Items manufactured by companies other than Smith-Root carry the original manufacturer’s warranty. Please contact product manufacturer for return instructions.
All Smith-Root, Inc. manufactured products are covered by a one year warranty.
Credit & Refund Policy: Customers returning equipment, in new condition, will be given credit to the applicable VISA/Master
Card account within five days from the date of the return. A return authorization must accompany returns. Valid equipment
returns include, but are not limited to, ordering incorrect equipment, funding deficits, and defective equipment returned for
reimbursement. All returns are subject to a restocking fee and applicable shipping charges. The restocking fee is figured at 10%
of the purchase price but not less than $20.00. Customers receiving equipment in damaged condition will be referred to the
shipping company for insurance reimbursement.
Rev.02
U.S. Patent 5,305,711
WARNING - Battery posts, terminals and related accessories contain lead compounds,
chemicals known to the state of California to cause cancer and reproductive harm. Wash
hands after handling.
Iwww.smith-root.com
USER’S GUIDE
INTRODUCTION
The
PES® Portable Electroanesthesia Systemis
a portable device intended for the generation of electrical energy to be used for the
electrical stunning of aquatic organisms.
It is designed for use in field applications, in hatcheries and on boats.
The PES®
Electroanesthesia
System consists of:
• PES® Control Unit
• Anesthesia Tank (with
electrodes)
• Electrode cables
• Remote Switch
• Battery Charger
• Manual
The unit must be connected to an electrical outlet to charge the battery prior to use.
®
The PES
The anode and cathode plates sit on either end of the anesthesia tank. Fish to be
anesthetized are introduced to the water between the anode and cathode.
The PES® is operated by means of a remote Switch. A bright LED flasher and audible
alarm provide a clear indication of when output is present.
is connected to the anesthesia tank via the anode and cathode cables.
ATTENTION!: THE PES® IS DESIGNED FOR BATTERY OPERATION
®
ONLY! The PES
operation. Operation of the PES® while connected to an electrical outlet
may damage internal components leading to failure of the system. Also,
The device must be OFF when the battery is being charged.
12015
should NOT be connected to an electrical outlet during
PES: PORTABLE ELECTROANESTHESIA SYSTEM
ON/OFF
UNIT IS ACTIVE WHEN LIGHT IS FLASHING
0-Off
I-On
RECHARGE INPUT
REMOTE OUTLET
26.4V DC3 Amps
TURNRIGHT TOACTIVATE
PUSH OFF
Select
Save
Setup
Load
Setup
Menu
Timer
Volts
Duty
Cycle
Freq.
Pulse
Type
PORTABLE ELECTROANESTHESIA SYSTEM
PES
PORTABLE ELECTROANESTHESIA INSTRUCTIONS
INSTRUCTIONS FOR CHARGING THE BATTERY
• Connect the external charging system to the PES via the Recharge Input plug.
• Connect the external charging system to an AC electrical outlet.
• Check the indicator lights on the external charging system to ensure that charge is being delivered.
• Disconnect the charging system from the PES and AC outlet after the battery is fully charged.
OPERATING INSTRUCTIONS
The PES is designed for battery operation only. The PES should NOT be connected to an electrical outlet during operation.
Red button – Device ON/OFF. For on, twist the button to the right. Press the button to turn the unit off.
LOAD SETUP – Review and select dose. Press LOAD SETUP, use the 5and 6buttons to review the electrical
parameters to be applied. Press SELECT to load the settings for use. The dose to be delivered is programmed by the user and is defined by PULSE TYPE, TIME, VOLTS, FREQ, and DUTY CYCLE.
Use the 5and 6buttons to review the parameters available for use. Use SELECT to load settings into memory.
Note: Any changes to the dose requires the user to review of ALL dose parameters (PULSE TYPE, TIME, VOLTS, FREQ, DUTY CYCLE). Use the 5and 6buttons to review the parameters and SELECT to load the parameters into memory. Display messages will prompt user for the required input. After all parameters have been reviewed, press SAVE SETUP, use the 5and 6buttons to select the memory location (page) where the newly defined dose will be saved, press SELECT. Up to 10 doses can be saved into memory.
ON/OFF
0-Off
I-On
RECHARGE INPUT
REMOTE SWITCH
& RECHARGE INPUT
TURNRIGHT TOACTIVATE
PUSH OFF
GETTING STARTED
BEFORE YOU USE THE PES®:
The PES® must be fully charged before you can use it. Make sure that the unit is OFF by
verifying that the large red knob is in the downward position, as shown in the diagram
on the front panel.
Charge the battery using the following instructions. Once the battery is fully charged,
you may begin using the PES® for anesthetizing fish.
CHARGING THE BATTERY
Prior to using the PES®,
connect the unit to the
battery charger using the
following procedure:
• Connect the external
charging system to an
AC electrical outlet (top
left).
• Connect the external
charging system to the
PES® via the Recharge
Input plug (top right).
• Check the indicator
light* on the external
charging system to
ensure that charge is
being delivered (bottom
right).
• Disconnect the charging
system from the PES
and AC outlet after the
battery is fully charged.
