SMC Networks HRS012-W*-10-*, HRS012-A*-20-*, HRS018-A*-10-*, HRS018-W*-10-*, HRS012-W*-20-* Operation Manual

...
Page 1
HRX-OM-M090
1
st
edition: Jul. 2009 Rev.M: Mar. 2013
Operation Manual
Original Instructions
Thermo chiller
Air-Cooled
refrigerated type
Water-Cooled
refrigerated type
HRS012-A∗-10-∗ HRS012-W∗-10-∗ HRS018-A∗-10-∗ HRS018-W∗-10-∗ HRS012-A∗-20-∗ HRS012-W∗-20-∗ HRS018-A∗-20-∗ HRS018-W∗-20-∗ HRS024-A∗-20-∗ HRS024-W∗-20-∗
Keep this manual available whenever necessary
© 2013 SMC CORPORATION All Rights Reserved
Page 2
To the users
Thank you for purchasing SMC’s Thermo chiller (hereinafter referred to as the “product”).
For safety and long life of the product, be sure to read this operation manual (hereinafter referred to as the “manual”) and clearly understand the contents.
Be sure to read and follow all instructions noted with “Warning” or “Caution” in this manual.
This manual is intended to explain the installation and operation of the product. Only people who
understand the basic operation of the product through this manual or who perform installation and operation of or have basic knowledge about industrial machines are allowed to work on the product.
This manual and other documents attached to the product do not constitute a contract, and will not affect any existing agreements or commitments.
It is strictly prohibited to copy this manual entirely or partially for the use by a third party without prior permission from SMC.
Note: This manual is subject to possible change without priornotice.
Page 3
HRX-OM-M090
Contents
HRS Series
Contents
Chapter 1 Safety Instructions .......................................................... 1-1
1.1 Before using the product ..........................................................................................1-1
1.2 Reading the Manual ...................................................................................................1-1
1.3 Hazards .......................................................................................................................1-2
1.3.1 Level of hazards.................................................................................................................. 1-2
1.3.2 Definition of “Serious injury” and “Minor injury”................................................................... 1-2
1.3.3 Types of hazard labels ........................................................................................................ 1-3
1.3.4 Locations of Hazard Labels................................................................................................. 1-4
1.4 Other Labels ............................................................................................................... 1-5
1.4.1 Product Label ...................................................................................................................... 1-5
1.5 Safety Measures.........................................................................................................1-5
1.5.1 Safety Instructions for Use .................................................................................................. 1-5
1.5.2 Personal Protective Equipment........................................................................................... 1-6
1.6 Emergency Measures ................................................................................................ 1-7
1.7 Waste disposal...........................................................................................................1-8
1.7.1 Disposal of refrigerant and compressor oil.......................................................................... 1-8
1.7.2 Disposal of product.............................................................................................................. 1-8
1.8 Material Safety Data Sheet (MSDS) ..........................................................................1-8
Chapter 2 Name and Function of Parts............................................. 2-1
2.1 Part number of product .............................................................................................2-1
2.2 Name and Function of Parts .....................................................................................2-2
2.3 Function of Parts........................................................................................................2-3
2.4 Operation display panel ............................................................................................2-4
Chapter 3 Transport and Setting Up ................................................. 3-1
3.1 Transport ....................................................................................................................3-1
3.1.1 Transportation using casters............................................................................................... 3-2
3.2 Installation .................................................................................................................. 3-3
3.2.1 Environment ........................................................................................................................ 3-4
3.2.2 Location (Required ventilation rate and facility water source) ............................................ 3-5
3.2.3 Installation and Maintenance Space ................................................................................... 3-6
3.3 Installation .................................................................................................................. 3-7
3.3.1 Mounting.............................................................................................................................. 3-7
3.3.2 Electrical wiring ................................................................................................................... 3-8
3.3.3 Preparation and wiring of power supply cable .................................................................... 3-9
3.3.4 Wiring of remote operation signal input..............................................................................3-11
3.3.5 Wiring of operation signal output and alarm signal output................................................ 3-13
3.3.6 RS-485 Communication wiring.......................................................................................... 3-14
3.3.7 RS-232C Communication wiring ....................................................................................... 3-15
3.4 Piping ........................................................................................................................ 3-16
Page 4
HRX-OM-M090 Contents
HRS Series
3.5 Fill of circulating fluid..............................................................................................3-18
3.6 Option J Piping ofAutomatic fluid filling....................................................... 3-19
3.7 Wiring of external switch ........................................................................................3-20
3.7.1 Reading of the External switch .......................................................................................... 3-21
3.7.2 Wiring................................................................................................................................. 3-22
3.7.3 Setting items ...................................................................................................................... 3-23
Chapter 4 Starting the Product.......................................................... 4-1
4.1 Before Starting........................................................................................................... 4-1
4.2 Preparation for Start ..................................................................................................4-2
4.2.1 Power supply .......................................................................................................................4-2
4.2.2 Setting of circulating fluid temperature ................................................................................4-2
4.3 Preparation of circulating fluid.................................................................................4-3
4.4 Starting and Stopping................................................................................................4-6
4.4.1 Starting the product..............................................................................................................4-6
4.4.2 Stopping the product............................................................................................................4-7
4.5 Check items after starting.........................................................................................4-8
4.6 Adjustment of Circulating Fluid ...............................................................................4-8
Chapter 5 Display and setting of various functions........................ 5-1
5.1 Function...................................................................................................................... 5-2
5.1.1 Key operations.....................................................................................................................5-2
5.1.2 List of parameters ................................................................................................................ 5-4
5.2 Main screen ................................................................................................................5-7
5.2.1 Main screen ......................................................................................................................... 5-7
5.2.2 Display on the main screen .................................................................................................5-7
5.3 Alarm display menu................................................................................................... 5-8
5.3.1 Alarm display menu .............................................................................................................5-8
5.3.2 Content of display of alarm display menu............................................................................ 5-8
5.4 Inspection monitor menu .......................................................................................... 5-9
5.4.1 Inspection monitor menu .....................................................................................................5-9
5.4.2 Checking of the Inspection monitor menu ...........................................................................5-9
5.5 Key-lock.................................................................................................................... 5-13
5.5.1 Key-lock ............................................................................................................................. 5-13
5.5.2 Key-lock setting / checking ................................................................................................5-14
5.6 Run timer, stop timer function................................................................................ 5-15
5.6.1 Run timer and stop timer function...................................................................................... 5-15
5.6.2 Setting and checking of Run timer and stop timer function...............................................5-17
5.7 Signal for completion of preparation (TEMP READY) ..........................................5-19
5.7.1 Signal for completion of preparation (TEMP READY).......................................................5-19
5.7.2 Signal for completion of preparation (TEMP READY) setting / checking.......................... 5-20
5.8 Offset function .........................................................................................................5-22
Page 5
HRX-OM-M090
Contents
HRS Series
5.8.1 Offset function ................................................................................................................... 5-22
5.8.2 Offset function setting and checking ................................................................................. 5-24
5.9 Function to recover from power failure .................................................................5-26
5.9.1 Function to recover from power failure.............................................................................. 5-26
5.9.2 Function to recover from power failure setting and checking........................................... 5-27
5.10 Anti-freezing function..............................................................................................5-28
5.10.1 Anti-freezing function......................................................................................................... 5-28
5.10.2 Anti-freezing function setting and checking....................................................................... 5-29
5.11 Key click sound setting...........................................................................................5-30
5.11.1 Key click sound setting...................................................................................................... 5-30
5.11.2 Key click sound setting and checking ............................................................................... 5-30
5.12 Temperature unit switching ....................................................................................5-31
5.12.1 Temperature unit switching ............................................................................................... 5-31
5.12.2 Temperature unit switching setting and checking............................................................. 5-31
5.13 Pressure unit switching ..........................................................................................5-32
5.13.1 Pressure unit switching ..................................................................................................... 5-32
5.13.2 Pressure unit switching setting and checking .................................................................. 5-32
5.14 Alarm buzzer sound setting....................................................................................5-33
5.14.1 Alarm buzzer sound setting............................................................................................... 5-33
5.14.2 Alarm buzzer sound setting and checking........................................................................ 5-33
5.15 Alarm customize function ....................................................................................... 5-34
5.15.1 Alarm customize function .................................................................................................. 5-34
5.15.2 Alarm customize function setting and checking ................................................................ 5-36
5.15.3 Setting of temperature alarm monitoring method and alarm generation timing................ 5-49
5.16 Data reset function...................................................................................................5-51
5.16.1 Data reset function ............................................................................................................ 5-51
5.16.2 Method of resetting data reset function............................................................................. 5-51
5.17 Accumulated time reset function ...........................................................................5-52
5.17.1 Accumulated time reset function ....................................................................................... 5-52
5.17.2 Method of resetting accumulated time reset function ....................................................... 5-52
5.18 Communication function......................................................................................... 5-54
5.18.1 Communication function.................................................................................................... 5-54
5.18.2 Communication function setting and checking.................................................................. 5-54
5.19 Option J [Automatic fluid filling] ............................................................................5-71
5.19.1 Option J [Automatic fluid filling] ......................................................................................... 5-71
5.20 Option [Drain pan set] .............................................................................................5-72
5.20.1 Option [Drain pan set] ....................................................................................................... 5-72
5.20.2 Option [Drain pan set] setting and checking ..................................................................... 5-73
5.21 Option [Electric resistivity sensor set] ..................................................................5-74
5.21.1 Option [Electric resistivity sensor set] ............................................................................... 5-74
Page 6
HRX-OM-M090 Contents
HRS Series
5.21.2 Option [Electric resistivity sensor set] setting and checking..............................................5-75
Chapter 6 Alarm indication and trouble shooting ........................... 6-1
6.1 Alarm Display............................................................................................................. 6-1
6.2 Alarm buzzer stop...................................................................................................... 6-3
6.3 Troubleshooting.........................................................................................................6-4
6.4 Other Errors................................................................................................................6-6
Chapter 7 Control, Inspection and Cleaning.................................... 7-1
7.1 Control of Circulating Fluid Quality ......................................................................... 7-1
7.2 Inspection and Cleaning ...........................................................................................7-2
7.2.1 Daily check ..........................................................................................................................7-2
7.2.2 Monthly check......................................................................................................................7-3
7.2.3 Inspection every 3 months................................................................................................... 7-4
7.2.4 Inspection every 6 months................................................................................................... 7-4
7.2.5 Inspection for winter season................................................................................................ 7-5
7.3 Consumables .............................................................................................................7-5
7.4 Stop for a Long Time ................................................................................................. 7-6
7.4.1 Discharge of the circulating fluid and facility water.............................................................. 7-6
Chapter 8 Documents......................................................................... 8-1
8.1 Specifications List .....................................................................................................8-1
8.1.1 Product specification............................................................................................................8-1
8.1.2 Communication specifications .............................................................................................8-3
8.2 Outline dimensions....................................................................................................8-4
8.3 Flow Chart ..................................................................................................................8-5
8.3.1 HRS012-A∗-∗0, HRS018-A∗-∗0, HRS024-A-20............................................................... 8-5
8.3.2 HRS012-W∗-∗0, HRS018-W∗-∗0, HRS024-W-20............................................................ 8-5
8.4 Cooling capacity ........................................................................................................8-6
8.4.1 HRS012-∗∗-10-(BJM) ..........................................................................................................8-6
8.4.2 HRS018-∗∗-10-(BJM) ..........................................................................................................8-6
8.4.3 HRS012-∗∗-20-(BJMT) ........................................................................................................8-7
8.4.4 HRS018-∗∗-20-(BJMT) ........................................................................................................8-7
8.4.5 HRS024-∗∗-20-(BJMT) ........................................................................................................8-7
8.4.6 Option G [High temperature type]........................................................................................ 8-8
8.5 Heating capacity ........................................................................................................8-9
8.5.1 HRS012-∗∗-10,HRS018-∗∗-10 ............................................................................................8-9
8.5.2 HRS012-∗∗-20,HRS018-∗∗-20, HRS024-∗∗-20................................................................... 8-9
8.6 Pump capacity..........................................................................................................8-10
8.6.1 HRS012-∗∗-10-(BJM), HRS018-∗∗-10-(BJM)....................................................................8-10
8.6.2 HRS012-∗∗-20-(BGJM), HRS018-∗∗-20-(BGJM), HRS024-∗∗-20-(BGJM)....................... 8-10
8.6.3 Option:-T............................................................................................................................ 8-11
8.6.4 Option:-MT......................................................................................................................... 8-11
Page 7
HRX-OM-M090
Contents
HRS Series
8.7 Required facility water flow (for water-cooled type)............................................. 8-12
8.8 Compliance...............................................................................................................8-13
8.9 Sample DoC..............................................................................................................8-14
8.10 Daily Check Sheet....................................................................................................8-15
Chapter 9 Product Warranty .............................................................. 9-1
Page 8
HRX-OM-M090 Contents
HRS Series
Page 9
HRX-OM-M090
Chapter 1 Safety Instructions
HRS Series 1.1 Before using the product
1-1
Chapter 1 Safety Instructions
1.1 Before using the product
z
This chapter is intended to specifically describe the safety related issues for handling the product. Read this before handling the product.
z
The product is a cooling device using circulating fluid. SMC does not take any responsibility for any problems that may arise from using the product for other purposes.
z
This product is for indoor use only and not to be used outdoors.
z
This product is not designed for a clean room. It generates dust from the internal components such as pump and fan motor.
z
The product is operated at high voltage and contains components which become hot and rotate. If a component needs to be replaced or repaired, contact a specialized vendor for parts and service.
z
All personnel who work with or around the product should read and understand the safety related information in this manual carefully before starting work.
z
The safety manager is responsible for strictly observing safety standards, but responsibility in respect to safety standards during daily work resides with each individual operator and maintainance personnel.
z
This manual must be kept available to operators whenever necessary.
1.2 Reading the Manual
This manual contains symbols to help identify important actions when installing, operating or maintaining the product.
Before using the product be sure to read and understand all the important actions highlighted in this manual.
This sign indicates actions that must be followed.
This sign indicates prohibited actions.
Page 10
HRX-OM-M090
Chapter 1 Safety Instructions
1.3 Hazards HRS Series
1-2
1.3 Hazards
1.3.1 Level of hazards
The instructions given in this manual aim to assure the safe and correct operation of the product, and to prevent injury of operators or damage to the product. These instructions are grouped into three categories, Danger, Warning and Caution, which indicate the level of hazard, damage and also the degree of emergency. All safety critical information should be carefully observed at all times.
“DANGER”, “WARNING” and “CAUTION” signs are in order according to severity (DANGER> WARNING> CAUTION).
1.3.2 Definition of “Serious injury” and “Minor injury”
“Serious injury”
This term describes injuries that result in after effects including loss of eyesight, burns, electrical shock, fracture, poisoning, etc. and requires long-term treatment or hospitalization.
“Minor injury”
This term describes injuries that do not need long-term treatment or hospitalization. (Others excluded from serious injury.)
“DANGER”: Hazard that WILL cause serious personal injury or death during operation.
“WARNING”: Hazard that MAY cause serious personal injury or death during operation.
“CAUTION”: Hazard that MAY cause minor personal injury.
“CAUTION without exclamation symbol”: Hazard that MAY cause damage or failure of the product, facility, devices, ect.
Page 11
HRX-OM-M090
Chapter 1 Safety Instructions
HRS Series 1.3 Hazards
1-3
1.3.3 Types of hazard labels
The product has various potential hazards and they are marked with warning labels. Be sure to read this section before starting any work on the product.
Warning related to electricity
Warning related to high temperatures
Warning related to rotating objects
Warning related to other general dangers
This symbol stands for a possible risk of electric shock.
The product is operated at high voltage and contains uncovered live terminals inside. z DO NOT operate the product without cover panels fitted. DO NOT work inside this product unless you have been trained to do
so.
This symbol stands for general danger.
Hazards Inside
Hot Surfaces Inside – See Hot Surface symbol Rotating Fan Inside – See Rotating Fan symbol (For air-cooled type) Pressurized Sytem Inside – The product contains pressurised fluid systems.
z DO NOT operate the product without cover panels fitted.
This symbol stands for a possible risk of hot surface and burns.
The product has surfaces that can reach high temperatures during operation. Even after the power is turned off, there can still be residual heat in the product.
z DO NOT operate the product without cover panels fitted. z DO NOT start working inside the product until the temperature has
decreased sufficiently.
