SMC Networks ISE70-*-L2 series,ISE71-*-L2 series,ISE70G Series,ISE75G Series,ISE77G Series Operation Manuals

No.PS※※-OMV0004-C
PRODUCT NAME
High-precision Digital Pressure Switch
MODEL / Series / Product Number
ISE70-#-L2 ISE71-#-L2
-1-
No.PS※※-OMV0004-C
Table of Contents
Safety Instructions 2 Model Indication and How to Order 8 Summary of Product parts 9
Definition and terminology 11
Mounting and Installation 14
Installation 14 Piping 15
Wiring 17 Outline of Settings [Measurement mode] 20 Pressure Setting 20 3-Step Setting Mode 21 Simple Setting Mode 23 Function Selection Mode 25
Function selection mode 25
Default setting 25
F 0 Display units, switch output specifications and diagnostic information selection function 27
F 1 Setting of OUT1 29
F 2 Setting of OUT2 32
F 3 Digital filter setting 34
F 4 Auto-preset function 35
F 6 Fine adjustment of display value 37
F10 Sub display setting 38
F11 Display resolution setting 44
F14 Zero cut-off setting 45
F80 Power saving mode 46
F81 Security code 47
F90 Setting of all functions 49
F96 Number of pressurizing error 51
F98 Output check 52
F99 Reset to default settings 54 Other Settings 55 IO-Link specifications 60
Outline of IO-Link function 60
Communication specifications 60
Process data 61
IO-Link parameter setting 64 Maintenance 72 Forgotten the security code 72 Troubleshooting 73 Specifications 82
Dimensions 84
-2-
No.PS※※-OMV0004-C
Safety Instructions
These safety instructions are intended to prevent hazardous situations and/or equipment damage. These instructions indicate the level of potential hazard with the labels of "Caution", "Warning" or "Danger". They are all important notes for safety and must be followed in addition to International Standards (ISO/IEC)*1), and other safety regulations.
*1) ISO 4414: Pneumatic fluid power -- General rules relating to systems.
ISO 4413: Hydraulic fluid power -- General rules relating to systems. IEC 60204-1: Safety of machinery -- Electrical equipment of machines. (Part 1: General requirements) ISO 10218: Manipulating industrial robots -Safety. etc.
Caution
Caution indicates a hazard with a low level of risk which, if not avoided, could
result in minor or moderate injury.
Warning
Warning indicates a hazard with a medium level of risk which, if not avoided,
could result in death or serious injury.
Danger
Danger indicates a hazard with a high level of risk which, if not avoided, will
result in death or serious injury.
Warning
1. The compatibility of the product is the responsibility of the person who designs the equipment or decides its specifications.
Since the product specified here is used under various operating conditions, its compatibility with specific equipment must be decided by the person who designs the equipment or decides its specifications based on necessary analysis and test results. The expected performance and safety assurance of the equipment will be the responsibility of the person who has determined its compatibility with the product. This person should also continuously review all specifications of the product referring to its latest catalog information, with a view to giving due consideration to any possibility of equipment failure when configuring the equipment.
2. Only personnel with appropriate training should operate machinery and equipment.
The product specified here may become unsafe if handled incorrectly. The assembly, operation and maintenance of machines or equipment including our products must be performed by an operator who is appropriately trained and experienced.
3. Do not service or attempt to remove product and machinery/equipment until safety is confirmed.
1. The inspection and maintenance of machinery/equipment should only be performed after measures to
prevent falling or runaway of the driven objects have been confirmed.
2. When the product is to be removed, confirm that the safety measures as mentioned above are
implemented and the power from any appropriate source is cut, and read and understand the specific product precautions of all relevant products carefully.
3. Before machinery/equipment is restarted, take measures to prevent unexpected operation and malfunction.
4. Contact SMC beforehand and take special consideration of safety measures if the product is to be used in any of the following conditions.
1. Conditions and environments outside of the given specifications, or use outdoors or in a place
exposed to direct sunlight.
2. Installation on equipment in conjunction with atomic energy, railways, air navigation, space, shipping,
vehicles, military, medical treatment, combustion and recreation, or equipment in contact with food and beverages, emergency stop circuits, clutch and brake circuits in press applications, safety equipment or other applications unsuitable for the standard specifications described in the product catalog.
3. An application which could have negative effects on people, property, or animals requiring special
safety analysis.
4. Use in an interlock circuit, which requires the provision of double interlock for possible failure by using
a mechanical protective function, and periodical checks to confirm proper operation.
-3-
No.PS※※-OMV0004-C
Safety Instructions
Caution
1.The product is provided for use in manufacturing industries.
