Smiths Medical CADD-Prizm PCS II, 6101 Technical Manual

CADD-Prizm® PCS II Ambulatory Infusion Pump
Model 6101
Technical Manual
1
Table of Contents
Limited Warranty ..............................................1
Exposure to Radiation or Magnetic Resonance
Imaging (MRI) ................................................... 1
2. CADD‑Prizm® PCS II Pump ...................... 2
Delivery Modes .................................................2
Specifications (Nominal) ..................................4
Battery Compatibility .......................................7
DURACELL® Alkaline Battery Life ................... 7
ULTRALIFE® Lithium Battery Life ....................7
Keyboard Circuitry .......................................... 10
Data Memory EEPROM ..................................10
Battery Backed RAM ...................................... 10
Time Base Circuitry ........................................ 10
LCD Circuitry .................................................. 10
LED Status Indicators ....................................11
Flash PROM Technology ................................11
Gate Array Circuitry ........................................11
Audible Alarm Circuitry .................................. 11
Watchdog Timer Circuit .................................11
Motor Driver/Motor Watchdog Circuit ...........11
Power Circuitry...............................................12
Voltage Reference Circuit...............................12
Pumping Mechanism .....................................13
Pumping Characteristics ...............................13
Air Detector .................................................... 14
Upstream Occlusion Sensor .......................... 14
6. Safety Features and Fault Detection ...... 15
Hardware Safety Features .............................15
Watchdog Timer Circuit .................................15
Motor Driver/Motor Watchdog Circuit ...........15
Cassette ‘Type’ Sensor Circuit .......................16
Latch/Lock Sensor Circuit .............................16
Voltage Detector Circuit .................................16
Software Safety Features ...............................17
Overview ......................................................... 18
Order of Error Code Events ............................18
8. Cleaning and Inspection Procedures...... 19
Inspection Recommendation ......................... 19
Cleaning .........................................................19
Visual Inspection ............................................ 19
Mechanical Inspection ...................................19
Functional Testing .......................................... 20
Air Detector Test (if Applicable) .....................22
Occlusion Tests ..............................................23
Accuracy Testing ............................................ 24
Cleaning and Functional Testing Checklist ...26
For detailed instructions, specifications, warnings, warranties and additional information on operating CADD have additional comments or questions concerning the operation of CADD
®
pumps, please refer to the Operator’s Manual supplied with the product. If you
®
pumps, please call this number 1‑800‑258‑5361. Our staff is available to help you 24 hours a day with the programming and operation of CADD
®
pump infusion systems.
The issue date of this Technical Manual is included for the user’s information. In the event one year has elapsed between the issue date and product use, the user should contact Smiths Medical to see if a later revision of this manual is available.
Issue Date: November 2010
2
1 Introduction
The Technical Manual is intended to provide a basic, but limited, understanding of the mechanical and electrical operation of the Smiths Medical CADD‑Prizm
®
PCS II Computerized Ambulatory Drug Delivery pump to persons familiar with this device. The CADD‑Prizm
®
PCS II Operator’s Manual should be used in conjunction with this publication for complete information.
This manual also outlines cleaning and functional testing procedures that can be performed on the CADD‑Prizm
®
PCS II pump.
This technical manual is applicable to the CADD‑Prizm
®
PCS II pump only.
IMPORTANT NOTICE
CADD‑Prizm® PCS II pump operations and safety features are based on a microcomputer design. Inadequate servicing or tampering with the safety features of the pump may seriously affect performance and safety.
For that reason, All servicing and repair of the
CADD-Prizm
®
PCS II pump must be performed
by Smiths Medical or its authorized agents.
The manufacturer’s warranty agreement shall become null and void if the pump is not used in accordance with the Operator’s Manual and Instructions for Use for the pump accessories; or, the pump is serviced by persons other than Smiths Medical or those authorized by Smiths Medical.
