SKC patented* user-fi lled ULTRA® Passive Sampler provides for low ppb to ppt
detection of organic vapors in ambient and indoor air. Filling the sampler is simple
and provides a wide variety of sorbent options and a longer shelf-life.
* U.S. Patent No. 6,607,581
Performance Profi le
Sampling Rates: Go to the SKC Passive Sampling Guide at
Analysis: Thermal desorption; gas chromatography
www.skcinc.com/catalog/passive-guide.php.
See Analysis Instructions at www.skcinc.com/instructions/40125.pdf.
Shelf-life: Sorbent vials containing Tenax
Carbopack X, or Anasorb
®
TA, Chromosorb® 106,
®
GCB1 have a 60-day shelf-life at
≤ 39.2 F (4 C).
Sorbent vials containing Charcoal have a one-year shelf-life
at ≤ 39.2 F (4 C).
www.skcinc.com
Page 2
Filling the Sampler
6
2
4a
1. Remove sampler from the resealable pouch.
2. With the back of the sampler facing upward and
the metal clip oriented to the left, remove the plug
from the sample sorbent compartment (on left
side, immediately above diff usion holes).
Clip
3. Align the funnel (Cat. No. 690-301) over the compartment access port on the
sampler and press down fi rmly.
Ensure funnel is securely installed on the access port.
If funnel does not sit properly, the sorbent will not fi ll the sampler.
4. Open one vial of selected sorbent.
a. Slowly pour small amounts of the sorbent
into the funnel. Note: If sorbent starts to back
up into the funnel, lift funnel and housing as one
unit and gently tap against a hard surface until
funnel clears.
Do not lift funnel until sorbent is completely
transferred into housing.
Plug
Sample sorbent
compartment
Sliding cover
b. Repeat until all sorbent has been transferred from the vial into the hous-
ing compartment.
5. Once all sorbent has been transferred from the vial, remove the funnel, and
replace the plug on the compartment access port.
6. If a blank/correction is required, remove the plug
from the blank/correction compartment (on right
side, immediately above SKC cover).
7. Repeat Step 3.
Sliding cover
Clip
Plug
Blank/
correction
compartment
8. Open second vial and repeat Steps 4a. and b.
9. Once the sampler has been fi lled, remove funnel, and replace plug on com-
partment access port.
Page 3
1a
1b
2
3
4
Sampling
5
6
1. Clip sampler(s) in appropriate location.
a. Indoors: Clip sampler(s) to stand accessory
(Cat. No. 690-302) and locate at appropriate
height.
b. Outdoors: Clip sampler(s) inside shelter
(open on two ends and bo om to allow air
to fl ow freely). Hang shelter securely at the
appropriate height.
Ensure diffusion holes in sampler are not
obstructed in any way.
2. Start sampling: Slide sampler cover to open
position (diff usion holes visible). Write sample
start time on pouch label.
3. Stop sampling: Slide sampler cover to closed
position (arrow and SKC visible). Write sample
stop time on pouch label.
Important: Measure and record ambient temperature
and atmospheric pressure. Include in information sent
to the laboratory.
4. a. Place sampler in its resealable pouch
immediately after sampling.
Or
b. Remove plug from sample sorbent
compartment. Remove lid from sorbent vial
and place opening of sorbent vial over the
compartment access port. Transfer sorbent
to vial by tipping the sampler upside down
while holding the vial in place securely beneath the sampler. Gently tap
the sampler to remove any remaining sorbent. Replace lid on sorbent
vial and place vial in resealable pouch. Repeat for blank/correction
sorbent.
Diffusion holes
(sample sorbent
compartment)
Blank/
correction
sorbent
compartment
Blank/
correction
compartment
Clip
Sliding cover
Sample compartment
Clip
Sliding cover
Ensure sample ID on sampler or vial matches sample ID on pouch.
5. Seal pouch(es). Place pouch(es) in container with
&
cool pack.
6. Use expedited shipping to send sample(s),
pertinent information, and Chain of Custody
form(s) to a laboratory for analysis.
For further information on analysis, see Analysis Instructions 40125.
Page 4
Storage
Before Sampling: • Vials of Tenax TA, Chromosorb 106, Carbopack X, or
• Vials of Charcoal: Up to one year at ≤ 39.2 F (4 C) in a
Anasorb GCB1: Up to 60 days at ≤ 39.2 F (4 C) in a clean,
organic solvent-free environment
clean, organic solvent-free environment
After Sampling: 21 days at ≤ 39.2 F (4 C) in a clean, organic solvent-free
Sampling Accessories
Transfer Funnel, for ULTRA only690-301
Stand for Indoor Sampling690-302
Shelter for Outdoor Sampling690-303
Analysis Accessory
Thermal Desorption Tube, Perkin Elmer, 0.25-inch OD x 3.5-inch
length, includes screens and end capsP226530
# Limited-shelf life; see Storage above
‡ Comparable to Carbopack B
¥ Go to osha.gov and search on ULTRA for additional sampling rates.
Ω Can be stored for up to one year at ≤ 39.2 F (4 C)
#‡
370 mg690-201
#
400 mg690-206
SKC Limited Warranty and Return Policy
SKC products are subject to the SKC Limited Warranty and Return Policy, which
provides SKC’s sole liability and the buyer’s exclusive remedy. To view the complete
SKC Limited Warranty and Return Policy, go to h p://www.skcinc.com/warranty.
Form 40128 Rev 1807
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.