Pressure measurements can easily be taken any time, anywhere.
Compact Manometer
Series PPA
Compact and lightweight
Por table type with a lightweight of only about 100 g (unit 50 g, battery 50 g) can also be held in the palm of the hand.
Long service life of 12 months continuous operation
One year of continuous operation is possible with 2 type AA batteries (3 V).
Convenient hand strap for carrying
Keeping practical use in mind, the hand strap is a standard feature.
Zero/span calibration is possible
Offset adjustment with the zero clear function, and span calibration with the trimmer can be performed.
Peak/Bottom hold function
With pressure being displayed, variations in supply pressure can be grasped instantly with one touch switching of the display from peak value to bottom value.
Auto power off function to save batteries
Power turns off automatically if not operated for more than 5 minutes.
Case holder is available
The case holder is provided as an option to allow for situations where portability is not required.
Back light for easy viewing in dark locations
14-12-1
1
Application Example
Confirmation of air line source pressure
Human reading error is eliminated by the ability to confirm line pressure on the digital display. It is also possible to check pulsation in the source pressure using the peak/bot­tom display funcion.
Setting a regulator can be performed more precisely than with a dial gauge by viewing the digital display while making the setting. Furthermore, power lines are not needed for this battery operated unit.
Related products for line pressure measurement
Can also be used as an energy saving related device
Tube coupler
Finger valve
For specifications, refer to page 14-12-4.
IN-442-5 Pressure probe
Confirmation of regulator set pressure
Compact Manometer
Series
PPA
Convenient for easy line pressure measurement without re­moving piping or stopping supply pressure, etc.
Pressure can be sup­plied or stopped by in­serting or removing the tube.
Pressure can be easi­ly applied or released by switching the con­trol.
Used in combination with the pressure probe (IN­442-5), total pressure can be easily measured.
Pressure measurements can easily be taken any time, anywhere.
Measures the total pressure received by
an air blown workpiece
Series PPA
14-12-2
2
F.R.L.
AV
AU
AF
AR
IR
VEX
AMR
IC
VBA
VE
VY1
G
PPA
AL
1. The compact manometer can be used for measurement of air and non-corrosive gases.
Please note that the accuracy of measurement for other fluids cannot be guaranteed. Furthermore, the construction is not explosion proof, and therefore, flammable gases should not be used.
2. Be certain to stay within the rated pressure range.
Operation outside the pressure range will cause failure.
3. Do not intentionally swing around by the hand strap.
If the strap breaks or comes loose, there is a danger of injury or damage, etc.
4. When installing or removing One-touch fittings on tubing, first confirm that the fluid to be measured is at atmospheric pressure.
If tubing is disconnected while the fluid to be measured is in a pressurized state, the tubing may jump causing a danger of injury or damage. Also when connecting tubing, confirm that it is securely attached.
5. Instruction manual
Read it carefully and understand the contents before using a product. Also, keep the manual in a location where it can be referred to at any time.
Handling
Warning
1. Keep condensate and foreign matter from getting into the fluid to be measured.
If condensate or foreign matter is mixed in the fluid to be measured, this may cause failure or air leakage. If there is a possibility of these being contained in the fluid, use the meter via a filter or mist separator.
2. Do not drop or strike the unit, etc.
Do not drop, strike or apply a large impact shock (980 m/s2), as this may result in a failure.
3. Be certain to perform the zero clear function with pressure released to the atmosphere.
When perfoming the zero clear function, this should be done with piping ports in an atmospheric release condition. If adjustment is performed at a pressure other than atmospheric pressure, the correct value will not be displayed.
4. Tighten One-touch fittings in accordance with the following.
One-touch fittings should first be tightened by hand, and then further tightened approximately 1/6 of a turn using a tightening tool. If screwed in too far, this may cause air leakage due to breaking of threads or distortion of the gasket, etc. If not screwed in far enough, this may cause a loose fitting or air leakage, etc.
Caution
1. Absolutely never use in an atmosphere where explosive gases are used.
The compact manometer does not have explosion proof construction. If used in an atmosphere of explosive gases, there is a possibility of causing an explosion, and therefore, should absolutely not be used under these conditions.
Operating Environment
Warning
1. Do not use where there is splashing of water or oil,
etc.
The compact manometer is not a dusttight and dripproof type, and should not be used where there is splashing water or oil, etc., as this may result in a malfunction.
Caution
1. Perform maintenance and inspection on a regular
basis.
If there is an unintended misaction, misoperation, etc., or calibration has not been performed, there is a possibility of an incorrect value being displayed, making it impossible to ensure safety.
2. Do not disassemble or modify the unit.
Maintenance and Other
Warning
1. Use manganese type AA dry batteries (R6) or alkaline type AA dry batteries (LR6).
Do not use batteries other than the above, as this may cause failiure.
2. Insert the plus (+) and minus (–) terminals of the batteries in the proper direction as indicated inside the unit.
If the batteties are inserted incorrectly, this may cause them to leak or explode and result in damage to the unit.
3. Do not use old and new batteries or mix different types of batteries together.
This may cause batteries to leak and result in damage to the unit.
4. Remove the batteries when the unit will not be used for a long period.
5. Do not use batteries if their voltage has dropped.
Continuing to use them may lead to the display of incorrect values.
6. Do not touch the span calibration trimmer except when performing span calibration.
Touching the trimmer may cause generation of an error in the measured pressure. Also do not turn it too hard (0.3 N·m or less) or press it too hard (5 N or less).
7. Use a soft cloth to wipe the blot off of the body.
In case of heavy soiling, wipe it off with a cloth soaked in a neutral detergent diluted with water after wringing it out thoroughly, and finish up by wiping with a dry cloth.
Caution

