Water Level Meter Faceplate Replacement
Instructions
For Mk2 Water Level Meters (Push-Release Tape Connection)
Note: For instructions for Mk1 Water Level Meters (Molex tape
connection, soldered connections to battery tray) see page 3.
Model 101 SC1000, SC2000, SC3000
7. Unscrew the two screws holding the circuit board (and leads
from light if you have an older style light – see photo below)
to the Sonalert (watch for proper connection/polarity).
8. If you have an older style light (see photo below) push the
light out of the faceplate.
Tools and Materials Needed
1. 101 Replacement Faceplate Assembly
• Faceplate and Handle (without electronics)
101 P2: #103565, #105299, or #105301
or 101 P7: #110502, #110503, or #110504
2. Phillips Screwdriver
3. Small Flat Screwdriver
4. Needle Nose Pliers
Sonalert
Light
Test
Button
Handle
Sensitivity
Switch
Battery
Tray
Leads
Sonalert
From Light
Sensitivity
Switch
Push-Release Fittings
Light
Battery
Tray
Back of Mk2 101 Faceplate with Older Style Light
Sensitivity
Switch
Sonalert
Push-Release
Test Button
Fittings
Light
101 P7 Replacement Faceplate Assembly
indicating the location of the components
(comes with Handle Assembly attached and 3 Phillips screws)
1. Place the reel on a flat surface, with the faceplate up.
Remove the battery from the Meter.
2. Use the Phillips screwdriver to undo the three screws
holding the faceplate to the hub. Remove the faceplate.
3. To disconnect the tape from the circuit board, press down
on the white terminals of the push-release fittings on the
circuit board and pull out to remove the tape leads.
4. Unscrew the nut from the test button and push the button
out of the faceplate.
5. Use the small flat screwdriver to unscrew the small brass
screw on the side of the sensitivity knob. Remove the
sensitivity knob. Use the pliers to unscrew the nut holding
the sensitivity switch and remove the switch from the
faceplate.
6. Use the Phillips screwdriver to undo the two screws holding
the battery tray to the faceplate. Remove the black and
red wires by pulling the quick-connect fittings off of the
terminals on the battery tray. Remove the battery tray.
(Page 1 of 3)
High Quality Groundwater and Surface Water Monitoring Instrumentation
Battery
Tray
Test
Button
Back of Mk2 101 Faceplate with New Style Light
9. Remove the circuit board assembly with connected wires,
test button, sensitivity switch, and new style light if (see
photo above).
10. If you have a new style light, unscrew the red lens from the
faceplate.
11. Unscrew the Sonalert retaining ring from the front of the
faceplate and remove the Sonalert.
12. Put the Sonalert through the new faceplate, and secure it
using the retaining ring. Position it as shown in the photos
above, depending on which light you have.
13. If you have a new style light, install the red lens in the new
faceplate and secure it using the white nut.

Water Level Meter Faceplate Replacement
14. If you have an older style light (see photo below), thread
the leads of the light assembly through the front of the
faceplate and snap the light in place.
Leads
Sonalert
From Light
Sensitivity
Switch
Push-Release Fittings
Light
Battery
Tray
Test Button
Back of Mk2 101 Faceplate with Older Style Light
15. Place the circuit board assembly back in position on top of
the Sonalert, ensuring the sides marked +ve and -ve are
lined up with the corresponding terminals on the Sonalert
and if you have a new style light, it is seated in the lens (see
photo below).
16. Use the two screws to connect the wires from the circuit
board (and light if you have the older style) to the Sonalert.
Make sure the wires are connected to the correct terminals
(+ve and -ve) on the Sonalert.
17. Insert the test button through the opening in the faceplate,
and secure it in place using the nut.
18. Insert the sensitivity switch through the faceplate with the
three connection prongs positioned towards the battery
drawer, slightly towards the centre of the reel. Tighten the
nut. Fasten the sensitivity knob on the front of the switch
using the brass screw.
19. Insert the battery tray through the faceplate and secure
using the screws. Connect the black wire with quickconnect fitting to the negative terminal on the battery
tray. Connect the red wire with quick-connect fitting to the
positive terminal. Replace the battery.
