SMC Networks IZH10 Installation And Maintenance Manual

I
ZH10-TFK19GB
Installation and Maintenance Manual
Handheld Electrostatic Meter Type IZH10
Safety Instructions
• This manual contains essential information for the protection of users and others from possible injury and/or equipment damage.
• Keep this manual in a safe place for future reference.
• These instructions indicate the level of potential hazard by label of "DANGER", "WARNING" or "CAUTION", followed by important safety information which must be carefully followed.
• To ensure safety of personnel and equipment the safety instructions in this manual and the product catalogue must be observed, along with other relevant safety practices.
WARNING
• Do not disassemble, modify (including the change of a printed circuit board) or repair.
An injury or failure can result.
• Do not operate outside of the specification.
Fire, malfunction or damage can result. Please confirm the specifications before use.
• Do not operate in an environment of flammable, explosive or corrosive gas.
Fire or an explosion can result. The meter is not an explosion proof type.
• This instrument is supplied for the measurement of high voltage up to ± 20 kV.
The user must take all necessary precautions to use it safely, as high voltages are dangerous.
• Heavy impacts on the meter.
Do not drop, allow any collision or apply excessive shock to the meter when handling. It can result in damage to the meter and accidents.
• Do not allow the sensor to come into contact with the measurement target.
Failure and accidents may occur.
• Handling of ground wire.
To ensure safety and accurate measurements always use the ground wire. If the ground wire is not connected properly, charge can accumulate in the sensor and/or the ground terminal and then discharge to operator. Handle the sensor and ground terminal carefully.
Safety Instructions (continued)
NOTE
Follow the instructions given below when handling the meter. If the instructions are not followed the meter may malfunction or become damaged.
Notes for use
• Do not carry or swing the meter by the sensor cable. If the sensor cable breaks, the meter can hit people or objects, causing injury or damage.
• Do not press setting buttons with sharp, pointed objects.
• Do not touch the detecting surface of the sensor by hand, directly or with a metal object. Such action can result in damage and accidents as well as a loss of specified function and performance.
• When measuring objects with a high charged potential, there is a risk of electrostatic discharge to the hand of the user holding the sensor. In this situation use the high voltage measuring handle (optional extra) and wear protective rubber gloves. Starting from a safe distance, slowly move the sensor closer to the measurement target. Stop immediately if the display shows "HHH" or "LLL" as the charged potential is out of range. This means the charged potential is very high, which can be highly dangerous. The display will not change if the sensor is moved closer to the charged object.
• Do not place objects and cables other than the measurement target near the detecting port of the sensor. Doing so will cause interference and result in an inaccurate display reading.
• Handle cables carefully so they do not get tangled with userís equipment, as this can be highly dangerous.
• The measurement distance is 50 mm. Refer to scales shown on the label attached to the sensor.
• The meter is designed to measure static electricity and must not be used for other purposes.
Environment for use/storage
• Do not use in areas where electromagnetic noise is generated. It can result in malfunction (display of incorrect value), deterioration and damage to internal components.
• Do not leave the meter in areas exposed to direct heat or sunlight, such as inside a closed car or near a heating appliance. High temperatures can cause deformation, discolouration and damage to the meter.
• Do not use in areas where the meter could be splashed by oil or chemicals.
Maintenance and Other Precautions
• Use 2 of AA (LR6) alkaline dry cell batteries in the meter. Other batteries are not suitable and their use may result in damage.
• When fitting the batteries pay particular attention to the polarity markings ("+" and "-") stamped on the body of the meter. If fitted incorrectly, the batteries may leak or even burst.
• Remove the batteries when the meter will not be used for extended periods.
• If the meter body becomes dirty, clean by wiping with a soft cloth. For more stubborn dirt wipe with a cloth dampened with a dilute solution of neutral detergent, then dry thoroughly with another clean, dry cloth.
Model Indication Method
Outline with Dimensions (mm)
Meter body
Sensor
POWER
LIGHT
kV
BEST DETECTING POSITION 50mm
Do not apply a strong impact on the sensing part.
40 20
(5)
110
POWER
M
easured Potential 20.0kV
B
ATTERY 1.5V x 2 LR6
G
AUTO POWER-OFF Extension
P
ress
for 6sec
P
EAK / BOTTOM MODE Display
Normal Bottom
POWER POWER
Press! Press!