This symbol stands for a possible risk of cutting fingers or hand, or entanglement by rotating fan (For air-cooled type). The product contains a cooling fan that rotates during operation of the product. The fan can start and stop intermittently and without warning. z DO NOT operate the product without cover panels fitted.
Page 12
HRX-OM-M090
Chapter 1 Safety Instructions
1.3 Hazards HRS Series
1-4
1.3.4 Locations of Hazard Labels
There are various warning labels on the product to show the potential hazards.
Fig. 1-1 Warning label position
For air-cooled type
Fig.
1-2 Warning label position
Rear
Front
Page 13
HRX-OM-M090
Chapter 1 Safety Instructions
HRS Series 1.4 Other Labels
1-5
1.4 Other Labels
1.4.1 Product Label
Information about the product, such as Serial No. and Model No. can be found on the model label. This information is needed when contacting an SMC sales distributor.
How to see the manufacturing code R
O 001January 2013
O O 001
Year Symbol Remarks Month Symbol Remarks
Serial
no.
2013 R
1 O
2014 S 2 P 2015 T 3 Q
Repeated from A to Z in alphabetical order
Repeated from O to Z in alphabetical order, with O for January and Z for December
Fig. 1-3 Position of product label
1.5 Safety Measures
1.5.1 Safety Instructions for Use
z
Read and understand this manual carefully before using the product.
z
Before starting maintenance of the product, be sure to lock out and tag out the breaker of the user's power supply.
z
If operating the product during maintenance, be sure to inform all workers nearby.
z
Use only the correct tools and procedure when installing or maintaning the product.
z
Use personal protective equipment where specified (“1.5.2 Personal Protective Equipment”)
z
Check all parts and screws are fitted correctly and securely after maintenance.
z
Avoid working in a drunken or sick condition, which might cause an accident.
z
Do not remove the panels except for the cases permitted in this manual.
z
Do not remove the panels during operation.
Follow the instructions below when using the product. Failure to follow the instructions may cause an accident and injury.
(It is an example of model "HRS018-A-20".)
Model Number
Serial Number (SERIAL No.)
The type and quantity of refrigerant (REFRIGERANT)
Page 14
HRX-OM-M090
Chapter 1 Safety Instructions
1.5 Safety Measures HRS Series
1-6
1.5.2 Personal Protective Equipment
This manual specifies personal protective equipment for each work.
Transport, Installing and Uninstalling
Handling of circulating fluid
Operation
Always use safety shoes, gloves and head protection when transporting, installing or uninstalling the product.
Always use safety shoes, gloves, mask, apron and eye protection when handling the circulating fluid.
Always use safety shoes and gloves when operating the product.
Page 15
HRX-OM-M090
Chapter 1 Safety Instructions
HRS Series 1.6 Emergency Measures
1-7
1.6 Emergency Measures
When emergency conditions such as natural disaster, fire and earthquake, or injury occurs, turn off the power supply switch. The switch is located at the front of the product.
1. Turn off the power supply switch at the back of the product to stop the operation of the
thermo-chiller.
Fig. 1-4 Location of the switch for the power supply
2. Be sure to shut off the breaker of the facility power supply (the power supply of the
user’s machine).
Fig. 1-5 Shut off of facility power supply
Even when the power supply swich is turned off, some of the internal circuits are still energized, unless the user’s power supply is shut off. Be sure to shut off the breaker of the user’s power supply.
Page 16
HRX-OM-M090
Chapter 1 Safety Instructions
1.7 Waste disposal HRS Series
1-8
1.7 Waste disposal
1.7.1 Disposal of refrigerant and compressor oil
The product uses hydrofluorocarbon type refrigerant (HFC) and compressor oil. Comply with the laws and regulations in each country for the disposal of refrigerant and compressor oil. The type and quantity of refrigerant is described on the 1.4.1 Product Label.
If these fluids need to be recovered, read and understand the instructions below carefully. If there is any unclear point, contact an SMC's sales distributor.
1.7.2 Disposal of product
The disposal of the product must be handled by a specialized industrial waste disposal agency in accordance with local laws and regulations.
1.8 Material Safety Data Sheet (MSDS)
If the material safety data sheets of chemicals used in this product are needed, contact an SMC's sales distributor.
Any chemicals used by the user must be accompanied by an MSDS.
Comply with the laws and regulations in each country for the disposal
of refrigerant and compressor oil.
The release of refrigerant in to the atmosphere is banned by law.
Recover it with specific equipment and dispose of it correctly.
Only people who have sufficient knowledge and experience about the
product and its accessories are allowed to recover the refrigerant and compressor oil.
Only maintenance personnel or qualified people are allowed to open
the cover panels of the product.
Do not mix the compressor oil with domestic waste for disposal. Also,
the disposal of the waste must only be conducted by specific facilities that are permitted for that purpose.
Page 17
HRX-OM-M090
Chapter 2 Name and Function of Parts
HRS Series 2.1 Part number of product
2-1
Chapter 2 Name and Function of Parts
2.1 Part number of product
The product can be ordered with the part number configured as shown below. The product needs to be handled in different ways depending on the part number. Refer to “1.4.1 Product Label” and check the part number of the product.
Fig. 2-1 Part number of product
10A012
012
HRS
Cooling capacity
Cooling method
Piping thread type
Power supply
-
Null
F
N
-
Rc
G (PT-G conversion fitting is included)
NPT (PT-NPTconversion fitting is included)
A Air-cooled refrigerator type
W Water-cooled refrigerator type
20 1-phase AC200230V(50/60Hz)
10
1-phase AC100V(50Hz)
1-phase AC100~115V(60Hz)
018
024
1 Power supply-10(100V Type)2 Power supply-20(200V Type)3 Power supplyOnly -20(200V Type)can be chosen.4 Cooling method-A(Air-cooled refrigerator type),
  ④Power supplyOnly -20(200V Type)can be chosen.
1100W/1300W (50/60Hz)
1500W/1700W (50/60Hz)(1)
1700W/1900W (50/60Hz)(∗2) 2100W/2400W (50/60Hz)(∗3)
-
Option
Null None
J Automatic water-supply
M Pure water piping
T High-pressure pump mounted
B Earth leakage breaker
G High temperature type(4)
Page 18
HRX-OM-M090
Chapter 2 Name and Function of Parts
2.2 Name and Function of Parts HRS Series
2-2
2.2 Name and Function of Parts
Fig. 2-2 Names of each part
Table 2-1 Accessories list
1
Quick manual
1pc.
(with a clear cover)
2
Alarm code list label
1pc.
3
Operation Manual
2pcs.
(Jpn: 1pc., Eng:1pc.)
4
Power supply connector
1pc.
5
Sequence I/O command signal connector
1pc.
6
Fitting (for drain port)*
1pc.
7
Ferrite core (for communication)
1pc.
Not included when option [High head pump] is selected.
Page 19
HRX-OM-M090
Chapter 2 Name and Function of Parts
HRS Series 2.3 Function of Parts
2-3
2.3 Function of Parts
The function of parts is as follows.
Table 2-2 Function of parts
Name Function
Operation display panel
Runs and stops the product and performs settings such as the circulating fluid temperature. For details, refer to ’’2.4 Operation display panel’’.
Power supply switch
Shuts off the power supply to the internal eqipment of product. Circuit protector 10A
Model label
Shows the part number of the product. For details, refer to ‘’1.4.1 Product Label’’.
Circulating fluid outlet port The circulating fluid flows out from the outlet port.
Circulating fluid return port The circulating fluid returns to the return port.
Drain port
This drain port to drain the circulating fluid out of the tank. (The plug is connected to standard pump type at the time of shipment. The ball valve is installed in the high pump head type.)
Facility water inlet
(For water-cooled type)
A facility water inlet to which the facility water is fed through piping. The pressure of facility water should be in a range of 0.3 to 0.5MPa.
Facility water outlet
(For water-cooled type)
A facility water outlet from which the facility water returns to the user’s machine through piping.
Automatic water-fill port (When automatic fluid filling [Option] is selected.)
Piping to the automatic fluid filling port enables easy supply of the circulating fluid through the built-in solenoid valve.The supply pressure should be in a range of 0.2 to 0.5MPa.
Overflow port (When automatic fluid filling [Option] is selected.)
This is necessary when automatic fluid filling function.Discharge excess circulating fluid when the fluid level in the tank rises.
Page 20
HRX-OM-M090
Chapter 2 Name and Function of Parts
2.4 Operation display panel HRS Series
2-4
2.4 Operation display panel
The operation panel on the front of the product controls the basic operation of the product.
Fig. 2-3 Operation display panel
Table 2-3 Operation display panel
No Description Function
Reference
page
PV
Displays the temperature and pressure of the circulating fluid and alarm codes.
Digital display
(7 segment,
4 digits)
SV
Displays the set temperature of the circulating fluid and the set values of other menus.
5.2
[ oC oF ] lamp Displays the unit of display temperature (℃ or oF). 5.12
[MPa PSI] lamp Displays the unit of display pressure (MPa or PSI). 5.13
[REMOTE] lamp Lights up during remote operation by communication. 5.18
[RUN] lamp
Lights up when the product is started and in operation. Goes off
when the product is stopped.
Flashes during stand-by for stop (Interval 0.5 seconds). Flashes during independent operation of the pump (Interval
0.3 seconds). Flashes during anti-freezing function (At standby: Interval 2 seconds, At operation: Interval 0.3 seconds).
4.4
[ALARM] lamp Flashes with buzzer when alarm occurs (Interval 0.3 seconds). 5.3
[ ] lamp Lights up when the tank level indicator falls below the LOW level. 4.3
[ ] lamp Lights up while the run timer or stop timer function is working. 5.6
[ ] lamp
Lights up when the product is in automatic operation.
5.9
[RUN/STOP] key
Makes the product start or stop.
4.4
[MENU] key
Shifts the main menu (display screen of temperature) the other menu (entry of set values and monitor screen).
[SEL] key
Changes the item in menu and enters the set value.
5.1
[] key
Decreases the set value.
[] key
Increases the set value.
-
[PUMP] key
When the [MENU] and [RUN/STOP] keys are held down simultaneously, the pump starts running independently.
4.3
[RESET] key
Keep the [] and [] keys pressed down simultaneously. This will stop the alarm buzzer and reset the [ALARM] lamp.
6.3
Page 21
HRX-OM-M090
Chapter 3 Transport and Setting Up
HRS Series 3.1 Transport
3-1
Chapter 3 Transport and Setting Up
3.1 Transport
The product is heavy and has potential danger at transport. Also, to prevent damage and breakage of the product, be sure to follow these instructions for transport.
Only persons who have sufficient knowledge and experience about the
product and system are allowed to transport and set up the product.
Especially pay attention to personal safety.
Never lay the product on its side. The compressor oil will leak in to the refrigerant piping, which may cause early failure of the compressor.
Drain the residual fluid from the piping as much as possible to prevent
any spillage.
Page 22
HRX-OM-M090
Chapter 3 Transport and Setting Up
3.1 Transport HRS Series
3-2
3.1.1 Transportation using casters
1. Release the lock levers of the front casters.
2. Push the handles on the right/left panel or the corner of the product to move the product
to the destination. Do not hold the product by the cap to move it. This will apply excess force to the piping of internal parts which may lead to malfunctions such as fluid leakage.
3. To push the front or rear panel, push it by the corner. Pushing by the center of the panel
may deform the shape.Care should be taken.
Fig. 3-1 Transportation using casters
Handle
Unlocking
Press
Corners
This product is heavy. Care should be taken when the product is transported on a slope.
Page 23
HRX-OM-M090
Chapter 3 Transport and Setting Up
HRS Series 3.2 Installation
3-3
3.2 Installation
Keep the product horizontal to a rigid and flat floor which can resist
the weight of the product, and take measures to prevent the product from tipping over. Improper installation may cause water leakage, tipping, damage of the product or injure the operator.
Keep the ambient temperature of the product between 5 to 40
o
C.
Operation below 5
o
C may cause the compressor failure, and
operation above 40
o
C may cause the product to overheat and shut
down.
The operating ambient temperature is 5 to 45deg. C for option G
[Hig
h temperature type].
Do not set up the product in places possibly exposed to leakage of
flammable gas. Should any flammable gas stay around the product, the product may cause a fire.
Do not use the product outdoors. If the product subjected to rain or
water splash it may cause electrical shock, fire or failure.
Page 24
HRX-OM-M090
Chapter 3 Transport and Setting Up
3.2 Installation HRS Series
3-4
3.2.1 Environment
The product must not be operated, installed, stored or transported in the following conditions. Potential malfunction or damage to the product may occur if these instructions are disregarded.
The product does not conform to any Clean room specifications. The pump and ventilating fan inside the product generate particles.
The installer/end user is responsible for carrying out an acoustic noise risk assessment on the equipment after installation and taking appropriate measures as required.
z
Location that is outside.
z
Location that is exposed to water, water vapour, steam, salt water or oil.
z
Location that is exposed to dust or powder material.
z
Location that is exposed to corrosive gas, organic solvent, chemical solution, or flammable gas (the product is not flame-proof)
z
Location where the ambient temperature is out of the following range: In transportation and In storage 0 to 50°C (with no water or circulating fluid in piping) In operation 5 to 40°C (Option G [High temperature type]:5 to 45°C)
z
Location where the ambient humidity is out of the following range or where condensation occurs: In transportation and storage 15 to 85% In operation 30 to 70%
z
Location that is exposed to direct sunlight or heat radiation.
z
Location that is near heat sources and poor in ventilation.
z
Location that is subjected to abrupt changes in temperature.
z
Location that is subjected to strong electromagnetic noise (intense electric field, intense magnetic field, or surges).
z
Location that is subjected to static electricity, or conditions where static electricity can discharge to the product.
z
Location that is subjected to strong high frequencies raditation (microwaves).
z
Location that is subjected to potential lightning srtike.
z
Location at altitudes of 1000m or higher (except for product storage and transport).
z
Location where the product is affected by strong vibrations or impacts.
z
Condition that applies external force or weight causing the product to be damaged.
z
Location without adequate space for maintenance as required.
Page 25
HRX-OM-M090
Chapter 3 Transport and Setting Up
HRS Series 3.2 Installation
3-5
3.2.2 Location (Required ventilation rate and facility water source)
Installation of multiple products
Keep sufficient space between products so that the air vented from one product will not be taken in by other products.
Installation Area Ventilation (For air-cooled type)
Facility having a large installation area (that can vent the air naturally)
Make an air vent on a wall at a high level and another air vent on a wall at a low level, to allow for adequate airflow.
Facility having a small installation area (that can not vent the air naturally) Make a forced air exhaust vent on a wall at a high level and an air vent on a wall at a low
level.
Table 3-1 Amount of radiation and required ventilation
Required ventilation amount m3/min
Model
Heat
Radiated
kW
Differential temp. of 3 oC
between inside and outside
of installation area
Differential temp. of 6 oC
between inside and outside
of installation area
HRS012-A∗-∗
Approx. 2
40 20
HRS018-A-
Approx. 4
70 40
HRS024-A-20
Approx. 5
90 50
Facility water source to be prepared (For water-cooled type)
Table 3-2 Facility water source to be prepared
Required facility water rate
l/min
Facility water temprature
Model
Heat
Radiated
kW
Facility water temp.
range
o
C
25 32 40
HRS012-W-20 Approx. 2
8 12 20
HRS018-W-20 Approx. 4
12 15 23
HRS024-W-20 Approx. 5
5to40
(Rating 25)
14 17 25
Do not install in a location which can be subjected to any of the
conditions in 3.2.1 Environment.
The product radiates heat from the air vent of the cooling fan. If the product is operated with insufficient air ventilation the internal temperature can exceed 40
oC∗
, which can cause an overload or affect the performance and life of the product. To prevent this ensure that suitable ventilation is available (see below). (Option G [High temperature type]:45°C)
The water-refrigerated thermo-cooler discharges heat by using facility water. Therefore, it is necessary to supply facility water from the source listed in the following table.
Page 26
HRX-OM-M090
Chapter 3 Transport and Setting Up
3.2 Installation HRS Series
3-6
3.2.3 Installation and Maintenance Space
It is recommended to keep the space around the product shown in
Fig. 3-2.