Limited Warranty
The limited warranty associated with the CADD‑Prizm in the product literature supplied with the product when originally purchased, which is incorporated herein by reference.
Smiths Medical specifically disclaims any other warranty, whether express, implied or statutory, including, without limitation, any implied warranty of merchantability or fitness for use. Smiths Medical further disclaims
responsibility for the suitability of the system for a particular medical treatment or for any medical complications resulting from the use of the system. The manufacturer shall not be responsible for any incidental damages or consequential damages to property, loss of profits, or loss of use caused by any defect or malfunction of the system.
®
PCS II pump can be found
If you wish to receive additional information about the extent of the warranty on these products, please contact your Smiths Medical representative or call Customer Service at 1‑800‑258‑5361.
All recommendations, information and literature supplied by Smiths Medical with respect to the CADD
®
product line are believed to be accurate and reliable, but do not constitute warranties. No agent, representative, or employee of Smiths Medical has authority to bind Smiths Medical to any representation or warranty, expressed or implied.
Exposure to Radiation or Magnetic Resonance Imaging (MRI)
CAUTIONS
1 The pump SHOULD NOT BE DIRECTLY IRRADIATED by therapeutic levels of ionizing radiation because of the risk of permanent damage to the pump’s electronic circuitry. The best procedure to follow is to remove the pump from the patient during therapeutic radiation sessions or diagnostic levels of radiographic and fluoroscopic radiation. If the pump must remain in the vicinity during a diagnostic or therapy session, it should be shielded, and its ability to function properly should be confirmed following treatment.
2 Magnetic fields produced by magnetic resonance imaging (MRI) equipment may adversely affect the operation of the pump. Remove the pump from the patient during MRI procedures and keep it at a safe distance from magnetic energy.
1
2 CADD-Prizm® PCS II pump
Clinician Bolus
(used here as a loading dose)
Demand Doses
Continuous Rate
Time
Dosage
Delivery Modes
The CADD‑Prizm® PCS II pump provides measured drug therapy to patients. CADD‑Prizm
®
PCS II pumps are indicated for intravenous, intra‑arterial, subcutaneous, intraperitoneal, epidural space or subarachnoid space infusion. Epidural administration is limited to short‑term infusion of anesthetics and either long‑ or short‑term infusion of analgesics. Subarachnoid administration is limited to short‑term infusion of analgesics. (See Figure 1.)
Indicator Lights
Amber
Display
Keypad
Green
¤
Power jack
Data In/Out jack
Œ ´ Å Í Î
PCA Delivery Profile
The PCA (patient‑controlled analgesia) delivery mode is used for therapies that require a continuous rate of infusion, patient‑ controlled demand doses or both, such as patient‑controlled analgesia. (See Figure 2.)
Polemount
Bracket Recess
2000-03-07 D. Zurn
«Prizm Rear 3/4 (dark BW)»
Battery
compartment
Air Detector
Port Cover
Air Detector
(optional)
Figure 1. Front and back views of the CADD‑Prizm® PCS II pump.
Figure 2. PCA mode delivery profile.
h k j
®
CADD-Prizm
Upstream Occlusion Sensor
PCS II
Cassette lock
Cassette latch
Cassette
Rear ViewFront View
2
Continuous Rate Scroll Ranges
Units Starting Increment Maximum
Milliliters 0.10 0.10 30.00
Milligrams & 10% of Mg only Values between 0.01 and 0.5 0.01 Concentration Micrograms concentration Mcg only Values between 0.1 and 0.5 0.1 x 30 Values between 0.5 and 100 0.1 Values between 100 and 1000 1.0 Values greater than 1000 10.0
Table 1. PCA delivery mode continuous rate scroll ranges.
Milligrams
Concentration
mg/ml
Demand Dose increment max.
Clinician Bolus increment max.