Peak
M
ADE IN JAPAN
Extends at 15min
SMC
600
87.5
4.8
9
16.6
13
0.5
8.8
6.5
7
(7.5)
BEST DETECTING POSITION 50mm
2 3.4
Names and Functions of Individual Parts
Meter body
Sensor
kV
LCD Charge amount Peak / Bottom hold value
LIGHT button Backlight ON/OFF
AA (LR6) dry cell battery
Battery saving: "L"
Peak hold: "P" Bottom hold: "b"
POWER button Power supply ON/OFF
POWER
L
IGHT
(2 pcs.)
POWER
Measured Potential 20.0kV
BATTERY 1.5V x 2 LR6
G
AUTO POWER-OFF Extension
Press for 6sec
PEAK / BOTTOM MODE Display
Normal Bottom
POWER POWER
Press! Press!
Peak
MADE IN JAPAN
Extends at 15min
Ground terminal
BEST DETECTING POSITION 50mm
Caution label for sensor
Do not apply a strong impact on the sensing part.
Sensor label
Detecting port
Specification
Note 1) AA alkaline dry cell batteries (2 pcs.) are not included, and need
to be supplied separately.
Note 2) With 2 new alkaline dry cell batteries at room temperature.
CAUTION
If instructions are not followed there is a possibility of injury or equipment damage.
WARNING
If instructions are not followed there is a possibility of serious injury or loss of life.
DANGER
In extreme conditions, there is a possibility of serious injury or loss of life.
IZH10 -
Option
Symbol Contents
Nil None
H High voltage measuring handle
• Accessories and Options/Part Number for Individual Parts
Description Ground wire (1.5 m) Soft case High voltage measuring handle
Part number
IZH-A-01 IZH-B-01 IZH-C-01
*The ground wire and soft case are included with all products.
Item IZH10
Rated charge range
+
20.0 kV
Display resolution
0.01 kV (0 to +0.99 kV)
0.1 kV (
+
1.0 kV to +20.0 kV)
Measurement distance
50 mm
(between sensor and measurement
target)
Power supply Note 1)
1.5 V AA alkaline dry cell battery (2 pcs.) (Battery life: >15 hours continuous use)
Note 2)
Display accuracy
Temperature range
Humidity range
Vibration resistance
+
5 %F.S. +1 digit
Operating: 0 to 40
o
C,
Storage: -10 to 60
o
C
(No freezing or condensation)
Operation and storage: 35 to 85 %R.H.
(No condensation)
10 to 150 Hz applied for 2 hours in each
direction of X, Y and Z with 1.5 mm
amplitude resistance and 98 m/s
2
acceleration (with power OFF)
100 m/s2applied 3 times in each
direction of X, Y and Z (with power OFF)
Environmental resistance
Enclosure IP40
Impact resistance
Material Weight Standard Accessory
Meter body: PC/ABS, Sensor: ABS
85 g (excluding batteries)
CE marking
Ground wire, soft case
Power ON
When the "POWER" button is pressed with the meter off, the meter will be turned on. All display segments are displayed for 1 second after the meter is turned on.
Power OFF
When the "POWER" button is pressed for 3 seconds or more with the meter on, the meter will be turned off. If no buttons are pressed for a set time the meter will automatically turn off. (For details, refer to Auto Power-off function.)
To conserve battery life turn the meter off immediately after use.
kV
P
OWER
L
IGHT
kV
P
OWER
L
IGHT
P
OWER
1 s after
kV
P
OWER
L
IGHT
All display
segments are on
Instantaneous
value display
3 s
kVkV
P
OWER
L
IGHT
P
OWER
L
IGHT
P
OWER
I
ZH10-TFK19GB
Auto Power-off
If no buttons are pressed for 5 minutes or more the meter will turn off automatically.
Auto Power-off Extension
When the "POWER" button is pressed for 6 seconds or more with the meter turned off, the auto power-off display will extend to 15 minutes. (When the auto power-off extension is activated, all the display segments will flash for 3 seconds.)
When the meter is turned off again, the auto power-off extension will be cleared.
5 min.
kVkV
P
OWER
LIGHT
P
OWER
LIGHT
3 s
after
6 s
POWER
kV
POWER
LIGHT
kV
POWER
LIGHT
kV
P
OWER
LIGHT
kV
POWER
LIGHT
kV
POWER
LIGHT
All display
segments are on
Instantaneous
value display
Instantaneous
value display
Flashing
Peak/Bottom Hold Value
When the "POWER" button is pressed with the meter turned on, the display will be changed to instantaneous value, peak hold value, bottom hold value and back to instantaneous value, in that order.
Note) Release the "POWER" button after "P" or "b" is displayed so the meter is not
t
urned off.
Peak hold value The maximum charged potential detected and "P" are displayed. The maximum charged potential is continuously detected and updated from when the peak hold is started. If a value greater than the displayed charged potential is detected, the display will update to show the new value.
