SMC Networks PSE300 Operation Manual

Page 1
No.PS※※-OMG0002-F
PRODUCT NAME
Pressure Sensor Controller
MODEL / Series / Product Number
PSE300
Page 2
-1-
No.PS※※-OMG0002-F
Table of Contents
Safety Instructions 2
Model Indication and How to Order 8
Summary of Product parts 9
Mounting and Installation 10
Installation 10
Wiring 12
Internal circuit and wiring example 14
Setting 16
Pressure Setting 19
Other Functions 21
Maintenance 23
Troubleshooting 24
Specification 31
Specifications 31
Dimensions 33
Page 3
-2-
No.PS※※-OMG0002-F
Safety Instructions
These safety instructions are intended to prevent hazardous situations and/or equipment damage. These instructions indicate the level of potential hazard with the labels of "Caution", "Warning" or "Danger". They are all important notes for safety and must be followed in addition to International standards (ISO/IEC)
1)
and other safety regulations.
1) ISO 4414: Pneumatic fluid power -- General rules relating to systems
ISO 4413: Hydraulic fluid power -- General rules relating to systems IEC 60204-1: Safety of machinery -- Electrical equipment of machines (Part 1: General requirements) ISO 10218-1: Manipulating industrial robots -Safety. etc.
Caution :
CAUTION indicates a hazard with a low level of risk which, if not avoided, could result in minor or moderate injury.
Warning :
WARNING indicates a hazard with a medium level of risk which, if not avoided, could result in death or serious injury.
Danger :
DANGER indicates a hazard with a high level of risk which, if not avoided, will result in death or serious injury.
Warning
1. The compatibility of the product is the responsibility of the person who designs the equipment or decides its specifications.
Since the product specified here is used under various operating conditions, its compatibility with specific equipment must be decided by the person who designs the equipment or decides its specifications based on necessary analysis and test results. The expected performance and safety assurance of the equipment will be the responsibility of the person who has determined its compatibility with the product. This person should also continuously review all specifications of the product referring to its latest catalog information, with a view to giving due consideration to any possibility of equipment failure when configuring the equipment.
2. Only personnel with appropriate training should operate machinery and equipment.
The product specified here may become unsafe if handled incorrectly. The assembly, operation and maintenance of machines or equipment including our products must be performed by an operator who is appropriately trained and experienced.
3. Do not service or attempt to remove product and machinery/equipment until safety is confirmed.
1. The inspection and maintenance of machinery/equipment should only be performed after measures to prevent falling or runaway of the driven objects have been confirmed.
2. When the product is to be removed, confirm that the safety measures as mentioned above are implemented and the power from any appropriate source is cut, and read and understand the specific product precautions of all relevant products carefully.
3. Before machinery/equipment is restarted, take measures to prevent unexpected operation and malfunction.
4. Contact SMC beforehand and take special consideration of safety measures if the product is to be used in any of the following conditions.
1. Conditions and environments outside of the given specifications, or use outdoors or in a place exposed to direct sunlight.
2. Installation on equipment in conjunction with atomic energy, railways, air navigation, space, shipping, vehicles, military, medical treatment, combustion and recreation, or equipment in contact with food and beverages, emergency stop circuits, clutch and brake circuits in press applications, safety equipment or other applications unsuitable for the standard specifications described in the product catalog.
3. An application which could have negative effects on people, property, or animals requiring special safety analysis.
4. Use in an interlock circuit, which requires the provision of double interlock for possible failure by using a mechanical protective function, and periodical checks to confirm proper operation.
Page 4
-3-
No.PS※※-OMG0002-F
Caution
The product is provided for use in manufacturing industries.
The product herein described is basically provided for peaceful use in manufacturing industries. If considering using the product in other industries, consult SMC beforehand and exchange specifications or a contract if necessary. If anything is unclear, contact your nearest sales branch.
Limited warranty and Disclaimer/Compliance Requirements
The product used is subject to the following "Limited warranty and Disclaimer" and "Compliance Requirements". Read and accept them before using the product.
Limited warranty and Disclaimer
1. The warranty period of the product is 1 year in service or 1.5 years after the product is delivered, whichever is first.
2)
Also, the product may have specified durability, running distance or replacement parts. Please consult your nearest sales branch.
2. For any failure or damage reported within the warranty period which is clearly our responsibility, a replacement product or necessary parts will be provided. This limited warranty applies only to our product independently, and not to any other damage incurred due to the failure of the product.
3. Prior to using SMC products, please read and understand the warranty terms and disclaimers noted in the specified catalog for the particular products.
2) Vacuum pads are excluded from this 1 year warranty.
A vacuum pad is a consumable part, so it is warranted for a year after it is delivered. Also, even within the warranty period, the wear of a product due to the use of the vacuum pad or failure due to the deterioration of rubber material are not covered by the limited warranty.
Compliance Requirements
1. The use of SMC products with production equipment for the manufacture of weapons of mass
destruction (WMD) or any other weapon is strictly prohibited.
2. The exports of SMC products or technology from one country to another are governed by the
relevant security laws and regulation of the countries involved in the transaction. Prior to the shipment of a SMC product to another country, assure that all local rules governing that export are known and followed.
Caution
SMC products are not intended for use as instruments for legal metrology.
Products that SMC manufactures or sells are not measurement instruments that are qualified by pattern approval tests relating to the measurement laws of each country. Therefore, SMC products cannot be used for business or certification ordained by the measurement laws of each country.
Page 5
-4-
No.PS※※-OMG0002-F
Operator
This operation manual is intended for those who have knowledge of machinery using pneumatic
equipment, and have sufficient knowledge of assembly, operation and maintenance of such equipment. Only those persons are allowed to perform assembly, operation and maintenance.
Read and understand this operation manual carefully before assembling, operating or providing
maintenance to the product.
■Safety Instructions
Warning
■Do not disassemble, modify (including changing the printed circuit board) or repair.
An injury or failure can result.
■Do not operate the product outside of the specifications.
Do not use for flammable or harmful fluids. Fire, malfunction, or damage to the product can result. Verify the specifications before use.
■Do not operate in an atmosphere containing flammable or explosive gases.
Fire or an explosion can result. This product is not designed to be explosion proof.
■Do not use the product in a place where static electricity is a problem.
Otherwise it can cause failure or malfunction of the system.
■If using the product in an interlocking circuit:
Provide a double interlocking system, for example a mechanical system Check the product regularly for proper operation
Otherwise malfunction can result, causing an accident.