For maintenance, move the thermo-chiller into a space where maintenance work is possible.
Fig. 3-2 Installation space
The temperature of the outlet of for the ventilation of the thermo-chiller and the panel surface may become approx. 50
o
C or higher. When
placing the thermo-chiller, ensure the thermo-chiller does not affect surrounding environment.
Page 27
HRX-OM-M090
Chapter 3 Transport and Setting Up
HRS Series 3.3 Installation
3-7
3.3 Installation
3.3.1 Mounting
z
Mount the product on a flat and stable floor with no vibrations.
z
Refer to’’8.2 Outline dimensions’’ for dimensional information of the
product.
How to mount the product
1. Move the product to the installation area.
2. After moving, lock the front casters again.
Fig. 3-3 Installation procedures
Fixture
Follow the procedure below when fixing the thermo-chiller to the floor or the mounting frame.
1. Prepare the fixing bracket shown below (Not included in the package).
Item
Part number
Anti-seismic brackets HRS-TK001
2. Use M8 foundation bolts to fix the product within the dimensions below.
Fig. 3-4 Anti-seismic brackets installing
4 (four) M8 foundation bolts should be prepared by the customer.
Lock
240 335
555
590
Page 28
HRX-OM-M090
Chapter 3 Transport and Setting Up
3.3 Installation HRS Series
3-8
3.3.2 Electrical wiring
Power supply cable and Earth Leakage Breaker
Prepare the power supply shown in the following table. For the connection between the product and power supply, use the power supply cable and earth leakage breaker shown below. If communication with the user’s machine is necessary, use the following signal cable.
Table 3-3 Power supply cable and Earth Leakage Breaker(Recommended)
Recommended earth
leakage breaker
Model
Power supply
voltage
Cable
qty.
x size
Rated
voltage
[V]
Rated
current
[A]
Sensitivity
of leak
current [mA]
HRS012-∗∗-10 HRS018-∗∗-10
1-phase
100V AC(50Hz)
1-phase
100V to 115V
AC(60Hz)
100 200
Sharing
15 15 or 30
HRS012-∗∗-20 HRS018-∗∗-20
HRS024-∗∗-20
10 30
HRS0∗∗-∗∗-20-∗∗T
(High pump head
[Optional])
1-phase
200-230V AC
(50/60Hz)
3 cores x
14AWG
3 cores x
2.0mm
2
(including
ground)
200,
230
15 30
Only qualified persons are allowed to wire the product. Be sure to shut off the user’s power supply. Wiring with the product
energized is strictly prohibitted.
The wiring must be conducted using cables complying with “Table
3-3” firmly and secured to the product to prevent the external force of cables being applied to the terminals. Incomplete wiring or improper securing of wiring may cause electrical shock, excessive
heat and fire. Ensure a stable power supply with no voltage surges. Ensure that an Earth Leakage Breaker is used in the power supply of
the product. See “Table 3-3”. Use a power supply suitable for the specifications of the product. Be sure to connect the ground connection. Ensure that a lock out facility is availble on the power supply. Each product must have its own separate Earth Leakage Breaker.
Otherwise there can be a risk of electric shock or fire.
Do not modify the intenal electrical wiring of the product. Incorrect
wiring may cause electrical shock or fire. Also, modifing the
internal wiring will void the product’s warranty. Do not connect the ground to water line, gas pipe or lightening
Page 29
HRX-OM-M090
Chapter 3 Transport and Setting Up
HRS Series 3.3 Installation
3-9
3.3.3 Preparation and wiring of power supply cable
Preparation
1. Strip the sheath from both ends of the cable.
2. Disassemble the power supply connector. Attach crimp terminals (mounting screw dia.:
M3.5) to one end of the cable to crimp them to L, N, E inside of the connector, then reassemble the power supply connector.
3. Connect the other end of the cable to a terminal (e.g. crimped terminal) that is
compatible to the secondary side of the earth leakage breaker.
Fig. 3-5 Power supply cable
Cable
Crimped terminal (Prepared by customer)
[Accessory] Power supply
connector
N
E (Earth)
L
L
E
N
Earth leakage breaker (Secondary side)
Crimp terminal 3pcs. Terminal screw is M3.5. Prepare the crimp terminal which is suitable for the cable diameter.
The electrical facilities should be installed and wired in accordance
with local laws and regulations of each country and by a person who has knowledge and experience.
Check the power supply. Operation with voltages, capacities and
frequencies other than the specified values can cause fire and electrical shock.
Wire with an applicable cable size and terminal. Forcibly mounting
with an unsuitable size cable may result in heat generation or fire
Plug (Prepared by customer)
L
E
N
Page 30
HRX-OM-M090
Chapter 3 Transport and Setting Up
3.3 Installation HRS Series
3-10
Wiring
1. Connect the power supply cable to the secondary side of the earth leakage breaker and
grounding.
2. Plug the power supply cable in to the power cable connector on the product.
Fig. 3-6 Wiring of power supply
L
N
E
Earth
Power cable hole
Earth leakage
breaker
E
Grounded socket
Power supply connector
Page 31
HRX-OM-M090
Chapter 3 Transport and Setting Up
HRS Series 3.3 Installation
3-11
3.3.4 Wiring of remote operation signal input
The remote signal input is to enable the product to be run and stopped remotely by applying a contact signal input. This chapter illustrates examples of wiring
Select DIO mode as the communication mode to activate the remote control signal input. After wiring, select DIO mode referring to the Communications Operation manual.
Tips
This product has two input signals. These can be customized depending on the customer’s application. Refer to the Operation Manual for communication for details.
1. For operation using remote signal, prepare the switch (source voltage: 24V, contact
capacity: 35mA or more, min. load current: 5mA), and cable (dia. 0.14 to 1.5mm2).
2. Connect the remote control cable and the switch to the connector for the contact
input/output, supplied as an accessory. Then, mount the ferrite core (3 turn) supplied as an accessory. (See below) (This is an example of wiring. Refer to the Operation Manual for communication for details.)
Fig. 3-7 Wiring of contact for remote signal inputs(Example)
Terminal no.1
Terminal no.12
Switch
Recommended accessory cable dia. is
0.14 to 1.5mm2
9
12 11 10
Terminal no.
〃 〃
The capacity of the output contact of the product is limited. If the capacity is not large enough, install a relay, etc. (to allow for larger capacity). At the same time, ensure the input current of the relay is small enough in relation to the contact capacity of the product.
Be sure to shut off the breaker of the facility power supply (the user’s machine power supply) before wiring.
Page 32
HRX-OM-M090
Chapter 3 Transport and Setting Up
3.3 Installation HRS Series
3-12
3. Plug the prepared connector in to the contact input/output connector that is on the back
of the thermo-chiller.
Fig. 3-8 Connecting the Remote control signal cable
Page 33
HRX-OM-M090
Chapter 3 Transport and Setting Up
HRS Series 3.3 Installation
3-13
3.3.5 Wiring of operation signal output and alarm signal output
The operation signal output and alarm signal output are the outputs generated by a contact signal to shown the status of the product.
The specifications of the contact for each signal output are as follows.
Table 3-4 Signal output contact spec. at the time of shipment
Contact output
Signal explanation
(Default setting)
Operation
At run
Contact closed
At stop
Contact open
Output current 1 Terminal no.5,6
Run status signal A
With power supply
shut off
Contact open
At remote
Contact closed
At non remote
Contact open
Output current 2 Terminal no.3,4
Remote signal A
With power supply
shut off
Contact open
At generation
Contact open
Not generated
Contact closed
Output current 3 Terminal no.1,2
Alarm signal B
With power supply
shut off
Contact open
Tips
This product has three output signals which can be customized depending on the customer’s application
Signals below can be output. Refer to the Communications Operation Manual for more details.
Signal for completion of preparation (TEMP READY)
Signal for operation stop alarm
Signal for continuing operation alarm
Signal for selection alarm
Signal for operation start timer setting
Signal for operation stop timer setting
Signal for power recovery setting
Signal for freezing prevention setting
Contact input signal detection
Be sure to shut off the breaker of the facility power supply (the user’s machine power supply) before wiring.
Page 34
HRX-OM-M090
Chapter 3 Transport and Setting Up
3.3 Installation HRS Series
3-14
3.3.6 RS-485 Communication wiring
Serial communication RS-485, operation start/stop, setting and reading of circulating fluid temperature, and reading of alarm condition can be done by remote control.
Refer to the Communications Operation Manual for more details.
Wiring of interface communication cable
z Connecting to PC
RS-485 cannot be directly connected to a normal PC. Use a RS-232C/RS485 converter which is available on the market.
Be sure to follow the wiring procedure below for connecting multiple thermo-chillers.
z
Configuration of connection One host computer : One thermo-chiller, or one host computer : N thermo-chillers. (Max. 31 thermo-chillers can be connected.)
1
SD+
5
SG
9
SD-
SD+
SD- SG
Terminal
resistance
Master This product
(first slave)
1
SD+
5
SG
9
SD-
This product
second slave
1
SD+
5
SG
9
SD-
This product
31
st
slaves
Terminal
resistance120Ω
Fig. 3-9 Connection of RS-485
Tips
Both ends of the communication connection (the end nodes) need to be connected to the host computer.
The terminal resistance of this product can be set by the operation display panel. Refer to “5.18 Communication function”.
Be sure to shut off the breaker of the facility power supply (the user’s machine power supply) before wiring.
Page 35
HRX-OM-M090
Chapter 3 Transport and Setting Up
HRS Series 3.3 Installation
3-15
3.3.7 RS-232C Communication wiring
Serial communication RS-232C, operation start/stop, setting and reading of circulating fluid temperature, and reading of alarm condition can be done by remote control.
Refer to the Communications Operation Manual for more details.
Wiring of communication cable
Be sure to wire as shown in the figure below.
z
Configuration 1 master : 1 thermo-chiller
2
3
5
RD
SD
SG
2
3
5
RD
SD
SG
Master This product
Fig. 3-10 Connection of RS-232C
Be sure to shut off the breaker of the facility power supply (the user’s machine power supply) before wiring.
Page 36
HRX-OM-M090
Chapter 3 Transport and Setting Up
3.4 Piping HRS Series
3-16
3.4 Piping
Piping port size
Table 3-5 Piping port size
Name Port size*1
Recommended
tightening torque
Recommended proof
pressure for piping
Circulating fluid supply Rc1/2
28 to 30N・m
0.4MPa more
Circulating fluid return Rc1/2
28 to 30N・m
0.4MPa more
Faiclity water inlet*2 Rc3/8
22 to 24N・m
Faiclity water outlet*2 Rc3/8
22 to 24N・m
1.0MPa more
(Facility water pressure 0.3 to
0.5 MPa)
Automatic water-fill port *3
Rc3/8 22 to 24N・m
1.0MPa more
Automatic water-fill pressure
0.2 to 0.5MPa
Overflow port *3
Rc3/4 28 to 30N・m
Inside diameter 19mm more of piping
1 For NPT and G thread, use a conversion connector available as an accessory separately. 2 For water-cooled type. 3 For automatic fluid filling [Option]
Connect piping firmly. Incorrect piping might cause leakage of
supplied or drained leakage and wet surrounding area and facility. Keep supply water pressure below 0.5MPa. Pay attention not to allow dust and foreign materials to enter into
water circuit etc. during connection of piping. When the thermo-chiller is water-cooled refrigerated type, the
temperature of the outlet of the facility water may become approx.
60 depending on operating conditions Hold the piping port firmly with specific wrench when tightening. The piping should be selected with due consideraton of pressure and
temperature. Otherwise, the piping can burst in service.
Check the model number of this product in "1.4.1 Product Label” of
this manual before connecting piping. Model number: HRS0∗∗-N-0
The transition connector from Rc to NPT is enclosed as an
accessory.
For NPT piping, be sure to use this connector. Model number: HRS0∗∗-F-0
The transition connector from Rc to G is enclosed as an accessory.
For G piping, be sure to use this connector.
Page 37
HRX-OM-M090
Chapter 3 Transport and Setting Up
HRS Series 3.4 Piping
3-17
How to connect piping
Tighten the piping to circulating fluid outlet / return port and facility water outlet / inlet.
Fig. 3-11 Tightening of piping
Recommended piping circuit
Fig. 3-12 Recommended piping circuit
No. Name Size
1 Valve Rc1/2 2 Y-shaped strainer or filter Rc1/2 (#1000) Rc1/2 (50μm) 3 Pressure gauge 0 to 50 L/min 4 Pressure gauge 0 to 1.0 MPa 5 Others (Pipe, hose, etc.) I.Dφ15more
Page 38
HRX-OM-M090
Chapter 3 Transport and Setting Up
3.5 Fill of circulating fluid HRS Series
3-18
3.5 Fill of circulating fluid
Turn the tank lid anticlockwise to open. Supply the circulating fluid up to the “H” mark on the fluid level indicator. Use tap water which satisfies the water quality standard shown inTable 7-1, or a 15% aqueous solution of ethylene glycol.
Fig. 3-13
Circulating fluid fill
15% aqueous solution of ethylene glycol
When a 15% aqueous solution of ethylene glycol is used, prepare the ethylene glycol aqueous solution separately. To control the concentration of the ethylene glycol aqueous solution, a concentration meter is available separately from SMC.
Item No Remarks
Ethylene glycol aqueous solution 60% HRZ-BR001
Please dilute to 15% with tap water and use it.
Densitometer HRZ-BR002
Tank lid
Circulating fluid fill
Liquid level
z Check the drain port is plugged or closed by the valve to prevent the
supplied circulating fluid from draining out.
z Supply the circulating fluid up to the “H” mark on the tank.
Operation will stop when the fulid level falls lower than “L”.
When the temperature of the circulating fluid is set to lower than 10oC,
use a 15% aqueous solution of ethylene glycol. Tap water may freeze in the thermo-chiller, leading to malfunction.
Page 39
HRX-OM-M090
Chapter 3 Transport and Setting Up
HRS Series 3.6 Option J Piping of【Automatic fluid filling】
3-19
3.6 Option J Piping ofAutomatic fluid filling
Automatic fluid filling port and over flow port need to be connected when optional [automatic fluid filling] is selected.
Piping to automatic fluid filling port
Connect the piping to supply the circulating fluid. Automatic fluid filling is performed by using the solenoid valve which is built in the thermo-chiller.
Piping to the overflow port
This is necessary when automatic fluid filling function is used. Discharge excess circulating fluid when the fluid level in the tank rises due to the failure of the solenoid valve etc.
Piping name Port size Piping specifications
Automatic water-fill inlet Rc3/8
Supply pressure0.2 to 0.5MPa
Overflow outlet Rc3/4
Piping should be within f19 of I.D. and 5m of length. Riser piping (trapping area) should be avoided.
Fig. 3-14 Piping to automatic fluid filling port and overflow port
Supplying fluid from the automatic fluid filling port
Fluid filling starts if the power supply switch is turned ON when the fluid level of the circulating fluid is “L” of the level indicator or lower (P XX). Fluid supply stops when the fluid level reaches the proper level. If the fluid level does not reach the proper level in an hour after starting supplying fluid, it causes the alarm “AL01; Low level in tank". Fluid supply continues while the alarm is generated. Fluid supply stops after the fluid level reaching proper level. Alarm continues after fluid supply is completed. Release the alarm referring Chapter 6.
Page 40
HRX-OM-M090
Chapter 3 Transport and Setting Up
3.7 Wiring of external switch HRS Series
3-20
3.7 Wiring of external switch
This product can be monitored by sampling the signal of the external switch prepared by the customer.
Table 3-6 Power supply, contact specifications
Name Terminal NO. Specification
12(DC 24V)
Power supply output
11(24V COM)
DC 24V ±10% 0.5A MAX
1
10(Contact input signal 1)
Contact input signal 1
9(Common of contact output
signal 1)
8(Contact input signal 2)
Contact input signal 2
7(Common of contact output
signal 2)
NPN open collector output PNP open collector output
*1:To use the power of the device, the total load current must be 500mA or less.
If the load is 500mA or more, the internal fuse will be cut to protect the product and the alarm [AL21 DC line fuse cut] will be generated. Refer to Chapter 6 for handling of alarms.
One external switch can be connected to contact input signal 1 and one to contact input signal 2. (Two in total) The external switch cannot be connected to the contact input signal 1 depending on the communication mode. Table 3-7 shows the setting.
