Contents Page
1 Table of contents 1
2 Introduction2
3 Before Starting
Guide to these Operating Instructions 3
Intended Use 3
Set -Up and Function 4
Safety instructions 5
Installation 5
Start –Up 6
4 Basic Functions
Programming Aids 7
Configuration of the device 8
Configuration of the printer 8
Data Input 9
Activation of the printing data 10
Special functions 11
Operation and validation 12
5 Troubleshooting and Maintenance
Troubleshooting Chart 13
Customer Service 13
Maintenance Chart 15
Parts Service 15
Parts Required for Maintenance and Wear 16
Replacement parts 17
Part numbers 18
Maintenance Information
Replacing ink ribbon 19
Replacing Teflon strip on guide rail 20
Replacing Teflon strip for heating die 21
Replacing pressure roller 22
Setting 23
Adjustment
Service date 24Standby function 24
Activate default settings function 24
6 Technical Data
Circuit and Wiring Diagram 230 V 25
Circuit and Wiring Diagram 115 V / 100 V 25
Specifications 26
9.696.038
ENGLISH
2
GB 2
SGR30S
Operating Instructions
Introduction
First we would like to thank you for purchasing the sealing machine.
These instructions include information on the machine, its function and operation.
! Please read these operating instructions thoroughly before using the machine so that you will
be familiar with the capabilities of the machine and can utilize its features to the optimum.
FAlways keep these instructions in the vicinity of the machine.
Important note:
This device is a packaging machine for the processing of thermoplastic films and composite films
according to the thermal – bonding or sealing process.
In this respect , please pay attention to the “Intended use” chapter in the operating instructions.
According to the intended use , the CE symbol was affixed on the basis
of the EU Directives named below:
89/336/EEC, 98/37/EC
and 73/23EEC amended by 93/68EEC.
The 93/42EEC Directive is not applicable to bonding and sealing devices.
The limit values stipulated in IEC 60601 – 1 cannot be applied in case of the electrical repetition tests.
The manufacturer will not assume any liability for damage caused by tests according to standards
which are not listed in the declaration of Conformity .
Note
Since we continuously improve our products we
reserve all rights to modify these operating instructions
and the functions described therein.
We make all possible efforts to keep these instructions
free of errors and complete. If you note any errors or
unclear points please advise us.
9.696.038
ENGLISH
3
GB
3
SGR30S
Before Starting
Guide to these Operating Instructions
Before Starting
Basic Functions
Troubleshooting
and
Maintenance
Technical Data
Intended Use
The machine is intended for commercial and industrial applications and should
be used only for the intended purpose.
The machine is a sealing machine for sealing sterilization packages
consisting of paper, paper laminate foil or Tyvek foil.
! It is not suitable for sealing pure polyethylene foils, soft PVC foils, hard PVC foils,
polyamide foils or polypropylene foils.
The machine capacity depends on the characteristics of the sealing material used.
This chapter contains information on the functions of the machine
and instructions for handling and installation.
BE SURE TO READ!
Adjustment and operation of the machine are explained here.
Instructions for finding causes of malfunctions and eliminating
them.
This chapter includes information on the required maintenance,
parts subject to wear and replacement parts.
Shows the interrelationship between various item numbers, the
circuit and diagram plans and machine specifications.
9.696.038
ENGLISH
3
GB
4
SGR30S
Before Starting
Set-Up and Function
Power switch
Power fuse
Power connection
Peel edge
adjustment
5 - 30 mm
Display
Input and
function keys
Sealing and Printing Operation Sequence
1. After inserting the sterilization package switch on transport.
2. The sterilization package is then transported and the area of the sealing seam is heated to the
sealing temperature set on the upper and lower heating dies.
3. The hot seam is presse d together by the sealing roller sealing it.
4. The printing operation is activated and the activated print data is printed onto the sealed package.
5. The finished sterilization package is transported to the discharge side.
6. If more packages to be sealed are not fed in the transport mechanism shuts off after
approx. 30 s.
9.696.038
ENGLISH
3
GB
5
•
Œ
SGR30S
Before Starting
General safety instructions
The device must not be installed or operated by persons under
14 years of age.
The device must not be operated unattendedly.
Installation
Do not install or operate the machine in explosion hazard areas.
Set up the machine only in dry areas.
Heavy dust, vapor, dripping or spraying water have a negative effect
on the machine functions.
I
I
I
Turn of the device or disconnect the power cord when it is not in use.
The device should only be cleaned with a dry or slightly damp cloth.
Warning! Never clean the device wet!
Use only power sockets with protective ground.
Always ensure that the operating power corresponds to the specifications
on the machine rating plate.
FPlease choose a socket where the line power is stable.
Please do not transport device on the peel edge setting!Œ
Leave a gap of 200 mm between the device and the wall. •
200 mm
9.696.038
ENGLISH
3
D
EU
10-03-08 09:30
GB
6
SGR30S
Start-Up
Before Starting
l Plug power cable into power socket
on machine
l Insert plug into wall socket
l Switch on machine
with power switch
FUNCTION DISPLAY
Machine switched on
Self-test runs for approx. 30 s
Default settings
When the device is switched on
for the very first time, the
following defaults must be
undertaken:
Language
also refer to page 8
Measurement unit
also refer to page 8
Date and time
also refer to page 9
Display for approx. 5 seconds:
V069.A27 = Program version
440001 = Machine number
If maintenance deadline is
exceeded
Acknowledge maintenance
message
Perform maintenance
Display of
date, time
Actual temperature.
Setemperature
After 5 minutes is the
set temperature reached
Machine is ready for operation
Please wait
set Language
Unit of measure
dd-mm-yy hh.mm
Control SGR30S V069.27
M A I N T E N A N C E
17-02-08 14:30
025°C
17-02-08 14:35
140°C
440001
Power switch
Switch
connection
9.696.038
4
GB
7
ENGLISH
SGR30S
Basic Functions
Programming Aids
FUNCTION KEYS DISPLAY
INPUT AIDS
Cursor control
Press - cursor right - key once
INPUT AIDS
Changing numbers and
Letters
Press key – position +1 - once
INPUT AIDS
Return to previous menu
point without activating entry
INPUT AIDS
Activate input
Cursor Cursor
1x
left right
Position +1 Position -1
1x
100°C
5Cursor flashes at 1st digit
100°C
5Cursor flashes at 2nd digit
100°C
51st digit is number 1
200°C
51st digit is number 2
INPUT AIDS
Space at selected point
Set piece counter to 0
9.696.038
GB
0 - 99
C
4
GB
8
ENGLISH
SGR30S
Basic Functions
Configuration
FUNCTION KEYS DISPLAY
CONTRAST SETTING
DATA INPUT
Selection: Language,
symbols,
temperature unit
Language:
D, GB, F, E, I, SF, S, PL,CZ,TR
Symbols: yes/no
Temperature unit: C / F
è
è
Language selection
yes / no
Celsius / Fahrenheit
Contrast é
Contrast
display - contrast
set language
pictogramm NO
temperature unit
Printer Set-Up
FUNCTION KEYS DISPLAY
PRINTER SET UP
Printing direction
0°
180°
Start printing
0 - 99 mm
Character width
0 – 5
Character
interval
0 – 5
èYes / No
Input
0 / 180
Input
Start printing
Input
Character width
Input
Character interval
printer set up
YES
print out
0
distance
00 0-
character width
0 0-
character distance:
0 0-
9.696.038
Loading...