Precautions

Be sure to read before handling. Refer to pages 14-21-3 to 14-21-4 for Safety Instructions and Common Precautions.
14-12-3
3

How to Order

None
With PPA-B
Nil
B
With the unit switching function
(1)
Fixed in SI unit
(2)
Nil
M
Option/Case holder
Unit specifications
PPA10
One-touch fitting type
ø4 (mm size)
ø6 (mm size)
Nil
04
06
0M
–0.1 to 1 MPa (For high pressure)
–101 to 10 kPa (For vacuum)
–10 to 100 kPa (For low pressure)
0 1 2
Applicable tubing size
Symbol
KJH04-M5
KJH06-M5
One-touch fitting
None
Applicable tubing material
Nylon
Soft nylon
Polyurethane
NoneNone
Model
Rated pressure range
Pressure display
Pressure display discrimination
Mini. display units
Error display
Function
Withstanding pressure
Fluid
Power supply
Battery life
Response speed
Display accuracy
Repeatability
Temperature characteristics
Piping port
Ambient operating temperature
Ambient operating humidity
Shock resistance
Enclosure
Weight

Specifications

–101 to 10 kPa
3 digit LCD back light
1/100
Over pressure, Memory data error, Change battery sign
200 kPa
Air, Non-corrosive gases
3 V (DC), Type AA dry cell x 2 pcs.
12 months continuous operation (Without backlighting, temperature conditions: at 25°C)
250 ms
±2% F.S. or less (Temperatue conditions: at 25°C)
±1% F.S. or less (Temperatue conditions: at 25°C)
±3% F.S. or less (0 to 50°C with 25°C standard)
M5 x 0.8
0 to 50°C (With no condensation)
35 to 85% RH (With no condensation)
980 m/s
2
X, Y, Z directions, 3 times each
IP40 (IEC standard)
Approx. 100 g (Unit 50 g, batteries 50 g)
–0.1 to 1 MPa
PPA100 for high press.
–10 to 100 kPa
Peak/bottom display, Backlight, Auto power OFF
Zero clear, Units display switching
PPA101 for vacuum
PPA102 for low press.
1.5 MPa 200 kPa
kPa
MPa
mmHg
kgf/cm
2
inHg
PSI
bar
——
0.01
——
0.1
——
1
0.1
1
——
5
0.01
0.2
0.1
0.01
1
——
——
0.01
——
0.1
0.01
Note)
Series PPA
PPA100/101/102
Pressure specifications
Note 1) With unit switching function is not
permitted to sell for the domestic use in Japan, because the new Weight and Measure Act has been implemented since October ’99.
Note 2) Fixed unit
For vacuum/compound pressure: kPa For positive pressure: MPa
2 pcs. of type AA dry batteries (manganese R6 or alkaline LR6) are not included. Note) For the unit switching function (Types without the unit switching function is fixed in SI unit (kPa or MPa).)
Compact Manometer
14-12-4
4
(PPA100 shown, Unit: MPa)
Description of Operating Parts
Operation and Functions
Power ON
Power OFF
• When pressed and held for 3 seconds or more, the power turns OFF.
• When there is no button operation for more than 5 minutes, the power turns OFF. (auto power OFF function)
Initial Setting
1. When the power is
applied, and “Err” is displayed on LED, cut the power off for a time. After turning OFF (i.e. the state in which nothing is displayed on LCD), then proceed to 2. Besides, in the case that nothing is displayed on LCD, proceed to 2 with doing nothing.
2. Press and hold for 6 seconds or more. The unit will go into zero clear. When this happens “CAL” will appear on the LCD.
3. When zero clear is finished, the unit can be operated.
3. Release the POWER button.
2. Press and hold the POWER button for 6 seconds or more.
1. Confirmation of
display
“P” for peak display “b” for bottom display
In lock mode
Power button
• Power ON/OFF
• Peak mode switching
LCD
