acceleration and deceleration.
A value used to estimate run time at a specific speed to move
particles in suspension (assumes at 20°C) from the rmin of the
rotor to rmax of the rotor. The lower the K Factor, the more efficient the rotor is in sedimenting particles.
A security number that provides user authorization to access
protected data.
(g force) The force applied to a suspension of particles during centrifugation relative to the force that the earth's gravity would have on them.
The upper display on the control panel which shows actual time, Jo^dt, temperature, and speed values.
A group of fields visible in the SET display selected by pressing
either the PROGRAM, CALCULATE, or DIAGNOSTICS key.
Hie lower display on the control panel that changes according to
the screen selected.
Transition
Speed
The speed at which the rotor changes from slow acceleration or
deceleration to maximum acceleration or deceleration when Slow Start or Slow Stop is used.
Index
+/- Key, 3-3
Brake,
Cabinet, Cleaning, 7-2
CALCULATE,
Calculator, 1-4, 3-9, Chamber, Rotor, 1-4 Cleaning,
CLEAR Key, 3-3 Compartment Loads, 4-12 Control Panel, Control Panel, Display & Indicator Test, 7-10 Control Panel Lock, 1-3, 3-8, 4-11 Decontamination Policy, 7-11 Diagnostics, 1-4, DIAGNOSTICS Key, 3-9,
Dimensions, Centrifuge, 1-2, 2-1
Display
Door Interlock Override, 7-8 Drive System, 1-3, 1-5 Electrical Requirements, 1-2, 2-2
Emergency Sample Recovery, 7-8 ENTER Keys, 3-3 Faults, Filter, Mist Eliminator, 7-4 HOLD Key, 3-3 Hold Run, 4-8 Hr:Min Field, 3-4 Indicators, Test, 7-10 Inspection, 2-1
Installation, 2-1 Jco2dt Calculation, 5-1 Jco2dtRun, 1-3, 3-7, 4-9 Integrator, 1-3, 3-7-3-8, 4-9 K Factor, Calculation, 5-2 Key Number, 1-4, 4-6, 6-10 Keys
Keypad, 3-3 Leveling, 2-2 Location, Centrifuge, 2-1 Lock, Control Panel, 1-3, 3-8, 4-11
3-5
Display Screens, 3-9, 5-1
Key, 3-9,
Air Inlet Louvers, 7-4
Cabinet, 7-2
Control Panel, 7-2
Condenser Coils, 7-4 Door Seal, 7-2 Drive Spindle, 7-2 Rotor Chamber, 7-2
Messages, Run, 3-1
Screens, CALCULATE, 3-9, 5-1 Screens, PROGRAM,
Set, 3-1 Test, 7-10
CALCULATE, 3-9,
CLEAR, 3-3 DIAGNOSTICS, 3-9, PROGRAM, HOLD, START, 3-3 STOP,
5-1-5-2
5-1-5-2
7-1-7-3
3-1-3-3
6-5-6-9
6-5-6-9
6-1-6-5
3-3-3-9
6-6
5-1-5-2
6-5-6-9
3-3-3-9
3-3
3-3
Lubrication, 7-4 Maintenance, 7-1 Mass,
Centrifuge, 1-2
Normal Run, 4-1
Oil,
Vacuum Pump, 7-4 ON/OFF Power, 4-1 On/Off Key, 3-3
Operator's Guide, 1-4 Ordering Information, 7-11 Override, Door Interlock, 7-8 Overtemperature, 1-2, 3-5 Parts,
Ordering, 7-11 Power, ON/OFF, 4-1 Precool/Preheat, 4-12 PROGRAM Key, Program Recall, 3-8 Program Save, 3-9
Programmed Run, 1-4, 3-8, 4-10
RCF
Calculation, Field, 3-6 Run, 1-3, 3-6, 4-5 Set, 3-6
Recall, Program, 3-8 Refrigeration System, 1-3, 1-5 Rotor
Field, 3-4 Information, 4-4 Management System, 1-2, 1-3, 1-5, 3-4
Run Display, 3-1 Safety Information, ill Sample Recovery, 7-8 Save,
Program, 3-9 Screens, Display. See Display Service, Decontamination Policy, 7-11 Set Display, 3-1
Slow Start/Slow Stop, 1-3, 3-6-3-7, 4-3 Speed
Calculation,
Control, 1-2 Field, 3-4 Reducing, 4-12 Set, 1-2, 3-4
Spin-Right® Rotor Management
System, 1-2,
START Key, 3-3
STOP Key, 3-3 Temp/Max Field, 3-5 Temperature, Ambient, 1-2, 2-2 Temperature, Control, 1-3 Temperature, Maximum, 3-5 Temperature, Set, 1-2, 3-5 Time,
Calculation, 5-1
Time,
Display, 1-2, 3-4
Time,
Set, 1-2, 3-4
Transition Speed, 3-6-3-7, 6-13
Vacuum, 1-2, 1-3, 1-6 Vacuum, Bleed, 7-8 Vacuum Pump Oil, 7-4 Warranty. See Appendix Zonal Field, 3-7 Zonal Run, 1-3, 3-7, 4-7
3-3-3-9
5-1-5-2
5-1-5-2
1-3-1-5,
3-4
PROGRAM
LOG
PROGRAM
LOG
PROGRAM
LOG
PROGRAM
LOG
.