®
PES® connected to AC
outlet via Battery Charger
*INDICATOR LIGHTS
• RED indicates that the
battery is charging at full
power
• YELLOW indicates that
the battery is charging at
normal charging power
• GREEN indicates that the
battery is fully-charged
and ready for use
Indicator
light
2www.smith-root.com
USER’S GUIDE
PES
PORTABLE ELECTROANESTHESIA INSTRUCTIONS
INSTRUCTIONS FOR CHARGING THE BATTERY
• Connect the external charging system to an AC electrical outlet.
• Connect the external charging system to the PES via the Recharge Input plug.
• Check the indicator lights on the external charging system to ensure that charge is being delivered.
• Disconnect the charging system from the PES and AC outlet after the battery is fully charged.
OPERATING INSTRUCTIONS
The PES is designed for battery operation only. The PES should NOT be connected to an electrical outlet during operation. Red button – Device ON/OFF. For on, twist the button to the right. Press the button to turn the unit off.
LOAD SETUP – Review and select dose.
Use the 5and 6buttons to review the parameters available for use. Use SELECT to load settings into memory.
Note: Any changes to the dose requires the user to review of ALL dose parameters (PULSE TYPE, TIME, VOLTS, FREQ, DUTY CYCLE). Use the 5and 6 buttons to review the parameters and SELECT to load the parameters into
FRONT PANEL ITEMS
Press LOAD SETUP, use the 5 and 6 buttons to review the electrical
parameters to be applied. Press SELECT to load the settings for use. The dose
to be delivered is programmed by the user and is defined by PULSE TYPE,
TIME, VOLTS, FREQ, and DUTY CYCLE.
memory. Display messages will prompt user for the required input. After all
parameters have been reviewed, press SAVE SETUP, use the 5 and 6 buttons
to select the memory location (page) where the newly defined dose will be
saved, press SELECT. Up to 10 doses can be saved into memory.
C
ON/OFF
ON/OFF
TURN
TURN
RIGHT TO
RIGHT TO
ACTIVATE
ACTIVATE
PUSH OFF
PUSH OFF
RECHARGE INPUT
RECHARGE INPUT
REMOTE SWITCH
26.4V DC3 Amps
& RECHARGE INPUT
REMOTE OUTLET
E
I-On
0-Off
26.4V DC
3 Amps
A
I-On
0-Off
UNIT IS ACTIVE WHEN LIGHT IS FLASHING
PORTABLE
ELECTROANESTHESIA
SYSTEM
Save
Setup
Freq.
Load
Setup
Duty
Cycle
Select
Menu
Timer
Pulse
Type
Volts
D
J
U
S
T
PUSH
DOWN
TO
UNLOCK
PUSH
DOWN
TO
UNLOCK
A
ANNUNCIATOR
ANODE
CATHODE
PES® front panel items
PUSH TO
PULL UP
UNLOCK
Locking anode and cathode connectors
Remote switch and
battery charger share
a common, electrically
isolated plug (below)
REMOTE SWITCH
DEVICE ON/OFF (A)
The large red button located on the left side
of the front panel turns the device on and off.
For on, twist the button to the right. Press the
B
button to turn the unit off. The battery charger
must be disconnected from the PES® prior to
turning the device on.
ANNUNCIATOR (B)
D
The annunciator provides an audible warning
that electrical energy is being output by the
®
. Volume of the audible warning can be
PES
adjusted by rotation of the annunciator. A test
beep sounds at power up.
INDICATOR LIGHT (C)
A flashing light (located in the center of the
front panel) provides a visual warning that
electrical energy is being output by the PES®.
A test flash is given at power up.
ANODE AND CATHODE CONNECTORS (D)
The receptacle labeled ANODE is the
connection for the lead attached to the
positive plate of the exposure tank. The
receptacle labeled CATHODE is for connection
of the lead attached to the negative plate of
the exposure tank.
The anode and cathode cables use locking
connectors. To unlock them, press downward
to disengage the locking mechanism and then
grasp the body of the plug and pull upwards.
The remote
switch
features a
momentary
push-button.
(above)
32015
REMOTE SWITCH/RECHARGE CONNECTOR
(E)
Activation of the remote switch initiates the
output of electrical energy. The remote switch
must remain activated for continued output.
Deactivation of the remote switch terminates
output. Termination of electric current
®
output by the PES
may also be controlled
automatically via the dose timer.
The Remote Switch and the Recharger share a
common connector, although their respective
conductors are isolated (refer to CHARGING
THE BATTERY for complete instructions).
PES: PORTABLE ELECTROANESTHESIA SYSTEM
HOOK-UP
®
SETTING UP YOUR PES
ANODE
CATHODE
ANODE
CATHODE
ARRAY
ANODE
ARRAY
ANESTHESIA
TANK
REMOTE
OUTLET
REMOTE SWITCH
HOOK-UP DIRECTIONS
1. Ensure that the PES® is OFF by pressing the red button
and disconnect the unit from the battery charger.
2. Connect the RED cable to the red-ringed output
connector on the control panel labeled ANODE.
3. Connect the BLACK cable end to the black output
connector labeled CATHODE.
4. Connect the cables to the electrode assemblies on
either side of the anesthesia tank using the connectors.