Table 3-7 Sets external switch
Communication mode 1
Contact input signal 1 Contact input signal 2
Local mode
MODBUS
Simple communication protocol 1
SERIAL mode
Simple communication protocol 2
×
DIO mode
×
*1:Refer to the Communications Operation Manual for more details of each mode.
Local mode: Mode allowing the product to be operated by the operation panel. (Default setting) SERIAL mode: Mode allowing the product to be operated by serial communication. DIO mode: Mode allowing the product to be operated by the contact input/output communication.
Page 41
HRX-OM-M090
Chapter 3 Transport and Setting Up
HRS Series 3.7 Wiring of external switch
3-21
3.7.1 Reading of the External switch
This product can be monitored during operation by reading the signal of the external switch prepared by the customer.
The product stops monitoring when it stops operation. This product generates an alarm and stops operating when a problem is
detected from the external switch. You can set the product to continue operation or not to detect the alarm.
Run status of this product
Run
Stop
External switch
CLOSE
OPEN
Operation start
Signal reading
start
Delay timer of reading
OFF detection timer
Alarm
generation
Operation stops due to the alarm.
Fig 3-15 Timing chart of external switch monitoring
Delay timer of reading
If the signal of the external switch prepared by the customer is not closed instantly when the product is operated, set the delay timer for reading. By setting this timer, the external switch monitoring starts after the time set by the delay time of reading since the operation start.
“0" is the default setting. Set a time which is suitable for your environment.
Example} When using a flow switch
When operation is started, it takes time for the fluid to reach the piping and the flow
switch to detect the flow. Set the time for the flow switch to start.
OFF detection timer
If you do not want the alarm to be generated instantly when the external switch prepared by the customer is in open status, but instead want the alarm to be generated after the switch has been open for a specific time (continuous open status), set the OFF detection timer.
This timer enables the alarm to be generated when the time set for OFF detection time passes after the switch is in OPEN status.
The default setting is 0 sec. Set a time which is suitable for your application.
Contact input
N.O type or N.C. type can be selected for the external switch. Set the signal which is suitable for the external switch prepared by the customer.
Page 42
HRX-OM-M090
Chapter 3 Transport and Setting Up
3.7 Wiring of external switch HRS Series
3-22
Be sure to shut off the breaker of the facility power supply (the user’s machine power supply) before wiring.
3.7.2 Wiring
As an example of connection of an external switch, the connecting method is shown below using the SMC flow switch (NPN, PNP). This chapter illustrates examples of wiring
Table3-8 External switches used in examples
Name Manufacturer Part NO. Out put type
Current
consumption
PF2W7□□-□□-27(-M) NPN open collector output 70mA or less
Flow switch SMC
PF2W7□□-□□-67(-M) PNP open collector output 70mA or less
1. Prepare the external switch
2. Depending on the external switch output type, wire the switch to the communication
connector for contact input / output (which is included as an accessory) as shown below. Mount the ferrite core (3 turns), which is included as an accessory. This is an example of wiringRefer to the Communications Operation Manual for more details.
NPN open collector output
Fig 3-16 Wiring of the external switch (NPN open collector output) (example)
PNP open collector output
Fig 3-17 Wire the communication cable to the connector included as an accessory.
Brown (DC 24V)
Brown (DC 24V)
Blue (DC COM)
Black (Output
1)
Flow switch
Blue (DC COM)
Black (Output
1)
Flow switch
Terminal NO.1
Terminal NO.12
Wire the communication cable to the connector included as an accessory.
Terminal NO.1
Terminal NO.12
Wire the communication cable to the connector included as an accessory.
Page 43
HRX-OM-M090
Chapter 3 Transport and Setting Up
HRS Series 3.7 Wiring of external switch
3-23
3. Plug the prepared connector in to the contact input/output connector that is on the back
of the thermo-chiller.
Fig 3-18 Connecting the connector
3.7.3 Setting items
Table 3-9 shows the setting items of the external switch. For details, refer to
5.18 Communication function.
Table 3-9 Setting list of the external switch
Displa
y
Item
Initial value
(Default setting)
Example
Reference
page
Category
Communication mode LOC LOC
Contact input signal 1 RUN SW_A Contact input signal 1 type ALT ALT Contact input signal 1 delay timer (time delay) of reading
0 0
Contact input signal 1 OFF detection timer
0 2
Contact input signal 2 OFF OFF Contact input signal 2 type ALT ­Contact input signal 2 delay timer (time delay) of reading
0 -
Contact input/output communication
Contact input signal 2 OFF detection timer
0 -
5.18
Communication setting menu
Example: Connect flow switch A to contact input signal 1 in local mode.
Page 44
HRX-OM-M090
Chapter 3 Transport and Setting Up
3.7 Wiring of external switch HRS Series
3-24
Page 45
HRX-OM-M090
Chapter 4 Starting the Product
HRS Series 4.1 Before Starting
4-1
Chapter 4 Starting the Product
4.1 Before Starting
Check the following items before starting the product.
Installation conditions
z
Check the product is installed horizontally.
z
Check that there are no heavy objects on the product, and the external piping is not applying excessive force to the product.
Connection of cables
z
Check the power, ground and communications (optional) cables are
correctly connected.
Circulating fluid
z
Check proper connection of piping at inlet and outlet.
Facility water piping (For water-cooled type)
z
Check proper connection of piping at the facility water inlet and outlet.
z
Check that the facility water source is in operation.
z
Check that the facility water circuit is not shut off by valves
[Tips]
The product is equipped with a water regulator, which does not allow the facility water to flow unless the product is started.
Automatic fluid filling piping (When option J [With automatic fluid filling] is
selected
z
Confirm that the automatic fluid filling port piping is connected properly.
Over flow port piping (When option J [With automatic filling] is selected).
z
This is necessary when automatic fluid filling function is used.
Fluid level indicator (for tank)
z
Ensure that the fluid level is on “H”.
Only people who have sufficient knowledge and experience about the product and its accessories are allowed to start and stop the product.
The facility water should be checked for whether or not it satisfies the water quality standard described in “7.1 Control of Circulating Fluid Qualit
y
” and requirements specified in “8.1 Specifications List”.
Page 46
HRX-OM-M090
Chapter 4 Starting the Product
4.2 Preparation for Start HRS Series
4-2
Press
4.2 Preparation for Start
4.2.1 Power supply
Supply the power and turn on the power supply switch. When the product is switched on, the operation panel displays the following
conditions.
z
The initial screen (HELLO screen) is displayed for 8 seconds on the operation display panel. Then, the display changes to the main screen which displays the circulating fluid outlet temperature.
z
The set value of circulating fluid temperature is displayed as SV on the panel.
z
The present value of circulating fluid temperature is displayed as PV on the
panel.
Fig. 4-1 Power supply
4.2.2 Setting of circulating fluid temperature
Press the [] and [] buttons on the operational panel to change the SV to required value.
When the temperature of the circulating fluid is set by the communication, refer to "Communication” in the supplementary operation manual.
Fig. 4-2 Setting of circulating fluid temperature
Page 47
HRX-OM-M090
Chapter 4 Starting the Product
HRS Series 4.3 Preparation of circulating fluid
4-3
4.3 Preparation of circulating fluid
When the circulating fluid tank is filled the user’s machine and piping remains empty. In that condition, the circulating fluid flows out to the user’s machine and piping and the tank level decreases and may require a refill. In that case, refill the circulating fluid in the following procedure.
1. Press the [PUMP] key on the operation display panel (press the [RUN/STOP] key and
[MENU] key simultaneously). The pump operates independently while the [PUMP] key is pressed. The [RUN] lamp
(green) flashes while the pump is operating independently, and the circulating fluid in the tank is supplied to the customer’s device and piping. This can be done to check for leakage, and to discharge air from the piping. If the fluid level in the tank reaches the lower limit, a buzzer will sound, and alarm no. “AL01 (tank fluid level is low)” is displayed on the digital display PV. The [ALARM] lamp (red) flashes, the [ ] lamp turns on, and the independent operation of the pump is stopped.
Fig. 4-3 Manual operation of the pump
Fig. 4-4 Low tank level alarm
Flash
Press to
g
ether
Flash
ON
If leakage occurs due to faulty piping including an opened fitting of eternal piping,
stop manual operation of the pump and fix the leak.
Page 48
HRX-OM-M090
Chapter 4 Starting the Product
4.3 Preparation of circulating fluid HRS Series
4-4
2. Press the [RESET] key ([] and [] keys simultaneously) to stop the alarm buzzer.
Fig. 4-5 Alarm receipt
3. Open the tank lid and supply the circulating fluid up to the “H” mark on the tank.
Fig. 4-6 Filling of Circulating Fluid
Flash
Press together
Tank lid
Filling of Circulating Fluid
Liquid
ON
Be sure to reset the alarm on the operation display panel of the alarm. Alarm reset is not accepted from any screen except the alarm display menu. Refer to 5.1.1 Key operations.
Supply the circulating fluid up to the “H” mark on the tank. Operation will stop when the fluid level falls lower than “L”.
Page 49
HRX-OM-M090
Chapter 4 Starting the Product
HRS Series 4.3 Preparation of circulating fluid
4-5
4. Press the [RESET] key ([] and [] keys simultaneously) to reset the alarm.
The alarm (tank fluid level is low) is reset, and the [ALARM] lamp and [ ] lamp are turned off. The display returns to the initial screen of the main menu: “Circulating fluid temp. / Circulating fluid set temp.” Press the [PUMP] key (press [RUN/STOP] key and [MENU] key simultaneously) to start independent operation of the pump.
Fig. 4-7 Alarm release
5. Repeat steps 1 to 4 to supply the circulating fluid to the customer’s device and piping.
The tank level must be “H” on the tank liquid level indicator
Press together
OFF
OFF
Be sure to reset the alarm on the operation display panel of the alarm. Alarm reset is not accepted from any screen except the alarm display menu. Refer to 5.1.1 Key operations.
Page 50
HRX-OM-M090
Chapter 4 Starting the Product
4.4 Starting and Stopping HRS Series
4-6
4.4 Starting and Stopping
4.4.1 Starting the product
Before starting, check the items specified in “4.1 Before Starting” If any alarm lamp remains on, refer to Chapter 6 Alarm indication and
trouble shooting”
1. Press the [RUN/STOP] key on the operation panel.
The [RUN] lamp lights up (in green) and the product starts running. The circulating discharge temperature (PV) is controlled to the set temperature (SV).
Fig. 4-8 Starting the product
ON
Press
When an alarm occurs Refer to “Chapter 6 Alarm indication and trouble shooting’’
Allow at least five minutes before restarting the product.
Page 51
HRX-OM-M090
Chapter 4 Starting the Product
HRS Series 4.4 Starting and Stopping
4-7
Except in case of emergency, do not turn off the power supply switch until the product has stopped completely. Doing so could cause failure.
4.4.2 Stopping the product
1. Press the [RUN/STOP] button on the operation panel.
The [RUN] lamp on the operation panel flashes green at 1 second intervals, and
continues operation to prepare to stop. After approx. 15 seconds, the [RUN] lamp goes
off and the product stops.
Fig. 4-9 Stopping the product
2. Turn off the power supply switch.
All LEDs go off.
Flash
OFF
Press
Be sure to shut off the breaker of the facility power supply (the user’s machine power supply) before wiring. Also, drain the circulating fluid, etc. from the product in accordance with and put into storage properly
(
Referring to "7.4.1 Dischargeofthe circulating fluid".)
Page 52
HRX-OM-M090
Chapter 4 Starting the Product
4.5 Check items after starting HRS Series
4-8
If the valve locates in circulating piping, do not fully close the valve (0L/min). The pump may be damaged.
4.5 Check items after starting
Check the following items after starting the product.
z
There is no leakage from piping.
z
There is no drain of circulating fluid from the tank drain port.
z
The circulating fluid pressure is within the specified range.
z
The tank level is within the specified range.
4.6 Adjustment of Circulating Fluid
Flow adjustment
If the flow rate is less than 7L/min it will not be able to achieve the specified cooling capacity. The adjustment of flow rate should be performed using a manual bypass valve and monitoring the pressure or flow rate in the customer’s device, referring to the recommended piping flow shown in Figure 3-12, until they reach the required value.
When an Alarm is seen, press the [STOP] button and then turn off the power supply switch to stop the product, and turn off the breaker of the user’s power supply to isolate the product.
Page 53
HRX-OM-M090
Chapter 5 Display and setting of various functions
HRS Series 4.6 Adjustment of Circulating Fluid
5-1
Chapter 5 Display and setting of
various functions
The product can have the displays and settings shown in table 5-1.
Table 5-1 List of function
NO Function Outline
Reference
page
1 Main display
Displays the current temperature of the circulating fluid, discharge pressure of the circulating fluidhange the circulating fluid temperature.
5.2
2
Alarm display menu
Indicates alarm number when an alarm occurs.
5.3
3
Inspection monitor menu
Product temperature, pressure and accumulated operating time can be checked as daily inspection. Use these for daily inspection.
5.4
4 Key-lock
Keys can be locked so that set values cannot be changed by operator error.
5.5
5
Timer for operation start /stop
Timer is used to set the operation start/stop.
5.6
6
Signal for the completion of preparation
A signal is output when the circulating fluid temperature reaches the set temperature, when using contact input/output and serial communication.
5.7
7 Offset function
Use this function when there is a temperature offset between the discharge temperature of the thermo-chiller and the customer’s device.
5.8
8
Reset after power failure
Start operation automatically after the power supply is turned on.
5.9
9
Anti-freezing function
Circulating fluid is protected from freezing during winter time or night time. Set beforehand if there is a risk of freezing.
5.10
10
Key click sound setting
Operation panel key sound can be set on/off.
5.11
11
Changing temp. unit
Temperature unit can be changed. Centigrade(
o
C) ⇔ Fahrenheit(oF)
5.12
12
Changing pressure unit
Pressure unit can be changed. MPa
PSI
5.13
13
Setting of alarm buzzer
Alarm sound can be set to on/off.
5.14
14
Alarm customizing
Operation during alarm condition and threshold values can be changed depending on the alarm type.
5.15
15 Data reset
Functions can be reset to the default settings (settings when shipped from the factory).
5.16
16
Accumulation time reset
Reset function when the pump, the fan, or the compressor is replaced. Accumulated time is reset.
5.17
17 Communication
This function is used for contact input/output or serial communication.
5.18
18
Option
J
[Automatic water-fill port ]
This function is available for customer who selected option [Automatic fluid filling] (Refer to P2-1 Name and Function of Parts)
5.19
19
Option [Drain pan set]
This function is available for customer who purchased drain pan set (part no.: HRS-WL001).
5.20
20
Option [Electric resistivity sensor set]
This function is available for customer who purchased the electric resistivity sensor set (part no.: HRS-Dl001).
5.21
Read and understand this manual carefully before changing the settings.
Page 54
HRX-OM-M090
Chapter 5 Display and setting of various functions
5.1 Function HRS Series
5-2
5.1 Function
5.1.1 Key operations
Fig. 5.1-1 “Key operation (1/2)” and Fig. 5.1-2 “Key operation (2/2)” show the operation of keys of the
thermo-chiller.
MENU
PV
SV
Circulating fluid outlet temperature Circulating fluid set temperature
Circulating fluid outlet pressure
Main display
Circulating fluid outlet temperature
Circulating fluid return port
Check monitor
menu
Temperature of the compressor inlet.
PV
SV
PV
SV
PV
SV
PV
SV
Circulating fluid outlet pressure
PV
SV
Pressure of higher pressure refrigerant circuit
PV
SV
Pressure of lower pressure refrigerant circuit
Accumulated operation time of the pump
Accumulated operation time of the fun motor
Accumulated operation time of the compressor
Accumulated operation time
PV
SV
PV
SV
PV
SV
PV
SV
Key-lock
Run timer
Setting menu
Stop timer
READY mode
READY band width
READY time
Offset mode
Offset temperature
Recover from power failure
Anti-freezing
PV
SV
PV
SV
PV
SV
PV
SV
PV
SV
PV
SV
PV
SV
PV
SV
PV
SV
PV
SV
Data reset
Pump accumulated operating time reset
Fun motor accumulated operating time reset
Compressor accumulated operating time reset
Key click sound
Temperature unit
Pressure unit
PV
SV
PV
SV
PV
SV
Alarm The latest
Alarm
Alarm display menu
Alarm
The oldest
PV
SV
PV
SV
PV
SV
Note) It changes only when the alarm is generated.