■The following instructions must be followed during maintenance:
Turn off the power supply Stop the air supply, exhaust the residual pressure and verify that the air is released before performing
maintenance
Otherwise an injury can result.
Page 6
-5-
No.PS※※-OMG0002-F
Caution
■Do not touch the terminals and connectors while the power is on.
Otherwise electric shock, malfunction or damage to the product can result.
■After maintenance is complete, perform appropriate functional inspections and leak tests.
Stop operation if the equipment does not function properly or there is a leakage of fluid. When leakage occurred from other parts except piping, the product might break. Cut off power supply and stop supplying fluid. Do not apply fluid at leaking condition. Safety cannot be assured in the case of unexpected malfunction.
■NOTE
○Follow the instructions given below when designing, selecting and handling the product.
●The instructions on design and selection (installation, wiring, environment, adjustment,
operation, maintenance, etc.) described below must also be followed.
Product specifications
Operate the product with the specified voltage.
Operation with a voltage beyond specifications can cause malfunction or damage of the product. Insufficient supply voltage may not drive a load due to a voltage drop inside the product. Verify the operating voltage of the load before use.
Use the pressure sensor within the specified ranges of the measurement flow rate and under the
specified operating pressure.
Otherwise it can cause damage to the product and an abnormal measurement.
Do not exceed the specified maximum allowable load.
Otherwise it can cause damage or shorten the lifetime of the product.
Reserve a space for maintenance.
Remember to leave space for maintenance when designing the piping plan.
The direct-current power supply to combine should use UL authorization power supply which is the
Class 2 power supply based on UL1310 or the power supply is using the transformer of a Class 2 based on UL1585.
The product is a UL approved product only if it has a
mark on the body.
●Product handling
Installation
Do no drop, hit or apply shock to the product.
Otherwise it can result in damage to the product causing failure or malfunction.
Do not pull cables or lift the body with cables.
Hold the body when handing. Otherwise it can result in damage of the product causing failure or malfunction.
Follow the specified tightening torque.
Excessive tightening torque can break the product, bracket, and mounting screws. Insufficient tightening torque can displace the product from the original position or loosen the mounting screws.
Do not apply excessive external force with joints such as hoses when installing with a panel mount
adapter.
Otherwise it can damage the pipe joint of the product or cause drop off from the panel mount adapter.
Connect frame-ground terminal (FG terminal) to the ground when using a switching power supply. Insert a noise filter (power line noise filter, ferrite core, etc.) between the switching power supply and
product when using analogue output.
Page 7
-6-
No.PS※※-OMG0002-F
Wiring
Do not bend or apply tensile stress to cables repeatedly.
Wiring with repetitive bending stress or tensile stress can cause breakage of the cables. Replace the product when damage to a cable is observed. The recommended bend radius of the cable is 6 times the outside diameter of the sheath, or 33 times the outside diameter of the insulation material, whichever is larger.
Connect wires and cables correctly.
Miswiring can break the product depending on a miswired circuit.
Do not attempt to insert or pull out the pressure sensor or its connector when the power is on. Switch
output may malfunction.
Do not connect wires while the power is on.
Otherwise it can break the circuit inside the product causing malfunction.
Do not lay wires or cables with power cable or high-voltage cable in the same wiring route.
Otherwise the wires to the product can be contaminated with noise or induced surge voltage from power lines or high voltage lines causing malfunction. Lay the wires to the product to a wire duct or in a protective tube other than those for power lines or high voltage lines.
Verify the insulation of wiring.
Poor insulation (interference with other circuit, poor insulation between terminals and etc.) can introduce excess voltage or current to the product causing damage.
Keep wiring as short as possible to prevent contamination from noise and induced surge voltage.
Do not use a cable longer than 10 m. Consult with SMC for the use with a cable longer than 10 m. Connect the 0 VDC wire (blue line) directly or as close as possible to the 0 VDC terminal of the DC power supply.
Environment
Do not use the product in an atmosphere containing corrosive gas, chemicals, seawater, water or vapor,
or in a place where there is a possibility of adhesion of those substances to the product.
It can cause failure or malfunction.
Avoid exposure of this product to direct sunlight.
Use sunshades if the product is exposed to direct sunlight. Otherwise it can cause failure or malfunction.
Do not use in a place where water, oil or chemicals splashes.
Otherwise it can cause failure or malfunction.
Do not use a product nearby a place where electric surges are generated.
Internal circuit elements of the product can deteriorate or break when equipment generating a large surge (electromagnetic lifter, high frequency induction furnace, motor, etc.) is located near the product. Provide surge preventives, and avoid interference.
Do not apply the product to the load that generates electric surge voltage.
Relays or solenoid values generate electric surge voltage. When applying the product to drive these loads directly, use the product equipped with surge absorber.
The product is not resistive to a lightning surge defined in CE marking. Take measures to protect
against a lightning surge at the load side.
Prevent foreign matter such as remnant of wires from entering this product.
Take proper measures for the remnant not to enter the product in order to prevent failure or malfunction.
Mount the product in a place that is not exposed to vibration or impact.
Otherwise it can cause damage or malfunction.
Follow the specified ranges of the operating fluid and maintain ambient temperatures.
The operating fluid and ambient temperatures should be in the range of 0 to 50 °C. When operating at low temperature of 5 °C or below, breakage or malfunction can occur to the product due to freezing of condensed water in the pressurized air. Take preventive measures against freezing. Do not use the product in a place where temperature suddenly changes even if it stays within the specified range.
Do not expose the product to heat radiation from a heat source located nearby.
It can cause malfunction.
Page 8
-7-
No.PS※※-OMG0002-F
Adjustment and Operation
Do not short-circuit the load. Do not press the buttons with a sharp object.
It can cause damage to the setting buttons.
A warm-up time of 20 to 30 minutes is needed for detection of low pressure.
The indication drifts about 1% soon after the power is on.
Do not touch the LCD during operation.
The indication on the LCD changes due to static electricity.
Maintenance
Before performing maintenance, make sure to turn off the power supply.
Otherwise an unexpected operation of the system component can occur.
Perform maintenance and check regularly.
Otherwise an unexpected malfunction of the system can occur due to a malfunction of the product.
Perform a proper functional check after maintenance.
Stop operation when an abnormality is observed such that the device does not work properly. Otherwise an unexpected malfunction of the system component can occur.
Do not use solvents such as benzene or thinner to clean the product body.