It is unimportant which one is cathode or anode, but
both cable ends must be electrically isolated from one
another.
CATHODE
Insert anode and cathode
connectors firmly
Connect anode and cathode
cables to each electrode
REMOTE OUTLET
5. Connect the Remote Switch cable to the 4-pin
connector labeled REMOTE OUTLET.
IMPORTANT!: DO NOT CONNECT BATTERY CHARGER TO UNIT WHEN
OPERATING THE UNIT! It is intended for charging the battery only. It
cannot power the unit during operation. Charge the battery with unit
shut down only. Also, make sure you only use the supplied battery
charger with the PES®. Using an incorrect charger will damage the unit.
Plug in the remote switch
4www.smith-root.com
USER’S GUIDE
ANESTHESIA TANK
The PES® is available with
a 150 quart tank (Part #
09879)
The electroanesthesia
tank has two electrode
assemblies. One is
permanently fixed to one
end of the tank and the
other may be placed at
varying distances using
the guides within the tank
interior.
The output cables
are attached to each
electrode by means of the
corresponding connectors.
It’s not important which
electrode is assigned
to either ANODE or
CATHODE, as both are
electrically isolated
from one another. Both
electrodes are enclosed
within a non-conductive
plastic screen to prevent
direct accidental contact
with the electrode surface.
CARE & MAINTENANCE
OF THE ELECTRODES
It may become necessary to
clean the electrodes after
periods of extended use.
Over time, the electrodes
accumulate a build-up
of oxidation, making
settings unpredictable
and shortening battery life
between recharges. To clean
the electrodes, remove the
protective screen covering
the electrode plate. On the
movable electrode unit,
carefully remove the two
plastic screws on the outer
edges of the top handle
assembly. The electrode
and handle can be slid out
of the screen assembly. On
the fixed electrode unit,
carefully remove the four (4)
screws that hold the screen
to the electrode and put
the screen aside. Remove
the two (2) stainless steel
screws holding the electrode
plate to the top mounting
bracket.
Remove any oxidation
with an abrasive pad
or sandpaper. It is not
recommended that
you use any cleaning
chemicals or solvents, as
they may remain on the
electrode assemblies and
unintentionally contaminate
the electroanesthesia tank.
Assemble both units in
reverse order, taking care
not to damage or crossthread the plastic screws.
52015
PES: PORTABLE ELECTROANESTHESIA SYSTEM
ELECTROANESTHESIA TANK
STORAGE POUCH
Inside of the lid of the
control box is a pouch for
the storage of the output
cables, remote switch and
manual for the PES®. It is
constructed of tough, water
resistant nylon and features
a hook-and-loop closure
system on the protective
flap. Care should be taken
not to overfill the pouch
with too many items as it
could interfere with the
closing of the lid.
The pouch may be removed
for cleaning and reattached
using the hook-and-loop
retention system inside the
lid.
While we have taken great
care to ensure that the
PES® is safe to use near
water, it is not designed to
be used in the rain, where
it could be splashed with
water and/or should never
be immersed in water.
Never use the PES® if it is
wet. Dry unit off with a dry
cloth before operating.
If necessary, wipe control
panel surface with a moist
towel. Solvents or cleansers
may damage the control
panel or controls. Avoid
spilling liquids on the
panel as they may damage
internal components.
OVERFLOW
STANDPIPE
Included with the PES®
anesthesia tank is a standpipe that allows users to
replenish the fresh water in
the tank by running a hose
into it, while letting the
over-flow water vent safely
out the tank’s drain hole.
The 90˚ elbow is attached
to the drain from the
inside of the tank and the
resulting overflow goes out
the top of the standpipe,
while maintaining desired
water level. The amount of
water in the tank can be
adjusted accordingly using
different lengths of vertical
pipe.
To attach the standpipe to
the outlet drain, press the
elbow over the drain tube
until it can’t go any further.
It is important that the
standpipe is not in place
when the PES® tank is in
transit. Otherwise, it may be
damaged by items shifting
around inside the tank
during shipping./
Electrode
Screen
MOVEABLE ELECTRODE
ElectrodeMetal Electrode
Plastic
screen
mounting
screws
FIXED ELECTRODE
Screws
Screen
6www.smith-root.com
USER’S GUIDE
OPERATIONS
INTERACTING WITH THE PES
Once the unit is turned on, a series of buttons and the centrally located
display on the front panel of the device are used to interact with the PES®
and to program the electrical doses to be output.
LOAD SETUP, SELECT, SAVE
SETUP
The LOAD SETUP button is used to review
the various electrical parameters defining
a dose. The PES® is pre-programmed with
10 stored, predefined doses (See PREPROGRAMMED DOSES, SPECS.).
After pressing LOAD SETUP, the and
buttons are used to scroll through
the predefined doses (defined by sets of
electrical parameters), and the SELECT
button is used to load the set of parameters
for use. These parameters can be modified
by the user and, if desired, these new
parameters can be stored for recall.
The dose to be delivered is programmed
by the user and is defined by PULSE TYPE,
TIME, VOLTS, FREQ, and DUTY CYCLE.
72015
Use the and buttons to review the
parameters and SELECT to load the
parameters into memory. Display messages
will prompt the user for the required input.
