Sonel P-3 Operating Manual

Page 1
Page 2
Page 3
OPERATING
MANUAL
VOLTAGE TESTER
P-3
SONEL S. A.
ul. Wokulskiego 11
58-100 Świdnica
Version 1.6 May 22, 2012
Page 4
2
TABLE OF CONTENTS
1
SAFETY .............................................. 3
2
MEASUREMENTS ........................... 4
2.1
TESTER OPERATION CONTROL ........ 4
2.2
AC OR DC VOLTAGE TEST .............. 4
2.3
RCD OPERATION TEST .................... 5
2.4
SINGLE-POLE PHASE TESTING .......... 5
2.5
CIRCUIT CONTINUITY TEST ............. 6
2.6
MEASUREMENT OF RESISTANCE ..... 6
2.7
PHASE SEQUENCE TEST ................... 7
2.8
PHASING ......................................... 7
2.8.1
Without transmitter ................. 8
2.8.2
With transmitter ...................... 9
2.9
LIGHTING OF THE MEASURED POINT
..
..................................................... 12
3
BATTERY REPLACEMENT ........ 12
4
CLEANING AND MAINTENANCE
.
........................................................... 13
5
STORAGE ........................................ 13
6
DISMANTLING AND
UTILIZATION
................................ 13
7
TECHNICAL DATA ....................... 14
8
MANUFACTURER ......................... 16
We appreciate your having purchased our volt­age tester. The P-3 tester is a modern high-quality, simple and safe device. However it is recommended to get acquainted with the present manual in order to avoid measuring errors and prevent possible problems related to operation of the meter.
Note:
The producer reserves the right to modify the
appearance, equipment and technical data of the device.
Page 5
3
1 Safety
The P-3 meter has been devised for the purpose of voltage and circuit continuity testing, measure­ment of resistance and indication of the phase spin direction and phase identification. In order to guar­antee proper operation and adequacy of the results the following recommendations must be observed:
• Before you proceed to operate the meter, ac- quaint yourself thoroughly with the present manual and observe the safety regulations and specifications determined by the producer.
• Any other application than those specified in the present manual may result in a damage to the device and constitute a source of danger for the user.
• The device must be operated solely by appro- priately qualified personnel with relevant certifi­cates that entitle them to realise measurements of electric installation. Operation of the meter realised by unauthorised personnel may result in damage to the device and constitute a source of danger for the user.
• The instrument must not be used for mains and in premises of special conditions, e.g. in a dan­gerous environment regarding the possibility of explosion and fire.
• It is unacceptable to operate the following: ⇒ A damaged meter which is completely or
partially out of order,
⇒ A meter with damaged cable insulation, ⇒ A meter stored for an excessive period of
time in disadvantageous conditions (e.g. ex­cessive humidity). If the meter has been
transferred from a cool to a warm envi­ronment of a high level of relative humid­ity, do not realise measurements until the meter has been warmed up to the ambi­ent temperature (approximately 30 min­utes).
• Do not operate a meter with an open or incor- rectly closed battery compartment or power it from other sources than those specified in the present manual.
• Voltage measurements must not last longer than 30s. Having concluded a 30-second meas­urement, the following measurement may be performed not sooner than after 240s.
• Limit value signalling is solely provided as warn- ing for the operator and not for the purpose of measurement.
Page 6
4
Note:
The device can be used in rainy conditions,
however at user’s own responsibility.
It is
recommended to use gauntlets.
2 Measurements
2.1 Tester operation control
Before each operation of the tester, its function­ing must be checked:
• Test the voltage indicator using a known volt- age source,
• Short-circuit the measurement probes – there should be a sound signal, the LED diode Ω
should be lit and there should be the
leg-
end in the display.
Notes:
The voltage indication function is active whether the batteries are charged or not. Other functions re­quire charged batteries.
2.2 AC or DC voltage test
Page 7
5
Notes:
In mains protected with an RCD switch, whose nominal current is 10mA or 30mA, the switch may be activated during a voltage measurement between L and PE. In order to avoid that connect the tester between L and N and after approximately 5s switch the probe over from N to PE.
2.3 RCD operation test
In order to check the operation of the RCD switch, whose nominal current is 10mA or 30mA, perform a voltage test directly between the L phase and the PE protective conductor.
2.4 Single-pole phase testing
Notes:
During single-pole phase testing for the purpose of identification of external conductors, under certain circumstances the functioning of the tester may be impaired (e.g. if insulated individual protective means are used or is the working station is insu­lated).
Note! Do not touch the measurement probe electrode L1 during the phase testing.
Page 8
6
Single-pole phase testing may not be sufficient to determine whether the circuit is live. Perform two­pole voltage test.
2.5 Circuit continuity test
Note!
The tested object must not be live.