It can damage the surface of the body and erase the indication on the body. Use a soft cloth to remove stains. For heavy stains, use a cloth soaked with diluted neutral detergent and fully squeezed, then wipe up the stains again with a dry cloth.
Page 9
-8-
No.PS※※-OMG0002-F
Model Indication and How to Order
PSE3 T
PSE3
: A unit label is attached. 1: The new Measurement Law prohibits use in Japan of products with a unit selection function. 2: Fixed unit for compound, vacuum, low and differential pressure is: kPa
for positive pressure is: MPa (kPa for 500 kPa range model)
○Options
Option products Model Note
Power and output cable ZS-28-A Length 2 m
Bracket ZS-28-B With set screws M3 x 5 L (2 pcs.)
Connector for sensor cable ZS-28-C 1 pc.
Panel mount adapter ZS-27-C With set screws M3 x 8 L (2 pcs.) Panel mount adapter + Front protective cover ZS-27-D With set screws M3 x 8 L (2 pcs.)
Front protective cover ZS-27-01 1 pc.
DIN rail and terminal base type
Option 1
Nil No option
L Power and output cable
Input/output specification
0 NPN open collector 2 outputs + 1 to 5 VDC Analogue output
1 NPN open collector 2 outputs + 4 to 20 mA Analogue output
2 NPN open collector 2 outputs + Auto-shift input
3 PNP open collector 2 outputs + 1 to 5 VDC Analogue output
4 PNP open collector 2 outputs + 4 to 20 mA Analogue output
5 PNP open collector 2 outputs + Auto-shift input
Unit specification
Nil Unit selection function provided
1
M SI units fixed
2
Option 2
Nil No option
A Bracket
B Panel mount adapter
D
Panel mount adapter + Front protective cover
Connector mounting type
Option
Nil No option
E Front protective cover
Input specification
0 Voltage input
1 Current input
Option 3
Nil No option
C Connector for sensor cable
Page 10
-9-
No.PS※※-OMG0002-F
Summary of Product parts
○Names of individual parts
Output (OUT1) lamp (Green): Lit when OUT1 is ON.
Output (OUT2) lamp (Red): Lit when OUT2 is ON.
LCD display: Displays the current status of pressure, setting mode, selected indication unit and error code.
Four display modes can be selected: display always in red or green only, or changing from green to red linked to output.
button (UP): Selects a mode and increases a set ON/OFF value.
Press this button to change to the peak display mode.
button (DOWN): Selects a mode and decreases a set ON/OFF value.
Press this button to change to the bottom display mode.
button (SET): Changes the mode and sets a set value.
Page 11
-10-
No.PS※※-OMG0002-F
Mounting and Installation
■Installation
<PSE3   >
○Mounting
●Mount the optional bracket and panel mount adapter to the product.
○Mounting by bracket
●Fix the bracket to the product with the set screws M3 x 5 L (2 pcs.) as attached.
●The tightening torque of the set screws must be 0.5 to 0.7 Nm.
○Mounting by panel mount adapter
●Fix the panel mount adapter to the product with the set screws M3 x 8 L (2 pcs.) as attached.
○Notice when removing the controller
●The controller with adapter for panel mounting can be
removed from facility by making hook of the controller wide as illustration after removing two screws. Pressure sensor controller and panel mount adapter may be damaged.
Page 12
-11-
No.PS※※-OMG0002-F
<PSE3   T>
●Mounting
Hang the Hook 1 at the bottom of the body on the DIN rail as shown in Fig. a, press it in arrowed direction
to fix.
●Removing
For removal, pull it with a flat driver in arrowed direction shown in Fig. b.
Page 13
-12-
No.PS※※-OMG0002-F
■Wiring
○Connection
●Connections should only be made with the power supply turned off.
●Use separate routes for the product wiring and any power or high voltage wiring. Otherwise, malfunction may result due to noise.
●Ensure that the FG terminal is connected to ground when using a commercially available switch-mode power supply. When a switch-mode power supply is connected to the product, switching noise will be superimposed and the product specification can no longer be met. This can be prevented by inserting a noise filter, such as a line noise filter and ferrite core, between the switch-mode power supply and the product, or by using a series power supply instead of a switch-mode power supply.
<PSE3   >
○Att
aching the connector to the cable
●Sensor wire is stripped as shown in the right figure. (Refer to the table below for correspondence between connector and electrical wire gauge.)
Cable table
AWG No. Conductor size (mm2 ) Overall diameter (mm) Colour of cover SMC product No.
0.8t to 1.0 Red ZS–28–C
1.0 to 1.2 Yellow ZS–28–C–1
24-26 0.14-0.2
1.2 to 1.6 Orange ZS–28–C–2
23 0.1-0.5 1.15 to 1.35 Blue ZS-28-CA-4
1.0 to 1.2 Green ZS–28– C–3 1.2 to 1.6 Blue ZS–28–C–4
20-22 0.3-0.5
1.6 to 2.0 Gray ZS–28–C–5
The correspondence table of each maker and SMC product number
SMC product No. Sumitomo 3M Ltd. product No. Tyco Electronics AMP K.K. product No.
ZS–28–C 37104–3101–000FL 1–1473562–4
ZS–28–C–1 37104–3122–000FL
ZS–28–C–2 37104–3163–000FL
ZS–28–CA–4
ZS–28–C–3 37104–2124–000FL
ZS–28–C–4 37104–2165–000FL
2–1473562–4
ZS–28–C–5 37104–2206–000FL
: The cable that is suitable with conductor composition may be different. Please contact with an electric wire manufacture for details.
Page 14
-13-
No.PS※※-OMG0002-F
●The core of the corresponding colour shown in the following table is put into the pin of the number stamped on the connector for sensor connection to the back.
Colour of cable core
PSE31  (Current input)
Number stamped
on connector
PSE30  (Voltage input)
Pressure sensor 2-wire type Pressure sensor 3-wire type
1 Brown (DC(+)) Brown (LINE(+)) Brown (DC(+))
2 N.C. N.C. N.C.
3 Blue (DC(-)) N.C. Blue (DC(-))
4 Black (OUT: 1 to 5 V) Blue (LINE(-)) Black (OUT: 4 to 20 mA)
●It checks that the above-mentioned preparation work has been performed correctly, and A part shown in right figure is pushed by hand and makes temporary connection.
●A part center is straightly pushed in by tools, such as pliers.
●Re-use cannot be performed once it connects the connector for sensor connection completely. When you
fail in the connection mistake of a core and a pin, or the plug of wire, please use the new connector for sensor connection.
