Smiths Medical 1818R Operation And Service Manual

Page 1
BCI PulseOx Handheld Manual
Get an original copy of the BCI PulseOx Handheld Manual for manufacturer information about service, available accessories and how to use and maintain your device.
Page 2
Oximeter
Operation and Service Manual
Catalog Number 1818R
Version 7, February 2005
© 2005 Smiths Medical family of companies. All rights reserved.
Page 3
Page 4
Table of Contents
Oximeter Operation and Service Manual i
Table of Contents
Chapter 1: Introduction................................................................ 1-1
About This Manual..........................................................................1-1
Proprietary Notice................................................................... .........1-1
Warranty..........................................................................................1-1
Limited Warranty......................................................................1-1
Disclaimer of Warranties ..........................................................1-2
Conditions of Warranty.............................................................1-2
Limitation of Remedies.............................................................1-2
Warranty Procedure.........................................................................1-3
CE Notice.........................................................................................1-3
Warnings, Cautions, and Notes .......................................................1-5
Chapter 2: General Description ................................................... 2-1
Intended Use....................................................................................2-1
Features............................................................................................2-1
Description.......................................................................................2-2
Theory of Operation .................................................................... ....2-4
Chapter 3: Using the Oximeter....................................................3-1
Install the Batteries....... .... ..... ..........................................................3-1
Installing or Replacing the Batteries......................................... 3-2
Care and Handling of Sensor ....................................................3-3
Choose the Sensor............................................................................3-4
Attach the Sensor to the Patient.......................................................3-5
Clean or Disinfect the Sensors..................................................3-5
Finger Sensor for Adult or Pediatric Finger..............................3-5
Attach the Sensor to the Oximeter............................................3-6
Measuring the Patient’s % SpO2 and Pulse Rate............................3-6
Patient Numbers and Spot Check Data............................................3-8
Manually Incrementing the Patient Number.............................3-8
Clearing All Spot Check Data...................................................3-8
Low Battery Indicator......................................................................3-8
Turning Off the Oximeter................................................................3-9
Checking the Oximeter’s Performance............................................3-9
Page 5
Table of Contents
ii Oximeter Operation and Service Manual
Chapter 4: Printer and Computer Interface.................................. 4-1
Computer Interface Description...................................................... 4-1
Equipment Required........................................................................4-1
Interface Instructions.......................................................................4-1
Spot Check Printouts........................................................ .... ..... .... ..4-3
Collecting Spot Check Data......................................................4-3
Manually Incrementing the Patient Number.............................4-3
Clearing All Spot Check Data ..................................................4-3
About the Oximeter’s Batteries and Spot Check Data .............4-3
Printing Spot Check Data .........................................................4-4
Chapter 5: Operator’s Maintenance............................................. 5-1
Batteries...........................................................................................5-1
Disposal of batteries and rechargeable batteries....................... 5-1
Sensors ............................................................................................5-1
Reusable Sensors......................................................................5-1
Disposable Sensors........................................................ ...........5-2
Cleaning the Oximeter’s Surfaces ...................................................5-2
Long Term Storage..........................................................................5-2
Chapter 6: Operator’s Troubleshooting Chart ............................. 6-1
EMI Interference.............................................................................6-3
Chapter 7: Optional Supplies and Accessories............................ 7-1
Ordering Information ......................................................................7-2
Chapter 8: Specifications............................................................. 8-1
Equipment Classification ................................................................8-1
Displays, Indicators, & Keys...........................................................8-1
SpO2................................................................................................8-1
Pulse Rate.................................................. ......................................8-2
Auxiliary Printer Output..................................................................8-2
Power Requirements ............................................................ ...........8-2
Battery Life .....................................................................................8-2
Dimensions......................................................................................8-2
Environmental Specifications .........................................................8-2
Page 6
Table of Contents
Oximeter Operation and Service Manual iii
Chapter 9: Service Maintenance and Repair................................ 9-1
General Description.........................................................................9-1
Power Supply and ON/OFF Circuitry .............................................9-1
Digital Section.................................................................................9-1
Analog Section..........................................................................9-3
LED Controller..........................................................................9-4
Signal Dictionary ......................................................................9-4
Test Equipment and Tools Required...............................................9-6
Connecting a DC Power Supply......................................................9-6
Voltage Test Points..........................................................................9-7
Waveform Test Points .....................................................................9-9
Appendix: Assembly Drawings, Schematics & Parts Lists.............1
Page 7
This page is intentionally left blank.
Page 8
Chapter 1: Introduction
Oximeter Operation and Service Manual 1-1
Chapter 1: Introduction
About This Manual
The operator’s instru ct ions provide installation, operation, and mainte­nance instructions. It is intended for health-care professionals trained in monitoring respiratory and cardiovascular activity.
The service maintenance and repair section contains circuit descriptions, voltage and waveform test points, detailed parts lists, and circuit diagrams. It is intended for persons trained in service, maintenance, and repair of modern medical equipment. Thorough knowledge of this equipment’s operation is required before attempting to repair this equipment.
Proprietary Notice
Information contained in this document is copyrighted by Smiths Medical PM, Inc. and may not be duplicated in full or part by any person without prior written approval of Smiths Medical PM, Inc. Its purpose is to provide the user with adequately detailed documentation to efficiently install, oper­ate, maintain and order spare parts for the device supplied. Every effort has been made to keep the information contained in this document current and accurate as of the date of publication or revision. However, no guarantee is given or implied that the document is error free or that it is accurate regard­ing any specification.
Warranty
Limited Warranty
Seller warrants to the original purchaser that the Product, not including accessories, shall be free from defects in materials and workmanship under normal use, if used in accordance with its labeling for two years from the date of shipment to the original purchaser.
Seller warrants to the original purchaser that the reusable oximeter sensors supplied as accessories, shall be free from defects in materials and work­manship under normal use, if used in accordance with its labeling for one year from the date of shipment to the original purchaser (USA only).
Page 9
Chapter 1: Introduction
1-2 Oximeter Operation and Service Manual
Disclaimer of Warranties
THE FOREGOING EXPRESS WARRANTY, AS CONDITIONED AND LIMITED, IS IN LIEU OF AND EXCLUDES ALL OTHER WARRAN­TIES WHETHER EXPRESS OR IMPLIED, BY OPERATION OF LAW OR OTHERWISE, INCLUDING BUT NOT LIMITED TO, ANY IMPLIED WARRANTIES OF MERCHANT ABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
Seller disclaims responsibility for the suitability of the Product for any par­ticular medical treatment or for any medical complications resulting from the use of the Product. This disclaimer is dictated by the many elements which are beyond Seller’s control, such as diagnosis of patient, conditions under which the Product may be used, handling of the Product after it leaves Seller’s possession, execution of recommended instructions for use and others.
Conditions of Warranty
This warranty is void if the Product has been altered, misused, damaged by neglect or accident, not properly maintained or recharged, or repaired by persons not authorized by Seller. Misuse includes, but is not limited to, use not in compliance with the labeling or use with accessories not manufac­tured by Seller. This warranty does not cover normal wear and tear and maintenance items.