0.1 0.01 0.99 0.01 2
0.2 0.02 1.98 0.02 4
0.3 0.03 2.97 0.03 6
0.4 0.04 3.96 0.04 8
0.5 0.05 4.95 0.05 10 1 0.05 9.9 0.05 20
2 0.10 19.8 0.10 40 3 0.15 29.7 0.15 60
4 0.20 39.6 0.20 80 5 0.25 49.5 0.25 100 6 0.30 59.4 0.30 120
7 0.35 69.3 0.35 140 8 0.40 79.2 0.40 160 9 0.45 89.1 0.45 180
10 0.50 99.0 0.50 200 11 0.55 108.9 0.55 220 12 0.60 118.8 0.60 240
13 0.65 128.7 0.65 260 14 0.70 138.6 0.70 280 15 0.75 148.5 0.75 300
20 1.00 198.0 1.00 400 25 1.25 247.5 1.25 500 30 1.50 297.0 1.50 600
35 1.75 346.5 1.75 700 40 2.00 396.0 2.00 800 45 2.25 445.5 2.25 900
50 2.50 495.0 2.50 1000 55 2.75 544.5 2.75 1100 60 3.00 594.0 3.00 1200
65 3.25 643.5 3.25 1300 70 3.50 693.0 3.50 1400 75 3.75 742.5 3.75 1500
80 4.00 792.0 4.00 1600 85 4.25 841.5 4.25 1700 90 4.50 891.0 4.50 1800
95 4.75 940.5 4.75 1900 100 5.00 990.0 5.00 2000
Table 2. Demand dose, clinician bolus scroll ranges, milligrams
Micrograms
Concentration
mcg/ml
Demand Dose increment max.
Clinician Bolus increment max.
1 0.05 9.9 0.05 20 2 0.10 19.8 0.10 40 3 0.15 29.7 0.15 60 4 0.20 39.6 0.20 80 5 0.25 49.5 0.25 100
6 0.30 59.4 0.30 120 7 0.35 69.3 0.35 140 8 0.40 79.2 0.40 160
9 0.45 89.1 0.45 180 10 0.50 99.0 0.50 200 11 0.55 108.9 0.55 220
12 0.60 118.8 0.60 240 13 0.65 128.7 0.65 260 14 0.70 138.6 0.70 280
15 0.75 148.5 0.75 300 20 1.00 198.0 1.00 400 25 1.25 247.5 1.25 500
30 1.50 297.0 1.50 600 35 1.75 346.5 1.75 700 40 2.00 396.0 2.00 800
45 2.25 445.5 2.25 900 50 2.50 495.0 2.50 1000 55 2.75 544.5 2.75 1100
60 3.00 594.0 3.00 1200 65 3.25 643.5 3.25 1300 70 3.50 693.0 3.50 1400
75 3.75 742.5 3.75 1500 80 4.00 792.0 4.00 1600 85 4.25 841.5 4.25 1700
90 4.50 891.0 4.50 1800 95 4.75 940.5 4.75 1900 100 5.00 990.0 5.00 2000
200 10.00 1980.0 10.00 4000 300 15.00 2970.0 15.00 6000 400 20.00 3960.0 20.00 8000 500 25.00 4950.0 25.00 10000
Table 3. Demand dose, clinician bolus scroll ranges, micrograms
Milliliters
Demand Dose Clinician Bolus increment max. increment max.
0.05 9.9* 0.05 20
Table 4. PCA delivery mode Demand dose, clinician bolus scroll ranges, milliliters
*The maximum Demand Dose is 20 with software revision E or higher.
