Kendro Laboratory Products
Asheville, North Carolina
U.S.A.
PN 04505-32
Issued January 2005
Page 4
Table of ContentsSORVALL® Centrifuges
!
This manual is a guide to the use of the
SORVALL® Cellwasher 2 Cell Washing System
Data herein has been verified and is believed adequate for the intended use of the Cellwasher 2.
Because failure to follow the recommendations set forth in this manual could produce personal
injury or property damage, always follow the recommendations set forth herein. Kendro does not
guarantee results and assumes no obligation for the performance of centrifuges or other products
that are not used in accordance with the instructions provided. This publication is not a license to
operate under, nor a recommendation to infringe upon, any process patents.
The SORVALL Cellwasher 2 is intended to be used in strict accordance with The American
Association of Blood Banks' guidelines defined in the most recent edition of the AABB Technical
Manual (published by AABB Press. Bethesda, MD).
Publications prior to the Issue Date of this manual may contain data in apparent conflict with that
provided herein. Please consider all data in this manual to be the most current.
WARNING, CAUTION
important and critical instructions.
WARNING
personal injury, affect the operator’s health, or contaminate the environment.
informs the operator of a hazard or an unsafe practice that could result in
informs the operator of an unsafe practice that could result in damage of
WARNING
, and
NOTE
are accompanied by a hazard symbol
within the text of this manual are used to emphasize
and appear in the left
ii
Page 5
Cellwasher 2Table of Contents
!
!
Important Safety Information
Important Safety Information
Certain potentially dangerous conditions are inherent to the use of all centrifuges. To ensure
safe operation of this centrifuge, anyone using it should be aware of all safe practices and take
all precautions described below and throughout this manual.
WARNING
When using radioactive, toxic, or pathogenic materials, be aware of all characteristics of the
materials and the hazards associated with them in the event leakage occurs during cell washing. If
leakage does occur, neither the Cellwasher 2 nor the rotor can protect you from particles dispersed
in the air. To protect yourself, we recommend additional precautions be taken to prevent exposure
to these materials, e.g., use of controlled ventilation or isolation areas.
Always be aware of the possibility of contamination when using radioactive, toxic, or pathogenic
materials. Take all necessary precautions and use appropriate decontamination procedures if
exposure occurs.
Never use any material capable of producing flammable or explosive vapors, or creating extreme
exothermic reactions.
Use only tubes that meet the criteria provided in this manual (see Tube Selection and Use, Chapter
3). Results may be affected if tubes are not within specifications. If tubes are of inadequate strength
so that tube breakage occurs, sample and test results will be lost, the rotor and distributor could be
damaged, and during required cleanup, contact with sharp glass fragments could result in personal
injury.
CAUTION
Do not operate Cellwasher 2 without a rotor properly installed on the drive shaft. See Rotor and
Distributor Installation, Chapter 2.
Do not operate the Cellwasher 2 with a rotor out of balance. To do so can cause damage to the
Cellwasher 2 drive assembly. See Rotor Loading and Balancing, Chapter 2.
We recommend using Coomb's control cells to confirm all negative results.
The Cellwasher 2 can be damaged if connected to the wrong voltage. Check the voltage before
plugging the instrument into a power source. Kendro is not responsible for incorrect installation.
Always maintain the Cellwasher 2 in the recommended manner. See Chapter 4, Maintenance.
This manual contains information required to operate and maintain the SORVALL® Cellwasher 2 Cell
Washing System. If you require additional information regarding operation or maintenance, please contact
Kendro for assistance. In the United States, call Kendro toll-free 1-800-522-7746; outside the United
States, contact the nearest Kendro office (see back cover) or your local representative for SORVALL
products. SORVALL
or http://www.kendro.de.
®
product information is available on our internet web site at http://www.kendro.com
Cellwasher 2 Description
The Cellwasher 2 is designed to perform the washing phase of the
Coomb's Procedure automatically. In normal AUTO mode operation, after placing two or three drops of a pre-prepared 2 to 5% red
blood cell/saline suspension* into each tube, the Cellwasher 2 will
automatically add saline, spin at high speed to concentrate the cells,
decant the saline, and agitate to resuspend the washed cells for the
addition of Coomb's reagent. The Cellwasher 2 is also capable of
being used manually as a low speed centrifuge, spinning at either of
two preset speed ranges (LOW speed or HIGH speed mode, see
Cellwasher 2 Specifications on next page).
®
The Cellwasher 2 combines a peristaltic saline pump and a wet
chamber centrifuge in one cabinet. The centrifuge portion uses a
dedicated DA-12 Rotor, which is a unique dual-angle swingingbucket rotor that is designed to accommodate up to 12 each of
either 3 ml or 5 ml tubes. The Cellwasher 2 is microprocessorcontrolled, featuring a front control panel with sensor-touch keys
and a digital timer for the wash cycle. Other features include a
saline-detect system with an audible low-saline warning, and an
agitate cycle that ensures complete resuspension of cells and
eliminates manual agitation.
Applications Infor mation
The Cellwasher 2 is intended to be used in routine blood work to
speed up and simplify the procedures for determining human blood
groups and types. Often, the results of blood tests made using the
Cellwasher 2 determine donor/patient compatibility for blood transfusions, and the results of these tests must be accurate to prevent
serious transfusion reactions. If testing procedures provided by the
reagent manufacturers are not followed exactly, or if test materials
are outdated, accurate test results cannot be guaranteed.
* The Cellwasher 2 is not designed to automatically separate donor whole blood to create
the necessary 2 to 5% red blood cell/saline suspension. The Cellwasher 2 may, however,
be used as a centrifuge to manually separate red blood cells from whole blood.
1-1
Page 10
DescriptionSORVALL® Centrifuges
!
!
False Coomb's readings can be caused by any of the following:
•Improper ratio of patient serum and cells.
•Incomplete washing of cells.
•Contaminated saline or supply tubing.
•Improper decanting of saline solutions.
•Inactive Coomb's serum.
•Failure to add Coomb's serum to a tube.
•Improper final centrifugation.
•Inaccurate reading or recording of results.
W A R N I N G
We recommend using
Coomb's control cells to confirm all
negative results.
W A R N I N G
The SORVALL Cellwasher 2
is intended to be used in strict accordance with The American Association
of Blood Banks' guidelines defined in
the most recent edition of the AABB
Technical Manual (published by AABB
Press. Bethesda, MD).
While the Cellwasher 2 provides accurate washing, decanting, and a
proper speed for the final spin, it does not eliminate all potential
errors. Proper laboratory procedures require adequate controls of all
tests performed with this instrument (read the WARNINGS). Refer to
Tables 4-5, 4-6, and 4-7 for applications troubleshooting information. When using the Cellwasher 2, always follow standard laboratory procedures in handling and disposing of blood serum, and
observe safety precautions and good laboratory practices.
Tubes must be able to withstand up to 1365 g (948 g in 50 Hz
operation) when run at a 45° angle
with tube walls point-loaded and otherwise unsupported. Using tubes of inadequate strength could result in tube
breakage: sample and test results will
be lost, the rotor and distributor could
be damaged, and cleanup will be required. During cleanup, be aware that
glass fragments are sharp, and contact could result in personal injury.
* At the start of a Manual run, the timer function is delayed for several seconds before
beginning, so that timing occurs only when the rotor is spinning; the timer is inactive
during the initial aggitation step that occurs before the spin starts.
** The maximum noise output with the instrument running at HIGH speed, measured 3 feet
from the front panel at an approximated operator's height.
*** Electrical requirements are indicated on the rating plate on the back of the instrument.
**** A set of Adapter Clips (Catalog No. 04330, 12/pkg., supplied) is required when using
3 ml, 10 mm diameter tubes in the DA-12 Rotor.
1-3
Page 12
InstallationSORVALL® Centrifuges
Accessories
The accessories listed in Table 1-1 are supplied with all Cellwasher 2
models.