Connect both probes to the tested object. A sound signal, diode Ω on, and the
legend indi-
cate continuity of the circuit (R<600kΩ).
Notes:
During the circuit continuity measurement, the voltage polarization at the L2 probe is negative!
2.6 Measurement of resistance
Note!
The tested object must not be live.
Press the Ω button for a short while and connect both probes to the tested object. Read the result from the LCD.
Notes:
During the resistance measurement, the voltage polarization at the L2 probe is negative!
Page 9
7
2.7 Phase sequence test
Notes:
The phase sequence may solely be determined in a three-phase system. Contact of the touch elec­trode during measurement in a single-phase system gives a random result.
2.8 Phasing
The phasing function facilitates comparison of voltage phases in two distant point of the installa­tion.
Page 10
8
2.8.1 Without transmitter
Page 11
9
Additional information displayed by the meter
Touch again the touch electrode. It is displayed if the previous contact was too short.
Notes:
- Correct phasing is possible after no more than 15s from the sound signal. Afterwards horizontal lines are displayed.
- If before 6s from the moment of synchronisation the tester is not disconnected from voltage, syn­chronization commences anew.
- When phase is indicated in another socket and its symbol is displayed, precise synchronization of the P-3 receiver commences automatically. Thus phas­ing between distant sockets is possible, if there are other sockets in between.
- If the tester is disconnected from voltage before
synchronization has concluded, the
symbol is
displayed.
2.8.2 With transmitter
The transmitter mated with the P-3 tester is a part of auxiliary equipment. The new LKN-710P transmitter is not factory­dedicated to any P-3 indicator, and therefore before phasing may commence, it is required to perform the so called mating (point 2.8.2.1). The transmitter during a given period may work with solely one P-3 indicator. Repeated mating is required in case an­other P-3 indicator is to be used. If the transmitter and indicator mating has not been performed, go on to point 2.8.2.1 of the man­ual.
Page 12
10
If both devices were mated before go on to point
2.8.2.2.
2.8.2.1 Mating
1. Connect the LKN-710P transmitter cables to the
source of alternating voltage 24...230V. After the connection has been performed, the red diode should blink signalling its operation in the conductor tracking mode (along with the LKO-710 received by SONEL S.A.).
2. Then, press the
button twice, until the diode
turns greed (continuous light).
3. Press fleetingly the L123 button on the P-3 dis­play. The following message should be displayed
and after a short while the symbol .
4. Then hold the L123 button until the message
is displayed
.
5. Keep the
button of the LKN-710P transmitter
until the green diode stops blinking.
6. Now the transmitter is synchronized with the P-3 indicator. The process of synchronization lasts up to
30 seconds. The symbol
may blink during syn-
chronization.
7. Correct mating of the transmitter with the indica­tor is signalled as follows:
• The message
is replaced by ,
while the symbol
is displayed continu-
ously,
• The green diode of the transmitter blinks faster than at the moment of mating, sig­nalling transmission of phasing signal.
From that moment on both devices are ready for work in tandem.
Note :
Green light signal indicated operation of the transmitter in the P-3 mode.
Page 13
11
2.8.2.2 Correct work of the P-3 indicator with the LKN-710P transmitter
Page 14
12
Notes:
If during the phase indication (L-1, L-2 or L-3)
the symbol
starts to blink and the
is dis­played, it means there are problems, which may be related to:
- excessive distance between the transmitter and the P-3 indicator,
- transmission interference, which is not related to distance. In such a case, the problem will be solved automatically in no more than 10-20 seconds.
The
symbol, which appears when the
symbol is displayed continuously means:
- no voltage at the P-3 indicator measurement ter­minals,
- incorrect connection of the LKN-710P transmitter or the P-3 indicator.
2.9 Lighting of the measured
point
The P-3 tester may light the measured area un­der difficult lighting conditions (e.g. in switchgears). In order to light the measured point, press the fol-
lowing button:
. The LCD is backlit simultane-
ously.
3 Battery replacement
The tester is powered from two 1.5V AAA bat­teries. A lack of sound signal after the probes have
been short-circuited or dim light after the button has been pressed, or the symbol
indicate the need to replace the batteries. In order to do so, per­form the following actions:
Page 15
13
4 Cleaning and maintenance
NOTE! Use solely the maintenance techniques speci­fied by the manufacturer in the present operat­ing manual.
The tester may be cleaned with a soft, damp cloth using all-purpose detergents. Do not use any solvents or cleaning agents which might scratch the casing (powders, pastes, etc.). The electronic system of the meter does not re­quire maintenance.
5 Storage
In the case of storage of the device, the follow­ing recommendations must be observed:
• Place protection on the probes,
• Make sure the tester is dry.
• Should the tester be stored for a prolonged period
of time, the batteries will be removed.