Rate.
Rate.
Date.
Speed
Temp_
not
turn ZONAL
LOG
Date.
_
PROGRAM
Operator
Rotor
Hr
Brake
RCF. Slow Start
Slow Stop
Notes
'
PROGRAM RECALL will
but
#.
_Min_
DoFF Zonal* Speed
it
will set speed & other parameters. but
PROGRAM
PROGRAM
Operator
*.
PROGRAM
Operator
Rotor
Hr
Brake
RCF.
Slow Start
SlowStopj^pRate.
Notes
On, * PROGRAM RECALL will
#.
Min
Zonal*Speed.
Rate.
it
will set speed & other parameters. but
PROGRAM
PROGRAM*
Operator
Date.
Speed_
Temp_
not
turn ZONAL
LOG
Date.
PROGRAM*
Operator
Rotor
°C
Hr
On, * PROGRAM RECALL will
.Min. ION
Brake
y^F Zonal* Speed
RCF Jco2cfr Slow Start R2£l Rate
Slow Stop
Notes
it
will set speed & other parameters. but
Rate
PROGRAM
PROGRAM*
Operator
Date.
Speed
Temp_
not
Date.
turn ZONAL
LOG
PROGRAM#.
Operator
Rotor
Hr
Min
Brake
RCF.
Slow Start
Slow Stop
Notes
On, * PROGRAM RECALL will
it
will set speed & other parameters.
Zonal* Speed
Rate.
PROGRAM
PROGRAM*
Operator
Date.
Speed
Temp_
not
Date.
turn ZONAL
LOG
On,
Rotor
Hr
Brake
Min
Speed
Temp
Zonal* Speed
RCF
SlowStartj^pRate.
SlowStop0°^FRate.
Notes
1
PROGRAM RECALL will
but
it
will set speed & other parameters. but
PROGRAM
PROGRAM
Operator
Rotor
Hr.
Brake
Min
HoFF
*
Zonal*
not
Date.
Speed
Temp__
Speed
turn ZONAL
LOG
_
Rotor
Hr
Min.
Brake
RCF_
Slow Start
Slow Stop
Notes
On, * PROGRAM RECALL will
it
will set speed & other parameters. but
Zonal*
_ Wdt
^
Rate.
Rate.
PROGRAM
PROGRAM*.
Operator
Rotor
_°C Hr.
Min
Brake
DoFF Zonal*Speed.
Speed
Temp_
Speed
not
turn ZONAL
LOG
Date.
Speed,
Temp_
_
Rotor
Hr
Brake
RCF_
Slow Start
SlowStopgg^pRate.
Notes.
On, * PROGRAM RECALL will
PROGRAM*.
Operator
Rotor
Hr
Brake
Speed
Min.
Temp.
Zonal*Speed.
Rate
g^p
not
it
will set speed & other parameters. but
PROGRAM
turn ZONAL
LOG
Date.
Speed
Min
0gFF Zonal*
Temp_
Speed
On, * PROGRAM RECALL will
_
Rotor
Hr
Min.
Brake
DoFF Zonal* Speed
RCF
Slow Start
Slow Stop
Notes
it
will set speed & other parameters.
Rate.
Rate
PROGRAM
PROGRAM*.
Operator
Rotor
Hr
Min
Brake
Zonal*Speed
Speed,
Temp_
not
turn ZONAL
LOG
Date.
Speed_
Temp_
On,
°C
RCF Slow Start
Slow Stop
Notes
PROGRAM RECALL will but
it
will set speed & other parameters. but
Rate
not
turn ZONAL
RCF_ Slow Start Qg
SlowStop^pRate
Notes
On, * PROGRAM RECALL will
it
will set speed & other parameters. but
^p
FJF
Rate
not
turn ZONAL
RCF.
Slow Start
Slow Stop
Notes
On, * PROGRAM RECALL will
it
will set speed & other parameters. but
Rate
not
turn ZONAL
RCF_
SlowStarty°^FRate.
SlowStop^pRate.
Notes
On, * PROGRAM RECALL will
it
will set speed & other parameters.
not
turn ZONAL
On,
DECONTAMINATION CERTIFICATE
Instructions on the reverse of this card must be completed
before field service
to the Du Pont/Sorvall service facility.
NAME DEPARTMENT INSTITUTION __ ADDRESS
CITY STATE ZIP.
INSTRUMENT SERIAL NUMBER
ROTOR - .-. SERIAL NUMBER PART PART NUMBER CONTAMINATE USED _ DECONTAMINATION CERTIFIED BY _
DATE DECONTAMINATED.
or the
return
of the
instrument
or
part
DECONTAMINATION CERTIFICATE
Instructions on the reverse of this card must be completed before field service or the return of the instrument or part to the Du Pont/Sorvall service facility.