Table 1-1. Accessories Supplied
Catalog
NumberDescription
43260Distributor Assembly – for 12 mm x 75 mm and
10 mm x 75 mm tubes
12977Tubing Kit containing:
1-255 mm (10 inch) long frosted Pump Tubing
1-1650 mm (65 inch) long, large diameter (5/16 inch i.d.,
7/16 inch o.d.) clear Tygon
1-Overflow Tubing 150 mm (6 inch) long
1-Discharge Tubing 200 mm (8 inch) long
1-Drain Tubing (cut desired length from remainder;
1220 mm [48 inch] supplied on new product)
1-2900 mm (114 inch) long, small diameter (1/4 inch i.d.,
3/8 inch o.d.) clear Tygon
1-Flow Tubing 305 mm (12 inch) long
1-Vent Tubing 150 mm (6 inch) long
1-Supply Tubing (cut desired length from remainder;
610 mm [24 inch] supplied on new product)
1-Pump Connector
2-Y-Connectors
2-Adjustable Tubing Clamps
2-Nylon Cable Ties
Installation Instructions
®
Tubing from which to cut:
®
Tubing from which to cut:
1-4
12925Flow Control Valve
04285DA-12 Rotor (includes a set of 12 Adapter Clips for use
with 3 ml 10 mm x 75 mm tubes, Catalog No. 04330)
12796Rotating Bowl Assembly
SORVALL
®
5 ml, 12 mm x 75 mm Pyrex® tubes (Catalog Number
03102, 50/box), and SORVALL® 3 ml, 10 mm x 75 mm Pyrex® tubes
(Catalog Number 03100, 50/box) are not supplied, but are recommended for use in the Cellwasher 2.
Page 13
Cellwasher 2Installation
!
!
Chapter 2: INSTALLATION
This chapter contains the information needed to unpack and install the SORVALL® Cellwasher 2.
Unpacking
As soon as you receive your Cellwasher 2, carefully inspect it for
any shipping damage that may have occurred. If you find any
damage, please report it immediately to the transportation company
and file a damage claim, then notify Kendro. Check all packages
received against the shipping list and, after unpacking, check the
accessories received against Table 1-1 on page 1-4; if any items are
missing, contact Kendro (see back cover).
C A U T I O N
Cooling air is drawn into the
Cellwasher 2 from all four sides of
the base. Be sure that the entrances
are not obstructed.
W A R N I N G
During operation, never
lean on or move the instrument, keep
the defined clearance area clear of
all objects (including all hazardous
and flammable materials), and do not
work within the clearance area.
Location
Place the Cellwasher 2 on a sturdy, level bench or table near a sink,
drain, or waste container that can receive the decanted saline. The
following factors should be considered when selecting a location:
•Allow a 150 mm (6 inch) clearance area behind the instrument
for tubing and a waste container. For safe operation, maintain a
50 mm (2 inch) "clearance envelope" on all sides.
Personnel should know that the Cellwasher 2 routinely deals
with significant energy levels and could move suddenly in the
unlikely event of a rotor failure. Laboratory management procedures should require that no person or any hazardous materials
are within the clearance envelope while the Cellwasher 2 is
operating. During operation, personnel should be instructed not
to lean on or move the Cellwasher 2, not to stay within the
clearance envelope longer than necessary for operational reasons, and not to deposit potentially hazardous materials within
the clearance envelope.
•Allow a 610 mm (24 inch) clearance above the tabletop or
bench surface for lid to open.
•Drainage in the Cellwasher 2 is accomplished by gravity, so
the discharge tubing must extend downward from the instrument to drain, sink, or waste receptacle.
•The preferred location for the saline supply is either at or
above instrument level.
•To obtain best results, minimize tubing length from the
saline supply to the Cellwasher 2 and from the Cellwasher 2
to the drain, sink, or waste receptacle.
2-1
Page 14
InstallationSORVALL® Centrifuges
!
Environmental Conditions
An ambient temperature range of 10°C to 38°C (50°F to 100°F)
should be maintained for Cellwasher 2 operation. The relative humidity can be up to 90%, non-condensing.
The Cellwasher 2 is intended for use in 1) a Pollution Degree 2
Environment, 2) an installation category II supply circuit, and 3) at a
maximum altitude of 2000 meters. The Cellwasher 2 is a Class A
product and is not intended for home use. If used in a domestic
environment, it may cause radio interference, in which case the user
may be required to take adequate measures.
Electrical Requirements
W A R N I N G
The receptacle used should
be a grounded 3-wire system. If it is
not, the equipment must be grounded
to reduce the possibility of electrical
shock.
The Cellwasher 2 has a 3-wire (2-pole and earth) power cord with a
universal keyed plug that inserts into a receptacle at the back of the
instrument. The other end must be connected to the appropriate
power supply as specified below and on the rating plate on the back
of the instrument.
120V ±10%, 60 Hz, 5A; 100V ±10%, 60 Hz, 5A;
100V ±10%, 50 Hz, 5A; or 230V ±10%, 50 Hz, 2.5A
Cellwasher 2 120V and 100V power cords use a NEMA 5-15P
three-prong molded plug with a ground pin and parallel blades that
will fit Hubbell receptacle No. 5261 for 120 volts. 230V power cords
are CE-rated, and are supplied with a 3-pin plug that fits a 16A CEE17 single-phase receptacle (2-pole and earth).
The ON/OFF main power switch has a 3-Amp fuse; however, for
emergency disconnect purposes, we recommend a separate means of
power interruption in a remote location.
Tubing Installation
NOTE
The Cellwasher 2 comes with the tubing already connected to the
instrument, but not installed in the peristaltic pump (do not open
the tubing replacement kit, save it for future use). To complete
the tubing installation, unwrap the bundle of tubing at the back of
the instrument and, referring to figure 2-1 and table 2-1, install
the tubing according to the procedure that follows.
2-2
The overflow tubing serves as an emergency liquid exit
to protect instrument components in the event of tube
breakage or drain blockage. To ensure that potentially
biohazardous liquid does not exit into the work surface,
route overflow tubing to a 250 ml (or larger) open waste
container.
Page 15
Cellwasher 2Installation
4
5
FLOW
CONTROL
10
VALVE
SALINE
2
6
1
3
9
PUMP HOUSING
THUMBSCREW
9
108
7
WASTE
Figure 2-1. Cellwasher 2 Tubing Diagram
Table 2-1. Parts Location: Tubing Diagram
(keyed to figure 2-1)
ItemDescription
1*Overflow Tubing (from overflow tray), 150 mm (6 inch)
2Vent Tubing, 150 mm (6 inch)
3Pump Tubing, 255 mm (10 inch), frosted
4Flow Tubing (to nozzle), 305 mm (12 inch)
5Discharge Tubing (from collector ring), 200 mm (8 inch)
6Supply Tubing, 610 mm (24 inch),
7Drain Tubing 1220 mm (48 inch),
8Pump Connector
9Y-Connector
10Adjustable Tubing Clamp
* In the event of saline residual in the rotor chamber (for example, from overfill of tubes or
tube breakage) saline will exit through this tube. Under normal operating conditions, saline
will exit through the drain tubing not through the overflow tube.
2-3
Page 16
InstallationSORVALL® Centrifuges
!
1. Unscrew the four knurled thumbscrews from the peristaltic pump,
and remove the front half of the pump housing (do not remove
the pump roller assembly).
2. Turn the roller assembly so that one of the rollers is in the 11
o'clock position.
3. Position the 255 mm (10 inch) frosted pump tubing (with pump
connector) in front of the pump so that the connector is positioned just above the left side of the pump, then without turning
the roller assembly, route the pump tubing around the rollers.
4. Pull up on the two ends of the pump tubing to fit the tubing
between the pump housing and the lower roller.
