SKC ULTRA 690-200 Operating Instructions Manual

Page 1
Operating Instructions
863 Valley View Road, Eighty Four, PA 15330 USA
Tel: 724-941-9701 • www.skcinc.com
User-fi lled ULTRA Passive Sampler
Cat. No. 690-200
Clip
Diffusion holes
(sample sorbent
compartment)
Blank/
correction
sorbent
compartment
Sliding cover in open
(sampling) position
Sliding
cover
Clip
Sample
compartment
Blank/
correction
compartment
Sliding cover in closed position
Sample sorbent
compartment
Sliding
cover
correction
compartment
Clip
Plugs
Blank/
sorbent
Back view
Applications
SKC patented* user-fi lled ULTRA® Passive Sampler provides for low ppb to ppt detection of organic vapors in ambient and indoor air. Filling the sampler is simple and provides a wide variety of sorbent options and a longer shelf-life.
* U.S. Patent No. 6,607,581
Performance Profi le
Sampling Rates: Go to the SKC Passive Sampling Guide at
Analysis: Thermal desorption; gas chromatography
www.skcinc.com/catalog/passive-guide.php.
See Analysis Instructions at www.skcinc.com/instructions/40125.pdf.
Shelf-life: Sorbent vials containing Tenax
Carbopack X, or Anasorb
®
TA, Chromosorb® 106,
®
GCB1 have a 60-day shelf-life at
39.2 F (4 C).
Sorbent vials containing Charcoal have a one-year shelf-life
at 39.2 F (4 C).
www.skcinc.com
Page 2
Filling the Sampler
6
2
4a
1. Remove sampler from the resealable pouch.
2. With the back of the sampler facing upward and the metal clip oriented to the left, remove the plug from the sample sorbent compartment (on left side, immediately above diff usion holes).
Clip
3. Align the funnel (Cat. No. 690-301) over the compartment access port on the sampler and press down fi rmly.
Ensure funnel is securely installed on the access port.
If funnel does not sit properly, the sorbent will not fi ll the sampler.
4. Open one vial of selected sorbent.
a. Slowly pour small amounts of the sorbent
into the funnel. Note: If sorbent starts to back
up into the funnel, lift funnel and housing as one unit and gently tap against a hard surface until funnel clears.
Do not lift funnel until sorbent is completely transferred into housing.
Plug
Sample sorbent
compartment
Sliding cover
b. Repeat until all sorbent has been transferred from the vial into the hous-
ing compartment.
5. Once all sorbent has been transferred from the vial, remove the funnel, and replace the plug on the compartment access port.
6. If a blank/correction is required, remove the plug from the blank/correction compartment (on right side, immediately above SKC cover).
7. Repeat Step 3.
Sliding cover
Clip
Plug
Blank/
correction
compartment
8. Open second vial and repeat Steps 4a. and b.
9. Once the sampler has been fi lled, remove funnel, and replace plug on com- partment access port.
Page 3
1a
1b
2
3
4
Sampling
5
6
1. Clip sampler(s) in appropriate location.
a. Indoors: Clip sampler(s) to stand accessory
(Cat. No. 690-302) and locate at appropriate height.
b. Outdoors: Clip sampler(s) inside shelter
(open on two ends and bo om to allow air to fl ow freely). Hang shelter securely at the appropriate height.
Ensure diffusion holes in sampler are not obstructed in any way.
2. Start sampling: Slide sampler cover to open position (diff usion holes visible). Write sample start time on pouch label.
3. Stop sampling: Slide sampler cover to closed position (arrow and SKC visible). Write sample stop time on pouch label.
Important: Measure and record ambient temperature
and atmospheric pressure. Include in information sent to the laboratory.
4. a. Place sampler in its resealable pouch
immediately after sampling.
Or b. Remove plug from sample sorbent
compartment. Remove lid from sorbent vial and place opening of sorbent vial over the compartment access port. Transfer sorbent to vial by tipping the sampler upside down while holding the vial in place securely beneath the sampler. Gently tap the sampler to remove any remaining sorbent. Replace lid on sorbent vial and place vial in resealable pouch. Repeat for blank/correction sorbent.
Diffusion holes
(sample sorbent
compartment)
Blank/
correction
sorbent
compartment
Blank/
correction
compartment
Clip
Sliding cover
Sample compartment
Clip
Sliding cover
Ensure sample ID on sampler or vial matches sample ID on pouch.
5. Seal pouch(es). Place pouch(es) in container with
&
cool pack.
6. Use expedited shipping to send sample(s), pertinent information, and Chain of Custody form(s) to a laboratory for analysis.
For further information on analysis, see Analysis Instructions 40125.
Page 4
Storage
Before Sampling: Vials of Tenax TA, Chromosorb 106, Carbopack X, or
Vials of Charcoal: Up to one year at 39.2 F (4 C) in a
Anasorb GCB1: Up to 60 days at 39.2 F (4 C) in a clean, organic solvent-free environment
clean, organic solvent-free environment
After Sampling: 21 days at 39.2 F (4 C) in a clean, organic solvent-free
environment
Ordering
User-fi lled ULTRA Passive Sampler Cat. No. Empty Housing, packaged in reusable pouch, ea, requires sorbent
vials listed below; select sorbent appropriate for the application 690-200
Sorbent Vials, pk/2
Contain purged sorbent (< 25 ng typical background levels)
Anasorb GCB1, Chromosorb 106,#¥ 285 mg 690-203 Tenax TA,# 253 mg 690-204 Charcoal,#Ω 500 mg, for solvent extraction 690-205 Carbopack X,
Sampling Accessories Transfer Funnel, for ULTRA only 690-301 Stand for Indoor Sampling 690-302 Shelter for Outdoor Sampling 690-303 Analysis Accessory Thermal Desorption Tube, Perkin Elmer, 0.25-inch OD x 3.5-inch
length, includes screens and end caps P226530
# Limited-shelf life; see Storage above
‡ Comparable to Carbopack B
¥ Go to osha.gov and search on ULTRA for additional sampling rates.
Ω Can be stored for up to one year at 39.2 F (4 C)
#‡
370 mg 690-201
#
400 mg 690-206
SKC Limited Warranty and Return Policy
SKC products are subject to the SKC Limited Warranty and Return Policy, which provides SKC’s sole liability and the buyer’s exclusive remedy. To view the complete SKC Limited Warranty and Return Policy, go to h p://www.skcinc.com/warranty.
Form 40128 Rev 1807