3
Specifications (Nominal)
General Pump Specifications
Resolution
CADD or CADD per pump stroke nominal
Size
4.4 cm x 10.4 cm x 14.1 cm [1.7 in x 4.1 in x 5.6 in] excluding cassette or other accessories
Weight
568 g [20 oz.] including 9 volt battery and empty 100 ml CADD excluding other accessories
Pump Alarms
Low battery power; depleted battery power; external power source low, faulty, depleted; pump stopped; pump fault; low reservoir volume; high delivery pressure; air in line; Air Detector faulty or detached (only with the use of the optional Air Detector); Air Detector Port Cover detached; delivery too slow; key stuck; cassette detached or unlocked; print failure, epidural cassette not used.
medication cassette reservoir
®
administration set, 0.050 ml
medication cassette reservoir,
System Operating Temperature
+2°C to 40°C (36°F to 104°F)
System Storage Temperature
‑20°C to 60°C (‑4°F to 140°F)
Power Pack Charging Temperature
+10°C to 35°C (50°F to 95°F)
System DeliveryAccuracy
± 6% (nominal)
System Definition
System is defined as a CADD‑Prizm pump with an attached CADD cassette reservoir and CADD
®
®
PCS II
medication
extension set with integral anti‑siphon valve, or an attached CADD administration set with integral or add‑on anti‑siphon valve.
Delivery Specifications
Reservoir Volume
1 to 9999 or Not In Use; programmable in 1 ml increments, displayed in 0.1 ml increments. Default: 1 ml
®
Bolus Volume at Occlu sion Alarm
Pressure 0.050 ml resolution administration sets/CADD
medication
cassette Reservoirs <0.25 ml
Power Sources
9 volt alkaline or lithium battery such as DURA CELL Lithium U9VL; CADD
®
Alkaline MN 1604 or ULTRALIFE®
®
External Power Source (EPS) Power Pack reorder number 21‑3801; AC Adapter. The expected life of a 9 volt battery is 12 hours at 100 ml/hour, or approximately 5 days at 10 ml/day (nominal). This estimate is based on laboratory tests conducted at room temperature using a new battery. Actual battery life will vary depending on the brand of battery, shelf life, temperature conditions, delivery rate, and frequency of screen display, backlighting and printing. It is recommended that a new 9 volt battery be kept available for replacement if necessary.
An internal battery powers the clock. When it is depleted, it cannot reliably maintain the clock time. This battery must be replaced by the manufacturer. The internal battery has an expected life of 5 years.
Units*
Milliliters (ml), milligrams (mg), micrograms (mcg). Default: milligrams
Concentration
Mg/ml:
0.1 to 0.5 mg/ml in increments of 0.1 1 to 15 mg/ml in increments of 1 mg/ml 20 to 100 mg/ml in increments of 5 mg/ml Default: 100 mg/ml
Mcg/ml: 1 to 15 mcg/ml in increments of 1 mcg/ml
15 to 95 mcg/ml in increments of 5 mcg/ml 100 to 500 mcg/ml in increments of 100 mcg/ml. Default: 500 mcg/ml
Continuous Rate
0 to 30 ml/hr (or the mg or mcg equivalent). Default: 0 mg/hr (See Table 1 for Scroll ranges)
* If programmed to be part of pump programming screens in Biomed Toolbox.
4
Demand Dose
0 to 9.9 ml* Delivery rate (Continuous Rate + Demand Dose) programmable from 40 to 125 ml/hr. Default: 0 ml (See Table 2, 3 & 4 for Scroll ranges)
*The Maximum Demand Dose is 20 with software revision E or higher.
Demand Dose Lockout
5 minutes to 24 hours in the following increments:
• 1 minute for values between 1 and 20 minutes
• 5 minutes between 20 minutes and 24 hours Default: 5 min
Set Delivery Limit
0.5 ml to 1000 ml (or the mg or mcg
equivalent), or “No Limit”:
0.01 from 0.01 to 0.1
0.1 from 0.1 to 100
1.0 from 100 to 1000
10.0 from 1000 to 10000
100.0 from 10000 to 100000
1000.0 from 100000 and up
Default: 0.5 ml or mcg or mg equivalent
Time
0000 to 2359
Air Detector
Turned On or Turned Off. Default: On
Biomed Toolbox Specifications
Custom Concentrations
All individual mg or mcg concentration settings may be enabled or disabled (at least one concentration must be enabled). Default: All On
Program Limits*
Maximum program limits may be programmed for Demand Dose, Continuous Rate, and Clinician Bolus.