The product herein described is basically provided for peaceful use in manufacturing industries. If considering using the product in other industries, consult SMC beforehand and exchange specifications or a contract if necessary. If anything is unclear, contact your nearest sales branch.
Limited warranty and Disclaimer/Compliance Requirements
The product used is subject to the following "Limited warranty and Disclaimer" and "Compliance Requirements". Read and accept them before using the product.
Limited warranty and Disclaimer
1. The warranty period of the product is 1 year in service or 1.5 years after the product is delivered, whichever is first.
2)
Also, the product may have specified durability, running distance or replacement parts. Please consult your nearest sales branch.
2. For any failure or damage reported within the warranty period which is clearly our responsibility, a replacement product or necessary parts will be provided. This limited warranty applies only to our product independently, and not to any other damage incurred due to the failure of the product.
3. Prior to using SMC products, please read and understand the warranty terms and disclaimers noted in the specified catalog for the particular products.
2) Vacuum pads are excluded from this 1 year warranty.
A vacuum pad is a consumable part, so it is warranted for a year after it is delivered. Also, even within the warranty period, the wear of a product due to the use of the vacuum pad or failure due to the deterioration of rubber material are not covered by the limited warranty.
Compliance Requirements
1. The use of SMC products with production equipment for the manufacture of weapons of
mass destruction (WMD) or any other weapon is strictly prohibited.
2. The exports of SMC products or technology from one country to another are governed by
the relevant security laws and regulation of the countries involved in the transaction. Prior to the shipment of a SMC product to another country, assure that all local rules governing that export are known and followed.
Caution
SMC products are not intended for use as instruments for legal metrology.
Products that SMC manufactures or sells are not measurement instruments that are qualified by pattern approval tests relating to the measurement laws of each country. Therefore, SMC products cannot be used for business or certification ordained by the measurement laws of each country.
-4-
No.PS※※-OMV0004-C
Operator
This operation manual is intended for those who have knowledge of machinery using pneumatic
equipment, and have sufficient knowledge of assembly, operation and maintenance of such equipment. Only those persons are allowed to perform assembly, operation and maintenance.
Read and understand this operation manual carefully before assembling, operating or providing
maintenance to the product.
Safety Instructions
Warning
Do not disassemble, modify (including changing the printed circuit board) or repair.
An injury or failure can result.
Do not operate the product outside of the specifications.
Do not use for flammable or harmful fluids. Fire, malfunction, or damage to the product can result. Verify the specifications before use.
Do not operate in an atmosphere containing flammable or explosive gases.
Fire or an explosion can result. This product is not designed to be explosion proof.
Do not use the product in a place where static electricity is a problem.
Otherwise it can cause failure or malfunction of the system.
If using the product in an interlocking circuit:
Provide a double interlocking system, for example a mechanical system Check the product regularly for proper operation
Otherwise malfunction can result, causing an accident.
The following instructions must be followed during maintenance:
Turn off the power supply Stop the air supply, exhaust the residual pressure and verify that the air is released before performing
maintenance
Otherwise an injury can result.
-5-
No.PS※※-OMV0004-C
Caution
Do not touch the terminals and connectors while the power is on.
Otherwise electric shock, malfunction or damage to the product can result.
After maintenance is complete, perform appropriate functional inspections and leak tests.
Stop operation if the equipment does not function properly or there is a leakage of fluid. When leakage occurs from parts other than the piping, the product might be faulty. Disconnect the power supply and stop the fluid supply. Do not apply fluid under leaking conditions. Safety cannot be assured in the case of unexpected malfunction.
NOTE
Follow the instructions given below when designing, selecting and handling the product.
The instructions on design and selection (installation, wiring, environment, adjustment, operation,
maintenance, etc.) described below must also be followed.
Product specifications
Use the specified voltage.
Otherwise failure or malfunction can result.
Do not exceed the specified maximum allowable load.
Otherwise it can cause damage or shorten the lifetime of the Pressure switch.
Design the product to prevent reverse current when the circuit is opened or the product is forced to operate for
operational check. Reverse current can cause malfunction or damage to the product.
Input data to the Pressure switch is not deleted, even if the power supply is cut off.
(Writing time: 10,000 times, Data duration: 20 years after power off)
Use the clean air.
This can cause operating failure. If compressed air containing condensate is used, install an air dryer or drain catch before the filter and perform drainage regularly. If drainage is not performed regularly and condensate enters the secondary side, it can cause operating failure of pneumatic equipment. If regular drainage is difficult, the use of a filter with an auto drain is recommended.
Applicable fluid is air, inert gases and incombustible gases.