Limitation of Remedies
The original purchaser’s exclusive remedy shall be, at Seller’s sole option, the repair or replacement of the Product. THIS IS THE EXCLUSIVE
REMEDY. In no event will Seller’s liability arising out of any cause whatsoever (whether such cause is based in contract, negligence, strict liability, tort or otherwise) exceed the price of the Product and in no event shall Seller be responsible for consequential, incidental or special damages of any kind or nature whatsoever, including but not limited to, lost business, revenues and profits.
Page 10
Chapter 1: Introduction
Oximeter Operation and Service Manual 1-3
Warranty Procedure
To obtain warranty service in the USA, you must request a Customer Ser­vice Report (CSR) number from Technical Service. Reference the CSR number when returning your Product, freight and insurance prepaid, to: Smiths Medical PM, Inc., N7 W22025 Johnson Drive, Waukesha, WI 53186-1856. Telephone: 1-800-558-2345. Facsimile: 262-542-3325. Seller will not be responsible for unauthorized returns or for loss or damage to the Product during the return shipment. The repaired or replaced Product will be shipped, freight prepaid, to Purchaser.
To obtain warranty service outside the USA, contact your local distributor. Keep all original packing material, including foam inserts. If you need to
ship the device, use only the original packaging material, including inserts. Box and inserts should be in original condition. If original shipping mate­rial in good condition is not available, it should be purchased from Smiths Medical PM, Inc.
Damages occurred in transit in other than original shipping containers are the responsibility of the shipper. All costs incurred returning devices for repair are the responsibility of the shipper.
CE Notice
Marking by the symbol C
0473
indicates compliance of this device to the
Medical Device Directive 93/42/EEC. Authorized Representative (as defined by the Medical Device Directive ):
Smiths Medical International Ltd. Colonial Way, Watford, Herts, UK, WD24 4LG
Phone: (44) 1923 246434 Fax: (44) 1923 240273
Page 11
Chapter 1: Introduction
1-4 Oximeter Operation and Service Manual
Definition of Symbols
SYMBOL DEFINITION
2
Attention, consult accompanying documents.
r
Type BF equipment
7
Refer servicing to qualified service personnel.
REF
Catalog Number
SN
Serial number
R
On
Q
Off
y
Date of Manufacture
1
Non AP Device
C
Use by
Page 12
Chapter 1: Introduction
Oximeter Operation and Service Manual 1-5
Warnings, Cautions, and Notes
2
WARNING! Federal law (USA) restricts this device to sale by, or on the order of, a physician.
2
WARNING! This device is not intended for continuous patient monitoring. This device is intended to measure the patient’s %SpO
2
and pulse rate values. There are no audi-
ble or visible alarms.
2
WARNING! This device must be used in conjunction with clinical signs and symptoms. This device is only int ended to be an adjunct in patient assessment.
2
WARNING! Do not use this device in the presence of flam­mable anesthetics.
2
WARNING! Do not use this device in the presence of mag­netic resonance imaging (MR or MRI) equipment.
2
WARNING! Prolonged use or the patient’s condition may require changing the sensor site periodic ally. Change sen­sor site and check skin integrity, circulatory status, and correct alignment at least every 4 hours.
2
WARNING! When connecting this monitor to any instru­ment, verify proper operation before clinical use. Refer to the instrument’s user manual for full instructions. Acces­sory equipment connected to the monitor’s data interface must be certified according to the respective IEC stan­dards, i.e., IEC 950 for data processing equipment or IEC 601-1 for electromedical equipment. All combinations of equipment must be in compliance with IEC 601-1-1 sys­tems requirements. Anyone connecting additional equip­ment to the signal input port or signal output port configures a medical system, and, therefore, is responsi­ble that the system complies with the requirements of the system standard IEC 601-1-1.
Page 13
Chapter 1: Introduction
1-6 Oximeter Operation and Service Manual
2
WARNING! IEC 950 approved equipment must be placed outside of the “patient environment.” The patient environ­ment is defined as an area 1.5 m (4.92 feet) from the patient.
2
WARNING! When attaching sensors with Microfoam®1 tape, do not stretch the tape or attach the tape too tightly. Tape applied too tightly may cause inaccurate readings and blisters on the patient’s skin (lack of skin respiration, not heat, causes the blisters).
CAUTION! Observe proper battery polarity (direction) when replacing batteries.
CAUTION! Do not autoclave, ethylene oxide sterilize, or immerse the sensors in liquid. Evidence that liquid has been allowed to enter the monitor voids the warranty.
CAUTION! This device is intended for use by persons trained in professional health care. The operator must be thoroughly familiar with the information in this manual before using the device. (Notice: This device is not approved for home use by a non-health care profes­sional.)
CAUTION! Connect only the printer adapter specifically intended for use with this device (see Optional Supplies and Accessories).
NOTE! Operation of this device may be adv ersely af fected in the presence of strong electromagnetic sources, such as electrosurgery equipment.
1. Microfoam is a registered trademark of the 3M Company.
Figure 1-1: Patient Environment
Page 14
Chapter 1: Introduction
Oximeter Operation and Service Manual 1-7
NOTE! Operation of this device may be adv ersely affect ed in the presence of computed tomograph (CT) equipment.
NOTE! Use only SpO
2
sensors supplied with, or specifi-
cally intended for use with, this device. NOTE! SpO
2
measurements may be adversely affected in
the presence of high ambient light. Shield the sensor area (with a surgical towel, for example) if necessary.
NOTE! Dyes introduced into the bloodstream, such as methylene blue, indocyanine green, indigo carmine, patent blue V (PBV), and fluorescein, may ad versely affe ct the accuracy of the SpO
2
reading.
NOTE! Any condition that restricts blood flow, such as use of a blood pressure cuff or extremes in systemic vas ­cular resistance, may cause an inability to determine accurate pulse rate and SpO
2
readings.
NOTE! Remove fingernail polish or false fingernails be fore applying SpO
2
sensors. Fingernail polish or false finger-
nails may cause inaccurate SpO
2
readings.
NOTE! The presence of dyshemoglobins, such as carbox­yhemoglobin (with CO-poisoning) or methemoglobin (with sulfonamide therapy) may advers ely affect the accuracy of the SpO
2
measurement.
NOTE! Hazards arising from software errors have been minimized. Hazard analysis was performed to meet EN1441: 1997.
NOTE! Optical cross-talk can occur when two or more sensors are placed in close proximity. It can be eliminated by covering each site with opaque material.
Page 15
This page is intentionally left blank.
Page 16
Chapter 2: General Description
Oximeter Operation and Service Manual 2-1
Chapter 2: General Description
Intended Use
The oximeter provides fast, reliable SpO2 and pulse rate measurements. It can be used in the hospital or clinical environment, and during emergency
air or land transport. The oximeter will operate accurately over an ambient temperature range of 32 to 131° F (0 to 55° C). The oximeter works with
all BCI
®
oximetry sensors providing SpO2 and pulse rate on all patients
from neonate to adult.
Features
Provides fast, reliable SpO2 and pulse rate measurements on any
patient, from neonates to adults.