PV
SV
PV
SV
PV
SV
PV
SV
Power supply ON
PV
SV
Press the 2 sec.
PV
SV
Electric resistivity
PV
SV
SEL
Option drain pan set
PV
SV
PV
SV
Electric resistivity
Option electric resistivity sensor set
SEL
Unused
PV
SV
Unused
Unused
PV
SV
PV
SV
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
MENU
MENU
Press the 2 sec.
MENU
Press the 2 sec.
MENUMENU
Unused
PV
SV
SEL
Fig. 5.1-1 Key operation (1/2)
Page 55
HRX-OM-M090
Chapter 5 Display and setting of various functions
HRS Series 5.1 Function
5-3
MENU
Changing of contact input 1 signal detection
Alarm buzzer sound
Changing of low tank level
Changing of circulating fluid discharge temperature rise
Detection temperature for circulating fluid discharge temperature rise
Changing of circulating fluid discharge temperature drop
Detection temperature for circulating fluid discharge temperature drop
Changing of circulating fluid discharge pressure rise
Detection pressure for circulating fluid discharge pressure rise
Changing of communicati on error
The monitoring time of communication error
PV
SV
PV
SV
PV
SV
PV
SV
PV
SV
PV
SV
PV
SV
PV
SV
Changing of circulating fluid discharge pressure drop
Detection pressure for circulating fluid discharge pressure drop
PV
SV
PV
SV
PV
SV
PV
SV
Changing of contact input 2 signal detection
PV
SV
PV
SV
Changing of DI level rise
Upper limit of DI level rise
Changing of DC line fuse cut
Changing of water leakage
PV
SV
PV
SV
PV
SV
Lower limit of DI level drop
Changing of DI level drop
PV
SV
PV
SV
PV
SV
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
Press the 2 sec.
MENU
Alarm setting
menu
MENU
Communicati on mode
Serial protocol
Communication specifications
RS-485 terminal
Slave address MODBUS
Communicati on speed MODBUS
Slave address Simple communicatio n protocol
Communication speed Simple communication protocol
Response delay time Simple communication protocol
Communicati on range Simple communicati on protocol
MENU
BCC Simple communicatio n protocol
Data length Simple communicati on protocol
Parity check Simple communicati on protocol
Stop bit length Simple communicatio n protocol
Digital input signal 1
Digital input signal 1 type
Digital input signal 2
Digital input signal 2 type
Digital output 1 function
Digital output 1 operation
Digital output 1 selected alarm
Digital output 2 function
Digital output 3 selected alarm
Digital output 2 operation
Digital output 2 selected alarm
Digital output 3 function
Digital output 3 operation
PV
SV
PV
SV
PV
SV
PV
SV
PV
SV
PV
SV
PV
SV
PV
SV
PV
SV
PV
SV
PV
SV
PV
SV
PV
SV
PV
SV
PV
SV
PV
SV
PV
SV
PV
SV
PV
SV
PV
SV
PV
SV
PV
SV
PV
SV
PV
SV
PV
SV
PV
SV
SV
Press the 2 sec.
Contact input signal 1 Delay timer (time delay) of reading
Contact input signal 2 OFF Detection timer
PV
SV
PV
SV
Contact input signal 2 Delay timer (time delay) of reading
Contact input signal 2 OFF Detection timer
PV
SV
PV
SV
Communication
setting menu
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
PV
Temperature alarm Monitoring method
PV
SV
PV
SV
Monitoring start timer
PV
SV
Range over Detection timer
SEL
SEL
SEL
SEL
Fig. 5.1-2 Key operation (2/2)
Page 56
HRX-OM-M090
Chapter 5 Display and setting of various functions
5.1 Function HRS Series
5-4
5.1.2 List of parameters
Table 5.1-1“List of parameter (1/3)” and Table 5.1-3“List of parameter (3/3)” show the parameters of the thermo-chiller.
Table 5.1-1 List of parameter (1/3)
Display Item
Initial value
(Default setting)
Reference
page
Category
Circulating fluid temperature (TEMP PV)
Temperature
Circulating fluid set temperature (TEMP SV)
20(68 F)
Circulating fluid outlet pressure Electric resistivity
5.2
Main display
X X
Alarm no.
5.3
Alarm display menu
Circulating fluid outlet temperature Circulating fluid return port temperature Temperature of the compressor inlet.
Circulating fluid outlet pressure Pressure of higher pressure refrigerant
circuit Pressure of lower pressure refrigerant circuit Electric resistivity
Accumulated operation time of the pump Accumulated operation time of the fun
motor Accumulated operation time of the compressor
Accumulated operation time
5.4
Check monitor menu
Key-lock OFF 5.5 Run timer 0.0H Stop timer 0.0H
5.6
READY mode OFF READY band width 0.0℃(0.0 ゚ F) READY time 10 sec
5.7
Offset mode OFF Offset temperature 0.0℃(0.0 ゚ F)
5.8
Recover from power failure OFF 5.9 Anti-freezing OFF 5.10 Key click sound ON 5.11 Temperature unit C 5.12 Pressure unit MPa 5.13 Data reset NO 5.16 Pump accumulated operating time reset
NO Fun motor accumulated operating time reset
NO Compressor accumulated operating
time reset
NO
5.17
OptionDrain pan set
OFF 5.20 OptionElectric resistivity sensor set
OFF 5.21 Unused
- -
Unused
- -
Unused
- -
Unused
- -
Setting menu
Page 57
HRX-OM-M090
Chapter 5 Display and setting of various functions
HRS Series 5.1 Function
5-5
Table 5.1-2 List of parameter (2/3)
Display Item
Initial value
(Default setting)
Reference
page
Category
Alarm buzzer sound ON 5.14 Changing of low tank level A.STP Changing of circulating fluid discharge
temperature rise
A.RUN
Detection temperature for circulating fluid discharge temperature rise
45.0(113.0 F)
Changing of circulating fluid discharge temperature drop
A.RUN
Detection temperature for circulating fluid discharge temperature drop
1.0(33.8 F)
Changing of circulating fluid discharge pressure rise
A.RUN
Detection pressure for circulating fluid discharge pressure rise
0.30MPa(44PSI)*
Changing of circulating fluid discharge pressure drop
A.RUN
Detection pressure for circulating fluid discharge pressure drop
0.05MPa(7PSI)
Changing of communication error OFF The monitoring time of communication
error
30 sec
Unused ­Changing of Contact input signal 2
detection
A.STP
Changing of DC line fuse cut A.STP Changing of water leakage A.STP
Changing of upper limit of electric resistivity
A.RUN
Upper limit of electric resistivity
4.5Mcm
Changing of lower limit of electric resistivity
OFF
Lower limit of electric resistivity
0.2Mcm
Temperature alarm Monitoring method
0
Monitoring start timer
----
Range over Detection timer
5
5.15
Alarm setting menu
Communication mode LOC
Serial protocol MDBS Communication specifications 485 RS-485 terminal OFF
Slave address 1
Mod bus
Communication speed 19.2 Slave address 1 Communication speed 9.6 BCC ON Data length 8BIT Parity check NON Stop bit length 2BIT Response delay time 0
Serial communication
Simple
communication
protocol
Communication range RW
5.18
Communicati on setting menu
Page 58
HRX-OM-M090
Chapter 5 Display and setting of various functions
5.1 Function HRS Series
5-6
Table 5.1-3 List of parameter (3/3)
Display Item
Initial value
(Default setting)
Reference
page
Category
Contact input signal 1 RUN Contact input signal 1 type ALT Unused ­Unused ­Contact input signal 2 OFF Contact input signal 2 type ALT Contact input signal 2 delay timer (time delay) of reading
0
Contact input signal 2 OFF detection timer
0
Contact output 1 function
RUN
Contact output 1 operation
A
Selected for contact output 1
AL.01
Contact output 2 function
RMT
Contact output 2 operation
A
Selected for contact output 2.
AL.01
Contact output 3 function ALM Contact output 3 operation B
Contact input/output communication
Selected for contact output 3 AL.01
5.18
Communicati on setting menu
Option T [High pump head] :0.70MPa102PSI,MT:0.60MPa(87PSI)
Page 59
HRX-OM-M090
Chapter 5 Display and setting of various functions
HRS Series 5.2 Main screen
5-7
5.2 Main screen
5.2.1 Main screen
Displays the current temperature and the set temperature of the circulating fluid. The set temperature can be changed on this screen.
5.2.2 Display on the main screen
The display on the main screen is as follows.
Current discharge temperature of circulating fluid Display
1. Turn ON the power supply switch.
Current temperature and set temperature are displayed on the digital display. Alarm display screen (See 5.3) appears when an alarm is generated.
PV
SV
Circulating fluid temperature Set
2. Change the set temperature by pressing the [▼][] key.
After changing the set temperature, set it by pressing the [SEL] key.
The set value flashes while it is being changed. If [SEL] key is not pressed, the value is reset after 3 sec.
Circulating fluid discharge pressure Diaplay
3. Press the [SEL] key.
The discharge pressure of the circulating fluid is displayed on the digital display.
PV
SV
Electric resistivity Display
4. Press the [SEL] key.
The electric resistivity is displayed on the digital display.
PV
SV
Electric resistivity is displayed as [- - - -] if electric resistivity sensor set [Part
NO.:HRS-DI001] is not set.
The unit of electric resistivity is [MΩ/cm] Electric resistivity is displayed within the range of 0.0 to 4.5MΩ/cm] electric
resistivity is 4.5MΩ/cm or more when flashing at 4.5.
Current discharge temperature
Set Temperature
Circulating fluid discharge pressure
Page 60
HRX-OM-M090
Chapter 5 Display and setting of various functions
5.3 Alarm display menu HRS Series
5-8
5.3 Alarm display menu
5.3.1 Alarm display menu
The alarm display screen appears when an alarm is generated.
The alarm display menu cannot be accessed when no alarm has been
generated.
Refer to “Chapter 6 Alarm indication and trouble shooting” for the content
of alarms.
5.3.2 Content of display of alarm display menu
The alarm display screen appears when an alarm is gernerated. When multiple alarms are generated, the latest alarm is displayed on the
screen. Each time the [SEL] key is pressed, the alarms are displayed in order,
starting from the latest one.
PV
SV
PV
SV
The main screen is displayed when the alarm is reset.
PV
SV
PV
SV
The main screen is displayed when [MENU] key is pressed while an alarm is output.
PV
SV
PV
SV
The alarm display screen is displayed if [MENU] key is pressed again.
A
larm code
Alarm history
SEL
MENU
Alarm
release
Page 61
HRX-OM-M090
Chapter 5 Display and setting of various functions
HRS Series 5.4 Inspection monitor menu
5-9
5.4 Inspection monitor menu
5.4.1 Inspection monitor menu
As a part of the daily inspection, the temperature, pressure and accumulated operating time can be checked. Please use this for confirmation of your daily inspection.
5.4.2 Checking of the Inspection monitor menu
The table below explains the check items of the inspection monitor menu.
Table 5.4-1 List of check items of Inspection monitor menu
Display Item Contents
Circulating fluid outlet temperature
Displays the temperature of the circulating fluid outlet. This temperature does not take offset into
consideration. Circulating fluid return port temperature
Displays the temperature of the circulating fluid return. Temperature of the inlet of the
compressor
Displays the temperature of the inlet of the compressor.
Circulating fluid outlet pressure
Displays the circulating fluid outlet pressure at the
outlet. Pressure of higher pressure refrigerant circuit
Displays the pressure of higher pressure side of the
refrigerant circuit. Pressure of lower pressure refrigerant circuit
Displays the pressure of lower pressure side of the
refrigerant circuit. Electric resistivity
Displays the electric resistivity. Accumulated operation time of the
pump
Displays the accumulated operation time of the pump. Accumulated operation time of the
fun motor
Displays the accumulated operation time of the fan
motor. (For air-cooled type) Accumulated operation time of the compressor
Displays the accumulated operation time of the
compressor. Accumulated operation time Displays the accumulated operation time.
Page 62
HRX-OM-M090
Chapter 5 Display and setting of various functions
5.4 Inspection monitor menu HRS Series
5-10
Check of the circulating fluid outlet temperature
1. Press and hold the [MENU] key for approx. 2 sec.
The temperature of the circulating fluid outlet
is displayed on the digital display.
PV
SV
PV
SV
Displays the temperature of the circulating fluid at the outlet from which the fluid is fed to the customer’s device. This temperature does not take temperature offset into consideration
Check of the circulating fluid inlet temperature
2. Press the [SEL] key once.
The temperature of the circulating fluid return is displayed on the digital display.
PV
SV
Displays the temperature of the circulating fluid returning from the customer’s device.
Check of the temperature of the inlet of the compressor.
3. Press the [SEL] key once.
The temperature of the refrigerant circuit compressor inlet is displayed on the digital display.
PV
SV
Displays the temperature of the compressor inlet.
Check of the circulating fluid outlet pressure
4. Press the [SEL] key once.
The circulating fluid outlet pressure is displayed on the digital display.
PV
SV
The pressure of the circulating fluid outlet from which the fluid is sent to the customer’s device is displayed.
Check of the pressure of the higher pressure side of the refrigerant circuit
5. Press the [SEL] key once.
The pressure of higher pressure refrigerant circuit is displayed on the digital display.
PV
SV
Displays the pressure of the higher pressure side of the refrigerant circuit.
MENU
Press and
Page 63
HRX-OM-M090
Chapter 5 Display and setting of various functions
HRS Series 5.4 Inspection monitor menu
5-11
Check of the pressure of the lower pressure side of the refrigerant circuit
6. Press the [SEL] key once.
The pressure of lower pressure refrigerant circuit is displayed on the digital display.
PV
SV
Displays the pressure of the lower pressure side of the refrigerant circuit.
Check of the electric resistivity
7. Press the [SEL] key once.
The electric resistivity is displayed on the digital display.
PV
SV
This function is available for customers who have purchased the electric resistivity
sensor set (part no.: HRS-Dl001).
Check of the accumulated operation time of the pump
8. Press the [SEL] key once.
The accumulated operation time of the pump is displayed on the digital display.
PV
SV
Displays the accumulated operation time of the pump. Refer to the table below for the display.
Table 5.4-2 List of time display
Cumulative time Indicated value
0h to 999h
to
1,000h to 99,999h
to
100,000h
Return to
AL28 Pump maintenance alarm is generated when the accumulated operation time of the pump reaches 20,000 hours (
) or more. For details, refer to Chapter 6 Alarm
indication and trouble shooting
.
Page 64
HRX-OM-M090
Chapter 5 Display and setting of various functions
5.4 Inspection monitor menu HRS Series
5-12
Check of the accumulated operation time of the fan motor
9. Press the [SEL] key once.
The accumulated operation time of the fun motor is displayed on the digital display.
PV
SV
Displays the accumulated operation time of the fan motor. Refer to Table 5.4-2 for the display.
AL29 Fan motor maintenance alarm is generated when the accumulated operation time of the fan motor reaches 20,000 hours (
) or more. For details, refer to
Chapter 6 Alarm indication and trouble shooting. Water-cooled refrigerated type does not have the fan motor. The accumulated time of
the digital display shows “
“.
Also, AL29 fan motor maintenance alarm is not generated.
Check of the accumulated operation time of the compressor
10. Press the [SEL] key once.
The accumulated operation time of the compressor is displayed on the digital display.
PV
SV
Displays the accumulated operation time of the compressor. Refer to Table 5.4-2 for the display.
AL30 Compressor maintenance alarm is generated when the accumulated operation time of the compressor reaches 50,000 hours (
) or more. For details, refer to
Chapter 6 Alarm indication and trouble shooting.
Check of the accumulated operation time
11. Press the [SEL] key once.
The accumulated operation time is displayed on the digital display.
PV
SV
Displays the accumulated operation time. Refer to Table 5.4-2 for the display.
Page 65
HRX-OM-M090
Chapter 5 Display and setting of various functions
HRS Series 5.5 Key-lock
5-13
5.5 Key-lock
5.5.1 Key-lock
The keys can be locked to prevent the set values being changed by operator error. Operation can be started/stopped by the “RUN/STOP” key even while key-lock is set.