Do not use a fluid containing chemicals, synthetic oils including organic solvent, salt and corrosive gases. Otherwise, damage to the product and malfunction can result. Check the details of the specifications before using.
Use the specified measurement flow rate and operating pressure.
Otherwise it can cause damage to the pressure switch or inability to measure correctly.
Reserve a space for maintenance.
Allow sufficient space for maintenance when designing the system.
-6-
No.PS※※-OMV0004-C
Product handling
Installation
Tighten to the specified tightening torque.
If the tightening torque is exceeded the mounting screws and brackets may be broken. If the tightening torque is insufficient, the product can be displaced and loosen the mounting screws.
Be sure to ground terminal FG when using a commercially available switch-mode power supply. Do not drop, hit or apply shock to the Pressure switch.
Otherwise damage to the internal parts can result, causing malfunction.
Do not pull the lead wire forcefully, not lift the product by pulling the lead wire. (Tensile force 35 N or less)
Hold the body when handling to avoid the damage of the Pressure switch which lead to cause the failure and malfunction.
For piping of the Pressure switch, hold the piping with a spanner on the metal part of the piping (Piping
attachment).
Holding other part with spanner leads to damage the Pressure switch.
Eliminate any dust left in the piping by air blow before connecting the piping to the product.
Otherwise it can cause damage or malfunction.
Do not insert metal wires or other foreign matter into the pressure measurement port.
It can damage the pressure sensor causing failure or malfunction.
Never mount a Pressure switch in a location that will be used as a foothold.
The product may be damaged if excessive force is applied by stepping or climbing onto it.
If the entering of foreign material to the fluid is possible, install and pipe the filter or the mist separator to
the inlet to avoid failure and malfunction.
Wiring
Do not pull the lead wires.
In particular, never lift a Pressure switch equipped with fitting and piping by holding the lead wires. Otherwise damage to the internal parts can result, causing malfunction or to be off the connector.
Avoid repeatedly bending or stretching the lead wire, or placing heavy load on them.
Repetitive bending stress or tensile stress can cause the sheath of the wire to peel off, or breakage of the wire. If the lead wire can move, fix it near the body of the product. The recommended bend radius of the lead wire is 6 times the outside diameter of the sheath, or 33 times the outside diameter of the insulation material, whichever is larger. Replace the damaged lead wire with a new one.
Wire correctly.
Incorrect wiring can break the Pressure switch.
Do not perform wiring while the power is on.
Otherwise damage to the internal parts can result, causing malfunction.
Do not route wires and cables together with power or high voltage cables.
Otherwise the product can malfunction due to interference of noise and surge voltage from power and high voltage cables to the signal line. Route the wires (piping) of the product separately from power or high voltage cables.
Confirm proper insulation of wiring.
Poor insulation (interference from another circuit, poor insulation between terminals, etc.) can lead to excess voltage or current being applied to the product, causing damage.
Design the system to prevent reverse current when the product is forced to operate for operational check.
Depending on the circuit used, insulation may not be maintained when operation is forced, allowing reverse current to flow, which can cause malfunction and damage the product.
Keep wiring as short as possible to prevent interference from electromagnetic noise and surge voltage.
Do not use a cable less than 30 m. Wire the DC(-) line(blue) as close as possible to the power supply.
Environment
Do not use the product in area that is exposed to corrosive gases, chemicals, sea water, water or steam.
Otherwise failure or malfunction can result.
Do not use in a place where the product could be splashed by oil or chemicals.
If the product is to be used in an environment containing oils or chemicals such as coolant or cleaning solvent, even for a short time, it may be adversely affected (damage, malfunction, or hardening of the lead wires).
-7-
No.PS※※-OMV0004-C
Do not use in an area where surges are generated.
If there is equipment which generates a large amount of surge (solenoid type lifter, high frequency induction furnace, motor, etc.) close to the Pressure switch, this may cause deterioration or breakage of the internal circuit of the Pressure switch. Avoid sources of surge generation and crossed lines.
Do not use a load which generates surge voltage.
When a surge-generating load such as a relay or solenoid is driven directly, use a Pressure switch with a built-in surge absorbing element.
The product is CE marked, but not immune to lightning strikes. Take measures against lightning strikes
in the system.
Mount the product in a place that is not exposed to vibration or impact.
Otherwise failure or malfunction can result.
Prevent foreign matter such as remnant of wires from entering the Pressure switch.
Take proper measures for the remnant not to enter the Pressure switch in order to prevent failure or malfunction.
Do not use the product in an environment that is exposed to temperature cycle.
Heat cycles other than ordinary changes in temperature can adversely affect the inside of the product.