W A R N I N G
Saline solution from some
manufacturers contain sodium azide
as a preservative. If this solution is
discharged directly down the drain,
explosive azide salts may form as
the sodium azide reacts with the
plumbing. Check with your saline supplier before discharging saline from
the Cellwasher 2 into a drain.
NOTE
Do not try to fit the pump tubing with a sharp object, this
could puncture the tubing.
5. With the pump tubing properly positioned in the back half of the
pump housing, reinstall the front half of the housing. Gently
stretch the ends of the pump tubing to eliminate slack while
fitting the front half of the housing in place.
6. Visually check to make sure that the front half of the housing is
properly aligned and fully seated against the back half (a gap
indicates that the pump tubing is pinched; if so, remove the front
half of the housing and repeat steps 3 through 6).
7. Install the four thumbscrews and hand-tighten.
8. Check the pump tubing installation by turning the roller assembly shaft counterclockwise using pliers (the rollers should rotate
smoothly without pinching tubing).
9. If the 610 mm (24 inch) supply tubing is too long, cut it to an
appropriate length, then slide the flow control valve (catalog
number 12925, supplied) onto the supply tubing, and connect the
end of the supply tubing to a saline supply reservoir.
NOTE
The preferred location for the saline supply is either at or
above instrument level. If, at some later time, the saline
reservoir is moved from its original location, recheck the
saline volume by priming the system as directed in
Chapter 3, Operation, Priming the System.
10. Make sure the drain tubing flows downward to an open sink,
drain, or waste receptacle because drainage is accomplished by
gravity. Use the shortest length of drain tubing possible (if the
1220 mm (48 inch) tubing is too long, cut it to an appropriate
length) and make sure the drain tubing is positioned so there is
no upward travel to result in a "trap" for collected waste fluid.
2-4
Page 17
Cellwasher 2Installation
!
11. Ideally, the drain tubing should be positioned so that it can not
become immersed in collected waste fluid. However, if there is a
possibility that the drain tubing will become immersed in waste
fluid, cut the drain tubing at a convenient point near the waste
receptacle and insert the Y-Connector. Always insert a Y-Connector in the drain line when using a waste receptacle.
NOTE
The Y-Connector prevents pressure from building up in
the drain tubing due to submerged tubing or air blockage. Failure to vent the drain tubing by installing the YConnector can result in improperly decanted samples.
12. If necessary, use additional tubing and connectors from the
Tubing Kit shipped with the Cellwasher 2 to supply saline and
discharge waste properly.
Rotor and Distributor Installation
a. DA-12 Dual Angle Rotor and Rotating
Bowl Installation
1. Open the lid of the Cellwasher 2.
NOTE
There may be one or more holes in the stainless steel
rotating bowl. The holes are for balancing purposes only
and will not affect the performance of the rotating bowl.
2. Position the rotating bowl assembly (catalog number 12796,
supplied) over the motor drive shaft. Align the two drive pins
in the bowl with the slot in the drive shaft (see figure 2-2).
Slide the bowl down the drive shaft until the two pins in the
bowl engage the rubber drive ring. Press the bowl down
firmly.
3. Slide the DA-12 Rotor (catalog number 04285, supplied)
over the drive shaft and lower it gently into the bowl.
C A U T I O N
If the rotating bowl assembly, rotor, and distributor are not
installed properly, damage to the instrument could result. Before
operating the Cellwasher 2:
• Make sure the drive pins in rotat-
ing bowl are engaged in the rubber
drive ring and in the rotor.
• Make sure the distributor is firmly
seated on the rotor with the distributor pins completely engaged
in the rotor (see figure 2-3).
4. Hold the edge of the bowl and turn the rotor slowly until the
drive pins in the bowl engage the rotor and the rotor drops
down onto the bottom of the bowl.
b. Distributor Installation
Place the Distributor (catalog number 43260, supplied) over the
rotor knob (see figure 2-3). Turn the distributor until its three
pins engage the three holes in the rotor. Press the distributor
down until it is completely seated on the rotor.
Results may be affected if
tubes are not within the specifications provided in Chapter 1 on page
1-3. Using tubes of inadequate
strength could result in tube breakage: sample and test results will be
lost, the rotor and distributor could be
damaged, and cleanup will be required. During cleanup, contact with
glass fragments could result in personal injury.
Rotor Loading and Balancing
The Cellwasher 2 can process any balanced load up to twelve tubes.
When using less than a full complement of tubes, place the tubes in
opposing compartments to ensure that the rotor is symmetrically
balanced (see figure 2-4). The volume of saline delivered through
the distributor will be the same; at the locations without tubes, the
saline will be dispensed into the basin and drained off. Read the
WARNING.
TWO
THREE
= TUBE
= EMPTY
FOUR
NINE
SIX
TEN
EIGHT
Figure 2-4. Tube Placement for Balancing Less Than Twelve Tubes
2-7/2-8
Page 20
InstallationSORVALL® Centrifuges
2-8
Page 21
Cellwasher 2Operation
!
Chapter 3: OPERATION
This chapter describes the controls, indicators and operating procedures for the SORVALL® Cellwasher 2.
Controls and Indicators
All controls and indicators for the Cellwasher 2 are located on the
W A R N I N G
The SORVALL Cellwasher 2
is intended to be used in strict accordance with The American Association
of Blood Banks' guidelines defined in
the most recent edition of the
Technical Manual (published by AABB
Press. Bethesda, MD).
AABB
front panel keyboard, except for the main power switch and the
remote saline flow control valve. The sensor-touch controls register
commands visually and/or audibly.
Power to the instrument is controlled by a switch located in the
lower right hand corner of the Cellwasher 2, beneath the keyboard.
The symbol "l" indicates ON, while the symbol "O" indicates the
OFF position. Switching the power off erases any manually programmed spin times from the memory.
Table 3-1 lists and describes the controls and indicators that can be
found on the front panel. Figure 3-1 identifies each item and shows
the location on the keyboard.
Table 3-1. Cellwasher 2 Controls and Indicators
(keyed to Figure 3-1)
ItemNameAcceptance SignalFunction
1AUTOLED* Lights up,Selects AUTO run mode for automatic wash
accompanied by one beep.cycle.
2HIGHLED* lights up,Selects HIGH speed run for duration of time
accompanied by one beep.selected, or indefinitely if HOLD is selected.
3LOWLED* lights up,Selects LOW speed run for duration of time
accompanied by one beep.selected; HOLD is not to be selected.
4CHECKLED* lights up,When in AUTO mode only, stops at the end
accompanied by one beep.of any step in the wash cycle to allow verifi-
cation of procedure.
5CYCLEPanel key illuminates,Selects 1, 2, 3 or 4 wash cycles during the
digital cycle displayAUTO mode.
accompanied by one beep.
*LED, or light emitting diode, refers to the small amber light adjacent to the panel button.
(continued)
3-1
Page 22
OperationSORVALL® Centrifuges
Table 3-1. Cellwasher 2 Controls and Indicators
ItemNameAcceptance SignalFunction
(continued)
6CYCLEDisplays the number of wash cycles
DISPLAYselected in the AUTO mode.
7TIMEDisplays the time selected for AUTO,
DISPLAYHIGH, or LOW mode in seconds.
8TIMEDigital time display,In HIGH or LOW mode only, selects
accompanied by one beep.duration of centrifugation time in seconds.
9STARTLED* lights up,Starts the mode selected.
accompanied by one beep.
10STOPLED* lights up,Stops the spin in progress. In AUTO,
accompanied by one beep.HIGH, or LOW mode, the program will
return to conditions at the start of the run.
11PRIMESingle beep.Primes the pump for 6 seconds.
12AGSingle beep.Agitates the rotor for 5 seconds to
resuspend cell button.
13CLEARSingle beep.In HIGH or LOW mode only, clears
digital display so that time can be reset.
14ALARMIn AUTO mode only, turns off the alarm.