If you try to change the set value with “” and ”” key while key-lock is set, “
” is displayed for 1 sec. on the screen. The set value cannot be
changed. (Refer to the figure below.)
PV
SV
PV
SV
During key-lock setting, no other setting is available. Release the key-lock setting for other setting.
Page 66
HRX-OM-M090
Chapter 5 Display and setting of various functions
5.5 Key-lock HRS Series
5-14
5.5.2 Key-lock setting / checking
The table below explains the setting items of the key-lock function and the initial values.
Table 5.5-1 List of key-lock
Display Item Contents
Initial value
(Default setting)
Key-lock
Sets key-lock. While key-lock setting is ON, no other setting is available.
OFF
1. Press and hold the [MENU] key for approx. 2 sec.
Repeat pressing until the setting display [
] for key-lock setting appears on the
digital display
PV
SV
PV
SV
PV
SV
Key-lock setting and checking
2. Select “ON” from the table below with [] key or [] key, and confirm with “SEL”.
Table 5.5-2 List of set value
Set value Explanation
Initial value
(Default setting)
Key-lock function OFF
Key-lock function ON
3. Press the [MENU] key once.
Return to the main screen (screen displaying the circulating fluid temperature).
PV
SV
MENU
Press and
hold
MENU
Press and
hold
Page 67
HRX-OM-M090
Chapter 5 Display and setting of various functions
HRS Series 5.6 Run timer, stop timer function
5-15
5.6 Run timer, stop timer function
5.6.1 Run timer and stop timer function
This function starts or stops operation automatically when a set time has passed. The time can be set according to the customer’s working hours. Set the circulating fluid temperature in advance.
[Run timer] is a function to start operation after a set time. Stop timer] is a function to stop operation after a set time. It is possible to set both [Run timer] and [Stop timer]. The set time of both the [Run timer] and [Stop timer] can be up to 99.5 hours, in 0.5 hour units.
When communication is used
If the communication mode is DIO REMOTE, SERIAL mode, this function does not operate. DIO REMOTE, SERIAL mode operation/stop signal has priority.
Run timer
Run timer] starts operation after the set time.
If the thermo-chiller is already operating or the pump is operating independently, this function does not operate even if the set time has passed.
Operation can start when the condition is normal and there is no alarm generated.
The [
] lamp lights up when the run timer is set. The [ ] lamp is turned
off when the operation is started by the run timer.
The [
] lamp is not turned off if the Stop timer is set.
The Run timer setting is reset when the main power supply is cut or a
power failure occurs. Please reset.
Stop timer
The [
] lamp lights when the Stop timer is set. The [ ] lamp is turned off
if the operation is stopped by the Stop timer.
The [
] lamp is not turned off if the Run timer is set.
The Stop timer setting is reset when the main power supply is cut or a
power failure occurs. Please reset.
Page 68
HRX-OM-M090
Chapter 5 Display and setting of various functions
5.6 Run timer, stop timer function HRS Series
5-16
Timer setting example
Now 1H 2H 3H 4H 5H 6H 7H 8H
Stopped Operating
・・
Run timer Set to start after 3hours
Operating
Now 1H 2H 3H 4H 5H 6H 7H 8H
Operating Stopped
・・
Stop timer Set to stop after 3 hours
Stopped
Now 1H 2H 3H 4H 5H 6H 7H 8H
Stopped Operating Stopped
・・
Run timer
Stop timer Set to start after 2 hours Set to stop after 5.5 hours
Operating Stopped
Now 1H 2H 3H 4H 5H 6H 7H 8H
Operating Stopped Operating
・・
Run timer
Stop timer
Set to stop after 2 hours Set to start after 5.5hours
Stopped Operating
z Set while the power supply switch is ON (while the power is
supplied).
z The setting is released when operation is started or stopped by the
timer. Resetting is necessary to use the timer next time.
z Run timer setting is released when the power supply switch or the
customer’s power supply facility is cut, or a power failure occurs. Please reset.
Page 69
HRX-OM-M090
Chapter 5 Display and setting of various functions
HRS Series 5.6 Run timer, stop timer function
5-17
5.6.2 Setting and checking of Run timer and stop timer function
The table below explains the setting items of the run/stop timer and the initial values.
Table 5.6-1 List of set run timer and stop timer
Display Item Contents
Initial value
(Default setting)
Run timer Sets run timer. 0.0H Stop timer Sets stop timer. 0.0H
This part explains the setting and checking items of both the run and stop timers. Please refer to the applicable items for the timer you wish to use.
1. Press and hold the [MENU] key for approx. 2 sec.
Repeat pressing the key until the setting screen for key-lock [
] appears on the
digital display.
PV
SV
PV
SV
PV
SV
Run timer Setting and checking
2. Press the [SEL] key once.
The set screen of run timer is displayed on the digital display.
PV
SV
3. Select run timer from the table below with [] key or [] key, and confirm by pressing
“SEL”.
Table 5.6-2 List of set value
Set value Explanation
Initial value
(Default setting)
Timer OFF
to
Operation starts after set time. Setting unit is 0.5 hour
E.g. Set at 5:30pm the previous day. Operation starts 14 hours later (7:30am the next day).
PV
SV
MENU
Press and
hold
MENU
Press and
hold
Page 70
HRX-OM-M090
Chapter 5 Display and setting of various functions
5.6 Run timer, stop timer function HRS Series
5-18
Stop timer Setting and checking
4. Press the [SEL] key once.
The set screen of stop timer is displayed on the digital display.
PV
SV
5. Select stop timer from the table below with [] key or [] key, and confirm by pressing
“SEL”.
Table 5.6-3 List of set value
Set value Explanation
Initial value
(Default setting)
Timer OFF
to
Operation stops after set time. Setting unit is 0.5 hour
E.g. Set at 4:30pm. Operation stops 1 hour and 30 minutes later (at 6:00pm).
PV
SV
6. Press the [MENU] key once.
Return to the screen displaying the circulating fluid temperature.
PV
SV
7. Once the run timer is set leave the power supply to the product on. The product will start
automatically in the set time.
When the stop timer is set, leave the product running. The product will stop automatically in the set time.
Page 71
HRX-OM-M090
Chapter 5 Display and setting of various functions
HRS Series 5.7 Signal for completion of preparation (TEMP READY)
5-19
5.7 Signal for completion of preparation (TEMP READY)
5.7.1 Signal for completion of preparation (TEMP READY)
This function sets the band width for the circulating fluid set temperature (upper/lower temperature range) in order to notify the customer by communication that the circulating fluid temperature has reached the band range (upper/lower temperature range).The default setting of this function is “OFF”.
Tips
This function is available when contact input/output and serial communication is used. Refer to the Communications Operation Manual for more details.
An example is shown below.
Circulating fluid set temperature
20
READY band width (upper/lower temperature range)
±2
READY time
60 sec.
Preparation is completed 60 seconds after the circulating fluid temperature reaches 18
o
C to 22oC.
Circulating set fluid 20
READY TEMP 22
READY TEMP 18
READY band width upper/lower temperature range 2
READY band width upper/lower temperature range 2
Circulating set fluid
READY time
60sec
Completion
of preparation
READY time
lower than 60sec
Contact outputState generation output SerialON output
Contact outputState not generation out put Serial OFF output
Completion of preparation
TEMP READY
Page 72
HRX-OM-M090
Chapter 5 Display and setting of various functions
5.7 Signal for completion of preparation (TEMP READY) HRS Series
5-20
5.7.2 Signal for completion of preparation (TEMP READY) setting / checking
The table below shows the explanation of and initial value of set items of signal for completion of preparation (TEMP. READY)
Table 5.7-1 List of set Signal for completion of preparation (TEMP READY)
Display Item Contents
Initial value
(Default setting)
READY mode
Sets signal for completion of preparation (TEMP READY)
OFF
READY band width (upper/lower temperature range
Sets temperature of signal for completion of preparation.
0.0
READY time
Sets time of signal for completion of preparation.
10 sec
1. Press and hold the [MENU] key for approx. 2 sec.
Repeat pressing the key until the setting screen for key-lock [
] appears on the
digital display.
PV
SV
PV
SV
PV
SV
READY mode Setting and checking
2. Press the [SEL] key 3 times.
The set screen of ready mode is displayed on the digital display.
PV
SV
3. Select ON from the table below with [] key or [] key, and confirm by pressing
“SEL”.
Table 5.7-2 List of set value
Set value Explanation
Initial
(Default setting)
Signal for completion of preparation (TEMP READY) function OFF
Signal for completion of preparation (TEMP READY) function ON
READY band width setting and checking
4. Press the [SEL] key once.
The set screen of READY band width (upper/lower temperature range) is displayed on the digital display.
PV
SV
MENU
Press and
hold
MENU
Press and
hold
Page 73
HRX-OM-M090
Chapter 5 Display and setting of various functions
HRS Series 5.7 Signal for completion of preparation (TEMP READY)
5-21
5. Select READY band width (upper/lower temperature range) from the table below with
[] key or [] key, and confirm by pressing “SEL”.
Table 5.7-3 List of set value
Set value Explanation
Initial
(Default setting)
Setting and checking are not available when READY mode setting is OFF.
Centigrade
to
Fahrenheit
to
Set READY band width (upper/lower temperature range) for the circulating fluid temperature
Temperature unit is Centigrade : Setting unit is
0.1
o
C
Temperature unit is Fahrenheit : Setting unit is
0.1
o
F
READY time Set and checking
6. Press the [SEL] key once.
The set screen of READY time is displayed on the digital display.
PV
SV
7. Select READY time from the table below with [] key or [] key, and confirm by
pressing “SEL”.
Table 5.7-4 List of set value
Set value Explanation
Initial value
(Default setting)
Setting and checking are not available when READY mode setting is OFF.
to
Sets ultimate time. Set unit is 1 sec.
8. Press the [MENU] key once.
Return to the main screen (screen displaying the circulating fluid temperature).
PV
SV
Page 74
HRX-OM-M090
Chapter 5 Display and setting of various functions
5.8 Offset function HRS Series
5-22
5.8 Offset function
5.8.1 Offset function
This function controls the circulating fluid discharge temperature with offset. Temperature offset may occur between the thermo-chiller and the
customer’s device depending on the installation environment. To correct the temperature offset, three types of offset functions are available (MODE1 to
3). The default setting of this function is “OFF”.
When communication is used
The circulating fluid temperature sent by serial communication is the circulating fluid temperature which is displayed on the thermo-chiller (the circulating fluid temperature after offset).
Example of temperature offset The discharge temperature of the circulating fluid is 30
o
C, but the fluid
temperature in the customer’s device is 29
o
C because of heat radiation
during sending of the fluid.
Circulating fluid display
temperature
30
Circulating fluid set
temperature
30
This produuct Your system
Serial communication
Discharge
temperature 30
Heat
radiation
1
29℃
Circulating fluid temperature 30
MODE Explanation
MODE1
Control the temperature so that the discharge temperature of the circulating fluid is circulating fluid set temperature + offset temperature. Circulating fluid temperature indicates the circulating fluid discharge temperature.
MODE2
Control the temperature so that the discharge temperature of the circulating fluid is circulating fluid set temperature. Circulating fluid temperature indicates the circulating fluid discharge temperature + offset temperature.
MODE3
Control the temperature so that the discharge temperature of the circulating fluid is circulating fluid set temperature + offset temperature. Circulating fluid temperature indicates the circulating fluid discharge temperature - offset temperature.
OFF
Control the temperature so that the discharge temperature of the circulating fluid is circulating fluid temperature set value.
Page 75
HRX-OM-M090
Chapter 5 Display and setting of various functions
HRS Series 5.8 Offset function
5-23
Example of MODE 1 When the offset temperature is 1
o
C, the thermo-chiller controls the temperature aiming at
31
o
C (circulating fluid set temperature + offset temperature.) Even if the discharge
temperature is 31
o
C, the circulating fluid temperature is 30oC at the customer’s device
because of 1
o
C of heat radiation during sending of the fluid. Recircualting fluid display
temperature and communication data is 31
o
C.
Circulating fluid display
temperature
31
Circulating fluid set
temperature
30
Offset temperature
1
This produuct Your system
Serial communication
Discharge
temperature 31
30℃
Circulating fluid temperature 31
Heat
radiation
1
Example of MODE 2 When the offset temperature is -1
o
C, circulating fluid display temperature and the
communication data is 29
o
C (circulating fluid discharge temperature + offset temperature),
and matches the circulating fluid temperature of the customer's device.
Circulating fluid display
temperature
29
Circulating fluid set
temperature
30
Offset temperature
-1
This produuct Your system
Serial communication
Discharge
temperature 30
29℃
Circulating fluid temperature 29
Heat
radiation
1
Example of MODE 3 When the offset temperature is 1
o
C, the thermo-chiller controls the temperature aiming at
31
o
C (circulating fluid set temperature + offset temperature) Even if the discharge
temperature is 31
o
C, the circulating fluid temperature is 30oC at the customer’s device
because of 1
o
C of heat radiation during sending of the fluid. The circulating fluid display
temperature and the communication data is 30
o
C (circulating fluid discharge temp. - offset
temp.), and matches the circulating fluid temperature of the customer's device.
Circulating fluid display
temperature
30
Circulating fluid set
temperature
30
Offset temperature
1
This produuct Your system
Serial communication
Discharge
temperature 31
30℃
Circulating fluid temperature 30
Heat
radiation
1
Page 76
HRX-OM-M090
Chapter 5 Display and setting of various functions
5.8 Offset function HRS Series
5-24
5.8.2 Offset function setting and checking
The table below explains the setting items of the offset function and the initial values.
Table 5.8-1 List of set offset function
Display Item Contents
Initial
(Default setting)
Offset mode Sets offset mode. OFF Offset temperature Sets offset temperature.
0.0
1. Press and hold the [MENU] key for approx. 2 sec.
Repeat pressing the key until the setting screen for key-lock [
] appears on the
digital display.
PV
SV
PV
SV
PV
SV
Offset mode Setting and checking
2. Press the [SEL] key 6 times.
The set screen of offset mode is displayed on the digital display.
PV
SV
3. Select offset mode from the table below with [] key or [] key, and confirm by pressing
“SEL”.
Table 5.8-2 List of set value
Set value Explanation
Initial value
(Default setting)
Offset function OFF
Offset mode 1 Offset mede 2 Offset mode 3
MENU
Press and
hold
MENU
Press and
hold
Page 77
HRX-OM-M090
Chapter 5 Display and setting of various functions
HRS Series 5.8 Offset function
5-25
Offset temperature Setting and checking
4. Press the [SEL] key once.
The set screen of offset temperature is displayed on the digital display.
PV
SV
5. Select offset temperature from the table below with [] key or [] key, and confirm by
pressing “SEL”.
Table 5.8-3 List of set value
Set value Explanation
Initial
(Default setting)
Setting and checking are not available when offset mode setting is OFF.
Centigrade
to
Fahrenheit
to
Sets offset temperature.
Temperature unit is Centigrade : Setting unit is
0.1
o
C
Temperature unit is Fahrenheit : Setting unit is
0.1
o
F
6. Press the [MENU] key once.
Return to the main screen (screen displaying the circulating fluid temperature).
PV
SV
z This function controls the offset temperature to the circulating fluid
discharge temperature
z Control range of the circulating fluid temperature is 5.0 to 40.0
(41.0 F to 104.0 F).
z When the circulating fluid temperature is set to 5.0
o
C (41oF) and the
offset temperature to -20.0
o
C (-36.0oF), the offset temperature is
automatically adjusted to 0.0oC (0.0
o
F) depending on the offset
mode.
Page 78
HRX-OM-M090
Chapter 5 Display and setting of various functions
5.9 Function to recover from power failure HRS Series
5-26
5.9 Function to recover from power failure
5.9.1 Function to recover from power failure
When the power supply is cut due to power failure etc., this function restarts the operation when the power supply recovers, retaining the conditions before the power cut.
When communication is used
If the communication mode is DIO REMOTE, SERIAL mode (MODBUS), this function does not start. The signal of start/stop of DIO REMOTE SERIAL mode (MODBUS) has priority.
The [
] lamp lights when the power failure recovery is set. The default
setting of this function is “OFF”.