Do not expose the product to direct sunlight.
If using in a location directly exposed to sunlight, shade the product from the sunlight. Otherwise failure or malfunction can result.
Keep within the specified fluid and ambient temperatures range.
The fluid and ambient temperatures should be 0 to 50 °C. Operation under low temperature (5 °C or less) leads to cause damage or operation failure due to frozen moist in the fluid or air. Protection against freezing is necessary. Air dryer is recommended for elimination of drain and water. Avoid sudden temperature change even within specified temperature.
Do not operate close to a heat source, or in a location exposed to radiant heat.
Otherwise malfunction can result.
When the Pressure switch is used in a place which is exposed to the splash of oil and solvent all the
time, protect the Pressure switch from direct splash.
It can corrode and swell the seal (FKM, HNBR) which causes the malfunction of the Pressure switch.
Adjustment and Operation
Turn the power on after connecting a load.
Otherwise it can cause excess current causing instantaneous breakage of the Pressure switch.
Do not short-circuit the load.
Although error is displayed when the Pressure switch load is short circuit, generated excess current lead to cause the damage of the Pressure switch.
Do not press the setting buttons with a sharp pointed object.
It may damage the setting buttons.
If using the product to detect very small pressure rates, warm up the product for 10 to 15 minutes first.
There will be a drift on the display of approximate 1% immediately after the power supply is turned on, within 10 minutes.
Perform settings suitable for the operating conditions.
Incorrect setting can cause operation failure. For details of each setting, refer to page 19 to 59 of this manual.
Do not touch the LCD during operation.
The display can vary due to static electricity.
Maintenance
Turn off the power supply, stop the supplied air, exhaust the residual pressure and verify the release of
air before performing maintenance.
There is a risk of unexpected malfunction.
Perform regular maintenance and inspections.
There is a risk of unexpected malfunction.
Perform drainage regularly.
If condensate enters the secondary side, it can cause operating failure of pneumatic equipment.
Do not use solvents such as benzene, thinner etc. to clean the Pressure switch.
They could damage the surface of the body and erase the markings on the body. Use a soft cloth to remove stains. For heavy stains, use a cloth soaked with diluted neutral detergent and fully squeezed, then wipe up the stains again with a dry cloth.
-8-
No.PS※※-OMV0004-C
Model Indication and How to Order
Accessories/Part numbers
If an option is required independently, order with the following part numbers.
Items
Part No.
Remarks
Bracket A
ZS-50-A
With 2 mounting screws M4 x 6
Bracket B
ZS-50-B
With 2 mounting screws M4 x 6
Lead wire with M12 connector
ZS-31-B
Straight: 5 m
ZS-31-C
Right angle: 5 m
-9-
No.PS※※-OMV0004-C
Summary of Product parts
Names of individual parts
Body
Power supply/ output connectors: Connects the power supply/ output lead wire. Operation/Display: Refer to the table below.
Operation/Display
Operation light: Displays the switch operating condition. Main display: Displays pressure measurement values and error codes. (2-colour display) Sub display (left): Displays items. (Orange) Sub display (right): Displays set values, peak and bottom values. (Orange) UP button: Increases mode and ON/OFF set values. DOWN button: Decreases mode and ON/OFF set values. SET button: Press this button to change mode and to confirm settings. IO-Link status indicator light: Displays OUT1 output communication status (SIO mode, start-up mode,
Pre-operation mode, operation mode) and presence of communication data.
-10-
No.PS※※-OMV0004-C
IO-Link indicator light operation and display
Communication
with master
IO-Link
indicator light
Status
Sub screen
display 1
Content
Yes
IO-Link
mode
Correct
Operate
Normal communication
status
(Reading of
measurement value)
Start up
When communication
starts up.
Preoperate
Abnormal
Version does not
match
Version of master and
IO-Link
does not match
2
Lock
Back-up and re-store
required due to data
storage lock
No
Communication
shut-off
Correct communication
was not received for
1 second or more.
SIO mode
General switch output
LCD display: “ ” OFF, Flashing, ON
1: “ModE - - - is displayed when selecting the modes on the sub screen. 2: When the product is connected to the master with version "V1.0", error Er15 is generated.
-11-
No.PS※※-OMV0004-C
Definition and terminology
Term
Definition
A
Auto-preset
Performs pressure setting automatically by detecting the increase and decrease in pressure. For example, if this function is used for a suction test, the pressure setting will be completed by performing suction and release of the workpiece.
B
Bottom value display (mode)
Shows the minimum pressure from when the power was supplied to the current time.
