SKC Vac-U-Go 228-9605, Vac-U-Go 228-9605B Operating Instructions Manual

Operating Instructions
863 Valley View Road, Eighty Four PA 15330 USA
skctech@skcinc.com
Sampling Pump
Catalog No. 228-9605 and 228-9605B
The Vac-U-Go non-compensating vacuum pump is a rotary vane-type pump
that provides ows up to 30 L/min and maintains sonic ow when used with the BioSampler. AC-powered, the Vac-U-Go Sampling Pump can be used for a variety of ambient and indoor air sampling applications including asbestos, bioaerosol, stack, VOC, and fenceline monitoring. It is ideal for sampling bioaerosols using the SKC BioSampler or the AGI-30 Impinger.
The line-operated Vac-U-Go Sampling pump has a permanently lubricated motor that is mounted in a housing with a vacuum pressure gauge to indicate sonic ow and a throling valve to adjust ow. Both models are supplied with an extension rod for mounting media and a power cord.
Performance Prole
Flow range: Non-compensated airow up to 30 L/min
Sonic ow with BioSampler (acts as a critical orice) is approximately 12.5 L/min
Motor: Oil-less rotary vane
Vane life: Approximately 6000 to 27,000 hrs depending on application
Power: Line power only
115 V, AC, 60 Hz (Cat. No. 228-9605) 230 V, AC, 50 Hz (Cat. No. 228-9605B)
Operating temperature range: 32 to 104 F (0 to 40 C)
Size: 5.5 x 10.5 x 6.5 ins (13.9 x 26.7 x 16.5 cm)
Weight: 12 lbs (5.4 kg)
Standards: None. Pump is not intrinsically safe
Not intended for use outdoors
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
15
30
25
5
20
10
Z
E
R
O
In HG VAC
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
15
30
25
5
20
10
Z
E
R
O
In HG VAC
Cautions
• Pump only clean, dry air.
• Protect pump from dirt and moisture.
• Pump is not intended for use outdoors.
• Protect surrounding items from pump exhaust air.
• Use only a three-wire grounded cable.
• Pump produces a high-pitched whine when operated.
• Donotpumpammableorexplosivegases.
• Do not operate pump in hazardous locations; not UL Listed for intrinsic safety
Operation
The Vac-U-Go Sampler can be operated in two dierent modes: non-compensating and BioSampler. Vac-U-Go is non-compensating, therefore, it is not suitable for compliance sampling in non-sonic mode. For indoor air studies, compliance, or other applications that require a quiet, compensating vacuum pump, use the QuickTake 30 or QuickTake 15 Sample Pump.
Non-compensating Mode
1. Use exible tubing to connect the outlet of a
representative sample medium to the pump inlet.
2. Connect a rotameter or calibrator to the
sample medium inlet.
3. Using the provided power cable, plug the pump into a wall outlet and press the pump power switch to the “on” position.
4. Use the valve adjust to set the desired ow
rate (see rotameter/calibrator operating instructions).
5. Remove rotameter or calibrator and replace representative sample medium with a new unexposed medium.
6. Sample for the desired sampling period.
7. Turn o pump and unplug from wall outlet.
8. Package sample appropriately for shipment to a laboratory.
Pump inlet
Valve adjust
to set ow
rate
BioSampler Mode
Pump inlet
1. Prepare BioSampler (see BioSampler Operating Instructions). Use exible
tubing to connect the BioSampler
outlet to the pump inlet.
Use tubing to connect BioSampler
2. Using the provided power cable, plug the pump into a wall outlet and press the pump power switch to the “on” position.
3. Use the valve adjust to open the valve fully. The pressure gauge
Valve adjust to open valve
should indicate > 15 inches Hg vacuum.
Note: The BioSampler features built-in critical orices. No other orice is needed to achieve sonic ow.