Page 15
-14-
No.PS※※-OMG0002-F
○Connector
Connecting/Disconnecting
●When connecting the connector, insert it straight onto the pin holding the lever and connector body between fingers and lock the connector by pushing the lever claw into the square groove in the housing until connector clicks.
●When disconnecting the connector, push down the lever by thumb to disengage the lever claw from the square groove. Then pull the connector straight out.
Pin No. of the connector for power and output cable
<PSE3   T>
○Applicable crimping terminal dimensions
●The terminal screw is M3.
●If using the crimping terminal, follow the specifications below.
●Tighten the terminal screw at a torque of 0.3 to 0.35 Nm.
■Internal circuit and wiring example
○Wiring example
PSE3   T (Voltage input, Current input: Pressure sensor
3-wire type)
PSE31
T
(Current input: Pressure sensor 2-wire type)
Page 16
-15-
No.PS※※-OMG0002-F
○Output specification
●When the cable with SMC power and output cable (Model: ZS-28-A) is used, the colors of wire (Brown,
Black, White, Gray, Blue) will apply as shown on circuit diagram.
PSE3
0(T)
NPN open collector output: 2 outputs
Max. 30 V, 80 mA
Residual voltage 1 V or less Analogue output: 1 to 5 V Output impedance: Approx. 1 kΩ
PSE3
1(T)
NPN open collector output: 2 outputs
Max. 30 V, 80 mA
Residual voltage 1 V or less Analogue output: 4 to 20 mA Max. load impedance: 300 Ω (12 VDC)
600 Ω (24 VDC)
Min. load impedance: 50 Ω
PSE3  2(T)
Auto shift input Voltage free contact NPN open collector output: 2 outputs
Max. 30 V, 80 mA Residual voltage 1 V or less
PSE3
3(T)
PNP open collector output: 2 outputs
Max. 80 mA
Residual voltage 1 V or less Analogue output: 1 to 5 V Output impedance: Approx. 1 kΩ
PSE3  4(T)
PNP open collector output: 2 outputs
Max. 80 mA
Residual voltage 1 V or less Analogue output: 4 to 20 mA Max. load impedance: 300 Ω (12 VDC)
600 Ω (24 VDC)
Min. load impedance: 50 Ω
PSE3
5(T)
Auto shift input Voltage free contact PNP open collector output: 2 outputs
Max. 80 mA Residual voltage 1 V or less
Page 17
-16-
No.PS※※-OMG0002-F
Setting
○Setting procedures
○Initial setti
ng
Keep pressing the button longer than two seconds. Remove the finger off the button when [Sor] is displayed and initial setting can get started.
1. Display colour setting Select a colour for the LCD display. When changing the display colour, press the
or
button to select a display colour.
Press the
button to set the desired
display colour.
2. Output linked to display colour setting (For selection of Sor and SoG only) Select output linked to display colour, press the
or button
and select output. Press the
button to set.
3. Pressure range setting Select the pressure range suitable for the sensor connected. Press the
or button and select
the pressure range. Press the
button to set.
(Refer to the following table for the labels printed the units stuck on the display part)
4. Selecting display unit (In case [M] is not assigned to unit specification in model indication) The indication unit can be selected freely. Pressing the
or button will change the unit and will
automatically convert set values. Press the
button to set and to move to setting the output mode.
LCD display
For compound and vacuum kPa kgf/cm2 bar psi inHg mmHg
For low pressure kPa kgf/cm2 bar psi
For positive
1
MPa.kPa kgf/cm2 bar psi
Unit
For low differential kPa mmH2O
1: MPa for 0 to 1 MPa range model, kPa for 0 to 500 kPa range model.
○Unit label
How to use the labels printed the units. In order to display the selected unit, the unit label according to the pressure range or the display unit is attached.
●In case [M] is assigned to unit specification in model indication.
Use the suitable label in the following labels by setup of the pressure range.
LCD display
Unit label kPa MPa kPa
Page 18
-17-
No.PS※※-OMG0002-F
5. Output method setting Four output mode can be selected by an operating mode and by output style. One of these four output
mode can be selected for each output.
OUT1 and OUT2 can be set independently. Refer to "List of output mode".
1) The operating mode for OUT1 is set.
Press the
or button and select the hysteresis mode or the window comparator mode. Press the button to set.
2) The output style for OUT1 is set.
Press the
or button and select the normally open or the normally closed mode. Press the
button
to set.
3) The operating mode and the output style for OUT2 is set.
Press the
or button and the button to
set, as in OUT1.
List of output mode
Only hysteresis mode can be set at auto preset. The following is given using OUT1 as an example. The descriptions for OUT2 are the same as those
for OUT1, under the conditions that [n_1] and [n_2] should be replaced by [n_3] and [n_4], [P_1] and [P_2] should be replaced by [P_3] and [P_4] and [H_1] should be replaced by [H_2].
Page 19
-18-
No.PS※※-OMG0002-F
6. Response time setting
A response time for switch output can be set as user desires. Set the optimum response time to prevent
the chattering of a switch.
The response time currently set will be displayed. Select a desired response time by pressing the
or
button. Press the button to set.
7. Pressure setting
There are two methods for pressure set-up: manual and auto
preset, either one of which can be selected. The auto preset is provided for an automatic optimum set-up by using a sample for a case in which switch output is used to check absorption.
An operation mode currently selected is displayed. Press the
or button to select the set-up method to be used.
Press the
button to set.
When both OUT1 and OUT2 are in window comparator mode, this item is not shown up.
8. Auto shift setting (PSE3  2(T)/3  5(T) models only)
1) Select the display mode of the pressure value at the time of auto shift operation.
Either [AS (Auto shift)] or [ASO (Auto shift zero)] can be selected.
AS (Auto shift): [AS] displays the differential pressure of the
atmosphere and measurement pressure.
ASO (Auto shift zero): [ASO] displays the differential pressure of
the measurement pressure and the measurement pressure at the time of auto shift signal input.
Press the
or button to select the auto shift or auto shift zero. Press the button to set.
2) Selecting the switches which Auto Shift mode apply, when the auto shift signal is inputted.
Press the
or button to select the A1, A2 or Ab. Press the button to set.
The initialize setting will be completed and return to the Measurement mode.
Page 20
-19-
No.PS※※-OMG0002-F
Pressure Setting
○Manual setting
Manually set a set value of the product.
1. Selection of OUT1 [P_1] setting mode
Press the
button during the Measurement mode to
display set values.