After all parameters have been reviewed,
press SAVE SETUP, use the and
buttons to select the memory location
(page) where the newly defined dose will
be saved, press SELECT. Use the SAVE
SETUP button to save the settings. Up to
10 sets of parameters (doses) can be saved
into memory.
Note: Because there is a potential
for interactions among the electrical
parameters defining a “dose”, a review of
ALL dose parameters is required when a
change is made to any of the parameters
defining a dose (PULSE TYPE, TIMER,
VOLTS, FREQ, DUTY CYCLE).
®
PES: PORTABLE ELECTROANESTHESIA SYSTEM
UNIT IS ACTIVE WHEN LIGHT IS FLASHING
ANNUNCIATOR
ANODE
CATHODE
A
D
J
U
S
T
Select
Save
Setup
Load
Setup
Menu
Timer
Volts
Duty
Cycle
Freq.
Pulse
Type
PORTABLE
ELECTROANESTHESIA
SYSTEM
PES
Use the 5and 6buttons to review the parameters available for use. Use SELECT to load settings into memory.
Note: Any changes to the dose requires the user to review of ALL dose parameters (PULSE TYPE, TIME, VOLTS, FREQ, DUTY CYCLE). Use the 5and 6buttons to review the parameters and SELECT to load the parameters into memory. Display messages will prompt user for the required input. After all parameters have been reviewed, press SAVE SETUP, use the 5and 6buttons to select the memory location (page) where the newly defined dose will be saved, press SELECT. Up to 10 doses can be saved into memory.
PULSE TYPE
The “PULSE TYPE” button provides
options for type of electric current output
by the PES®. After pressing PULSE TYPE
use the and buttons to scroll through
the choices. When the desired output is
displayed on the screen, press SELECT to
load the option into memory.
Options for output: Ramped DC, Standard
pulsed DC, and Bursts of pulses.
RAMPED-DC
This is when the output ramps up
from zero to maximum (voltage level
programmed by the user) over a period of
1 second. This term applies to continuous
DC only and not to pulsed waveforms.
When using the dose TIMER, upon
completion of the exposure, the ramped
DC output will also decrease back to zero
over a period of 1 second. Total exposure
time includes the periods for ramping
the output up and down; minimum total
exposure time is 2 seconds.
When output is controlled manually (via
remote switch), upon activation of the
remote switch the output current starts
low and takes 1 second to reach the
selected maximum; if the output voltage
was set for 100 volts the output would take
1 second to raise the voltage to 100 volts
level. The output will remain at this level
until interrupted by release of the switch.
Release of the remote switch deactivates
the output immediately. Example: If the
pulse timer is disabled when using Ramped
DC, the DC output increases from 0 to
the prescribed voltage level over a period
of 1 second, when the remote switch is
activated. The output will remain at this
level until the remote switch is deactivated
(released). The output immediately
decreases to zero upon release of the
remote switch.
OPERATIONS
PES® Control panel with central display
If the pulse timer is enabled, the output
increases from 0 to the prescribed voltage
level over a period of 1 second, remains
at that level for exposure time minus 1
second, then decreases to 0 over a period
of the final second.
If, however, the remote switch is released
before the exposure-time has elapsed the
output is immediately turned off.
STANDARD PULSED DC
Output immediately goes to maximum
voltage and stays constant for the entire
pulse duration, then drops to zero. This
type of pulse is known as “rectangular”.
BURST OF PULSES
Output is a series of dual-frequency
rectangular pulses. This type of output may
be referred to as dual-frequency pulsed
DC, gated burst pulsed DC, or Complex
Pulsed System. The pulses occur at two
frequencies, an inner (burst) frequency and
an outer (cycle).
8www.smith-root.com
USER’S GUIDE
CONTROLS & SETTINGS
VOLTS
The amplitude of the PES® output can be
varied between 20 and 400 volts, in 5 volt
increments. After pressing VOLTS, use the
and buttons to scroll through the
available voltages. Press SELECT to load
the voltage setting into memory.
FREQ
Multiple options available are for the
frequency of the pulsed DC output by
the PES®. Pulse frequency can be varied
between 5 Hz and 100 Hz, in increments of
5 Hz, and then 250Hz, 500Hz and 1000Hz.
After pressing FREQ, use the and
buttons to scroll through the frequency
options. Press SELECT to load the desired
frequency setting into memory.
When the Pulse-Type is Burst-of-Pulses
the Burst Frequency (inner frequency)
is displayed first followed by the Cycle
(outer) Frequency. Each frequency must be
selected individually by using the and
buttons.
When the Pulse-Type is Standard Pulse,
only the Cycle Frequency is displayed.
When the Pulse-Type setting is RampedDC, the message “Not used in DC” is briefly
displayed, the screen then returns to the
previous menu.
DUTY CYCLE
Pressing DUTY CYCLE displays the
calculated “percentage of time” for the
selected pulse type and frequency.
When the Pulse-Type is Standard Pulse, the
calculated on-time (the ratio of the length
of the pulse to the period of the waveform)
is displayed. Options for duty cycle vary
between 1 and 99%. After pressing DUTY
CYCLE, use the UP () and DOWN () to
scroll through the available options. Press
SELECT to load the desired setting into
memory.