Page 79
HRX-OM-M090
Chapter 5 Display and setting of various functions
HRS Series 5.9 Function to recover from power failure
5-27
5.9.2 Function to recover from power failure setting and checking
The table below explains the setting items of the power recovery function and the initial values.
Table 5.9-1 List of set function to recover from power failure
Display Item Contents
Initial value
(Default setting)
Recover from power failure
Sets recover from power failure.
OFF
1. Press and hold the [MENU] key for approx. 2 sec.
Repeat pressing the key until the setting screen for key-lock [
] appears on the
digital display.
PV
SV
PV
SV
PV
SV
Recover from power failure Setting and checking
2. Press the [SEL] key 8 times.
The set screen of recover from power failure is displayed on the digital display.
PV
SV
3. Select recover from power failure from the table below with [] key or [] key, and
confirm by pressing “SEL”.
Table 5.9-2 List of set value
Set value Explanation
Intial value
(Default setting)
Function to recover from power failure OFF
Function to recover from power failure ON
4. Press the [MENU] key once.
Return to the main screen (screen displaying the circulating fluid temperature).
PV
SV
MENU
Press and
hold
MENU
Press and
hold
Page 80
HRX-OM-M090
Chapter 5 Display and setting of various functions
5.10 Anti-freezing function HRS Series
5-28
5.10 Anti-freezing function
5.10.1 Anti-freezing function
The product can prevent the circulating fluid from freezing during winter time. If there is a possibility of the circulating fluid freezing due to changes in the installation and operating environment (operating period and weather), set the protection in advance.
z
If the circulating fluid temperature falls below 3, the pump will start operating automatically.
z
The heat generated by the pump operation will warm up the circulating fluid. When the circulating fluid temperature reaches 5 or higher, the pump will stop operating automatically.
z
As a result, the circulating fluid maintains a temperature of between 3 and 5, preventing freezing.
If the anti-freezing function is set, [RUN] lamp flashes for 2 sec. during waiting (pump does not operate). The [RUN] lamp flashes at 0.3 sec. intervals during the automatic operation of the pump. The default setting of this function is “OFF”.
z This function starts in stand-by condition (power supply switch is
ON).
z Fully open the valve or manual bypass valve arranged by the
customer so that the circulating fluid can circulate when the pump starts automatic operation.
z In extremely cold weather conditions, the heat generated by the
pump as described above may not be enough to prevent freezing.
z During the automatic operation, the pump does not stop even if you
press "RUN/STOP" key.
z In an emergency, stop operation by shutting off the power supply.
Page 81
HRX-OM-M090
Chapter 5 Display and setting of various functions
HRS Series 5.10 Anti-freezing function
5-29
5.10.2 Anti-freezing function setting and checking
The table below explains the setting items of the anti-freezing function and the initial values.
Table 5.10-1 List of set anti-freezing function
Display Item Contents
Initial value
(Default setting)
Anti-freezing Sets anti-freezing OFF
1. Press and hold the [MENU] key for approx. 2 sec.
Repeat pressing the key until the setting screen for key-lock [
] appears on the
digital display.
PV
SV
PV
SV
PV
SV
Anti-freezing Setting and checking
2. Press the [SEL] key 9 times.
The set screen of anti-freezing is displayed on the digital display.
PV
SV
3. Select anti-freezing from the table below with [] key or [] key, and confirm by
pressing “SEL”.
Table 5.10-2 List of set value
Set value Explanation
Initial value
(Default setting)
Anti-freezing function OFF
Anti-freezing function ON
4. Press the [MENU] key once.
Return to the main screen (screen displaying the circulating fluid temperature).
PV
SV
MENU
Press and
hold
MENU
Press and
hold
Page 82
HRX-OM-M090
Chapter 5 Display and setting of various functions
5.11 Key click sound setting HRS Series
5-30
5.11 Key click sound setting
5.11.1 Key click sound setting
Set whether or not a click sound is made when keys on the operation panel are pressed.
The default setting is key sound “on”.
5.11.2 Key click sound setting and checking
The table below explains the setting items of the key click sound and the initial values.
Table 5.11-1 List of set key click sound
Display Item Contents
Initial value
(Default setting)
Key click sound Sets key click sound. ON
1. Press and hold the [MENU] key for approx. 2 sec.
Repeat pressing the key until the setting screen for key-lock [
] appears on the
digital display.
PV
SV
PV
SV
PV
SV
Key click sound Setting and checking
2. Press the [SEL] key 10 times.
The set screen of key click sound is displayed on the digital display.
PV
SV
3. Select key click sound from the table below with [] key or [] key, and confirm by
pressing “SEL”.
Table 5.11-2 List of set value
Set value
Explanation
Initial value
(Default setting)
No key click sound Key click sounds
4. Press the [MENU] key once.
Return to the main screen (screen displaying the circulating fluid temperature).
PV
SV
[MENU]
Press and
hold
[MENU]
Press and
hold
Page 83
HRX-OM-M090
Chapter 5 Display and setting of various functions
HRS Series 5.12 Temperature unit switching
5-31
5.12 Temperature unit switching
5.12.1 Temperature unit switching
The temperature unit of the thermo-chiller can be set to centigrade (oC) or Fahrenheit (
o
F). This setting determines the temperature unit which is
displayed/output. The default setting is centigrade (
o
C).
5.12.2 Temperature unit switching setting and checking
The table below explains the setting items of the temperature unit switching and the initial values.
Table 5.12-1 List of set temperature unit switching
Display Item Contents
Initial value
(Default setting)
Temperature unit Sets temperature unit.
o
C
1. Press and hold the [MENU] key for approx. 2 sec.
Repeat pressing the key until the setting screen for key-lock [
] appears on the
digital display.
PV
SV
PV
SV
PV
SV
Temperature unit Setting and checking
2. Press the [SEL] key 11 times.
The set screen of temperature unit is displayed on the digital display.
PV
SV
3. Select temperature unit from the table below with [] key or [] key, and confirm by
pressing “SEL”.
Table 5.12-2 List of set value
Set value Explanation
Initial value
(Default setting)
Sets temperature unit is centigrade (oC).
Sets temperature unit is fahrenheit (
o
F)
4. Press the [MENU] key once.
Return to the main screen (screen displaying the circulating fluid temperature).
PV
SV
[MENU]
Press and
hold
[MENU]
Press and
hold
Page 84
HRX-OM-M090
Chapter 5 Display and setting of various functions
5.13 Pressure unit switching HRS Series
5-32
5.13 Pressure unit switching
5.13.1 Pressure unit switching
The pressure unit of the thermo-chiller can be set to MPa or PSI. This setting determines the pressure unit which is displayed/output. The default setting is MPa.
5.13.2 Pressure unit switching setting and checking
The table below explains the setting items of the pressure unit switching and the initial values.
Table 5.13-1 List of set Pressure unit switching
Display Item Contents
Initial value
(Default setting)
Pressure unit Sets pressure unit. MPa
1. Press and hold the [MENU] key for approx. 2 sec.
Repeat pressing the key until the setting screen for key-lock [
] appears on the
digital display.
PV
SV
PV
SV
PV
SV
Pressure unit Setting and checking
2. Press the [SEL] key 12 times.
The set screen of pressure unit is displayed on the digital display.
PV
SV
3. Select pressure unit from the table below with [] key or [] key, and confirm by
pressing “SEL”.
Table 5.13-2 List of set value
Set value Explanation
Initial value
(Default setting)
Sets pressure unit is MPa.
Sets pressure unit is PSI.
4. Press the [MENU] key once.
Return to the main screen (screen displaying the circulating fluid temperature).
PV
SV
[MENU]
Press and
hold
[MENU]
Press and
hold
Page 85
HRX-OM-M090
Chapter 5 Display and setting of various functions
HRS Series 5.14 Alarm buzzer sound setting
5-33
5.14 Alarm buzzer sound setting
5.14.1 Alarm buzzer sound setting
This sets whether a warning sound is made or not when alarm signal is output.
The default setting is buzzer sound ON.
5.14.2 Alarm buzzer sound setting and checking
The table below explains the setting items of the alarm buzzer sound and the initial values.
Table 5.14-1 List of set alarm buzzer sound
Display Item Contents
Initial value
(Default setting)
Alarm buzzer sound
Sets
alarm buzzer sound.
ON
1. Press and hold the [MENU] key for approx. 2 sec.
Repeat pressing the key until the setting screen for alarm buzzer sound [
]
appears on the digital display.
PV
SV
PV
SV
PV
SV
PV
SV
Alarm buzzer sound Setting and checking
2. Select alarm buzzer sound from the table below with [] key or [] key, and confirm by
pressing “SEL”.
Table 5.14-2 List of set value
Set value Explanation
Initial value
(Default setting)
No alarm buzzer sound Alarm buzzer sound
3. Press the [MENU] key once.
Return to the main screen (screen displaying the circulating fluid temperature).
PV
SV
[MENU]
Press
and hold
[MENU]
Press
and hold
[MENU]
Press
and hold
Page 86
HRX-OM-M090
Chapter 5 Display and setting of various functions
5.15 Alarm customize function HRS Series
5-34
5.15 Alarm customize function
5.15.1 Alarm customize function
The operation and the threshold when alarm signal is output can be customized. Customers should set them depending on their applications.
The alarms below can be customized.
AL01 Low level in tank(Refer to Table 5.15-3) Operation: When this alarm signal is output, user can choose to
stop/continue the operation. "Operation stop" is default setting.
AL03 Circulating fluid discharge temp. rise (Refer to Table 5.15-4 and Table 5.15-5) Operation: When this alarm signal is output, user can choose to
stop/continue operation, or not to detect this alarm.
"Operation continue" is default setting.
Threshold change: Temp. setting at which the alarm is generated can be
changed, and also, the alarm generating conditions can be set.
“45.0
o
C" is default setting.
AL04 Circulating fluid discharge temp. (Refer to Table 5.15-6 and Table 5.15-7) Operation: When this alarm signal is output, user can choose to
stop/continue operation, or not to detect this alarm.
"Operation continue" is default setting.
Threshold change: Temp. setting at which the alarm is generated can be
changed, and also, the alarm generating conditions can be set.
“1.0
o
C" is default setting.
AL08 Circulating fluid discharge pressure rise (Refer to Table 5.15-8 and Table
5.15-9) Operation: When this alarm signal is output, user can choose to
stop/continue operation, or not to detect this alarm.
"Operation continue" is default setting.
Threshold change: Pressure at which this alarm is generated can be
changed.
“0.30MPa" is default setting. (High pump head [Optional] -T:「0.70MPa」,
-MT0.60MPa)
AL09
Circulating fluid discharge pressure drop (Refer to Table 5.15-10 and Table
5.15-11) Operation: When this alarm signal is output, user can choose to
stop/continue operation, or not to detect this alarm.
"Operation continue" is default setting.
Threshold change: Pressure at which this alarm is generated can be
changed.
“0.05MPa" is default setting.
Page 87
HRX-OM-M090
Chapter 5 Display and setting of various functions
HRS Series 5.15 Alarm customize function
5-35
AL19 Communication error (Refer to Table 5.15-12 and Table 5.15-13) Operation: When this alarm signal is output, user can choose to
stop/continue operation, or not to detect this alarm.
“Not to detect “is default setting.
Threshold change: Time at which this alarm is generated can be
changed.
“30sec" is default setting.
AL31 Contact input signal 1 detection(Refer toTable 5.15-14) Operation: When this alarm signal is output, user can choose to
stop/continue operation, or not to detect this alarm.
"Operation continue" is default setting.
AL32 Contact input signal 2 detection (Refer to Table 5.15-15) Operation: When this alarm signal is output, user can choose to
stop/continue operation, or not to detect this alarm.
"Operation continue" is default setting.
AL21 DC fuse cutRefer to Table 5.15-16 Operation: When this alarm signal is output, user can choose to
stop/continue the operation.
"Operation stop" is default setting.
AL33
Water leakage(Refer to Table 5.15-17
Operation: When this alarm signal is output, user can choose to
stop/continue the operation.
"Operation stop" is default setting.
AL34 Electric resistivity riseRefer to Table 5.15-18Table 5.15-19 Operation: When this alarm signal is output, user can choose to
continue operation, or not to detect this alarm.
"Operation continue" is default setting.
Threshold change: Electric resistivity at which this alarm is generated
can be changed.
“4.5Mcm " is default setting.
AL35 Electric resistivity dropRefer to Table 5.15-20Table 5.15-21 Operation: When this alarm signal is output, user can choose to
continue operation, or not to detect this alarm.
“Not to detect “is default setting.
Threshold change: Electric resistivity at which this alarm is generated
can be changed.
“0.2Mcm" is default setting.
"Operation stop" is the default setting for alarm “AL01 Low tank level”. If the customer changes this setting to continue operation, fill the circulating fluid immediately after the alarm is generated.Operation without filling the circulating fluid leads to cause malfunction.
Page 88
HRX-OM-M090
Chapter 5 Display and setting of various functions
5.15 Alarm customize function HRS Series
5-36
5.15.2 Alarm customize function setting and checking
The table below explains the setting items of the alarm customize function and the initial values.
Table 5.15-1 List of set alarm customize function(1/2)
Display Item Contents
Initial value
(Default setting)
Changing of
low tank level
Set the operation when alarm No. AL01 "Low tank level" is generated.
A.STP
Changing of circulating fluid discharge temperature rise
Set the operation when alarm No. AL03 "circulating fluid discharge temperature rise" is generated.
A.RUN
Detection temperature for circulating fluid discharge temperature rise
Sets the detection temperature for the alarm of alarm NO. AL03 "circulating fluid discharge temperature rise ".
Alarm signal is generated when the temperature becomes higher than this temperature.
45.0 (113.0 F)
Changing of circulating fluid discharge temperature drop
Set the operation when alarm No. AL04" circulating fluid discharge temperature drop “is generated.
A.RUN
Detection
temperature for
circulating fluid discharge temperature drop
Sets the detection temperature for the alarm of alarm NO. AL04 "circulating fluid discharge temperature drop ". Alarm signal is generated when the temperature becomes lower than this temperature.
1.0 (33.8 F)
Changing of circulating fluid discharge pressure rise
Set the operation when alarm No. AL08 "circulating fluid discharge pressure rise " is generated.
A.RUN
Detection pressure for circulating fluid discharge pressure rise
Sets the detection pressure for the alarm of alarm NO. AL08 "circulating fluid discharge pressure rise ".
Alarm signal is generated when the pressure becomes higher than this pressure.
0.30MPa (44PSI)
Changing of circulating fluid discharge pressure drop
Set the operation when alarm No. AL09 "circulating fluid discharge pressure drop" is generated.
A.RUN
Detection
pressure for
circulating fluid discharge pressure drop
Sets the detection pressure for the alarm of alarm NO. AL09 "circulating fluid discharge pressure drop ".
Alarm signal is generated when the pressure becomes lower than this pressure.
0.05MPa (7PSI)
Changing of operation when
communication error
Set the operation when the alarm No. AL19 “Communication error” is generated.
OFF
The monitoring time of communication error
Set the alarm monitoring time when the alarm No. AL19 “Communication error” is generated. Alarm signal is generated when the monitoring time is exceeded.
30 sec
Changing of contact input signal 1 detection
Set the operation when the alarm No. AL31 “contact input signal 1 detection” is generated.
A.STP
Changing of contact input signal 2 detection
Set the operation when the alarm No. AL32 “contact input signal 2 detection” is generated.
A.STP
Changing of DC line fuse cut
Set the operation when the alarm No. AL21 “DC line fuse cut” is generated.
A.STP
Changing of water leakage
Set the operation when the alarm No. AL33 “water leakage” is generated.
A.STP
Changing of electric resistivity rise
Set the operation when the alarm No. AL34 “Electric resistivity rise” is generated.
A.RUN
Upper limit of electric resistivity rise
Sets the detection level for the alarm of alarm NO. AL34 "Electric resistivity rise".
Alarm signal is generated when the level becomes higher than this level.
4.5Mcm
Page 89
HRX-OM-M090
Chapter 5 Display and setting of various functions
HRS Series 5.15 Alarm customize function
5-37
Table 5.15-2 List of set alarm customize function(2/2)
Display Item Contents
Initial value
(Default setting)
Changing of electric resistivity drop
Set the operation when the alarm No. AL35 “Electric resistivity drop” is generated.