Ideally suited for use in intensive care units, in outpatient clinics,
in emergency rooms, or during emergency air or land transport.
Portable and lightweight. Weighs only 9 ounces (255 grams) with-
out the batteries.
Ergonomically designed to fit comfortably in the palm of your
hand.
Uses three standard alkaline batteries (type LR 14) or three
rechargeable (type KR27/50) NiCad “C” cell batteries.
Battery life is approximately twenty-four (24) hours in continuous
mode or eighty (80) hours in spot check mode.
Bright, easy-to-read LED displays indicate SpO
2
and pulse rate
measurements.
An eight-segment LED bar graph indicates pulse strength.
Automatically turns off after patient’s finger is removed from the
sensor.
Low battery indicator lights when about two hours of battery use
remains.
Optionally connects to an external printer or computer, providing
spot check printouts of SpO
2
and pulse rate readings.
Collects up to seventeen (17) hours of spot check data for up to
ninety-nine (99) patients for printout later.
Page 17
Chapter 2: General Description
2-2 Oximeter Operation and Service Manual
Description
PATIENT CABLE/SENSOR
The sensor connects here. If you need the extra length of the patient cable, attach the sensor to the patient cable, then attach the patient cable to the oximeter’s PATIENT CABLE/SENSOR connector. The serial printer or PC communication cable is also connected here.
% SpO
2
DISPLAY
The % SpO
2
value is shown here. Dashes ( -- ) indicate the oximeter is
unable to calculate the SpO
2
value.
PULSE RATE DISPLAY
The pulse rate value is shown in beats per minute (BPM). Dashes ( --- ) indicate the oximeter is unable to calculate the pulse rate value. Flashing 255 indicates the pulse rate value is greater than 255.
Figure 2-1: Oximeter Controls
Patient Cable/Sensor
% SpO
2
Display
Pulse Rate Display Pulse Strength Bar Graph
R and Q Keys
Low Battery Indicator
Batteries and Access Door
Page 18
Chapter 2: General Description
Oximeter Operation and Service Manual 2-3
PULSE STRENGTH BARGRAPH
The eight-segment bar graph “sweeps” with the patient’s pulse beat, indi­cating pulse strength. The bar graph is logarithmically scaled to indicate a wide range of pulse strengths.
R and Q KEYS
Press “R to turn on the oximeter. Press Q to turn off the oximeter. While the oximeter is on, momentarily pressing the R key increments the
patient number. While the oximeter is on, pressing and holding the R key for about six seconds clears all the spot check data and resets the patient number to P1.
The oximeter turns off automatically two minutes after the sensor is removed from the patient or after the sensor is disconnected from the oximeter. This feature extends the battery use time.
LOW BATTERY INDICATOR
When about two hours of battery use time remains, the left-most bar graph segment lights. The oximeter will continue to operate normally until the batteries no longer have sufficient power to operate the oximeter. At that point, the oximeter automatically turns off.
BATTERIE S AND ACCESS DOOR
The oximeter’s three “C” cell batteries are accessed through this door on the backside of the oximeter. See Installing or Replacing the Batteries for details on installing or replacing the batteries.
Page 19
Chapter 2: General Description
2-4 Oximeter Operation and Service Manual
Theory of Operation
The oximeter determines SpO2 and pulse rate by passing two wavelengths of light, one red and one infrared, through body tissue to a photodetector.
During measurement, the signal strength resulting from each light source depends on the color and thickness of the body tissue, the sensor place­ment, the intensity of the light sources, and the absorption of the arterial and venous blood (including the time varying effects of the pulse) in the body tissues.
The oximeter processes these signals, separating the time invariant parame­ters (tissue thickness, skin color , light intensity , and venous blood) from the time variant parameters (arterial volume and SpO
2
) to identify the pulse
rate and calculate oxygen saturation. Oxygen saturation calculations can be performed because oxygen saturated blood predictably absorbs less red light than oxygen depleted blood.
Figure 2-2: Oximetry Theory of Operation
Low intensity red and infrared LED light sources
Detector
Page 20
Chapter 3: Using the Oximeter
Oximeter Operation and Service Manual 3-1
Chapter 3: Using the Oximeter
Unpack the Oximeter
Carefully remove the oximeter and its accessories from the shipping car­ton. Save the packing materials in case the oximeter must be shipped or stored.
Compare the packing list with the supplies and equipment you received to make sure you have everything you’ll need.
Install the Batteries
The oximeter uses three standard “C” cell batteries (type LR 14). You can use disposable alkaline batteries or rechargeable (type KR27/50) batteries.
If you use disposable batteries, be sure to dispose of them in compliance with your institution’s guidelines and local ordinances.
If you use rechargeable batteries, it’s best to have two sets of batteries on hand. That way, you can use one set of batteries in the oximeter while the other set of batteries is recharging.
NOTE! If you’ve collected spot check data for printing, make sure you print the spot check data before removing and replacing the oximeter’s batteries. Removing the bat­teries erases spot check data from the oximeter’s memory.
Page 21
Chapter 3: Using the Oximeter
3-2 Oximeter Operation and Service Manual
Installing or Replacing the Batteries
1. Turn over the oximeter so its back is facing you.
2. Push on the thumb grip and slide the door open.
3. If you are replacing the batteries, remove the old batteries from the battery compartment.
If the old batteries are disposable, be sure to dispose of them in compli­ance with your institution’s guide­lines and local ordinances.
If the old batteries are rechargeable, be sure to charge them right away so they’ll be ready to use again as s oon as possible.
4. Install three batteries in the oximeter battery compartment. Make sure the batteries are installed in the proper direction.
5. Slide the battery door closed, push­ing firmly until it snaps into place.
NOTE! It is easiest to install the batteries in the sequenc e shown in figure 3-2.
Figure 3-1: Open the Battery Door
Figure 3-2: Install the Batteries
Page 22
Chapter 3: Using the Oximeter
Oximeter Operation and Service Manual 3-3
Care and Handling of Sensor
Hold the connector rather than the cable when connecting or disconnecting the finger sensor to the oximeter as shown in figure 3-3.
Do not use excessive force, unnecessary twisting, or kinking when con­necting, disconnecting, storing, or when using the sensor.
When placing the sensor on the patient, allow the cable to lay across the palm of the hand and parallel to the arm of the patient as shown in figure 3-4.
Upon completion of patient monitoring, detach the sensor and loosely coil finger sensor cable. Do not wrap the finger sensor cable around the oxime­ter.
2
WARNING! Misuse or improper handling of the sensor and cable could result in damaging the sensor. This may cause inaccurate readings.
Figure 3-3: Connect the Finger Sensor
Figure 3-4: Position the Sensor Cable
Page 23
Chapter 3: Using the Oximeter
3-4 Oximeter Operation and Service Manual
Choose the Sensor
Choose the appropriate sensor from the following chart. Refer to the sensor insert for application instructions.