Default: maximum program limits.
Dosing Limit*
Delivery Limit, a Maximum Doses per Hour, or neither. Default: Neither
Key Beeps
On or Off. Default: On
Res Vol Trip Point
1 to 999 ml in increments of 1 ml, or “Standard.” Default: Standard
Given
0 to 99999.99 in 0.01 unit increments.
Clinician Bolus
0.1 ml to 20.00 ml (or mg or mcg equivalent)
Delivery rate (Continuous Rate + Clinician Bolus): 125 ml/hr nominal (See tables 2, 3 & 4 for Scroll ranges)
High Pressure Alarm
18 ± 9 psi [1.24 ± 0.62 bar]
Air Detector Alarm
Single bubble greater than 0.100 ml
Options Specifications
Lock Level
LL0, LL1, LL2. Default: LL2
Epidural Mode
On or Off. Default: Off
Units*
Milliliters (ml), milligrams (mg), micrograms (mcg). Default: milligrams
Res Vol Empty Alarm*
Single or Insistent alarm. Default: Single
Pump Stopped Alarm*
Beep or Two‑tone alarm. Default: Beep
AutoLock
Not In Use, LL1 Key/Code, LL2 Key/Code, LL1 No Key or LL2 No Key. Default: Not In Use
PM (Preventive Maintenance) Reminder 1 to 24 months in 1 month increments, or “Not In Use.” Default: Not In Use
Custom Lock Level Code
001 to 899 (excluding preset code) in increments of 1. Default: 061
Units Selection*
Program units to appear in the Programming Units screens. Default: All Programming Units
Units Location
Options, Program or Biomed Toolbox. Default: Programming screens
* If programmed to be part of Options settings in the Biomed Toolbox.
5
Date Format
US Standard (mm/dd/yy) or European Standard (dd/mm/yy). Default: U.S. Standard
Custom Main Display
Display:
• Res Vol or Continuous Rate
• Power Source Always or Low 9 volt battery only Default: Res Vol and Low 9V
Auto Review*
Select the automatic program review feature during the pump’s power‑up sequence. Default: On
• Dose Counters (0 to 999 Given and/or Attempted). Default: On
• Given. Default: On
• Doses Hour By Hour (up to 48 hours in increments of 1 hour). Default: Off
• Patient Review. Default: Off
• Pain Scale (subjective pain scale rating of 0 to 10 in increments of 1). Default: On
• Pain Scale Log (0 to 500 entries). Default: On
• Delivery Log (0 to 500 events). Default: Off
• Event Log (0 to 500 events). Default: Off
• New Patient Marker. Default: On
New Patient Marker*
Reports/No Clear, Power‑up/No Clear, Reports/Clear, Power‑up/Clear
Default: Reports/No Clear
Air Detector Required
Required or Not Required. Default: Not Required
Default:ing the Lock Level Code & Clinician Bolus Code
The standard Lock Level Code (061) can be changed to a customized code using the Biomed Toolbox Custom Lock Code feature. See the Operator’s Manual supplied with the pump for instructions on customizing the Lock Level Code. If it becomes necessary to change a customized code back to the standard Lock Level Code, do the following:
1. Press the OPTIONS key until the Lock Level screen appears
2. Press the ENTER key twice
3. Scroll to 911
4. Press the OPTIONS key
Compatible Reservoirs and Administration Sets
• 50‑ml or 100‑ml CADD™ medication cassette reservoir, used with the CADD extension set with anti‑siphon valve.