C
Chattering
The problem of the switch output turning ON and OFF repeatedly around the set value at high frequency due to the effect of pulsation.
Chattering prevention function
A function to delay the response time of switch output in order to prevent chattering.
D
Delay time
The setting time from when the pressure applied to the pressure switch reaches the set value, to when the ON-OFF output actually begins working. Delay time setting can prevent the output from chattering. The response time indicates when the set value is 90% in relation to the step input.
digit (Min. setting unit)
Shows how precisely the pressure can be displayed or set by the digital pressure switch. When 1 digit = 1 kPa, the pressure is displayed in increments of 1 kPa, e.g., 1, 2, 3, …, 99, 100.
Digital filter
Function to add digital filtering to the fluctuation of pressure value. Smooth the fluctuation of displayed value for sharp start up or fall of the pressure. When the function is valid, digital filtering is reflected to the ON/OFF of the switch output. Output chattering or flicker in the measurement mode display can be reduced by setting the digital filter.
Display accuracy
Shows The maximum deviation between the displayed pressure value and the true pressure.
Display colour
Indicates the colour of the number of digital display. Always green, always red, green (switch OFF) red (switch ON), red (switch OFF)  green (switch ON) are available.
Display resolving power
Indicate in how many the rated pressure range can be divided to display. (Example: When the value can be displayed down to 0.001 MPa for the product for 0 to 1 Mpa, the resolution is 1/1000)
Display value fine adjustment (function)
Displayed pressure value can be adjusted within the range of 5%R.D. (5% of displayed value). It is used if the true pressure value is known, or to eliminate differences between the displayed values of different instruments that are measuring the same pressure.
E
Error displayed
The code number displayed, identifying the error detected by the self-diagnosis function of the pressure switch. Refer to "Error indication function" on page 81 for details of the errors.
Error output
Switches the switch output to ON/OFF when an error is displayed. Refer to "List of output modes" on page 31 for operating conditions. Refer to "Error indication function" on page 81 for details of the errors.
-12-
No.PS※※-OMV0004-C
Term
Definition
F
F.S. (full span/full scale)
Abbreviation of full span and full scale; difference between the minimum and maximum rated pressure values. means the maximum fluctuation range of the pressure switch rated value. For example, when the rated pressure range is -0.100 to 1.000 [MPa]: F.S. = 1.000 - (-0.100) = 1.100 [MPa] (Reference: 1%F.S. = 1.100 x 0.01 = 0.011 [MPa])
Fine adjustment mode
Refer to "Display value fine adjustment (function)".
Fluid contact part (or wetted part)
Part of the pressure switch which contacts detected fluid. Pressure sensor, seal and fitting are included.
Function selection mode
A mode in which setting of functions is performed. It is a separate menu from the pressure setting. If any function settings need to be changed from the factory default, each setting can be selected with "F". The setting items are: display colour, operation mode, output type, digital filter, display resolution, display value fine adjustment, use of auto preset, use of power saving mode, security code, etc.
H
Hysteresis
Difference between the points at which the pressure switch is turned ON and OFF.
Hysteresis mode
Refer to the "List of output modes" on page 31.
I
Insulation resistance
Insulation resistance of the product. The resistance between the electrical circuit and the case.
K
Key-lock function
Function that prevents changes to the settings of the Pressure switch (disables button operation).
M
Manual setting
Manual pressure setup without using auto preset. This term is used to distinguish between manual and auto preset pressure setup.
Maximum applied voltage
The maximum voltage that can be connected to the output of an NPN device.
Maximum load current
The maximum current that can flow to the output (output line) of the switch output.
Measurement mode
Operating condition in which pressure is being detected and displayed, and the switch function is working.
Min. setting unit
Refer to "digit".
N
Normal output
One of the switch output types. In hysteresis mode the switch output is turned ON when pressure equal to or greater than the switch output set value is detected. In window comparator mode, the switch output is turned ON when pressure between the switch output set values (P1L to P1H) is detected. (Refer to the "List of output modes" on page 31.)
O
Operation light
A light that turns on when the switch output is ON.
Operation mode
Either hysteresis mode or window comparator mode can be selected.
Output style
The operation principle of the switch output. Normal output and reverse output can be selected. Please refer to the" List of output modes" on page 31 operating conditions.
-13-
No.PS※※-OMV0004-C
Term
Definition
P
Peak value display (mode)
Shows the maximum pressure from when the power was supplied to the current time.
Port size
The diameter of the connecting part of the switch for connecting with the object to be measured.
Power saving mode
Operating mode in which the digital display turns off and power consumption is reduced.
Pressure setting
The set pressure value that determines the point at which the pressure switch turns ON and OFF.