• Present pressure display
• Peak/bottom value display
• Units display
LIGHT button
• Turns on back light
MPa
POWER
LIGHT
MPa
POWER
LIGHT
POWER
LIGHT
MPa
POWER
LIGHT
MPa
POWER
LIGHT
MPa
POWER
LIGHT
• The power comes ON as it is pressed.
• When pressed and held for 6 seconds or more, the unit goes into zero clear.
Press the POWER button.
Press and hold the POWER button for 3 seconds or more.
Be certain to perform initial setting when using for the first time and after changing batteries, as the unit will indicate memory data error.
14-12-5
Compact Manometer Series PPA
F.R.L.
AV
AU
AF
AR
IR
VEX
AMR
IC
VBA
VE
VY1
G
PPA
AL
5
Turning on the Backlight
It normally lights up while the button is be­ing pressed. In the lock mode, it lights up when pressed and turns off when pressed again. However, the maxi­mum lighting time is approximately one min­ute.
Zero Clear
The zero point dis­played at atmospheric pressure can be auto­matically adjusted. By this means it is possi­ble to eliminate a dis­play discrepancy at at­mospheric pressure.
• Turn the power OFF.
• Release the supply pressure to the at­mosphere.
• When continuously pressed for 6 sec­onds or more, zero clear is performed and “CAL” is dis­played on the LCD.
Lock Mode (Auto power OFF cancel)
The auto power OFF function is canceled by activating the lock mode (auto power OFF cancel). When continuously pressed for 6 seconds or more, “L” is dis­played on the LCD. Moreover, when the power is turned OFF, the lock mode is re­leased.
Auto Power OFF Function
When the power is turned ON and there is no button operation for more tham 5 minutes, the power will turn OFF. Note) For cancelling this function, refer to the functions and oper­ation of the lock mode (below).
MPa
POWER
LIGHT
MPa
POWER
LIGHT
MPa
POWER
LIGHT
Peak/Bottom Display
Do this when pressure is being displayed.
Peak display
Displays the maximum pressure value and “P” appears on the LCD. The display will change if pressure increases beyond the pressure value that is being held.
Bottom display
Displays the minimum pressure value and “b” appears on the LCD. The display will change if pressure falls below the pressure value that is being held. (These modes are convenient for con­firming pressure fluc­tuations.)
Press the POWER button.
MPa
POWER
LIGHT
MPa
POWER
LIGHT
MPa
POWER
LIGHT
Press the POWER button.
Press the POWER button.
POWER
LIGHT
Press the LIGHT button.
Press the POWER button for 6 seconds or more.
Unit Display Switching
1.When pressed con-
tinuously for 3 sec­onds or more, the unit on the LCD will flash.
2.
The unit will change.
(See the table below.)
3. The unit is set, and switching is finished.
MPa
MPa
bar
Note) The “inHg” unit cannot be displayed.
High pressure
(PPA100)
MPa barPSI kgf
Vacuum
(PPA101)
kPabar PSIinHg mmHg
Low pressure
(PPA102)
kPa barPSI kgf
kgf
POWER
LIGHT
kgf
POWER
LIGHT
MPa
POWER
LIGHT
2. Press the LIGHT button.
3. Press the POWER button.
(PPA100 shown, Unit: MPa)
Operation and Functions
Note) This operation cannot be done for
the type which does not have the unit switching function.
Note) Since this is combined with power
OFF operation, the button should be released at the point when “P” or “b” is displayed.
1. Press and hold the POWER and LIGHT buttons for 3 sec­onds or more.
Press and hold the POWER and LIGHT but­tons for 6 seconds or more.
14-12-6
Series PPA
6