When the Pulse-Type is Burst-of-Pulse,
the duty cycle of the two frequencies is
displayed. The duty cycle of the inner burst
is displayed first, followed by the frequency
of the outer cycle and the duty cycle of the
CYCLE frequency is displayed second.
When the Pulse-Type is Ramped-DC the
message “Not used in DC” is displayed.
TIMER
Termination of the output of the PES® can
be automated with the TIMER or controlled
manually. The TIMER can be adjusted from
1 to 255 seconds or “disabled”.
The output of the PES® is always controlled
by the hand switch.
When PULSE TYPE is STANDARD PULSES
or GATED BURST the minimum time for
output is 1 second and the maximum is 255
seconds.
When PULSE TYPE is Ramped-DC the
minimum time is 2 seconds and the
maximum time is 256 seconds: One second
is needed to ramp up the voltage and one
second to ramp it back down.
DUAL MODE
The PES® can be configured to output
two different doses, in sequence. In
this configuration, a secondary dose is
delivered following a primary dose, when
the primary waveform TIMER elapses on
a single activation of the remote switch.
Each of the phases of the exposure the
primary phase and the secondary phase
are programmed by the user using the
PULSE TYPE, DUTY CYCLE, FREQ,
92015
PES: PORTABLE ELECTROANESTHESIA SYSTEM
TIMER, and VOLTS buttons, as previously
described for a single phase output.
Releasing the remote switch at any time
will stop the output of the PES.
In Dual-Mode the prefix “Pri” is displayed
for the Primary dose setting. Pressing
“SELECT” advances to display the prefix
“Sec” for the secondary dose. Pressing
“SELECT” again sets the displayed option
for the Secondary dose type.
Dual mode requires the pulse TIMER
for the primary waveform. Disabling the
primary waveform timer disables the dual
mode.
CONTROLS & SETTINGS
DEFAULT SETTINGS
The PES® comes with 10 sets of
parameters pre-programmed as default
doses. The individual parameters of the
doses can be changed. The 10 storage
locations (identified on the screen) can be
reviewed, changed, and stored in memory
for future use, by the user.
To restore the factory default settings
press the “MENU” button, press the “UP”
arrow to cycle to “System Menu 9/10”,
“Restore factory default settings?” and
then press “SELECT”. Press the “UP” arrow
to confirm the operation.
When powered on, the unit initializes with
the dose settings previously loaded by the
user. The dose settings to be used should
be reviewed prior to operation of the unit.
10www.smith-root.com
USER’S GUIDE
PES
®
MENU GUIDE
1 STATUS DISPLAY
On Startup and at the end of various other
menu operations the PES® displays the current
Status. There are 3 or 4 Status displays. If a
menu button is pressed while in another menu
the PES® returns to the status display. The
default is to display the current waveform
settings such as these:
The up or down arrow (or) will cycle the to
one of the other Status displays like these:
2 LOAD SETUP
There are 10 settings in memory. Pressing
LOAD SETUP will display one of these settings.
The setting number is shown in the upper
right corner of the display. The up and down
arrows will cycle through all settings from 0
to 9. Pressing SELECT will load the displayed
setting, display a message to that effect and
return to the Status Display.
Note: For the Dual mode settings the Primary
(Pri) waveform is displayed first regardless of
which arrow is pressed. The Secondary (Sec)
wave form is only displayed if the up arrow is
pressed while viewing the Primary waveform
setting.
Example settings:
3 SAVE SETUP
In the save setting menu only the setting
number is display in the upper right corner.
Pressing SELECT will overwrite the setting
at that memory location with the current
waveform parameter settings. The display will
then return to a Status Display.
Example:
4 TIMER
The TIMER button allows the pulse timer to be
changed. When the pulse TIMER is enabled
the output is only active for the set number of
seconds after the output switch is pressed and
then the output automatically stops. With the
Wave Time disabled the output remains active
for as long as the output button is held on.
Pressing the Timer button displays the current
setting which can be changed by pressing the
up or down arrow until the desired setting is
displayed and then pressing SELECT. After
pressing SELECT the display will then return to
Status Display.
Like most wave settings menus if Dual mode
waveform is enabled the first display will
show the Primary waveform setting for that
parameter. Pressing SELECT, whether or not a
change is made, will switch the display to the
Secondary waveform setting.
112015
PES: PORTABLE ELECTROANESTHESIA SYSTEM
PES® MENUS
Note: For dual waveform output to function
the Dual mode setting must be enabled and
the Primary Wave Timer must not be disabled.
The PES® switches from the Primary waveform
to the secondary waveform when the Primary
Wave Time expires. Even if Dual mode is
enabled the output will not switch if the
Primary Wave Time is disable. The Secondary
Wave Time can be disabled. In that case the
output will switch from the Primary waveform
to the Secondary waveform when the Primary
Wave Time expires and the PES® will continue
to output the Secondary waveform for as long
as the REMOTE SWITCH is held.
5 PULSE TYPE
Press the PULSE TYPE menu button to select
one of the three output waveform types;
Ramped DC, Standard (Rectangular Pulses), or
Burst of Pulses. On pressing the PULSE TYPE
menu button the current setting is displayed.