Bottom hold value The minimum charged potential detected and "b" are displayed. The minimum charged potential is continuously detected and updated from when the bottom hold is started. If a value less than the displayed charged potential is detected, the display will update to show the new value.
kV kVkV
Instantaneous value Peak hold value Bottom hold value
POWER
L
IGHT
POWER
L
IGHT
POWER
L
IGHT
P
OWER
P
OWER
P
OWER
Zero Clear
The displayed value can be adjusted to zero for a measured charged potential if it is in the range of the factory zero setting
+
5 % F.S. (There may be a slight deviation from the factory setting, due to variations in the sensor itself and the ambient environment where the meter is used.)
When "POWER" and "LIGHT" buttons are pressed simultaneously for 6 seconds or more (with the meter on), the displayed value is reset to zero and the meter will then return to measurement mode. Once the meter is turned off, the offset value for Zero Clear is cleared.
Backlight
When the "LIGHT" button is pressed while the meter is in measurement mode, the backlight will turn on. Pressing the "LIGHT" button again will turn the backlight off.
kV kV
LIGHT
P
OWER
LIGHT
LIGHT
P
OWER
6 s
P
OWER
Instantaneous
value display
kV kV
POWER
LIGHT
POWER
LIGHT
LIGHT
LIGHT
Battery LOW
When the power from the batteries is running low, "L" will appear on the display. Depending upon how much power is left in the batteries, "L" will be displayed differently.
Display resolution change
The display resolution changes depending upon the measured charged potential value.
<Display example>
kVkV
P
OWER
LIGHT
P
OWER
LIGHT
kV
P
OWER
LIGHT
kV
P
OWER
L
IGHT
Display resolution: 0.1 kV Display resolution: 0.01 kV
1.0 kV to 20.0 kV 0 to 0.99 kV
Error Indication
When a malfunction or an error occurs, information about the affected part and type of error will be displayed.
Contact
URL http://www.smcworld.com (Global) http://www.smceu.com (Europe)
Specifications are subject to change without prior notice from the manufacturer.
© 2009 SMC Corporation All Rights Reserved.
Error Indication (continued)
Display
The batteries are low. Prepare to replace with new batteries.
The batteries are very low. Replace with new batteries immediately.
L (ON continuously)
L (
Fl
a
s
hi
n
g
)
Error name Error display Error contents Troubleshooting
A charged potential greater than
+
5 %F.S. of the factory zero setting was present when Zero Clear was performed. *The error display lasts for
approx. 1 s and then the meter returns to measurement mode. There may be a slight deviation from the factory setting, due to variations in the sensor itself and the ambient environment where the meter is used.
Return to uncharged condition, and perform Zero Clear again.
There is an internal data error.
Turn the meter off and on again. If the error has not cleared, contact the local SMC sales branch.
Zero Clear error
System error
The sensor is broken.
Stop using immediately and contact the local SMC sales branch.
Sensor error
The value of charged potential being measured is higher than the upper limit for the sensor or the sensor is too close to the measurement target.
Eliminate static electricity until the charge is within the measurable voltage range. Also, check if the sensor is the correct distance from the measurement target.
The value of charged potential being measured is lower than the bottom limit for the sensor or the sensor is too close to the measurement target.
Measurement error
Correct measurements cannot be made if any of the wires in the cable are broken. Even if the sensor is still, functioning and detects a charged potential the display may not change.
Stop using immediately and contact the local SMC sales branch.
Cable breakage
-
Er
r
o
r
n
a
m
e
Er
r
o
r
d
i
s
p
l
a
y
Er
r
o
r
c
o
n
t
e
n
t
s
Tr
o
u
b
l
e
s
h
o
o
t
i
n
g
AUSTRIA (43) 2262 62280 NETHERLANDS (31) 20 531 8888 BELGIUM (32) 3 355 1464 NORWAY (47) 67 12 90 20 CZECH REP. (420) 541 424 611 POLAND (48) 22 211 9600 DENMARK (45) 7025 2900 PORTUGAL (351) 21 471 1880 FINLAND (358) 207 513513 SLOVAKIA (421) 2 444 56725 FRANCE (33) 1 6476 1000 SLOVENIA (386) 73 885 412 GERMANY (49) 6103 4020 SPAIN (34) 945 184 100 GREECE (30) 210 271 7265 SWEDEN (46) 8 603 1200 HUNGARY (36) 23 511 390 SWITZERLAND (41) 52 396 3131 IRELAND (353) 1 403 9000 UNITED KINGDOM (44) 1908 563888 ITALY (39) 02 92711
Setting
Loading...