Proof pressure
Pressure limit that if exceeded will result in mechanical and/or electrical damage to the product.
R
R.D.
Current read value For example, when the display value is 1.000[MPa], 5%R.D. is 5% of
1.000[MPa], which becomes 0.05[MPa]. When the display value is
0.800[MPa], 5%R.D. is 5% of 0.800[MPa], which becomes 0.04[MPa].
Rated pressure range
The pressure range within which the product will meet all published specifications. Values outside of this range can be set as long as they are within the set pressure range, but the specifications cannot be guaranteed.
Repeatability
Variation in repeated measurement of pressure display or ON-OFF output point when the pressure changes at 25 centigrade.
Residual voltage
The difference between the ideal ON voltage and the actual voltage when the switch output is on. Varies with load current. Ideally should be 0 V.
Resolution
Refer to "Display resolution".
Reversed output
One of the switch output types. In hysteresis mode the switch output is turned ON when pressure less than or equal to the switch output set value is detected. In window comparator mode, the switch output is turned ON when pressure is outside the switch output set values (n1L to n1H) is detected. (Refer to the "List of output modes" on page 31.)
Ripple
A type of chattering.
S
Set pressure range
The pressure range that can be set for switch output.
Switch output
Sometimes referred to as "ON-OFF output".
U
Units selection function
A function to change the units in which the measured pressure value is displayed. The display units can only be changed if the product is equipped this function. It is not possible to purchase the product with this function if the product is used in Japan. The product for Japan is displayed in SI only.
W
Window comparator mode
An operating mode in which the switch output is turned on and off depending on whether the flow is inside or outside the range of two set values. (Refer to the "List of output modes" on page 31.)
Withstand voltage
A measure of the product’s resistance to a voltage applied between the
electrical circuit and case. Durability in withstanding voltage. The product may be damaged if a voltage over this value is applied. (The withstand voltage is not the supply voltage used to power the product.)
Z
Zero-clear function
This function to adjust the displayed pressure to zero.
-14-
No.PS※※-OMV0004-C
Mounting and Installation
Installation
○Mounting with bracket
●Mount the bracket to the body with mounting screws (Cross recessed round head screw: Nominal size
M4 x 6L (2 pcs)), then set the body to the specified position.
: Tighten the bracket mounting screws to a torque of 0.76±0.1 Nm.
•Bracket A (Part No.: ZS-50-A)
•Bracket B (Part No.: ZS-50-B)
-15-
No.PS※※-OMV0004-C
Piping
Piping specification: -02 and -N02
After hand tightening, tighten the fitting using a spanner on the flat surfaces of the fitting. The tightening torque must be 8 to 12 Nm.
Piping specification: -F02 After hand tightening, tighten the fitting using a spanner on the flat surfaces of the fitting. The tightening torque must be 4 to 5 Nm.
When tightening, do not hold the upper part of the product (Display) with a spanner.
-16-
No.PS※※-OMV0004-C
The upper part (display) of the product can be rotated by 336°.
Rotating the display with excessive force will damage the end stopper.
336° rotation
-17-
No.PS※※-OMV0004-C
Wiring
Connections should be made with the power supply turned off.
Use a separate route for the product wiring and any power or high voltage wiring. Otherwise, malfunction
may result due to noise.
If a commercially available switching power supply is used, be sure to ground the frame ground (FG)
terminal. If the switching power supply is connected, switching noise will be superimposed and it will not be able to meet the product specifications. In that case, insert a noise filter such as a line noise filter/ferrite between the switching power supplies or change the switching power supply to the series power supply.
How to use connector
Align the cable connector key groove with the product connector key to insert and rotate the knurled part
of the connector.
Connect the wires of the lead wire with M12 connector as shown below.
M12 connector (Port Class A)
Used as switch output device
No.
Name
Lead wire colour
Function
1
DC(+)
Brown
12 to 24 VDC
2
OUT2
White
Switch output 2
3
DC(-)
Blue
0 V 4 OUT1
Black
Switch output 1
Used as IO-Link device
No.
Name
Lead wire colour
Function
1
L+
Brown
18 to 30 VDC
2
DO
White
Switch output 2
3
L-
Blue
0 V
4
C/Q
Black
Communication data (IO-Link)/Switch output 1 (SIO)
-18-
No.PS※※-OMV0004-C
Wiring examples
Used as switch output device
Setting of NPN open collector 2 outputs
Setting of PNP open collector 2 outputs
Used as IO-Link device
: Numbers in the figures show the connector pin layout.
-19-
No.PS※※-OMV0004-C
Outline of Settings [Measurement mode]
Power is supplied
The product code is displayed for approximately 3 seconds after power is supplied.