Pressing the up or down arrows changed
the display and pressing the SELECT button
changes the setting to the one displayed and
return the display to Status Display.
If Dual mode waveform is enabled the first
display will show the Primary Waveform pulse
type. Pressing SELECT will set the type for the
Primary waveform and the display will then
show the Secondary waveform type. Pressing
SELECT again would set the Secondary
waveform type and the display will return to
Status Display. Graphical display are shown
only for example and are in no specific order.
6 VOLTS
Pressing the VOLT menu button displays the
current voltage setting. The up and down
arrows change the voltage value in the display
and the SELECT button sets that voltage for
the waveform. If Dual mode is enabled the
display first shows the Primary wave voltage.
Pressing SELECT sets the voltage for the
Primary and switches the display to show the
Secondary waveform voltage setting which
can be changed with the arrow buttons and set
with the SELECT button.
7 FREQ.
Pressing the FREQ. button displays the current
frequency setting, if the pulse type is set for
DC, in which case the display moves to the
next waveform of a Dual mode setting or
briefly displays an error message and then
returns to Status Display. With some frequency
displays, the bottom line of the display also
shows the calculated width of the pulse in
milliseconds using the current setting for Duty
Cycle. The displayed waveform frequency
setting can be varied up or down with the
up or down arrows and the setting of the
waveform is changed to that displayed value
when the SELECT button is pressed. The range
of frequencies may be limited by the current
Duty Cycle setting, as the PES® has limits for
pulse width.
If BURST OF PULSES wave type is set, the
first display in the frequency menu is for the
Burst Frequency. Pressing SELECT sets that
frequency and switches the display to the
Cycle Frequency display where it can be raised,
lowered or left unchanged. Pressing SELECT
again would then set the Cycle Frequency and
the display would return to Status Display.
12www.smith-root.com
USER’S GUIDE
If Dual mode is enabled the first display shows
the frequency setting of the Primary waveform.
If that waveform is a Burst of Pulses type the
display would show the setting of the Burst
Frequency of the Primary Waveform otherwise
the display will simply show the Cycle
Frequency. Changes to the displayed setting
can be made with the up or down arrows.
Pressing SELECT would save that setting and
change the display to the next menu item.
Menu items follow the following order
depending on the Dual mode setting and
the Pulse Type settings. The display cycles
to the next menu item when SELECT is
pressed.
• Cycle Frequency > Status Display
• Burst Frequency > Cycle Frequency
>Status Display
• Primary Cycle Frequency > Secondary
Cycle Frequency > Status Display
• Primary Burst Frequency > Primary Cycle
Frequency > Secondary Cycle Frequency
> Status Display
• Primary Cycle Frequency > Secondary
Burst Frequency > Secondary Cycle
Frequency > Status Display
• Primary Burst Frequency > Primary Cycle
Frequency > Secondary Burst Frequency
> Secondary Cycle Frequency > Status
Display
• Primary Cycle Frequency > Status Display
• Secondary Cycle Frequency > Status
Display
• Primary Burst Frequency > Primary Cycle
Frequency > Status Display
• Secondary Burst Frequency > Secondary
Cycle Frequency > Status Display
If one of the waveform types was set to
Ramped DC in the last four sequences, the
menu state that would have displayed and
that set’s waveform frequency would be
skipped.
Note that in DUAL MODE, if the Primary
wave Pulse Type is set to Ramped DC, the
frequency menu starts with the frequency
setting of the Secondary waveform. The
following graphical displays are shown only
for example and are in no specific order.
8 DUTY CYCLE
The DUTY CYCLE button is use to view
and set the duty cycle, percent of the
wave cycle (where the output is at the
set voltage). If the pulse type is BURST
OF PULSES, the DUTY CYCLE button
sets the number of pulses in a burst. The
pulse width, calculated using the current
frequency setting, is shown on the last line
of the DUTY CYCLE display.
If the wave type is set for DC and DUAL
MODE is not enabled or the secondary
wave type is also DC, an error message is
shown briefly before returning to Status
Display.
If the wave type is set for STANDARD
PULSE than the current waveform’s duty
cycle setting is displayed.
Changes to the displayed setting can be
made with the arrow buttons, limited by
the currently set frequency, and the PES®’s
132015
PES: PORTABLE ELECTROANESTHESIA SYSTEM
PES® MENUS
pulse width limitations. The SELECT button
saves that display setting and the display
returns to the Status Display.
If the current pulse type is BURST OF
PULSES, the first display shows the number
of pulses per burst. The number shown can
be changed with the arrow keys. After the
parameter is set with the SELECT button,
the display will show the current duty cycle
setting. This displayed setting can also be
changed with the arrow keys and selected
with the SELECT button.
As in other menus if Dual mode is enabled
the display will show the setting(s) of the
Primary waveform, which can be changed
with the arrow keys. After pressing the
SELECT button, the Secondary wave’s
setting will be displayed. After pressing
the SELECT button for the secondary
waveform’s duty cycle, the display returns
to Status Display. The menu cycles through
the following items based on the setting
of Dual mode and Pulse Type(s) with each
press of the SELECT button.