Patient Site Description
Adult
>45 Kg
Finger 3044: Sensor, Reusab le, Adult
3444: Sensor, Reusable, Comfort Clip
®
1310: D.O.T. Reusable Oximetry Sensors
Finger or Toe 3043: Sensor, Universal “Y”
1300: Sensor, Disposable, Adult, Finger
Ear 3078: Sensor, Ear
Pediatric
15-45 Kg
Finger 3044: Sensor, Reusab le, Adult
3444: Sensor, Reusable, Comfort Clip
®
3178: Sensor, Reusable, Pediatric
Finger or Toe 3043: Sensor, Universal “Y”
1301: Sensor, Disposable, Pediatric, Finger
Ear 3078: Sensor, Ear
Infant
3-15 Kg
Hand or Foot Toe Finger or Toe
3043: Sensor, Universal “Y” 3025: Sensor, Wrap, Infant 1303: Sensor, Disposable, Infant
Neonate
<3 Kg
Hand or Foot Foot
1302: Sensor, Disposable, Neonate 3026: Sensor, Wrap, Neonate
Page 24
Chapter 3: Using the Oximeter
Oximeter Operation and Service Manual 3-5
Attach the Sensor to the Patient
Clean or Disinfect the Sensors
Clean or disinfect the reusable sensors before attaching a new patient.
Clean the sensor with a soft cloth moistened in water or a mild soap solu­tion. To disinfect the sensor, wipe the sensor with isopropyl alcohol.
Finger Sensor for Adult or Pediatric Finger
Attach the finger sensor to the patient as shown. Be sure to fully insert the patient’s finger into the sen­sor. For patients with long fingernails, use the universal “Y” sensor.
2
WARNING! Prolonged use or the patient’s condition may require changing the sensor site periodically. Change sen­sor site and check skin integrity, circulatory status, and correct alignment at least every 4 hours.
2
W ARNING! When at t aching s ensors with Microfoam® tape, do not stretch the tape or attach the tape too tightly. Tape applied too tightly may cau se in acc urat e rea din gs a nd blis­ters on the pat ient’s skin (lack of skin respiration, not heat, causes the blisters).
2
WARNING! Do not autoclave, ethylene oxide sterilize, or immerse the sensors in liquid.
2
CAUTION! Unplug the sensor from the monitor before cleaning or disinfecting.
Figure 3-5: Attach the Finger Sensor
Page 25
Chapter 3: Using the Oximeter
3-6 Oximeter Operation and Service Manual
Attach the Sensor to the Oximeter
Attach the sensor connector to the oximeter’s PATIENT CABLE/
SENSOR connector.
If you need the extra length of the patient cable, attach the sensor
to the patient cable, then attach the patient cable to the oximeter’s PATIENT CABLE/SENSOR connector.
Measuring the Patient’s % SpO2 and Pulse Rate
To begin measuring the patient’s SpO2 and pulse rate, press the “R” key. When turned on, the oximeter goes through this power-up sequence:
The pulse strength bar graph segments light one at a time.
The oximeter's software revision is momentarily displayed.
The patient number for spot check printouts is momentarily dis-
played. The format for the patient number display is “P” followed by the number. For example, P14 means the patient number is 14.
Figure 3-6: Attach the Sensor to the Oximeter
Page 26
Chapter 3: Using the Oximeter
Oximeter Operation and Service Manual 3-7
After a few seconds the % SpO2 value, pulse rate, and pulse strength bar graph should be shown. If not, see Operator’s Troubleshooting Chart for
help.
The SpO
2
display shows the patient’s blood oxygen saturation, calculated
as a percentage. The pulse rate display shows the patient’s pulse rate in beats per minute (BPM). The pulse strength bar graph shows the patient’s pulse strength; the bar graph is scaled logarithmically to indicate a wide range of pulse strengths.
Figure 3-7: SpO2 Pulse Rate and Pulse Strength Bar Graph
% SpO
2
Value Displayed
Pulse Rate Value Displayed Pulse Strength Bar Graph
Sweeping with Patient’s Pulse
Press On
Low Battery Indicator
Page 27
Chapter 3: Using the Oximeter
3-8 Oximeter Operation and Service Manual
Patient Numbers and Spot Check Data
Whenever the oximeter is on, it stores one SpO2 and pulse rate reading every thirty (30) seconds. The stored readings are called spot check data.
The oximeter remembers spot check data for up to ninety-nine (99) patients and seventeen (17) hours of run-time. The spot check data then can be printed at any time on the optional printer.
Spot check data is saved for each patient number. When you turn on the oximeter, the patient number is automatically incremented and displayed during the power-up sequence if valid spot check data was collected from the previous patient. When the patient number is incremented, then spot check data is saved for the new patient number. If no valid spot check data was collected from the previous patient, the patient number is displayed only and is not incremented. The oximeter remembers all the spot check data and all the patient numbers for up to ninety-nine (99) patients and sev­enteen (17) hours of run-time.
Manually Incrementing the Patient Number
Press the “R” key while the oximeter is on to manually increment the patient number . The new patient number is momentarily displayed and spot check data for the new patient is automatically saved.
Clearing All Spot Check Data
Press and hold the “R” key for about six seconds while the oximeter is on to clear all spot check data and reset the patient number to
P1. While you are
holding the “R” key, the message
[{r flashes on the display to notify you
that the spot check data for all patients is about to be cleared. When the spot check data is cleared, the display shows
P1.
See Spot Check Printouts for more information on printing spot check data.
Low Battery Indicator
When about two hours of battery use time remains, the left-most bar graph segment lights. The oximeter will continue to operate normally until the batteries no longer have sufficient power to operate the oximeter. At that point, the oximeter automatically turns off.
Page 28
Chapter 3: Using the Oximeter
Oximeter Operation and Service Manual 3-9
Turning Off the Oximeter
Press the Q key to turn off the oximeter. The oximeter turns off automatically two minutes after the sensor is
removed from the patient or after the sensor is disconnected from the oximeter. This feature extends the battery use time.
Checking the Oximeter’s Performance
Pulse oximeters do not require user calibration. To check the function of the device, an optional Oximeter/ECG Patient Simulator is available as an
accessory (BCI
®
Cat# 1606HH). The simulator attaches to the oximeter in
place of the sensor or patient cable. It provides a known SpO
2
and pulse
rate signal to the oximeter, allowing the oximeter's performance to be checked.
Follow the instructions included with the Oximeter/ECG Patient Simulator.
NOTE! At power up, when the sensor is disconnected and patient data is stored, the oximeter will turn off automati­cally two minutes after the oximeter executes its print patient routine.
Page 29
This page is intentionally left blank.
Page 30
Chapter 4: Printer and Computer Interface
Oximeter Operation and Service Manual 4-1
Chapter 4: Printer and Computer Interface
Computer Interface Description
The oximeter can transfer stored data to a PC. Information Data Log and Spot Checks will be displayed/saved to a file in the format as shown in the Spot Check Printouts section of this manual.
Equipment Required
1. Hand Held Oximeter
2. 3350 Printer Cable
3. 3339 PC Adapter Cable
4. Communication settings:
Interface Instructions
1. Disconnect the sensor from the oximeter.
2. Connect the printer adapter cable (part #3350) to the sensor connector
on the unit.