OFF
15STEPIn AUTO mode only, bypasses one step
at a time in the wash cycle for each push
of the button.
16HOLDSingle beep.In HIGH mode only, sets the centrifuge
to spin indefinitely; not to be selected in
LOW mode.
17ALARMSingle beep, ALARMIn AUTO mode only, activates an alarm
lights up.that sounds at the end of the cycle.
18SALINESALINE lights up.Lights up when saline supply is
interrupted during AUTO fill.
19LIDLED* lights up.**Opens the lid latch at the end of a run.
20ALARMALARM lights up.In AUTO mode only, lights up to indicate
the alarm will sound at the end of the run.
*LED, or light emitting diode, refers to the small amber light adjacent to the panel button.
**The Cellwasher 2 can only be opened when the LID light is illuminated.
3-2
Page 23
Cellwasher 2Operation
Figure 3-1. Cellwasher 2 Controls and Indicators
Principles of Operation
The SORVALL® Cellwasher 2 has three modes of operation: AUTO,
HIGH, and LOW. In the AUTO mode, the washing steps of the
Coomb's Procedure are performed automatically. In the HIGH and
LOW modes, the Cellwasher 2 spins at a fixed speed (see Chapter 1,
Cellwasher 2 Specifications, page 1-2).
NOTE
a. The AUTO Mode
The Cellwasher 2 is not designed to be able to automatically separate donor whole blood to create the 2 to 5%
red blood cell/saline suspension that is required for
washing cells. The Cellwasher 2 may, however, be operated as a centrifuge to manually separate red blood
cells from whole blood (refer to b. HIGH and LOWModes on page 3-4). To prevent loss of blood sample
and avoid the subsequent blood cleanup required, donot attempt to use the AUTO mode to separate whole
blood.
In the AUTO mode, the Cellwasher 2 performs a 80-second wash
cycles* that include the following steps: fill, high speed spin,
decant, and agitate. The AUTO mode proceeds automatically
through the number of wash cycles selected, unless interrupted
by the operator or by the SALINE alarm.
* 100V 50Hz models perform 90-second wash cycles.
3-3
Page 24
OperationSORVALL® Centrifuges
The operator can interrupt the AUTO mode at any point to check the
progress of the wash cycle by pressing the CHECK button. When the
CHECK button is activated, the Cellwasher 2 will stop at the end of
the washing step in progress (fill, spin, or decant).
The instrument may also stop if the saline supply is interrupted. The
SALINE alarm will sound, and the cycle in progress will stop after
the filling step. To restart the instrument, the pump must be primed
until the saline flow is continuous (see Auto Operation, step b.
SALINE alarm, page 3-8). When the Cellwasher 2 is restarted, it
begins at the next step of the wash cycle – the high speed spin.
The information programmed into the Cellwasher 2 at the start of the
AUTO cycle remains in memory until the POWER switch is turned
off.
b. HIGH and LOW Modes
NOTE
The HIGH speed mode allows the Cellwasher 2 to be
used as a centrifuge to separate red blood cells from
donor whole blood, for use in the creation of the 2 to 5%
red blood cell/saline suspension that is required for
washing cells. To do so, fill tubes with the desired proportions of whole blood and saline, then perform a 30 to
60 second run in the HIGH speed mode according to
standard protocols.
The HIGH and LOW modes are used for agglutination spins. These
spins are preceded by a 5 second agitation to allow reagent and
sample to mix. After agitation, the samples are spun at a constant
rate of speed for the amount of time selected. The spin time programmed by the operator remains in the memory for that mode,
appearing automatically in the digital TIME display whenever that
mode is selected. Once the run begins, the TIME displays counts
down the spin time in seconds to zero. At the end of the spin, the
rotor will come to a stop. The Cellwasher 2 lid cannot be opened
until the light next to the LID button is illuminated.
NOTE
Depending on run conditions and component variability
from instrument to instrument, the rotor may still be
rotating slowly after the LID light is illuminated.
3-4
Page 25
Cellwasher 2Operation
!
Tube Selection and Use
The Cellwasher 2 is designed to use either 5 ml 12 x 75 mm, or 3 ml 10 x
75 mm*, glass centrifuge tubes (saline flow must be adjusted to one size
tube or the other). In normal operation, tubes are subjected to significant
stress, making the use of high-strength centrifuge tubes necessary to avoid
tube breakage. The Cellwasher 2 spins tubes at a 45° angle with tube
walls point-loaded and otherwise unsupported. Maximum RCF generated may be as high as 1365 g (948 g in 50 Hz operation), and adequate
tubes would typically be strength-rated for an even higher g-force, as tube
manufacturers' strength ratings are based on the tubes being fully supported.
W A R N I N G
Tubes must be able to withstand up to 1365 g (948 g in 50 Hz
operation) when run at a 45° angle
with tube walls point-loaded and otherwise unsupported. Using tubes of inadequate strength could result in tube
breakage: sample and test results will
be lost, the rotor and distributor could
be damaged, and cleanup will be required. During cleanup, be aware that
glass fragments are sharp, and contact could result in personal injury.
Use only centrifuge tubes that meet the following criteria.
Tube Material . . . . . . . . . . . . . . . . . . . . . . . . Type 1, class A borosilicate
or equivalent high-strength
rolled glass (such as Pyrex®)
As additional criteria, do not use tubes that have been
dropped, scratched, or that have a visible defect, as
they may not be able to withstand centrifugation.
Be advised that, even if you do not intend to reuse tubes, you may
find that centrifuge-rated tubes of adequate strength are often labeled reusable, not disposable. SORVALL® 5 ml 12 x 75 mm Pyrex
tubes (Catalog Number 03102, 50/box) and SORVALL® 3 ml 10 x
75 mm Pyrex® tubes (Catalog Number 03100, 50/box) are not
supplied, but are recommended for use in the Cellwasher 2.
®
* Before using 3 ml, 10 mm diameter tubes, a set of Adapter Clips (Catalog No. 04330, 12/
pkg., supplied) must be installed on the DA-12 Rotor.
3-5
Page 26
OperationSORVALL® Centrifuges
Priming the System & Adjusting
Saline Fill Volume
NOTE
If sterilization is required, follow the Decontamination
procedure in Chapter 4 on page 4-2, then prime the
system as described below.
The Cellwasher 2 must be primed and adjusted using the following
procedure at installation, and whenever the saline supply reservoir
or the tubing is changed.
NOTE
When the saline flow is interrupted, prime the Cellwasher
2 as described in Auto Operation, step b. SALINE alarm,
page 3-8.
To Prime the System:
1. Open the lid (press LID button).
2. Select:
MODECYCLE
AUTO2
■■
■
■■
■■
■
■■
3. Holding a graduated cylinder under the saline dispenser in the
lid, press:
CMD
FLOW CONTROL
VALVE (located on
saline supply tubing)
PRIME
■■
■
■■
NOTE
If air bubbles remain in the tubing, repeat the PRIME
command. Air bubbles may cause uneven distribution of
saline in the test tubes.
4. Once air bubbles have been removed, empty the graduated cylinder and repeat step 3 to check total saline volume. If an adjusted
total fill volume (based on tube fills) has previously been determined, compare the collected volume to that volume; if not,
initial recommended coarse-adjust total volumes are 54 ml for
12 mm (5 ml) tubes, or 36 ml for 10 mm (3 ml) tubes.
5. If total volume is not correct, adjust the flow control valve (turn
counterclockwise to allow more saline or clockwise to allow
less), then repeat steps 3 and 4 until the desired total volume is
obtained. This completes coarse-adjustment based on nominal
total volume; a fine-adjustment based on tube fills is still required.
6. Place 12 tubes in the DA-12 rotor, close the lid, press START,
and then press CHECK. After filling is complete, the lid light
will come on. Press the LID button, open the lid and check the
saline level in each tube.