NAME DEPARTMENT INSTITUTION __„_ ADDRESS
CITY ________ STATE ZIP.
INSTRUMENT
ROTOR. SERIAL NUMBER PART PART NUMBER CONTAMINATE USED. DECONTAMINATION CERTIFIED BY. DATE DECONTAMINATED
„„_-_____
SERIAL NUMBER
DECONTAMINATION CERTIFICATE
Instructions on the reverse of this card must be completed before field service or the return of the instrument or part to the Du Pont/Sorvall service facility.
NAME DEPARTMENT INSTITUTION ADDRESS
CITY STATE ZIP. INSTRUMENT SERIAL NUMBER ROTOR SERIAL NUMBER
PART PART NUMBER CONTAMINATE USED DECONTAMINATION CERTIFIED BY DATE DECONTAMINATED
INSTRUCTIONS
When an instrument that has been used with radioactive or pathogenic
material requires servicing by Du Pont personnel either at the customer's laboratory or at Du Pont facilities, the following procedure must be
complied with to insure safety of our personnel:
1.
The instrument or part to be serviced shall be cleaned of all blood
and other encrusted material and decontaminated prior to servicing by our representative. No radioactivity shall be detectable by survey equipment.
2.
A Decontamination Information Certificate (Sorvall Instruments
Form No. IPDP-59) shall be completed and attached to the instrument
or part
If an instrument or part to be serviced does not have a Decontamination
Information Certificate attached to it, and, in our opinion, the instrument or part presents a potential radioactive or biological hazard, our representa­tive will not service the equipment until proper decontamination and
INSTRUCTIONS
When an instrument that has been used with radioactive or pathogenic material requires servicing by Du Pont personnel either at the customer's laboratory or at Du Pont facilities, the following procedure must be complied with to insure safety of our personnel:
1 The instrument or part to be serviced shall be cleaned of all blood
and other encrusted material and decontaminated prior to servicing by our representative. No radioactivity shall be detectable by survey equipment.
2.
A Decontamination Information Certificate (Sorvall Instruments
Form No. IPDP-59) shall be completed and attached to the instrument or part.
If an instrument or part to be serviced does not have a Decontamination Information Certificate attached to it. and, in our opinion, the instrument or
part presents a potential radioactive or biological hazard, our representa­tive will not service the equipment until proper decontamination and
certification has been completed. If an instrument is received at our Service facilities and, in our opinion, is a radioactive or biological hazard, the sender will be contacted for instructions as to disposition of the equipment. Disposition costs will be borne by the sender.
Decontamination Information Certificates are included with these Oper­ation Instructions. Additional certificates are available from your local technical or customer service representative. In the event these certifi­cates are not available, a written statement certifying that the instrument or part has been properly decontaminated and outlining the procedures used will be acceptable.
NOTE
Service representatives will indicate on a Customer Service
Repair Report if decontamination was required, and if so, what the contaminate was and what procedure was used. If no decontamination was required, it should be so stated.
certification has been completed. If an instrument is received at our
Service facilities and, in our opinion, is a radioactive or biological hazard, the sender will be contacted for instructions as to disposition of the equipment. Disposition costs will be borne by the sender.
Decontamination Information Certificates are included with these Oper­ation Instructions Additional certificates are available from your local technical or customer service representative. In the event these certifi­cates are not available, a written statement certifying that the instrument or part has been properly decontaminated and outlining the procedures used will be acceptable
NOTE
Service representatives will indicate on a Customer Service
Repair Report if decontamination was required, and if so, what the contaminate was and what procedure was used. If no decontamination was required, it should be so stated.
INSTRUCTIONS
When an instrument that has been used with radioactive or pathogenic material requires servicing by Du Pont personnel either at the customer's laboratory or at Du Pont facilities, the following procedure must be complied with to insure safety of our personnel:
1.
The instrument or part to be serviced shall be cleaned of all blood
and other encrusted material and decontaminated prior to servicing by our representative. No radioactivity shall be detectable by survey equipment.
2.
A Decontamination Information Certificate (Sorvall Instruments
Form No. IPDP-59) shall be completed and attached to the instrument
or part.
If an instrument or part to be serviced does not have a Decontamination Information Certificate attached to it, and, in our opinion, the instrument or part presents a potential radioactive or biological hazard, our representa-
tive will not service the equipment until proper decontamination and
certification has been completed. If an instrument is received at our
Service facilities and, in our opinion, is a radioactive or biological hazard,
the sender will be contacted for instructions as to disposition of the
equipment. Disposition costs will be borne by the sender.
Decontamination Information Certificates are included with these Oper­ation Instructions. Additional certificates are available from your local technical or customer service representative. In the event these certifi­cates are not available, a written statement certifying that the instrument or part has been properly decontaminated and outlining the procedures used will be acceptable.
NOTE
Service representatives will indicate on a Customer Service Repair Report if decontamination was required, and if so, what the contaminate was and what procedure was used. If no decontamination was required, it should be so stated.
IPDP-59
Loading...