3. Connect 9-pin connector of PC adapter cable (part #3350) to the
printer adapter cable (part #3339). Be sure to connect the 3339 to the end of the 3350 cable marked as “printer.”
Baud rate 9600 Data bits 8 Stop bits 1 Parity none I/O Port: Serial RS-232C Data Type: ASCII Data Format: 9600 baud, 1 start bit, 8 data
bits, 1 stop bit, no parity
I/O Connector: Standard DB-9
Page 31
Chapter 4: Printer and Computer Interface
4-2 Oximeter Operation and Service Manual
4. Using the PC adapter cable (part #3339), connect the 9-pin connector of the PC adapter cable to the COM port on the PC as shown in figure 4-1.
2
WARNING! When connecting this monitor to any instru­ment, verify proper operation before clinical use. Refer to the instrument’s user manual for full instructions. Acces­sory equipment connected to the monitor’s data interface must be certified according to the respective IEC stan­dards, i.e., IEC 950 for data processing equipment or IEC 601-1 for electromedical equipment. All combinations of equipment must be in compliance with IEC 601-1-1 sys­tems requirements. Anyone connecting additional equip­ment to the signal input port or signal output port configures a medical system, and, theref ore, is responsible that the system complies with the requirements of the sys­tem standard IEC 601-1-1.
Figure 4-1: Connect the Oximeter to a PC
Connect to Oximeter
Cable (catalog # 3350)
Cable (catalog # 3339)
Printer Adapter db-9 Connector
PC Adapter db-9 Connector
Connect to IBM Compatible PC or Printer
Page 32
Chapter 4: Printer and Computer Interface
Oximeter Operation and Service Manual 4-3
Spot Check Printouts
Collecting Spot Check Data
Whenever the oximeter is on, it stores one SpO
2
and pulse rate reading
every thirty (30) seconds. The stored readings are called spot check data. The oximeter remembers spot check data for up to ninety-nine (99) patients and seventeen (17) hours of run-time. The spot check data then can be printed at any time on the optional printer .
Spot check data is saved for each patient number. When you turn on the oximeter, the patient number is automatically incremented and displayed during the power-up sequence if valid spot check data was collected from the previous patient. When the patient number is incremented, then spot check data is saved for the new patient number . If no valid spot check data was collected from the previous patient, the patient number is displayed only and is not incremented. The oximeter remembers all the spot check data and all the patient numbers for up to ninety-nine (99) patients and sev­enteen (17) hours of run-time.
Manually Incrementing the Patient Number
Press the “R” key while the oximeter is on to manually increment the patient number . The new patient number is momentarily displayed and spot check data for the new patient is automatically saved.
Clearing All Spot Check Data
Press and hold the “R” key for about six seconds while the oximeter is on to clear all spot check data and reset the patient number to
P1. While you are
holding the “R” key, the message
[{r flashes on the display to notify you
that the spot check data for all patients is about to be cleared. When the spot check data is cleared, the display shows
P1.
About the Oximeter’s Batteries and Spot Check Data
NOTE! If you’ve collected spot check data for printing, make sure you print the spot check data before removing and replacing the oximeter’s batteries. Removing the batteries erases spot check data from the oximeter’s memory.
Page 33
Chapter 4: Printer and Computer Interface
4-4 Oximeter Operation and Service Manual
Printing Spot Check Data
1. Set up the oximeter and the printer as previously described.
2. Disconnect the SpO
2
sensor
from the oximeter.
3. Turn on the printer.
4. Turn on the oximeter. The oximeter prints the spot check data for each patient, from patients 1-99, as shown in the sample printout.
5. If there is no spot check data at all, the message **** is printed.
6. The oximeter does not automati­cally turn itself off when print­ing spot check data.
7. If no valid data is collected for a patient number, **** is printed.
8. If valid data is collected for a patient number for less than one minute, only the last measure­ment is printed.
9. If data is collected for a patient number for more than one minute, the relative time since the first measurement is shown for that patient.
10. Dashes indicate invalid or unavailable data (for example, the patient’s finger was removed from the SpO
2
sensor).
Figure 4-2: Print Spot Check Data
Page 34
Chapter 5: Operator’s Maintenance
Oximeter Operation and Service Manual 5-1
Chapter 5: Operator’s Maintenance
Batteries
When about two hours of battery use time remains, the left-most bar graph segment lights. The oximeter will continue to operate normally until the batteries no longer have sufficient power to operate the oximeter. At that point, the oximeter automatically turns off.
Replace weak or dead batteries with new disposable batteries or freshly charged NiCad batteries. See Installing or Replacing the Batteries for instructions.
Disposal of batteries and rechargeable batteries
Batteries and rechargeable batteries
do not throw into fire - risk of explosion.
do not force open - danger of corrosion.
do not recharge normal batteries.
Batteries and rechargeable batteries must be treated as special waste.
dispose of in accordance with local waste disposal regulations.
For information contact the local environmental protection offices or licensed waste disposal contractors.
Sensors
Reusable Sensors
Clean the reusable sensor’s surfaces with a soft cloth moistened in a mild soap solution. If disinfection is required, wipe the reusable sensor’s sur­faces with a soft cloth moistened in isopropyl alcohol. Do not allow liquid to enter any of the sensor’s openings.
2
WARNING! Prolonged use or the patient’s condition may require changing the sensor site periodically. Change sen­sor site and check skin integrity, circulatory status, and correct alignment at least every 4 hours.
CAUTION! Do not autoclave, ethylene oxide sterilize, or immerse the sensors in liquid.
Page 35
Chapter 5: Operator’s Maintenance
5-2 Oximeter Operation and Service Manual
Disposable Sensors
Disposable sensors are for single-patient use only. Disposable sensors can be repositioned on the same patient as long as the correct SpO
2
and pulse
rate values are shown after repositioning. Disposable sensors cannot be cleaned or repaired; discard them if they become soiled or no longer work.
Cleaning the Oximeter’s Surfaces
Clean the oximeter’s surfaces with a soft cloth moistened in a mild soap solution. If disinfection is required, wipe the oximeter’s surfaces with a soft cloth moistened in isopropyl alcohol. Do not allow any liquid to enter any of the oximeter’s openings.
Long Term Storage
Remove the batteries from the oximeter before long term storage, or if the oximeter won't be used for 6 months or more. This protects the oximeter from damage due to batteries leaking acid.
Store the oximeter in its original shipping carton and packing materials to help protect the oximeter from damage during storage.
CAUTION! Do not autoclave, ethylene oxide sterilize, or immerse the oximeter in liquid.
Page 36
Chapter 6: Operator’s Troubleshooting Chart
Oximeter Operation and Service Manual 6-1
Chapter 6: Operator’s Troubleshooting Chart
Problem Possible Cause Corrective Action
No pulse shown on the bar graph.
Patient cable or sen­sor is disconnected from the oximeter. Sensor is incorrectly positioned on the patient. Poor patient perfu­sion. Defective sensor or patient cable.
Check sensor connections to the patient cable and to the oximeter.
Reposition the sensor.
Reposition the sensor.
Try a new sensor or contact your authorized repair center for help.
Pulse rate is erratic, intermittent, or incorrect.
Sensor incorrectly positioned.