• CADD
®
administration set with integral anti‑siphon valve, with or without bag spike (allows use of flexible plastic bag or sterile vial with injector)
• CADD
®
administration set with add on anti‑siphon valve and bag spike (allows for gravity priming before attaching the add on anti‑siphon valve)
®
Remote Dose Cord
Smiths Medical provides a Remote Dose Cord for the PCA delivery mode. The push button switch is a Single Pole Double Throw (SPDT). When the Remote Dose Cord is attached to the pump, the patient may press the Remote Dose button to receive a Demand Dose. The clinician may use the Remote Dose button to deliver a clinician bolus. For easy access, the Remote Dose cord may be fastened to the patient’s clothing or bedsheet with the attached clip.
NOTE
To detach the Remote Dose cord from the pump, grasp the Remote Dose cord connector and pull back using a straight, steady motion. Do not twist or turn the connector, or use any instrument to remove it. For additional specifications refer to the Operator’s Manual provided with the product.
Note The CADD‑Sentry® Medication Software programming mode must be used to change a customized code back to the standard lock level code with CADD‑Prizm® PCS II pumps with software revision E or higher.
* If programmed to be part of pump programming screens in Biomed Toolbox.
6
3 Batteries
Battery Compatibility
Recommended Batteries
Nine‑volt alkaline or lithium batteries are recommended for use in the CADD‑Prizm
®
PCS II pump. Carbon‑zinc, mercury, nickel‑ cadmium, or zinc‑air 9‑volt batteries should not be used.
Battery Life
The CADD‑Prizm
®
PCS II pump has been designed to provide optimal battery life. The expected battery life in the CADD‑Prizm
®
PCS
II pump depends on the following factors:
• Programmed delivery rate
• Operating temperatures
• Frequency of display backlighting
• Frequency of printing
• Battery type and brand
• Battery age
DURACELL® Alkaline Battery Life
The following tables may be used to predict typical alkaline battery life at different delivery rates when an alkaline battery is used in the CADD‑Prizm battery life decreases as the delivery rate increases. These tables are based on laboratory tests using fresh DURACELL alkaline batteries in CADD‑Prizm pumps while the pumps were operating at room temperature.
Actual battery life may be significantly shorter depending on the operating temperature and the storage conditions of the battery.
Battery life is shortened significantly at very low operating temperatures. For example, at 0°C (32°F), an alkaline battery will yield approximately 30% of its normal capacity.
Alkaline batteries do not need to be stored in a refrigerator. After four years of storage at 21°C (70°F), an alkaline battery retains approximately 86% of its original capacity. Battery life will be shorter if the battery is stored above room temperature. An alkaline battery stored at 43°C (110°F) will be down to approximately 80% of its capacity within one year.
®
PCS II pump. As expected,
®
PCS II
®
Recommended storage conditions are 10°C to 25°C (50°F to 77°F) with no more than 65% relative humidity noncondensing.
The following tables are based on laboratory tests conducted at room temperature using fresh DURACELL
®
alkaline batteries and a CADD® administration set. Actual battery life will vary depending on the brand of battery, battery shelf life and temperature conditions.
ULTRALIFE® Lithium Battery Life
The following tables may be used to predict typical lithium battery life at different delivery rates when a lithium battery is used in the CADD‑Prizm life decreases as the delivery rate increases. These tables are based on laboratory tests using fresh ULTRALIFE CADD‑Prizm were operating at room temperature.
Actual battery life may be significantly shorter depending on the operating temperature and the storage conditions of the battery. Lithium battery life is dependent upon the temperature and relative humidity of storage. Recommended storage conditions are less than 20°C (68°F) with a desiccant to ensure less than 10% relative humidity.
The following tables are based on laboratory tests conducted at room temperature using fresh ULTRALIFE
®
CADD depends upon the brand of battery selected, the particular battery selected, battery shelf life, and temperature conditions. Smiths Medical’s testing indicates a large variability in battery life.
®
PCS II pump. As expected, battery
®
®
PCS II pumps while the pumps
®
lithium batteries and a
lithium batteries in
administration set. Actual battery life
7
Loading...
+ 20 hidden pages