6 Dismantling and utilization
Worn-out electric and electronic equipment should be gathered selectively, i.e. it must not be placed with waste of another kind. Worn-out electronic equipment should be sent to a collection point in accordance with the law of worn-out electric and electronic equipment. Before the equipment is sent to a collection point, do not dismantle any elements.
Observe the local regulations concerning dis­posal of packages, worn-out batteries and accumu­lators.
Page 16
14
7 Technical data
The abbreviation „i.v.” in the basic uncertainty definition means the indicated value.
Constant voltage measurement (on the display):
* - from 6,0V for Sr. no. up to 681793 Additionally the voltage is indicated by the diode bar for the following values: 12, 24, 50, 120, 230, 400, 690V along with the signalling of the voltage polari­zation (the diode „+” or „-” is lit). The diode indicator functions also without batteries. Input resistance
Uin
Rin
12V, 24V, 50V
~ 6kΩ
120V
~ 20kΩ
230V
~ 70kΩ
400V
~ 150kΩ
690V
~ 240kΩ
Alternating voltage measurement within the range between 20 and 400Hz (on the display):
* - from 6,0V for Sr. no. up to 681793 Additionally the voltage is indicated by the diode bar for the following values: 12, 24, 50, 120, 230, 400, 690V along with the signalling of alternating voltage (diodes „+” and „-” are lit simultaneously). The diode indicator functions also without batteries. The measurement voltage frequency for the diode bar: 15…400Hz.
Resistance measurement:
Range Resolution
Basic
uncertainty
0...1999Ω 1Ω
±(3% i.v. + 8 dig-
its)
Other data:
a) Kind of insulation: double, in accordance with EN 61010-1 b) Measurement category: III 1000V (IV 600V) in
accordance with EN 61010-1
Range Resolution
Basic
uncertainty
+1,5/-2,5*
...49,9V
0,1V
±(2% i.v. + 3
digits)
50...750V
1V
Range Resolution
Basic
uncertainty
1,5*...49,9V 0,1V
±(3% i.v. + 4 dig-
its)
50...750V 1V
±(2% i.v. + 3 dig-
its)
Page 17
15
c) Casing protection grade in accordance with EN
60529: IP65, with an open battery compartment:
IP10 d) Voltage measurement range on the LCD:
6…750V AC/DC e) Voltage indication for the diode bar: 12V, 24V,
50V, 120V, 230V, 400V, 690V f) Minimum indicator activation voltage: 6V g) Accuracy of voltage indications: in accordance
with EN 61243-3 h) Voltage frequency range for the LCD:
20…400Hz i) Voltage frequency range for the diode bar:
15…400Hz j) Maximum current: I
S
<0,2A/IS(5s)<3,5mA k) Maximum period of continuous operation: 30s l) Minimum interruption of operation after 30 sec-
onds of work: 240s
m) Voltage range for a single-pole phase indicator:
50…690V
n) Frequency range for a single-pole phase indica-
tor: 50…400Hz
o) Continuity tester range: 0…600kΩ (400kΩ for
Sr. no. up to 681793)
p) Accuracy of the continuity tester activation
threshold: ±50%
q) Continuity tester measurement current: 7μA r) Voltage range for two-pole phase sequence in-
dicator: 100…690V
s) Frequency range for two-pole phase sequence
indicator: 50…60Hz
t) Time to automatic turn-off (Auto-OFF):
approximately 7s
during phasing approximately 30s u) Display: LCD 3 1/2 digits v) Tester power supply: 2x1,5V AAA/LR03 (alkaline batteries are recommended) w) Dimensions: approximately 240x60x30 mm x) Mass of the tester with batteries:
approximately 0,2kg y) Working temperature: -10..+55°C z) Storage temperature: -30..+70°C aa) Quality standard: design and production in ac-
cordance with ISO 9001 bb) the product meets the EMC requirements (im-
munity for industrial environment) according to
the following standards EN 61326-1:2006 and
EN 61326-2-2:2006
Transmitter data:
a) Kind of insulation: double, in accordance with
PN-EN 61010-1 b) Measurement category: III 300V in accordance
with PN-EN 61010-1 c) Casing protection grade in accordance with PN-
EN 60529: IP40 d) Voltage range: 24…250V AC
Page 18
16
e) Voltage frequency range: 45…65Hz f) Dimensions: approximately 128 x 66 x 28 mm g) mass: approximately 0,25kg
8 Manufacturer
The manufacturer of the device, which also pro­vides guarantee and post-guarantee service is the following company:
SONEL S.A.
ul. Wokulskiego 11
58-100 Świdnica
Poland tel. +48 74 858 38 60 fax +48 74 858 38 09
E-mail:
Web page:
www.sonel.pl
Note: Service repairs may be realised solely by the manufacturer.
Page 19
Page 20
Loading...