OFF
Lower limit of electric resistivity drop
Sets the detection level for the alarm of alarm NO. AL35 "Electric resistivity drop".
Alarm signal is generated when the level becomes lower than this level.
0.2Mcm
Temperature alarm Monitoring method
One alarm monitoring method can be selected from four methods for AL04 "Detection temp for the circulating fluid discharge temp. increase" and AL06 "Detection temp. for circulating fluid discharge temp. drop".
0
Monitoring start timer
Alarm will not be generated during the set period of time after starting operation. Alarm monitoring starts when it reaches the set time.
----
Range over Detection timer
After starting the alarm monitoring, the alarm will not be generated right away and will be kept not generated for the set period of time for AL04 "Detection temp for the circulating fluid discharge temp. increase" and AL06 "Detection temp. for circulating fluid discharge temp. drop", when the temperatures goes out of the set range.
5
*1This function is available for customers who have purchased the drain pan set (part no.: HRS-WL001). *2 This function is available for customers who have purchased the electric resistivity sensor set (part no.:
HRS-Dl001).
1. Press and hold the [MENU] key for approx. 2 sec.
Repeat pressing the key until the setting screen for alarm buzzer sound [
]
appears on the digital display.
PV
SV
PV
SV
PV
SV
PV
SV
Low tank level Setting and checking
2. Press the [SEL] key once.
The set screen of changing of low tank level is displayed on the digital display.
PV
SV
3. Select changing of low tank level from the table below with [] key or [] key, and
confirm by pressing “SEL”.
Table 5.15-3 List of set value
Set value Explanation
Initial value
(Default setting)
Operation continues when this alarm signal is generated. Operation is stopped when this alarm signal is generated.
[MENU]
Press and
hold
[MENU]
Press and
hold
[MENU]
Press and
hold
Page 90
HRX-OM-M090
Chapter 5 Display and setting of various functions
5.15 Alarm customize function HRS Series
5-38
Changing of circulating fluid discharge temperature rise Setting and checking
4. Press the [SEL] key once.
The set screen of changing of circulating fluid discharge temperature rise is displayed on the digital display.
PV
SV
5. Select changing of circulating fluid discharge temperature rise from the table below with
[] key or [] key, and confirm by pressing “SEL”.
Table 5.15-4 List of set value
Set value Explanation
Initial value
(Default setting)
This alarm signal is not detected. Operation continues when this alarm signal is
generated.
Operation is stopped when this alarm signal is generated.
Detection temperature for circulating fluid discharge temperature rise Setting and checking
6. Press the [SEL] key once.
The set screen of detection temperature for circulating fluid discharge temperature rise is displayed on the digital display.
PV
SV
7. Select detection temperature for circulating fluid discharge temperature rise from the
table below with [] key or [] key, and confirm by pressing “SEL”.
Table 5.15-5 List of set value
Set value Explanation
Initial value
(Default setting)
Setting/checking are not available if the setting of the circulating fluid discharge temperature rise is OFF.
Centigrade
to
Fahrenheit
to
Sets detection temp for the circulating fluid discharge temperature rise.
Temperature unit is Centigrade : Setting unit is
0.1
o
C
Temperature unit is Fahrenheit : Setting unit is
0.1
o
F
Page 91
HRX-OM-M090
Chapter 5 Display and setting of various functions
HRS Series 5.15 Alarm customize function
5-39
Changing of circulating fluid discharge temperature drop setting and checking
8. Press the [SEL] key once.
The set screen of changing of circulating fluid discharge temperature drop is displayed on the digital display.
PV
SV
9. Select changing of circulating fluid discharge temperature drop from the table below with
[] key or [] key, and confirm by pressing “SEL”.
Table 5.15-6 List of set value
Set value Explanation
Initial value
(Default setting)
This alarm signal is not detected. Operation continues when this alarm signal is
generated.
Operation is stopped when this alarm signal is generated.
Detection temperature for circulating fluid discharge temperature drop Setting and checking
10. Press the [SEL] key once.
The set screen of detection temperature for circulating fluid discharge temperature drop is displayed on the digital display.
PV
SV
11. Select detection temperature for circulating fluid discharge temperature drop from the
table below with [] key or [] key, and confirm by pressing “SEL”.
Table 5.15-7 List of set value
Set value Explanation
Initial value
(Default setting)
Setting/checking are not available if the setting of the circulating fluid discharge temperature drop is OFF.
Centigrade
to
Fahrenheit
to
Sets detection temp for the circulating fluid discharge temperature drop.
Temperature unit is Centigrade : Setting unit is
0.1
o
C
Temperature unit is Fahrenheit : Setting unit is
0.1
o
F
Page 92
HRX-OM-M090
Chapter 5 Display and setting of various functions
5.15 Alarm customize function HRS Series
5-40
Changing of circulating fluid discharge pressure rise Setting and checking
12. Press the [SEL] key once.
The set screen of changing of circulating fluid discharge pressure rise is displayed on the digital display.
PV
SV
13. Select changing of circulating fluid discharge pressure rise from the table below with [▲]
key or [] key, and confirm by pressing “SEL”.
Table 5.15-8 List of set value
Set value Explanation
Initial value
(Default setting)
This alarm signal is not detected. Operation continues when this alarm signal is
generated.
Operation is stopped when this alarm signal is generated.
Detection temperature for circulating fluid discharge pressure rise Setting and checking
14. Press the [SEL] key once.
The set screen of detection temperature for circulating fluid discharge pressure rise is displayed on the digital display.
PV
SV
15. Select detection temperature for circulating fluid discharge pressure rise from the table
below with [] key or [] key, and confirm by pressing “SEL”.
Table 5.15-9 List of set value
Set value Explanation
Initial value
(Default setting)
Setting/checking are not available if the setting of the circulating fluid discharge pressure rise is OFF.
MPa
to
PSI
to
Sets detection temp for the circulating fluid discharge pressure rise.
・Option -T Set range:0.05 to 0.70MPa(7 to 102PSI) Default setting:0.70MPa(102PSI) ・Option -MT Set range:0.05 to 0.60MPa(7 to 87PSI) Default setting:0.60MPa(87PSI)
Pressure unit is MPa : Setting unit is 0.01MPa Pressure unit is PSI : Setting unit is 1PSI
Page 93
HRX-OM-M090
Chapter 5 Display and setting of various functions
HRS Series 5.15 Alarm customize function
5-41
Changing of circulating fluid discharge pressure drop Setting and checking
16. Press the [SEL] key once.
The set screen of changing of circulating fluid discharge pressure drop is displayed on the digital display.
PV
SV
17. Select changing of circulating fluid discharge pressure dro from the table below with [▲]
key or [] key, and confirm by pressing “SEL”.
Table 5.15-10 List of set value
Set value Explanation
Initial value
(Default setting)
This alarm signal is not detected. Operation continues when this alarm signal is generated.
Operation is stopped when this alarm signal is generated.
Detection temperature for circulating fluid discharge pressure drop Setting and checking
18. Press the [SEL] key once.
The set screen of detection temperature for circulating fluid discharge pressure drop is displayed on the digital display.
PV
SV
19. Select detection temperature for circulating fluid discharge pressure drop from the table
below with [] key or [] key, and confirm by pressing “SEL”.
Table 5.15-11 List of set value
Set value Explanation
Initial value
(Default setting)
Setting/checking are not available if the setting of the circulating fluid discharge temperature drop is OFF.
MPa
to
PSI
to
Sets detection temperature for circulating fluid discharge pressure drop.
・Option -T Set range:0.05 to 0.70MPa(7 to 102PSI) Default setting:0.70MPa(102PSI) ・Option -MT Set range:0.05 to 0.60MPa(7 to 87PSI) Default setting:0.60MPa(87PSI)
Pressure unit is MPa : Setting unit is 0.01MPa Pressure unit is PSI : Setting unit is 1PSI
Page 94
HRX-OM-M090
Chapter 5 Display and setting of various functions
5.15 Alarm customize function HRS Series
5-42
Changing of operation when communication error Setting and checking
20. Press the [SEL] key once.
The set screen of changing of operation when communication error is displayed on the digital display.
PV
SV
21. Select changing of operation when communication error from the table below with [▲]
key or [] key, and confirm by pressing “SEL”.
Table 5.15-12 List of set value
Set value Explanation
Initial value
(Default setting)
This alarm signal is not detected.
Operation continues when this alarm signal is generated. Operation is stopped when this alarm signal is generated.
The monitoring time of communication error Setting and checking
22. Press the [SEL] key once.
The set screen of the monitoring time of communication error is displayed on the digital display.
PV
SV
23. Select the monitoring time of communication error from the table below with [] key or
[] key, and confirm by pressing “SEL”.
Table 5.15-13 List of set value
Set value Explanation
Initial value
(Default setting)
Setting/checking are not available if the setting of the communication error is OFF.
to
Sets communication error. Set unit is 1 sec. unit.
Changing of Contact input signal 1 detection Setting and checking
24. Press the [SEL] key once.
The set screen of the changing of contact input signal 1 detection is displayed on the digital display.
PV
SV
Page 95
HRX-OM-M090
Chapter 5 Display and setting of various functions
HRS Series 5.15 Alarm customize function
5-43
25. Select the changing of contact input signal 1 detection from the table below with [▲]
key or [] key, and confirm by pressing “SEL”.
Table 5.15-14 List of set value
Set value Explanation
Initial value
(Default setting)
This alarm signal is not detected. Operation continues when this alarm signal is
generated. Operation is stopped when this alarm signal is generated.
Changing of Contact input signal 2 detection Setting and checking
26. Press the [SEL] key once.
The set screen of the changing of contact input signal 2 detection is displayed on the digital display.
PV
SV
27. Select the changing of contact input signal 2 detection from the table below with [▲]
key or [] key, and confirm by pressing “SEL”.
Table 5.15-15 List of set value
Set value Explanation
Initial value
(Default setting)
This alarm signal is not detected. Operation continues when this alarm signal is
generated. Operation is stopped when this alarm signal is generated.
28. Press the [MENU] key once.
Return to the main screen (screen displaying the circulating fluid temperature).
PV
SV
Changing of DC line fuse cut Setting and checking
29. Press the [SEL] key once.
The set screen of the changing of DC line fuse cut is displayed on the digital display.
PV
SV
Page 96
HRX-OM-M090
Chapter 5 Display and setting of various functions
5.15 Alarm customize function HRS Series
5-44
30. Select the changing of DC line fuse cut from the table below with [] key or [] key,
and confirm by pressing “SEL”.
Table 5.15-16 List of set value
Set value Explanation
Initial value
(Default setting)
Operation continues when this alarm signal is generated. Operation is stopped when this alarm signal is generated.
Changing of water leakage Setting and checking
31. Press the [SEL] key once.
The set screen of the changing of water leakage is displayed on the digital display.
PV
SV
32. Select the changing of water leakage from the table below with [] key or [▼] key, and
confirm by pressing “SEL”.
Table 5.15-17 List of set value
Set value Explanation
Initial value
(Default setting)
Setting/checking are not available if the setting of the water leakage option is OFF. Operation continues when this alarm signal is generated. Operation is stopped when this alarm signal is generated.
Changing of electric resistivity rise Setting and checking
33. Press the [SEL] key once.
The set screen of the changing of electric resistivity rise is displayed on the digital display.
PV
SV
Page 97
HRX-OM-M090
Chapter 5 Display and setting of various functions
HRS Series 5.15 Alarm customize function
5-45
34. Select the changing of electric resistivity rise from the table below with [] key or [▼]
key, and confirm by pressing “SEL”.
Table 5.15-18 List of set value
Set value Explanation
Initial value
(Default setting)
Setting/checking are not available if the setting of the DI option is OFF. This alarm signal is not detected.
Operation continues when this alarm signal is generated.
Changing of upper limit of electric resistivity rise Setting and checking
35. Press the [SEL] key once.
The set screen of the changing of upper limit of electric resistivity rise is displayed on the digital display.
PV
SV
36. Select the changing of upper limit of electric resistivity rise from the table below with []
key or [] key, and confirm by pressing “SEL”.
Table 5.15-19 List of set value
Set value Explanation
Initial value
(Default setting)
Setting/checking are not available if the setting of the DI option is OFF and changing of Upper limit of electric resistivity rise is OFF.
to
Sets Upper limit of electric resistivity rise. Set unit is 0.1Mcm unit.
Changing of electric resistivity drop Setting and checking
37. Press the [SEL] key once.
The set screen of the changing of electric resistivity drop
is displayed on the digital
display.
PV
SV
Page 98
HRX-OM-M090
Chapter 5 Display and setting of various functions
5.15 Alarm customize function HRS Series
5-46
38. Select the changing of electric resistivity dropfrom the table below with [] key or [▼]
key, and confirm by pressing “SEL”.
Table 5.15-20 List of set value
Set value Explanation
Initial value
(Default setting)
Setting/checking are not available if the setting of the DI option is OFF. This alarm signal is not detected.
Operation continues when this alarm signal is generated.
Changing of lower limit of electric resistivity drop Setting and checking
39. Press the [SEL] key once.
The set screen of the changing of lower limit of electric resistivity drop is displayed on the digital display.
PV
SV
40. Select the changing of lower limit of electric resistivity drop from the table below with
[] key or [] key, and confirm by pressing “SEL”.
Table 5.15-21 List of set value
Set value Explanation
Initial value
(Default setting)
Setting/checking are not available if the setting of the DI option is OFF and changing of lower limit of electric resistivity drop is OFF.
to
Sets lower limit of electric resistivity drop. Set unit is 0.1Mcm unit.
How to monitor the temperature alarm Setting and Checking
41. Press the [SEL] key once.
Setting screen of the temperature alarm monitoring method is displayed on the digital display.
PV
SV
Page 99
HRX-OM-M090
Chapter 5 Display and setting of various functions
HRS Series 5.15 Alarm customize function
5-47
42. Select temperature alarm monitoring method from the table below with [] key or [▼]
key, and confirm by pressing “SEL”.
Table 5.15-22 List of set value
Set value Item Explanation
Initial value
(Default setting)
Continuous monitoring
Alarm monitoring starts at the same time as operation starts.
Automatic monitoring
When the circulating fluid temperature is outside of the alarm threshold range at the time of operation start, the alarm will not be generated until the temperature comes inside the alarm threshold range.
Monitoring start timer
Alarm will not be generated until it reaches the time set for the AS.22 "Monitoring start timer" after the operation starts. Alarm monitoring starts when it reaches the set time.
Automatic monitoring + Monitoring start timer
Alarm will not be generated until it reaches the time set for the AS.22 "Monitoring start timer" after the operation starts. Alarm monitoring starts when it reaches the set time. When the circulating fluid temperature enters the alarm threshold range before it reaches the set time, the alarm monitoring will be started at that time.
Settings of this function and example of alarm generating timing for 5.15.3 "Setting of temperature alarm
monitoring method and generation timing".
Monitoring start timer Setting and Checking
43. Press the [SEL] key once.
Setting screen of the monitoring start timer is displayed on the digital display.
PV
SV
44. Select monitoring start timer from the table below with [] key or [] key, and confirm
by pressing “SEL”.
Table 5.15-23 List of set value
Set value Explanation
Initial value
(Default setting)
Setting and checking are impossible when "0: Continuous monitoring" or "1: Automatic monitoring" is selected for the setting of AS21 "Temperature alarm monitoring method".
to
Sets the time when alarm monitoring starts Setting unit is 1 minute.
Settings of this function and example of alarm generating timing for 5.15.3 "Setting of
temperature alarm monitoring method and generation timing".
Page 100
HRX-OM-M090
Chapter 5 Display and setting of various functions
5.15 Alarm customize function HRS Series
5-48
Range over detection timer; Setting and Checking
45. Press the [SEL] key once.
Setting screen of the range over detection timer is displayed on the digital display.
PV
SV
46. Select range over detection timer from the table below with [] key or [] key, and
confirm by pressing “SEL”.
Table 5.15-24 List of set value
Set value Explanation
Initial value
(Default setting)
to
Sets time from detecting the alarm to generation of the alarm. Setting unit is 1 second.
Settings of this function and example of alarm generating timing for 5.15.3 "Setting of
temperature alarm monitoring method and generation timing".
47. Press the [MENU] key once.
Return to the main screen (screen displaying the circulating fluid temperature).
PV
SV
Loading...