Dimensions

ø4 ø6 ø1/4"
123 124 126
Note) Dimension X includes
the One-touch fittng
(Back)
20
5
Fitting port diameter M5 x 0.8 Thread depth 5
40
20
LIGHT
POWER
180
X Refer to Note)
110
(Front)
BATTERY 1.5 Vx2 LR6, R6P
MADE IN JAPAN
AX
LIGHT
POWER
LOCK MODE(AUTO POWER OFF CANCEL)
PPA100
Connecting tube port
Accessory One-touch fitting
Body
Span calibration
Type AA dry batteries (2 pcs.)
+
+
Option/Case holder
SMC
44
9.2
4.5
152518
120
ø4.5
2-(R)
2
6.8
29
14-12-7
Compact Manometer Series PPA
F.R.L.
AV
AU
AF
AR
IR
VEX
AMR
IC
VBA
VE
VY1
G
PPA
AL
7
Error Correction Maintenance
When errors occur, they should be corrected as shown below.
Display Contents Corrective action
Pressure being applied is above the rating.
Memory data has probably been corrupted in some way.
Operate within the rated pressure range.
Perform auto zero adjustment.
Battery voltage is low. Replace the batteries.
Entire display flashes
Caution
Caution
Trimmer
Standard
pressure gauge
Compact
manometer
Compressor
Related Products Useful for Measuring Line Pressure
Valve type
Fluid
Proof pressure
Maximum operating pressure
Operating vacuum pressure
Ambient and fluid temperature
Applicable tubing material
Note)
Option
2 port valve, 3 port valve
Air
1.5 MPa
1.0 MPa
–100 kPa
0 to 60°C
Nylon, Soft nylon, Polyurethane
Bracket
Finger valve
Series
VHK
Specifications
Note) Use caution with soft nylon and polyurethane at the maximum operating
pressure. (For details, refer to Best Pneumatics Vol. 15.)
For a vacuum application, use VHK2 (2 way valve).
JIS Symbol
Tubing material
Tubing O.D.
Self-seal fittings
Series
KC
Applicable Tubing
Fluid
Maximum operating pressure
Proof pressure
Ambient and fluid temperature
Thread
Seal (Thread portion)
Adaptor for copper family incompatible parts
Air
1.0 MPa
3.0 MPa
0 to 60°C
JIS B 0203 (Taper threads for piping)
JIS B 0211 Class 2 (Metric fine thread)
With seal (Standard)
Part C3604BD (Electroless nickel plated)
Specifications
Nylon, Soft nylon, Polyurethane
ø4, ø6, ø8, ø10, ø12
Mounting section
Nut section
Body
Stud
Chuck spring
Guide
Collet release bushing
Valve retainer
Stopper
Seal O-ring
C3604BD, PBT
C3604BD (Thread portion)
Stainless steel 304
C3604BD, Polyacetal (POM)
POM
POM
C3604BD, Polyacetal (POM)
NBR
Principal Parts Material
These products are convenient for measuring line pressure easily without the need to remove piping or stop supply pressure, etc.
2 port valve 3 port valve
A
II
P
RP
A
For details, refer to Best Pneumatics Vol. 15.
For details, refer to Best Pneumatics Vol. 15.
Span calibration method
Do not touch the span calibration trimmer except when performing span calibration.
1. Perform zero clear at atmospheric pressure.
2. Apply the maximum rated pressure, and calibrate the span while
comparing with a standard pressure gauge.
3. If the display value of the compact manometer is “0” after returning to atmospheric pressure, then calibration is complete. If the display value is not “0”, calibrate again by repeating steps 1 and 2.
Replacing the batteries
When battery voltage becomes low the entire LCD will flash. When the LCD flashes replace the batteries. Use 2 pcs. of type AA dry batteries.
To replace the batteries, turn the power OFF and replace them within approximately 30 seconds. When not completed within 30 seconds, “Err” will be displayed. In that case, perform zero clear once again.
In the event that the display runs out of control, remove the batteries for one minute or longer, and then perform zero clear again for inserting the batteries and turning on the power.
Switching between pressurization and atmospheric re­lease can be easily performed by switching the control.
Pressure can be supplied or stopped by inserting or re­moving a tube.
14-12-8
Series PPA