Then, measurement mode will be displayed.
: Within approximately 0.2 seconds after power-on, the switch starts.
[Measurement mode]
Detects the pressure after power is supplied, and indicates the display and switch operating status. This is the basic mode; other modes should be selected for set-point changes and other function settings.
Measurement mode screen
Sub display
In measurement mode, the sub display can be temporarily changed by pressing the UP or DOWN buttons.
: One arbitrary display mode can be added to the sub display by setting the [F10] sub display.
If the sub display is switched during the arbitrary display setting, the display will be returned to the arbitrary display 30 seconds later. (The default setting does not include arbitrary display.)
Press the SET button once.
Press the SET button between 1 and 3 seconds
Press the SET button between 3 and 5 seconds
[3 step setting
mode]
Set either of set
value or hysteresis.
(Refer to page 21)
[Simple setting
mode]
Select the set
value, hysteresis
and delay time.
(Refer to page 23)
[Function selection
mode]
Change the
function settings.
((Refer to page 25)
[Other Settings]
•Zero-clear function
•Key-lock function
((Refer to page 55)
: The outputs will continue to operate during setting. : If a button operation is not performed for 30 seconds during the setting, the display will flash.
(This is to prevent the setting from remaining incomplete if, for instance, an operator were to leave during setting.)
: 3 step setting mode, simple setting mode and function selection mode settings are reflected each other.
-20-
No.PS※※-OMV0004-C
Pressure Setting
Default settings
When the pressure exceeds the set value, the switch will be turned on. When the pressure falls below the set value by the amount of hysteresis or more, the switch will be turned off. The default setting is to turn on the pressure switch when the pressure reaches the center of the atmospheric pressure and upper limit of the rated pressure range. If this condition, shown to the below, is acceptable, then keep these settings.
Default setting per pressure range
Item
ISE70
ISE71
[P_1] Set value of OUT1 [MPa]
0.500
0.800
[H_1] Hysteresis of OUT1 [MPa]
0.050
0.080
: OUT2 will have the same set value and hysteresis.
-21-
No.PS※※-OMV0004-C
3-Step Setting Mode
3 step setting mode
In this mode, the set values can be input in just 3 steps. Use this mode if the product is to be used straight away, after changing only the set values. (The current pressure value is displayed on the main display.)
<Operation>
[3 step setting mode (hysteresis mode)] In the 3 step setting mode, the set value (P_1 or n_1) and hysteresis (H_1) can be changed. Set the items on the sub display (set value or hysteresis) with the UP or DOWN button. When changing the set value, follow the operation below. The hysteresis setting can be changed in the same way.
(1) Press the SET button once when the item to be changed is displayed on the sub display.
The set value on the sub display (right) will start flashing.
(2) Press the UP or DOWN button to change the set value.
The set value can be increased with the UP button and can be reduced with the DOWN button.
Press the UP button once to increase the value by one digit, press and hold to continuously increase.
Press the DOWN button once to reduce the value by one digit, press and hold to continuously
reduce.
When the UP and DOWN buttons are pressed and held simultaneously for 1 second or longer, the
set value is displayed as [- - -], and the set value will be the same as the current pressure value automatically (snap shot function (Refer to page 55.)). Afterwards, it is possible to adjust the value by pressing the UP or DOWN button.
(3) Press the SET button to complete the setting.
-22-
No.PS※※-OMV0004-C
The Pressure switch turns on within a set pressure range (from P1L to P1H) during window comparator mode. Set P1L, the lower limit of the switch operation, and P1H, the upper limit of the switch operation and WH1 (hysteresis) following the instructions given on page 21. (When reversed output is selected, the sub display (left) shows [n1L] and [n1H].) Please refer to the "List of output modes" on page 31 for the relationship between the set values and operation.
: Set OUT2 in the same way.
Setting of the normal/reverse output switching and hysteresis/window comparator mode switching are performed with the function selection mode [F 1] OUT1 setting and [F 2] OUT2 setting.
-23-
No.PS※※-OMV0004-C
Simple Setting Mode
<Operation>
[Simple setting mode (hysteresis mode) In the simple setting mode, the set value, hysteresis and delay time can be changed while checking the current pressure value (main display).
(1) Press and hold the SET button between 1 and 3 seconds in measurement mode. [SEt] is displayed on
the main display. When the button is released while in the [SEt] display, the current pressure value is displayed on the main display, [P_1] or [n_1] is displayed on the sub display (left), and the set value is displayed on the sub display (right) (Flashing).