3-6
Page 27
Cellwasher 2Operation
7. Fine-adjust saline volume based on tube fills as follows:
Fine-Adjusting Saline Fill Volume
Based On Tube Fills
This adjustment procedure is required when the system is initially set
up, when priming the system, or whenever saline fill volume adjustments are required.
5 mm
TOP OF TUBE
FILL RANGE
HIGHEST FILL IS
TO BE WITHIN
1. Fine-adjust the
saline fill volume
until the level
of saline in the
highest tube fill
is between 5 &
8 mm from the
top of the tube.
8 mm
NOTE
Correct tube fills are necessary for proper cell washing;
we recommend that you check fill volume periodically.
Fill volume should also be checked whenever the saline
supply reservoir is moved or replaced.
To check, install tubes with samples in all 12 places,
start an AUTO run, and press CHECK. The Cellwasher 2
will stop after the fill step – open the lid and examine the
level of saline in each tube. Fill volume is considered
acceptable if:
• When using the recommended tubes (page 3-5), the
level of saline in the highest tube fill is between 5 and 8
mm (0.2 and 0.3 inch) from the top of the tube.
• the difference between the highest fill tube and lowest
fill tube is not more than 10 mm (0.4 inch).
Using a graduated cylinder (following the procedure on the previous
page), first adjust the flow control valve to the previously-determined adjusted total fill volume (if an adjusted total fill volume has
not yet been established, set the volume to 54 ml for 12 mm diameter
[5 ml] tubes, or to 36 ml for 10 mm diameter [3 ml] tubes). After
adjusting to achieve the desired total fill volume, fine-adjust the
volume based on tube fills.
HIGHEST FILL
10 mm RANGE
ALL FILLS ARE
TO BE WITHIN
MINIMUM LINE
– LOWEST FILL
TO BE ABOVE
2. Make sure that
the difference
between the
highest tube fill
and the lowest
tube fill does not
exceed 10 mm.
Install tubes in all 12 places, and after filling, check each tube to
determine the highest tube fill –the highest fill should be between 5
and 8 mm (0.2 and 0.3 inch) from the top of the tube when recommended tubes are used – adjust the flow control valve and recheck
fills as required. After the highest tube fill is correct, make sure that the
difference between the highest tube fill and the lowest tube fill is not
more than 10 mm (0.4 inch), equating to a maximum variance of
approximately 0.7 ml in 12 mm diameter tubes, or 0.45 ml in 10 mm
diameter tubes – troubleshoot if the fill variance is too great.
If, after initially coarse-adjusting the total fill volume (to 54 or 36
ml), the flow control valve had to be readjusted based on tube fills,
again use a graduated cylinder to measure the new adjusted total fillvolume and record the new value to use as an initial coarse-adjustment point for subsequent checks (volume adjustments will continue to change due to tube lot tolerances and normal changes to
system flow over time).
3-7
Page 28
OperationSORVALL® Centrifuges
!
W A R N I N G
During operation, never
lean on or move the instrument, keep
the defined clearance area clear of
all objects (including all hazardous
and flammable materials), and do not
work within the clearance area (see
Location in Chapter 2).
Auto Operation
a. The basic operating sequence in the AUTO mode is as follows:
1. Press the LID button to open the lid.
2. Install the rotor, tubes, and distributor as described in Chapter 2, Installation.
3. Close the lid.
4. Select:
5. For an audible signal at the end of the run, press ALARM
ON.
6. Press START to begin the run.
NOTE
If you wish to check for proper fill, press CHECK after
beginning the run. At the end of the fill step, the
Cellwasher 2 will stop. Examine the level of saline in
each tube. Fill is considered acceptable if:
• When using the recommeded tubes (page 3-5), the
• the difference between the highest fill tube and lowest
Press START to continue run.
MODECYCLE
AUTO1-4
■■
■
■■
level of saline in the highest tube fill is between 5 and 8
mm (0.2 and 0.3 inch) from the top of the tube.
fill tube is not more than 10 mm (0.4 inch).
7. At the end of the run, if the alarm sounds, press ALARM
OFF to turn off the alarm.
8. Once the lid LED is illuminated, press LID to open the lid.
b. SALINE alarm – this alarm will sound during the filling step if
the saline supply is interrupted. To deactivate the alarm and
correct the problem, prime the Cellwasher 2 as follows:
1. Open the lid (press LID button).
2. Holding a graduated cylinder under the saline dispenser in
the lid, press:
CMD
NOTE
PRIME
The SALINE alarm will not shut off until the PRIME
button is pressed.
■■
■
■■
3-8
Page 29
Cellwasher 2Operation
3. Close the lid and press START to continue the wash cycle in
progress.
NOTE
The SALINE alarm may also indicate broken or obstructed tubing. If the saline reservoir is adequate, check
the tubing. Clean or replace tubing if necessary (see
Chapter 4, Maintenance).
To ensure adequate washing, perform one additional
wash cycle at the completion of the cycles in progress.
c. During AUTO operation, the following features may also be
used:
•CHECK button – the automatic wash cycle can be checked
after any step (fill, spin, or decant) by pressing the CHECK
button during that step. The Cellwasher 2 will automatically
stop at the end of the step in progress, and the lid may then be
opened. After checking the tubes, close the lid and press
START to continue run.
•STOP button – when STOP is pressed, the run in progress
will stop immediately. When the LID light goes on, the lid
may be opened. When the Cellwasher 2 is restarted, it will
begin the programmed cycle over again unless steps are
bypassed (see STEP button, Table 3-1).
•STEP button – if the AUTO mode is interrupted by a STOP
command, it will restart at the beginning of the programmed
run. To move the sequence ahead (to avoid refilling tubes for
instance), press the STEP button once for each step in the
cycle to be bypassed.
3-9
Page 30
OperationSORVALL® Centrifuges
!
W A R N I N G
During operation, never
lean on or move the instrument, keep
the defined clearance area clear of
all objects (including all hazardous
and flammable materials), and do not
work within the clearance area (see
Location in Chapter 2).
High and Low Speed Operation
a. To operate the Cellwasher 2 in the HIGH or LOW speed mode,
proceed as follows:
1. Press the LID button to open the lid.
2. Install the rotor, tubes, and distributor as described in Chapter 2, Installation.
3. Select:
NOTE
When HIGH or LOW is pressed, the time selected for the
previous run will appear automatically in the digital time
display. In the HIGH and LOW modes, the spin time that is
input remains in the memory of that mode until a new time
is input, or until the power is switched off.
MODETIMECMD
HIGH1-999
SECONDS
LOW
START
After the START button is pressed, agitation will occur automatically for 5 seconds before the run begins. The time will count down
in seconds. At the end of the run, the lid light will come on. Press
LID to open the lid.
NOTE
NOTE
Depending on run conditions and component variability
from instrument to instrument, the rotor may still be
rotating slowly after the lid is opened.
The Cellwasher 2 is not intended to perform repetitive
Low Speed TIMED runs without allowing a rest period of
at least 10 minutes between runs with the door open to
allow the motor to cool. Failure to allow the motor to cool
between repetitive Low Speed runs can cause the motor to overheat and shut down.
b. During HIGH or LOW speed operation, the following features
may also be used:
•AGITATE – pressing the AG button, when the rotor is not
spinning, will provide an additional 5 seconds of agitation.
•HOLD – after selecting HIGH mode only, press HOLD to
provide an indefinite spin period; HOLD is not to be selected
in LOW speed mode. To end the run, press STOP.
NOTE
The Low Speed Mode is not intended for continuous
use; performing lengthy Low Speed HOLD runs could
cause the motor to overheat and shut down. Because of
this, Low Speed HOLD runs are not recommended.
3-10
Page 31
Cellwasher 2Operation
!
!
T ube Breakage
Necessary Action to Prevent Instrument
Malfunction
W A R N I N G
In the event of tube breakage, be careful to avoid personal injury due to sharp fragments. Always
make sure all glass is removed, including any pieces that may have
become lodged in the overflow and
drain tray exits and tubing. Also, make
sure that the overflow container is
empty before resuming use.
