Smeg PAD2R User manual [GB]

PAD2
PAD2, PAD2-60
PAD2X, PAD2X-60
DEMINERALISED WATER
INSTALLATION AND INSTRUCTIONS MANUAL
TRANSLATION OF THE ORIGINAL INSTRUCTIONS
PAD2/PAD2X
INSTRUCTIONS MANUAL
2
READ CAREFULLY
This document refers to the Smeg accessories “PAD2, PAD2-60, PAD2X and PAD2X-60”. The accessories can only be installed and used on Smeg devices. In order for the accessory to be used properly and safely, this manual must be used in conjunction with the manual of the device, instrument washer or glassware washer in which it is installed. Failure to read, understand or correctly interpret the instructions contained in this manual may cause damage to the appliance, and as well as being a source of danger for the operator, this may considerably reduce the performance of the machine.
The manufacturer declines all liability for uses other than those indicated in this manual. Installation, maintenance and repairs must only be carried out by personnel authorised by the manufacturer.
The guarantee may become void if the device is used in a way other than that indicated by the manufacturer.
This manual is for informational purposes only. The contents of this manual and the equipment described herein may be subject to modification without prior notice.
INFORMATION AND ASSISTANCE FOR PRODUCTS FROM THE
SMEG INSTRUMENTS DIVISION
Our Sales Office staff will be happy to provide you with all the information you may require regarding prices and offers. Our Technical Assistance Department can provide you with instructions about how to operate your appliance correctly or if necessary, put you in contact with your nearest Authorised Assistance Centre.
19 390 5338 08
15/11/2019
Rev.
Date
PAD2/PAD2X
INSTRUCTIONS MANUAL
3
Contents
1 DESCRIPTION .............................................................................................................................................................. 4
2 COMPATIBILITY BETWEEN THE ACCESSORY AND DEVICES ........................................................................................ 5
3 PACKAGE CONTENTS .................................................................................................................................................. 6
4 INSTALLATION INSTRUCTIONS .................................................................................................................................... 7
4.1 PAD2 – PAD2X on the GW6090/6290 ............................................................................................................... 7
4.1.1 Mounting on GW6090 .................................................................................................................................. 7
4.1.2 Mounting on GW6290 .................................................................................................................................. 8
4.2 PAD2X on the GW3060 and the GW4090 ......................................................................................................... 8
4.3 PAD2/PAD2X on the 6010, 7010 and the 7015 ................................................................................................. 9
4.3.1 EXTERNAL ACCESSORIES ELECTRICAL CONNECTION ENCLOSURE ................................................................ 9
4.3.2 PAD ELECTRICAL CONNECTION ................................................................................................................... 13
4.3.3 PAD2 PRESENCE ACTIVATION ..................................................................................................................... 15
5 ELECTRICAL JUNCTION BOX AND CONNECTIONS ..................................................................................................... 15
6 PRECAUTIONS FOR USE AND MAINTENANCE ........................................................................................................... 16
7 MANUFACTURER’S DETAILS...................................................................................................................................... 16
PAD2/PAD2X
INSTRUCTIONS MANUAL
4
1 DESCRIPTION
PAD2, PAD2X
Booster pumps for non-pressurised demineralised water. These allow the machine to be supplied with demineralised water drawn from a non-pressurised tank placed on the floor. This pump ensures that the demineralised water enters the machine with sufficient pressure for the machine to work properly. The capacity of the tank must be adequate for the machine that is installed. Consider both the type and characteristics of the products when determining the correct size of the tank and the system for the production of the demineralised water. Contact our After-Sales Service for further details.
CHARACTERISTICS OF THE PRODUCTS
PAD2
PAD2X
Pump with painted cast-iron body and brass impeller.
Pump with stainless steel and engineering thermoplastics (fibreglass reinforced PPO) body.
The capacity of the tank must be adequate for the machines supplied by it. Consider both the number and the characteristics of the products when determining the correct size of the tank and the system for the production of the demineralised water, in order to calculate the required flow rate correctly. Contact our After-Sales Service for further details.
PAD2/PAD2X
INSTRUCTIONS MANUAL
5
2 COMPATIBILITY BETWEEN THE ACCESSORY AND DEVICES
The PAD2 and PAD2X accessories are installed external to the appliance. On some models, the accessory can be installed only in conjunction with the PAD2R auxiliary kit, as specified in the following tables. (e.g. on GW1160 it is necessary to install the PAD2 with the kit PAD2R. On GW3060, the auxiliary kit is not required because the machine is already set up for the PAD2 to be installed).
The accessory can be installed on the following devices:
GW RANGE – LABORATORY GLASSWARE WASHER
MACHINE MODEL
PAD model that can be installed
AUXILIARY KIT REQUIRED
GW0160
PAD2/PAD2X
PAD2R
GW1060
PAD2/PAD2X
PAD2R
GW1160
PAD2/PAD2X
PAD2R
GW3060
PAD2/PAD2X
-
GW4060
PAD2/PAD2X
PAD2R
PW4060
PAD2/PAD2X
PAD2R
GW4090
PAD2/PAD2X
-
GW4190
PAD2/PAD2X
PAD2R
GW6090
PAD2/PAD2X
-
GW6290
PAD2/PAD2X
-
GW6010
PAD26010/PAD2X6010
-
GW7010
PAD26010/PAD2X6010
GW7015
PAD26010/PAD2X6010
WD RANGE – WASHER DISINFECTORS FOR HOSPITAL USE
MACHINE MODEL
PAD model that can be installed
AUXILIARY KIT
WD1050
PAD2/PAD2X
PAD2R
WD1160
PAD2/PAD2X
PAD2R
WD4060
PAD2/PAD2X
PAD2R
WD4190
PAD2/PAD2X
PAD2R
WD6090
PAD2/PAD2X
-
WD5090
PAD2/PAD2X
-
WD6010
PAD26010/PAD2X6010
-
WD7010
PAD26010/PAD2X6010
-
WD7015
PAD26010/PAD2X6010
-
Note: The 60Hz variants of the 230V or 400V Smeg machine models must be equipped with the corresponding 60Hz variant: PAD2-60 and PAD2X-60.
For example, model GW4090-60 must be used with the accessory: PAD2-60 or PAD2X-60.