• Primary Pulses/Burst > Primary Duty
Cycle > Status Display
• Secondary Duty Cycle > Status Display
• Secondary Pulses/Burst > Secondary
Duty Cycle > Status Display
The last four cycles occur when either the
Primary or the Secondary wave type is set
for DC. Graphical display are shown only for
example and are in no specific order.
9 MENU BUTTON
The Menu Button allows access to a set of
system operating parameters. The menu
always starts with the same parameter
and cycles forward with the up arrow
and backward with the down arrow. In
most menu items the display shows the
parameter and the setting change that
would occur if the SELECT Button is
pressed. To leave the setting unchanged
cycle to another menu parameter with the
arrow buttons or abort the System Menu
by pressing the menu button again and
the display will return to Status Display. If
the SELECT button is pressed the system
setting is changed a message may be
briefly displayed before the display returns
to the Status Display.
14www.smith-root.com
USER’S GUIDE
SYSTEM MENU ITEMS ARE (IN ORDER):
1. TIMER RESET - Pressing SELECT in this
menu will reset the seconds counter in
the Status Display to 0.
2. QUICK SETUP - Pressing SELECT will
initiate the Quick Setup sequence.
3. DUAL MODE - Pressing SELECT will
toggle the current setting of Dual Mode;
enabling or disabling it.
4. POWER LIMIT - Press SELECT to enter
the power limit menu where the max
power setting can be changed. Pressing
Select in this menu will return the display
back to Status Display.
5. LIMIT MODE - In Power Limiting
or Thermal Fold-back the power is
reduced by changing the voltage or
the frequency, while retaining the pulse
width. This menu allow the selection of
which method us used. Note: Frequency
limiting is not possible with a DC Pulse.
8. EXTENDED TESTS - Pressing SELECT
in this menu item will initiate a basic
test of the PES® unit. This should only
be performed either by or under the
direction of a engineer or technician.
9. FACTORY DEFAULTS - Pressing SELECT
in this menu item will set the current
waveform and all stored waveforms to
the Default Factory Setting.
10. HISTORY - Press SELECT in this menu
item to access the event history of the
PES® unit. Press the up or down arrow
buttons to go forward or backward in
history. Press SELECT or Menu to return
to the Status Display
6. THERMAL FOLDBACK - If the PES® gets
too hot this mode, if enabled, will reduce
the output power like Power Limit mode.
Press SELECT in this menu to toggle the
current setting of this mode: Enabling or
Disabling it.
7. LOW OUTPUT CURRENT - As a safety
feature; If the PES® output is activated
but the unit does not detect any
significant output current the output
will be shut off and an error message
displayed. In this Menu the test for Low
Output Current can be Disabled. Press
SELECT in this menu to toggle this
system setting; Disabling or Enabling it.
152015
PES: PORTABLE ELECTROANESTHESIA SYSTEM
PRE-PROGRAMMED DOSES
SYSTEM MENUS
SETTING 0: Ramped-DC 100 VOLTS
With the activation of the remote switch
the unit outputs a low voltage that raises
to 100 volts in 1 second and stays at that
level until the remote switch is released.
SETTING 1: Standard 100-Volt Pulse of
30Hz with a 25% duty
The PES® outputs 100 volts pulses 30
times a second. The pulse period is 33 1/3
milliseconds so the output voltage is 100
volts for 8 1/3 milliseconds and off for 25
milliseconds.
SETTING 2: Standard 100-Volt Pulse of
60Hz with a 25% duty
60 time each second the unit outputs 100
volts for approximately 4.17 milliseconds
with 12.5 milliseconds off time between
pulses.
SETTING 3: Standard 100-Volt Pulse of
15Hz with a 25% duty
15 pulses per second of at 100 volts for
16 2/3 milliseconds and output off for 50
milliseconds.
SETTING 4: Standard 100-Volt Pulse of
90Hz with a 25% duty
Approximately 2.78 milliseconds of 100
volts followed by no output for about 8.33
milliseconds 95 time each second.
SETTING 5: Standard 100-Volt Pulse of
100Hz with a 25% duty
100 volts for 25 milliseconds and 75
milliseconds off time 100 time each
second.
SETTING 6: Burst of 3 100-Volt pulses of
250Hz with a 62% duty at 15Hz
3 pulses at 100 volts for 2.48 milliseconds
with 1.52 milliseconds between pulses
repeated 15 time each second.
SETTING 7: Burst of 3 100-Volt pulses of
500Hz with a 62% duty at 30Hz
3 pulses of 1.24 milliseconds at 100 volts
with 0.760 off time between pulses
repeated 30 times each second.
SETTING 8: Burst of 3 50-Volt pulses
of 250Hz with a 62% duty at 15Hz for 1
second followed by a Burst of 3 100-Volt
pulses of 250Hz with a 62% duty at 15Hz
for 2 seconds.
3 pulses at 50 volts for 2.48 milliseconds
with 1.52 milliseconds off time repeated
15 times a second for 1 second and then
followed by similar pulses at 100 volts for 2
seconds and then the output stops.