C A U T I O N
The distributor is balanced
as an assembly. In extreme cases, if
it must be disassembled for cleaning,
all parts must be marked for position/
orientation so that it will be in balance
after reassembly.
Tube breakage followed by an inadequate cleanup can be a major
cause of instrument malfunction. Residual glass fragments can block
drains and cannulas, leading to flooding of internal components and
an accelerated buildup of encrusted salt/media. Every time tube
breakage occurs, or if fluid exits through the overflow tubing or
appears from under the cabinet, a thorough cleaning and inspection
should be completed before continued operation. Disconnect power
and decontaminate the Cellwasher 2 as required, then perform the
cleaning and inspection as follows:
1. In the lid, separate the collector ring from the retaining ring by
removing the four mounting screws and stepwashers (for parts
identification, refer to Figure 4-1 on page 4-3). Clean all glass
and residue from both sides of the collector ring, and from inside
its drain nozzle and the connected discharge tubing. Using water, be sure the collector ring drains freely through the tubing – if
not, clear blockage and recheck as required. Clean the surface of
the retaining ring and its O-ring, then reassemble, making sure
the collector ring is fully seated on the retainer ring before
securing it with screws and stepwashers.
2. Remove and clean the distributor assembly, the rotor, and the
rotating bowl assembly. On the distributor assembly, make sure
that no fragments have fallen down inside the center area, to
ensure that cannulas will not become obstructed, and make sure
cannulas have not been damaged. On the rotating bowl assembly, make sure that no fragments have become situated between
parts by checking that the lift plate mechanism slides up and
down freely.
3. Clean the wet guard (black plastic liner) that covers the chamber
and top deck, then remove it. Locate the small white overflow
tray, which is positioned directly under, and obstructed by, the
decant coil (a black ring around the top of the motor assembly).
Clean all glass and residue from the overflow tray, and from
inside its drain nozzle and the connected overflow tubing. Using
water, be sure the overflow tray drains freely through the tubing
(being careful not to overflow the tray, if drainage is still impaired) – if not, clear blockage and recheck as required.
4. Reinstall the wet guard, then, being sure to engage the drive pins
on each, reinstall the rotating bowl assembly, the rotor, and the
distributor assembly. If the problem recurs, contact Kendro Service.
3-1 1
Page 32
OperationSORVALL® Centrifuges
!
!
W A R N I N G
To avoid exposure to a
spinning rotor and potential for personal injury, if a momentary power
failure terminates a run, wait at least
ten seconds (from when power was
lost) before trying to open the lid.
W A R N I N G
This procedure is included
for emergency sample recovery only
and should never be used for any
purpose other than those explained
in this section.
When the main power shuts off the
brake will not operate. Wait until the
rotor stops spinning before using the
mechanical override.
Momentary Power Interruption
If a momentary power interruption occurs when a run is in progress
with the rotor spinning, it can take up to ten seconds before the rotor
comes to a stop. Do not open the lid with the rotor spinning.
Emer gency Sample Recovery
If, while the rotor is spinning, the Cellwasher 2 main power shuts off
due to a power failure or system malfunction, the lid is designed not
to open, to keep the operator safe. For emergency situations such as
these, where the rotor has stopped spinning but the LID button will
not allow access to the chamber, a mechanical override is provided
to allow sample recovery.
The mechanical override is located on the left side of the Cellwasher
2 as shown in figure 3-2, below. To operate the override, insert a
small screwdriver or similar object into the hole and push to cause
the latch pin to disengage the striker plate and release the lid.
HOLE FOR
MECHANICAL
OVERRIDE
Figure 3-2. Location of Mechanical Override
3-12
Page 33
Cellwasher 2Maintenance
!
!
Chapter 4: MAINTENANCE
This chapter describes routine maintenance procedures that should be performed on a regular basis, with
the specific time interval or duty cycle determined by the user, and based on product use. It is also the
responsibility of the user to make certain that these procedures are followed whenever necessary. If further
service is needed, contact Kendro or your local representative for SORVALL® products.
W A R N I N G
There are no user-serviceable items inside the Cellwasher 2.
Due to the hazards involved, repair
should only be attempted by a qualified technician who is familiar with
electronics and trained in the servicing of this product.
Preventive Maintenance
The following preventive maintenance checks and cleaning procedures should be performed regularly. If etiologic or biologically
hazardous materials are processed in the Cellwasher 2, be sure
appropriate decontamination procedures have been followed before
inspecting or cleaning the instrument.
a. Inspection
•Inspect the vents on the bottom of the Cellwasher 2 to be sure
they are not blocked, and remove any obstructions.
•Check all sealing surfaces, tubing, liners, and the collecting ring
assembly for cleanliness and good condition.
•Inspect the distributor, rotor, and stainless steel rotating bowl:
1. Remove the distributor from the top of the rotor; check for
cracks and make sure cannulas are not bent, damaged, or
clogged. From underneath, make sure the metal clip that
holds the distributor to the rotor is not cracked or broken.
2. Remove the DA-12 rotor from the rotating bowl assembly.
Make sure that the rotor does not wobble on its base. Make
sure that the pivot pin at the top of each tube holder is not
W A R N I N G
Continued use of a rotor
with tube holder band gaps greater
than 2 mm, or that has had bands
compressed to close the gaps, can
cause tube breakage, possibly resulting in instrument damage or personal
injury.
NOTE
loose. Also, check the gaps in the tube holder bands – gaps
may widen over time due to centrifugal force, and the rotor
should not be used if the gaps become greater than 2 mm. Donot compress the tube holder bands to close the gaps; bands
are precisely shaped to support tubes during centrifugation,
and tube support will be compromised if the bands are
deformed in this manner (read the WARNING).
3. Remove the stainless steel rotating bowl assembly (located in
the rotor chamber, directly beneath the rotor), and inspect the
rotating bowl assembly for signs of cracks or corrosion.
If inspection reveals that any part of the Cellwasher 2 is
not functioning properly, do not use the instrument until
it is repaired.
4-1
Page 34
MaintenanceSORVALL® Centrifuges
!
b. Cleaning
C A U T I O N
All saline solutions have
long term corrosive effects. Routine
cleaning and maintenance are essential to ensure safe and efficient
operation.
•Clean the Cellwasher 2 cabinet with a damp cloth and mild
detergent.
•Wipe up spills from the interior and exterior whenever they
occur.
•Remove the rotor, distributor, and rotating bowl and wash them
with warm water and mild detergent.
•Remove the collecting ring assembly and tubing for routine
cleaning as follows (see figure 4-1):
1. From the back of the instrument, open the adjustable tubing
clamp that secures the Y-Connector in place and remove the
discharge tubing from the Y-Connector (see figure 2-1).
2. Remove the four mounting screws and step washers securing
the collecting ring assembly to the lid.
3. Pull the collecting ring assembly away from the lid, and pull
the discharge tubing up through the wet guard (molded black
liner).
4. Disconnect the flow tubing from the spray nozzle in the lid
and pull it through the hole in the retainer ring.
5. Separate the collector ring assembly by pulling the collector
ring off.
6. Remove discharge tubing and flow tubing, wash with warm
water and mild detergent, and reinstall (see figure 2-1).
7. Wash the retainer ring, O-ring seal, and collector ring with
warm water and mild detergent.
8. To reassemble, place the retainer ring on work surface (small
opening down), and work the O-ring seal around the large
opening of the retainer ring until it is fully seated in the
groove. Then, place the collector ring so that it rests concentrically over the O-ring seal and press firmly into place.
NOTE
Make sure the nozzle for the discharge tubing on the
collector ring aligns with the groove in the wet guard when
the collecting ring assembly is reinstalled in the lid.