[P_1] and the current set value will displays in turn. (In case
the Normally Closed mode is selected in initialization, [n_1] and the set value will displays in turn.)
Press the
or button to enter into the value changing
mode, then change the set value. (See "Value setting")
Check the corrected value, then press the
button to set the set value.
2. Selection of OUT1 [P_2] setting mode (Window comparator mode selected)
[P_2] and the current set value will displays in turn. (In case the
normally closed mode selected in initialization, [n_2] and the set value will displays in turn.)
Press the
or button to enter into the value changing
mode, then change the set value. (See "Value setting")
Check the corrected value, then press the
button to set
the set value.
3. Selection of OUT1 [H_1] setting mode
[H_1] and the current set value will displays in turn. Press the
or button to enter into the value changing
mode, then change the set value. (See "Value setting")
Check the corrected value, then press the
button to set the set value.
4. Selection of OUT2 setting mode
Set the set values [P_3] [P_4] and [H_2] of OUT2 as in OUT1.
[P_3] [P_4] or [H_2] and current set value will displays in turn. (In case the normally closed mode selected in initialization, [n_3] [n_4] or [H_2] and set value will displays in turn.)
Press the
or
button to enter into the Value changing mode, then change the set value.
(See "Value setting")
Check the corrected value, then press the
button to set the set value.
5. Auto shift compensation value setting (PSE3
2(T)/3  5(T) models only)
[C_5] and Auto shift corrected value will displays in turn. Check the corrected value, then press the
button.
The pressure setting mode will be completed and return to the measurement mode.
○Value setting
To input a value for pressure setting or other purposes:
1. Press the
or button to enter the set value change mode. The 1st digit will flicker.
2. Press the
or button to set a desired value. (No operation within thirty seconds after the set value change mode was selected results in automatic setting of the value appearing in the display window and in charging of the mode from set value change mode to set value indication mode.)
3. Press the
button to make the value one digit higher flicker.
(If the highest place is zero, [
] or [  ] will flicker.)
[
] means "+zero", [  ] means "-zero".
(In the case that the
button is pressed in the highest place, the 1st digit will flicker.)
4. Press the
button continuously for longer than one second to memorize
the set value and to return to displaying set values.
○Setting example
Page 21
-20-
No.PS※※-OMG0002-F
○Auto-preset
In case auto preset is selected in Initialize, this function stores in the memory a pressure setting value which is calculated from a measurement pressure as a reference value. The set value of product is automatically set to an optimum value by repeating absorption and non-absorption several times with a sample which is to be set up.
1. Selection of OUT1 auto preset mode
Press the
button to display [AP1].
(In case OUT1 setting is not necessary, press
button and button at the same time longer
than one second.)
2. Preparation of unit for OUT1
Prepare a unit for which pressure for OUT1 is to be set.
3. Selection of auto preset value of OUT1 setting
Press the
button to display [A1L].
Operate system so that pressure may change. Detection will be made and a set value will be stored in the memory
automatically and display [A1H].
4. Selection of OUT2 auto preset mode
Press the
button to set [P_1], [H_1] ([n_1], [H_1] in normally closed mode) and display [AP2].
(In case OUT2 setting is not necessary, press
and button at the same time longer than one second.)
5. Preparation of unit for OUT2 and pressure setting
Prepare a unit for which pressure for OUT2 is to be set. Press the
button to display [A2L].
[A2L] is displayed and Detection will be made and a set value will be stored in the memory automatically
and display [A2H].
6. Set up of OUT2 auto preset value
Press the
button to set the set value of [P_3], [H_2] ([n_3], [H_2] in normally closed mode), and
auto preset mode is finished.
The mode will return to the Measurement mode.
A pressure setting value in auto preset is as follows in normally open mode with OUT1. (P_1 is n_1 in Normally Closed mode with OUT1.)
P_1 = A – (A – B)/ 4 A = maximum pressure value H_1 = (A – B)/ 2 B = minimum pressure value
For OUT2 set-up, above P_1, n_1, and H_1 become P_3, n_3, and H_2 respectively.
○Fine adjustment mode (Fine adjustment function of display value)
1. Press the button and buttons simultaneously for longer than two seconds in the measurement mode. [FSt] and current pressure measurement value will displays in turn.
2. Press the
or button to change the set
value. (The range to be changed: 5%R.D.)
3. If no operation is made for longer than three seconds or press the
button, the product will display the current pressure measurement value which will then displays in turn with [FSt].
4. Press the
button to display an adjusted amount (percentage), which will then displays in turn with [FSC].
5. Press the
button to return to the normal measurement mode.
Page 22
-21-
No.PS※※-OMG0002-F
Other Functions
○Auto shift function
In case the source pressure fluctuates too much, the product may not be able to operate normally. Auto shift is provided to compensate for the fluctuation of the source pressure. While measured pressure becomes standard pressure value when auto shift input is received, this function correct set value of the switches.
With auto shift
Set auto shift input as Lo at the time pressure source change, in order to memorize the pressure change and to correct pressure set value, so that correct decision emerge.
○Conditions and explanations for auto shift function
Keep constant pressure for 5ms or more from the close signal of auto shift input. At auto shift input, the pressure at that time is memorized to [C_5] as corrected value, and the switch
operates with the value which applied compensation value to setting value. Display indicates [ooo] for approx.1 sec.
The switch set as auto shift mode at the time of initial setting operates with the value which applied
corrected value [C_5] to setting value.
OUT1 will operate with Auto shift function, when "A1" is selected.
The operating value of OUT1 applies corrected value [C_5] to [P_1], [P_2] or [n_1], [n_2].
OUT2 will operate with Auto shift function, when "A2" is selected.
The operating value of OUT2 applies corrected value [P_3], [P_4] or [n_3], [n_4].
Both OUT1 and OUT2 will operate with Auto shift function, when "Ab" is selected.
The operating value of OUT1 and OUT2 applies corrected value [P_1] to [P_4] or [n_1] to [n_4].
Span is 10ms or less until switch output perform soon after auto shift input. When corrected set value exceed the set pressure range with auto shift input, corrected value is not
memorized and displays [o.r] for approx.1sec.
Correct value [C_5] after auto shift input vanish when off the power. Correct value [C_5] for auto shift input function is reset as zero (Initial value) when re-supplied power. In case of auto shift zero selected, the display indicates [0] (zero)] if the auto shift signal is inputted.
: No EEPROM in the memory of corrected value.