Patient motion
Reposition the sensor.
SpO
2
value is
erratic, intermittent, or incorrect.
Poor patient perfu­sion.
Patient motion.
Reposition the sensor
Patient must remain still to obtain an accurate measurement.
The oximeter doesn't turn on.
Batteries weak.
Batteries not installed or batteries incorrectly installed.
Replace the batteries.
Ensure the batteries are installed correctly.
Page 37
Chapter 6: Operator’s Troubleshooting Chart
6-2 Oximeter Operation and Service Manual
The oximeter turns off unexpectedly.
The oximeter turns itself off automati­cally two minutes after the sensor is removed from the patient or after the sensor is discon­nected from the oximeter. This fea­ture extends the bat­tery use time.
Batteries are weak or dead.
None.
Replace the batteries.
No printout on optional printer.
Printer power not connected or printer power switch is Off.
Connect the printers power source and turn on the printer.
Printer interface assembly not securely connected.
Push on both ends of the printer interface assembly to ensure all the connections are tight.
Printer interface malfunction.
Contact your authorized repair center for help.
No trends in mem­ory.
Collect trend data.
Sensor must be dis­connected for spot check output.
Disconnect sensor.
E00 ROM Error
7 Refer servicing to qualified
service personnel. Contact Smiths Medical PM, Inc. Service
E01 RAM Error
7 Refer servicing to qualified
service personnel. Contact Smiths Medical PM, Inc. Service
Page 38
Chapter 6: Operator’s Troubleshooting Chart
Oximeter Operation and Service Manual 6-3
EMI Interference
The monitor is designed for use in environments in which the signal can be obscured by electromagnetic interference. During such interference, mea­surements may seem inappropriate or the monitor may not seem to operate correctly.
The monitor generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with these instructions, may cause harmful interference with other devices in the vicinity. Disruption may be evidenced by erratic readings. Cessation of operation, or other incorrect function. If this occurs, the site of use should be surveyed to determine the source of this disruption, and actions taken to eliminate the source:
Turn equipment in the vicinity off and on to isolate the offending equipment.
Reorient or relocate the other receiving device.
Increase the separation between the interfering equipment and this equipment.
If assistance is required, contact Smiths Medical PM, Inc. Service Depart­ment or you local representative.
2
CAUTION! This device has been tested and found to com­ply with the limits for medical devices to the IEC 601-1­2:1993, EN 60601-1-2:1994, Medical Device Directive 93/42/ EEC. These limits are designed to provide reasonable pro­tection against harmful interference in a typical medical installation. However, because of the proliferation of radio­frequency transmitting equipment and other sources of electrical noise in the heath-care and home envir onments (for example, cellular phone, mobile two-way radios, elec­trical appliances), it is possible that high levels of such interference due to close proximity or strength of a so urce, may result in disruption of performance of this device.
Page 39
This page is intentionally left blank.
Page 40
Chapter 7: Optional Supplies and Accessories
Oximeter Operation and Service Manual 7-1
Chapter 7: Optional Supplies and Accessories
Cat. No Description Quantity
1300 Sensor, Oximetry, Disposable, Adult Finger 10/box 1301 Sensor, Oximetry, Disposable, Pediatric,
Finger, 15-45 Kg
10/box
1302 Sensor, Oximetry, Disposable, Neonate,
< 3 Kg
10/box
1303 Sensor, Oximetry, Disposable, Infant,
3-15 Kg
10/box
1310 D.O.T. Reusab le Oximetry Sensors 3/box 1311 D.O.T. Sensor Replacement Tape 24/pkg 1606HH Simulator & Cable, Oximeter, 5 ft. each 1818R Manual, Operation (English and Spanish)
and Service (English)
each
3025 Sensor, Oximetry, Wrap, Infant, 3-15 Kg each 3026 Sensor, Oximetry, Wrap, Neonate, < 3 Kg each 3043 Sensor, Oximetry, Universal “Y” each 3044 Sensor, Oximetry, Finger each 3049 Strips, Adhesive 40/pkg 3078 Sensor, Oximetry, Ear each 3134 Tap e, Attachment, Neonatal 50/pkg 3135 Tap e, Attachment, Infant 50/pkg 3136 Tape, Attachment, Neonatal 100/pkg 3137 Tape, Attachment, Infant 100/pkg 3138 Posey Wrap, Attachment, Universal "Y" 10/pkg 3178 Sensor, Oximetry, Pediatric Finger each 3311 Cable, Patient, 5 ft. each 3315 Case, Carrying each 3318 Protective Boot with Strap each 3339 PC Adapter Cable each 3444
Sensor, Oximetry, Finger, Comfort Clip
®
each
3350 Cable, printer interface each
Page 41
Chapter 7: Optional Supplies and Accessories
7-2 Oximeter Operation and Service Manual
Ordering Information
For ordering information, contact your local distributor or the Smiths Med­ical PM, Inc. customer service department.
Smiths Medical PM, Inc.
N7 W22025 Johnson Drive
Waukesha, WI 53186 USA
Phone: 262 542 3100
Fax: 262 542 3325
Email Address: info@smiths-bci.com
Page 42
Chapter 8: Specifications
Oximeter Operation and Service Manual 8-1
Chapter 8: Specifications
Equipment Classification
Displays, Indicators, & Keys
SpO
2
Type of Protection Against Electric shock: Internally Powered Mode of operation: Continuous Degree of Protection Against ingress of Liquids: IPX1, drip proof Degree of Mobility: Portable Degree of Protection Against Electric Shock: Type BF Safety Requirements: EN60601-1; 1990
SpO
2
: LED numeric display, 0.46 inches (11.68 mm) high
Pulse Rate: LED numeric display, 0.38 inches (9.65 mm) high Pulse Strength: Logarithmically scaled 8-segment LED bar graph Keys: R & Q keys
Range: 0-99% Accuracy: ±2% at 70-99% Averaging: 8 pulse beat average Display Response The display is to functional saturation. The pulse
strength bar graph is not proportional to pulse vol­ume.
Display Update Rate
1 Hz (SpO
2
); 60 Hz (pulse strength)
Calibration Factory calibrated over range 70% to 100% SpO
2
using human blood samples to functional saturation. Test methods available upon request. No in-service calibration required.
Sensors Red 660nm, 2mW (typical)
Infrared 905nm, 2-2.4mW (typical)
Page 43
Chapter 8: Specifications
8-2 Oximeter Operation and Service Manual
Pulse Rate
Auxiliary Printer Output
Data saved every thirty (30) seconds can be printed (spot check).
Power Requirements
Uses three standard alkaline (type LR 14) or nickel-cadmium “C” cell bat­teries (type KR27/50).
Battery Life
Approximately 24 hours in continuous mode or 80 hours in spot check mode.