20. To reconnect the tape to the circuit board, press down on
the white terminals and insert the tape leads. Release the
terminals and the leads should be secured. The lead on the
bottom of the tape (numbers facing up, may be black mark
on lead) is inserted into the terminal with a white square
below it on the circuit board. See photo below.
Sensitivity
Switch
Sonalert
Battery
Tray
Back of Mk2 101 Faceplate with New Style Light
Push-Release
Fittings
Light
Test Button
Square
21. With the probe in a glass of tap water, turn the Water Level
Meter to the ‘ON’ position. If the connections are correct
the buzzer and light will activate. If the buzzer or light do
not activate, check the polarity of the battery and that all
the electronics are connected properly.
22. Screw the new faceplate to the reel using the new Phillips
screws.
(Page 2 of 3)
High Quality Groundwater and Surface Water Monitoring Instrumentation

Instructions
For Mk1 Water Level Meters (Molex Tape Connection)
Note: Instructions for Mk2 Water Level Meters are on page 1
(tape connection using push-release fittings on circuit board
and quick-connect fittings to battery tray).
Test Button
Light
Sonalert
Tools and Materials Needed
1. 101 Replacement Faceplate Assembly
• Faceplate and Handle (without electronics)
101 P2: #103565, #105299, or #105301
or 101 P7: #110502, #110503, or #110504
2. Phillips Screwdriver
3. Small Flat Screwdriver
4. Needle Nose Pliers
5. Wire Strippers and Cutters
6. Soldering Iron and Wire
Sonalert
Light
Test
Button
101 P2 Replacement Faceplate Assembly
indicating the location of the components
(comes with Handle Assembly attached and 3 Phillips screws)
1. Place the reel on a flat surface, with the faceplate up.
Remove the battery from the Meter.
2. Use the Phillips screwdriver to undo the three screws
holding the faceplate to the hub.
3. Remove the faceplate and disconnect the Molex connector
that attaches the faceplate electronics to the tape.
4. Unscrew the test button and push it out of the faceplate.
5. Unscrew the two screws holding the wires from the light,
through the circuit board and onto the Sonalert (watch for
proper polarity). Push the light out of the faceplate.
Sensitivity
Switch
Battery
Tray
Handle
Battery
Tray
Molex
Connector
Sensitivity
Switch
Back of Mk1 101 faceplate showing wiring connections
and the location of the components
6. Use the small flat screwdriver to unscrew the small brass
screw on the side of the sensitivity knob. Remove the
sensitivity knob. Use the pliers to unscrew the nut holding
the sensitivity switch and remove the switch from the
faceplate.
7. Use the Phillips screwdriver to undo the two screws holding
the battery tray to the faceplate. Cut the red and black wires
from the back of the battery tray close to the terminals (other
end of black wire is connected to the sensitivity switch, and
red wire connects to the circuit board). Remove the battery
tray. Strip the wires about 1/4" (6 mm).
8. Unscrew the Sonalert retaining ring from the front of the
faceplate and remove the Sonalert.
9. Put the Sonalert through the new faceplate, and secure it
using the retaining ring.
10. Replace the sensitivity switch, light, and test button in the
new faceplate.
11. Reconnect the two wires, from the light, using the two
screws through the circuit board and onto the Sonalert.
12. Insert the battery tray through the new faceplate. Re-solder
the red and black wires to the correct terminals labeled on
the back of the battery tray. Red wire from the circuit board
to the positive terminal and black wire from sensitivity switch
to the negative terminal. Use the two screws to secure the
tray to the faceplate. Replace the battery.
13. Connect the Molex connector from the faceplate to the
tape.
14. With the Probe in a glass of tap water, turn the Water Level
Meter to the ‘ON’ position. If the connections are correct
the buzzer and light will activate. If the buzzer or light do
not activate, check the polarity of the battery and Molex
connector, and the soldered connections.
15. Screw the new faceplate to the reel using the new Phillips
screws.
Printed in Canada
March 20, 2018
(#109336)
(Page 3 of 3)