Using with auto shift input, accepted set range is like below
Set pressure range Accepted set range
For compound
-
101.0 to 101.0 kPa
-
101.0 to 101.0 kPa
For vacuum 10.0 to -101.0 kPa
-
101.0 to 101.0 kPa
For low pressure
-
10 to 100.0 kPa
-
100.0 to 100.0 kPa
-
0.1 to 1.000 MPa
-
1.000 to 1.000 MPa
For positive
-
50 to 500 kPa
-
500 to 500 kPa
For low differential
-
0.2 to 2.00 kPa
-
2.00 to 2.00 kPa
Page 23
-22-
No.PS※※-OMG0002-F
○Peak and bottom hold display function
Maximum and minimum values are always detected and updated during measurement. Displayed values can be held. In peak hold, press the
button for longer than one second to make flicker and to hold the maximum
pressure value. To reset holding, press the
button again for more than one second. The
measurement mode will be set.
In bottom hold, press the
button for longer than one second to make flicker and to hold the minimum
pressure value. To reset holding, press the
button again more than one second. The Measurement
mode will be set.
Press continuously the
and buttons simultaneously more than one second during displaying
the peak/bottom hold values to reset the maximum or minimum pressure value.
○Key lock function
This function prevents errors such as changing a set value by mistake.
Lock
Keep p
ressing the
button longer than four seconds, remove the finger off the button when [UnL] is
displayed.
Press the
or buttons to set the display to [LoC].
Press the
button and return to the Measurement mode.
Unlock
P
ress the
button longer than four seconds, remove the finger off the button when [LoC] is displayed.
Press the
or buttons to change the display to [unL].
Press the
button and return to the Measurement mode.
○Zero clear function
A displayed value can be adjusted to zero when pressure to be measured is within 7%F.S. of the atmospheric pressure. (There is variation in 4 digits according to a product characteristic.) Press continuously the
and buttons simultaneously more than one second to reset to "0" on the
display.
The mode will return to the Measurement mode automatically.
Page 24
-23-
No.PS※※-OMG0002-F
Maintenance
How to reset the product for power cut or forcible de-energizing
The setting of the product is remained as that before power cut or de-energizing. The output condition is also basically recovered to that before power cut or de-energizing, but may change depending on the operating environment. Therefore, check the safety of whole facility before operating the product. If the facility is under accurate control, wait until it has warmed up (20 to 30 minutes).
Page 25
-24-
No.PS※※-OMG0002-F
Troubleshooting
○Troubleshooting
Applicable pressure switch: PSE300
If a cause applicable to the failure cannot be identified and normal operation can be recovered by replacement with a new Pressure switch, this indicates that the Pressure switch itself was broken. The Pressure switch breakage can be caused by operating environment (network construction, etc.), and so consult with SMC separately to obtain countermeasures.
Yes
No
The Pressure switch
does not operate
normally
Refer to reference
No.1
The switch output is
on
The indication light
is on
The indication light
works improperly
The switch output is
off
The switch output
generates chattering
The indication light
is off
The indication light
is off
The indication light
is on
The product is
broken
Refer to reference
No.2
Refer to reference
No.1
Refer to reference
No.4
Refer to reference
No.3
The product is
broken
The product is
broken
Analogue output is not
provided (Specified
accuracy is not
satisfied)
Refer to reference
No.6
Slow switch output
response
Refer to reference
No.5
An error signal
comes on
Refer to reference
No.7
Page 26
-25-
No.PS※※-OMG0002-F
The unit cannot be
changed
The buttons cannot be
operated
The display fluctuates
There is a noise
The display disappears
The display flashes
Pressure indication
difference when using
two or more pressure
switches
Refer to reference
No.8
The dis
play
is abnormal
Refer to reference
No.9
Refer to reference
No.9
Refer to reference
No.11
Refer to reference
No.14
Refer to reference
No.13
Refer to reference
No.15
The display breaks off
Refer to reference
No.16
Refer to reference
No.10
The indication
accuracy does not
satisfy the
specifications
Refer to reference
No.12
Yes
No
Page 27
-26-
No.PS※※-OMG0002-F
○Cross-reference for troubleshooting
Reference
No.
Problem Possible cause Investigation method Countermeasure
Wrong pressure setting
(1)Check the set pressure. (2)Check the settings of the
operation mode, hysteresis and output style.
(Hysteresis mode/window comparator mode, normal output/reversed output)
(1)Reset the pressure setting. (2)Reset the setting of
function.
1
Output remains
on.
Indication light remains on.
Output remains
off.
Indication light remains off.
Product failure
Replace the product.
Incorrect wiring
Check the wiring of the output line. Check if the load is connected directly to DC(+) or DC(-).
Correct the wiring.
2
Output remains on. Indication light works correctly.
Product failure
Replace the product.
Incorrect wiring
Check the wiring of the output line. Check if the load is connected directly to DC(+) or DC(-).
Correct the wiring.
Unsuitable model selection
Check if PNP is used even though NPN should have been selected, or the other way around.
Review the selected model (output type).
Cable breakage
Check if there is bending stress applied to any parts of the cable. (Bending radius and tensile force applied to the cable)
Correct the wiring conditions. (Adjust the tensile force and widen the bending radius.)
3
Output remains off. Indication light works correctly.
Product failure
Replace the product.
Incorrect wiring
Check the wiring. Check if the brown and blue wires are connected to DC(+) and DC(-) respectively, and if the output line is about to come off (contact failure).
Correct the wiring.
Wrong setting
(1)Check the set pressure. (2)Check if the hysteresis range is
too narrow.
(3)Check the response time set as
initial setting.
Check if the response time is too short.
(1)Reset the pressure setting. (2)Widen the hysteresis. (3)Reset the setting of
function.
4
Switch output generates chattering.
Product failure
Replace the product.
5
Slow switch output response
Incorrect pressure setting
(1)Check the pressure setting.
Check if the detected pressure and the set pressure value have the same value or are too close.
(2)Check the response time set as
initial setting.
Check if the response time is too long.
(1)Reset the pressure setting.
Set up the pressure setting value so it is not too close to the detected pressure value.
(2)Reset the setting of
function.
Page 28
-27-
No.PS※※-OMG0002-F
Reference
No.
Problem Possible cause Investigation method Countermeasure
Incorrect wiring
Check if the analogue output line is connected with a load.
Correct the wiring.
Non-compliance with the load spec.
(1) Check if the proper load is
connected.
(2) Check if input impedance of
input equipment (A/D transformer) is proper.
Connect a proper load.