9. Reinstall the collecting ring assembly by reversing steps 1
through 4 of this procedure.
NOTE
4-2
When replacing the four mounting screws, make sure
the flat side of each step washer faces the screw head.
Page 35
Cellwasher 2Maintenance
!
SPRAY
NOZZLE
LID
COLLECTOR RING
ASSEMBLY
O-RING
GROOVE
RETAINER
RING
O-RING
SEAL
COLLECTOR
RING
STEP
WASHER
MOUNTING
SCREW
Figure 4-1. Collecting Ring Assembly, Exploded View
W A R N I N G
If tubes have been broken
in the Cellwasher 2, be aware of the
potential for residual glass fragments
and take extra precautions.
The Cellwasher 2 is intended for use
with human blood. Always observe
Universal Precautions.
FLOW
TUBING
DISCHARGE
TUBING
TO FLOW
DETECTOR
TO WASTE
CONTAINER OR
T-CONNECTOR
c. Decontamination
The Cellwasher 2 is intended for use in Coomb's testing and incorporates the use of human blood. If the Cellwasher 2 has been used
with other hazardous materials, be aware of all hazards involved and
decontaminate using additional appropriate methods. As a guide, the
SORVALL® Product Guide contains descriptions of commonly used
decontamination methods and a chart showing method compatibility
with various materials. Otherwise, observe Universal Precautions
and proceed as directed on the next page.
4-3
Page 36
MaintenanceSORVALL® Centrifuges
If the Cellwasher 2 is oper ational
1. Prepare a 3-10% solution of household liquid chlorine bleach.
2. Soak a wash rag or sponge with the bleach solution and gently
wipe the bowl assembly, the wet guard, and the inner portion of
the Cellwasher 2, making sure all encrusted material is gone. Do
this step twice.
3. Disconnect the supply tubing from the saline supply and set the
tubing to draw from the bleach solution.
4.Pump the bleach solution through the Cellwasher 2 for four cycles.
5. Set the tubing to draw from clean water and follow the bleach
solution with 4 to 12 water washes, so that no bleach remains.
6. Reconnect the tubing to the saline supply and follow the water
washes with four cycles of saline.
7. Check the pH level to make sure it is normal. If not, continue
with saline cycles until pH checks normal.
8. Open the lid, turn the main power switch OFF, and unplug the
power cord. Wipe all parts (including parts inside the lid and
chamber, the cabinet, the control panel, and the front surface of the
main power switch) with a 70% ethanol solution and allow to dry
(do not flood the switch or power cord connector areas).
If the Cellwasher 2 is not oper ational
1. Turn the power switch OFF, unplug the power cord, and open the lid
using the mechanical override (see page 3-12, Emergency Sample
Recovery). Remove tubes (if any) from the DA-12 rotor.
2. Prepare a 3-10% solution of household liquid chlorine bleach.
Disassemble the collector ring assembly (see Cleaning, page 4-2);
immerse components for 10 minutes in the bleach solution.
3. Remove the discharge tubing, vent tubing, and drain tubing
along with any Y-connectors, draining any excess saline from the
pieces. Immerse and completely fill the tubing and connectors in
the bleach solution (try not to leave any air in the tubing); keep
them immersed for 10 minutes.
4-4
NOTE
Depending on the age or condition of the tubing, as well
as the type or amount of contamination, you may choose
to replace tubing instead of decontaminating it. If so,
discard contaminated tubing in an appropriate manner,
then, after the instrument has been decontaminated,
install a Tubing Replacement Kit (Catalog No. 12977,
supplied), following instructions supplied in the kit.
Page 37
Cellwasher 2Maintenance
4. Individually remove the distributor, the DA-12 rotor, and the
stainless steel bowl. Immerse the distributor for 10 minutes in
the bleach solution. Wipe the DA-12 rotor and stainless steel
bowl using a 70% solution of ethanol, making sure that all
encrusted material is gone, then rinse with water. Repeat the
ethanol wash a second time.
5. Remove the wet guard (black plastic liner) from the chamber.
Soak a wash rag or a sponge in the bleach solution and gently
wipe the wet guard and the inner portion of the unit (including
the nozzle and inside of the lid, the motor shaft and the decant
coil). Make sure that all encrusted material is gone.
6. Make sure a container is positioned to collect fluid from the
overflow tubing. Pouring slowly so that the tray doesn't overflow, pour some bleach solution into the white plastic overflow
tray (below the decant coil) this should drain out freely through
the overflow tubing. Follow with clean water to rinse.
7. Wipe the entire cabinet, including the control panel and the front
surface of the main power switch, with 70 % ethanol (do not
flood the power switch or power cord connector areas).
8. After 10 minutes in the bleach solution, remove all components
that had been soaking and rinse with water several times.
9. Allow all components to dry, then reassemble the unit as before.
NOTE
When reinstalling the four mounting screws of the collector ring assembly, make sure that the flat side of each
step washer is against the screw head.
Ser vice Decontamination Policy
If a Cellwasher 2 requires servicing by Kendro personnel, either at
the customer’s laboratory or at a Kendro facility, comply with the
following to ensure the safety of all personnel.
If the Cellwasher 2 has been used with other hazardous materials, be
aware of all hazards involved and decontaminate using additional
appropriate methods. As a guide, the SORVALL® Product Guide
contains descriptions of commonly used decontamination methods
and a chart showing method compatibility with various materials.
Otherwise, if a Cellwasher 2 requires servicing by Kendro, proceed
as follows:
1. Decontaminate the Cellwasher 2 to be serviced, and clean it of
all encrusted material prior to servicing by a Kendro representative or returning it to the Kendro facility. There must be no
radioactivity detectable by survey equipment. Decontaminate
according to the instructions that begin on page 4-3.
4-5
Page 38
MaintenanceSORVALL® Centrifuges
2. Complete and attach a Decontamination Information Certificate
(in the back of this manual) to the Cellwasher 2 before servicing.
In addition to those included with in this book, Decontamination
Information Certificates are available from the local Kendro
Representative or Field Service Engineer. In the event that certificates are not available, a signed, written statement certifying
that the unit has been properly decontaminated, identifying what
the contaminants were and outlining the decontamination procedures used will be acceptable.
NOTE
The Field Service Engineer will note on the Customer
Service Repair Report if decontamination was required
and, if so, what the contaminant was and what procedure was used. If no decontamination was required, it
will be so stated.
If a Cellwasher 2 to be serviced does not have a Decontamination
Information Certificate attached and, in Kendro's opinion presents a
potential radioactive or biological hazard, the Kendro representative
will not service the equipment until proper decontamination and
certification is complete.
If the Cellwasher 2 must be returned to a Kendro facility:
1. Contact your Kendro representative to obtain an Equipment
Return Decontamination Form; be prepared with the product
name (Cellwasher 2), serial number and the repairs required.
2. Complete the Equipment Return Decontamination Form and
return it to Kendro. Upon receipt of a completed form, a Returned Material Authorization Number (RMA Number) will be
issued to you.
4-6
3. With the RMA Number clearly marked on the outside of packaging, send the items to the address obtained from your Kendro
representative.
NOTE
United States federal regulations require that parts and
instruments must be decontaminated before being transported. Outside the United States, check local regulations.
If equipment is received at Kendro facilities without a valid RMA
Number on the outside of the shipping container and a completed
Equipment Return Decontamination Form on file, the equipment
will be treated as a potential contamination hazard, and will not be
serviced until decontamination certification has been completed.
The sender will be contacted for instructions regarding disposition
of the equipment in question; all disposition costs will be borne by
the sender. If contaminated equipment is received at Kendro facilities, both the carrier and appropriate authorities shall be notified.
Page 39
Cellwasher 2Maintenance
!
Par ts Replacement
To order replacement parts, telephone toll-free 800-522-7746 in the
United States. Outside the United States, contact Kendro or your
local representative for SORVALL
partial list). Be prepared to supply the product name (Cellwasher 2),
serial number, and a description of the parts required.