SETTING 9: Burst of 3 100-Volt pulses
of 250Hz with a 62% duty at 15Hz for 2
second followed by a Burst of 3 50-Volt
pulses of 250Hz with a 62% duty at 15Hz
for 1 seconds.
3 pulses at 100 volts for 2.48 milliseconds
with 1.52 milliseconds off time repeated
15 times a second for 2 second and then
followed by similar pulses at 50 volts for 1
seconds and then the output stops.
16www.smith-root.com
USER’S GUIDE
SAFETY CHECKLIST:
Before operating unit, make sure that:
• At least 2 PEOPLE on the crew know how to administer first aid for
electric shock.
• Unit is clean and dry.
• All cables are in good condition, with no cracks or abrasions and that all
connectors are in place with no missing or damaged parts.
• Safety control switches are in good condition and functioning properly.
• Indicators and alarms are functioning properly.
• All personal protective equipment is in good condition.
• All electrical connections in the system are clean and tight.
• Electrodes are clean and in good condition, with no loose or damaged
parts.
• Treatment tanks clean and in good condition, with no leaks or damaged
areas.
• Fish handling system is clean and in good condition, with no worn or
damaged parts.
• Power outlet is in good condition and is properly wired.
• Work area is clear, with no trip hazards.
• Floor has a non-slip surface.
• ALL CREW MEMBERS know where the nearest phone is and the
emergency services number is clearly posted by the phone.
172015
PES: PORTABLE ELECTROANESTHESIA SYSTEM
SPECIFICATIONS
PES
®
SPECIFICATIONS
Conductivity Range:
Input Voltage:
Input Current:
Input Monitoring:
Output Voltage:
Output Current:
Output Waveforms:
Output Frequency:
Duty Cycle:
Waveform Storage:
Output Power:
Overload Protection:
100 to 1500 microsiemens
25.9VDC Nominal
20 Amps Max.
Input voltage and current plus fuel gauge type display for
battery voltage
20 to 400V in 5V steps
40A peak max. 4A continuous at 100V
Ramped DC, Pulsed DC, Burst of Pulses DC
5 to 100Hz in 5Hz steps, 250, 500, and 1000Hz
1 to 99% in 1% steps
Save voltage, frequency, duty cycle, and pulse type for 10
different waveforms
400W max. continuous, 16,000W peak max.
Excessive peak current, average current, or temperature
will shutdown the unit before damage can occur. Resets
automatically when condition is corrected
Output Indicators:
Output Timer:
Dose Timer:
Safety Devices:
Dimensions:
Weight:
Audio tone with increasing pulse rate for output power, Flashing
red light, Status display for output voltage both average and
peak, output current both average and peak and output power
0 to 999,999 seconds reset-able via menu.
2 to 256 seconds in 1 second steps
Low output current sensing, Remote safety hand switch,
Emergency stop switch and Electronic circuit breaker
19.63”W X 15.5”H X 8.75”D 49.86mm X 39.37mm X 22.23mm
28.5lbs 12.96kg
18www.smith-root.com
USER’S GUIDE
PULSED DC WAVEFORMS
The Pulse Width of these waveforms are
the duty cycle times the Period. For a
standard waveform the duty is multiplied
by the Cycle Period but for a Burst
Waveform the duty is multiplied by the
Burst Period. For the Standard wave the
Pulse Width would be 0.25 times 200
milliseconds which is 50 milliseconds.
The Pulse Width for the Burst Waveform
would be the duty cycle time the Burst
Period, 0.50 times 50 milliseconds or 25
milliseconds.
Figure 1
Figure 1 shows a graphic representation
of two pulsed DC waveforms. The top
waveform is a Standard Pulsed Waveform
with a Cycle Frequency of 5 Hertz and a
25% duty cycle. The bottom waveform
depicts a Burst of 3 pulses with a burst
frequency of 20Hz, a duty cycle of 50%
and a cycle frequency of 5 Hertz.
Other characteristics of a Pulsed DC
Waveform are the Period and the Pulse
Width.
The Period of a waveform is simply
the inverse of the Frequency. For the
Standard Pulse in Figure 1 the period is
1-second divided by 5-Hertz which is 200
milliseconds. The Cycle period of the Burst
Waveform of Figure 1 is the same as the
cycle period of the standard waveform:
200 milliseconds. The Burst Period is
1-second divided by the burst frequency of
20-Hertz which make the Burst Period 50
milliseconds.
192015
When selecting the parameters for a
waveform it is best to determine, in
advance, the setting that will produce
the desired output. When changing
the settings of an existing waveform,
consider the side effect that a change,
such as frequency, will have on another
characteristic, such as pulse width. Note
also that when configuring for a burst
of pulses that the cycle period must be
greater than the product of the burst
period time the number of pulses. (While
in theory the relation may be greater or
equal, but the resultant waveform would
then be identical to a standard waveform.
The Burst waveform in Figure 1 could
conceivably have 4 pulses in each burst
but that would make it a standard wave at
20hz and in reality the timing granularity
may not allow it.)
PES: PORTABLE ELECTROANESTHESIA SYSTEM
USER’S GUIDE
APPENDIX
CYCLE FREQUENCY
Following is a chart that shows the maximum number of pulses in a burst for
combinations of Cycle Frequency and Burst Frequency.