Insufficient warm-up
Check if the product satisfies the specified accuracy in 20 minutes after supplying power.
After energizing, indication and output can drift. For detecting fine pressure, warm up the product for 20 to 30 minutes.
6
Analogue output is not provided. (Specified accuracy is not satisfied.)
Product failure Replace the product.
Over current to the output (Er1 and 2)
(1)Check if a current of 80 mA or
more is flowing to the output.
(2)Check if the connected load
satisfies the specifications, and if the load is shorted.
(3)Check if a relay without a surge
voltage suppressor is connected.
(4)Check if the wiring is in the same
route as (or bundled together with) a high-voltage line or the power line.
(1),(2)Connect the load as
specified.
(3)Use a relay with a surge
voltage suppressor or take a measure to prevent noise.
(4)Separate the wiring from
the high-voltage line and/or power line.
Improper transaction of the internal data of the Pressure switch (Er4, 6, 7, and 8)
(1)Check if there is noise
interference such as static electricity. Check if there is a noise source
(2)Check if the power supply
voltage is in the range of 12 to 24 VDC.
(1)Remove the noise and the
noise source (or take measures to prevent noise interference), and turn off the power supply. Then, supply the power again.
(2)Supply power voltage of 12
to 24 VDC.
Applied pressure is over the upper limit (HHH).
(1)Check if the pressure gets over
the upper limit of the set pressure range.
(2)Check if foreign matter got into
the piping.
(3)The connector for sensor may
not be connected correctly.
(1)Bring the pressure back
within the set pressure range.
(2)Take measures to prevent
foreign matter from getting into the piping.
(3)Check the wires and contact
of the connector for sensor.
Applied pressure is under the lower limit (LLL).
(1)Check if the pressure gets below
the lower limit of the set pressure range.
(2)Check if foreign matter got into
the piping.
(3)The connector for sensor may
not be connected correctly.
(1)Bring the pressure back
within the set pressure range.
(2)Take measures to prevent
foreign matter from getting into the piping.
(3)Check the wires and contact
of the connector for sensor.
Pressure is not atmospheric pressure at zero-clear operation (Er3)
Check if the pressure over ±7%F.S. of the atmospheric pressure is applied.
Return the applied pressure to atmospheric pressure, and retry the zero clear operation.
7
An over current
error (Er1 and
2) is indicated.
System error
(Er4, 6, 7, and
8) is indicated.
The display
shows "HHH".
The display
shows "LLL".
Zero-clear error
(Er3) is indicated.
Product failure Replace the product.
Page 29
-28-
No.PS※※-OMG0002-F
Reference
No.
Problem Possible cause Investigation method Countermeasure
Incorrect power supply
Check if the power supply voltage is within the range of 12 to 24 VDC.
Supply power supply voltage of 12 to 24 VDC.
8
Indicated values fluctuate.
Incorrect wiring
Check the wiring to the power supply. Check if the brown and blue wires are connected to DC(+) and DC(-) respectively and if the output line is about to come off (contact failure).
Correct the wiring.
Incorrect power supply
Check if the power supply voltage is within the range of 12 to 24 VDC.
Supply power supply voltage of 12 to 24 VDC.
Power saving mode
Check if the power saving mode is selected.
Reset the setting of function.
9
Indicator turns
off.
A part of the
indication misses.
Product failure Replace the product.
The peak value/bottom value indication mode is selected.
Check if the peek value or bottom value indicating mode is selected.
Turn off the peak value/bottom value indication mode.
10
Indicator is blinking.
Wiring failure
(1)Check the power supply wiring. (2)Check if bending stress is being
applied to a specific part of the cable.
(1)Correct the wiring. (2)Correct the wiring (bending
radius and stress).
Dispersion within the indication accuracy range
Check if the dispersion is within the indication accuracy range.
Use the fine adjustment mode to adjust the indication if the dispersion is within the indication accuracy range.
11
Pressure indication difference when using two or more Pressure switches.
Product failure Replace the product.
Foreign matter
Check if foreign matter has entered the pressure port.
Install a 5 m filter to prevent foreign matter from getting into the pressure port. Also, clean the filter regularly to prevent drainage deposits.
Air and liquid leakage
Check if air and liquid are leaking from the piping.
Rework the piping. If excessive tightening torque over the specified range is applied, a mounting screw, mounting bracket, and product may be broken.
Insufficient warm-up
Check if the product satisfies the specified accuracy 20 minutes after supplying power.
After energizing, indication can drift. For detecting fine pressure, warm up the product for 20 to 30 minutes.
12
The pressure indication accuracy does not satisfy the specifications.
Product failure
Replace the product.
Page 30
-29-
No.PS※※-OMG0002-F
Reference
No.
Problem Possible cause Investigation method Countermeasure
Improper model selection (Selection of model "without unit conversion function")
Check if there is a "-M" at the end of the part number printed on the product
"M" in the part number means that the unit cannot be changed.
: The unit change function is not
available in Japan due to a new measurement law.
: It is fixed to the SI unit "kPa",
"MPa".
13
The unit cannot be changed.
Product failure Replace the product.
Key lock mode
Check if the key lock mode is turned on.
Turn off the key lock mode.
14
The buttons cannot be operated.
Product failure Replace the product.
Air and liquid leakage
Check if air liquid are leaking from the piping.
Rework the piping. If excessive tightening torque over the specified range is applied, a mounting screw, mounting bracket, and product may be broken.
15 Noisy.
Product failure Replace the product.
Effect of pressure source fluctuation due to small hysteresis or too early of a response time
(1)Check the set pressure
(hysteresis)
(2)Check the response time
(1)Check the pressure
setting.
(2)Reset the setting of
function.
Incorrect wiring/ cable breakage
(1)Check the power supply wiring. (2)Check if bending stress is
applied to a specific part of the cable.
(bending radius and tensile force applied to the cable)
(1)Correct the wiring (2)Correct the wiring
conditions. (Adjust the tensile force and widen the bending radius.)
16
The operation is unstable. (Chattering)
Product failure Replace the product.
Page 31
-30-
No.PS※※-OMG0002-F
○Error indication function
This function displays error location and nature when a problem or an error occurs.
Error Name Error Display Error Type Troubleshooting
OUT1
Over current Error
OUT2
A load current of switch output is 80 mA or more.
Turn the power off and remove the output factor for the over current. Then turn the power on.
Residual pressure Error
During zero clear operation, pressure over ±7%F.S. is applied. After 3 s, the mode will reset to the measurement mode. ±4 digits of the zero clear range varies with individual product differences.