(2) Change the set value with the UP or DOWN button, and press the SET button to set the value. Then,
the setting moves to hysteresis setting. (The snap shot function can be used. (Refer to page 55.))
(3) Change the set value with the UP or DOWN button, and press the SET button to set the value. Then,
the setting moves to the delay time of the switch output. (The snap shot function can be used. (Refer to page 55.))
(4) The delay time of the switch output can be selected by pressing the UP or DOWN button at the ON and
OFF point of the switch output. Delay time setting can prevent the output from chattering. The delay time can be set in the range 0.00 to 60.00 sec. in 0.01 sec. increments.
(5) Press the SET button for less than 2 seconds to complete the OUT1 setting.
[P_2] or [n_2] is displayed on the sub screen (left). Continue with setting the OUT2.
Press and hold the SET
button for 2 seconds or longer to complete the setting. The product will return
to measurement mode.
1: Selected items (1) to (4) become valid after pressing the SER button. 2: After enabling the setting by pressing the SET button, it is possible to return to measurement mode by pressing the SET
button for 2 seconds or longer.
3: When the output mode (refer to page 29) is set to error output or output OFF, the simple setting mode cannot be used.
(The setting changes to measurement mode by releasing the button when [SEt] is displayed.)
-24-
No.PS※※-OMV0004-C
In the window comparator mode, set P1L, the lower limit of the switch operation, and P1H, the upper limit of the switch operation, WH1 (hysteresis) and dtH/dtL (delay time) following the instructions given on page 23. (When reversed output is selected, the sub display (left) shows [n1L] and [n1H].) Please refer to the "List of output modes" on page 31 for the relationship between the set values and operation.
: Set OUT2 in the same way.
-25-
No.PS※※-OMV0004-C
Function Selection Mode
Function selection mode
In measurement mode, press the SET button between 3 and 5 seconds, to display [F 0]. Select to display the function to be changed [F□□]. Press and hold the SET button for 2 seconds or longer in function selection mode to return to measurement mode.
: Some products do not have all the functions. If no function is available or selected due to configuration of other functions, [- - -] is
displayed on the sub display (right).
Default setting
The default setting is as follows. If no problem is caused by this setting, keep these settings. To change a setting, enter function selection mode.
[F 0] Display units, switch output specifications and diagnostic information selection function Page 27
Item
Default setting
Display units
Units specification ["Nil" or M]: MPa
Switch output specifications
PNP
Diagnostic information
ALL
[F 1] Setting of OUT1 Page 29
Item
Explanation
Default setting
Output mode
Either hysteresis mode, window comparator mode, error output or switch output OFF can be selected.
Hysteresis mode
Reversed output
Selects which type of switch output is used, normal or reversed.
Normal output
Pressure setting
Sets the ON and OFF point of the switch output.
ISE70: 0.500 MPa ISE71: 0.800 MPa
Hysteresis
Appropriate setting of the hysteresis will prevent the switch output from chattering.
ISE70: 0.050 MPa ISE71: 0.080 MPa
Delay time
Delay time of the switch output can be selected.
0.00 sec.
Display colour
Select the display colour.
Output ON: Green Output OFF: Red
(Linked to OUT1)
-26-
No.PS※※-OMV0004-C
[F 2] Setting of OUT1 Page 32
Item
Explanation
Default setting
Output mode
Either hysteresis mode, window comparator mode, error output or switch output OFF can be selected.
Hysteresis mode
Reversed output
Selects which type of switch output is used, normal or reversed.
Normal output
Pressure setting
Sets the ON and OFF point of the switch output.
ISE70: 0.500 MPa ISE71: 0.800 MPa
Hysteresis
Appropriate setting of the hysteresis will prevent the switch output from chattering.
ISE70: 0.050 MPa ISE71: 0.080 MPa
Delay time
Delay time of the switch output can be selected.
0.00 sec.
Display colour
Select the display colour.
Output ON: Green Output OFF: Red
(Linked to OUT1)
Other parameter settings
Item
Page
Default setting
[F 3] Digital filter setting
Page 34
0.00 ms
[F 4] Auto-preset
Page 35
OFF
[F 6] Fine adjustment of display value
Page 37
0.0%
[F10] Sub display setting
Page 38
std (Standard)
[F11] Display resolution setting
Page 44
1000-split
[F14] Zero cut-off setting
Page 45
0.0
[F80] Power saving mode
Page 46
OFF
[F81] Security code
Page 47
OFF
[F90] Setting of all functions
Page 49
OFF
[F96] Number of pressurizing errors
Page 51
[F98] Output check
Page 52
N/A (normal output)
[F99] Reset to default settings
Page 54
OFF