Dimensions
Environmental Specifications
Range: 30-254 BPM Accuracy: ±2% at 30-254 BPM Averaging: 8 second average Display Update Rate 1 Hz
Width: 3.25 inches (82.6 mm) Height: 6.3 inches (160 mm) Depth: 1.25 inches (31.75 mm) Weigh t: 9 ounces (255 grams) without the batteries
Operating Temp.: 0-55° C (32-131° F) Storage Temp.: -34-70° C (-29-158° F) Relative Humidity: 10-95% (storage), non-condensing
15-95% (operating), non-condensing
Page 44
Chapter 9: Service Maintenance and Repair
Oximeter Operation and Service Manual 9-1
Chapter 9: Service Maintenance and Repair
The service maintenance and repair section contains circuit descriptions, voltage test points, and waveform test points. The Appendix contains detailed parts lists and circuit diagrams. It is intended for persons trained in service, maintenance, and repair of modern medical equipment. Thorough knowledge of this equipment’s operation is required before attempting to repair this equipment.
General Description
The oximeter contains one circuit board. This board contains all power supply, microprocessor, date acquisition, and display circuitry.
Power Supply and ON/OFF Circuitry
The power supply is built around a switch mode regulator IC U10. It is configured as a fly-back regulator to convert input voltage in the range of
2.7-4.5 VDC into +5VDC output voltage. Negative voltage used by the analog section is generated as a “by-product” of the main +5VDC supply. Charge pump C55-D4-C69 is used to convert pulsatile voltage on the out­put LX of U10 into negative voltage in the range of –4VDC to –6VDC. The V-RAM voltage used by the on-off circuitry and RAM is generated either from the battery, or from VCC when unit is on. V-RAM is taken straight from the battery due to the maximum voltage being 4.5V. V-RAM may approach +3VDC as the battery power gets depleted, however, this voltage is high enough to retain trend data in RAM. Capacitor C70 is used to maintain V-RAM when batteries are being replaced. It can keep infor­mation in RAM for more than 1.5 minutes.
On/Off circuitry is built around U7A-U7B flip-flop, which directly con­trols regulator IC U10.
Digital Section
Microprocessor U11 is Hitachi HD64180. Its address and data buses are connected to the One Time Programmable Read Only Memory (OTP ROM) U12 and Random Access Memory (RAM) U24. Memory Address Decoder is implemented using U17A demultiplexer chip.
Page 45
Chapter 9: Service Maintenance and Repair
9-2 Oximeter Operation and Service Manual
Memory address map is shown in the following table:
Display Select Signal DSP-LD is used to latch displayed data into U22 and U23 LED drivers.
Address decoder for Input/Output Interface devices is implemented using U17B demultiplexer chip.
The I/O address space is divided according to the following table:
Watch dog timer U18 is used to reset microprocessor in case of ESD and at power up. Its internal comparator is used to compare battery voltage offset by Vr voltage coming from digitally co ntrolled potentiometer with internal U18 voltage reference. The output of the comparator (signal BATT) is used to detect battery voltage. The following battery states are distinguished by this measurement system:
RS-232 communication interface is built around Q3, Q5, and Q6. It con­verts CMOS voltage levels into RS-232 bipolar voltage levels. Inverter U16C is used to disable communication interface circuitry when oximetry probe is plugged in to reduce power consumption and eliminate cross-talk between communication and probe reading.
Device Address Space
OTP ROM 0..FFFF RAM 10000..1FFFF Display Select Signal 20000..2FFFF
I/O Address Space Device
0..3F Used by Z180 for internal registers
40..7F Input port
80..BF Output port
Battery Voltage Condition
2.7V Dead battery level-units shuts itself off
3.0V Low battery level-signal to user
4.5V Ok battery level
Page 46
Chapter 9: Service Maintenance and Repair
Oximeter Operation and Service Manual 9-3
Internal timer 1 of U11 is used as a tone generator to drive piezo speaker through connector J5.
Analog Section
Analog to Digital Converter (ADC) U6 is a 12 bit, successive approxima­tion sampling ADC. The ADC+IN signal input comes from TP22. VREF is used here as a pseudo ground. The ADC is configured as a differential input converter with the –IN input connected to VREF. This allows the +IN signal to range from 2.5V to 5V. The ADC transfers data with a 3 wire serial interface. ADCLK synchronizes the data transfer with each bit being transmitted on the falling ADCCLK edge and captured on the rising ADCLK edge. /ADCS must beheld low when the ADC transmits data.
LED Drive circuitry is used to turn on and off the oximetry probe LED’s, precisely controlling the current through them. The second half of the vari­able potentiometer U5 is used to generate voltage between 0 and 2.5VDC. U5 is controlled, serially, by the processor. Operational amplifier U2B together with transistor Q7 create a constant current sink. The current is proportional to the voltage generated by variable potentiometer. The H­bridge Q8-Q9 is used to activate either LED, red or infrared, inside the oximeter probe. The following table describes the states of the H-bridge:
In this table, 1 indicates logic 1, 0 indicates logic 0, and x indicates either logic 0 or 1.
The “Open circuit” state is used to check for a possible probe cable fault. In case of probe cable fault, one of the LED wires can short to the ground shield causing high current through the probe LED. To prevent this, before turning any LED on, the processor checks the cable by “floating” the H­bridge. If the cable is shorted CON3 and CON2 connections are pulled low , and comparator U9A generates /PRB_FAULT signal, which causes the processor to shut down the LED excitation cycle and generate the appropri­ate message.
RED-DRV IR-DRV FET-ON Function
101Red LED is on 011Infrared LED is
on
X X 0 “Open circuit”
state
Page 47
Chapter 9: Service Maintenance and Repair
9-4 Oximeter Operation and Service Manual
The differential photo amplifier formed by U1 and U3A, converts the pho­todetector’s current output to a voltage at T10 (VAMB). Amplifier U3B offsets the signal at T10 allowing a wider signal range. Comparator U9B is used to inform the microprocessor if U3A is saturated by an excess of ambient light. VAMB is passed through blocking capacitor C77 to remove the signal’s DC component. The signal is then buffered and amplified by U27A. Gain is defined by the first channel of digital potentiometer U5. The output of U27A is routed to the integrator-filter-of fset circuitry U27B, con­trolled by analog switch S53. The output of the integrator is then routed to the ADC for measurement.
LED Controller
LED controller is built around U22 and U23 buffers. Data is latched into buffers by DSP-LD signal. Signal/DSP-EN is used to disable LED’s at power up to limit initial surge current.
Signal Dictionary
This section lists, in alphabetical order, the signal names used on the sche­matics. The signal’s origin, destination, and purpose are described.
Signal Description
ADCLK ADOUT /ADCS
3 wire serial interface for analog to digital converter U6.
/AMB­FAULT
Signal informing processor about excess of ambient light.
ANA+5 Filtered VCC. ANA+5 powers the analog circuitry. ANA-5 ANA-5 po wers the analog circuitry. BASE Keeps integrator U27B fro saturating during data acquisitions. BATT Signal from watchdog timer U18 to inform processor of low bat-
tery condition.
CAP­GND
Signal used to short blocking capacitor C77 to ground.
CTS0 Serial Port Clear To Send handshake signal from external RS-232
communication device.
Page 48
Chapter 9: Service Maintenance and Repair
Oximeter Operation and Service Manual 9-5
CON2, CON3
Probe LED’s driving signals.
/DSP­EN
Enable display drivers signal.