Perform zero clear operation again after restoring the applied pressure to an atmospheric pressure condition.
Pressure over max. limit of set pressure range is applied or it is over the display range.
Applied pressure Error
The sensor may not be connected or connected incorrectly. Pressure over min. limit of set pressure range is applied or it is over the display range.
Check proper connection of the sensor and make sure the applied pressure to the value is within the set pressure range. Set back an applied pressure into within set pressure range. While using the auto shift, even if it exceeds the display continuously.
Auto shift Error
The measured pressure at auto-shift input exceeds the set pressure range.
: After 1 s, measurement mode returns
automatically.
Auto shift input is invalid by connected equipment and machine. Check the connected equipment and machine.
System Error
Displayed in the case of an internal data error.
Turn the power off and turn it on again. If resetting fails, an investigation by SMC Corporation will be required.
If the error can not be reset after the above measures are taken, then please contact SMC.
Page 32
-31-
No.PS※※-OMG0002-F
Specification
■Specifications
Model No. PSE3  
Applicable pressure sensor
For compound For vacuum
For low
pressure
For positive
For low
differential
Rated pr essure range -100 to 100 kPa 0 to -101 kPa 0 to 100 kPa 0 to 1 MPa 0 to 500 kPa 0 to 2 kPa
Set pressure range
1
-101 to 101 kPa 10 to -101 kPa -10 t o 100 kPa -0 .1 to 1 MPa -50 to 500 kPa -0.2 to 2 kPa
Set pressure resolution 0.2 kPa 0.1 kPa 0.1 kPa 0.001 MPa 1 kPa 0.01 kPa
Power supply voltage 12 to 24 VDC, ripple (p-p) 10% or less (Protected against inverse connection)
Current consumption 50 mA or less (No load)
Sensor input signal
PSE30
: Voltage input 1 to 5 VDC (Input impedance: 1 MΩ)
PSE31
: Current input 4 to 20 mADC (Input impedance: 100 Ω)
Input point 1 input
Input protection With over voltage protection (Max. 26.4 V)
Hysteresis Variable
Switch output NPN or PNP open collector output, 2 outputs
Max. load current 80 mA
Max. applied voltage 30 VDC (NPN output)
Residual voltage 1 V or less (80 mA load current)
Output protection Short circuit protection
Response time 1 ms or less
Chattering-proof func. Chattering-proof func. working: 20 ms, 160 ms, 640 ms, 1280 ms selected Repeatability 0.1%F.S.
Voltage output
2
Output voltage: 1 to 5 V (within rated pressure range), Output impedance: approx.1 kΩ Linearity: 0.2%F.S. (Without sensor), Response time: 150 ms or less
Accuracy 0.6%F.S. 1.0%F.S. 1.5%F.S.
Current output
2
Output Current: 4 to 20 mA (within rated pressure range), Max. load impedance: 300 Ω (power supply voltage: 12 VDC)
600 Ω (power supply voltage: 24 VDC)
Min. load impedance: 50 Ω, Linearity: 0.2%F.S. (Without sensor), Response time: 150 ms or less
Analogue output
Accuracy 1.0%F.S. 1.5%F.S. 2.0%F.S.
Indicator accuracy
0.5%F.S.
2 digits
0.5%F.S.  1digit
LCD display 3 1/2 digits 7-segment display, Dual-colour display (Red/Green)
Indicator OUT1: Illuminate ON (Green), OUT2: Illuminate ON (Red)
Auto shift input
2
Non-Voltage input (reed or solid state), Low level input: 5 ms or more, Low level: 0.4 V or less
1: Using with auto shift input, accepted set range is different depends on the set pressure ranges. (Refer to page 21.) 2: Auto shift input is not applied when analogue output type is selected. And analogue output is not applied when auto shift input type is
selected.
3: Fixed unit is as below
for compound and vacuum: kPa, kgf/cm
2
, bar, psi, mmHg, inHg
for low and positive pressure: MPa,
kPa, kgf/cm2, bar, psi,
for low differential pressure: kPa,
mmH2O
Page 33
-32-
No.PS※※-OMG0002-F
Model No. PSE3  
Enclosure IP40 Ambient temp. range Operation: 0 to 50 C, Storage: -10 to 60 C (No condensation or freezing)
Ambient humidity range Operation, Storage: 35 to 85%RH (No condensation)
Withstand voltage 1000 VAC, 1 minute Between lead block and case
Environment
Insulation resistance 50 MΩ or more at 500 VDC Between lead block and case
Temp. characteristic 0.5%F.S. or less of detected pressure (25 C)
Connection method PSE3   : Power supply and output 5P connector. Sensor 4P connector, PSE3   T: Te r mi n al bl o ck
Material Front case: PBT, Rear case: PBT (PSE3   ), M-PPE (PSE3   T)
With power and output cable
PSE3
 
: 85 g
Weight
Without power and output cable
PSE3
 
: 30 g, PSE3   T: 50 g
Power supply/Output connection cable
Oil resistance vinyl cabtyre cable
5 cores 4.1, 2 m
Sectional area of conductor: 0.2 mm
2
Outside diameter of insulator: 1.12 mm
Standard CE, UL/CSA, RoHS
○Analogue output
Range Rated pressure range A B E
For vacuum 0 to -101 kPa 0 -101 kPa 10.1 kPa
For compound -100 kPa to 100 kPa -100 kPa 100 kPa -
For low pressure 0 to 100 kPa 0 100 kPa -10 kPa
0 to 1 MPa 0 1 MPa -0.1 MPa
For positive pressure
0 to 500 kPa 0 500 kPa -50 kPa
Range Rated pressure range C D F
For low differential pressure 0 to 2 kPa 0 2 kPa -0.2 kPa
Page 34
-33-
No.PS※※-OMG0002-F
■Dimensions
○Body dimensions
○Mounting using mounting option
Mounting by bracket
Mounting by panel mount
Page 35
-34-
No.PS※※-OMG0002-F
○Panel cutout dimension
: Panel thickness: 0.5 to 6 mm
Page 36
-35-
○PSE3   T
No.PS※※-OMG0002-F
Page 37
No.PS※※-OMG0002-F
Revision history
C: Revision and Format change. D: Contents revised in several places. E: Contents revised in several places. F: Contents revised in several places.
Note: Specifications are subject to change without prior notice and any obligation on the part of the manufacturer.
© 2007-2014 SMC Corporation All Rights Reserved
Loading...