®
products (see back cover for a
Tubing Replacement
If tubing is damaged and must be replaced, install tubing supplied in
the Tubing Replacement Kit (Catalog No. 12977) following installation instructions supplied with the kit.
Fuse Replacement
W A R N I N G
There are no user-serviceable items inside the Cellwasher 2.
Due to the hazards involved, repair
should only be attempted by a qualified technician who is familiar with
electronics and trained in the servicing of this product.
To replace the fuse in the power supply adapter:
1. Unplug the power cord.
2. Remove the power cord from the power supply adapter.
3. Slide the clear plastic guard to the left of the power cord connection.
4. Pull out the FUSE PULL lever. The end of the fuse will pop out.
5. Remove the fuse, and replace it with the appropriate slow-blow
fuse. (Fuse rating is indicated on rear nameplate).
6. Move the guard to the right, and replace the power cord.
Troubleshooting
Tables 4-1 through 4-4 contain troubleshooting information for
some common problems that occur during operation of the Cellwasher
2. The customer should not attempt any further troubleshooting.
Tables 4-5 through 4-7 contain applications troubleshooting information which identify possible causes of false negative results, false
positive results, or inconsistent reaction strength.
4-7
Page 40
MaintenanceSORVALL® Centrifuges
Table 4-1. Tr oubleshooting Chart: Problem A
Instrument completely
inoperative and displays not
illuminated.
Is instrument plugged in?
Yes
Is power switch on?
Yes
Is AC line fuse blown?
No
Contact Kendro
Representative
No
No
Yes
Connect power cord as
required.
Place power switch in the
ON position.
Replace fuse. See fuse replacement
procedure on page 4-7.
Table 4-2. Troubleshooting Chart: Prob lem B
Is saline volume correct?
Yes
Is distributor dirty? Are cannulas
clogged?
No
Are tubes the correct length and
diameter?
No
Replace with tubes of correct length
and diameter.
Improper saline fill of tubes.
No
Yes
Yes
Adjust volume (see page 3-7, Fine-Adjusting
Saline Fill Volume Based on Tube Fills).
Clean distributor and each cannula.
Contact Kendro Representative.
4-8
Page 41
Cellwasher 2Maintenance
Table 4-3. Troubleshooting Chart: Pr oblem C
Instrument will not operate and displays
Is lid properly latched?
Yes
Contact Kendro Representative
are illuminated.
No
Table 4-4. Troubleshooting Chart: Pr oblem D
(see figures 4-2 and 4-3)
Decant Malfunction.
Close lid properly.
Is decant coil aligned properly?
(Alignment is improper if lift plate ring
shows abnormal rubbing on one side.)
Yes
Does lift plate ring move
freely?
Yes
Is the drive ring loose of
deformed?
No
Does motor turn
counterclockwise?
No
Contact Kendro
Representative
Yes
No
Yes
No
Align decant coil to be concentric
with motor shaft.
Clean ring – look for glass
fragments, rust, salt residue, or
other foreign substances.
Secure or replace
drive ring.
Faulty clutch – replace.
4-9
Page 42
MaintenanceSORVALL® Centrifuges
DRIVE
RING
MOTOR
SHAFT
Figure 4-2. Location of Decant Coil and Drive Ring
DECANT
COIL
4-10
BOWL
Figure 4-3. Location of Lift Plate Ring
LIFT
PLATE
RING
Page 43
Cellwasher 2Maintenance
Table 4-5. Applications Troubleshooting Char t:
False Negative Results
False negative results.
Is the red blood cell suspension
concentration correct (2-5%)?
Yes
Was the antiglobulin serum
omitted?
No
Are the reagents no longer
reactive due to improper
storage?
No
Was the reagent contaminated
with human serum?
No
Was cell-coating inadequate
during incubation?
No
Yes
Yes
Yes
Yes
Modify the cell suspension
concentration.
Repeat the test, noting
antiglobulin step.
Repeat the test with fresh reagent.
Repeat the test with fresh reagent.
1. Check incubation time.
2. Check the volume of
reagent added.
No
Were the samples undercentrifuged
in the manual mode?
No
Contact Kendro Representative
Yes
Increase spin time.
4-1 1
Page 44
MaintenanceSORVALL® Centrifuges
Table 4-6. Applications Troubleshooting Chart:
False Positive Results
False positive results.
Is the saline solution or the
tubing contaminated?
No
Are the reagent cells
contaminated by bacteria?
No
Does the reagent contain
extraneous antibodies?
No
Is septicemia present in the
patient serum?
No
Is extreme reticulocytosis
present?
Yes
Yes
Yes
Yes
Yes
1. Change the saline supply.
2. Clean and decontaminate the
tubing and the distributor.
Repeat the test with fresh
reagent cells.
Crossmatch versus fresh
reagent.
Test the patient serum for
bacterial contamination.
1. Test versus fresh reagent.
2. Observe patient red blood
cells microscopically.
4-12
No
Were the samples overcentrifuged in
the manual mode?
No
Contact Kendro Representative.
Yes
Decrease spin time.
Page 45
Cellwasher 2Maintenance
Table 4-7. Applications Troubleshooting Char t:
Inconsistent Reaction Strength
Inconsistent reaction
strength from tube to tube.
Were the same number of
cells added to each tube?
Yes
Was the same amount of
reagent added to each tube?
Yes
Is the antiglobulin serum
weak or outdated?
No
Contact Kendro Representative.
No
No
Yes
Repeat the test, adding the
same number of cells to
each tube.
Repeat the test, adding the
same amount of reagent to
each tube.
Repeat the test with fresh
antiglobulin serum.
4-13/4-14
Page 46
Page 47
Cellwasher 2Appendix
APPENDIX
Page 48
AppendixSORVALL® Centrifuges
War r anty
Kendro Laboratory Products makes no warranty of any kind, expressed or implied, except as stated in this
warranty policy.
The SORVALL® Cellwasher 2 Cell Washing instrument and DA-12
Rotor are warranted (subject to the conditions specified below and in
the warranty clause of the Kendro terms and conditions of sale in
effect at the time of sale) to be free from defects in material and
workmanship for a period of one (1) year from the date of delivery.
Kendro will repair or replace and return free of charge any part
which is returned to its factory within said period, transportation
prepaid by user, and which is found upon inspection to have been
defective in materials or workmanship. This warranty does not
include normal wear from use, it does not apply to any instrument or
part which has been altered by anyone other than an employee of
Kendro, nor to any instrument which has been damaged through
accident, negligence, failure to follow operating instructions, the use
of electric currents or circuits other than those specified on the plate
affixed to the instrument, misuse or abuse.
Kendro reserves the right to change, alter, modify or improve any of
its instruments without any obligation whatever to make corresponding changes to any instrument previously sold or shipped.
The foregoing obligations are in lieu of all other obligations and
liabilities including negligence and all warranties, of merchantability or otherwise, expressed or implied in fact or by law, and state our
entire and exclusive liability and buyer's exclusive remedy for any
claim or damages in connection with the sale or furnishing of goods
or parts, their design, suitability for use, installation or operation.
Kendro will in no event be liable for any special or consequential
damages whatsoever, and our liability under no circumstances will
exceed the contract price for the goods for which liability is claimed.
Page 49
Cellwasher 2Appendix
Index
Accessories Supplied, 1-4
Adjusting Fill Volume, 3-6
Applications Information, 1-1
Balancing, Rotor, 2-7
Cellwasher 2
description, 1-1
specifications, 1-2
Cleaning, 4-2
Controls and Indicators, 3-1
Kendro Laboratory Products
Beijing Representative Office
8th Floor, Office Building
Hong Kong Macau Center
No. 2 Chao Yang Men Bei Da Jie
Beijing 100027, P.R. China
Tel.: 86-(10)-6501-1573