DSP-LD Signal used to load data into display drivers. FET-ON Used to control MOSFET H-bridge Q8 and Q9, which powers
probe LED’s. INTGAT Signals used to enable integrator U27B. IR-DRV Used to control MOSFET H-bridge Q8 and Q9, which powers
probe LED’s. /NMI Non-Maskable Interrupt that signals watch dog timer time-out. PON Power on signal. CLK Clocks data into digital potentiometer chip U5. POT-LD POT-LD is used to select digital potentiometer chip U5. SDI Serial Data Input to digital potentiometer chip U5. PRB-
DET
PRB-DET is used to inform processor if probe is plugged in.
/PRB­FAULT
Signal used to inform processor about probe cable problem.
QUIET Signal used to shut down power down supply during data acquisi-
tion. RX External RS-232 receive (CTS0) line. RED-
DRV
Used to control MOSFET H-bridge Q8 and Q9, which powers
probe LED’s. RESET System reset signal generated by the watch dog tim e r U18 . RST-
INT
Resets integrator U27B by shorti n g cap acit o r C50.
SHT­DWN
Output of the microprocessor to turn unit off.
TX External RS-232 transmit line. /TEN Signal which enables RS-232 driver. Is connected to PRB-DET
signal.
Page 49
Chapter 9: Service Maintenance and Repair
9-6 Oximeter Operation and Service Manual
Test Equipment and Tools Required
To diagnose and repair the full extent of possible malfunctions on the oximeter and display board, you will need the following test equipment and tools:
DMM volts/ohms/amps, with M input impedance or greater.
Oscilloscope, 50 MHz, with 10 M input impedance or greater.
Variable output DC power supply, 0-5 VDC at 200 mA or greater.
Small Phillips screwdriver
Small flat blade screwdriver
Needle nose pliers
Diagonal cutters
Clip leads
Low-power microscope or magnifying glass
Soldering iron and solder
Solder wick or solder remover
Connecting a DC Power Supply
To verify so me of the voltage measurements that follow, you’ll need to connect a variable output DC power supply in place of the three “C” cell batteries. You must be careful to observe the polarity of J9 when doing this. Follow these steps to connect a DC power supply in place of the “C” cell batteries:
TXA0 Asynchronous serial communication transmit signal to RS-232. VAMB VAMB is the output of the front-nd differential amplifier. V-BT Battery voltage VR Output voltage from the digitally controlled potentiometer. VREF 2.5VDC reference to voltage. V-RAM Voltage used to maintain trend information in RAM. VCC,
VDD
5VDC used for digital output supply voltage.
/WDT Watch dog timer strobe
Page 50
Chapter 9: Service Maintenance and Repair
Oximeter Operation and Service Manual 9-7
1. Remove the three “C” cell batteries.
2. Connect the DC power supply’s “+” output to J9-1.
3. Connect the DC power supply’s “-” or “ground” output to J9-2.
Voltage Test Points
Unless otherwise noted, all voltages are measured with respect to ground at P1-2.
CAUTION! Be sure to observe the polarity of the battery connections. Failure to observe the polarit y may cause fus e F1 to blow.
Location Condition Nominal and Range
T19
Oximeter on. 5 VDC, ± 0.1 VDC ANA+5 (T18) Same as T19. Same as T19. FLTR+5 Same as TP14. Same as T19. ANA-5 Same as TP15. Same as TP15. FLTR-5 (T16) Same as TP15. -4.8 VDC to –6.2 VDC V-REF (T17) Oximeter on. 2.5 VDC V-RAM Oximeter on.
Oximeter off.
T19 minus 0.7 VDC.
V_BT minus 0.7 VDC.
U11-71 Oximeter on. 3.072 MHz, 50% duty
cycle.
Low battery LED. When battery voltage
drops to 3.00 VDC.
Low battery LED first turns on at 3.00 VDC, ±0.2 VDC.
Low battery shut off. Oximeter on, battery
voltage slowly drop-
ping.
Oximeter shuts off when battery voltage drops to 2.7 VDC, ± 0.2 VDC.
Page 51
Chapter 9: Service Maintenance and Repair
9-8 Oximeter Operation and Service Manual
Automatic shut off. Turn on oximeter and
place finger in sensor. Remove finger from sensor.
Remove sensor from oximeter. Turn on oximeter.
Oximeter should turn off 2 minutes (± 5 sec­onds) after finger is removed from sensor.
Oximeter should turn off in 2 minutes (± 5 seconds).
Current draw through
J9-1 (total battery cur­rent).
Oximeter on, sensor attached to oximeter, no finger in the sensor, with battery voltage at
3.5 VDC, ±0.2 VDC.
110 mA, ± 15 mA
Page 52
Chapter 9: Service Maintenance and Repair
Oximeter Operation and Service Manual 9-9
Waveform Test Points
The following oscilloscope screens show typical waveforms from the sen­sor detector’s amplifier and data acquisition circuits. Unless otherwise noted:
The waveforms are measured with respect to ground at J9-2.
The oscilloscope is triggered on T2, falling edge.
The waveforms are measured with the finger sensor attached to the oximeter, but no finger inserted in the sensor.
Connect CH1 probe to T10, T21, T22, T24. Verify the signal shown in the respective figures.
Measurement at T2 2V/DIV
0.4mS/Div
0V
Page 53
Chapter 9: Service Maintenance and Repair
9-10 Oximeter Operation and Service Manual
Measurement at T22 2V/DIV
0.4mS/Div
0V
Measurement at T24 2V/DIV
0.4mS/Div
0V
Page 54
Chapter 9: Service Maintenance and Repair
Oximeter Operation and Service Manual 9-11
Remove all scope connections from the oximeter board. Functional Testing Disconnect the oximetry test fixture. Turn oximeter board off and then on
again. Connect Finger probe to oximeter board and put your finger in the probe. Verify SpO
2
and Pulse Rate readings are displayed.
Verify that the pulse tone is heard when the pulse strength bar lights up. Remove finger from finger probe. Verify “dashes.”
Measurement at T24 2V/DIV
0.4mS/Div
0V
0V
Page 55
This page is intentionally left blank.
Page 56
Appendix
Oximeter Operation and Service Manual Appendix-1
Appendix
Assembly Drawings, Schematics & Parts Lists
Number Description Item
Number of pages
71000A23 F/ASM OXIMETER REDESIGN
BOM
A-1 1
71000A23 F/ASM OXIMETER REDESIGN A-2 1 71264B1 PWB ASM MAIN BOARD
REDESIGN
A-3 1
71264B1 PWB ASM MAIN BOARD
REDESIGN PARTS LIST
A-4 3
71264S1 MAIN BOARD REDESIGN
SCHEMATIC
A-5 5
Page 57
This page is intentionally left blank.
Page 58
Page 59
Smiths Medical PM, Inc.
Patient Monitoring and Ventilation
m
Authorized Representative (as defined by the Medical Device Directive ):
Smiths Medical International Ltd. Colonial Way, Watford, Herts, UK, WD24 4LG
Phone: (44) 1923 246434 Fax: (44) 1923 240273
Loading...