Smeg CSK6B User manual [GB]

CSK6B
CART FOR INSTRUMENT BASKETS WITH
INSTRUCTION MANUAL
READ CAREFULLY
CSK6B
INSTRUCTION MANUAL
2
This document is related to Smeg’s device “CSK6B CART”. This appliance can be mounted exclusively on Smeg’s devices.
For correct and safe use, this manual has to be used in accordance with the manual of the device, instrument washer or glassware washer, on which the device is installed. If not read, possible incomprehension or wrong interpretation of the instructions included in this manual might damage the appliance, as well as represent a source of danger to the operator and considerably reduce the machine’s performance.
The manufacturer shall not be held liable for uses which are different from those specified in this manual. Installation, maintenance and possible repair operations must be carried out exclusively by personnel authorized by the manufacturer.
In case the machine is used without complying with the specifications stated by the manufacturer, its warranty might be affected.
The material included in this manual is meant for information. The content and the device described might be subject to changes without any notification.
INFORMATION AND SUPPORT FOR THE PRODUCTS OF
SMEG INSTRUMENTS DEPARTMENT
Our Sales Department staff will give you information about prices and offers. Our Technical Services Department will give you directions for a proper functioning of the appliance and will direct you to the closest authorized service centre.
instruments@smeg.itservice.instruments@smeg.it
Our full product range can be viewed on the website:
www.smeg-instruments.com
International customers, please contact your local SMEG distributor.
19 390 4108 02
23/05/2017
Rev.
Date
CSK6B
INSTRUCTION MANUAL
3
Table of Contents
1 DESCRIPTION .............................................................................................................................................................. 4
2 COMPATIBILITY OF THE ACCESSORY – DEVICES ......................................................................................................... 4
3 EQUIPMENT IN THE BOX ............................................................................................................................................ 4
4 ASSEMBLY INSTRUCTIONS ......................................................................................................................................... 4
5 MODE OF USE ............................................................................................................................................................ 5
6 CART START UP .......................................................................................................................................................... 7
6.1 CART’S SPRAY ARMS.......................................................................................................................................... 7
6.2 CART’S WATER SUPPLY...................................................................................................................................... 8
6.3 DRYING OF THE INSTRUMENTS ......................................................................................................................... 9
7 PRECAUTIONS FOR USE .............................................................................................................................................. 9
8 TROUBLE SHOOTING ................................................................................................................................................ 10
9 MANUFACTURER’S REFERENCE ............................................................................................................................... 10
CSK6B
INSTRUCTION MANUAL
4
1 DESCRIPTION
CSK6B is a cart designed to process surgical instruments through thermal disinfection, using baskets.
2 COMPATIBILITY OF THE ACCESSORY – DEVICES
The accessory can be mounted on the following devices:
GAMMA WD – INSTRUMENT WASHER HOSPITAL SECTOR
WD1160
WD3060
WD4060
WD4190
WD5090
3 EQUIPMENT IN THE BOX
N.1 cart CSK6B
4 ASSEMBLY INSTRUCTIONS
GENERAL WARNINGS FOR THE INSTALLATION
Carefully read both warnings and indications on safety for thermal disinfection devices where this cart is used, as well as the warnings indicated by the manufacturer of the instruments or any other component which is meant to be processed with this cart.
Badly positioned objects or potential breaking of glassware during the wash cycle might affect the efficiency of the process and cause serious injury.
If needed, the washing performance may be verified through both visual and other kind of tests, on the basis of the cycle used, the load processed, and the supports used.
FREE SPACE: It is necessary to have a free space of about 1m2 in front of the machine to be able
to correctly work on it.
PROTECTION DEVICES: Use personal protective equipment, if necessary.
CSK6B
INSTRUCTION MANUAL
5
5 MODE OF USE
In the next pictures, the baskets proposed by SMEG
CSKDIN cod. 900018
Dimension lxpxh mm 245 x 480 x 50
CSK2 cod. 900020
Dimension lxpxh mm 160 x 435 x 50
CSK1 cod. 900019
Dimension lxpxh mm 450 x 450 x 55
The trolley CSK6B can be configured with different trays (load-carriers). Typical configurations of the trolley:
Configuration with n°4 baskets CSKDIN (standard size).  Configuration with n°4 baskets CSK2.  Configuration with n°2 baskets CSK1.
In the next pictures, the area between levels
The trolley CSK6B as stand alone trolley cannot wash surgical instruments, For a correctly washing, in support of CSK6B, baskets restraints should always be used.
CSK6B
INSTRUCTION MANUAL
6
AREA BETWEEN LEVELS CSK6B
CSK6B
INSTRUCTION MANUAL
7
6 CART START UP
6.1 CART’S SPRAY ARMS
After inserting the instruments, it is important to make sure that the spray arms are able to rotate freely. Failing that, a correct washing performance is not guaranteed and the machine cannot report this fault.
IMAGE
NOTES
Opposite is a schematic example of how a surgical instrument can block the correct rotation of the spray arm and, consequently, reduce the washing performance of all instruments.
The figure opposite shows that the instruments, if correctly positioned in the basket, do not interfere with the spray arm.
Every time the cart is loaded, it is necessary to manually test the spray arms to check that ALL are able to rotate freely (see photo below).
INTERFERENCE
CSK6B
INSTRUCTION MANUAL
8
6.2 CART’S WATER SUPPLY
For the correct use of the CSK6B, another important step is to check the perfect contact between the nozzle supply trolley and that of the machine. If necessary, adjust the height by screwing / unscrewing ONLY the nozzle trolley. In the next picture you can see how to raise the nozzle trolley.
CSK6B
INSTRUCTION MANUAL
9
6.3 DRYING OF THE INSTRUMENTS
Some models of thermal disinfection devices are equipped with a drying system, which can be distinguished by the central hole placed inside, on the back of the tank (see photo).
It is extremely important that the air inlet is neither fully nor partially obstructed by any instruments at the moment of loading. This may result in an improper drying of the instruments.
7 PRECAUTIONS FOR USE
For a correct washing performance, particular attention should be paid to the following aspects:
An efficient washing performance is given by the correct loading of the baskets, inserts, supports etc. Jointed
instruments must be positioned open.
Baskets must not be overloaded, in order to allow a proper rinsing of the instruments.
Large instruments must be placed in the baskets so that “grey areas” are avoided and the cleaning of other
instruments is not impeded.
Instruments must be placed and arranged taking into account their mechanical sensitivity so as to prevent
any damage.
CSK6B is a cart designed to clean surgical instruments through thermal disinfection; it is therefore necessary to make sure that the instruments inserted in the machine are suitable for the thermal disinfection treatment.
AIR INLET
CSK6B
INSTRUCTION MANUAL
10
8 TROUBLE SHOOTING
In the case of poor quality wash performance:
1. Verify the cleanliness of the nozzles / sprayers – according to the described procedure.
2. Check the instructions for the correct procedure, as supplied by the manufacturer of the instruments.
3. Verify that the correct cycle has been selected.
4. Verify the functioning of the machine, check the manual.
5. Check the quantity and quality of the available detergent.