Read the safety warnings and instructions for the thermodisinfectors in which the rack is used with
care, together with the safety instructions of the producer of the glassware and any other item placed on the rack. Badly positioned items or glass which shatters during washing may cause dangerous injuries. If necessary, washing results must be checked and subjected to non-visual inspections.
CSK8B – USER MANUAL
Page 2
-EN- CSK8B -
19390404000_CSK8B-EN.docx
2
CSKDIN cod. 900018
Dimension lxpxh mm 245 x 480 x 50
CSK2 cod. 900020
Dimension lxpxh mm 160 x 435 x 50
CSK1 cod. 900019
Dimension lxpxh mm 450 x 450 x 55
Only the trolley CSK8B cannot wash surgical instruments
1 MODALITA’ DI UTILIZZO
For a correctly washing, in support of CSK8B, should always be used baskets restraints In the next pictures, the baskets proposed by SMEG
CSK8B containing :
N.8 baskets CSKDIN (standard size) N.8 baskets CSK2 N.4 baskets CSK1
Page 3
-EN- CSK8B -
19390404000_CSK8B-EN.docx
3
AREA BETWEEN LEVELS CSK8B
31
102
716480
In the next pictures, the area between levels
Page 4
-EN- CSK8B -
19390404000_CSK8B-EN.docx
4
IMMAGINE
NOTE
In the near picture, a schematic exemple for incorrect positioning of surgical instrument. You can see that the instrument blocks the spray arms, with consequent bad washing
In the near picture, you can see a correct positioning of surgical instrument, that don’t stop the sprayarms
Dopo aver inserito lo strumentario, è molto importante accertarsi che gli irroratori siano liberi di ruotare. In caso contrario il corretto lavaggio non è garantito e la macchina non può segnalare tale anomalia
INTERFERECE
2 INITIAL OPERATION OF THE TROLLEY
2.1 Trolley spray arms
After inserting the instruments, it is very important to ensure that the spray arms are free to rotate. Otherwise the correct washing is not guaranteed and the machine cannot report this anomaly.
Page 5
-EN- CSK8B -
19390404000_CSK8B-EN.docx
5
it is necessary, whenever the truck is loaded, try manually the free rotation of ALL spray arms (next photo)
2.2 Hydraulic supply of trolley
For the correct use of the CSK8B, another important step is to check the perfect contact between the nozzle supply trolley and that of the machine. If necessary, adjust the height by screwing / unscrewing ONLY the nozzle trolley. In the next picture you can see how to raise the nozzle trolley
Page 6
-EN- CSK8B -
19390404000_CSK8B-EN.docx
6
DRYER INLET
2.3 Drying surgical instruments
Some models of Washer-disinfectors, are provided with the drying system. You can identify by the central hole, internally of the washer chamber. (see photo).
It’s very important that the instruments do not obstruct the air inlet. This could compromise the correct drying of the instruments.
3 ATTENTION
For proper clearing, pay attention to this recommendations:
A prerequisite for an effective washing, is given by the correct loading of the baskets, inserts,
holders, etc.. Articulated instruments must be positioned open.
The baskets must not be overloaded, to allow proper rinsing of the instruments
The large instruments should be stored in baskets in order to avoid "dead spots" that impede the
cleaning of other instruments.
The instruments must be stored or placed, having regard to their mechanical sensitivity, so as to
avoid possible damage.
The CSK8B is a trolley for the thermal disinfection of surgical instruments, be necessary to ensure that the tools introduced in the machine are suitable for the treatment of thermal disinfection.